P.O. Box 489 1350 Tolland Road Rollinsville, CO 80474 Phone: (303) 258-0100 FAX: (303) 258-0775 www.criteriontech.com CONFIDENTIAL AND PROPRIETARY NVLAP LAB CODE 100396-0 EMC QUALIFICATION TEST REPORT ZYNEX MEDICAL, INC. NEURO-MOVE, NM-900 TESTED TO CONFORM WITH: EMISSIONS STANDARDS IMMUNITY STANDARDS FOR INDUSTRIAL, SCIENTIFIC AND MEDICAL (ISM) Test Report Number: 071010-1176 Date of Issue: February 13, 2008 Date of Test Completion: February 6, 2008 Manufacturer’s Address: 8022 Southpark Circle, Unit 100 Littleton, CO 80120 Phone: 303-707-0203 Approved by: Laboratory Director

Transcript of EMC QUALIFICATION TEST REPORT ZYNEX … · criterion technology emc qualification test report sheet...

• P.O. Box 489 • 1350 Tolland Road • Rollinsville, CO 80474 • Phone: (303) 258-0100 • FAX: (303) 258-0775 •

• www.criteriontech.com •

CONFIDENTIAL AND PROPRIETARY

NVLAP LAB CODE 100396-0

EMC QUALIFICATION TEST REPORT

ZYNEX MEDICAL, INC.

NEURO-MOVE, NM-900

TESTED TO CONFORM WITH:

EMISSIONS STANDARDS IMMUNITY STANDARDS

FOR

INDUSTRIAL, SCIENTIFIC AND MEDICAL (ISM)

Test Report Number: 071010-1176

Date of Issue: February 13, 2008

Date of Test Completion: February 6, 2008

Manufacturer’s Address: 8022 Southpark Circle, Unit 100

Littleton, CO 80120

Phone: 303-707-0203

Approved by:

Laboratory Director

Sheet 2 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

DISCLAIMERS

This report is the confidential property of the client. For the protection of our clients and ourselves, extracts from this test report cannot be produced without prior written approval from Criterion Technology. Reproduction of the complete report can be performed at the client’s discretion.

The client is aware that Criterion Technology has performed testing in accordance with the applicable standard(s). Test data is accurate within ANSI parameters for Emissions testing, unless a specific level of accuracy has been defined in writing prior to testing, by Criterion Technology and the client.

Criterion Technology reports apply only to the specific Equipment Under Test (EUT) sample(s) tested under the test conditions described in this report. If the manufacturer intends to use this report as a document demonstrating compliance of this model, additional models of this product must have electrical and mechanical characteristics identical to the device tested for this report. Criterion Technology shall have no liability for any deductions, inferences, or generalizations drawn by the client or others from Criterion Technology issued reports.

Total liability is limited to the amount invoiced for the testing of this EUT and the contents of this report are not warranted.

Compliance with the appropriate governmental standards is the responsibility of the manufacturer. Any questions regarding this report should be directed to:

Laboratory Director Criterion Technology Corp. P.O. Box 489 1350 Tolland Road Rollinsville, Colorado 80474 Phone: (303) 258-0100 Fax: (303) 258-0775 mailto:[email protected]

NVLAP Note: Criterion Technology is accredited by the National Voluntary Laboratory Accreditation Program (NVLAP) for the specific scope of accreditation under Lab Code 100396-0. Test methods included in Lab Code 100396-0 are

12/CIS11a - IEC/CISPR 11, edition 3.1 (1999-08) 12/CIS11b - IEC/CISPR 11 (2003) & EN 55011 (1998), A1 (1999), A2 (2002) 12/CIS11c - IEC/CISPR 11 (1997), A1 (1999), A2 (2002) 12/CIS11d - EN 55011 (1998), A1 (1999), A2 (2001) 12/CIS22 - IEC/CISPR 22 (1997) and EN 55022 (1998) + A1 (2000) 12/CIS22a - IEC/CISPR 22 (1993) and EN 55022 (1994), Amendment 1:1995 & Amendment 2: 1996 12/CIS22b - CNS13438 (1997) 12/EM02a - IEC 61000-3-2, Edition 2.1 (2001-10) and EN 61000-3-2 (2000), and AS/NZS 2279.1 (2000) 12/EMO2b - IEC 61000-3-2, Second Edition (2000-08) 12/EMO2c - BS EN 61000-3-2, ED. 2 (2001); IEC 610003-2, ED. 2 (2000) 12/EM03 - ICE 61000-3-3 (1995); EN 61000-3-3 (1995); AS/NZS 2279.3 (1995) 12/EM03a - ICE 1000-3-3 (1994-12) 12/EM03b - ICE 61000-3-3 Edition 1.1 (2002-03) & EN 61000-3-3, A1 (2001) 12/EM03c - ICE 61000-3-3 (1994) with Amendment 1 (2001) 12/EM03d - ICE 61000-3-3 (1995) + A1 (2001) 12/FCC15b - ANSI C63.4 (2003) with FCC Method 47 CFR Part 15, Subpart B 12/T51 - AS/NZS CISPR 22 (2002) and AS/NZS 3548 (1997) 12/I01 – IEC 61000-4-2, Ed. 2.1 (2001), A1, A2; EN 61000-4-2 12/I02 – IEC 61000-4-3, Ed. 2.0 (2002-03); EN 61000-4-3 (2002) 12/I03 – IEC 61000-4-4 (1995), A1 (2002), A2 (2001); EN 61000-4-4 12/I04 – IEC 61000-4-5, Ed. 1.1 (2001-04); EN 61000-4-5 12/I05 – IEC 61000-4-6, Ed. 2.0 (2003-05); EN 61000-4-6 12/I06- - IEC 61000-4-8, Ed. 1.1 (2001); EN 61000-4-8 12/I07 – IEC 61000-4-11, Ed. 1.1 (2001-03); EN 61000-4-11

The NVLAP Logo on the front cover of this report applies only to data taken for the above test methods.

This report may contain data which is not covered by the NVLAP accreditation.

This report must not be used by the client to claim product endorsement by NVLAP or any agency of the U.S. Government.

Criterion Technology has been accredited by the following groups: NVLAP(#100396-0), FCC(#90688), BSMI(#SL2-IN-007R), VCCI(#1255) 3&10 meter site (#R-2097), Immunity Shield room(#C-2269), Open Area Site(#C-2270), Nemko(#ELA-214), NMi (EU Competent Body Accreditation) and Industry Canada(#IC 3301). The National

Institute for Standards and Technology (NIST) has designated Criterion Technology a Conformity Assessment Body (CAB) for Taiwan (BSMI # SL2-IN-E-007R). riterion Technology instrumentation and accessories used to test products for compliance to the indicated standards are calibrated regularly in

accordance with ISO 9002, ISO 17025, ANSI/NCSL Z540-I-1994 and are traceable to national standards.

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 3 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

TABLE OF CONTENTS 1.0 EXECUTIVE SUMMARY.............................................................................................................................................................. 4

1.1 PURPOSE..........................................................................................................................................................................................4 1.2 CONFORMITY ................................................................................................................................................................................4 1.3 EQUIPMENT UNDER TEST (EUT) ...........................................................................................................................................4

2.0 EMISSIONS TEST STANDARDS................................................................................................................................................ 5 2.1 RADIATED EMISSIONS – 30 MHZ TO 1000 MHZ ..............................................................................................................5

3.0 IMMUNITY STANDARDS ............................................................................................................................................................ 6 3.1 IMMUNITY TEST STANDARDS...............................................................................................................................................6 3.2 PERFORMANCE CRITERIA .......................................................................................................................................................6 3.3 ELECTROSTATIC DISCHA RGE (ESD)...................................................................................................................................7 3.4 RADIATED RF ELECTROMAGNETIC (EM) FIELD IMMUNITY..................................................................................8

4.0 APPENDIX A: EUT PHOTOGRAPHS ..................................................................................................................................... 9 4.1 RADIATED EMISSIONS – FRONT VIEW ...............................................................................................................................9 4.2 ELECTROSTATIC DISCHA RGE..............................................................................................................................................10 4.3 RADIATED RF ELECTROMAGNETIC FIELD IMMUNITY...........................................................................................11

5.0 APPENDIX B: DATA SHEETS .................................................................................................................................................12 5.1 RADIATED EMISSIONS PLOT – 30 MHZ TO 1 GHZ........................................................................................................12 5.2 RADIATED EMISSIONS TABLE – 30 MHZ TO 1 GHZ ....................................................................................................13 5.3 ELECTROSTATIC DISCHA RGE (EN61000-4-2, IEC1000-4-2, IEC 801-2) ..................................................................17 5.4 RADIATED SUSCEPTIBILITY (EN 61000-4-3, IEC 1000-4-3, IEC 801-3) ...................................................................18

6.0 APPENDIX C: PRODUCT INFORMATION FORM.........................................................................................................19

7.0 APPENDIX D: TES T EQUIPMENT AND CALIBRATION STATUS ..........................................................................21

8.0 APPENDIX E: TEST DIRECTIVES, STANDARDS AND METHODS ........................................................................22

Sheet 4 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

EMC QUALIFICATION TEST REPORT NEURO-MOVE, NM-900

1.0 EXECUTIVE SUMMARY 1.1 PURPOSE

The purpose of this report is to present EMC test data and demonstrate conformity to the requirements of the prescribed standards for Emissions and/or Immunity.

1.2 CONFORMITY

The test article was tested to the standards listed in Table I with the indicated conformity status. All test methods were performed in accordance to with the standards listed.

TABLE I. EMISSIONS CONFORMITY SUMMARY

TEST TYPE COMPLIANCE STANDARD

TESTING TECHNIQUE

TEST DESCRIPTION

PRODUCT CLASSIFICATION

CONFORMITY STATUS

EMISSIONS

FCC Part 15

ICES-003

AS/NZ 3548

IEC/EN 55022

Radiated Emissions

Class B

PASSED

TABLE II. IMMUNITY CONFORMITY SUMMARY

TEST TYPE COMPLIANCE STANDARD

TESTING TECHNIQUE

TEST DESCRIPTION

MINIMUM PERFORMANCE

CRITERIA

CONFORMITY STATUS

61000-4-2 Electrostatic Discharge B PASSED 61000-4-3 Radiated, RF Electromagnetic Field Amplitude Modulated PASSED

IMMUNITY EN 60601-1-2

ENV 50204

Radiated, RF Electromagnetic Field

Pulse Modulated

A PASSED

1.3 EQUIPMENT UNDER TEST (EUT)

EUT NAME: NEURO-MOVE

EUT MODEL/PART NUMBER(S): NM-900

EUT SERIAL NUMBER(S): 2690979 & 2791123

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 5 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

2.0 EMISSIONS TEST STANDARDS

EN 55022 for IT Equipment Class B

FCC Part 15, Subpart B Class B ICES-003 Class B AS/NZ 3548 Class B

2.1 RADIATED EMISSIONS – 30 MHZ TO 1000 MHZ

Measurements for Radiated Emissions were performed over the frequency range of 30 MHz to 1000 MHz in the horizontal and vertical antenna polarities to the requirements of: EN 55022 for IT Equipment Class B

Testing Conditions Date of Test: January 21, 2008

Temperature: 20°C

Relative Humidity: 15%

Test Voltage: 7.4 V Battery

Test Operator: lws

Test Location

Criterion Technology Open Area Test Site

Test Distance

Antenna Distance: 10 meter(s) Final Measurement(s)

Test Equipment

Hewlett-Packard Spectrum Analyzer, HP 8566B Hewlett-Packard Quasi-Peak Adapter, HP 85650A

Hewlett-Packard Tracking Generator, HP 85645A

Rohde and Schwarz Receiver, ESHS-30 Rohde and Schwarz Receiver, ESVS-30

Mini Circuits Pre-Amp #2 Veratech Pre-Amp #3

Chase BiLog Antenna, Model 1121 Antenna Research, Horn Antenna, Model DRG118/A

EMCO BiConnical Antenna, Model 3108 EMCO Log Periodic Antenna, Model 3146 Test Results of Radiated Emissions Test Status: PASSED Frequency Range: 30 MHz to 1000 MHz

Remarks

See: APPENDIX A for EUT Photographs APPENDIX B for Data Sheets APPENDIX D for Test Equipment Calibration Status

Minimum Margin to Limit: -10.27 dB at 30.4997 MHz

Sheet 6 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

3.0 IMMUNITY STANDARDS EN 60601-1-2: Medical Electrical Equipment

3.1 IMMUNITY TEST STANDARDS.

TABLE II. IMMUNITY TESTS

BASIC STANDARDS TESTED ENVIRONMENTAL

PHENOMENA SPECIFICATIONS/UNITS REQUIRED PERFORMANCE

EN 61000-4-2 Electrostatic Discharge ±2,4,8 kV Air

±2,4 kV Contact

Performance Criterion B

EN 61000-4-3 Radiated, RF Electromagnetic Field -Amplitude Modulated

3 V/m (unmodulated, RMS) 80%, 1 kHz AM 80 MHz - 1 GHz

ENV 50204 Radiated, RF Electromagnetic Field -Pulse Modulated

3 V/m (unmodulated, RMS) 50% duty cycle 200 Hz repetition frequency 900 ±5 MHz

Performance Criterion A

3.2 PERFORMANCE CRITERIA 3.2.1 Performance Criterion A

The apparatus shall continue to operate as intended. No degradation of performance or loss of function is allowed below a performance level specified by the manufacturer, when the apparatus is used as intended. The performance level may be replaced by a permissible loss of performance. If the minimum performance level or the permissible performance loss is not specified by the manufacturer then either of these may be derived from the product description and documentation and what the user may reasonably expect from the apparatus if used as intended.

3.2.2 Performance Criterion B

The apparatus shall continue to operate as intended after the test. No degradation of performance or loss of function is allowed below a performance level specified by the manufacturer, when the apparatus is used as intended. The performance level may be replaced by a permissible loss of performance. During the test, degradation of performance is however allowed. No change of actual operating state or stored data is allowed. If the minimum performance level or the permissible performance loss is not specified by the manufacturer then either of these may be derived from the product description and documentation and what the user may reasonably expect from the apparatus if used as intended.

3.2.3 Performance Criterion C

Temporary loss of function is allowed, provided the loss of function is self recoverable or can be restored by the operation of the controls.

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 7 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

3.3 ELECTROSTATIC DISCHARGE (ESD)

Measurements of immunity against ESD were performed to the requirements of EN 61000-4-2.

Testing Conditions

Date of Test: February 6, 2008

Temperature: 17°C

Relative Humidity: 24%

Atmospheric Pressure: 73.87kPa

Test Voltage: 7.4 V Battery

Test Operator: lws

Test Location

Criterion Technology Semi-Anechoic Chamber

Test Equipment

Haefely Trench PESD, 1600

Test Setup Air Contact

Discharge Type:

Discharge Voltages: ±2,4,8 kV ±2,4 kV

Discharge Polarity: Positive/Negative Positive/Negative

Discharge Factor: ≥1 second ≥1 second

Discharge Number: ≥10 ≥10

Discharge Impedance: 330 ohms/150 pF

Discharge Locations: Human-Interface Accessible

See Photographs APPENDIX A

Test Results of ESD

Test Status: PASSED Performance Criterion A

Remarks

See: APPENDIX A for EUT Photographs APPENDIX B for Data Sheets APPENDIX D for Test Equipment Calibration Status

Sheet 8 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

3.4 RADIATED RF ELECTROMAGNETIC (EM) FIELD IMMUNITY

Measurements of immunity against Radiated RF EM Fields were performed to the requirements of:

EN 61000-4-3 ENV 50204

Testing Conditions

Date of Test: January 14, 2008

Temperature: 18°C

Relative Humidity: 28%

Atmospheric Pressure: 75.06kPa

Test Voltage: 7.4 V Battery

Test Operator: lws

Test Location

Criterion Technology Semi-Anechoic Chamber

Test Equipment

Amplifier Research Field-Strength Monitoring System, FM2000/FP2000

Amplifier Research Power Amplifier, 100W1000M1

Amplifier Research Power Amplifier, 150A100 Amplifier Research Power Amplifier, 10S1G4

Amplifier Research Log Periodic Antenna, Model AT1080

EMCO Log Periodic Antenna, Model 3146

HP Signal Generator, HP8648D HP Spectrum Analyzer, HP8566B/85662A Test Specifications

Frequency Range: 80 MHz to 1 GHz 900 ±5 MHz

Field Strength: 3 V/m Other: 3 V/m

Modulation: AM - 1 kHz, 80% sinewave Pulse ON/OFF, 100%, 200 Hz None

Step: 1% 3.0 second(s) Dwell Time Antenna Distance: 3 meter(s)

Antenna Polarization: Horizontal Vertical EUT Position: Front Left Top Back Right Bottom

Test Results of Radiated RF EM Field Immunity

Test Status: PASSED Performance Criterion A

Remarks

See: APPENDIX A for EUT Photographs APPENDIX B for Data Sheets APPENDIX D for Test Equipment Calibration Status

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 9 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

4.0 APPENDIX A: EUT PHOTOGRAPHS 4.1 RADIATED EMISSIONS – FRONT VIEW

Sheet 10 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

4.2 ELECTROSTATIC DISCHARGE

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 11 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

4.3 RADIATED RF ELECTROMAGNETIC FIELD IMMUNITY

Sheet 12 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

5.0 APPENDIX B: DATA SHEETS 5.1 RADIATED EMISSIONS PLOT – 30 MHZ TO 1 GHZ Criterion Technology Date: January 21, 2008 EUT: Neuro-Move, NM-900 S/N: 2690979 Manufacturer:Zynez Medical, Inc. Tester: lws SpiD: 071010-1176 EUT Information: tabletop, 7.4 V battery Test Information: continuous operation, 10M, EN 55022 per EN 60601-1-2, Class B Test Cond: Temp: 20°C Humidity: 15%

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 13 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

5.2 RADIATED EMISSIONS TABLE – 30 MHZ TO 1 GHZ Notes: The third column below contains alpha characters which pertain to the type of measurements made. The following are the definitions for those characters: q = Quasi Peak, m = Maximized (cable, rotation and antenna height), s = scanned but no data taken, and a = average. For the first character in column four, a ‘-’ indicates that value is below the limit while an ‘*’ indicates that value is above the limit If the list is sorted using “I-sort”, then quasi-peak and average levels are weighted higher than peak levels and are moved to the front of the scan list. The following keys help to better understand the data: TT: Turntable position in degrees Hght: Height of antenna in centimeters Az: Azimuth, V = Vertical, H= Horizontal Criterion Technology Mon Jan 21 18:04:06 2008 EUT: Neuro-Move, NM-900 S/N: 2690979 Manufacturer: Zynex Medical, Inc. Tester: lws Special ID: 071010-1176 EUT Information: tabletop, 110 VAC 60 Hz & 7.4 V battery Test information: continuous operation, 10m, EN 55022 per EN 60601-1-2, Class B

Table 1: Scan List, sorted by margin to limit 22-B, -40.0dB filter

Freq, MHz Value dBuV/m Sts Margin to 22-B limits (dB) TT Hght Az Comment 30.4997 19.73 m -10.27 273 101 V . 440.5977 26.67 m -10.33 0 101 H . 167.3467 17.78 q -12.22 73 101 V . 512.2977 22.73 q -14.27 0 246 H . 160.9267 11.84 q -18.16 -2 101 V . 157.3507 9.23 q -20.77 0 101 V . 434.0677 15.99 m -21.01 90 101 V . 177.0007 8.82 q -21.18 270 162 V . 199.4827 8.64 q -21.36 270 401 H . 178.0517 8.31 q -21.69 0 101 V . 74.0777 8.08 m -21.92 270 402 H . 414.6927 14.82 q -22.18 270 101 V . 408.3777 14.59 q -22.41 90 101 V . 180.2057 7.18 q -22.82 270 162 V . 196.2657 6.96 q -23.04 90 401 H . 189.8497 6.79 q -23.21 0 401 H . 193.0587 6.69 q -23.31 270 401 H . 183.3977 6.63 q -23.37 0 162 V . 186.6097 6.50 q -23.50 90 101 V . 87.0357 5.90 q -24.10 180 401 H . 353.7427 12.27 q -24.73 270 98 V . 347.2877 12.16 q -24.84 90 401 H . 263.7377 12.09 q -24.91 180 401 H . 302.2877 11.68 q -25.32 0 401 H . 295.8727 11.00 q -26.00 180 401 H . 321.5927 10.87 q -26.13 0 401 H . 254.0977 10.84 q -26.16 0 401 H . 292.6627 10.81 q -26.19 180 401 H . 315.1577 10.75 q -26.25 180 401 H . 289.4127 10.67 q -26.33 0 101 V . 308.7177 10.45 q -26.55 180 401 H .

Minimum Margin to Limit: -10.27 dB at 30.4997 MHz

Sheet 14 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

283.0327 10.41 q -26.59 180 401 H . 270.1627 10.35 q -26.65 270 401 H . 276.6077 10.33 q -26.67 0 401 H . 247.6527 9.85 q -27.15 0 401 H . 241.2727 9.33 q -27.67 0 401 H .

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 15 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

Table 2: Scan List for 22-B, sorted by Frequency, -40.0dB filter Freq, MHz Final Value dBuV/m Sts Margin to 22-B limits (dB) TT Hght Az Comment 30.4997 19.73 m -10.27 273 101 V . 74.0777 8.08 m -21.92 270 402 H . 87.0357 5.90 q -24.10 180 401 H . 157.3507 9.23 q -20.77 0 101 V . 160.9267 11.84 q -18.16 -2 101 V . 167.3467 17.78 q -12.22 73 101 V . 177.0007 8.82 q -21.18 270 162 V . 178.0517 8.31 q -21.69 0 101 V . 180.2057 7.18 q -22.82 270 162 V . 183.3977 6.63 q -23.37 0 162 V . 186.6097 6.50 q -23.50 90 101 V . 189.8497 6.79 q -23.21 0 401 H . 193.0587 6.69 q -23.31 270 401 H . 196.2657 6.96 q -23.04 90 401 H . 199.4827 8.64 q -21.36 270 401 H . 241.2727 9.33 q -27.67 0 401 H . 247.6527 9.85 q -27.15 0 401 H . 254.0977 10.84 q -26.16 0 401 H . 263.7377 12.09 q -24.91 180 401 H . 270.1627 10.35 q -26.65 270 401 H . 276.6077 10.33 q -26.67 0 401 H . 283.0327 10.41 q -26.59 180 401 H . 289.4127 10.67 q -26.33 0 101 V . 292.6627 10.81 q -26.19 180 401 H . 295.8727 11.00 q -26.00 180 401 H . 302.2877 11.68 q -25.32 0 401 H . 308.7177 10.45 q -26.55 180 401 H . 315.1577 10.75 q -26.25 180 401 H . 321.5927 10.87 q -26.13 0 401 H . 347.2877 12.16 q -24.84 90 401 H . 353.7427 12.27 q -24.73 270 98 V . 408.3777 14.59 q -22.41 90 101 V . 414.6927 14.82 q -22.18 270 101 V . 434.0677 15.99 m -21.01 90 101 V . 440.5977 26.67 m -10.33 0 101 H . 512.2977 22.73 q -14.27 0 246 H .

Sheet 16 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

Table 3: Complete Scan List Sorted by Frequency Freq, MHz I-val before xducr

factors dBuV Final Value dBuV/m

Sts TT Hght Az Time Comment

30.4997 24.02 19.73 m 273 101 V Mon Jan 21 16:58:31 2008 .

74.0777 23.18 8.08 m 270 402 H Mon Jan 21 17:47:27 2008 .

87.0357 19.40 5.90 q 180 401 H Wed Jan 16 16:02:08 2008 .

157.3507 19.94 9.23 q 0 101 V Wed Jan 16 16:22:03 2008 .

160.9267 22.82 11.84 q -2 101 V Mon Jan 21 17:43:12 2008 .

167.3467 29.26 17.78 q 73 101 V Mon Jan 21 17:29:29 2008 .

177.0007 20.46 8.82 q 270 162 V Wed Jan 16 15:57:20 2008 .

178.0517 20.15 8.31 q 0 101 V Wed Jan 16 16:22:12 2008 .

180.2057 19.38 7.18 q 270 162 V Wed Jan 16 15:57:24 2008 .

183.3977 18.90 6.63 q 0 162 V Wed Jan 16 16:20:21 2008 .

186.6097 18.56 6.50 q 90 101 V Wed Jan 16 16:12:12 2008 .

189.8497 18.70 6.79 q 0 401 H Wed Jan 16 16:17:55 2008 .

193.0587 18.49 6.69 q 270 401 H Wed Jan 16 16:00:47 2008 .

196.2657 18.47 6.96 q 90 401 H Wed Jan 16 16:16:01 2008 .

199.4827 19.75 8.64 q 270 401 H Wed Jan 16 16:00:51 2008 .

241.2727 18.42 9.33 q 0 401 H Wed Jan 16 16:18:03 2008 .

247.6527 18.22 9.85 q 0 401 H Wed Jan 16 16:18:06 2008 .

254.0977 18.54 10.84 q 0 401 H Wed Jan 16 16:18:08 2008 .

263.7377 20.08 12.09 q 180 401 H Wed Jan 16 16:02:47 2008 .

270.1627 18.29 10.35 q 270 401 H Wed Jan 16 16:01:02 2008 .

276.6077 18.18 10.33 q 0 401 H Wed Jan 16 16:18:14 2008 .

283.0327 18.08 10.41 q 180 401 H Wed Jan 16 16:02:54 2008 .

289.4127 18.12 10.67 q 0 101 V Wed Jan 16 16:22:43 2008 .

292.6627 18.14 10.81 q 180 401 H Wed Jan 16 16:02:58 2008 .

295.8727 18.24 11.00 q 180 401 H Wed Jan 16 16:03:00 2008 .

302.2877 19.17 11.68 q 0 401 H Wed Jan 16 16:18:25 2008 .

308.7177 18.11 10.45 q 180 401 H Wed Jan 16 16:03:05 2008 .

315.1577 18.01 10.75 q 180 401 H Wed Jan 16 16:03:07 2008 .

321.5927 17.99 10.87 q 0 401 H Wed Jan 16 16:18:31 2008 .

347.2877 18.13 12.16 q 90 401 H Wed Jan 16 16:16:35 2008 .

353.7427 17.98 12.27 q 270 98 V Wed Jan 16 15:54:39 2008 .

408.3777 18.27 14.59 q 90 101 V Wed Jan 16 16:12:56 2008 .

414.6927 18.41 14.82 q 270 101 V Wed Jan 16 15:56:26 2008 .

434.0677 19.35 15.99 m 90 101 V Mon Jan 21 17:38:17 2008 .

440.5977 29.95 26.67 m 0 101 H Wed Jan 16 16:27:22 2008 .

512.2977 24.65 22.73 q 0 246 H Mon Jan 21 17:24:20 2008 .

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 17 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

5.3 ELECTROSTATIC DISCHARGE (EN61000-4-2, IEC1000-4-2, IEC 801-2)

TEST NUMBER: 071010-1176 TEST ARTICLE: Neuro-Move

MODEL NUMBER: NM-900 SERIAL NUMBER: 2791123

TEMPERATURE: 17ºC HUMIDITY: 24%

ATMOSPHERIC PRESSURE: 73.78kPa TEST PERSONNEL: lws

TEST RESULTS : Complies ( X ) Does Not Comply ( )

EUT OPERATING VOLTAGE: 7.4 V battery TEST DATE: 2/6/08 DISCHARGE METHOD: Air (A) Contact (C)

TEST POINT

DESCRIPTION

DISCHARGE

VOLTAGE

DISCHARGE

NOTE

REQUIRED PERFORMANCE

(A,B,or C)*

TEST PERFORMANCE

(A,B, or C)*

PASS/ FAIL

OBSERVED RESPONSE

OF THE EUT

Front panel push buttons (4) A ±2, 4, 8 kV 1 B A PASS Normal Operation

Front panel LCD screen corners (4)

A ±2, 4, 8 kV 1 B A PASS Normal Operation

Front panel LCD screen center (4 pl)

A ±2, 4, 8 kV 1 B A PASS Normal Operation

Front panel stim indicator A ±2, 4, 8 kV 1 B A PASS Normal Operation

Bezel to enclosure joint A ±2, 4, 8 kV 1 B A PASS Normal Operation

Clam shell joint A ±2, 4, 8 kV 1 B A PASS Normal Operation

Rear panel connector A ±2, 4, 8 kV 1 B A PASS Normal Operation

Rear panel corners (4) A ±2, 4, 8 kV 1 B A PASS Normal Operation

Horiz Coupling Plane

EUT Front Left C ±2, 4 kV 1 B A PASS Normal Operation

EUT Front Right C ±2, 4 kV 1 B A PASS Normal Operation

EUT Right Front C ±2, 4 kV 1 B A PASS Normal Operation

EUT Right Back C ±2, 4 kV 1 B A PASS Normal Operation

EUT Rear Left C ±2, 4 kV 1 B A PASS Normal Operation

EUT Rear Right C ±2, 4 kV 1 B A PASS Normal Operation

EUT Left Front C ±2, 4 kV 1 B A PASS Normal Operation

EUT Left Back C ±2, 4 kV 1 B A PASS Normal Operation

Vert. Coupling Plane

EUT Front Left C ±2, 4 kV 1 B A PASS Normal Operation

EUT Front Right C ±2, 4 kV 1 B A PASS Normal Operation

EUT Right Front C ±2, 4 kV 1 B A PASS Normal Operation

EUT Right Back C ±2, 4 kV 1 B A PASS Normal Operation

EUT Rear Left C ±2, 4 kV 1 B A PASS Normal Operation

EUT Rear Right C ±2, 4 kV 1 B A PASS Normal Operation

EUT Left Front C ±2, 4 kV 1 B A PASS Normal Operation

EUT Left Back C ±2, 4 kV 1 B A PASS Normal Operation

*NOTE: Performance Criteria categories A, B, and C are as defined in EN 60601-1-2.

The Performance Criterion for passing is as shown in Table II of this report.

Refer to Setup Photos to see the test points. Discharge Notes: 1. No perceived discharge, and no observed response in the EUT. 2. Discharge observed, but no observed response in the EUT. 3. Discharge observed, and the EUT was affected.

Sheet 18 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

5.4 RADIATED SUSCEPTIBILITY (EN 61000-4-3, IEC 1000-4-3, IEC 801-3)

TEST NUMBER: 071010-1176 TEST ARTICLE: Neuro-Move MODEL NUMBER: NM-900 SERIAL NUMBER: 2791123 TEMPERATURE: 18ºC HUMIDITY: 28% ATMOSPHERIC PRESSURE: 75.06 kPa DWELL TIME: 3 Seconds

TEST DATE: 1/14/08 TEST PERSONNEL: lws EUT OPERATING VOLTAGE: 7.4 V battery

TEST FREQ.

(MHz) FIELD

STRENGTH (V/m)

MODULATION FREQ. %

FIELD POLARITY

TESTED SIDE

OF EUT

TEST PERFORMANCE

(A,B, or C)

TEST PERFORMANCE

(A,B, or C)

(PASS/ FAIL)

OBSERVED RESPONSE OF THE EUT

80 to 1000 3 1kHz 80%AM Horizontal Front A A PASS Normal Operation

SPOT 3 1kHz 80%AM Horizontal Front A A PASS Normal Operation

900 3 200 Hz pulse Horizontal Front A A PASS Normal Operation

900 3 200 Hz pulse Horizontal Right A A PASS Normal Operation

SPOT 3 1kHz 80%AM Horizontal Right A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Horizontal Right A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Horizontal Rear A A PASS Normal Operation

SPOT 3 1kHz 80%AM Horizontal Rear A A PASS Normal Operation

900 3 200 Hz pulse Horizontal Rear A A PASS Normal Operation

900 3 200 Hz pulse Horizontal Left A A PASS Normal Operation

SPOT 3 1kHz 80%AM Horizontal Left A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Horizontal Left A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Vertical Left A A PASS Normal Operation

SPOT 3 1kHz 80%AM Vertical Left A A PASS Normal Operation

900 3 200 Hz pulse Vertical Left A A PASS Normal Operation

900 3 200 Hz pulse Vertical Rear A A PASS Normal Operation

SPOT 3 1kHz 80%AM Vertical Rear A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Vertical Rear A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Vertical Right A A PASS Normal Operation

SPOT 3 1kHz 80%AM Vertical Rear A A PASS Normal Operation

900 3 200 Hz pulse Vertical Right A A PASS Normal Operation

900 3 200 Hz pulse Vertical Front A A PASS Normal Operation

SPOT 3 1kHz 80%AM Vertical Front A A PASS Normal Operation

80 to 1000 3 1kHz 80%AM Vertical Front A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Horizontal Front A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Horizontal Right A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Horizontal Rear A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Horizontal Left A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Vertical Left A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Vertical Rear A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Vertical Right A A PASS Normal Operation

1 to 2.5GHz 3 1kHz 80%AM Vertical Front A A PASS Normal Operation

NOTE: Performance Criteria categories A, B, and C are as defined in EN 60601-1-2.

The Performance Criterion for passing is as shown in TABLE II of this report.

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 19 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

6.0 APPENDIX C: PRODUCT INFORMATION FORM

CRITERION TECHNOLOGY PRODUCT INFORMATION FORM General Information Date: 2/12/08 Company Name: Zynex Medical, Inc. Company Address: 8022 Southpark Circle, Unit 100 Littleton, CO 80120 Contacts: Compliance Engineer: Jim Arnold Phone: 303-707-0203 Email: [email protected] Design Engineer: Alex Russo Phone: 303-707-0203 Email: [email protected] Test Description De-Bug Formal (initial) X Formal (re-verification) Market Information (check all that apply) USA X Canada X Euro. Union X Taiwan Japan New Zealand X Australia X Other Product Information Name Neuro-Move Model Number NM-900 Serial Number 2690979 and 2791123 Product Dimensions: 7.9” x 6.8” x 3.0” Weight: 22 oz. Product Power Source: Battery Type Lithium ion battery pack, 7.4 V nominal, two, 3.7 V cells connected in series (2S1P) Redundant Power Supplies N/A AC Supply Input Voltage Range(s) 100-240 VAC Phases 1 Delta N/A Wye N/A Current 31 mA AC max. Frequency 50 – 60 Hz Manufacturer Globtek Model Number GTM21089-1512 -T2 Topology Linear Switching Mode X Switching Frequency 100 kHz Support Equipment (if used): I/O Cables – Manufacturer, P/N, Length: Serial Port N/A Parallel Port N/A SCSI Port N/A Other EMG Cable, Zynex Medical, P/N 104148 Operation Software: Name Firmware Version Number 3.52-147 Operating Modes: (please include cycle time) 1. Normal (EMG triggered stimulation). Cycle time: 0.5 sec. 2. Stimulation only. Cycle time: 0.1 sec. Time necessary for EUT to be exercised and able to fully respond: 1 second. Operation Pass/Fail Criteria: Any interruption that does not result in a hard component failure and can be corrected by cycling the power is considered passable.

Sheet 20 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

Emissions Testing (please check all that apply): Information Technology Equipment Class A Class B Oscillator/Clock Frequencies (MHz) Industrial, Scientific, Medical Equipment Class A Class B X Oscillator/Clock Frequencies (MHz) 3.2768 MHz Unintentional Radiator Class A Class B Oscillator/Clock Frequencies (MHz) Receiver Type (Regen., Superhet., Direct Conv., Homodyne) Local Oscillator Frequencies Frequency Range Intentional Radiator Fundamental Frequency Range Local Oscillator Frequencies Power Output (to antenna) Integral Antenna (Yes/No) Modulation Type (AM, CM, Pulse, Spread Spectrum) Control Circuits (Microprocessor/Micro-controller) Oscillator/Clock Frequencies (MHz) Immunity Testing Test Type (please check all that apply): EN 61000-6-2 – Electromagnetic Compatibility- Part 6-2. Generic Standard-Immunity for Industrial

Environments EN 55024 – Information Technology Equipment – Immunity Characteristics - Limits and Methods

of Measurement Requirements X EN 60601-1-2 – Medical Electrical Equipment, Collateral Standard: EMC Requirements and Test EN 61326 - Electrical Equipment for Measurement, Control and Laboratory Use – EMC Requirements X EN 61000-4-2 (ESD) Number of metallic test points touchable by equipment operator: 1 Number of non-metallic test points touchable by equipment operator: 5 Is the product enclosure completely plastic? No Is the product enclosure partly plastic? Yes Are there any additional ESD voltages required for testing? If so, list herein: X EN 61000-4-3 & ENV 50204 (Radiated Susceptibility Testing, 80 - 1000 MHz) What is the maximum time necessary for the product to respond? 1 sec. During normal operations, what parameter will be monitored to determine susceptibility of the product? Stimulation output.

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 21 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

7.0 APPENDIX D: TEST EQUIPMENT AND CALIBRATION STATUS Manufacturer Name/Description Model Number Serial Number Cal. Due Date

Haefely Trench EFT Tester PEFT Junior 583-333-51 2/10/2008 Haefely Trench Surge Coupler FP-Surge 32.1 083-925-05 2/10/2008 Haefely Trench Surge Generator PSURGE 6.1 083-906-07 2/10/2008 Haefely Trench Interrupter tester Pline 1610 083-970-07 2/10/2008 Haefely Trench ESD Gun PESD 1600 H605100 3/25/2008 Hewlett Packard Spectrum Analyzer Display HP 85662A 2403A07322 4/4/2008 Hewlett Packard Quasi Peak Adapter 85650A 2403A07322 4/4/2008

Hewlett Packard Spectrum Analyzer HP 8566B 2421A00527 4/4/2008

Amplifier Research E-Field Probe FP2000 19682 5/13/2008

Amplifier Research E-Field Probe FP2080 20236 6/19/2008 Antenna Research 1-18 GHz Horn DRG118/A 1056 6/19/2008 EMCO Active Loop 6502 2626 6/19/2008 Rohde/ Schwarz VHF/UHF Receiver ESVS-30 863342014 7/8/2008

Rohde/ Schwarz LISN ESH2-Z5 828739-001 7/8/2008 Rohde/ Schwarz HF Receiver ESHS-30 826003/011 7/8/2008 Veratech Preamp (AMP2) unknown N/A 7/18/2008 EMCO biconnical antenna 3108 9103-2441 7/22/2008 EMCO log periodic antenna 3146 9004-2763 7/23/2008 Chase Bilog 30 - 1000 MHz CB6111 1121 7/23/2008 FCC CDN FCC-801-M3-25 9714 10/8/2008 FCC EM Clamp F2031 309 10/8/2008 Amplifier Research Directional Coupler DC2600 302981 10/8/2008 Amplifier Research Power Amplifier 100W1000M1 20214 11/4/2008 Amplifier Research Power Amplifier 150A100A 20183 11/19/2008 Tegam Current Probe 925236-1 12588 11/19/2008 Haefely Trench Test Mag Mag 100 80162 12/12/2008 Hewlett Packard Tracking Generator HP85645A 3210A00124 1/8/2009 Califorina Instruments AC Power Source Pacs-1 5001iX-CTS-411 55637/ 72242 2/21/2009

EMCO Horn 3160-08 1147 11/26/2009 Microwave Technologies Standard Gain Horn & Harmonic Mixer 12A-18 & HP1197OK 19527JE & 2332A01314 11/26/2009 Hewlett Packard Signal Generator HP 8648D 3642000145 1/7/2010

Sheet 22 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

8.0 APPENDIX E: TEST DIRECTIVES, STANDARDS AND METHODS 8.1.1 EUROPEAN DIRECTIVES, STANDARDS AND METHODS

89/336/EEC: Council Directive of 03 May 1989 on the Approximation of the Laws of the Member States Relating to Electromagnetic Compatibility, OJEC No. L 139/19-26, Aug 1993.

BS DD ENV 50204 (CENELEC): Testing and Measurement Techniques; Radiated Electromagnetic Field from Digital Radio Telephones - Immunity Test, 1996.

EN 55011 (CENELEC): ISM Radio-Frequency Equipment Radio Disturbance Characteristics - Limits and Methods of Measurement, 2007.

EN 55014-1 (CENELEC): Part 1. Electromagnetic Compatibility Requirements for Household Appliances, Electric Tools and Similar Apparatus - Part 1. Emission - Product Family Standard, 2006.

EN 55022 (CENELEC): ITE - Radio-Frequency Equipment Radio Disturbance Characteristics - Limits and Methods of Measurement, 2006.

EN 55024 (CENELEC): ITE - Immunity Characteristics - Limits and Methods of Measurement, 2003. EN 55103-1: Product Family standard for audio, video, audio - visual and entertainment lighting control apparatus for professional use. Part 1: Emissions, April 1997. EN 55103-2: Product Family standard for audio, video, audio - visual and entertainment lighting control apparatus for professional use. Part 2: Immunity, April 1997.

EN 60601-1-2 (CENELEC): Medical Electrical Equipment. Part 1. General Requirements for Safety - Section 1.2. Collateral Standard: Electromagnetic Compatibility - Requirements and Tests, A1:2006 , A2: 2007. EN 61000-6-1: EMC- Part 6-1. Generic Standard-Immunity for residential, commercial and light-industrial Environments 2007. EN 61000-6-2: EMC- Part 6-2. Generic Standard-Immunity for Industrial Environments, 2005.

EN 61000-6-3: EMC- Part 6-3. Generic Standard-Emissions for residential, commercial and light-industrial Environments 2007.

EN61000-6-4 (CENELEC): EMC - Generic Emission Standard, Part 6-4: Industrial Environment, 2007.

EN 61000-3-2 (CENELEC): EMC - Part 2. Limits for Harmonic Current Emissions (Equipment Input Current ≤16 A per phase), with Amendment 14, 2006.

EN 61000-3-3 (CENELEC): EMC - Part 3. Limitation of Voltage Fluctuation and Flicker in Low-Voltage Supply Systems for Equipment with Rated Current ≤16 A, 1998, A1:2001, A2:2005, A3:2006. EN 61000-4-7 (CENELEC): EMC – Part 4-7 Testing and measurement techniques – General guide on harmonics and interharmonics measurements and instrumentation, for power supply systems and equipment connected thereto: 2002, incorporating corrigenda Nos. 1:2004 and 2:2005.

EN 61000-4-2 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 2. Electrostatic Discharge Immunity Test, with Amendments 1 & 2, 2001.

EN 61000-4-3 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 3. Radiated, Radio-Frequency, Electromagnetic Field Immunity, 2006.

EN 61000-4-4 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 4. Electrical Fast Transient/Burst Immunity Test, incorporating corrigendum no. 1: January 2007.

CRITERION TECHNOLOGY EMC QUALIFICATION TEST REPORT Sheet 23 of 24 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

EN 61000-4-5 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 5. Surge Immunity Test, 2006. EN 61000-4-6 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 6. Immunity to Conducted Disturbances, Induced by Radio-Frequency Fields, 2005, A1: 2007.

EN 61000-4-8 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 8. Power Frequency Magnetic Field Immunity Test, 1993 with the incorporation of amendment A1:2001.

EN 61000-4-11 (CENELEC): EMC - Part 4. Testing and Measurement Techniques; Section 11. Voltage Dips, Short Interruptions and Voltage Variations Immunity Tests, 2004

EN 61326 (CENELEC): Electrical Equipment for Measurement, Control and Laboratory Use - EMC Requirements, 1997, with the incorporation of amendments A1:1998, A2:2001 and A3:2003.

8.1.2 47 CFR FCC PART 15 RADIO FREQUENCY DEVICES: OCT 2006

Subpart A General.

Subpart B Unintentional Radiators.

Subpart C Intentional Radiators.

Subpart D Unlicensed Personal Communications Service Devices. 8.1.3 47 CFR FCC PART 22 PUBLIC MOBILE SERVICES: OCT 2006 8.1.4 47 CFR FCC PART 24 PERSONAL COMMUNICATIONS SERVICES: OCT 2006 8.1.5 JAPAN

VCCI V-3 8.1.6 CANADA

ICES-001: Interference-Causing Equipment Standard - ISM RF Generators, 2006.

ICES-003: Interference-Causing Equipment Standard - Digital Apparatus, 2004. 8.1.7 AUSTRALIA/NEW ZEALAND

SAA AS/NZ 3548: Limits and Methods of Measurement of Radio Disturbance Characteristics of ITE, 1997. AS/NZS CISPR22

8.1.8 CHINA

CNS13438, 1997.

Sheet 24 of 24 EMC QUALIFICATION TEST REPORT CRITERION TECHNOLOGY 071010-1176 FOR ZYNEX MEDICAL, INC.

CONFIDENTIAL AND PROPRIETARY

THIS PAGE INTENTIONALLY LEFT BLANK