    DRY HEAT STERILIZERS/OVENS

    USERS MANUAL

    NVE SANAY MALZEMELERMALAT VE TCARET A..

  • 8/2/2019 dry oven

    2/15

    2

    NVE SANAYI MALZEMELERIMALAT VE TICARET A..

    Esenboa Yolu 22 kmAkyurt - ANKARA / TURKEYTel : (00 90 312) 399 28 30 ( 3 lines )Fax : (00 90 312) 399 21 97e-mail: [email protected]

    WARRANTY CERTIFICATE

    1. Nve warrants that the equipment delivered is free from defects during material andworkmanship. This warranty is provided for a period of two years. The warranty periodbegins from the delivery date.

    2. Warranty does not apply to parts normally consumed during operation or generalmaintenance or any adjustments described in the operating instructions provided withthe equipment.

    3. Nve does not accept any liability in the case where the goods are not used in

    accordance with their proper intent.

    4. The warranty may not be claimed for damages occurred during the shipment, fordamages resulting from improper handling or use, the defects in maintenance,negligence, bad functioning of auxiliary equipment, in the case of force majeure oraccident and incorrect power supply.

    5. In the event of failure, Nve shall be under no liability for any injury, or any loss ordamage as the result of the failure other than the guarantee conditions.

    BEFORE OPERATING THE INSTRUMENT THIS MANUAL SHOULD BE READCAREFULLY.

    INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NVE. ITMAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT HIS PERMISSION.

    THE VALIDITY OF THE GUARANTEE IS SUBJECT TO THE OBSERVATION OF THEINSTRUCTIONS AND PRECAUTIONS DESCRIBED IN THIS MANUAL.

  • 8/2/2019 dry oven

    3/15

    3

    TABLE OF CONTENTS Page

    1. INTRODUCTION

    1.1 Use and Function

    4

    4

    2. TECHNICAL SPECIFICATIONS 5

    2.1. Technical Specifications Table 52.2. General Presentation 62.3. Optional Accessories 6

    3. INSTALLATION PROCEDURE 7

    3.1. Lifting and Transport 73.2. Unpacking 7

    3.3. Environmental Conditions3.4. Mains Supply3.5. Positioning3.6. Prior to Operation

    7778

    4. OPERATING PRINCPLES

    4.1. Switching On4.2. Control Panel4.3. Programming Summary4.4. Completion of Work

    5. PERIODIC MAINTENANCE AND CLEANING

    5.1. Periodic Maintenance5.2. Cleaning

    6. TROUBLESHOOTING

    8

    891011

    11

    1111

    11

    7. ELECTRICAL CIRCUIT DIAGRAMS 13

    7.1. FN 300 137.2. FN 400

    7.3. FN 5007.4. FN 400 P7.5. FN 500 P

    13

    141415

  • 8/2/2019 dry oven

    4/15

    4

    1. INTRODUCTION

    1.1 Use and Function

    The FN series dry heat sterilizers/ovens are designed to sterilize metal and glass

    materials and heat or dry samples in laboratories. They offer excellent sterilization, dryingand heating conditions. Glass, metal and other materials can be sterilized in FN seriesdry heat sterilizers.

    They maintain temperatures between ambient temperature + 5 C and 250 C and keepthe temperature stable within the given tolerances.

    FN 300/400/500 dry heat sterilizers/ovens provide homogeneous temperature distributionby means of the sheet heaters placed onto three outer surfaces of the useful volume.

    Differently from FN 300 / 400 and FN 500, inner sides of FN 400P and FN 500P are

    made of stainless steel and their heaters are circular heaters which provide thehomogeneity of the temperature by the circulation fans placed on their centers.

    The insulation of the dry heat sterilizers/ovens not only provides economical workingconditions but also helps to achieve the excellent homogeneity.

    The FN series dry heat sterilizers/ovens provide precise operation conditions with thePID microprocessor controlled system and with the timer. Adjustable safety thermostatoffers an additional safety.

    The FN series dry heat sterilizer/ovens are manufactured according to the followingstandards,

    EN 61010-1, EN 60601-1, EN 60601-1-2

    Do not operate the instrument for purposes other than main purpose.

    The instrument is only to be used by authorized people after the users manual hasbeen read carefully. Only technical personnel can handle the product in case of abreakdown.

  • 8/2/2019 dry oven

    5/15

    5

    2. TECHNICAL SPECIFICATIONS

    2.1. Technical Specifications Table

    FN 300 FN 400 FN 400 P FN 500 FN 500 PTemperature range Ambient temperature + 5C / 250CTemperature sensor Fe-ConstControl system PID Programmable MicroprocessorTemperature set & display sensitivity 1CTemperature variation(110C-190C) 2CTemperature fluctuation 1CTimer 1 minute + 99 hours 54 minutes+ Hold positionNumber of shelves (max/min) 2/6 2/7 2/7 2/10 2/10Safety thermostat Gas expansion thermostat (50C-300C)Power consumption 500 W 800 W 850 W 1500 W 1550 WPower Supply 230V, 50 /60 HzUseful volume 22 liters 44 liters 42 liters 120 liters 110 liters

    Internal materialElectro-acid

    coatedaluminum

    Electro-acidcoated

    aluminum

    StainlessSteel

    Electro-acidcoated

    aluminum

    StainlessSteel

    External material Epoxy-polyester painted steelInternal dim.s (wxdxh) mm 300x240x300 420x320x360 420x320x360 490x490x500 500x490x500External dim.s (wxdxh) mm 555x380x460 700x470x540 705x470x540 780x627x680 790x630x670Net/Packed weight 22/25 35/39 35/39 49/59 49/59Packing dims.(wxdxh)mm 670x480x510 795x565x605 810x550x615 870x770x720 890x720x730

  • 8/2/2019 dry oven

    6/15

    6

    2.2. General Presentation

    Inner side of FN 400 P / 500 PFigure 1

    1. Control panel 7. Mains supply inlet and the fuses2. Safety thermostat 8. Shelf carrier holder3. On / Off switch 9. Shelf carrier4. Ventilation hole5. Shelf6. Chamber gasket

    10. Circulation fan11. Heater12. Heater protecting sheet

    2.3 Optional Accessories

    - Shelves and shelf carriers

    1

    2

    3

    4

    5

    6

    7

    9

    8

    12

    11

    10

  • 8/2/2019 dry oven

    7/15

    7

    3. INSTALLATION PROCEDURE

    3.1. Lifting and Transport

    Because of the heavy weight of the dry heat sterilizer/oven, all lifting and transport must

    be carried out using proper handling equipment. The dry heat sterilizer/oven must besupported from underneath and never turned over.

    3.2. Unpacking

    Remove the packing cardboard box and the second nylon packing around the dry heatsterilizer. The below written are provided with the instrument, please check them;

    Users manual 1 piece electrical cable 2 pieces of shelves 4 pieces of shelf carriers

    3.3. Environmental Conditions

    Please pay special attention to the followings,

    Indoor use only Temperature from 5C to 40 C Humidity level % 80 up to 22 C Maximum height 2000 m. Temperature range for maximum performance: 15C / 25C

    3.4. Mains Supply

    The dry heat sterilizer/oven requires 230, 50 / 60 Hz.

    Please make sure that the supplied mains matches the required power ratings. If no,provide an extra line to support.

    Always plug the dry heat sterilizer/oven to properly earthed sockets.

    A supply fitted with a circuit breaker should be used for protection against indirectcontact in case of an insulation fault.

    3.5. Positioning

    Check that no damage has occurred during transport. Lift the dry heat sterilizer/oven underneath and carry it to its place carefully. Balance the dry heat sterilizer/oven on the four pedestals. If necessary, provide stable

    standing by adjusting the pedestal heights. Insert the shelf carriers and then the shelves.

    Check the followings,

  • 8/2/2019 dry oven

    8/15

    8

    The proposed site is suitable for users, The dry heat sterilizer/oven does not occupy the utilisation space of others or damage

    them.

    3.6. Prior to Operation

    Check the followings,

    Make sure that the safety thermostat is adjusted to the temperature which is higherthan the set temperature.

    Liquids are not heated in sealed containers. The boiling points of the samples are higher than the set temperature. The freezing points of the samples are lower than the set temperature. The liquids which may expand during heating do not overflow from their containers. The set temperature does not destroy the structure of the samples. The vapours and gases which are generated during the operation are not harmful to

    human health or flammable or explosive.

    4. OPERATING PRINCIPLES

    4.1. Switching On

    Push on/off switch. See that the microprocessor control system activates.

    Set the values and start the operation.

  • 8/2/2019 dry oven

    9/15

    9

    4.2. Control Panel

    1. Temperature Display: This display shows the chamber temperature during theoperation and the set temperature value during programming. The error codes arealso shown on this display.

    2. Time Display: This display shows the elapsed time during the operation and the setvalue during programming.

    3. Alarm Mute Key: This key is pushed to mute the alarm which sounds when theprogram ends and if any failure occurs duringthe operation.

    4. Start / Stop Key: This key is pushed to start the program or to stop the runningprogram.

    5. Temperature Set Key: This key is pushed to set the temperature.

    6. Temperature Value Increase/Decrease Keys: These keys are pushed to increaseor decrease the values on the temperature display.

    1

    9

    5

    6

    2

    7

    4

    8

    10

    3

  • 8/2/2019 dry oven

    10/15

    10

    7. Time Set Key: This key is pushed to set the time (01 minute 99.9 hours and Holdposition). The number after the decimal point is multiplied by 6 minutes to find the realtime. (Ex: 17.6 = 17 hours 36 minutes)

    8. Time Value Increase/Decrease Keys: These keys are pushed to increase or

    decrease the values on the time display.

    9. Heat Led: This led flashes during the heating process.

    10. Alarm Led: It lights up when the program ends and if any failure occurs during theoperation.

    4.3. Programming Summary

    NOT: During the operation of the program, the time starts to count up after the

    instrument has reached to the set temperature.

    Push the temperature set key.

    Push the temperature set key again to save thetemperature value.

    Push the time set key.

    Set the temperature by pushing the valueincrease/decrease keys on the temperatureadjustment side.

    See t n on the temperature display. Set the time valueby pushing the value increase/decrease keys on the timeadjustment side (01 minute to 99 hours 54 minutes orHold)

    Push the time set key again.

    Push the start/stop key to start the program.

    See dly on the temperature display. Set the delay time,after which the program starts, by pushing the valueincrease/decrease keys on the time adjustment side (01

    minute to 99 hours 54 minutes)Push the time set key again to save the settings.

  • 8/2/2019 dry oven

    11/15

    11

    4.4. Completion of the Work

    See that the program is over. Take the samples out. Wipe the chamber surface to clean if needed. You may leave the dry heat sterilizer / oven on stand-by position or switch it off.

    Warning!!The samples may be hot after the operation, please be careful while handlingthem!!

    5. PERIODIC MAINTENANCE AND CLEANING

    5.1. Periodic Maintenance

    The dry heat sterilizer/oven does not require any periodical maintenance which is carriedout by the operator.

    Please contact to Nuve agent for an authorised service or maintenance.

    5.2. Cleaning

    After unplugging the equipment and the equipment is at the room temperature, wipedown the chamber to remove any undesirable effects of the operation, for examplespillage.

    You may use a soft brush to clean the chamber.

    For the external body, you may use a piece of cloth. Mild detergent use is recommendedto remove difficult dust and dirt.

    Protect your chamber against rust coming from outside.

    Please be aware of the undesirable effects of the chemicals and be careful whileapplying them.

    6. TROUBLESHOOTING

    If the dry heat sterilizer/oven does not operate, check the followings,

    the on/off switch is on, the plug is plugged-in properly, the plug is not defective, the mains supply is present, fuses of the instrument are sound, the fuse of the installation on which the plug is mounted is not defective

  • 8/2/2019 dry oven

    12/15

    12

    The incubator does not heat, check the followings,

    The program is started, The safety thermostat is adjusted higher than set temperature.

    Error Codes

    OFL: 1. The chamber temperature exceeds 255C.2. The temperature sensor (Fe-Const) endings are broken.

    Switch off the dry heat sterilizer/oven and call for an technical help.

    EOF: This error code is displayed when power is interrupted and the temperature in thechamber goes out of the 2 C alarm limit. The sterilization program is cancelled.You may re-start the dry heat sterilizer/oven by pushing the start/stop key.

    PLEASE CONTACT TO AN AUTHORIZED NUVE AGENT TO SEEK TECHNICALHELP IF AN ERROR OCCURS.

  • 8/2/2019 dry oven

    13/15

    13

    7. ELECTRICAL CIRCUIT DIAGRAMS

    7.1. FN 300

    7.2. FN 400

  • 8/2/2019 dry oven

    14/15

    14

    7.3. FN 500

    7.4. FN 400 P

  • 8/2/2019 dry oven

    15/15

    15

    7.5. FN 500 P