Electronic Fetal Monitoring Dr. Augustinus BT SpOG

description

journal of obgyn

Transcript of EFM

Page 1: EFM

Electronic Fetal Monitoring

Dr. Augustinus BT SpOG

Page 2: EFM

HISTORY - EFM 1818-Francios Major in Geneva-DDx between FH and

Maternal Pulse 1827- John C Ferguson –described FHR sounds. 1849-Killian indicated FHr parameters requiring interventions. 1876-Pinard produced his design for a fetal stethoscope. 1893-Winkel set normal FHR120-169 bpm. 1958-Hon in USA and Hammacher in Europe introduced first

EFM. 1964- Doppler ultrasound scan replaced phonocord. 1966- Saling in Berlin introduced FBS. 1968-Hamacher and Hewitt-Packard developed first fetal

monitor. 1985- Dublin RCT changed terminology for the CTG

interpretation.

Page 3: EFM

Electronic Fetal Monitoring Pioneered in 1958 by Hon.in USA and

Hammacher in Europe Commercially available 1968 Continues monitoring of FH in low risks

pregnant -lacks evidence Needs to be viewed in conjunction with

other assessment e.g. FBS and clinical situation.

Should be used for the right reason and with appropriate degree of skill.

Page 4: EFM

EFM-ISSUES

Detect fetal hypoxia i.e reduce and avoid harm to the fetus and improve fetal and baby out-come.

Severe acidosis may result in FHR changes. Could occur in Normal physiological response in

labor. Misunderstanding the physiological and

pathphysiological CTGs will improve the Mx.

Page 5: EFM

EFM Problems and Realities

Electronic Intra-partum FHR Monitoring is now considered mandatory for high-risk pregnancies.

Difficulties with interpretation include over confidence and not-only difference in opinion between practitioners but, also when the same practitioner examines the same CTG twice.

Increases CS rates 1.41%rr. Increases operative vaginal delivery 1.20%rr. And no change in incidence of C Palsy. Reduction in Neonatal seizures rates 0.51% No difference in APGAR scores. ? About the efficacy.

Page 6: EFM

EFM- Facts

Reliability of interpretation-50-75% are false positive .

False positive Dx reduces to 105 with FBS. FBS 93% sensitivity, 6% false positive. PH Vs Lactate -39% Vs 2.3(rr 16.7).

Page 7: EFM

Electronic Fetal Monitoring-Indications

Indications for the continuous EFM

High risk pregnancies

IOL and Augmentation of Labour.

Reduced FM. Premature

labour/TPL. APH/IPH

Oligohydramnios Hypertension. Abnormal FHR

detected. Malpresentation and in

labour. DM,Multiple Gestation. Previous CS. Abdominal Trauma. Prolonged ROM. Meconium Liq.

Page 8: EFM

EFM- Interpretation

Consider : Intrapartum/antepartum trace. Stage of labour. Gestation. Fetal presentation, ?Malpresentation. Any augmentation,?IOL Medications Direct or indirect monitoring/

Page 9: EFM

EFM- 4 Basic Features of FH Trace

Page 10: EFM

EFM-4 Basic Features.

Baseline FHR - Mean level of FHR when this is stable, excluding Accelerations and Decelerations (110-160 bpm)-Tachycardia-Bradycardia

Baseline Variability-5 bpm or greater than or equal to 5bpm, between contractions-Normal-Non-reassuring-Less than 5 bpm or less but less than 30 min-Abnormal-less than 5 bpm for 90 min or more.

Page 11: EFM

Baseline variability CTG

Baseline variability

Page 12: EFM

Baseline variability

The minor fluctuations on baseline FHR at 3-5 cycles p/m produces Baseline variability.

Examine imin segment and estimate highest peak and lowest trough.

Normal is more than or equal to 5 bpm.

Page 13: EFM

Factors affecting Baseline variability.

Para-Sympathetic affects short term variability whilst Long Term is more Symp.

CNS ,Drugs reduce Variability High gestation increases variability Mild Hypoxia may cause both S and

para S stimulation.

Page 14: EFM

Non-reassuring Baseline variability.

NRCTGs- reduced or less than 5 bpm for 40 min or more but less than 90 mins..

B-B or short Term V is varying intervals between successive heart beats .

Long Term v is irregular waves on the CTG 3-5 bpm.

Normal is 5-25 bpm– this indicates N-CNS.

Page 15: EFM

EFM-Accelerations

Accelerations- transient increase in FHR of 15 bpm or more lasting for 15 sec.

Absence of accelerations on an otherwise normal CTG remains un clear.

Presence of FHR Accelerations have Good outcome.

Page 16: EFM

EFM Decelerations

Decelerations- transient slowing of FHR below the baseline level of more than 15 bpm and lasting for 15 sec. Or more.

Page 17: EFM

Electronic Fetal Monitoring

a) Early Decelerations (fig 3) Head compression Begins on the onset of contraction

and returns to baseline as the contraction ends.

Should not be disregarded if they appear early in labor or Antenatal.

Clinical situation should be r/v

Page 18: EFM

Fig 3 Early Decelerations

Page 19: EFM

Late Decelerations.

Uniform periodic slowing of FHR with the on set of the contractions .

Repetitive late decels increases risk of Umbilical artery acidosis and Apgar score of less than 7 at 5 mins and Increased risk of CP.

Page 20: EFM

Electronic Fetal Monitoring

b) Late Decelerations (Fig 4)• Due to acute and chronic feto-placental

vascular insufficiency Occurs after the peak and past the length of uterine

contraction, often with slow return to the baseline. Are precipitated by hypoxemia Associated with respiratory and metabolic acidosis Common in patients with PIH, DM, IUGR or other

form of placental insufficiency.

Page 21: EFM

Fig 4 Late Decelerations

Page 22: EFM

Late Decelerations

Reduces Baseline variability together with Late Decelerations or Variable Decelerations is associated with increased risk of CP.

Page 23: EFM

EFM- Variable Decelerations

Variable intermittent periodic slowing of FHR with rapid onset recovery and isolation.

They can resemble other types of deceleration in timing and shape.

Atypical VD are associated with an increased risk of umbilical artery acidosis and Apgar score less than 7 at 5 min

Additional components: Loss of 1 degree or 2 degree rise in baseline Rate Slow return to baseline FHR after and end of contraction. Prolonged secondary rise in Base FHR Biphasic deceleration Loss of variability during deceleration Continuation of base line at a lower level.

Page 24: EFM

Electronic Fetal Monitoring

c) Variable Deceleration (Vagal activity) (Fig 5) Inconsistent in configuration, No uniform temporal r-ship to the onset of contraction, are

variable and occur in isolation. Worrisome when Rule of 60 is exceeded (i.e. decrease of

60 bpm,or rate of 60 bpm and longer than 60 sec) Caused by cord compression of the umbilical cord Often associated with Oligo-hydroaminos with or without

ROM Can cause short lived RDS if they MILD Acidosis if prolonged and Recurrent.

Page 25: EFM

Fig 5 Variable Decelerations

Page 26: EFM

EFM Prolonged deceleration

Prolonged Deceleration (Fig 6) Drop in FHR of 30 bpm or More lasting for at

least 2 min Is pathological when crosses 2 contractions i.e 3

mins. Reduction in O2 transfer to placenta. Associated with poor neonatal outcome.

Page 27: EFM

EFM- Prolonged DecelerationsCAUSES

Cord prolapse. Maternal hypertension Uterine Hypertonia Followed by a VE or ARM or SROM

with High PP.

Page 28: EFM

Fig 6 Prolonged Deceleration

Page 29: EFM

EFM Mx Prolonged Deceleration

Maternal position IV fluids V.E to exclude cord prolapse Assess BP FBS if cx dilated and well applied PP Mx Depending on the clinical situation.

Page 30: EFM

Baseline Bradycardia

FH below 110bpm(FIGO ). less than 100bpm (RANZCOG).Causes.Postdates, Drugs, Idiopathic,Arrythmias, hypothermia(increased Vagal

Tone)Cord Compression (Acute Hypoxia, congenital

H/disease and Drugs).Mx depends on the clinical situation.(FBS,VE

Observation or expedite delivery)

Page 31: EFM

Types

Moderate Bradycardia 100-109 bpm Abnormal bradycardia less than

100bpm. Tachycardia 161-180 bpm Abnormal Tachycardia more than 180

bpm Ranzcog Australian more than 170

bpm

Page 32: EFM

Baseline tachycardia and Bradycardia.

Uncomplicated baseline tachycardia 161-180 bpm or bradycardia 101-109 do not appear to be associated with poor NN outcome.

Page 33: EFM

Causes of B Tachycardia.

Asphyxia Drugs Prematurity Maternal Fever Maternal thyrotoxicosis Maternal Anxiety Idiopathy Mx depends on the clinical situation

Page 34: EFM

Electronic Fetal Monitoring

Baseline Bradycardia FH Rate below 110bpm (FIGO Recommended) Postdates Drugs Idiopathic Arrhythmia's Hypothermia.(Increased Vagal tone), Cord compression(Acute Hypoxia,Congenital

H/disease, and drugs)Mx depends on the clinical situation. (FBS, VE, Observation or expedite Delivery).

Page 35: EFM

Electronic Fetal Monitoring

Baseline Tachycardia Asphyxia Drugs Prematurity Maternal fever Maternal thyrotoxicosis Maternal Anxiety IdiopathyMx depends on the clinical situation

Page 36: EFM

Fig 2 Sinusoidal pattern

Interpretation of the CTG

Page 37: EFM

EFM-Sinusoidal Pattern

Regular Oscillation of the Baseline long-term Variability resembling a Sine wave ,with no B-b Variability (Fig 2),

Has fixed cycle of 3-5 p min. with amplitude of 5-15 bpm and above but not below the baseline.

Should be viewed with suspicion as poor outcome has been seen (eg Feto-maternal haemorrhage)

Page 38: EFM

Electronic Fetal Monitoring

Sinusoidal pattern - distinctive smooth undulating Sine-wave baseline with no B-b variability ( Fig 2 ) 0.3 % (Young 1980) cord compression hypovolemia ascites idiopathic(fetal thumb sucking) Analgesics Anaemia Abruption Mx r/v clinical situation

Page 39: EFM

EFM- Saltatory pattern

Seen During Fetal thumb sucking. Could be associated with Hypoxia. See slide 11.

Page 40: EFM

NR CTGs

Difficult to interpretation,leads to Increased rate of C Section.

50% CTG in Labour have 1 abnormal feature

15-20% Nr CTGs (pathological). ?? To reduce CS….

Page 41: EFM

F O R E M O S T Feta oximetry randomised evaluation Multi Oxygen saturation Trial

To evaluate the role of F Sats P02 Had a Multicentre Randomised trial- RWB Brisbane MMH Brisbane RWH Melbourne RWH Sydney

Page 42: EFM

Endpoints

To reduce Operative deliveries for the Nr CTG by 1/3 The Sample size was 600 participants. Secondary Endpoint- Investigate the incidence of Dystocia F/up outcomes Low Umbilical arterial PH NICu admissions HIE (Enchelopathy) Maternal perception of monitoring Clinician perception of sensor placement.

Page 43: EFM

NR CTG (Foremost)

Any of the following for more than 15min-Persistent Late decelerationsSinusoidal patternVariable decelsLess than 70 bpm for more than 60 secPersistent slow return to the baselineLong term variability less than 5 bpmTachycardia more than 160 bpm.Recurrent prolonged deceleration 2 or more in15 min and less than

70bpm for 90 sec .Any on of the following for more than 60 minTachycardia with variability less than 5bpmPersistent reduced baseline variability less than 5 bpm for more

than 60min.

Page 44: EFM

Inclusion Criteria

NR CTG Consent Gest age more than 36 k Early or active labour ROM or eligible for

Page 45: EFM

Exclusion criteria

Multiple gestation Non-Vertex Pl Praevia APH Fetal anomaly HIV other Thrombocytopenia.

Page 46: EFM

Clinical Mx Protocol

CTG Only Reassuring –continue labour unless

otherwise indicated NR CTG –Evaluate and manage NR

CTG Suspicious CTG Deliver for NR Fetal Status.

Page 47: EFM

CTG and Oximetry

Reassuring CTG Continue NR CTG Continue unless otherwise

indicated( FSp02 Reassuring) If FSp02 less than 30%- Further Evaluation and MX

Suspicious CTG -Deliver

Page 48: EFM

Conclusion

R/v NR CTG Mx tools & decision making

Responsibility to assess new technology before widespread introduction

Overview of Multicentre RCT Case presentation.

Page 49: EFM

Reliability of CGT Interpretation

50%-75% false positive False Positive Dx is reduced to 10%

with FBS PH Vs lactate 39% Vs 2.3%(rr16.79).

Page 50: EFM

Fetal Blood Sampling

information on the acid base balance Has 6%-20% false- normal 8-10% False Low valuesWhy and when to do-Persistent Abnormal CTG after reversible factors have been

corrected, Persistent late decels and 2 abnormal other features e.g baseline tachycardia or reduced B-B variability or just difficult to interpret the CTG

Rom, PP accessible and well applied Cx dilatation at least 3 cm or more

Left lateral maternal positionSterile environment and good light and equipment.Good analgesia

Page 51: EFM

FBS-PH

Introduced by Sailing –fairly reliable predictor of fetal PH in presence of abnormal FHR

Has an error of 15% falsely low and 5 % falsely increased, however in presence of abnormal FHR a scalp PH of less 7.2 is pathologonomic of fetal asphyxia.

Page 52: EFM

FBS-Cord PH

FBS-Arterial Normal-7.25-7.35 Normally good agreement

between Apgar,cord PH . Less than 7.20-significant asphyxia Values between 7.2 and 7.24 need further

evaluation Low –normal PH should be repeated in 30

min Less than 7.20- eminent delivery

Page 53: EFM

FBS- Lactate

FBS easier to interpret, difficult to perform Anaerobic metabolism can lead to metabolic

acidosis Lactate is major end product of metabolic

acidosis Lactate levels more specific for degree of

metabolic acidosis than Ph Lactate rises quicker takes longer to resolve

than Ph.

Page 54: EFM

FBS ContraindicationsFetalPremature –less than 34 ksActive HerpesKnown HIV,Hep B,C positive status.Thrombocytopenia.Maternal-Unfavourable CxMobile PPMalpresentation(face etc) uncertain??Pl Praevia or APHSepsis

Page 55: EFM

FBS-Sampling errors

Between decelerations if possible Avoid Excess pressure on PP reduces

perfusion Do not sample on the caput. Failure of scalp to bleed –due to

peripheral shut down.

Page 56: EFM

EFM-Summary

Normal - CTG with all 4 Features Suspicious -one non reassuring category

and reminder are reassuring Pathological -2 or more non-reassuring

categories or one or more abnormal categories.

Page 57: EFM

At Birth

Need to ConsiderCord PH if CTG suspicious Preterm labourMec LiquorFBS intrapartumFlat baby at deliveryOperative or instrumental delivery.

Page 58: EFM

Mx of FBS Results

If PH /Lactate normal –Observe Pre-acidotic were CTG remains to be

Suspicious –Rpt after 30 min. Acidotic-eminent delivery Instrumental or CS depending on the clinical

situation.Lactate 4.8 upper limit-average 2.8 in 1st stage–

changes in labour

Page 59: EFM

Electronic Fetal Monitoring

Education and training Improves Knowledge for all staff Improves clinical skills Training should include instruction on

documenting traces and storage Training should include appropriate clinical

responses to suspicious or pathological traces Training should include local guidelines

relating to fetal monitoring both intermittent and EFM

Page 60: EFM

EFM-Legal Issues

CTG in Litigation Unsatisfactory or Missing Abnormal CTG ignored or not recognized Traces not done.Risk Mx EFM traces should be kept up to 25 years. If removed for teaching purposes or etc,

should be easily located Minimize incidence of adverse outcome

Page 61: EFM

Legal Issues

In the U.K 70% of cases are due to EFM. In U S A 45.8% involve E F M, 3.8% more $1Million payouts and

only17% are due to delay or failure to diagnose fetal distress.

Page 62: EFM

What Influences Litigation

Consumer Expectation The profession –education The employer (policies/procedures) Legislation (duty of care/scope of

practise/ registration)

Page 63: EFM

Legal issues- Consumer expectation

Are for a good outcome (healthy baby/mother)

Bad outcome Someone to blame Someone must pay

Page 64: EFM

Professional Responsibility

To act within scope of practise. To seek support and guidance Work within organisational standards Duty of care to the woman and

employer Maintain knowledge and skills

(evidence Based Practise) Be prepared to defend ones practise.

Page 65: EFM

When EFM is the focus of Malpractice

Comparison of consistency of documentation contained on the trace and in the chart will be made

Lapse in documentation may leave doubt about the quality of care.

Hospital policy and procedure manuals will be examined

Competency levels will be evaluated expert witness (plaintiff/defence)-to determine if acceptable standards were applied.

Page 66: EFM

Major Omission in Liability

Failure to appropriately monitor the mothers and fetus status

In appropriate Syntocinon monitoring Failure to notify the physician in a

timely manner. Initiation of procedures without

adequate client information or consent

Page 67: EFM

Elements of successful Malpractice Action

The clinician had a duty to the woman The clinician had committed a breach

of duty The woman suffered damages There was a casual connection

between the clinician’s actions and the woman's damage.

(MacRae,1993)

Page 68: EFM

Legal issues

If you are going to use the CTG You must be able to Interpret the trace and respond accordingly-

It’s a permanent record One that is scrutinized in litigation case May be pivotal in determining liability

Page 69: EFM

Legal issues

A normal CTG can be used to indicate that that were no abnormalities and no indication for intervention.

An abnormal CTG or suspicious trace may provide evidence that inappropriate or lack of treatment may give rise to litigation,whilst CTGs could be viewed as part of “defensive medicine”, litigation is reported to be on the increase but there is not enough data reporting system in Australia to support this.

Page 70: EFM

References

Manual Obs and Gyn. by Niswander, MD

Fetal Monitoring RCOG UK CTGs RANZCOG Literature review articles American

Family Physician CTG Made Easy