Electromyography. EMG Measures a muscle’s electric potential – Surface EMG – Intramuscular EMG.
EMG Physiology
-
Upload
shauki-ali -
Category
Documents
-
view
218 -
download
0
Transcript of EMG Physiology
-
7/27/2019 EMG Physiology
1/16
4
CHAPTER 2
PHYSIOLOGY OF EMG
Electromyography (EMG), also referred to as myoelectric activity, measures the
electrical impulses of muscles at rest and during contraction. As with other
electrophysiological signals, an EMG signal is small and needs to be amplified with an
amplifier that is specifically designed to measure physiological signals. This signal can be
recorded or measured with an electrode, and is then displayed on an oscilloscope, which
would then provide information about the ability of the muscle to respond to nerve stimuli
based upon the presence, size and shape of the wave the resulting action potential. While
the electrode could be inserted invasively into the muscle (needle electrodes), a skin surface
electrode is often the preferred instrument, because it is placed directly on the skin surface
above the muscle without employing the method of pinch insertion into the test subject.
When EMG is measured from electrodes, the electrical signal is composed of all the action
potentials occurring in the muscles underlying the electrode. This signal could either be of
positive or negative voltage since it is generated before muscle force is produced and occurs
at random intervals.The EMG signal is first picked up by electrode and amplified. Frequently more than
one amplification stages are needed, since before the signal could be displayed or recorded, it
must be processed to eliminate low or high frequency noise, or any other factors that may
affect the outcome of the data. The point of interest of the signal is the amplitude, which can
range between 0 to 10 millivolts (peak-to-peak) or 0 to 1.5 millivolts (rms). The frequency of
an EMG signal is between 0 to 500 Hz. However, the usable energy of EMG signal is
dominant between 50-150 Hz [7]. The extended study of EMG signal characteristics can be
found at [6] and [8].
In order to obtain a signal that yields the maximum information, the method
employed and the implementation device has to be considered. There are many dependent
factors that could affect a surface EMG since the signal is susceptible to noise interference
such as hum, signal acquisition such as clipping and baseline drift, skin artifacts, processing
-
7/27/2019 EMG Physiology
2/16
5
errors, and interpretation problems. For example, the contact of electrode to the skin could
distort a recording signal. The inadequate amplification of the signal could cause a recorder
detection problem. A wrong filter could efface some of desirable information of a signal.
Moreover, there are other factors such as the distance between electrodes as well as the
recording times used in the experiment. The device utilized in the measuring of the signal
must also be considered since low-level input into a recording device could also affect data
and yield inaccurate results.
2.1EXCITABLE TISSUE AND ACTION POTENTIAL
There are two main types of tissue in the nervous system: excitable tissue and non-
excitable tissue. The excitable tissue, which is composed of neurons, responds to and
transmits nerve stimuli. The non-excitable tissue, composed of glial cells, does not responseto voltage or any other conventional stimulus, since glial cells are non-conducting, and
function only as support cells in the nervous system.
Excitable tissue can be divided into four components: sensory receptors, neuron cell
bodies, axons, and muscle fibers [45]. In a situation involving a harmful stimulus such as
contact with a sharp pebble or a hot surface, the resulting pain and pressure are transmitted
by sensory receptors. The pain is by a receptor potential, which is the transmembrane
potential difference of a sensory cell. Produced by sensory transduction, a receptor potential
results from inward current flow, which will bring the membrane potential of the sensory
receptor toward the threshold to trigger the neuron into generating a rapid burst of voltage
pulses called the action potential(APs).
As shown in Figure 2.1, triggered by a constant high-pressured stimulus, the sensory
receptor generates an initially high receptor potential that rapidly decreases to a much lower,
steady level. This decrease in the receptor potential is called adaptation. The action potential
produced by the neuron has a magnitude of 0.1 volts [45], which is a value that is shared by
every animal, from a squid to a human.
To step off that pebble, the neuron sends a message along a nerve axon to the base of
the spinal cord. The axon, or nerve fiber, is the slender projection of a neuron that conducts
electrical impulses away from the soma, or the nerve cell body. There are two types of axons:
the afferentaxon and efferentaxon. The afferent axon, or sensory axon, leads to the central
-
7/27/2019 EMG Physiology
3/16
6
Figure 2.1: Typical time variations associated with a sudden, steady stimulus.
nervous system, and carries messages from sensory receptors at the peripheral endings to the
spinal cord or brain. The efferent axon, or motor axon, originates at the spinal cord and
carries information through the body parts, synapse with muscle fibers to stimulate muscular
contraction as well as the muscle spindles to alter proprioceptive sensitivity, which is a key
factor in muscle memory and hand-eye coordination. Because these two types of axons are
designed to relay high-speed messages, their diameter is between 0.001 and 0.022
millimeters, which is longer than ordinary axons, which have a diameter between 0.0003 and
0.0013 millimeters [45]. When compared with the ordinary axons, the efferent and afferent
axons also have a thicker layer of myelin, an electrically insulating fatty layer that increases
the speed of impulses by means of saltatory conduction. Therefore, by inhibiting charge
leakage, myelinated axons propagate action potentials that recur at successive nodes rather
than waves, and thus hop along the axon, thereby increasing the speed of the impulse. With
a large diameter and thick of myelin sheaths, all signals can thus travel through the afferent
and efferent axons at speeds as high as 120 meters per second, or 270 miles per hour [45]. On
the other hand, the ordinary axons, which are solely responsible for simple activities such as
-
7/27/2019 EMG Physiology
4/16
7
reporting pain and temperature changes, have small diameters and unmyelinated fibers,
which are adequate to carry slow-speed signals.
As shown in Figure 2.2, the afferent axon carries the action potential burst from the
neuron to the interneuron, a neuron that communicates only to other neurons, or to the motor
neuron. This causes a chemical transmitter to be released across a narrow fluid gap called
synapses. The latter are specialized junctions that allow neurons to signal to their target cell,
which could be another neuron or a non-neuronal cell such as a muscle or gland. The action
potential crosses this junction to either another interneuron or a motoneuron, triggering
another action potential burst as the process repeats until the message reaches the efferent
axon, which then carries the action signal back down to the leg muscle. Once the signal
reaches the muscle tissue, the message instructs the muscle to contract, resulting in lifting the
foot off the pebble.
Figure 2.2: Excitable tissue called into play when a person steps on a sharp pebble.
-
7/27/2019 EMG Physiology
5/16
8
2.2GENERATION OF ACTION POTENTIAL
As stated earlier, after a sensory receptor generates information, this electric signal is
transmitted to its intended target by traveling through an axon. However, an axon is a
relatively poor conductor because it rapidly attenuates the electrical signal. The potential can
decrease to 37 % of its original value after traveling a distance of only 0.15 millimeters along
an axon, resulting in an unusable potential value [46]. This distance in which the potential
becomes unusable is called the length constant. The length constant is dependent upon the
size of the axon, as it is proportional to the square root of an axon diameter.
To overcome this tendency of signal attenuation, the nervous system uses a method to
increase the strength of the electric signal. When the potential decreases to a threshold level,
such as eight millivolts, the neuron will fire another 100 millivolts action potential [46].
However, the action potential will keep decreasing after travel through the axon, which in
effect will stimulate the neuron to fire one burst of action potential after action potential, a
process that is referred to asfrequency modulation. For example, in order to make a potential
increase to 10 millivolts, the neuron might fire ten times per second, although the neuron is
also able to extinguish voltage in order to end action potential.
To get more insight in this process, one must understand the structure of the axon.
There are ions arranged in constant random thermal motion inside an axon, with protein
molecules being one of the main components of the axon membrane. Under normalconditions, sodium (Na
+) and calcium (Ca
2+) are more concentrated in the extracellular fluid,
while potassium (K+) is more concentrated within the cell. In effect, K
+is the key
determinant of the resting membrane potential, since the resting cell membrane is more
permeable to K+
than to the Ca2+
and Na+
molecules [14]. However, while it plays a small
part in the resting membrane potential, Na+
is a key player in the generation of electric
signals. When a cell goes from a resting to an excited state, orfiring level, the cell increases
its Na+
permeability. This causes Na+
molecules to enter the cell through voltage-gated
channels, thus moving down its chemical gradient. This addition of the positive charge of
Na+
to the intracellular fluid causes the cell to become depolarizedand initiates an action
potential. (M.S. Gordon, 1972). The extinguishinglevel that marks the falling phase of the
action potential is the result of an increase in K+
permeability in the cell. However, the
closing and opening of the voltage-gated channels is regulated by the jostling of the atoms
-
7/27/2019 EMG Physiology
6/16
9
within the cell, which results in randomness in the train of the generated action potentials.
Consequently, any undesired departure from a perfectly ordered system may give rise to so-
called noise. Higher receptor potential will initiate less noisy action potentials. However, a
noisy system is not always bad, as it enables living things to be able to adjust themselves to
changing environment [46].
The action potential is not only in the shape of narrow spike. Alan L. Hodgkin and
Andrew F. Huxley (1952) also suggest another model of action potential as illustrated at the
top of Figure 3-6 [46]. They applied a +20 millivolts trigger at zero time to the giant axon of
the squid, and found that during an action potential, an ion moves to the axon membrane by
using its protein molecules to create bridge to the membrane. The protein molecules are
unique for each ion species, with the Na+
ions trying to diffuse into the axon while the K+
ions are trying to diffuse out of the axon membrane.
Figure 2.3: Action potential model described by Alan L. Hodgkin and Andrew F.
Huxley.
-
7/27/2019 EMG Physiology
7/16
10
From Figure 2.3, the sodium in curve is negative because the current that flows into
the axonplasm is defined as negative. On the other hand, the current that flows out of the
axonplasm is defined as positive. The sodium carrier proteins convey Na+
ions into the axon
in accordance with the sodium in curve. When crossing the membrane, there is only low
voltage left to drive the sodium ions into the bridges of the transport proteins. Consequently,
the dip of the sodium curve exists at the peak of action potential curve. On the opposite side
of the sodium curve, the potassium carrier proteins convey K+
ions out of the axon in
accordance with the potassium out curve. To the left of the sodium dip, the Na+
current in
is much greater than the K+
current. As a result, the voltage rapidly rises to 100 millivolts
above the resting potential. To the right of the dip, the potassium ions are small excess to the
sodium ions, which marks the slow drop in voltage.
2.3PROPAGATION OF ACTION POTENTIAL
In this section, the focus will be on how unmyelinated and myelinated axons generate
their action potentials. Regenerating nerve fibers could either be unmyelinate or myelinated
[47]. As stated earlier, unmyelinated fibers have thin membranes to carry slow-speed signals.
On the other hand, myelinated axons have thick membrane allowing them to carry high-
speed signals.
In unmyelinated fibers, the AP propagates in the form of an ocean wave. When the
voltage across the membrane rises above the threshold level of eight millivolts, a
regenerating axon starts to generate an action potential. The thin membrane of the
unmyelinated axon allows ions to easily move across the membrane. Figure 2.4 shows the
AP waveform initiated by an unmyelinated axon.
Please note that the net ion current of Figure 2.4 is based on the model of the impulse
of an RC cable, which is a simpler and less accurate model than the Hodgkin and Huxley
model. Therefore, the net ion current curve in this section is different from that in section 2.2.
Figure 2.4 (a) depicts the net ion current that generates the action potential. The
voltage rises to its peak at 100 millivolts before decreasing in value. Figure 2.4 (b) shows the
curve of voltage and current versus distance. It is interesting to see the distance and time
curve are mirror image of each other. While generally the distance curve can differ from the
-
7/27/2019 EMG Physiology
8/16
11
time curve, the myelinated axon can generate the distance curve and time curve of the action
potential that has the same shape and amplitude [47].
Figure 2.4: Net ion current curve and action potential curve of an unmyelinated axon.
In comparison to unmyelinated axons, myelinated axons have a thicker wall. This
makes it impossible for sodium and potassium ions to move across the axon membrane. As a
result, regeneration of the action potential cannot occur. However, the thick membrane also
enables the axon to carry high voltage messages without breaking down. (P. Morell and W.T.
Norton, 1980)
The thick, non-regenerating myelin membrane is offset by periodic nodes, which are
also known as nodes ofRanvier[47]. The nodes are 100 outside diameter apart, as illustrated
at the top of Figure 2.5.
From Figure 2.5, the upper-wave form is an action potential, initiated by the first
node on the left. Then, theoretically, the action potential would fall to the dashed curve as
showed in the bottom of the picture. Instead, when the action potential reaches eight
millivolts, as indicated by the . (dots), the second node fires to generate a new action
potential. as shown in the lower wave form. This generation at the node is the same as that of
the unmyelinate axon.
-
7/27/2019 EMG Physiology
9/16
12
Figure 2.5: Nodes in the myelinated axon and action potential that generated.
2.4EMGELECTRODES
The EMG electrode could be explained by a receiving antenna concept. A receiving
antenna is an electrical device that detects oscillating magnetic fields, which are generated
from various sources. Then the signal is transmitted through the air from source to the
receiving antenna, a concept that is used to engineer the design of electrode. In terms of
recording the EMG signal, the muscle fiber is a biological signal generator, spreading out
over voltage fields to the volume-conductivity surrounded by fluid [18]. This fluid serves to
convey an EMG signal to an electrode, like air carry signals to an antenna.
The EMG recording starts from the beginning of the bioelectrical events as shown in
Figure 2.6. The changing conductivities in the membranes will make action currents flow
across the membranes as well as into the extracellular fluids around active cells. The
extracellular currents will then generate potential gradients as they flow through the resistive
extracellular fluids. The changing potential gradients, subsequently, will produce electrical
currents in the electrode leads by capacitive conductance across the metal/electrolyte
-
7/27/2019 EMG Physiology
10/16
13
interface of the electrode contacts. These weak currents will then flow through the high-
impedance circuits of the amplifier input stages, which will then convert these currents into
large output voltages.
Figure 2.6: The series of bioelectrical events.
The EMG electrodes can be classified by using its geometry. There are six classes of
EMG electrodes: monopolar electrode, bipolar electrode, tripolar electrode, multipolar
electrode, barrier or patch electrode, and belly tendon electrode.
A monopolar electrode takes potential from electrode and ground as the inputs to the
differential amplifier. When measuring, only a bare electrode is placed, without utilizing
other electrical connection. Because the ground yields a negative input to differential
amplifier, the potential from electrode is always based on ground.
-
7/27/2019 EMG Physiology
11/16
14
A bipolar electrode is used to measure the voltage different between two specific
points. It generally must be used with a differential amplifier. A bipolar electrode has two
contacts that are not connected to each other. Therefore, one node will be used for positive
input, and the other will be used for a negative input for the differential amplifier. Because
the differential amplifier treats both inputs equally, it will yield an accurate output. However
the distance between the electrodes could affect the measurement result. Placing the
electrodes too far from one another could yield a weak signal. On the other hand, placing
them too close may also result in unusable data, since the amplifier preprocesses each inputs
signal separately before subtracting those signals for output. In addition, there must be
another contact used as a reference point for these two inputs.
A tripolar electrode has three electrodes that are placed at equal intervals along a
straight line. The central electrode is usually connected to the positive input of a differential
amplifier, while the electrodes on the sides are usually connected to the negative input of a
differential amplifier. This configuration also requires another electrode to serve as a
reference. The tripolar electrode is often used to record nerve potentials, as its configuration
holds the advantage of being able to reject some forms of biological noise.
A multipolar electrode consists of rows of bipolar electrodes where an equal lead is
connected each side of bipolar electrodes to serve as a positive and negative input for a
differential amplifier. Besides, another electrode must be applied as a reference point. The
multipolar electrode is often used to record the activity of certain motor units based on
idiosyncrasies in their fiber locations.
The barrier or patch electrodes are typical bipolar electrodes that are closely
connected to a dielectrical barrier. The dielectrical is a non conductive substance that is
placed between the electrodes. This configuration redirects currents in extracellular flowing
around the tissue nearby. The patch also keeps the currents that are generated from tissues on
each side to prevent them from spreading into each other. Consequently, the potential
gradient of a desired action is larger, and the potential gradient of an undesired action is
smaller.
A belly tendon electrode is one of the fields of interest in the clinical EMG. Its
geometry is an interesting hybrid of the monopolar and bipolar approaches. In this technique,
the first electrode is placed in or over the middle point of the muscle of the belly, which
-
7/27/2019 EMG Physiology
12/16
15
serves as the positive input to the amplifier. The second electrode is placed over the tendon
of the same muscle, which is usually about the end of contractile elements, and serves as the
negative input to the amplifier. This arrangement gives a clean leading negative waveform,
since there is no virtual active contribution from tendon electrode. A belly tendon electrode is
employed specifically for tendon applications although it is not used for measuring a
selective muscle EMG recording during physiological activity.
All of electrode geometries discussed above could be considered as a dipole antenna
in term of electrical behavior. Monopolar electrodes are used to measure the EMG signal of
very small muscle. This is a good approach for sampling a signal that occurs near the surface
of an active single fiber. On the other hand, tripolar electrodes and multipolar electrodes are
used for sampling some large muscles.
2.5AMPLIFIER OF EMGSIGNALS
The amplifier is an electronic device that serves to boost low power signal to higher
power signal that is usable to perform work. There are two reasons to amplify the signal [19].
First, amplification increases the level of signal enough to protect an electrical interference
during transmission. Second, the signal is amplified so that it could be stored in a storage
device, or displayed by a measurement device like oscilloscope. In case of an EMG signal, an
amplifier is necessary. There are no such devices that can measure EMG signal without
amplification.
A differential amplifier is used to amplify an EMG signal, as it has the ability to
eliminate the noise from the signal [7]. As shown in Figure 2.7 the differential amplifier
takes two inputs, subtracts them and amplifies the different. In this case, if there is noise
interference through the input wires, the noise could be circuitry canceled out so long the
transmission of the two inputs is completely symmetrical manner.
It is difficult to make an amplifier with perfect subtraction. The Common Mode
Rejection Ratio (CMRR) could measure the accuracy of subtraction in each amplifier. It is
suggested to have a CMRR value at 90dB in order to sufficiently discard a contaminated
noise [7]. Yet with modern technology, the differential amplifier could make a CMRR value
of 120dB. However, even though a differential amplifier has the ability to reduce unwanted
noise signals that occur from both sides of the input wires, contaminated noises could still
-
7/27/2019 EMG Physiology
13/16
16
exist. This noise could have been injected into the signal by a stray capacitance that has been
amplified, and thus, degrading the signal [19].
Figure 2.7 A schematic of the differential amplifier configuration. The EMG signal is
represented by m and the noise signal by n.
Every electronic component or even an amplifier itself behaves as an effective filter,
since there are no such electronic devices that can transfer all frequency range. The electrode
itself tends to have lower impedance for a higher frequency and have higher impedance for a
lower frequency [19]. The connection of electrode, cable and amplifier creates an implicit
filter effect. The electrode contacts are connected in series to an amplifier; they function
similar to that of a capacitor, while an impedance of amplifier is similar to that of a resistor.
This connection visualizes a High-Pass filter circuit. The low frequency voltage tends to be
attenuated and drop the highest voltage across the electrode contacts rather than the
amplifier. On the other hand, the cables that connect electrodes to an amplifier have a stray
capacitor behavior. It is considered that this capacitor is connected to ground, which
simulates a Low-Pass filter circuit. The stray capacitor will provide low impedance, at which
the high frequency picked up by electrodes tend to drop their voltage here. Therefore an
amplifier will see an attenuation of high frequency.
The implicit filter could cause signal problems if it is not considered carefully in the
design of an amplifier. The explicit filter with real components (resistor and capacitor)
functions by using the same concept of an implicit filter, and could help in increasing the
signal-to-noise ratio. Since a signal is desired to be within in some frequency range, it is good
-
7/27/2019 EMG Physiology
14/16
17
idea to have an explicit filter for that particular band. Therefore, noise with the frequency
outside a desired frequency band will be distorted. (see Figure 2.8)
Figure 2.8 Implicit RC filter.
The implicit filter is important in designing a differential amplifier. To reduce an
implicit capacitance effect, the electrode contact should be placed close to an amplifier. In
other word, an amplifier should be located as close to the signal source as possible.
A raw EMG signal is an AC signal. Its bandwidth could occur anywhere from a few
tens of cycles per second to 3000 cycles per second. Therefore, sometime a large amount of
DC voltage appeared at the output of preamplifier [19]. A DC offset can be removed by
adding a series capacitor to the output, which also allows the AC signal to pass through. The
DC offset can also happen between the recording electrodes themselves, especially if they are
made from different materials. A battery-powered source sometime could overpower a
sensitive preamplifier, which could also disgrace the contact of electrodes or damage
surrounding tissue. Placing a capacitor between inputs of AC coupling could prevent this
-
7/27/2019 EMG Physiology
15/16
18
problem. (In chapter 3, Figure 3.6, a small capacitor is placed at the inputs to the preamplifier
to prevent this problem.)
A DC-offset-adjust potentiometer is often added to sensitive equipment, since it is
easy to adjust the DC-offset. However, the DC offset also dependents on temperature [19].
Therefore, it is better to power up the equipment for sometime before adjusting the DC
offset.
2.6PROBLEM WITH NOISE AND ARTIFACTS
In the process of recording an EMG signal, the source of the generated signal is not
only from bioelectrical generator or active cell, but also from any electrical fields that occur
around an electrode and lead cables. These electrical fields produce some signals that could
also be added to an EMG signal, causing a form of interference that is called noise.Interference noise can be produced from anything that has an electrical field such as power
lines, computer monitors, transformers, or EMG amplifier itself. Once noise has occurred, it
could cause problem in recording an EMG signal. Therefore while planning out the design of
the amplifier device and recording the EMG signal, noise factors should be taken into
consideration, since noise could come from a variety of sources such as electronic
components, recording devices, ambient noise, motion artifacts, or inherent instability of the
signal [7].
Any electronic devices can produce noise. The noise frequency is range between 0 Hz
to a 1000 Hz [7]. This kind of noise cannot be eliminated. Using an intelligent circuit design
and a good quality of electronic components to construct the device can only reduce the
noise.
Ambient noise could be generated from any electronic device that created an
electromagnetic field such as televisions, computer monitors, motors, electrical power lines,
fluorescent lamps or light bulbs. In fact there are radio waves and magnetic fields floating all
over our body. It is virtually impossible to drain these radiators to ground (earth surface). The
ambient noise also cannot be avoided. The ambient noise frequency occurs primarily within
the range 50 Hz or 60 Hz [7] [44], while the amplitude of an ambient noise is about one to
three times grater than that of an EMG signal.
-
7/27/2019 EMG Physiology
16/16
19
Motor artifacts come from two sources [7]; first, from the contact of an electrode to
skin; second, from the connection of cable from electrode to the amplifier. When performing
a grasping experiment, a movement of the wire alone could cause a noise problem. This
electrical noise from both sources has the frequency range between zero to twenty hertz [7].
However, a proper design of circuitry with a good connector and stable electrode contact
could reduce this motion artifact problem.
Inherent instability of the signal is caused by a nature of EMG signal. The amplitude
of the EMG signal frequency range between zero hertz to twenty hertz is particularly
unstable due to the quasi-random nature of the firing rate of motor units. It is suggested to
consider an EMG signal frequency in this range as an unwanted noise signal [7].
There is no boundary in what amplitude of an EMG signal is good for yielding
accurate recordings. While a certain amount of noise could be tolerated, the question is
exactly how much noise could be allowed. In other words, there is a question about the
tolerable levels of thesignal-to-noise ratio. The signal-to-noise ratio is determined by taking
the ratio of the amplitude of desire signal over the amplitude of added noise signal. The main
concern now is in the lever of the signal-to-noise ratio that could degrade an analysis result.
If the noise is produced from thermal motion, then twice of desired signal amplitude over the
noise amplitude is good enough. However, if the noise occurred periodically (pause like) and
forms a pattern similar to the desired signal, it is suggested that ten times grater of the desire
signal amplitude than the noise amplitude would be clarified a confusing event [20]. In order
to determine signal-to-noise ratio, the study of noise behavior alone may be required to see
how noise could affect the real signal.