Dose Recalculation and the Dose-Guided Radiation Therapy (DGRT) Process Using Megavoltage Cone-Beam...
-
Upload
joey-cheung -
Category
Documents
-
view
219 -
download
0
Transcript of Dose Recalculation and the Dose-Guided Radiation Therapy (DGRT) Process Using Megavoltage Cone-Beam...
Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 2, pp. 583–592, 2009Copyright � 2009 Elsevier Inc.
Printed in the USA. All rights reserved0360-3016/09/$–see front matter
doi:10.1016/j.ijrobp.2008.12.034
PHYSICS CONTRIBUTION
DOSE RECALCULATION AND THE DOSE-GUIDED RADIATION THERAPY (DGRT)PROCESS USING MEGAVOLTAGE CONE-BEAM CT
JOEY CHEUNG, B.A.,* JEAN-FRANCOIS AUBRY, M.S.,*y SUE S. YOM, M.D., PH.D.,*
ALEXANDER R. GOTTSCHALK, M.D., PH.D.,* JUAN CARLOS CELI, PH.D.,xyy AND JEAN POULIOT, PH.D.*
*Department of Radiation Oncology, University of California San Francisco, Comprehensive Cancer Center, San Francisco, CA;yDepartement de physique, genie physique et d’optique, Universite Laval, Quebec, Quebec, Canada; xOncology Care Systems Group,
Siemens Medical Solutions USA, Inc., Erlangen, Germany; yycurrently at IBA Dosimetry, Schwarzenbruck, Germany
Purpose: At the University of California San Francisco, daily or weekly three-dimensional images of patients intreatment position are acquired for image-guided radiation therapy. These images can be used for calculatingthe actual dose delivered to the patient during treatment. In this article, we present the process of performingdose recalculation on megavoltage cone-beam computed tomography images and discuss possible strategies fordose-guided radiation therapy (DGRT).Materials and Methods: A dedicated workstation has been developed to incorporate the necessary elements ofDGRT. Patient image correction (cupping, missing data artifacts), calibration, completion, recontouring, anddose recalculation are all implemented in the workstation. Tools for dose comparison are also included. Examplesof image correction and dose analysis using 6 head-and-neck and 2 prostate patient datasets are presented to showpossible tracking of interfraction dosimetric endpoint variation over the course of treatment.Results: Analysis of the head-and-neck datasets shows that interfraction treatment doses vary compared with theplanning dose for the organs at risk, with the mean parotid dose and spinal cord D1 increasing by as much as 52%and 10%, respectively. Variation of the coverage to the target volumes was small, with an average D5 dose differ-ence of 1%. The prostate patient datasets revealed accurate dose coverage to the targeted prostate and varyinginterfraction dose distributions to the organs at risk.Conclusions: An effective workflow for the clinical implementation of DGRT has been established. With these tech-niques in place, future clinical developments in adaptive radiation therapy through daily or weekly dosimetricmeasurements of treatment day images are possible. � 2009 Elsevier Inc.
Dose-guided radiation therapy, Megavoltage cone-beam CT, Dose calculation, Treatment replanning, Image-guided radiation therapy.
INTRODUCTION
With recent advancements in radiation treatment planning
and delivery systems such as intensity-modulated radiation
therapy (IMRT), very sharp dose gradients can be achieved
to attain optimal dose distributions. These advancements,
however, require accurate patient positioning and setup to en-
sure the proper treatment delivery (1, 2). This problem has
been addressed through the development of in-room patient
imaging systems that can image the patient on the treatment
table, which can then be used to verify and correct for patient
misalignments or target organ movement. This technique is
5
known as image-guided radiation therapy (IGRT) and has
quickly become a popular solution for patient setup (3, 4).
At the University of California San Francisco Comprehen-
sive Cancer Center, IGRT has been implemented using
a Siemens MVision megavoltage cone-beam computed
tomography (MVCBCT) system (5–8) to correct for patient
setup errors. For head-and-neck patients, a thermoplastic pa-
tient-specific mask is used to position the patient daily, and
weekly cone-beam images are acquired to correct for setup
errors and to assess the long-term patient positioning accu-
racy. For prostate patients, daily low-dose cone-beam images
Reprint requests to: Joey Cheung, University of California SanFrancisco Dept. of Radiation Oncology, Helen Diller Family Com-prehensive Cancer Center, 1600 Divisadero St., Suite HM006, SanFrancisco, CA 94115. Tel: (415) 353-7179; E-mail: [email protected]
Partly supported by Siemens Oncology Care Systems.Acknowledgments—The authors would like to thank ChandrasekharNunna, Thomas Boettger, and Venkataramana Abbaraju for their
83
aid in the development of the workstation and Sherry Leeper forher contribution in this work. One of the authors (J.F.A.) acknowl-edges financial support from the Natural Sciences and EngineeringResearch Council of Canada (NSERC).
Received Sept 17, 2008, and in revised form Nov 26, 2008.Accepted for publication Dec 19, 2008.
584 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 2, 2009
Fig. 1. Overview of the dose-guided radiation therapy (DGRT) workflow and its integration into the image-guided radi-ation therapy (IGRT) workflow. Dashed lines show possible feedback loops into the clinical process.
are acquired and the patients are realigned based on the posi-
tion of the prostate determined by implanted gold seeds (9,
10). For some prostate patients, weekly cone-beams of
slightly higher dose are also acquired to validate alignment
with soft tissue information or for research purposes.
Because these images are taken directly before treatment,
it has been proposed that they can be used for daily assess-
ment of the actual delivered dose to the patient, providing
feedback to the radiation physician and therapists on the ac-
curacy of the treatment delivery. As a result, deviations from
the plan that are not taken into account by IGRT, such as tu-
mor shrinkage, weight loss, and other internal anatomical
changes, can be assessed. Furthermore, the verification of
the actual treatment dose can help the physician track the evo-
lution of the treatment and determine if and when a replan or
plan adaptation is needed (11, 12).
Recently, our group has worked to develop a workstation
that integrates all of the necessary elements of dose-guided
radiation therapy (DGRT) in an attempt to streamline and in-
troduce the process into the clinical workflow (13, 14). The
workstation is currently built on top of a Syngo-based soft-
ware platform and is codeveloped by the Siemens Oncology
Care Systems innovation group and members of the Univer-
sity of California San Francisco research staff. This article
will go through the steps required to perform dose recalcula-
tion on MVCBCT treatment images and the possible strate-
gies that can be developed using DGRT. In addition,
several examples are chosen to illustrate how this process
can be performed on patient images. Case studies for tracking
daily delivered patient dose are explored.
MATERIALS AND METHODS
Acquisition protocolsSix head-and-neck and 2 prostate patient datasets were used to
illustrate the possibilities of the dedicated DGRT workstation. These
patients received MVCBCT imaging with low-dose image
Fig. 2. The conversion process for converting cone-bean (CB) num-bers in CB images to computed tomography (CT) numbers. The cal-ibration curves are obtained from measurements of phantoms withinserts of known densities. The CB numbers are converted to CTnumbers by matching the calibrated physical densities betweenboth systems.
DGRT process using MVCBCT d J. CHEUNG et al. 585
Table 1. Tumor location, diagnosis, and treatment for head-and-neck and prostate cases
# Age/Gender Location (Primary Site) Stage Prescription (cGy) No. Fraction Total Dose (Gy) Replan? Chemo?
1 55/m Nasopharynx T1N1M0 IIB 212 33 69.96 ConcurrentUndifferentiated carcinoma
2 69/m Oral cavity T1N2bM0 IVA 200 33 66 ConcurrentSquamous cell carcinoma
3 48/m Left tonsillar pillar T3N2b 212 33 69.96 ConcurrentSquamous cell carcinoma
4 48/m Unknown primary with metastasis to leftneck
TXN2bM0 IV 212 33 69.96 No
Squamous cell carcinoma5 59/m Base of tongue T2N3M0 212 33 69.96 Yes Concurrent
Squamous cell carcinoma6 64/m Melanoma of right ear pT1N0 IA 600 5 30 No
Malignant melanoma
# Age/gender Location (primary site) Stage Prescription (cGy)No.
FractionTotal Dose
(Gy) Postradiotherapy CyberKnife Boost
7 70/M Prostate adenocarcinoma cT1c 180 25 45 2 fractions 9.5 GyGleason 4+4 per fractionPSA 4.27
8 57/M Prostate adenocarcinoma cT2a/uT3a 180 25 45 2 fractions 9.5 GyGleason 4+5 per fractionPSA 4.3
acquisition protocols under institutional review board approval.
A total of 31 head-and-neck cone-beams and 8 weekly prostate
cone-beams were used. In the following discussion, the necessary
procedures to perform dose recalculation are outlined in detail.
A proposed clinical workflow diagram of the DGRT process is
shown in Fig. 1.
Image processing and reconstructionBefore reconstruction, a diffusion filter is applied to each cone-
beam projection image using a built-in function in MeVisLab (Me-
Vis Medical Solutions) to improve the contrast-to-noise ratio (15,
16). The cone-beam images are then reconstructed offline through
a modified Feldkamp-Davis-Kress back-projection reconstruction
algorithm and imported to the DGRT workstation (17). Because
of the geometry of the MVCBCT system setup, the center of the im-
age is the isocenter (7).
Image correction and calibration methodsAs with most cone-beam systems, a typical cupping artifact,
resulting from scatter and beam hardening, is present in the images
obtained from our MVCBCT system (18). In addition, a missing
data artifact also appears if the section of the patient being imaged
is larger than the field-of-view of the system. For the dose to be ac-
curately calculated using these images, these artifacts need to be cor-
rected and the images need to be calibrated to the correct electron
densities.
Several methods have been developed to correct for these artifacts
in head-and-neck cases (19, 20). The workstation uses the method
developed by Aubry et al. (20) in which a reference conventional
kVCT image is used to create a set of patient-specific correction
factors for the MVCBCT images. The slice-by-slice correction is ap-
plied to the cone-beam image based on the CT number values of the
blurred and filtered kVCT image. After applying the correction
factors, the corrected image is already calibrated to the correct CT
numbers because the planning kVCT image was used as a reference.
These CT numbers can then be calibrated in the dose calculation
engine to the correct physical and electron densities through mea-
surements of phantoms with inserts of known densities.
For the prostate patients, a different method developed by Aubry
et al. (21) was used that utilizes a set of correction factors calibrated
using custom-built pelvic-shaped water phantoms of various sizes.
The correction factors are determined by interpolating between the
values obtained from the water phantoms based on measured aver-
age radiological thicknesses of the patients. Unlike the head-and-
neck correction, however, the resulting image is not calibrated to
the correct CT numbers because of the nonlinear response of the
kVCT and MVCBCT systems to the density of imaged materials
and further calibration is needed for image correction. This calibra-
tion is done in the process of MVCBCT completion.
MVCBCT completionThe maximum field of view of our MVCBCT system is 27 cm in
diameter in the axial plane and 27 cm in length along the craniocau-
dal axis. Because the field of view of our MVCBCT system is
smaller than most patients, the images exclude part of the patient’s
anatomy. The resulting treatment beam attenuation for dose calcula-
tion will be incorrect if this missing information is not accounted for.
To correct for this, the cone-beam image is completed using the
planning kVCT image as a reference. The cone-beam images are
registered to the kVCT image based on bony alignment in the
area of interest. A new image is created called a cone-beam+
(cone-beam plus) image that takes the kVCT image and replaces
the overlapping region of the two images with the cone-beam image.
In creating the cone-beam+, the machine isocenter from the original
cone-beam is preserved.
For the prostate patients, the calibration curves of both imaging
systems are used to convert the CT numbers from the MVCBCT im-
age (CB numbers) to CT numbers of the planning kVCT image. This
process is shown in Fig. 2.
RecontouringFor this study, two radiation oncologists specialized in head-and-
neck and prostate radiation therapy recontoured several of the
586 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 2, 2009
Fig. 3. Images of a head-and-neck cone-beam (a) uncorrected, (b) corrected, and (c) completed using a kVCT image.
organs at risk on the cone-beam+ images. For head and neck, the
contours used were: brain stem, spinal cord, parotid glands, tempo-
romandibular joints (TMJ), mandible, and larynx. For the prostate
patients, they were: prostate, rectum, bladder, and seminal vesicles.
However, because these images were initially used only for patient
positioning and not for dose recalculation, some of the images were
taken with a reduced field of view to protect critical organs from ad-
ditional exposure. Thus, in some cases, a few of the organs of inter-
est were outside the field of view of the cone-beam and were
therefore excluded from this study.
In addition to the organs at risk, the target volumes for the head-
and-neck datasets were also recontoured. At the University of
California San Francisco, the technique for replanning patients in
the middle of treatment involves maintaining the original gross
tumor volume (GTV) in the second plan while removing any vol-
umes that extend beyond the skin. Similar attempts are made for
the clinical tumor volumes (CTV). This ensures that adequate
dose is given to the high-risk volumes to eliminate the cancer and
that original planned coverage to the targets are not compromised
(13). The same process was performed on the cone-beam+ images
in this study for the GTV and the CTV59.4 (CTV receiving a pre-
scription dose of 59.4 Gy).
Image registration and dose calculationThe DGRT workstation uses a pencil beam dose calculation en-
gine (22–25) based on the KonRad Inverse Planning Software (Sie-
mens Medical) commissioned to a 6MV Siemens ONCOR linear
accelerator. The original treatment plan for each patient was created
using the Pinnacle3 Radiation Therapy Planning System (Philips
Medical Systems), which uses a collapse cone algorithm for dose
calculation. To accommodate for this difference in dose calculation
DGRT process using MVCBCT d J. CHEUNG et al. 587
Fig. 4. Comparisons of the dose different maps for Patient V on the first week of treatment and the third week overlaid onthe planning kVCT image. The colors are windowed to show PDDiffs of greater than 5% (red), less than –5% (blue), andwithin �5% (green). Although visual verification of local changes to parotid dose may not be accurate, overall dosechanges to the spinal cord, larynx, brain stem, and mandible can be useful.
engines, a new plan for the kVCT image is created on the worksta-
tion that preserves the beam shapes, angles, and energies of the orig-
inal plan but renormalizes all the beam weights by the same
multiplicative factor in order to give the same dose to the isocenter
as the original plan. The dose distribution on the planning kVCT is
then recalculated using this renormalized plan so that differences in
treatment dose will be due to anatomical and positioning changes
rather than differences in dose calculation algorithms. This new
plan is copied from the kVCT to the cone-beam+ treatment image
based on the actual treatment registration to ensure that the treatment
isocenter for dose recalculation purposes is the same as it was on the
day of treatment. Note that the renormalization process is only per-
formed once per patient on the planning kVCT image and not on the
cone-beam images.
Dose comparisonSeveral methods of comparing the planned dose to the treatment
dose have been integrated into the DGRT workstation. A simple
side-by-side comparison of the two dose distributions using the
same isodose markers allows for a visual overview of the difference
in doses between the two images. The percentage dose difference
(PDDiff) of the local treatment dose to the reference planning
dose can also be calculated and viewed as an overlay on top of
the cone-beam+ image with the contoured anatomy. Although
voxel-by-voxel dose differences may not accurately represent
dose changes on the cellular level, the PDDiff does allow the user
to see the local difference in overall dose to different parts of the pa-
tient anatomy if the anatomy has had a minimal shift in location. The
dose difference is normalized to the reference plan dose so that
underdoses and overdoses can be easily visualized and understood.
The dose and PDDiff information with user-defined limiting param-
eters for the contoured anatomy can be extracted as a text file and
analyzed. Furthermore, dose–volume histograms (DVHs) can be
plotted and viewed on the workstation.
RESULTS
Head-and-neck casesThe six head-and-neck patient datasets used for this study
were all newly diagnosed with cancers in different primary
tumor sites. Table 1 lists the treatment information along
with the location and stages of the tumors.
Figure 3 shows an example of the application of the correc-
tion method and image completion on a head-and-neck pa-
tient image. For this particular dataset, the cone-beam was
taken on the first week of treatment. The cupping and missing
data artifacts are clearly seen on the images on the left (a).
After correction (b), the images are uniform and the density
values are calibrated to the correct CT numbers needed for
dose calculation. The image can then be completed (c) using
the original planning kVCT image to complement the miss-
ing data.
Figure 4 shows the evolution of the dose differences map of
one head-and-neck patient (Patient V) throughout treatment. The
color overlay is windowed to show PDDiffs greater than 5%
(red), less than –5% (blue), and within�5% (green) of the plan-
ning dose. A number of factors can result in larger overdosed
(red) regions including patient misalignments, tumor or organ re-
gression, more global anatomical changes, and weight loss.
Prostate casesBoth prostate cases were treated with concurrent hormone
therapy. The treatment information can be found in Table 1.
Figure 5 shows the application of the correction method and
image completion on a pelvic patient. As with head-and-neck
cases, the cone-beam correction and calibration (B) and com-
pletion (C) on pelvic images result in a final image that is den-
sity accurate.
Figure 6 shows a side-by-side dose grid comparison of
a prostate planning kVCT image with a treatment cone-
beam+ image for Patient VII. Analysis shows that the volumes
of the prostate, rectum, seminal vesicles, and bladder in this
image changed by –4%, –4%, –44%, and 36%, respectively.
Dosimetric end point trackingFigure 7 shows the time evolution of the percent difference
in mean dose compared to planning dose to the left and right
588 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 2, 2009
Fig. 5. Images of a pelvic cone-beam (a) uncorrected (b) corrected and (c) completed using a kVCT image.
parotids for three of the 6 patients (Patients I, II, and V) and
the difference in D1 dose (Dx dose is the minimum dose re-
ceived by the highest dosed x-percent of the volume) to the
spinal cord for all 6 patients. The D1 dose was used instead
of the maximum dose because of the high sensitivity and var-
iation of the maximum dose. For the parotids, 3 head-and-
neck patients were omitted because their cone-beam images
were taken with a limited field of view that resulted in the
parotids being incompletely imaged or not imaged at all.
An average net increase in dose to both parotids is observed
for all 3 patients. In addition, all 6 patients also received an
average net increase in dose to the spinal cord over the course
of treatment. A summary of the percent dose differences to
the target and various recontoured organs at risk averaged
across all 6 patients is shown in Table 2.
For the prostate patients, DVHs were plotted to observe the
differences in dose to the various recontoured organs. Figure 8
shows the DVHs of the bladder, prostate, seminal vesicles,
and rectum for Patient VII (a) and Patient VIII (b). The dotted
lines show the original planned DVHs to the regions of inter-
est. Figure 9 shows the percentage volume changes from the
planning kVCT volume for the bladder, prostate, rectum, and
seminal vesicles for the 2 prostate patients in our study.
DISCUSSION
Of the different head-and-neck structures that were con-
toured, the parotids tended to be the only ones that exhibited
a long-term unidirectional change in volume and geometry.
As has been shown in other studies (26–28), the irradiated pa-
rotids tended to shrink in size over the course of treatment
and migrate inward toward the center of the skull. This moves
the parotids into higher dose regions over time and a general
increase in dose to the right parotid was seen across all
DGRT process using MVCBCT d J. CHEUNG et al. 589
Fig. 6. A side-by-side comparison of the dose maps from the planning kVCT and the treatment cone-beam+ images. Bothimages use the same dose range colors and the values are the lower bounds of the ranges. The yellow contour is the bladder,purple is the prostate, green is the seminal vesicles, and brown is the rectum.
3 patients after the third week of treatment. Patient I showed
some weight loss during treatment and received a higher dose
to the right side. Patient II received a right unilateral treat-
ment, which could explain the large increase in dose to the
right parotid over the course of the treatment. Patient V ex-
hibited a large amount of weight loss (about 4 cm of soft tis-
sue on both sides of the neck) and tumor regression, which
can be seen in the increase in dose to both parotids.
Weekly variation of the spinal cord dose seemed to be
fairly random but tended to be shifted toward overdoses
(Fig. 7). These errors could be attributed to daily setup errors,
weight loss, or changes in anatomy between the date of the
planning kVCT and the start of the treatment. One notable ex-
ception in the study is Patient V, who had lost a large amount
of weight between the cone-beam images, resulting in a large
(10.7%) increase in D1 spinal cord dose.
In the study, the TMJ was too small and too close to the
edge of the cone-beam images for 2 of the patients to give ac-
curate results. Of the 4 patients left, there was an average net
decrease in dose to both TMJ (Table 2). The D1 dose to the
mandibles tended to vary randomly based on the data. This
is likely based on patient positioning rather than weight
loss. Slight movement in the patient’s jaw position was ap-
parent between the weekly cone beams and the planning
kVCT as the anatomy was recontoured.
Brain stem variability appeared to be the smallest of all the
anatomy studied. This is likely due to the relative position
and size of the brain stem. Because the base of the skull is
fairly easy to align for IGRT, the brain stem is likely to be po-
sitioned accurately during treatment and receive close to the
planned dose. The larynx, on the other hand, showed high
variability for all 6 patients, most likely from the non-rigidity
of the organ and its close location to the skin surface where
high dose variability can occur.
Dose coverage to the target volumes remained fairly consis-
tent for all of the cone-beam images. The D95 dose to the GTV
changed by an average (standard deviation) of 0.1% (0.8%)
with a maximum decrease of 0%, showing a consistent overall
coverage to the GTV. High-dose coverage to the GTV also re-
mained fairly consistent, with a mean dose increase of 1.0%
(1.7%) and a maximum decrease of –4.0% for the D5 dose. Sim-
ilar results were found for the CTV59.4 with values for the D95
and D5 with a maximum absolute change of 3.0% and 5.1% and
an average change of 0.1% and 0.7%, respectively.
590 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 2, 2009
Fig. 7. Time evolution of the percentage mean dose difference to the parotids (a) and the D1 percentage dose difference tothe spinal cord (b).
For the prostate patients, the variation of dose to the blad-
der and rectum are likely the result of volume changes. The
current procedure used in our institution aligns the patients
based on daily prostate images. Therefore, although the pros-
tate may be well aligned for each fraction, the bladder, rec-
tum, and other organs may not be well aligned, leading to
varying fractional dose distributions to these regions. It is
also important to consider the accuracy of the recontoured
anatomy on the lower contrast treatment day images when
evaluating the DVH curves. The same oncologist who pre-
pared the treatment plan recontoured the images used in
this study. Figure 9 examines the changes in volume for
the bladder, prostate, rectum, and seminal vesicles. These
Table 2. Statistics for head-and-neck targets and organs atrisk percentage dose differences
Mean Standard Minimum Maximum
Spinal cord D1* 2.2 2.9 –3.3 10.7Left parotid mean dose 5.8 9.5 –4.2 37.9Right parotid mean dose 18.0 18.0 –2.1 51.8Mandible D1* 1.8 4.1 –6.1 12.2Brain stem D1* 1.1 2.7 –4.8 10.1Left TMJ D1* –1.6 8.0 –16.0 10.0Right TMJ D1* –0.5 5.1 –12.3 5.2Larynx mean dose 6.7 9.3 –11.5 25.4GTV D95* 0.1 0.8 0.0 2.1GTV D5* 1.0 1.7 –4.0 5.8CTV59.4 D95* 0.1 1.1 –3.0 2.3CTV59.4 D5* 0.7 1.3 –1.2 5.1
Abbreviations: TMJ = temporomandibular joint; GTV = grosstumor volume; CTV = clinical tumor volume.
* D1 and D5 are the minimum doses received by at least 1% and5% of the volume that received the largest dose, respectively. Thiswas used instead of the maximum dose due to the high sensitivityand variation of the maximum dose. D95 is the maximum dose re-ceived by at least 95% of the volume. These values were chosenbased on the planning dosimetric considerations for the organs atrisk. All values are given as percent increases from the planningdose.
variations in volume will likely affect the dose received by
these organs as reflected by the variation in DVHs in
Fig. 8. However, because these volume changes occur in
three dimensions, the actual dosimetric consequences of
these volume changes are much more complicated and
must be evaluated on a case-by-case basis.
The dose to the prostate and seminal vesicles tended to be
fairly accurate when compared with the original planned
dose. On one of the cone-beam days, Patient VIII received
3.3% less dose to 95% of the prostate volume (D95). How-
ever, on closer observation of the images themselves, the
prostate was not aligned properly on that day of treatment,
which explains the diminished prostate dose coverage. Be-
cause the patients are aligned daily based on prostate location
using gold seed implants, the DGRT process provides a good
check that the primary tumor volume is getting the prescribed
dose throughout treatment.
The DGRT process presented can be used for verification
of the accuracy of the treatment delivery and for an assess-
ment of the daily trend of dose differences over the course
of treatment. Dose accumulation is not currently imple-
mented in the process. Although non-rigid deformation algo-
rithms could be used as an approximation for organ
movement, it is still insufficient in providing dose accumula-
tion on the cellular level. The DGRT process as presented
does not implement non-rigid deformation in order to reduce
calculation times for clinical use, but the process can be easily
integrated if desired and is an avenue that we are currently ex-
ploring for future clinical use. However, even without the
ability to perform dose accumulation, DGRT can provide
valuable feedback about the dose delivered during each treat-
ment fraction. Trends in dose differences can reveal impor-
tant consequences of anatomy change or setup errors.
CONCLUSION
All of the functionalities required to provide an effective
workflow for the clinical implementation of DGRT have
DGRT process using MVCBCT d J. CHEUNG et al. 591
been integrated in our workstation, enabling the regular com-
parison of the dose of the day with the plan. In the examples pre-
sented, it was found that although dose coverage to the target
volumes have remained fairly consistent for head-and-neck
cases, doses to various organs at risk, including the spinal
cord, parotids, mandible, brain stem, TMJ, and larynx, have
been observed to vary between fractions. For the prostate cases,
we have shown that accurate prostate positioning using IGRT
has allowed for accurate dose coverage to the targeted prostate.
However, doses to the organs at risk including the bladder, rec-
tum, and seminal vesicles are observed to vary. With DGRT,
this variation can be monitored to ensure that the dose to the
organs at risk do not surpass critical limits.
Regular monitoring of the daily patient dose based on im-
ages acquired of the patient on the treatment table can be
a powerful tool for tracking the progress and accuracy of
the treatment. In addition, with IGRT implemented in
many facilities, DGRT can be easily integrated into the
clinical workflow. Access to information about the dose-
of-the-day will open the doors to many areas of research
and clinical improvements in treatment verification and
dose accumulation. A systematic approach can also be de-
veloped to determine when a patient requires replanning
based on dosimetric considerations for either the organs at
risk or the target tumor volume. Furthermore, dose–
response relations can be developed based on long-term pa-
tient follow-up studies. The studies presented in this article
are only for a limited set of patients and serve as an exam-
ple of the many possible case studies that can be performed
using the proposed DGRT process.
Fig. 8. Dose–volume histograms of the contoured anatomy for the prostate patients: (a) Patient VII, (b) Patient VIII. Thedotted line shows the original planning dose–volume histogram.
Fig. 9. Plots of the percentage volume change from planning of the bladder, prostate, rectum, and seminal vesicles for the 2pelvic patients.
592 I. J. Radiation Oncology d Biology d Physics Volume 74, Number 2, 2009
REFERENCES
1. O’Daniel JC, Garden AS, Schwartz DL, et al. Parotid glanddose in intensity-modulated radiotherapy for head and neck can-cer: is what you plan what you get? Int J Radiat Oncol Biol Phys2007;69:1290–1296.
2. Webb S. Intensity-modulated radiation therapy. Bristol, UK:Institute of Physics Publishing; 2001.
3. Ma CM, Paskalev K. In-room CT techniques for image-guidedradiation therapy. Med Dosim 2006;31:30–39.
4. Verellen D, Ridder MD, Storme G. A (short) history of image-guided radiotherapy. Radiother Oncol 2008;86:4–13.
5. Pouliot J, Bani-Hashemi A, Chen J, et al. Low-dose megavolt-age cone-beam CT for radiation therapy. Int J Radiat Oncol BiolPhys 2005;61:552–560.
6. Sillanpaa J, Chang J, Mageras G, et al. Developments in mega-voltage cone beam CT with an amorphous silicon EPID: reduc-tion of exposure and synchronization with respiratory gating.Med Phys 2005;32:819–829.
7. Morin O, Gillis A, Chen J, et al. Megavoltage cone-beam CT:System description and clinical applications. Med Dosim2006;31:51–61.
8. Morin O, Gillis A, Descovich M, et al. Patient dose consider-ations for routine megavoltage cone-beam CT imaging. MedPhys 2007;34:1819–1827.
9. Pouliot J, Aubin M, Langden K, et al. (Non)-migration of radi-opaque markers used for on-line localization of the prostate withan electronic portal imaging device. Int J Radiat Oncol BiolPhys 2003;56:862–866.
10. Balter JM, Sandler HM, Lam K, et al. Measurement of prostatemovement over the course of routine radiotherapy using im-planted markers. Int J Radiat Oncol Biol Phys 1995;31:113–118.
11. Hansen E, Bucci MK, Quivey JM, et al. Repeat CT imaging andreplanning during the course of IMRT for head-and-neck can-cer. Int J Radiat Oncol Biol Phys 2006;64:355–362.
12. Lu W, Olivera GH, Chen Q, et al. Deformable registration of theplanning image (kVCT) and the daily images (MVCT) for adap-tive radiation therapy. Phys Med Biol 2006;51:4357–4374.
13. Pouliot J. From dose to image to dose: IGRT to DGRT. In:Mould RF, editor. Choices in advanced radiotherapy. Nucle-tron, BV: The Netherlands; 2007. p. 243–250.
14. Chen J, Morin O, Aubin M, et al. Dose-guided radiation therapywith megavoltage cone-beam CT. Br J Radiol 2006;79:S87–S98.
15. Monteil J, Beghdadi A. New adaptive nonlinear anisotropic dif-fusion for noise smoothing. Volume 3 of the IEEE International
Conference on Image Processing, page 254. Chicago, IL, USA,1998. IEEE Comp Soc, Los Alamitos, CA.
16. Perona P, Malik J. Scale-space and edge detection using aniso-tropic diffusion. IEEE Trans Pattern Anal Machine Intell 1990;12:629–639.
17. Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algo-rithm. J Opt Soc Am A 1984;1:612–619.
18. Spies L, Ebert M, Groh BA, et al. Correction of scatter in mega-voltage cone-beam CT. Phys Med Biol 2001;46:821–833.
19. Morin O, Chen J, Gillis A, et al. Dose calculations using mega-voltage cone-beam CT. Int J Radiat Oncol Biol Phys 2007;67:1202–1210.
20. Aubry JF, Pouliot J, Beaulieu L. Correction of megavoltagecone-beam CT images for dose calculation in the head andneck region. Med Phys 2008;35:900–908.
21. Aubry JF, Cheung J, Gottschalk A, et al. Correction of mega-voltage cone-beam CT images of the pelvic region based onphantom measurements for dose calculation purposes. J ApplClin Med Phys 2009;10:2852.
22. Mohan R, Chui C, Lidofsky L. Differential pencil beam dosecomputation model for photons. Med Phys 1986;13:64–73.
23. Bortfeld T, Schlegel W, Rhein B. Decomposition of pencilbeam kernels for fast dose calculations in three-dimensionaltreatment planning. Med Phys 1993;20:311–318.
24. Boudreau C, Heath E, Seuntjens J, et al. IMRT head and necktreatment planning with a commercially available Monte Carlobased planning system. Phys Med Biol 2005;50:879–890.
25. Yang J, Li J, Chen L, et al. Dosimetric verification of IMRTtreatment planning using Monte Carlo simulations for prostatecancer. Phys Med Biol 2005;50:869–878.
26. Barker JL, Garden AS, Ang KK, et al. Quantification of volu-metric and geometric changes occurring during fractionated ra-diotherapy for head-and-neck cancer using an integrated CT/linear accelerator system. Int J Radiat Oncol Biol Phys 2004;59:960–970.
27. Vasquez Osorio EM, Hoogeman MS, Al-Mamgani A, et al.Local anatomic changes in parotid and submandibular glandsduring radiotherapy for oropharynx cancer and correlationwith dose, studied in detail with nonrigid registration. Int JRadiat Oncol Biol Phys 2008;70:875–882.
28. Lee C, Langen KM, Lu W, et al. Assessment of parotid glanddose changes during head and neck cancer radiotherapy usingdaily megavoltage computed tomography and deformableimage registration. Int J Radiat Oncol Biol Phys 2008;71:1563–1571.