Draft Guidelines tga - custom made medical devices 2008

19
Australian Guidance Document Number xx Cathy Thomas Professional Representation OT Australia WA

description

Draft Guidelines tga - custom made medical devices 2008. Australian Guidance Document Number xx Cathy Thomas Professional Representation OT Australia WA. Draft Guidelines custom made medical devices 2008. Therapeutic Goods Administration web site - PowerPoint PPT Presentation

Transcript of Draft Guidelines tga - custom made medical devices 2008

Page 1: Draft Guidelines  tga -   custom made  medical devices 2008

Australian Guidance Document

Number xx

Cathy Thomas

Professional Representation

OT Australia WA

Page 2: Draft Guidelines  tga -   custom made  medical devices 2008

Therapeutic Goods Administration web sitehttp://www.tga.gov.au/devices/drcustom.pdfDisclaimer- “This document is provided for

guidance only”

Australian regulatory guidelines for medical devices (Jun 2010)- Custom made medical devices

section is “yet to be drafted”

Page 3: Draft Guidelines  tga -   custom made  medical devices 2008

Purpose:The regulatory framework for medical devices

contains provisions for regulation of custom made medical devices and the components of customised medical devices, relating to

Demonstrating compliance -essential safety and performance

Documentation of manufacturing process Labelling of finished devices Record keeping Post market reporting Notification manufacturing activities to

TGA

Page 4: Draft Guidelines  tga -   custom made  medical devices 2008

Regulation of medical devices is intended to ensure that all devices have been shown by the manufacturer to comply with the Essential Principles of Safety and Performance

Devices that are assembled, adapted, processed or used in accordance with the manufacturer’s intentions will comply with the Essential Principles

Page 5: Draft Guidelines  tga -   custom made  medical devices 2008

A person who customises a device from components supplied by others IS NOT REQUIRED to perform the relevant investigations to determine if the finished product complies with the Essential Principles

Page 6: Draft Guidelines  tga -   custom made  medical devices 2008

A Manufacturer is the person who is responsible for the design, production, packaging and labelling of a device before it is supplied

A person is NOT the Manufacturer if the person assembles or adapts the device for an individual patient And, the device has already been supplied,And, the assembly or adaption of the device DOES NOT CHANGE THE PURPOSE INTENDED FOR THAT DEVICE

Page 7: Draft Guidelines  tga -   custom made  medical devices 2008

CUSTOMISED MEDICAL DEVICES

VS

CUSTOM MADE MEDICAL DEVICES

Page 8: Draft Guidelines  tga -   custom made  medical devices 2008

CUSTOMISED MEDICAL DEVICES

Page 9: Draft Guidelines  tga -   custom made  medical devices 2008

A CUSTOMISED MEDICAL DEVICE is a medical device that is prepared/assembled by a person who is not the manufacturer of the device, but

The person assembles or adapts the device for an individual patient

The device has already been supplied The assembly of the device does not change

the purpose intended for the use of the device (by means of information supplied, labelling,

instructions for use, advertising material)

Page 10: Draft Guidelines  tga -   custom made  medical devices 2008

CUSTOMISED MEDICAL DEVICES are considered to be:

Assembled from components which are mass produced, with the intent these components be assembled into a finished medical device for an individual patient

The device should perform as intended if assembled in accordance with manufacturer’s instructions

Should comply with the Essential Principles

Page 11: Draft Guidelines  tga -   custom made  medical devices 2008

Examples of Customised Medical devices are:

Plaster or fibreglass splint used to support a fracture using materials specifically intended for this purpose

Purchased cushioning pre-shaped to accommodate user specific seating and positioning requirements, may be fixed or in a wheelchair

Artificial limbs constructed of component parts

Page 12: Draft Guidelines  tga -   custom made  medical devices 2008

CUSTOM MADE MEDICAL DEVICES

Page 13: Draft Guidelines  tga -   custom made  medical devices 2008

CUSTOM MADE MEDICAL DEVICES are considered to be:

Constructed from materials, components or elements where there was no specific intention by the manufacturer that the materials be used in the manufacture of a medical device

May include metals, adhesives, plastics, fibreglass, glass etc from commercial sources

Page 14: Draft Guidelines  tga -   custom made  medical devices 2008

“Healthcare practitioners who create and provide custom made medical devices to their patients are therefore considered MANUFACTURERS of custom made medical devices”

Page 15: Draft Guidelines  tga -   custom made  medical devices 2008

Examples of Custom Made medical devices are:

Plaster cast using commercially available plaster of Paris and gauze bandage material

A complete wheelchair or seating specifically constructed to accommodate seating and positioning of the user

A dynamic splint where components using welding rod, paper clips and fishing line have been added

The custom fitted socket for an artificial limb (so even only part of a customized device may be custom made)

Page 16: Draft Guidelines  tga -   custom made  medical devices 2008

The Manufacturer (the OT) of Custom Made medical devices (or part thereof)

must be able to demonstrate that as far as possible the finished device will be in accordance/compliance of the Essential Principles of safety and performance for medical devices

andmust establish and maintain a process for identifying potential hazards associated with their custom made device, associated risks, controlling risks and monitoring effectiveness of control methods

Page 17: Draft Guidelines  tga -   custom made  medical devices 2008

Information to be provided with the device

The manufacturer’s name, or trading name, and address

The intended purpose of the device, the intended user of the device, and the kind of patient on whom the device is intended to be used (if this information is not obvious)

Sufficient information to enable a user to identify the device, or if relevant, the contents of packaging

Any particular handling or storage requirements applying to the device

Any warnings, restrictions, or precautions that should be taken, in relation to use of the device

Any special operating instructions for the use of the device

If applicable, an indication that the device is intended for a single use only

An indication that the device has been custom-made for a particular individual and is intended for use only by that individual

If applicable, an indication that the device is intended to be used only for clinical or performance investigations before being supplied

For a sterile device, the word ‘STERILE’ and information about the method that was used to sterilise the device

The batch code, lot number or serial number of the device

If applicable, a statement of the date (expressed in a way that clearly identifies the month and year) up to when the device can be safely used

If the information provided with the device does not include the information mentioned in item 12 — a statement of the date of manufacture of the device (this may be included in the batch code, lot number or serial number of the device, provided the date is clearly identifiable)

Page 18: Draft Guidelines  tga -   custom made  medical devices 2008

Some suggestions for what we can start to do to comply:

Have procedures and protocol files up-to-date and patterns drawn up so your regular items are reproducible by any therapist

Document any safety considerations incl precautions, do’s and don’ts

Have patient handouts providing instruction on care, safety, use, when to return to therapist, intention of use, date of manufacture, batch no, contact details

Page 19: Draft Guidelines  tga -   custom made  medical devices 2008

Record batch numbers, exact type of product manufactured, supplier details (if relevant) against patient’s name

Provide suppliers with patient details when ordering specific or specialised products in case of need for recall