Eleva Workspot 1.x - Training User Interface & Workflow State: April 2007

Transcript of Eleva Workspot 1.x - Training User Interface & Workflo€¦ · – stitching with EVRAD –...

  • Eleva Workspot 1.x - Training

    User Interface & Workflow

    State: April 2007

  • Purpose of this presentationPurpose of this presentation

    • Understand the single workspot user interface and workflow

    • Not included:Not included:– multi-workspot– multi-reader

    networked solutions– networked solutions– stitching with EVRAD– advanced user functions, settings, QA ...

    • Just make X-Rays and process them with Eleva Workspot ....

    2Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • ContentContent

    • Overview user interface

    • General properties

    • Overview system configuration tools

    • Standard workflow video (Essenta DR)Standard workflow video (Essenta DR)

    • Move to position video (Essenta DR)

    • Workflow support details

    3Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Eleva Workspot has simply 5(4) +1 screens = task areasEleva Workspot has simply 5(4) 1 screens task areas

    (option)

    4Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva

    ELEVA WORKFLOW

    (Option)

    5Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Patient List TabPATIENT LIST• Patient folder status.

    Patient folder ready for:Patient folder ready for::Examination:Review (once completed):Print (If outstanding print jobs):Print (If outstanding print jobs)

    • Worklist filter “Completed Today”

    • Detector/System status:Workspot errors:Workspot errors:System detector status

    -Date of calibration due:Reader status (PCR Option)

    6Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    :Reader status (PCR Option)

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Examination TabIMAGE DISPLAY• Two step process

    :First step: Sub-sampled Pre Image:Second step: Fully Processed Image

    • Blue progress bar to display status of image• Sub-sampled Pre-Imagep g

    :Whole detector is displayed:Time to display approx 4 seconds

    • Fully Processed Imagey g:Final Image processing –what you see is what you get!:Image is zoomed “Fit to Shutter”g(system configuration) or “WholeDetector“:Time to Display pre image to fully

    7Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    p y p g yprocessed 6 seconds

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Examination TabIMAGE DISPLAY• All displayed images can be configured

    via EVA in regards to:via EVA in regards to::Collimation – Rectangular or

    Polygon:Image Rotation:Image Rotation:Automatic Markers and AutomaticAnnotation:Fit to Shutter or Fit to Detector (System tab UI):Image Processing Protocol:Image Processing Protocol

    Please refer to

    8Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    the EVA section for full details

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Examination TabIMAGE DISPLAY AND EI_s• Goals for having an EI_s:

    ClinicalClinical1. Did the detector get sufficient “energy” for a good image quality ?2. Did I overexpose this image ?3 Did I underexpose this image ?3. Did I underexpose this image ?4. If the exposure was wrong, by how far ?5. What is the correct exposure for this anatomy ?

    Technical1. Constancy check.

    Assuming the X ray source works correctly:Assuming the X-ray source works correctly:Does the imaging system response properly to a standard exposure

    Assuming the imaging system works correctly:

    9Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    Assuming the imaging system works correctly:Does the X-Ray source generate a proper standard exposure ?

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Examination TabIMAGE DISPLAY AND EI_s• The EI_s gives an indication of the image

    receptor dose right in front of the detector p gor cassette

    • Under calibrated conditions:

    70kV, 21mmAl phantom

    EI_s = Dose@ROI / µGy * 100

    e g EI s = 250 for a detector dose of 2 5µGye.g. EI_s = 250 for a detector dose of 2.5µGy

    And is calculated for the ROI = Region of Interest (“green pixels”)

    10Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot ElevaCALCULATION OF EI_s• Ranger determines “Region of Interest”Ranger determines Region of Interest

    as the anatomy, where the highest contrast (film curve) should be.

    • This Region is highlighted in theThis Region is highlighted in the Ranger panel by green pixels(switch on button to see).

    • These green pixels have all (almost)These green pixels have all (almost) the same raw data value.

    • From calibration measurement in thefactory Philips knows the necessary detectorfactory, Philips knows the necessary detectordose to obtain this raw data pixel value.

    • EI_s is the inverse calculation from the “green pixels” into detector dose forthe green pixels into detector dose for these pixels (air kerma /µGy *100). (provided calibration beam qualityand some other factors see next slides)

    EI_s = detector dose / µGy * 100(for calibration beam quality)

    11Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    and some other factors, see next slides)

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Examination TabEI_s TARGET VALUES

    • FOR BOTH DR AND CR:

    EI_s = 100 to 300Where the relationship is PROPORTIONAL to the dose

    • This is across all body areas. Chest X-Rays can go to 450 due to the “Auto Mode”

    If the number on the image is higher, you can reduce your dose givenIf the number on the image is lower, you have to increase your dose used

    and you should check that the quality of the image is good enough for diagnosis.

    • Note: for an 800 speed system, the EI_s target values are 50 to 150

    12Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    respectively

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Examination TabIMAGE DISPLAY AND EI_s• If the EI_s value is not within the expected range the following

    conditions have to be checked:conditions have to be checked:

    1/ Is the diagnostic relavant body part collimated correctly?2/ Is there too much direct radiation within the field?2/ Is there too much direct radiation within the field?3/ Is the “green snow” that defines the ranger within the diagnostic relevant body part?4/ Was an AEC used and was the AEC positioned correctly?

    • Note: 2x detector dose means 2x higher EI s valueNote: 2x detector dose means 2x higher EI_s value:The exposure index may vary by up to ±50% even for correct exposure due to patient variations

    13Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Exposure Index Scalesinverse proportional

    p

    800800 16001600 11001100 1.31.31.25 µGy1.25 µGy 190 1.31.6125

    400400 800800 14001400 1.61.62.5 µGy2.5 µGy 380 1.61.9250

    200200 400400 17001700 1.91.95 µGy5 µGy 1700760 1.92.2500

    EI

    100100

    S

    200200 20002000

    EI

    2.22.2

    lgM

    10 µGy10 µGy 20001520

    EXI

    2.22.5

    lgM

    1000

    EI s EI(Philips)

    S(Fuji)

    EI(Kodak)

    lgM(Agfa)

    SC= 200

    EXI(Siemens)

    lgM(Agfa)

    SC= 400

    EI_s(Philips)

    14Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    IEC Standard

  • Select a patient in Patient List

    15Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Patientl tiselection

    remainsover allother task areas

    PerformcompletecompleteexaminationinExaminationtask area.

    16Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Review andE i tiExaminationare similar

    In Review,you work only withonly withacquiredimages,while inwhile in Examinationyou acquirethem

    17Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Print isth t kthe taskarea whereyou create manualprintouts

    18Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • iSite (option)is the taskarea with access to thePhilips (Stentor)

    1.2.Philips (Stentor)

    iSite EnterpriseWeb Viewer

    Attention:3 steps1. new login2. Different Patient context.

    Patient has to bePatient has to be reselected(will be enhancedin near future )

    3.

    19Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    in near future...)3. PACS image viewing

  • POST PROCESSINGPOST PROCESSING

    20Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Post ProcessingBASIC IMAGE PROCESSING

    Contrast & BrightnessContrast & Brightness

    Image Rotation

    Annotation

    Collimation

    Zoom & PanZoom & Pan

    21Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Post ProcessingPOST PROCESSING

    • Can be performed in both the Examination tab and the Review tabCan be performed in both the Examination tab and the Review tab• Basic and Advanced Post Processing functions• Displayed icons can be configured.

    :Free Rotation:Simple Ranger Tool:Simple Ranger Tool:Full Screen (Review tab only)

    22Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Post Processing

    One Degree RotationOne Degree Rotation

    23Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Post Processing

    Simple Ranger ToolSimple Ranger Tool

    Can be used instead of contrast/brightnesscontrast/brightness

    Manual adjustment of OROI.

    Corresponding change of image impressionEI_s value will change

    24Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Essenta DR Workspot ElevaEssenta DR Workspot Eleva Post ProcessingCONFIGURATION OF POST PROCESSING ICONS.

    • In both the Examination and the Review Tab, the post processing iconsIn both the Examination and the Review Tab, the post processing icons can be configured to the customers requirements.

    • Can only be performed by Application Specialist or Advanced UserConfiguration can only be performed when an image is available in the• Configuration can only be performed when an image is available in the Examination Tab

    • Select the post processing iconSelect the post processing icon • Click on Customise• Click on the icon, then click on thePost Processing Panel to placePost Processing Panel to place function

    • To remove a function, select the “blank” icon then click on the function

    25Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    blank icon then click on the functionyou want to remove

  • System isf lfor generalsystem functions,settings etc.

    26Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • General properties of Eleva WorkspotGeneral properties of Eleva Workspot• Advanced user concept

    Normal user as many as you like– Normal user as many as you like1 factory user “user", password “user"

    Advanced user "eleva" factory password "eleva"

    normalnormal

    advanced useradvanced user

    ii– Advanced user eleva , factory password eleva – Service user password cannot be changed– Demo user "demo", pw "eleva4demo"

    needed to activate the demo mode

    serviceservice

    needed to activate the demo mode can be deleted and re-created

    • Operating system hardening– Windows XP embedded based– Application starts directly after PC startup

    and PC shuts down after application exitand PC shuts down after application exit– Operating system (desktop, explorer) not accessible by anybody!

    > no manual file editing or exchange– Configuration backup restore import export etc only via

    27Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    Configuration, backup, restore, import, export etc. only via provided Eleva service tools (FS dongle may be required)

  • Eleva Workspot has 3 tools for customization &

    normalnormal advanced useradvanced user serviceservice

    Eleva Workspot has 3 tools for customization & configuration

    • System– InformationInformation– Queues– Quality Assurance– SettingsSettings ...

    • EVA tool (see RF/RC.. Eleva)– Advanced user only

    A t i l d t badvanced useradvanced user serviceservice

    – Anatomical database– Processing protocol usage– Auto-export

    A t i t start in– Auto-print– RIS Code mapping ...

    • FSF (Field Service Framework)

    sta tSystem / Settings / EVA

    ( )– only with service "dongle"– System parameters– DICOM configuration

    FieldService

    Framework

    serviceservice

    28Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    g– Backup & restore ...

    Framework start inSystem / General

  • Get in touch:Get in touch:Geometry (DR Essenta)

    Please, click on image to show video

    Move to position (I) Move to position (II)

    Changing SID Move up and downwards

    29Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    C a g g S Move up and downwards

  • Workflow and Functions in detailWorkflow and Functions in detail

    ... this is not described here.

    Please look into the Instruction for Use.Please look into the Instruction for Use.

    30Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Workflow support:Workflow support:Confirm, reject, complete & auto-export and print• Eleva Workspot supports workflow using different states for

    image/exam/patient• Here: viewHere: view

    Marked for auto-export

    Marked for auto-

    Image is confirmed

    print or selected in manual print

    I iImage is rejected

    31Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Confirm or reject a view triggers auto-export/printConfirm or reject a view triggers auto export/print"Complete" = Implicit confirm of all views

    Auto-export

    Auto-print

    Explicitconfirm

    Auto print

    Auto-export

    Auto-print

    Reject ? ... depends:Icons may betaken awaytaken away

    32Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

  • Autofunctions triggered by confirm / completeAutofunctions triggered by confirm / complete

    On complete, confirm or reject is triggered:(depending on configuration and manual overwrite)

    • Auto exportAuto export– image is sent to all DICOM export nodes configured for the export

    scheme in for this examination programmed in EVA• Auto print• Auto print

    – image is sent to a print job, using template and position as programmed for this examination / view in EVA

    – print job is printed on a printer and medium as configured in the print medium mapping in

    – confirm/reject: if this was the last image for one page, the page is j g p g , p gsent to the printer

    – complete: even incomplete pages are sent to printer after sending all images to the print jobs as programmed

    33Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    all images to the print-jobs as programmed

  • Complete a patient (study examination)Complete a patient (study, examination)With the user completes one or more examinations. Why ?

    to confirm the unconfirmed images altogether– to confirm the unconfirmed images altogether– to finally start printing of uncomplete films– to send the "MPPS completed" message to the RIS– to let the affected examinations get a proper status, so in theto let the affected examinations get a proper status, so in the

    patient list, the patient (study) entry gets the proper state and appears in the proper worklist(s)

    If there is more than one examination for one patient the system asksIf there is more than one examination for one patient, the system asks "Complete all or just the selected examination ?"

    – this question is about "RIS-examinations", that means:– the "selected examination" may be more than just the selected blue single y j g

    entry in the examination list, but all user interface examinations that are booked based on the the same SPS from RIS, and are hence reported in the same MPPSthi i t d li itl l b th i t f t t f th– this is not made explicitly clear by the user interface not to confuse the majority of users, where it is irrelevant, but it may confuse somebody ...

    You can "discontinue" an examination with proper MPPS message.S ki

    34Company confidential Eleva Workspot – Training Material – T.Dorn - April 2007

    See networking ppt.