Introduction

The differential amplifier can measure as well as amplify small signals that are buried in much larger signals.

Page 3: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

There are two input terminals, labeled () input, and (+) input.

Page 4: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Superposition

If E1 is replaced by a short circuit, E2 sees an inverting amplifier with a gain of m.

• Therefore, the output voltage due to E2 is mE2.

mE2

Page 5: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Now let E2 be short-circuited:

• E1 divides between R and mR to apply a voltage of E1m/ (1+ m) at the op amp’s (+) input.

Page 6: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

This divided voltage sees a noninverting amplifier with a gain of (m + 1).

• The output voltage due to E1 is the divided voltage:

• E1m/(1 + m) times the noninverting amplifier gain, (1 + m), which yields mE1.

mE1

Page 7: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Therefore, E1 is amplified at the output by the multiplier m to mE1.

• When both E1 and E2 are present at the (+) and () inputs, respectively.

• Vo is mE1 mE2.

Page 8: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

The output voltage of the differential amplifier, Vo, is proportional to the difference in voltage applied to the (+) and () inputs.

Multiplier m is called the differential gain and is set by the resistor ratios.

Page 9: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

When E1 = E2 the output voltage is 0.

• To put it another way, when a common (same) voltage is applied to the input terminals, Vo = 0.

Page 10: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Lab 6_Differential Amplifier

The gain of the amplifier below can be determined using the Superposition Principle.

'

'

+

2.2 k

4.7 k

22 k

RS

Ri

Rf

VOUT

RD

Page 11: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Inverting Amplifier

Forcing V2 to 0 develops an inverting amplifier with an output, VOUT of:

'

'

+

2.2 k

4.7 k

22 k

RS

Ri

Rf

VOUT

RD

V1

1111 1022

22V

k

kV

R

RVV

i

fOUT

.

i

fOUT R

RVV 11

Page 12: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Non-inverting Amplifier

Forcing V1 to 0 develops a non-inverting amplifier.

'

'

+

2.2 k

4.7 k

22 k

RS

Ri

Rf

VOUT

RD

V2

Page 13: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Applying Thevenin’s Theorem:

'

'

+

2.2 k

4.7 k

22 k

RS

Ri

Rf

VOUT

RD

V2

22 VRR

RVV

DS

Dopen

DS

DSTH RR

RRR

Page 14: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

The output of the non-inverting amplifier is:

'

'

+

2.2 k

22 k

RTh

Ri

Rf

VOUT

DS

D

RR

RV

2

i

fOUT R

RVV 122

i

fOUT R

RVV 122

Page 15: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

The total output is the sum:

To balance the circuit, we set the coefficients to add to zero.

i

f

i

fOUTOUT R

RV

R

RVVV 1212

i

f

i

f

R

R

R

R

Page 16: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

fi

f

fi

i

i

f

i

f

i

fi

i

f

i

f

i

i

i

f

i

f

RR

R

RR

R

R

R

R

R

R

RR

R

R

R

R

R

R

R

R

R

R

Page 17: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

1212

12

1

2

2

12

VVR

RV

RR

RRV

R

R

VRR

R

RR

RV

R

R

R

RV

RRR

R

RRR

RRV

R

RV

R

R

RR

R

RR

RVV

i

f

fi

fi

i

f

fi

f

fi

i

i

f

i

f

fii

f

fii

fi

i

f

i

f

fi

f

fi

fOUT

Page 18: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

So the balanced condition yields

• and the differential gain Ad is

12 VVR

RV

i

fOUT

i

fD R

RA

Page 19: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Common-mode rejection of 60 cycle power line interference in medical instrumentation which measures difference potentials on the body is a fundamental problem. • Power-line interference may exceed the level

of the signal being measured. • This bad news is often cancelled by the fact that

the interfacing signal appears equally intense at both input terminals of the diff amp, and is therefore called a common-mode signal.

Page 20: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

If the diff amp is not perfectly balanced, as is always the case in the real world, then the common-mode signal input will cause an output signal that then constitutes interference with the desired amplified signal. • Since one of the functions of the diff amp is to reject

the common-mode signal, we define a figure of merit, the common-mode rejection ratio (CMRR), which measures how well the rejection occurs.

Page 21: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

The common-mode rejection ratio CMRR is defined as the magnitude of the ratio of the differential voltage gain Ad to the common-mode voltage gain Ac.

Page 22: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Ad equals VOUT divided by V1 when node 2 is grounded, and V1 is applied to node 1.

• Also, AC equals VOUT divided by V1 when node 1 is connected to node 2, and V1 is applied again.

V Vwhen

grounded is Vwhen

12

2

OUT

OUT

V

VCMRR

Page 23: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

In practice the CMRR is measured in the following steps:

1.Ground V2, and apply a voltage V1 to the upper terminal.

2.Measure the resulting VOUT.

3.Lift V2 from ground and short the two input leads, then apply the same value of V1.

4.Measure the resulting VOUT.5.To compute CMRR, divide the results of step 2 by the

result of step 4, and take the magnitude.

Page 24: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

The CMRR is a voltage ratio, and therefore in decibel units we may define CMRRdb as

CMRRCMRRdb log20

Page 25: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Common-Mode Voltage

The simplest way to apply equal voltages is to wire inputs together and connect them to the voltage source.

Page 26: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

For such a connection, the input voltage is called the common-mode input voltage, ECM.

Page 27: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Now Vo will be 0 if the resistor ratios are equal (mR to R for the inverting amplifier gain equals mR to R of the voltage-divider network.)

Page 28: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

Practically, the resistor ratios are equalized by installing a potentiometer in series with one resistor.

Page 29: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

The potentiometer is trimmed until Vo is reduced to a negligible value. • This causes the common-mode voltage gain,

Vo/ECM to approach 0.

• It is this characteristic of a differential amplifier that allows a small signal voltage to be picked out of a larger noise voltage.

Page 30: ENTC 4350 BIOMEDICAL INSTRUMENTATION I BASIC DIFFERENTIAL AMPLIFIER.

It may be possible to arrange the circuit so that the larger undesired signal is the common-mode input voltage and the small signal is the differential input voltage. • Then the differential amplifier’s output voltage

will contain only an amplified version of the differential input voltage.