Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth...

12
Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth Erickson, X. Allen Li Medical College of Wisconsin

Transcript of Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth...

Page 1: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Entesar Dalah, CungenYang, Ion Moraru,

Eric Paulson, Beth Erickson, X. Allen Li

Medical College of Wisconsin

Page 2: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Aim Correct for partial volume effect (PVE) on PET

images for patients with pancreatic cancer.

Methods A voxel-by-voxel PVE correction using a

recovery coefficient (RC) approach, the use of 4

different PET thresholding methods and manual

delineation on 2 MRI methods.

Results SUVmax values before and after PVE

correction, diameter and volumetric analysis.

Summary

Contents

Page 3: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

What is partial volume effect (PVE)?

The term PVE represents differences in intensity values in

the image, which effects images quantitatively and

qualitatively, and is referred to two distinct phenomenon :

� Size dependent 3D blurring effect (i.e., predominant in

small lesions that are < 3 times the FWHM of PET

scanner).

� This effect is pronounced with

PET due to the poor spatial

resolution

� Tissue sampling effect

� Affects also CT & MRI images

Soret et al., J Nucl Med, 6, 932-945 (2007)

Page 4: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

For the RC approach : Three phantom set-up were made using the following spheres to background ratios i.e., 8:1, 6:1, 4:1.

Now that RC is known, the voxel-by-voxel

PVE corrected activity is given by :

Srinivas et al., Ann Nucl Med, 23, 341-348 (2009)

RC is defined using the following formula :

Page 5: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Continue – NEMA phantom

Profile analysis (PA) study : to define a cut-off value

(COV) to threshold PET images. A vertical and a horizontal profile were

drawn through the centre of each sphere, followed by profile analysis and

mathematical interpolation along the x-axis at the FWHM value to determine sphere

diameter in mm. We used a spheres to background ratio of 4:1

The COV is given by following formula:

Note : The COV is point spread function (PSF )

Dependent and size dependent. Dalah et al., ISBN 978-3-659-26628-7, pp 1-177 (2012)

Page 6: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

For sphere to background ratio of 4:1

0.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

0.80

0.90

1 1.5 2 2.5 3 3.5

RC

Spheres diameter (cm)

Sphere

(D) mm

RC COV%

10 0.15 81

13 0.26 62

17 0.47 48

22 0.68 37

28 0.83 28

37 0.85 22

Page 7: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

After PVE correction

SUVmax = 6.56 g/ml

Tumor diameter

on CT

SUVmax before PVE

mean & range

SUVmax after PVE

mean & range

RC

mean & range

(1.9 – 5.8) cm 8.2 (3.9 – 13.1) 10.6 (6.56 – 16.37) 0.72 (0.58 – 0.81)

SUVmax = 3.93 g/ml

Before PVE correction Define background in patients

� A duplicated ROI of that

used on the tumor was

drawn in a region inside the

liver. Blake et al., Appl Radiol, 34: 8-18 (2010)

� The mean and range of

SUVs in the liver were 1.63

(1.2 – 2.2), 9 patients.

Page 8: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Pancreatic tumor on different modalities

DWI-MRI (b=1000 s/mm2)

PETDCE-MRI

ADC-DWI-MRI

Diameter CT with

contrast

ADC-DW DCE PET SUV2.5

before corr.

PET SUV2.5

after corr.

cm 1.9 2.31 1.58 1.61 1.95

Page 9: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Findings : Tumor contours on PET using

different methods

Before PVE correction After PVE correction

Thresholding

methods

Threshold value Volume cm3

Before corr. After corr. Before corr. After corr.

SUV2.5 2.5 2.5 0.33 0.64

42%SUVmax 1.65 2.75 2.29 0.79

COV% 1.71 2.86 1.93 0.73

S/B 2.44 2.44 13.78 14.00

maximum diameter on CT = 1.9 cm

Page 10: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Findings : tumor volume on different modalities &

different PET thresholding methods

General observations� SUV2.5 & S/B ratio yield

larger volumes after

PVE correction.

� 42%SUVmax & %COV

yield smaller volumes

after PVE correction.

� In all modalities, S/B

ratio demonstrated the

highest tumor volume

except for CT.

� ADC-MRI yield volumes

that sits between PET &

CT.0

10

20

30

40

50

60Average volume ± SD

0

10

20

30

40

50

60

1 2 3 4

Vo

lum

e (

cc)

Patients

SUV2.5 (B)

SUV2.5(A)

42%SUVm(B)

42%SUVm(A)

S/B(B)

S/B(A)

COV%(B)

COV%(A)

DCE

ADC

CT 50%

Page 11: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Findings : maximum diameter on different

modalities & different PET thresholding methods

0

1

2

3

4

5

6

7

8

9

Dia

me

ter

(cm

)

Patients

SUV2.5 (B)

SUV2.5 (A)

S/B (B)

S/B (A)

42%SUVm (B)

42%SUVm (A)

COV% (B)

COV% (A)

General observation� No trend was seen on

maximum diameter

measurements

before & after PVE

correction using the 4

PET thresholding

methods

0

1

2

3

4

5

6

7

CT DCE ADC SUV2.5 S/B 42%SUVm COV%

Average diameter ± SDAxial plane profile

Page 12: Entesar Dalah, CungenYang, Ion Moraru, Eric Paulson, Beth ...chapter.aapm.org/nccaapm/z_meetings/2013-10-10_and_10-11/fall2013/Dalah.pdfOct 10, 2013  · Entesar Dalah, CungenYang,

Summary

• PVE yield a considerable impact on 3D tumor

volume measured on PET images.

• PVE corrected PET images reveals larger

volumes except for 42%SUVmax and %COV

methods.

• Corrected PET images reveal a reduced

volumetric difference between CT & ADC-DW

MRI.

• Further studies combined with pathological

specimens are required to identify the optimal

PET thresholding method.