Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories

Transcript of Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo...

Page 1: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Dr. Meliyanthi GunatillakaConsultant Chemical Pathologist

National Hospital of SrilankaColombo

Application of Westgard multi rules in medical

laboratories

Page 2: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 2

Components of Quality Assurance

Internal Quality Control (IQC)Steps taken by all health care professionals in their day to day activities to ensure generation of reliable laboratory results. (Prospective activity)

Pre-analytical phase

Analytical phase

Post-analytical phase

Page 3: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 3

Components of Quality Assurance cont…

External Quality Assessment (EQA)

Organized Inter-laboratory comparison

Tool to assess IQC to improve performance Performed by an independent agency Retrospective and periodic

Page 4: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 4

Control of analytical phase• Validated analytical methods

• Calibration of analytical procedure using traceability established calibration material

• Quality control material in the assay Monitor the performance by Levy Jennings charts Application of Westgard multi rules

• Equipment & Reagents

Page 5: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Measurement Procedure

• Is used to determine the value of a quantity

• This estimate contains a measurement error, which is the difference between the obtained value and the true value of the measurand.

• Measurement error has 2 components: Random error & systematic error

April 21, 2023 5

Page 6: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Measurement Accuracy

Is the extent of the agreement between the result of a measurement and the true value.

( True value of an analyte is obtained by using a reference method and CRM with stated uncertainties and traceability.)( Practically not possible)

(Approximations are obtained using QC material)

Accuracy encompasses the concepts of precision and trueness

( Earlier it was only the trueness)April 21, 2023 6

Page 7: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• ACCURACY

• PRECISION Is the extent of agreement between independent results of repeated measurements

• Imprecision :Expressed as SD or Coefficient ofVariation CV mmol/L or CV %

• TRUNESS Is the extent of agreement between the mean value of repetitions and the true value of the measured

• Expressed as a % of difference between the mean value of multiple repetitions and the true value.

April 21, 2023 7

Page 8: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 8

Random error

Is an unpredictable analytical variation which influence each measurement differently in either a positive or negative direction and to a different extent in magnitude.

Page 9: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

•Possible causes of imprecision are

• Wrong pipetting technique

• Variable reaction timing and temperature

of measurement procedures

• Instrument instability

April 21, 2023 9

Page 10: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 10

Systematic error (bias)Constant systematic bias denotes a constant difference between the true value and the observed value regardless of the concentration level.

Proportional systematic bias denotes a difference between the true value and the observed value, which changes proportionally as the concentration level changes.

Page 11: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 11

Possible causes of systematic errors

Errors in the assigned value to the calibrator

Deterioration of calibration material

Incorrect sample or reagent volume pipetted

Incorrect reaction timing or temperature

Incorrect instrument setting (wavelength)

Calculation errors

Presence of interferents in samples(affect the

individual samples)

Page 12: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 12

• Calculations

• Mean

If the distribution of results is a normal Gaussian curve the mean is the total score of all the measurements divided by the number of measurements.

If not, remove the outliers (values with > 3SD) and recalculate the mean.

Page 13: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 13

• Standard deviation

A descriptor of the extent of dispersion of a population of test values (set of measurement data)

_

SD = √(X – X )2 ÷ (N – 1 )

Coefficient of variation

Relates the SD to the actual measurement so that measurements at different levels can be compared.

CV = SD CV = SD x 100

mean mean

Page 14: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Selection of Quality control material

Control materials should be used as they were patients samples and each and every one of the steps and stages that make up measurement procedure should be followed

Essential characteristics

Homogenicity: No significant differences between the parts that make up a lot of the control material

(dispensing errors, variability in lyophylization due to residual water and reconstitution errors)

( Check the procedure for reconstitution of QC)April 21, 2023 14

Page 15: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Stability

The component that is analysed must be stable over a

sufficient period of time. Once recontituted the aliquoted vials should be stored in appropriate temperatures according the manufacturers instructions. Labels : date of reconstituition

No freezing and thawing cycles.

( ALP increases if stored at 2- 80 C)

April 21, 2023 15

Page 16: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Value of the control quantity

• Value should be close to the clinical decision level or cut off levels

• If various cut off values exist, a number of control material can be used with different concentrations

• Physiological level : upper limit of reference range

• Pathological ranges : low, medium & high

• Commutability

The QC material must respond to the changes or alterations in the measurement procedure in the same way as patients samples.

April 21, 2023 16

Page 17: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Types of QC material

Defined matrix material : non biological or systhetic matrix in the form of buffers, stabilizing agents

Biological matrix: Derived from biological fluids, blood,plasma,serum & urine

Matrix varies from lot to lot

Human origin : especially for immunoassays

Animal origin: cheaper and less risk of infections

April 21, 2023 17

Page 18: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Quality Control materials with assigned values• Manufacturers assigned value

Method & instrument specific value

Details of the certified reference material used

( with UM & stated references)

Range and standard deviation should be given

Laboratory assigned value

The mean value obtained following minimum of 20 repititions should fall within the 2SD of the method & instrument specific assigned value of the manufacturer ( preferably within the UM)

If not : peer review world wide values : local agent

iif

April 21, 2023 18

Page 19: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Perform the glucose test under optimum conditions for minimum 20 times • Example 01- Obtained Readings• 1-120, 2-120, 3-120, 4-120, 5-119, 6-120, 7-120, 8-121, 9-120,10 -120, 11-120, 12 -120, 13-120, 14-120, 15-119, 16- 120, 17-120, 18- 121, 19-120,20 -120,

• Mean = 120.0 • Standard Deviation = 0.45 • +1 s = 120.0 + 0.45 = 120.45• - 1s = 120.o - 0.45 = 119.55

• +2s = 120.0 + 2x 0.45 = 120.9• - 2s = 120.0 - 2x0.45 = 119.1

• +3s = 120.0 + 3x 0.45 = 121.35• -3s = 120.0 - 3x 0.45 = 118.65

How do we find the mean glucose value of the in-house QC sample?

April 21, 2023 19

Page 20: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Mean glucose under routine conditions

• Perform the glucose under routine conditions for 30 days• Example 03• 1-120, 2-74, 3-124, 4-120, 5-117, 6-116, 7-122, 8-126, 9-191,10 -126, 11-120, 12 -123, 13-124, 14-120, 15-117, 16- 116, 17-122, 18- 126, 19-130, 20 -

126, 21-64, 22 -123, 23-124, 24-120, 25-117, 26- 116, 27-122, 28- 126, 29-200, 30 -126,

Corrected (After removal of outliers)• Mean 123.2 120.6• SD 24.0 19.5 • ±1s 147.2 – 99.2 140.1-101.1• ±2s 171.2- 75.2 159.6 -81.6• ±3s 195.2 – 51.2 179.1- 62.1•

April 21, 2023 20

Page 21: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 21

QC Charts and rules Shewhart/Levey and Jennings chart

Analyze the QC material by the analytical method to be controlled on at least 20 times under optimal conditions and calculate the mean, standard deviation and CV % (OCV) Remove the outliers

Construct the control chart :

Y axis-control value

X axis-days

Page 22: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 22

Mean

+2SD

+3SD

-2SD

-3SD

QC Charts and rules cont…

Warning

Warning

Action

Action

Page 23: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

L-J Chart

DaysWithin ±2SD – accept > 3SD –reject between 2SD & 3SD - warning

A Trend is when six or more consecutive plots all fall in one direction, specifically upward or downward.

A shift is when six or more consecutive plots all fall on/above or below the mean line.

April 21, 2023 23

Page 24: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 24

QC Charts and rules cont…

Introduce a control specimen daily to the analytical run, plot each value on the chart.

After 20 days calculate the mean, SD and RCV, remove the outliers & recalculate the mean,SD and construct the chart, plot daily control value and follow the rules to accept/reject the run.

Page 25: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 25

QC Charts and rules cont…

Westgard Multirule Chart and rules

Introduce two control specimens into each analytical run one for each of the two concentrations ( NORMAL & ABNORMAL)

Plot the charts with mean and SD.

12s ‘’ 1’’ Number of control observations 2s Type of error s – Standard deviation

Page 26: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 26

QC Charts and rules cont…

If both control results are within 2SD from their target the batch is accepted

If at least one control result is more than 2SD from the target, the remaining rules are evaluated in turn, and the batch is rejected if any one rule is satisfied. If none is satisfied the batch is accepted.

The situation should be investigated before the next batch is analyzed

Page 27: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

• Automatic chemistry analysers• Normal & abnormal QC : chemistry

Daily & 50th /100th and at the end : day time

night time : at the start• Low,medium & high QC: immunoassays

Daily all 3 levels• Procedure : calibration of the method

QC:all results within 2SD : accept:run the samples

IF one QC is out, check the violations of rules in other levels. Find the cause. ( calibration) Re run the QC : IF out again : new aliquot of QC

All QC levels satisfactory – Run the samples

April 21, 2023 27

Page 28: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

01

April 21, 2023 28

Page 29: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

12s refers to the control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus/minus 2s. In the original Westgard multirule QC procedure, this rule is used as a warning rule to trigger careful inspection of the control data by the following rejection rules.

02

April 21, 2023 29

Page 30: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

22s - reject when 2 consecutive control measurements exceed the same mean plus 2s or the same mean minus 2s control limit.

03

April 21, 2023 30

Page 31: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

R4s - reject when 1 control measurement in a group exceeds the mean plus 2s and another exceeds the mean minus 2s.

04

April 21, 2023 31

Page 32: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

05

41s - reject when 4 consecutive control measurements exceed the same mean plus 1s or the same mean minus 1s control limit.

April 21, 2023 32

Page 33: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

10x - reject when 10 consecutive control measurements fall on one side of the mean.

06

April 21, 2023 33

Page 34: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

8x - reject when 8 consecutive control measurements fall on one side of the mean.

07

April 21, 2023 34

Page 35: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models…..

Westgard multirules

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

12x - reject when 12 consecutive control measurements fall on one side of the mean.

April 21, 2023 35

Page 36: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models….

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   

2of32s - reject when 2 out of 3 control measurements exceed the same mean plus 2s or mean minus 2s control limit;

What are other common multirules?

In situations where 3 different control materials are being analyzed, some other control rules fit better and are easier to apply, such as:

April 21, 2023 36

Page 37: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models….

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   What are other common multirules?

31s - reject when 3 consecutive control measurements exceed the same mean plus 1s or mean minus 1s control limit.

April 21, 2023 37

Page 38: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models….

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   What are other common multirules?

6x - reject when 6 consecutive control measurements fall on one side of the mean.

April 21, 2023 38

Page 39: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models….

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   What are other common multirules?

9x - reject when 9 consecutive control measurements fall on one side of the mean.

April 21, 2023 39

Page 40: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

Symbolic Models Used in Internal Quality Control

variance

                   Analogous models….

13s refers to a control rule that is commonly used with a Levey-Jennings chart when the control limits are set as the mean plus 3s and the mean minus 3s. A run is rejected when a single control measurement exceeds the mean plus 3s or the mean minus 3s control limit.

                                                                   What are other common multirules?

7T - reject when seven control measurements trend in the same direction, i.e., get progressively higher or progressively lower.

April 21, 2023 40

Page 41: Dr. Meliyanthi Gunatillaka Consultant Chemical Pathologist National Hospital of Srilanka Colombo Application of Westgard multi rules in medical laboratories.

April 21, 2023 41