EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO...

12
EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted in Odel using Dynalizer High Voltage Divider, Dynalizer Digital Display and Endeavour’s Computer signal kV feedback. The test focuses the linearity of kV range. Test has been conducted every 5 kV in the interval 40kV..145kV using the 100mA, 80ms loads Dynalizer Digital Display Setup: - trigger @ 75% of peak kV - measure blind window set to 10ms Oscilloscope Setup: Channel 1 = High Voltage Divider output (A+C) 1V=20kV Channel 2 = MCU feedback test point (A+C) 100mV = 40V MEASUREs: TEK MAX for both CH1 and CH2 (equivalent to kV peak measurement) Measures has been recorded and analyzed using MS Excel in order to trace a scatter graph, compute the linearization on the curve and check the angular coefficient of the intercepta: theoretical value = 1 Than has been computer the mean quadratic root to have an idea how how the real curve is far from its intercepta. Bothe the curve and the R2 is directly displayed on the graphs.

Transcript of EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO...

Page 1: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

1

di 

 

Test has been conducted in Odel using Dynalizer High Voltage Divider, Dynalizer Digital Display and Endeavour’s Computer signal kV feedback.

The test focuses the linearity of kV range.

Test has been conducted every 5 kV in the interval 40kV..145kV using the 100mA, 80ms loads

Dynalizer Digital Display Setup:

- trigger @ 75% of peak kV

- measure blind window set to 10ms

Oscilloscope Setup:

‐ Channel 1 = High Voltage Divider output (A+C) 1V=20kV ‐ Channel 2 = MCU feedback test point (A+C) 100mV = 40V

MEASUREs: TEK MAX for both CH1 and CH2 (equivalent to kV peak measurement)

Measures has been recorded and analyzed using MS Excel in order to trace a scatter graph, compute the linearization on the curve and check the angular coefficient of the intercepta: theoretical value = 1

Than has been computer the mean quadratic root to have an idea how how the real curve is far from its intercepta. Bothe the curve and the R2 is directly displayed on the graphs.

Page 2: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

2

di 

 

Analysis between Control Console display and Dynalyzer Digital Display measurements

kV Console kV Dynalizer DD40 40,8145 45,3450 50,2155 55,9060 61,6865 65,5070 71,4675 76,3780 81,4685 86,0090 91,3795 94,81100 100,60105 104,70110 111,20115 115,80120 120,90125 125,80130 131,00135 135,90140 140,70145 144,90

Page 3: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

3

di 

 

Analysis between theoretical Dynalizer Voltage Divider (1V/20kV) and Tektronix MAX measurement on Dynalizer’s voltage divider signal

Expected kV signal kV Dynalizer2 2,00

2,25 2,282,5 2,522,75 2,80

3 3,083,25 3,323,5 3,563,75 3,80

4 4,084,25 4,324,5 4,564,75 4,84

5 5,045,25 5,285,5 5,605,75 5,84

6 6,086,25 6,366,5 6,606,75 6,88

7 7,087,25 7,28

Page 4: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

4

di 

 

Analysis between theoretical Feedback signal and Tektronix MAX measurement on generator’s kV feedback signal

Expected Feedback Measured Feedback in MAX m100,00 108,00112,50 120,00125,00 140,00137,50 148,00150,00 160,00162,50 172,00175,00 188,00187,50 200,00200,00 208,00212,50 224,00225,00 236,00237,50 252,00250,00 260,00262,50 272,00275,00 288,00287,50 300,00300,00 313,00312,50 328,00325,00 340,00337,50 352,00350,00 364,00362,50 372,00

Follows test data

Page 5: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

5

di 

 

40kV

45kV

50kV

Page 6: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

6

di 

 

55kV

60kV

65kV

Page 7: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

7

di 

 

70kV

75kV

80kV

Page 8: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

8

di 

 

85kV

90kV

95kV

Page 9: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

9

di 

 

100kV

105kV

110kV

Page 10: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

10

di 

 

115kV

120kV

125kV

Page 11: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

11

di 

 

130kV

135kV

140kV

Page 12: EEN-037 V1 R1 kV Linearity test · EEN-037 kV Linearity test V1 R1 2011-07-08 DOCUMENTO FOGLIO NOMEFILE CLASSE EEN‐037 V1 R1 kV Linearity test.doc 13 1 di Test has been conducted

 

EEN-037 kV Linearity test

V1 R1

2011-07-08

 

DOCUMENTO FOGLIO

NOMEFILE CLASSE

EEN‐037 V1 R1 kV Linearity test.doc  13 

12

di 

 

145kV