Dual Energy CT:Approaches and Applications
Dianna Cody, Ph.D.
Professor
Department of Imaging Physics
![Page 2: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/2.jpg)
Objectives• How does it work?
– Approach by vendor
• What is it good for?– Established applications
– Emerging applications
![Page 3: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/3.jpg)
Dual Energy CT• Recycled idea from 1980’s
• Bone QCT Densitometry
• Acquisition of attenuation data at two distinct energy levels
• Produce two sets of images
• Data processing for basis materials
![Page 4: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/4.jpg)
Material DecompositionInterleaved High- and
Low-kVp ProjectionsData Acquisition
Low kVp Projections High kVp Projections
80 kVp 140 kVp
split
Filtered Back Projection
![Page 5: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/5.jpg)
40keV –140keV
....)()()()()()()()()( 1
2
1
2
1111 iPiPiPiPiPiiPiiP highlowhighlowhighlow
....)()()()()()()()()( 2
2
2
2
2222 iPiPiPiPiPiiPiiP highlowhighlowhighlow
water
iodine
FBP
Attenuation-to-material density transformation
Monochromatic Generation
Iodine Density Water Density
70 keV
Iodine Projections Water Projections
![Page 6: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/6.jpg)
Blended Images
![Page 7: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/7.jpg)
![Page 8: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/8.jpg)
How to acquire data at two energies? Varies by vendor
• GE – FAST kVp switching
• Philips – Two layer detector
• Siemens – Two x-ray tubes
• Toshiba – Two sequential rotations
![Page 9: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/9.jpg)
Important Issues
• Temporal match of data acquired at separate kVs
• Adjust mA for kV? (Increase for low kVp, decrease for high kVp)
• Field of View limitations?
• Dose impact?
• Post-processing options
![Page 10: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/10.jpg)
GE Approach
• Newest scanner model only (HD750)
• Fast kVp switching: 140 kV & 80 kV
• Full field of view
• 1/3 projections obtained at 140 kV
• 2/3 projections obtained at 80 kV
• Excellent temporal matching
• “Tissue Signature” function
![Page 11: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/11.jpg)
GE Approach - Dose• Measured on HD750 June 2009
• 120kVp, 300mAs, 40mm beam– CTDIw 13.7 mGy
• 80 & 140 kVp, 600mAs, 40mm beam– CTDIw 31 mGy
• DE dose about 13% higher than 120kVp predicted by mAs change alone
![Page 12: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/12.jpg)
Dose Penalty?• YES – difficult to obtain dual energy CT
images at same dose as single average 120kVp scan
• BUT – if standard practice is C-/C+ scans, a single pass dual energy CT scan may provide virtual non-contrast images
• Small dose increase may justify the additional information obtained
![Page 13: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/13.jpg)
Philips Approach• Two-layer detector
• No extra x-ray tube technology
• Excellent temporal matching
• Full field of view
• Dose?
![Page 14: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/14.jpg)
Siemens Approach
• Available on Definition
• Field of view limited (depends on model)
• Two x-ray tubes operated simultaneously at different kVps
• Adjust mA value independently
• Temporal matching somewhat limited (90° out of sync)
• Cross-scatter
• Dose?
![Page 15: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/15.jpg)
Toshiba Approach• Dual rotations
• One at low kVp, next at high kVp
• Then move forward
• Limited to axial mode?
• Adjust mA for kVp?
• Temporal matching worst-case
• Full Field of view
• Dose?
![Page 16: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/16.jpg)
Established Applications
Kidney Stone Characterization
![Page 17: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/17.jpg)
Automated Bone Removal (Angiography)
![Page 18: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/18.jpg)
Potential Applications
Liver & Kidney Lesion Characterization
![Page 19: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/19.jpg)
75 keV - With Contrast
Images Courtesy of Dr. Amy Hara, Mayo Clinic, Scottsdale, AZ
Water Density(virtual non-contrast)
Iodine Density
![Page 20: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/20.jpg)
Liver Fat Quantification
![Page 21: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/21.jpg)
Virtual Non-Contrast Images
![Page 22: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/22.jpg)
Calcified Plaque Processing
![Page 23: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/23.jpg)
Pulmonary Embolism Detection
![Page 24: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/24.jpg)
Metal Artifact & Beam Hardening Reduction
80 kVp 80 keV
![Page 25: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/25.jpg)
Metal Artifact Reduction
140 kVp 70 keV w/MAR
![Page 26: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/26.jpg)
80kVp
140kVp
80keV
![Page 27: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/27.jpg)
More?• Improve detection of small pancreatic
lesions
• Discriminate between mucinous and nonmucinous pancreatic cystic lesions
• Characterize solid lesions (tumor vsfibrosis)
• Proton dose estimation in spine (fatty replacement of bone marrow)
![Page 28: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/28.jpg)
Tissue Signature
70 keV (mono)
m1 (water) m2 (iodine)
![Page 29: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/29.jpg)
Tissue Signature
• Optimize imaging parameters for new targeted tracers?
• Useful for discriminating solid lesions?
• Other applications yet to emerge?
![Page 30: Dual Energy CT: Approaches and Applicationschapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z... · 2012-05-03 · Dual Energy CT •Recycled idea from 1980’s •Bone QCT](https://reader034.fdocuments.us/reader034/viewer/2022042300/5ecaa3bcd9706f5db33b38bc/html5/thumbnails/30.jpg)
Summary
• Vendors have diverged in approach to dual energy CT
• Advantages & Disadvatages
• Established applications
• MANY emerging applications…!