DOPPLER ECHOCARDIOGRAPHY-1

141
DOPPLER ECHOCARDIOGRAPHY-1 BASIC PHYSICS,PULSE WAVE AND CONTINUOS WAVE DOPPLER DR PRADEEP SREEKUMAR

description

DOPPLER ECHOCARDIOGRAPHY-1. BASIC PHYSICS,PULSE WAVE AND CONTINUOS WAVE DOPPLER. DR PRADEEP SREEKUMAR. Sound is a mechanical vibration transmitted through an elastic medium Ultrasound-portion of sound spectrum having frequency greater than 20,000 cycles /sec - PowerPoint PPT Presentation

Transcript of DOPPLER ECHOCARDIOGRAPHY-1

Page 1: DOPPLER ECHOCARDIOGRAPHY-1

DOPPLER ECHOCARDIOGRAPHY-1

BASIC PHYSICS,PULSE WAVE AND CONTINUOS WAVE DOPPLER

DR PRADEEP SREEKUMAR

Page 2: DOPPLER ECHOCARDIOGRAPHY-1

• Sound is a mechanical vibration transmitted through an elastic medium

• Ultrasound-portion of sound spectrum having frequency greater than 20,000 cycles /sec

• Use of ultrasound to study the structure and function of heart and great vessels-echocardiography

• Advantages of ultrasoundCan be directed as a beam and focussedObeys laws of reflection and refractionProduce longitudnal waves

Page 3: DOPPLER ECHOCARDIOGRAPHY-1

Generation Of An Ultrasound Image

Page 4: DOPPLER ECHOCARDIOGRAPHY-1
Page 5: DOPPLER ECHOCARDIOGRAPHY-1
Page 6: DOPPLER ECHOCARDIOGRAPHY-1

Machines

There are 5 basic components of an ultrasound scanner that are required for generation, display and storage of an ultrasound image.

1. Pulse generator - applies high amplitude voltage to energize the crystals

2. Transducer - converts electrical energy to mechanical (ultrasound) energy and vice versa

3. Receiver - detects and amplifies weak signals

4. Display - displays ultrasound signals in a variety of modes

5. Memory - stores video display

Page 7: DOPPLER ECHOCARDIOGRAPHY-1

• Depicted as sine wave-peaks and troughs• One cylce=one compression + one rarefaction• Distance between 2 similar points represent

wavelength• 0.15 to 1.5 mm in soft tissue• Frequency- number of wavelengths per unit time• V=f X λ(v=velocity,f =frequency, λ is wavelength)

Page 8: DOPPLER ECHOCARDIOGRAPHY-1

• Velocity of sound=1540 m/sec in soft tissue

• Wavelength=1.54/f

• AmplitudeMeasure of strength of the sound waveIndicated by height of sine wave above and below baseline

Page 9: DOPPLER ECHOCARDIOGRAPHY-1
Page 10: DOPPLER ECHOCARDIOGRAPHY-1
Page 11: DOPPLER ECHOCARDIOGRAPHY-1

• Higher the frequency greater the resolution

• Higher frequency,lesser the penetration

• Loss of ultrasound as it propogates through a medium is called attenuation

Page 12: DOPPLER ECHOCARDIOGRAPHY-1

PRICIPLES OF PEIZO ELECTRIC CRYSTALS

The charges in a piezoelectric crystal are exactly balanced, even if they're not symmetrically arranged.

The effects of the charges exactly cancel out, leaving no net charge on the crystal faces

the electric dipole moments—vector lines separating opposite charges—exactly cancel one another out.

If you squeeze the crystal , you force the charges out of balance.

Page 13: DOPPLER ECHOCARDIOGRAPHY-1

• Now the effects of the charges (their dipole moments) no longer cancel one another out and net positive and negative charges appear on opposite crystal faces.

• By squeezing the crystal, voltage is produced across its opposite faces- piezoelectricity

• The piezoelectric effect was discovered in 1880 by two French physicists, brothers Pierre and Paul-Jacques Curie, in crystals of quartz, tourmaline, and Rochelle salt (potassium sodium tartrate). They took the name from the Greek work piezein, which means "to press."

Page 14: DOPPLER ECHOCARDIOGRAPHY-1

• The phenomenon of generation of a voltage under mechanical stress is referred to as the direct piezoelectric effect

• mechanical strain produced in the crystal under electric stress is called the converse piezoelectric effect.

Page 15: DOPPLER ECHOCARDIOGRAPHY-1

• Ferro electrics,barium tianate,lead zirconate titanate are used as peizo electric crystals.

• Dampening material-shortens the ringing responseAlso absorbs backward and laterally transmitted acoustic energy

• Frequency emitted by transducer is directly proportional to propagation speed within crystal and inversely related to thickness

Page 16: DOPPLER ECHOCARDIOGRAPHY-1

• Important feature of ultrasound is ability to direct or focus the beam as it leaves the transducer

• Proximal cylindrical and distally divergent• Proximal zone –Fresnel zone• Divergent field is called Fraunhofer zone• Imaging is optimal in near field• Decreasing wavelength or increasing transducer size

increase near field

Page 17: DOPPLER ECHOCARDIOGRAPHY-1
Page 18: DOPPLER ECHOCARDIOGRAPHY-1

Haemo”dynamics”

• Blood flow is a complex phenomenon• Not a uniform liquid• Flow pulsatile• Vessel walls are elastic

Page 19: DOPPLER ECHOCARDIOGRAPHY-1

Properties of Blood

• Density-mass of blood per unit volume• Measure of resistance to accelaration• Greater the density,greater the resistance to

flow• Viscosity:resistance to flow offered by fluid in

motion• 0.035 poise at 37 degree.

Page 20: DOPPLER ECHOCARDIOGRAPHY-1
Page 21: DOPPLER ECHOCARDIOGRAPHY-1

Factors determining flow

• Flow rate is determined by– Pressure gradient– Resistance • Viscosity of blood• Radius of lumen• Length of vessel

Page 22: DOPPLER ECHOCARDIOGRAPHY-1

Types of flow

Laminar flowShape of parabolaConcentric layers,each parallel to vessel wallVelocity of each layer differsMaximal velocity is at centre of vesselDecreasing profile towards peripheries

Page 23: DOPPLER ECHOCARDIOGRAPHY-1
Page 24: DOPPLER ECHOCARDIOGRAPHY-1
Page 25: DOPPLER ECHOCARDIOGRAPHY-1
Page 26: DOPPLER ECHOCARDIOGRAPHY-1

Turbulent flow

• Obstruction produce increased velocities, flow vortices• Whirlpools shed off in different directions producing

variable velocities- chaos• Predicted by Reynolds number• Reynolds number depends on

Re=( ρ x c x D)/v ρ-Density of bloodD-Vessel diameterc-Velocity of flowV-viscosity

Page 27: DOPPLER ECHOCARDIOGRAPHY-1

The Reynolds number is dimensionless

If Re is less than 1200 the flow will be -laminar

1200-2000 flow is described as -transitional

Greater than 2000 -turbulent

Page 28: DOPPLER ECHOCARDIOGRAPHY-1

Doppler Principle

Page 29: DOPPLER ECHOCARDIOGRAPHY-1

First described by Johann Christian Doppler, an Austrian mathematician and scientist who lived in the first half of the19th century.Doppler’s initial descriptions referred to changes in the wavelength of light as applied to astronomical events.

In 1842, he presented a paper entitled "On the Coloured Light of Double Stars and Some Other Heavenly Bodies" where he postulated that certain properties of light emitted from stars depend upon the relative motion of the observer and the wave source.

Page 30: DOPPLER ECHOCARDIOGRAPHY-1

• Doppler effect describes the frequency shift of the signal in relation to the relative motion of a source and an observer.

• The wave generated by a source that moves away from an observer/receiver appears to him to be of lower frequency than the wave generated by a stationary source, or generated by a source moving toward the observer.

• The frequency of the signal detected by the receiver moving toward the still source is higher, compared to the frequency detected by the still receiver, or a receiver moving away from the source.

Page 31: DOPPLER ECHOCARDIOGRAPHY-1

• Applied in echocardiography to determine flow direction,flow velocities,flow characteristics

• Stationary rbc-zero doppler shift(received frequency= transmitted frequency)

• Positive doppler shift-RBCs moving towards transducer ,received frequency >transmitted frequency

• Negative doppler shift:RBC’s moving away from transducer- transmitted frequency more than receiving frequency

Page 32: DOPPLER ECHOCARDIOGRAPHY-1
Page 33: DOPPLER ECHOCARDIOGRAPHY-1

• Doppler shift represents difference between received and transmitted frequencies ,which occur due to motion of RBC’s relative to the ultrasound beam

• Fd = (2f V cos Ø)/C

Page 34: DOPPLER ECHOCARDIOGRAPHY-1

Why the factor 2?

• Double doppler shift

1st shift-transducer stationary source,RBC the moving receiver

2nd shift is when,RBCs are moving source and transducer is the stationary receiver.

Page 35: DOPPLER ECHOCARDIOGRAPHY-1

Doppler equation- rearranged

Page 36: DOPPLER ECHOCARDIOGRAPHY-1

Factors affecting doppler equation

• Estimation of blood flow velocity is dependent on incident angle between ultrasound beam and blood flow

• When RBCs parallel-maximum velocity• When RBCs perpendicular-no doppler shift• When angle between ultrasound beam and blood

flow is less than or equal 20 degree,cosine close to 1 and percent error is less than or equal to 7%

Page 37: DOPPLER ECHOCARDIOGRAPHY-1

The Effect of Angle

Page 38: DOPPLER ECHOCARDIOGRAPHY-1

Angle Cosine Percentage error0 1 010 0.98 220 0.94 730 0.87 1360 0.5 5090 0 100

Page 39: DOPPLER ECHOCARDIOGRAPHY-1

Angle correction

• It is possible to correct for angle

• Not recommended as in most cases its possible to align ultrasound beam parallel by utilising multiple views, serial assessment difficult unless same angle correction used

• It is assumed that angle between ultrasound beam and direction of blood flow is parallel

Page 40: DOPPLER ECHOCARDIOGRAPHY-1

• By adjusting according to the direction of assumed flow, it changes the angle calculations in the Doppler equation resulting in different estimates of flow velocity.

• The use of this control does not actually change the direction of the Doppler beam and its use does not alter the quality of either the audio output or the spectral recording

Page 41: DOPPLER ECHOCARDIOGRAPHY-1

Effect of frequency

• Lower the frequency,higher the velocity detected

• A 2 MHz transducer detects higher velocity compared to a 5 MHz transducer

Page 42: DOPPLER ECHOCARDIOGRAPHY-1

SPECTRAL DOPPLER DISPLAY

Flow velocity

• Displayed on y axis

• Velocity of RBCs within sampled volume is calculated

• Absence of velocity-zero baseline

Page 43: DOPPLER ECHOCARDIOGRAPHY-1

Spectral velocity recordings

Page 44: DOPPLER ECHOCARDIOGRAPHY-1

Direction of flow

• Flow direction also displayed on Y axis

• Positive doppler shift-flow towards transducerTraditionally displayed above baseline

• Negative doppler shift-flow away from transducerDisplayed below zero baseline

Page 45: DOPPLER ECHOCARDIOGRAPHY-1

Intensity or amplitude

• Blood cells do not move at equal velocities• Produce different frequency shifts• Amplitude or intensity of doppler signal reflects

number of blood cells moving within a range of velocities at a particular point of time

• Bright signal-strong doppler shift frequency at a particular point of time .

• Darker regions-weak doppler shift

Page 46: DOPPLER ECHOCARDIOGRAPHY-1

Timing

• Time is displayed along x axis

• Displayed along with ECG.

• Change in blood velocity,flow direction can be accurately timed in relation to cardiac cycle.

Page 47: DOPPLER ECHOCARDIOGRAPHY-1

Doppler Audio signals

• Doppler shift frequencies are in audible range

• Guide for localising blood flow and for proper aligning ultrasound beam parallel to flow

• Laminar flow-smooth tone

• Turbulent flow-harsh sound.

Page 48: DOPPLER ECHOCARDIOGRAPHY-1

Pulsed and Continuous Wave Doppler

Page 49: DOPPLER ECHOCARDIOGRAPHY-1
Page 50: DOPPLER ECHOCARDIOGRAPHY-1

Continuous Wave Doppler

• older and electronically more simple• continuous generation of ultrasound waves• continuous ultrasound reception• two crystal transducer• Blood flow along entire beam is observed

Page 51: DOPPLER ECHOCARDIOGRAPHY-1
Page 52: DOPPLER ECHOCARDIOGRAPHY-1

• ADVANTAGEability to measure high blood velocities accurately

• DISADVANTAGE1)lack of selectivity or depth discrimination2)no provision for range gating to allow selective placing of a given Doppler sample volume in space

Page 53: DOPPLER ECHOCARDIOGRAPHY-1

Pulsed Wave Doppler

• Ultrasound impulses are sent out in short bursts or pulses

• transducer that alternates transmission and reception of ultrasound

• ability to provide Doppler shift data selectively from a small segment along the ultrasound beam- sample volume can be selected.

Page 54: DOPPLER ECHOCARDIOGRAPHY-1

• The transducer functions as receiver for a limited time period

• Time corresponds to the interval required for sound to return from specified area.

• Another burst of sound waves are not transmitted until previous impulses are received.

• Pulse repetition frequency (PRF)–frequency at which transducer transmits pulses.

• PRF determines sampling rate.

Page 55: DOPPLER ECHOCARDIOGRAPHY-1
Page 56: DOPPLER ECHOCARDIOGRAPHY-1

Inability to accurately measure high blood flow velocities- aliasing

“Alias” means false

Page 57: DOPPLER ECHOCARDIOGRAPHY-1

Fig.1.24

Aliasing

• The aliasing phenomenon occurs when the velocity exceeds the rate at which the pulsed wave system can record it properly

Page 58: DOPPLER ECHOCARDIOGRAPHY-1
Page 59: DOPPLER ECHOCARDIOGRAPHY-1

Aliasing is represented on the spectral trace as a cut-off of a given velocity with placement of the cut section in the opposite

channel or reverse flow direction

Page 60: DOPPLER ECHOCARDIOGRAPHY-1

Nyquist Limit

• The Nyquist limit defines when aliasing will occur using PW Doppler.

• The Nyquist limit specifies that measurements of frequency shifts (and, thus, velocity) will be appropriately displayed only if the pulse repetition frequency (PRF) is at least twice the maximum velocity (or Doppler shift frequency) encountered in the sample volume.

Page 61: DOPPLER ECHOCARDIOGRAPHY-1
Page 62: DOPPLER ECHOCARDIOGRAPHY-1

The Nyquist Limit The simplest sound wave is an oscillation between two amplitudes. A sampled waveform thus needs at least two sample points per cycle. Thus the wave's frequency must not be above half the sampling frequency. This limit is called the Nyquist limit of a given sampling frequency

Page 63: DOPPLER ECHOCARDIOGRAPHY-1

• Shannon's sampling theorem(Claude E. Shannon, born 1916, American mathematician) Also known as the Nyquist criterion, a general "rule" for measurement of frequencies, stating that the measurement (sampling) frequency must be at least twice the maximum frequency to be measured. Whenever Shannon's sampling theorem is not fulfilled, aliasing occurs

Page 64: DOPPLER ECHOCARDIOGRAPHY-1

• Nyquist limit specifies the maximum velocity that can be recorded without aliasing.

Page 65: DOPPLER ECHOCARDIOGRAPHY-1

Avoiding aliasing

Increase the Nyquist limit-1)altering variables in Doppler equation 2)high PRF mode3 )Change from PW to CW

Page 66: DOPPLER ECHOCARDIOGRAPHY-1

• V = C × PRF4 f COS Ø

(PRF= Δf× 2= 2f V cos Ø × 2) C

Max velocity can be increased by1)Increasing PRF2)Decreasing transmitted frequency3)Increasing speed of sound in tissue4)Decreasing cosØ

Page 67: DOPPLER ECHOCARDIOGRAPHY-1

High PRF Doppler

Page 68: DOPPLER ECHOCARDIOGRAPHY-1

• It is desirable to use as high a PRF as possible for recording abnormally elevated velocity jets.

• Maximum PRF is limited by the distance the sample volume is placed into the heart.

Page 69: DOPPLER ECHOCARDIOGRAPHY-1

Range ambiguity

• some of the range selectivity used in precisely locating the sample volume is lost.

• pulsing sequence is carried on over and over, some data is returned to the transducer

• data from all these volumes are added together

Page 70: DOPPLER ECHOCARDIOGRAPHY-1

Baseline shift ("zero shift" or "zero off-set" )

Electronic cut and paste

Moves the aliased doppler signal upward or downward(unwrapping)

Repositioning baseline effectively increases the maximum velocity at the expense of other direction.

Page 71: DOPPLER ECHOCARDIOGRAPHY-1

"baseline shift"

Page 72: DOPPLER ECHOCARDIOGRAPHY-1

Pulse Wave VS Continuous Wave Doppler

CW PWDepth resolution no yesSample volume large smallHigh velocity detection yes noAliasing no yes Sensitivity more

lessControl of sample volume placement poor

good

Page 73: DOPPLER ECHOCARDIOGRAPHY-1

Continuity equation

• The continuity equation states that the amount of blood flow through one cardiac chamber (or valve orifice) is the same as the blood flow through the other chambers and orifices

• It is based on the principle of conservation of mass.

• “Whatever mass flows in must flow out.”

Page 74: DOPPLER ECHOCARDIOGRAPHY-1

The Continuity Equation

Page 75: DOPPLER ECHOCARDIOGRAPHY-1

VELOCITY TIME INTEGRAL• Calculation of volumetric flow is complex as

flow velocity is not constant• Blood flow is pulsatile• Hence integrated velocity over time is taken• VTI is equal to area under curve cm/sec X sec.• Measure of distance that blood moves with

each heart beat.• VTI is also referred to as “stroke distance”

Page 76: DOPPLER ECHOCARDIOGRAPHY-1
Page 77: DOPPLER ECHOCARDIOGRAPHY-1

• Volumetric flow or stroke volume can be thus calculated as

SV=CSA X VTIContinuity equation can be rewritten as-CSA1 X VTI 1=CSA2 X VTI 2

CSA2 =(CSA1 x VTI 1) /VTI 2

Page 78: DOPPLER ECHOCARDIOGRAPHY-1

• VTI is obtained by tracing the leading edge of velocity spectrum.

Page 79: DOPPLER ECHOCARDIOGRAPHY-1

Clinical Applications

• Calculation of valve areas

• Calculation of regurgitant volumes and fractions

• Calculation of regurgitant orifice areas

• Calculation of intra cardiac shunt ratios

Page 80: DOPPLER ECHOCARDIOGRAPHY-1

LIMITATIONS OF CONTINUITY PRINCIPLE

Erroneous determination of CSA• Measurement of diameter during wrong phase of

cardiac cycle.• Inconsistent annulus measurements

Erroneous determination of VTI• Incorrect placement of pulse wave sample volume• Significant angle between doppler beam and blood

flow• Incorrect filter/gain settings

Page 81: DOPPLER ECHOCARDIOGRAPHY-1

• Bernoulli's principle is named after Swiss mathematecian Daniel Bernoulli who published his principle in his book Hydrodynamica in 1738

Page 82: DOPPLER ECHOCARDIOGRAPHY-1

Bernoulli equation• Bernoulli described the conversion of energy in a fluid from one

form to another, as occurs when fluid flow in a tube that suddenly either increases or decreases its diameter.

• Bernoulli's law states that total energy at all points along a tube is the same—conservation of energy.

• Energy (the ability to do work) is composed of pressure energy (pressure x the volume of fluid),kinetic energy (fluid in motion has kinetic energy ,which is proportional to the mass and the velocity squared; mass is density x volume), and gravitational energy (the product of density, volume, height above a surface, and the gravitational constant).

Page 83: DOPPLER ECHOCARDIOGRAPHY-1
Page 84: DOPPLER ECHOCARDIOGRAPHY-1
Page 85: DOPPLER ECHOCARDIOGRAPHY-1

Limitations of Bernoulli Equation

• Significant flow acceleration-prosthetic valves

• Significant viscous forces-muscular VSD,tunnel subaortic stenosis

• Increased proximal velocity-AS+AR,AS + HOCM

• Altered blood viscosity-polycythemia

Page 86: DOPPLER ECHOCARDIOGRAPHY-1

Discrepancies between Catheter derived and Doppler derived Pressure Gradients

• Pressure gradients derived at cardiac cathetrisation and measured by Doppler may differ.

• Catheter derived gradient is peak to peak gradient between LV and Aorta

• Non simultaneous measurement• Doppler gives peak instantaneous gradient

and is greater than peak to peak gradient.

Page 87: DOPPLER ECHOCARDIOGRAPHY-1

PRESSURE RECOVERY PHENOMENON

• Complex hemodymanic concept- pressure of fluid decreases as velocity increases.

• Once flow passes through a narrowing pressure drops and increases towards original value

• Rate and magnitude of pressure recovery is variable• In prosthetic valves,3 effective orifices-2 large orifices

by sides and a central small orifice.• Maximum velocity and lowest pressure is at

narrowest orifice.

Page 88: DOPPLER ECHOCARDIOGRAPHY-1

• Immediately distal to orifice, pressure increases (recovers)and velocity decreases.

• Doppler gradients are measured at narrowest orifice, while catheter gradients are recorded downstream to prosthetic valve where the pressure has already recovered.

• Hence, Doppler derived pressure gradients are more compared to catheter derived pressure gradients.

Page 89: DOPPLER ECHOCARDIOGRAPHY-1
Page 90: DOPPLER ECHOCARDIOGRAPHY-1

Determination of Pressure Gradients

Pressure gradients are useful for assessment of severity of valvular stenosis and estimation of intracardiac pressures .

Pressure gradients commonly determined by Doppler include

1)Maximum instantaneous pressure gradient

2)Mean pressure gradient

Page 91: DOPPLER ECHOCARDIOGRAPHY-1

• Maximum Instantaneous pressure gradient =4v2

(v=peak velocity)

• Mean pressure gradient cannot be obtained from the mean (or average) velocity, but must be calculated by making multiple instantaneous gradient (and hence pressure calculations) measurements, and then averaging those instantaneous pressures.

Page 92: DOPPLER ECHOCARDIOGRAPHY-1
Page 93: DOPPLER ECHOCARDIOGRAPHY-1

Estimation of RVSP by TR

• TR doppler signal represents the pressure difference between RV and RA during Systole

• RVSP-RAP=4(V TR 2 )

• In absence of RVOT obstructionRVSP=Pulmonary artery systolic pressure.

Page 94: DOPPLER ECHOCARDIOGRAPHY-1
Page 95: DOPPLER ECHOCARDIOGRAPHY-1

RVSP= BP SYSTLOLIC -4(V vsd )2

Page 96: DOPPLER ECHOCARDIOGRAPHY-1

PASP=BP SYSTOLIC -4(V PDA )2

Page 97: DOPPLER ECHOCARDIOGRAPHY-1

Mean Pulmonary Artery Pressure=4 (V PR-PEAK)2

Page 98: DOPPLER ECHOCARDIOGRAPHY-1

PAEDP=4 (V PR-ED)2 + RVEDP

Page 99: DOPPLER ECHOCARDIOGRAPHY-1

LAP= BP SYSTOLIC -4 (V MR)2

Page 100: DOPPLER ECHOCARDIOGRAPHY-1

LVEDP =BP DIASTOLIC - 4 (V AR-ED)2

Page 101: DOPPLER ECHOCARDIOGRAPHY-1

Pressure half-time anddeceleration time

The pressure half-time (PHT) is defined as the time (in milliseconds) required for the peak initial pressure to drop by one half .

Doppler signal measures velocity

The modified Bernoulli equation is applied to convert velocity to pressure.

PHT is the time (in milliseconds) for the velocity to drop to 0.707 of the maximum velocity . V2 = (0.707) x V1

Page 102: DOPPLER ECHOCARDIOGRAPHY-1
Page 103: DOPPLER ECHOCARDIOGRAPHY-1

The DT is the time required for the velocity,beginning with the peak value, when extrapolated, to cross the zero baseline

The PHT is clinically used most frequently in evaluating mitral stenosis and aortic regurgitation

Page 104: DOPPLER ECHOCARDIOGRAPHY-1
Page 105: DOPPLER ECHOCARDIOGRAPHY-1

MVA = 220/PHT

PHT = 0.29 xDT

MVA =220/0.29 x DT=759/DT

Page 106: DOPPLER ECHOCARDIOGRAPHY-1

Limitations of PHT in calculation of MVA

• Non linear early diastolic slope

• Post balloon mitral valvuloplasty

• Significant Aortic regurgitation

• Cardiac rhythm disturbances

Page 107: DOPPLER ECHOCARDIOGRAPHY-1

Non linear early diastolic slope

• Non linear or curvilinear decay –lead to erroneous calculation of PHT.

• Part of slope considered most representative should be chosen.

Page 108: DOPPLER ECHOCARDIOGRAPHY-1

Post balloon mitral valvuloplasty

• Accuracy of calculated MVA by PHT declines immediate post BMV.

• PHT is directly related to chamber compliance and peak transmitral gradient

• Following BMV,abrupt changes in left atrial pressure and compliance occur altering the relationship between PHT and Mitral valve area.

• This effect on PHT lasts 24 to 48 hours.

Page 109: DOPPLER ECHOCARDIOGRAPHY-1

Significant Aortic regurgitation

• Severe AR shortens PHT.• This is due to markedly elevated LVEDP which

reduces diastolic pressure gradient between LA and LV.

• Mitral valve area is thus overestimated.

Page 110: DOPPLER ECHOCARDIOGRAPHY-1

Cardiac rhythm disturbances

• Tachycardia-deceleration slope is shortens,PHT shortens and thus valve area is overestimated.

Page 111: DOPPLER ECHOCARDIOGRAPHY-1

Proximal Isovelocity Surface Area (PISA)

• Proximal Isovelocity Surface Area (PISA) method is based on the continuity equation.

• When a flow passes through a narrow orifice, as it approaches the narrowest region, there is a flow convergence and flow acceleration.

• PISA is the surface area of the hemisphere at the aliasing region of the flow convergence.

• PISA increases with lower aliasing velocity.

Page 112: DOPPLER ECHOCARDIOGRAPHY-1
Page 113: DOPPLER ECHOCARDIOGRAPHY-1
Page 114: DOPPLER ECHOCARDIOGRAPHY-1

• Radius is measured from the orifice to point of colour change.

• If the flow convergence is not a true hemisphere, the angle subtended by the flow convergence at the orifice has to be measured and divided by 180 to get a correction .

• Good correlation between angiographic estimates of regurgitant flow and PISA based estimates have been reported.

Page 115: DOPPLER ECHOCARDIOGRAPHY-1

Valve area calculation

A0=(2 Л r2x VN)/V0

A0=area of narrowed orifice

V0=velocity at narrowed orifice

R=radius of shellVN=aliased velocity identified as Niquist limit

Page 116: DOPPLER ECHOCARDIOGRAPHY-1

Angle correction

PISA principle is for flow approaching narrow planar surfaceIn Mitral Stenosis,mitral leaflets may be funnel shaped To account for altered shape,angle corection factor ά /180 is applied.MVA= ( ά /180 ) x (2 Л r2x VN)/VMS

Page 117: DOPPLER ECHOCARDIOGRAPHY-1
Page 118: DOPPLER ECHOCARDIOGRAPHY-1

Limitations of PISA method

Errors in measurement of radiusValve area is proportional to square of radius-even small errors are magnified.

Errors in measurement of angle Measurement is done offline using a protractor

Angle measured in one dimension may not be true representation of valve leaflet geometry.

Page 119: DOPPLER ECHOCARDIOGRAPHY-1

THANK YOU

Page 120: DOPPLER ECHOCARDIOGRAPHY-1

MCQs

Page 121: DOPPLER ECHOCARDIOGRAPHY-1

Maximum detectable velocity for a 5 MHz probe ,(assuming Doppler shift frequency as 5 KHz ,speed of sound in tissue as 1500m/sec and doppler beam being parallel to blood flow) ,would be

a)0.62 m/secb)0.75 m/secc)0.92 m/secd)1.0 m/sec

Page 122: DOPPLER ECHOCARDIOGRAPHY-1

Percentage error when ultrasound beam makes an angle of 20 degrees to blood column

a)3b)7c)15d)30

Page 123: DOPPLER ECHOCARDIOGRAPHY-1

When frequency of ultrasound probe is reduced,

a)Resolution increasesb)Maximum detectable velocity increasesc) Attenuation increasesd)All of the above

Page 124: DOPPLER ECHOCARDIOGRAPHY-1

Following statements are true about pulse wave doppler compared to continuous wave doppler except

a)Better depth resolutionb)Larger sample volumec)More problem of aliasingd) Less sensitivity

Page 125: DOPPLER ECHOCARDIOGRAPHY-1

Resistance to flow through a tube is directly proportional to

a)Fourth power of radiusb)Second power of radiusc)Length of tubed)None of the above

Page 126: DOPPLER ECHOCARDIOGRAPHY-1

Turbulent flow typically occurs when reynolds number just exceeds

a)800b)1000c)2000d)5000

Page 127: DOPPLER ECHOCARDIOGRAPHY-1

Pulse wave is preferred to continuous wave doppler to assess flow in all except

a)Pulmonary veinb)Superior venacavac)Pulmonary valved)Aortic valve

Page 128: DOPPLER ECHOCARDIOGRAPHY-1

Doppler shift is directly proportional to all excepta)Speed of sound in tissueb)Velocity of blood flowc)Frequency of transducerd)Cosine of angle between interrogating beam and

blood flow

Page 129: DOPPLER ECHOCARDIOGRAPHY-1

True about Mitral valve area assessment by PHT in severe AR is

a)PHT increasesb)Valve area is underestimatedc)MS severity is underestimatedd)Increases diastolic gradient between LA and

LV.

Page 130: DOPPLER ECHOCARDIOGRAPHY-1

Mean pulmonary artery pressure of a patient with TR jet velocity of 3 m/sec, Peak PR velocity of 2m/sec is (in mm Hg)

a)36b)16c) 9d)20

Page 131: DOPPLER ECHOCARDIOGRAPHY-1

KEY

Page 132: DOPPLER ECHOCARDIOGRAPHY-1

Maximum detectable velocity for a 5 MHz probe ,assuming Doppler shift frequency as 5 KHz ,speed of sound in tissue as 1500m/sec and doppler beam being parallel to blood flow, would be

a)0.62 m/secb)0.75 m/secc)0.92 m/secd)1.0 m/sec

Page 133: DOPPLER ECHOCARDIOGRAPHY-1

Percentage error when ultrasound beam makes an angle of 20 degrees to blood column

a)3b)7c)15d)30

Page 134: DOPPLER ECHOCARDIOGRAPHY-1

When frequency of ultrasound probe is reduced,

a)Resolution increasesb)Maximum detectable velocity increasesc) Attenuation increasesd)All of the above

Page 135: DOPPLER ECHOCARDIOGRAPHY-1

Following statements are true about pulse wave doppler compared to continuous wave doppler except

a)Better depth resolutionb)Larger sample volumec)More problem of aliasingd) Less sensitivity

Page 136: DOPPLER ECHOCARDIOGRAPHY-1

Resistance to flow through a tube is directly proportional to

a)Fourth power of radiusb)Second power of radiusc)Length of tubed)None of the above

Page 137: DOPPLER ECHOCARDIOGRAPHY-1

Turbulent flow typically occurs when reynolds number just exceeds

a)800b)1000c)2000d)5000

Page 138: DOPPLER ECHOCARDIOGRAPHY-1

Pulse wave is preferred to continuous wave doppler to assess flow in all except

a)Pulmonary veinb)Superior venacavac)Pulmonary valved)Aortic valve

Page 139: DOPPLER ECHOCARDIOGRAPHY-1

Doppler shift is directly proportional to all excepta)Speed of sound in tissueb)Velocity of blood flowc)Frequency of transducerd)Cosine of angle between interrogating beam and

blood flow

Page 140: DOPPLER ECHOCARDIOGRAPHY-1

True about Mitral valve area assessment by PHT in severe AR is

a)PHT increasesb)Valve area is underestimatedc)MS severity is underestimatedd)Increases diastolic gradient between LA and

LV.

Page 141: DOPPLER ECHOCARDIOGRAPHY-1

Mean pulmonary artery pressure of a patient with TR jet velocity of 3 m/sec, Peak PR velocity of 2m/sec is (in mm Hg)

a)36b)16c) 9d)20