EFSUMB – European Course Book Editor: Christoph F. Dietrich
Nephrology, Transplantation and Shunts
FM Drudi, F Malpassini, N Di Leo
Corresponding Author Prof. F. M. Drudi [email protected]
Ultrasound nephrology transplantation …. 28.07.2010 18:16 2
Content
Content ....................................................................................................................................... 2 Introduction ................................................................................................................................ 2 Evaluation of kidney transplants using gray-scale sonography and color Doppler ................... 2 The usefulness of the resistive index.......................................................................................... 4 Arteriovenous fistula as a complication of percutaneous graft biopsy ...................................... 6 US contrast agents...................................................................................................................... 6 Magnetic resonance imaging...................................................................................................... 7 The use of ultrasound in the assessment of arteriovenous fistulae for hemodialysis access ..... 8 References ................................................................................................................................ 11
Introduction
Renal transplantation is an established cost-effective treatment in patients with end-stage renal
disease [1]. However, after the first year, graft survival curves show an exponential decline in
numbers of functioning grafts. Many causes of renal graft dysfunction are treatable, thus
making prompt detection and diagnosis of complications mandatory. Acute tubular necrosis
(ATN) may cause immediate oliguria and may follow an initial short period of graft function
as well as acute rejection. Finally, there is a possibility of cyclosporine or tacrolimus toxicity.
Evaluation of kidney transplants using gray-scale sonography and color
Doppler
Renal graft measurements (length, width) and anatomical characteristics (corticomedullary
differentiation, the existence of hydronephrosis affecting the graft, perinephric fluid
collections or masses) [Mov.1] as well as vascular flow [3,4] [Mov.2, Fig.1] can be defined
using gray-scale sonography and color Doppler.
Ultrasound nephrology transplantation …. 28.07.2010 18:16 3
Figure 1 Color Doppler in a normal patient shows blood flow and vascular
morphology (renal artery and renal vein are both visible)
During the first 24 hours after surgery, the graft must therefore be explored using gray-scale
and color Doppler as the first imaging technique according to the protocol or clinical
indication. Absence of perfusion in the graft is a sign of renal artery occlusion. Reverse
diastolic flow in the arteries due to retrograde blood flow during diastole [Fig.2] is a sign of
complete allograft vein thrombosis or acute rejection. Incomplete vein thrombosis is more
difficult to identify. When color Doppler cannot completely rule out arterial occlusion,
stenosis or vein thombosis, CT scans or even arteriography is required [7].
Figure 2 Acute rejection. Typical Color Doppler aspect. The flow under the baseline
indicates the retrograde arterial blood flow during diastole.
Ultrasound nephrology transplantation …. 28.07.2010 18:16 4
Infarctions can be diagnosed at Doppler ultrasound or power Doppler examination by
demonstrating a lack of blood flow to the infarcted region of the parenchyma. In these cases,
exploration using ultrasound contrast agent is the best option.
Sonographic exploration cannot differentiate between ATN and acute rejection, so ultrasound
guided allograft biopsy is required.
Renal artery stenosis occurs at a rate of 1% to 10%, usually at the site of the surgical
anastomosis. Diagnosis is made by demonstration of a focal and segmental region of flow
abnormality characterized by elevated peak systolic velocity (PSV; normal value 250 cm/s)
with associated turbulence at an adequate insonation angle. The ratio of renal artery PSV
compared to that of the iliac artery can be more useful, because PSV may be variable in the
allograft artery. In addition to this, tardus-parvus waveform abnormalities can be observed in
the renal parenchyma. After sonographic diagnosis of possible renal artery stenosis, magnetic
resonance imaging or computed tomography angiography may be indicated to confirm the
diagnosis before percutaneous transluminal angioplasty is performed.
The usefulness of the resistive index
Resistive index (RI) is usually determined as a standard practice in clinical monitoring. The
RI value depends on the graft vessels, but it is possible that it is even more influenced by the
recipient's vessels and their elasticity. An isolated elevated RI has limited value and is
nonspecific. In the postoperative days, elevated RI values (0.9) [Fig. 3] can be found in
several types of graft dysfunction such as acute rejection, calcineurin inhibitor toxicity, severe
ATN [Fig.4], renal vein obstruction, uretheral obstruction and pyelonephritis. Periodic RI
measurement is therefore useful in patients affected by these complications to help monitor
the graft function [8] [Fig.5].
Ultrasound nephrology transplantation …. 28.07.2010 18:16 5
Figure 3 RI results 0.90; this is a value which might indicate acute pathology
Figure 4 Sample volume used to calculate Resistive Index is positioned at the level of
the interlobar artery.
Ultrasound nephrology transplantation …. 28.07.2010 18:16 6
Figure 5 Sample volume used to calculate Resistive Index is positioned at the level of
the arcuate artery.
In chronic allograft function, RI has limited value, because it is not a particularly sensitive
marker of chronic graft pathology and cannot be used in clinical decision-making.
However, it has been shown that an elevated RI value (0.8), measured 3 months
postoperatively, is a predictor of subsequent poor graft function and/or death [9].
Arteriovenous fistula as a complication of percutaneous graft biopsy
Arteriovenous fistula is a possible complication after percutaneous graft biopsy. Doppler
ultrasound depicts arteriovenous fistulae as a localized area of disorganized color that extends
outside the confines of the normal vessel caused by perifistula aliasing [12]. Waveform
analysis demonstrates a high-velocity, low-resistance flow in the supplying artery and a high
pulsatility flow in the draining vein (arterialization).
US contrast agents
Recently, the introduction of second generation contrast agents has given new possibilities
and perspectives to US imaging and quantification of renal blood flow as well as
microvascular tissue perfusion [Mov.3]. CEUS has overcome the limitations of color Doppler
ultrasonography and permits depiction of parenchymal blood perfusion.
CEUS provides the best visualization of perfusion deficit indicating its extension and
allowing the characterization of indeterminate renal lesions, atypical cystic lesions, and the
identification of acute pyelonephritis [17] [Mov.4, Fig.6a, 6b].
Ultrasound nephrology transplantation …. 28.07.2010 18:16 7
Some authors have concluded that renal perfusion patterns of normal and abnormal tissue can
be visualized using contrast-enhanced phase-inversion ultrasound imaging.
New diagnostic possibilities of CEUS include evaluation of both cortical and medullar vessels
as well as functional evaluation of renal perfusion. Measuring the microbubble transit time is
useful in the diagnosis of renal artery stenosis and in the differential diagnosis between ATN
and acute rejection in transplanted kidneys [13].
Magnetic resonance imaging
Magnetic resonance imaging (MRI) has emerged as an alternative imaging modality in renal
transplant graft assessment [Fig.6c]. MRI is promising due to its multiplanar capabilities and
lack of ionizing radiation, invasiveness, and contrast medium-induced nephrotoxicity. Recent
studies have shown the utility of MRI in the evaluation of the renal graft and peritransplant
region, but its role is not yet firmly established [5]. At MRI, loss of corticomedullary
differentiation on T1-weighted images is the most useful finding indicating rejection.
Figure 6 In the upper part of this image, CEUS clearly shows the parenchymal region
with a perfusion deficit (a). Transplanted patient with infarction in the
central portion of the kidney; CEUS shows a parenchymal perfusion defect
(b). MRI using contrast agent confirms the presence of a hypovascular area
(c).
a
b
Ultrasound nephrology transplantation …. 28.07.2010 18:16 8
c
The use of ultrasound in the assessment of arteriovenous fistulae for
hemodialysis access
Color Doppler ultrasonography has proved to be effective in the evaluation of arteriovenous
fistulae for hemodialysis access both in the preoperative phase, for the assessment of
anatomical vascular features, and in the postoperative phase, to detect clinically presumed
arteriovenous fistula complications [Fig.7].
Ultrasound nephrology transplantation …. 28.07.2010 18:16 9
Figure 7 An arteriovenous fistula for hemodialysis access assessed using color Doppler
ultrasonography.
Color Doppler ultrasonographic vein mapping allows the identification of veins that are not
clinically evident. The inability of clinical examinatio8 alone to predict adequacy of venous
outflow causes the use of suboptimal veins and arteries, and this results in a high failure rate
of arteriovenous fistulae and in an increasing use of arteriovenous grafts [11].
Selection criteria for arteriovenous fistula insertion should be the following [12, 13]:
1. Cephalic or basilic veins measuring at least 2 mm in diameter in continuity with the deep
system.
2. An appropriate artery measuring at least 2 mm in diameter with peak systolic velocity of
50 cm/s or greater.
3. Patent ipsilateral subclavian vein.
The arteriovenous fistula complications that occur most frequently are thrombosis-correlated
stenosis and aneurysm [14, 15, 16].
Hemodynamical data with higher diagnostic significance are the mean flow volume (1204
ml/min), the mean maximum velocity in anastomosis (2.7 m/s), and the mean maximum
velocity in the brachial artery (1.35 m/s) [15].
Turbulent blood flow causes extensive vessel wall and perivascular tissue vibration. This
localized tissue vibration causes artifactual color assignment of the perivascular soft tissues,
which precludes adequate visualization of the venous anastomosis [14] [Fig.8].
Ultrasound nephrology transplantation …. 28.07.2010 18:16 10
Figure 8 Tissue vibration due to turbulent blood flow causes artifactual color
assignment to the perivascular soft tissues.
Diagnostic criteria indicating stenosis at color Doppler ultrasonography are visible narrowing
of the lumen, increased velocity of flow greater than 100% compared with that of an adjacent
normal segment in color Doppler study [17, 18].
Arteriovenous fistula waveform is arterial with systolic velocities of 100-400 cm/sec [19] but
demonstrates high end-diastolic flow because of the low-resistance runoff in the draining
veins [20].
An increase of mean maximum velocity in anastomosis (4.35 m/s) indicates narrowing of the
vessels with the degree of stenosis related to the increase in velocity [20, 21]. On the contrary,
both the mean flow volume and the mean maximum velocity in the brachial artery are found
to be lower in fistulae with venous stenosis [15].
In more than 3% of hemodialysis vascular accesses a false aneurysm is present, and in more
than 50% of cases there is an aneurysm (vein diameter >6 mm). Almost 100% of these
alterations are located at the puncture site [15].
Available data indicate that the mean flow volume is significantly higher in the fistulae with
aneurysms. A strong correlation has been found between aneurysm and calcifications and
aneurysm and fistula age [15].
Increased flow rates, more than 1500 ml/min, are associated with steal syndrome or venous
hypertension.
The Doppler finding noted in steal syndrome was marked reduction of flow in the distal
artery, while high flow rates reaching 600 ml/min in the distal venous limb of the
arteriovenous fistula are associated with venous hypertension [14].
Ultrasound nephrology transplantation …. 28.07.2010 18:16 11
References
1) Baxter GM. Imaging in renal transplantation. Ultrasound Q 2003;19:123-138
2) Schubert RA, Gockeritz S, Mentzel HJ, et al. Imaging in ureteral complications of renal
transplantation: value of static fluid MR urography. Eur Radiol 2000;10:1152-1157
3) Tublin ME, Dodd GD III. Sonography of renal transplantation. Radiol Clin North Am
1995;33:447-459
4) Howard RJ, Patton PR, Reed AI et al. The changing causes of graft loss and after kidney
transplantation. Transplation 2002; 73: 1923–1928.
5) Radermacher J, Mengel M, Ellis S et al. The renal arterial resistive index and renal
allograft survival. N Engl J Med 2003; 349:115–124.
6) F.M.Drudi, R. Pretagostini, S.Padula et al. Color Doppler Ultrasound in renal transplant:
role of Resistive Index versus Renal Cortical Ratio in the evaluation of renal transplant
diseases Nephron Clin Pract 2004; 98: c67-c72
7) Young LS, Regan MC, Barry MK, Geraghty JG et al. Methods of renal blood flow
measurement. Urol Res 1996; 24: 149–160.
8) Schwenger V, Korosoglou G,Hinkel UP, et al. Real-time contrast-enhanced sonography of
renal transplant recipients predicts chronic allograft nephropathy. Am J Transplant. 2006
Mar;6(3):609-15
9) Correas JM, ClaudonM, et al The kidney: imaging with microbubble contrast agents.
Ultrasound Q 2006;22(1):53-66
10) Neimatallah MA, Dong Q, Schoenberg SO, et al. Magnetic resonance imaging in renal
transplantation. J Magn Reson Imaging 1999;10:357-368.
11) Silva MB Jr, Hobson RW 2nd, Pappas PJ, et al. A strategy for increasing use of
autogenous hemodialysis access procedures: impact of preoperative noninvasive evaluation. J
Vasc Surg 1998; 27:302 – 308.
12) Sedlacek M, Teodorescu V, Falk A, et al. Hemodialysis access placement with
preoperative noninvasive vascular mapping: comparison between patients with and without
diabetes. Am J Kidney Dis 2001; 38:560 – 564.
13) Vassalotti JA, Falk A, Cohl ED, et al. Obese and non-obese hemodialysis patients have a
similar prevalence of functioning arteriovenous fistula using pre-operative vein mapping.
Clinical Nephrol 2002; 58:211 – 214.
14) Moghazy KM. Value of color Doppler sonography in the assessment of hemodialysis
access dysfunction Saudi J Kidney Dis. Transpl. 2009 Jan;20(1):35-43
Ultrasound nephrology transplantation …. 28.07.2010 18:16 12
15) Radoslaw P, Marzena J, Wojciech Z, et al. Doppler ultrasound assessment of well-
functioning mature arteriovenous fistulae for haemodialysis access Eur J Radiol. 2005
Jul;55(1):113-9. Epub 2004 Nov 19
16) Rodrigo P. Campos, Marcelo M. Do Nascimento, Domingos C. Stenosis in hemodialysis
arteriovenous fistula: Evaluation and treatment. Hemodial. Int. 2006; 10:152–161
17 Weber M,Kuhn FP,Quintes W, et al. Sonography of arteriovenous fistulae in hemodialysis
patients.Clin Nephrol 1984;22(5):258-61
18) Dousset V, Grenier N, Douws C, et al. Hemodialysis grafts: Color Doppler flow imaging
correlated with digital subtraction angiography and functional status. Radiology
1991;181(1):89-94
19) Finlay DE, Longley DG, Foshager MC, et al. Duplex and color Doppler sonography of
hemodialysis arteriovenous fistulas and grafts. RadioGraphics 1993; 13:983-999.
20) Elcheroth J, de Pauw L, Kinnaert P, et al. Elbow arteriovenous fistulae for chronic
hemodialysis. Br J Surg 1994;81(7):982-4
21) Polak JF. Peripheral arterial disease evaluation with color flow and Doppler sonography.
Radiol Clin North Am 1995; 33:71-90.