Application Highlight
Features/Benefits
Overview
Related Product
ConnectPort™ TS 4x4
Digi Connect ES provides mission-critical serial over Ethernet connectivty. It is the first IEC 60601/EN60601 compliant device of its kind and consists of four, eight or 16 galvanic isolated RS-232 serial ports, with a 10/100 Mbps network interface and optional Ethernet switch. Applications include providing Ethernet connections from serial devices such as ventilators, EKGs, patient monitoring systems, infusion pumps and glucose meters to the central data management system.
Galvanic isolation provides extended electrical safety. There is no electrical path for current to earth ground, ensuring no electrical shock when making physical contact with the Digi Connect ES. There is no electrical path from port to port, ensuring a ground fault will not affect the operation of the Digi Connect ES or the operation of any device connected to it.
Digi’s patented RealPort technology makes it possible to establish a connection between the host and a networked serial device by creating a local COM or TTY port on the host computer. Incoming/outgoing and Telnet sessions on each port give system administrators a high level of control over networked serial devices.
Extended safety terminal server is designed for serial over Ethernet applications that require isolation on and between ports.
• 4/8/16-portRS-232serial-to-networkconnectivity
• IEC60601compliant
• Galvanicisolationonallports
• Wallmountable
• PatentedDigiRealPort®technologyprovides
localserialportfunctionality
• RealPortwithencryptionavailableforSSL-based
protectionforadditionalsecurity
• Portsmaybesharedbetweenmultiplehosts
runningdifferentoperatingsystems
Data Management
System
EKG
SphygmomanometerVentilator
Infusion Pump
RS-23298.74˙
98.74˙
Digi Connect® ES
Ethernet
Terminal Server
Digi Connect® ES(Extended Safety)
-
Features
Management HTTP/HTTPS, CLI or Telnet, Optional secure enterprise management via Digi Connectware® Manager
Protocols UDP/TCP, DHCP, Extended Telnet RFC 2217, Telnet, Reverse Telnet, Rlogin & Auto Connect, TFTP
Security SSL, SSL/TLS, SSHv2, FIPS 197 (serial ports), SNMPv2
Software Device-initiated patented RealPort COM port redirector and RFC 2217, Python scripting, Custom development environment
Status LEDs Power, Locator, Serial and Ethernet link/activity
Operating SystemsAIX, HP-UX, Linux®, SCO® OpenServer™ 5, SCO® OpenServer™ 6, Solaris™ Intel, Solaris™ SPARC, Windows® XP,
Windows Server® 2003, Windows Server® 2008, Windows Vista®
Dimensions (L x W x D) 9.30 in x 10.60 in x 2.10 in (23.5 cm x 26.9 cm x 4.20 cm)
Weight 3.0 lb (1.34 kg)
Interfaces
Serial
Ports 4/8/16 separately isolated RS-232 ports (2.5kV)
Throughput Up to 230 Kbps
Signal Support TXD, RXD, RTS, CTS, DTR, DSR, DCD
Ethernet
Ports 1 uplink, Optional 4-port switch
Physical Layer 10/100 Base-T separately isolated (1.5kV)
Mode Full or half duplex
Power Requirements
Power Supply Internal 100-240VAC, 50-60 Hz 0.2A max
Surge Protection4 kV burst (EFT) per-4-4, 2 kV surge per EN61000-4-4
4kV isolation input to output2kV surge per EN61000-4-5
Environmental
Operating Temperature 0° C to 55° C (32° F to 130° F)
Storage Temperature -30° C to +85° C (-22° F to +185° F)
Relative Humidity 5 to 95% (non-condensing)
Serial Port Protection (ESD) +15 kV human body model
Regulatory Approvals
Emissions EN55011, EN55022, AS/NZS CISPR 22, ICES-003, FCC P15 Subpart B Class A
Safety IEC/EN60950-1, CAN/CSA 22.2 60950-1-3, UL 60950-1, IEC 60601-1, UL 60601-1, EN60601-1, EN60601-1-2
Immunity EN55024, ICES-003
Specifications Digi Connect® ES