Electrosurgery
• 1926- Cushing and Bovie• Alternating radiofrequency current.• 500,000 to 2 million Hz.• Electrocautery is not same as
electrosurgery.
Physics of electrosurgery
• Current takes path of least resistance.
• Tissue containing maximal water has least resistance.
• Energy density – amperes/cm2
Ohms Law
• I= V/R• Current = I : flow of electrons/time
(amps)• Resistance (Impedance) = R :
Obstacle to the flow of current (ohms)
• Voltage = V: force pushing current through the resistance (volts)
• Power = V x I (watts)
Monopolar
• ADVANTAGES– Ease of use– Cutting and coagulation currents– Dissecting capability.
• DISADVANTAGES– Larger volume of tissue injured.– Interferes with electrical implants.– Requires ground pad.
Bipolar
• ADVANTAGES.– Small volume of tissue injured.– Less risk of burns.– Safe with electrical implants.– Effective in wet fields.
• DISADVANTAGES–More skill required.– Coagulation only.– No dissecting capability
Cutting
• Non contact activity.• Non modulated sine wave almost
100 percent of time – best for cutting.
Fulgration and coagulation
• Non contact.• Short bursts of high voltage with
pauses. • Cause coagulative necrosis of tissues
and fluid.
Dessication
• Tissue is heated and water in the cell dries out.
• Can use cutting or coagulation current by contact with tissue.
Complications of electrosurgery
• Direct coupling – Active electrode in contact with metal instrument.
• Capacitative coupling – current is transferred from the active electrode to nearby tissue through the intact insulation
• Insulation failure.
New generation
• Higher current, less voltage.• Less thermal spread.• Impedence feedback, end point
feedback.• Grasp, dissect, coagulate and
transect.• More efficient but costly.
Newer energy sources
• Advanced Bipolar devices.– Gyrus ACMI (Olympus)– Enseal (Ethicon Endo-Surgery)– Ligasure (covidien)
vessel sealing
• Ultrasonic scalpel– Harmonic scalpel ( Ethicon Endo-
Surgery)
Gyrus ACMI
• Plasma kinetic energy delivered in pulses
• Intermittent tissue cooling prevents thermal damage and sticking.
• 2 modes– Vapor pulse coagulator– PK cutting
• Audible impedence and end point indicators.
• PK energy to tissue forms vapor zonescurrent passes around the vapor zone vapor zone collapses causing tissue sealing
Enseal
• Adjusts energy according to tissue impedance.
• Smart electrode technology – millions of nano conductive particles each with discrete thermostatic switch.
• Each particle interrupts current flow to a specific tissue when temperature rises.
• If temperature below fusion level particle turns back on.
• Less heat required. Tissue volume reduced by compression.
• Vessel walls fused by compression, protien denaturation and then renaturation.
• Seals upto 7 mm vessels. Thermal spread 1 mm
• Withstands upto 7 times systolic pressures.
Ligasure
• “Electrosurgical collagen welding” combination of pressure and energy denature collagen and elastin permanent seal.
• Hydrothermal rupture of collagen cross links
• 60-90 degree C.
• Cooling allows renaturation of entangled unwound collagen strands.
• High uniform mechanical compression increases entanglement/ recross linking upon thermal relaxation.
• Permanently fuses tissue and vessels without dessication.
• Average seal cycle is 2-4 sec.• Seal can withstand upto 3 times
normal systolic pressure.• Can ligate vessels upto 7 mm.• Impedence feedback – adjust energy
levels according to tissue impedence.• Feedback controlled response system
to detect end of seal cycle.
Harmonic Scalpel
• Converts ultrasonic energy→mechanical energy.• Piezoelectric crystal produces
~55000 vibrations per second.• Lysis of H-bond → heat →
denaturation of protien → separation of tissue.
• Tissue separation at 50-60 degree C.
Settings on a harmonic scalpelHigher the setting → longer excursion
of the blade --> more mechanical energy more cutting speed/more thermal energy
Advantages of harmonic scalpel
• Minimal thermal spread.
• No dessication or charring of tissue.
• No risk of electrical injury
• No smoke.
Clipless lap cholecystectomy uses a harmonic scalpel.
Disadvantages of harmonic scalpel
• inability to coagulate vessels greater than 5mm.
• Operator dependence of settings of the blade according to tissue.
• Increased cost.
CUSA
• Cavitron ultrasound surgical aspirator.
• Causes cavitational fragmentation of cells.
• Ultrasonic generator at 23000 Hz.• With coupled powerful aspiration
device.• Used in liver surgery and
neurosurgery• No coagulation or cutting.• Only tissue gets fragmented.
• Harmonic scalpel has the least thermal spread.
• Ligasure is the highest rated.
J Endourol. 2008 Oct;22(10):2307-12
Sealing time (s)
Burst pressure (hg mm)
Smoke/visibility (ppm)
Harmonic scalpel
14.3 204 2.88 ppm
Ligasure V 10 385 12.5 ppm
Gyrus PK 11.1 290 74.1 ppm
Enseal 19.2 255 21.6 ppm
Argon beam coagulator
• Inert, non combustible, easily ionisable.
• Used along with monopolar electrosurgery.
• Electric current ionises argon facilitates current flow.
• Not suitable for laparoscopy.
LASER
• Precise application of energy.• Less lateral damage and stray
currents.• Energy induces molecular vibration
and thermal vibrations.• Mediums – CO2, Argon, NdYAG.
Microwave ablation
• Generator probe at 2450 MHz.• Created alternating electrical field
induces motion of polar molecules kinectic energy converted into heat coagulation necrosis.
• Applications – liver, prostatic hyperplasia, endometrial bleeding, partial nephrectomy.
• Only small area ablated.
Radiosurgery
• Gamma knife used in neurosurgery.• High energy gamma radiation
arranged in circular fashion.• Focussed stereotactically