ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

55
ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE DR SASINTHAR SR, CARDIOLOGY JIPMER

Transcript of ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Page 1: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

DR SASINTHARSR, CARDIOLOGY

JIPMER

Page 2: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

VARIOUS ECHO MODALITIES

• TTE• TEE• 3D ECHO• ICE

Page 3: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

WHEN TO SUSPECT IN 2D ECHO

•RIGHT VENTRICULAR DILATIONABNORMAL MOTION OF IVS- brisk anterior movement in early systole or flattened movement throughout systole•? IAS DROP OUT IN APICAL 4C VIEW•RELATIVE ATRIAL INDEX

Page 4: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

2D ECHO

RA RV VOLUME OVERLOAD

SEPTAL FLATTENING IN DIASTOLE

Page 5: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

The Relative Atrial Index (RAI)—A Novel, Simple, Reliable, and Robust Transthoracic Echocardiographic Indicator of Atrial Defects

Cutoff value of >0.92 predicted patients with ASDs v/s matched controls with 99.1% sensitivity and 90.5% specificity

Natalie A Kelly -Journal of the American Society of EchocardiographyVolume 23, Issue 3 , Pages 275-281, March 2010

Page 6: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

SUB COSTAL 4C VIEW• Keeps the atrial septum perpendicular to the ultrasound beam• Distinguishes OS , OP & SV ASDs• SV ASD are consistently visualised in the SUBCOSTAL 4C VIEW• Anomalous drainage of pulmonary veins • Atrial septal aneurysm• Viewed with breath held in inspiration- index marker in 3o` clock

position

SUB COSTAL SHORT AXIS

• Index marker at 12o`clock position and sweeping the transducer from midline to Rt side of patient

Page 7: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

SUBCOSTAL 4C VIEW

SUB COSTAL SHORT AXIS VIEWALSO SHOWS IVC DRAINING TO RA AND EUSTACHIAN VALVE

Page 8: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Other TTE -views for ASD

• PSAX-IAS separates Rt &Lt atrium and runs posteriorly from NCC of aortic valve. Not seen in entirety as a result of drop out artefact

• APICAL 4C- Posterior aspect of Interatrial septum is clearly delineated in this view but drop out artefact is seen in region of fossa ovalis.

• Pulmonary venous drainage- 3 veins draining to LA• APICAL 5C VIEW- Anterior aspect of interatrial

septum

Page 9: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

PSAX VIEW IAS AGAINST NCC OF AORTA

APICAL 4C VIEW SHOWING THE IAS AND 3 VEINS DRAINING TO LA, RT LOWER PULMONARY VEIN IS USUALLY NOT SEEN

Page 10: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 11: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

ANOMALOUS PULMONARY VEIN

• Can be associated with ASD or can occur as an isolated anomaly• 95% of SV ASD a/w RUPV-SVC • RUPV-SVC; LUPV- innominate vein ;

RLPV- IVC• Isolated LLPV – extremely rare

Page 12: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

`Crab view` showing absent Rt upper and Rt lower pulmonary vein

Page 13: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

En face view in 2D

• First the apical 4c view was taken. The image index marker was at approximately kept at 1 o'clock. Keeping the atrial septum and ASD in the region of interest, the transducer was rotated counterclockwise approximately 45° to 60°

Xinseng et al Journal of the American Society of Echocardiography Volume 23, Issue 7 , Pages 714-721, July 2010

Page 14: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

A-4c view & B-En face view

Page 15: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Morphological variations1.MC- Deficient aortic rim (42.1%)

2.Central defects (24.2%) 3.Deficient Inferoposterior rim (12.1%)

4.Perforated aneurysm of the septum (7.9%) 5.Multiple defects (7.3%)

6.Combined deficiency of mitral and aortic rims (4.1%), 7.Deficient SVC rim (1%),

8.Deficient coronary sinus rim (1%).

Podnar T, Martanovic P, Gavora P,Masura J. Morphological variations of secundum-type atrial septal defects: feasibility for percutaneous closure using Amplatzer septal occluders. Catheter Cardiovasc Interv 2001;53:386 –91.

Page 16: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 17: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

ATRIAL SEPTAL ANEURYSM

CRITERIA

A-PROTRUSION OF ANEURYSM ATLEAST15MM OF PLANE OF IAS OR IAS SHOWING15MM OF PHASIC EXCURSION DURINGCARDIORESPIRATORY CYCLEB- BASE WIDTH≥ 15MM

Page 18: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

TRANSVERSE PLANE VIEWS VERTICAL PLANE SWEEP

TEE

Page 19: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Centrally located ASD imaged at 0°

Page 20: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 21: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 22: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

ASD with deficient Aortic margin

Page 23: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 24: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Large ASD with deficient posterior and Aortic margins

Page 25: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Multiple ASDs; larger anterior defect (block arrow) and a smaller posterior defect

Page 26: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

RIMS OF ASD

Aortic - Superoanterior Atrioventricular (AV) valve -mitral or inferoanterior Superior Vena Caval SVC – Superoposterior Inferior venacaval (IVC or Inferoposterior) Posterior (from the posterior free wall of the atria).

Page 27: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

IVC AND SVC RIMS

Page 28: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

OTHER RIMS

TEE

TTE

Page 29: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Measurement of the ASD rims

• Atleast 5 mm• IVC rim-most important

Schematic representation of the locations of the ASD rims

Page 30: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

TEE 4-chamber view depicting an adequate posterior rim for percutaneous closure of

20 mm.

Transesophageal 4-chamber view: The AV rim measures 9.5 mm, which is

adequate for PCT

Page 31: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

TEE upper-esophageal 4-chamber view with rightward (clockwise) rotation of the probe revealing an adequate RUPV rim of 15 mm . Beside, Doppler color image shows in red the inflow of the RUPV (white arrow). Note the correct ECG timing of the measure at the end of the ventricular systole while the atrio-ventricular valves are still closed.

Page 32: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Mid-esophageal short axis view of the aortic rim at 56 degrees with an adequate aortic rim (11 mm) for percutaneous closure

Page 33: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

• Absent aortic rim makes the procedure more challenging but does not, preclude device closure of the defect

Page 34: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Mid-esophageal bi-caval view at 97 degrees, an adequate SVC rim is noted,

measuring 13 mm .

Mid-esophageal bi-caval view at 97 degrees with an adequate IVC rim

of 10 mm

Page 35: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Special tee views for Inferoposterior rims

No Infero posterior rim with probe in normal position

Page 36: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Catheter Closure of Atrial Septal Defects With Deficient IVC Rim Under TEE GuidanceK.S. Remadevi, MD, FNB, Edwin Francis, DM, and Raman Krishna Kumar, DM, FACC . Catheterization and Cardiovascular Interventions (2008)

Retroflexed probe in the stomach and bought towards the esophagus and viewedIn the 70-90o view

Page 37: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

3D ECHO

• Matrix transducers – pyramid shaped volumes• Full volume 3D dataset in 4-7 cardiac cycles• Ideal window is the mid esophageal basal long

axis (bicaval view)• Subcostal 4c view- enface septum• Low parasternal 4c view case of suboptimal

windows• 3D tee overcomes 3D TTE if suboptimal windows

Page 38: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

• Real-time 3D imaging demonstrates the changing shape of the ASD during a cardiac cycle, with maximum size in diastole

• As we take the Bicaval view structures – we first remove the right atrial free wall .

• Images are taken with suspended respiration and ECG gating with optimal gain settings

• Low gain – drop outs and high gain – blurring of structural details

Page 39: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Gain settingsFor Best view

Cropping toGet the IAS

Page 40: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

TUPLE (TILT UP & LEFT)-ENFACE VIEW OF IAS FROM LT ATRIAL PERSPECTIVE

Page 41: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

RIMS OF ASD

ASD IN VARIOUS PHASES OF CARDIAC CYCLE

Page 42: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

ATRIAL SEPTUM ANEURYSM WITH ASD

MULTIPLE ASDs

Page 43: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

DEFECT NEAR THE IVC

Page 44: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

3D echo- En face 3D reconstruction of a secundum ASD with a relatively deficient

IVC and posterior rim

Page 45: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

multiple ASDs with the thin atrial septum (*) separating the 2 defects

Page 46: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 47: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

The correlations between the ASD maximal diameter by RT-3DE and operation or balloon sizing were excellent (r > 0.95). All surrounding rims of the atrial septum could be assessed on 3D reconstruction; except for the aortic rim, a

cross-sectional reconstruction was created mimicking the transesophageal echocardiographic cross section (r > 0.92)

Page 48: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

• Maximal criteria for transcatheter closure with ASO device are

(1) ASD secundum with a maximum TEE diameter of 34 mm (2) rims, except the anterosuperior rim, of at least 5 mm, and (3) the dimensions of the total length of the atrial septum were not smaller than the left atrial disk of the chosen device

Page 49: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

Measurement of ASD size

• Maximal ASD diameter must be measured at the end of ventricular systole

• Atleast two orthogonal views

• SBP = Max in TEE + 4 to 6mm

Mid-esophageal 4-chamber view at 0 degree depicting an ostium secundum ASD with a maximal transverse diameter of 18 mm .

Mid-esophageal bi-caval view at 97 degrees showing an ASD with a maximal longitudinal diameter of 14 mm

Page 50: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 51: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 52: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 53: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE
Page 54: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

• Max size of device used -44 mm• Device embolisation in 3/169 patients • 2- deficient posterior rim and large size (38

mm, 35 mm) were the reasons for instability, • In the third patient, the complete absence of

aortic rim with malaligned septum made the procedure difficult

Page 55: ECHO ASSESSMENT OF ASD FOR DEVICE CLOSURE

CONCLUSION

• Proper case selection• It is important to have inferior and posterior

rims• An anterior rim is not as important as the

device will grasp the aorta• A superior rim is less important as the device

will grasp the SVC orifice