DIGITAL HIGH-RESOLUTION HEART PHANTOMS Abstract Background: X-ray based imaging modalities employ ionizing radiation with the potential for causing harmful deterministic and stochastic effects to the patients. Accurate simulation to optimize image quality with respect to radiation dose requires detailed, sex-specific anthropomorphic phantoms with anatomically realistic heart and associated vasculature. Such phantoms are not widely available. This work describes our open-source heart-phantom development platform as well as example imaging simulations. Methods: To create a computerized phantom from a clinical coronary CT angiography (CTA), the relationship between the intensity distributions of the myocardium, the ventricles and the coronary arteries are identified via histogram analysis of the CTA images. By further refining the segmentation using operations such as active contours and vesselness, we are able to automatically generate high resolution volumetric heart models. The phantoms are obtained by fitting triangular meshes to the segmented data. A Graphical User Interface was developed to add virtual stenotic lesions to the segmented coronaries. The heart phantoms generated so far[1] have been cross-registered into the mesh-based Virtual Family[2] of phantoms. These phantoms can be used to obtain clinically realistic projection images of the heart using our Monte Carlo code penMesh[3]. Songxiang Gu, Andreu Badal and Iacovos Kyprianou Division of Imaging and Applied Mathematics/OSEL/CDRH/FDA Reference: [1] S. Gu, R. Gupta and I. Kyprianou, Computational High-Resolution Heart Phantoms for Medical Imaging and Dosimetry Simulations, Phys. Med. Bio., Accepted, 2011. [2] A. Christ and et al., The virtual family development of surface based anatomical models of two adults and two children for dosimetric simulations. Phys. Med. Bio., 55(2):23–38, 2010. [3] A. Badal, I. Kyprianou, A. Badano, and J. Sempau, penmesh–monte carlo radiation transport simulation in a triangle mesh geometry. IEEE Trans. Med. Imag., 28(12):1894–1901, 2009. The derived heart phantoms have high resolution coronary arteries, the myocardium, and the lumen in the ventricles. The phantom quality is sufficient for generating realistic simulated projections using Monte Carlo simulations. Adult Female Heart Phantom + + + = + + + = Adult Male Heart Phantom Stenose s · Histogram · Skeleton Extraction · Water-tight Mesh to Volume · Volume to Water-tight Mesh · Mesh + Volume View · Cutting Tool · Pick the Largest Part · File format change · …… ls0.asc ls3.asc ls1.asc ls2.asc ls4.asc Developed Heart Phantoms Other Heart Phantoms (a) Left and right coronary arteries; (b) Myocardium; (c) Ventricles and atriums. Stenosis Synthesis Body Phantoms with High-resolution Hearts Monte Carlo Simulations Female Heart Male Heart Female Heart with stenosis on left coronary artery Note: The material of coronary arteries is defined as blood with 10% iodine per volume and the ventricle material is defined as blood without iodine to make the vessels clearly identifiable. Conclusion Software Description Note: Two high-resolution heart phantoms, the source code of the related software, and the readme file are distributed freely at hades.googlecode.com . Female Body Phantom Male Body Phantom Automatic stenosis synthesis tool (the small box shown above) can generate stenoses with different morphology. More generated stenoses are shown to the right. The details of the synthesis strategy are described in [1]. (i) (ii) (iii) (iv) (v) (v) i (i) (ii) (iii) (iv) (v) (v) i (i) (ii) (iii) (iv) (v) (v) i (a) (b) (c)

Transcript of DIGITAL HIGH-RESOLUTION HEART PHANTOMS Abstract Background:X-ray based imaging modalities employ...

Page 1: DIGITAL HIGH-RESOLUTION HEART PHANTOMS Abstract Background:X-ray based imaging modalities employ ionizing radiation with the potential for causing harmful.

DIGITAL HIGH-RESOLUTION HEART PHANTOMS

AbstractBackground: X-ray based imaging modalities employ ionizing radiation with the potential for causing harmful deterministic and stochastic effects to the patients. Accurate simulation to optimize image quality with respect to radiation dose requires detailed, sex-specific anthropomorphic phantoms with anatomically realistic heart and associated vasculature. Such phantoms are not widely available. This work describes our open-source heart-phantom development platform as well as example imaging simulations.

Methods: To create a computerized phantom from a clinical coronary CT angiography (CTA), the relationship between the intensity distributions of the myocardium, the ventricles and the coronary arteries are identified via histogram analysis of the CTA images. By further refining the segmentation using operations such as active contours and vesselness, we are able to automatically generate high resolution volumetric heart models. The phantoms are obtained by fitting triangular meshes to the segmented data. A Graphical User Interface was developed to add virtual stenotic lesions to the segmented coronaries. The heart phantoms generated so far[1] have been cross-registered into the mesh-based Virtual Family[2] of phantoms. These phantoms can be used to obtain clinically realistic projection images of the heart using our Monte Carlo code penMesh[3].

Songxiang Gu, Andreu Badal and Iacovos Kyprianou

Division of Imaging and Applied Mathematics/OSEL/CDRH/FDA

Reference:[1] S. Gu, R. Gupta and I. Kyprianou, Computational High-Resolution Heart Phantoms for Medical Imaging and Dosimetry Simulations, Phys. Med. Bio., Accepted, 2011.[2] A. Christ and et al., The virtual family development of surface based anatomical models of two adults and two children for dosimetric simulations. Phys. Med. Bio., 55(2):23–38, 2010.[3] A. Badal, I. Kyprianou, A. Badano, and J. Sempau, penmesh–monte carlo radiation transport simulation in a triangle mesh geometry. IEEE Trans. Med. Imag., 28(12):1894–1901, 2009.

The derived heart phantoms have high resolution coronary arteries, the myocardium, and the lumen in the ventricles. The phantom quality is sufficient for generating realistic simulated projections using Monte Carlo simulations.

Adult Female Heart Phantom

+ + + =

+ + + =

Adult Male Heart Phantom

Stenoses

· Histogram· Skeleton Extraction· Water-tight Mesh to Volume· Volume to Water-tight Mesh· Mesh + Volume View· Cutting Tool· Pick the Largest Part· File format change · ……

ls0.asc

ls3.ascls1.asc

ls2.asc

ls4.asc

Developed Heart Phantoms

Other Heart Phantoms

(a) Left and right coronary arteries; (b) Myocardium; (c) Ventricles and atriums.

Stenosis Synthesis

Body Phantoms with High-resolution Hearts

Monte Carlo Simulations

Female Heart Male Heart Female Heart with stenosis on left coronary artery

Note: The material of coronary arteries is defined as blood with 10% iodine per volume and the ventricle material is defined as blood without iodine to make the vessels clearly identifiable.

Conclusion

Software Description

Note: Two high-resolution heart phantoms, the source code of the related software, and the readme file are distributed freely at hades.googlecode.com.

Female Body Phantom

Male Body Phantom

Automatic stenosis synthesis tool (the small box shown above) can generate stenoses with different morphology. More generated stenoses are shown to the right. The details of the synthesis strategy are described in [1].