Early Results of a Highly Porous Titanium Cementless...

1
Discussion & Conclusions • The use of cementless TKAs has gained popularity due to the increased life expectancy in our patient population 1 . • Cementless TKAs have not been adopted as readily as THAs due to the challenges and uncertainty of tibial fixation and past failures 2 . • With the advent of new technologies, the results of cementless TKAs with the potential for long term biologic fixation may now be equivalent or better than cemented TKAs 3,4 . • A novel highly porous tibial baseplate (Figure 1) was developed based on proximal tibial anatomy using CT scans using 3D printing technology with ideal length, location and design of press-fit pegs (Figure 2). • The purpose of this study was to review the early results with respect to fixation and complications using a new cementless tibial baseplate. • Study results indicate cementless fixation using this baseplate is a viable option with good short term clinical results and no cases of asceptic loosening at 2 years. • Early stability, pain relief and good ROM was shown. • Long term data will be required to determine the overall benefits of this highly porous TKA with biologic fixation versus cemented arthroplasty. Introduction Methods • 100 primary cementless TKAs were retrospectively reviewed (Stryker, Triathlon Tritanium baseplate, Peri- apetite Femoral Component and Porous Backed Tritanium Patella) • Patients were evaluated at 2 weeks, 3 months, 1 year, and 2 years for clinical and radiographic outcomes. • Radiographs were evaluated using zonal analysis for biological fixation and radiolucent lines. • Early (90 day and 2 year) complications were assessed. • The same pain management and rehabilitation protocol was assigned for all patients. Immediate full weight-bearing, as tolerated, was prescribed post-operatively. Results Early Results of a Highly Porous Titanium Cementless Tibial Baseplate in TKA Langan Smith 1 , Manoshi Bhowmik-Stoker 2 , Jeffrey Stimac 1 1. KentuckyOne Health, Louisville, KY, USA 2. Stryker, Mahwah, NJ, USA Fig 6. IntraOp photo of ideal rim contact. Fig 3. Pre-Op A/P and Lateral radiographs of a 67 year old male with OA. Fig 4. Radiographs at 1 year f/u. Fig 5. Radiographs at 2 year f/u. • 86 unilateral and 7 bilateral cases were enrolled in this study. Demographics and outcome scores were collected (Table 1). Mean patient follow up was 24.8 months (range 15 to 33) and mean LOS 3.2 days. Radiographic Results – Good short term fixation was noted. (Figure 3-5). Tibial coverage was maximized for all cases (Figure 6). Within 90 Day Adverse Events – • No cases of loosening, infections, or postoperative transfusions • 1 DVT, 1 PE non fatal • 1 revision for instability – CR femur converted to PS, tibial baseplate not revised Latest Follow-Up Adverse Events – • No revision of the tibial baseplate or femoral components. • 2 cases had revision of the liner only: Instability + extensor mechanism rupture: liner change + medial imbrication and lateral release (18 months post- op) MPFL rupture with subluxating patella: liner exchange with quad tendon repair (8 months post-op) Fig 1. Cementless Knee System References: 1. Kurtz et al. Clin Orthop Relat Res . 2009; 467(10):2606–2612. 2. Dodd et al. Clin Orthop Relat Res. 1990; 260:66–70. 3. Harwin et al. Orthopedics. October 2015. 38(10):620-624. 4. Bagsby et al. JOA. (2016) 1-5. Fig 2. Cruciform peg design. Demographics Age 63.8 (range 40-84) BMI (mean) 33.7 Sex 68 Female / 32 Male Knee Society Score Pre –Op (mean) 40 Post-Op (mean) 96 Knee Society ROM (Flexion / Extension) Pre-Op (mean) 105.8° / 1.9° Post-Op (mean) 117.9 / 0.1° Table 1 – Demographic and Clinical Results Fig 4. Fig 3. Fig 5. TRITAN-POS-16

Transcript of Early Results of a Highly Porous Titanium Cementless...

Page 1: Early Results of a Highly Porous Titanium Cementless ...az621074.vo.msecnd.net/syk-mobile-content-cdn/global-content... · Immediate full weight ... was prescribed post-operatively.

Discussion & Conclusions

• The use of cementless TKAs has gained popularity due to the increased life expectancy in our patient population1.

• Cementless TKAs have not been adopted as readily as THAs due to the challenges and uncertainty of tibial fixation and past failures2.

• With the advent of new technologies, the results of cementless TKAs with the potential for long term biologic fixation may now be equivalent or better than cemented TKAs3,4.

• A novel highly porous tibial baseplate (Figure 1) was developed based on proximal tibial anatomy using CT scans using 3D printing technology with ideal length, location and design of press-fit pegs (Figure 2).

• The purpose of this study was to review the early results with respect to fixation and complications using a new cementless tibial baseplate.

• Study results indicate cementless fixation using this baseplate is a viable option with good short term clinical results and no cases of asceptic loosening at 2 years.

• Early stability, pain relief and good ROM was shown. • Long term data will be required to determine the overall

benefits of this highly porous TKA with biologic fixation versus cemented arthroplasty.

Introduction

Methods • 100 primary cementless TKAs were retrospectively

reviewed (Stryker, Triathlon Tritanium baseplate, Peri-apetite Femoral Component and Porous Backed Tritanium Patella)

• Patients were evaluated at 2 weeks, 3 months, 1 year, and 2 years for clinical and radiographic outcomes.

• Radiographs were evaluated using zonal analysis for biological fixation and radiolucent lines.

• Early (90 day and 2 year) complications were assessed. • The same pain management and rehabilitation protocol was

assigned for all patients. Immediate full weight-bearing, as tolerated, was prescribed post-operatively.

Results

Early Results of a Highly Porous Titanium Cementless Tibial Baseplate in TKA

Langan Smith1, Manoshi Bhowmik-Stoker2, Jeffrey Stimac1

1. KentuckyOne Health, Louisville, KY, USA 2. Stryker, Mahwah, NJ, USA

Fig 6. IntraOp photo of ideal rim contact.

Fig 3. Pre-Op A/P and Lateral radiographs of a 67 year old male with OA.

Fig 4. Radiographs at 1 year f/u.

Fig 5. Radiographs at 2 year f/u.

• 86 unilateral and 7 bilateral cases were enrolled in this study. Demographics and outcome scores were collected (Table 1). Mean patient follow up was 24.8 months (range 15 to 33) and mean LOS 3.2 days.

• Radiographic Results – Good short term fixation was noted. (Figure 3-5). Tibial coverage was maximized for all cases (Figure 6).

• Within 90 Day Adverse Events – • No cases of loosening, infections, or postoperative

transfusions • 1 DVT, 1 PE non fatal • 1 revision for instability – CR femur converted to PS, tibial

baseplate not revised • Latest Follow-Up Adverse Events –

• No revision of the tibial baseplate or femoral components. • 2 cases had revision of the liner only:

• Instability + extensor mechanism rupture: liner change + medial imbrication and lateral release (18 months post-op)

• MPFL rupture with subluxating patella: liner exchange with quad tendon repair (8 months post-op)

Fig 1. Cementless Knee System

References: 1. Kurtz et al. Clin Orthop Relat Res . 2009; 467(10):2606–2612. 2. Dodd et al. Clin Orthop Relat Res. 1990; 260:66–70. 3. Harwin et al. Orthopedics. October 2015. 38(10):620-624. 4. Bagsby et al. JOA. (2016) 1-5.

Fig 2. Cruciform peg design.

Demographics Age 63.8 (range 40-84)

BMI (mean) 33.7 Sex 68 Female / 32 Male

Knee Society Score Pre –Op (mean) 40 Post-Op (mean) 96

Knee Society ROM (Flexion / Extension) Pre-Op (mean) 105.8° / 1.9° Post-Op (mean) 117.9 / 0.1°

Table 1 – Demographic and Clinical Results

Fig 4. Fig 3.

Fig 5.

TRITAN-POS-16