Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018...
Transcript of Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018...
![Page 1: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/1.jpg)
1
THERMAL ABLATION
TATO SYSTEM
Laura Sarti Product Specialist
![Page 2: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/2.jpg)
Definition
Thermal Ablation (TA) is the use of extreme hiperthermic temperatures to destroy diseased tissues.
T.A. is performed in Interventional Radiology as an alternative to the surgical treatment of tumors in:
2
![Page 3: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/3.jpg)
Different TA technologies
• Radiofrequency
• Microwaves
• Laser
• Cryoablation
33 3
![Page 4: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/4.jpg)
RF Current Flow
• RF Currents that goes through the tissue from the tip of the applicator to the ground pad.
Side effect: skin burning. 4 4
![Page 5: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/5.jpg)
MW Energy
Microwaves radiate only near the tip of the applicator.
An applied MW field causes water molecules to rotate bilions of time per second , which leads to
heat generation.
5
![Page 6: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/6.jpg)
Comparison between RF and MW thermal ablation
6
• Heat sink effect
• Ground pad required
• Difficult to control the increase
of ablation
• Less influenced by veins
• Ground pad is not required
• Predictable and controllable ablation
• Compatible with pacemaker
![Page 7: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/7.jpg)
7
TATO :
Multiprobe microwave
thermal ablation System
Rev. 09 del 27.10.2016
![Page 8: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/8.jpg)
8
Main features of the system
1 • No cooling system
2
• Ergonomic and easily transportable
3
• Multi-probes with detachable cable
![Page 9: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/9.jpg)
1. No cooling system
The NIC (Not Internally Cooled) coax needles have a high energy efficiency that allows operating with less MW power for treatment up to 15 minutes.
Cooling system not needed
Less complexity
Higher reliability
Greater safety
Better applicator handling
Reducing procedure dead time
Easy control of the procedure
9
![Page 10: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/10.jpg)
10
Switch Unit Roll Stand
2.Ergonomic and easily transportable
Main Unit- Microwave generator
Articulated Arm extendable up to 1,5 m
Tato cart version Tato travel version
Switch Unit Roll Stand
![Page 11: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/11.jpg)
11
Main unit Handle
On/off button
Touchscreen 7,1 inch
Yellow Led: lighting indicates that MWs are ON
Red Led: when lighting up, it indicates a fault
2,45 GHz
120 W total power
IEC standards
compliant
USB port
![Page 12: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/12.jpg)
Extendable arm and switch cover
12
Extendable up to 1.5 m
Sterile cover for arm
• Total power 120 W • 1 to 4 power outputs
![Page 13: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/13.jpg)
TATO1 vs TATO2
13
FEATURES TATO1 TATO2
TOTAL POWER 100W 120W
DATA LOGGING
DATA
DOWNLOAD
TRACK ABLATION
PROTOCOL
REFLECTED
POWER
MONITORING
SELECTION
NEEDLE GAUGE
![Page 14: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/14.jpg)
14
11 G
14 G
17 G
18 G
Multi-Gauge
![Page 15: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/15.jpg)
15
DIAMETER
G mm
3 mm
2,1 mm
1,5 mm
1,2 mm
11 G
14 G
17 G
18 G
The thinnest applicator currently
available in the world market
![Page 16: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/16.jpg)
Version with infusion tip
Holes 0,3 mm Ø
16
• 14 G and 17 G version • Applicator with 4 micro-holes in the tip for fluid infusion • A syringe can be used to inject fluids during the treatment
![Page 17: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/17.jpg)
17
• Simple radiating monopole
• High efficiency
• Lightweight
(~35g w/o cable, 50g w/cable)
• Detachable feeding cable:
advantage under CT scan
Main structural features
![Page 18: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/18.jpg)
Massa - Patient with HCC
Usability of TATO under CT-scan
Extendable arm
18
![Page 19: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/19.jpg)
19
Echogenic needle
![Page 20: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/20.jpg)
Video of needle insertion
20
![Page 21: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/21.jpg)
• All steel tips are tested with an applied force of 5kg.
• No fragile ceramic parts are used
• All probes are Teflon covered, non-sticky.
• Available in different lengths from 10 up to 50 cm.
• Centimeter marked cannula
21
Main structural features
![Page 22: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/22.jpg)
How to insert the needle
22
E
Lesion
Ablation
E
Insert the tip of the needle into the lesion and use length E parameter (indicated in the chart) as reference.
![Page 23: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/23.jpg)
23
Control of increasing ablation volume : By employing lower MW power, the procedure’ s safety is increased, thus eliminating a lot of bubble gas and ensuring a better control under ultrasound.
Gauge Max power W
11 60
14 40
17 30
18 15
Main functional features
In order to operate safely, antennas must be used respecting the assigned power limits:
![Page 24: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/24.jpg)
Increasing ablation
24
![Page 25: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/25.jpg)
TATO 2 HMI
25
![Page 26: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/26.jpg)
26
TATO2 HMI Home Screen
![Page 27: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/27.jpg)
Setting
27
Language
Date and time
LOG Download
Services
Output mating cycles :
1
4
3
2
![Page 28: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/28.jpg)
28
TATO2 Home Screen
![Page 29: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/29.jpg)
Possibility to insert Patient ID, Physician ID and note
29
![Page 30: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/30.jpg)
Needle choice
30
![Page 31: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/31.jpg)
Therapy Setting
31
![Page 32: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/32.jpg)
Therapy summary
32
![Page 33: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/33.jpg)
Therapy running
33
REFLECTED POWER COEFFICIENT OK : NEEDLE IS WORKING WELL AIR : NEEDLE IS IN THE AIR MISMATCH : NON-OPTIMAL FUNCTIONING OF NEEDLE NOT CONNECTED : NEDDLE IS NOT CONNECTED
![Page 34: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/34.jpg)
Therapy ended - Summary
34
TRACK
ABLATION
![Page 35: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/35.jpg)
Track Ablation
35
![Page 36: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/36.jpg)
Video for Track ablation
• 14 G, 60 W, 25 sec
36
![Page 37: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/37.jpg)
Ex – vivo and in – vivo charts of Tato System
37
![Page 38: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/38.jpg)
CHARTS TATO2
38
ABLATION ZONE EXAMPLES* – 1 APPLICATOR Ex – vivo Bovine Liver
11 G 60 W
Duration (min) 10:00 15:00
L (cm) 6,0 7,0
D( cm) 4,3 5,0
E (cm) 1,3 1,5
14 G 40 W
Duration (min) 5:00 10:00 15:00
L (cm) 5,0 5,4 6,2
D( cm) 2,6 4,0 4,5
E (cm) 0,9 1,0 1,1
17 G 30 W
Duration (min) 5:00 10:00 15:00
L (cm) 4,6 5,0 5,6
D( cm) 2,5 3,4 3,8
E (cm) 0,8 1,0 1,2
18 G 10 W 15 W
Duration (min) 5:00 10:00 5:00 10:00
L (cm) 2,0 2,6 2,3 3,3
D( cm) 1,4 1,7 1,5 2,3
E (cm) 0,4 0,6 0,6 0,6
![Page 39: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/39.jpg)
CHARTS TATO2
39
![Page 40: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/40.jpg)
CHARTS TATO2
40
![Page 41: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/41.jpg)
CHARTS TATO2
41
![Page 42: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/42.jpg)
E.g. HCC or liver metastasis
• Examples of MW applications in different liver tumors with same values of watt and time.
• HCC : «oven effect» due to the tissue’s consistency that is rich in water and to the encapsulated tumor’s morphology . The result is a bigger ablation.
• Metastasis: no «oven effect» because of the tissue has less water content and is not encapsulated. The result is a smaller ablation.
42
![Page 43: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/43.jpg)
Frequently used parameters In vivo most frequently used values of power (W) and time (min) • D = diameter
43
Diameter
of tumor
0,5 <D<1 1<D<2 2<D<2,8 2,8 <D<4 D>4
Gauge
11
60 W
Open
surgery
1 needle
10 min
2 needles
15 min
14
40 W
1 needle
10 min
1 needle
15 min
2 needles
15 min
3 needles
15min
4 needles
10 min
17
30 W
1 needle
10 min
2 needles
15 min
18
15W
1 needle
10 min
![Page 44: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/44.jpg)
Microwave ablation system Cross- reference
44
![Page 45: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/45.jpg)
Padova – TATO use in laparoscopy
• Lesion: 7 cm
• 2 x 14 G, 35 W, 15 min
• And Repositioning 1 x 14G, 35 W, 4.5 min
45
![Page 46: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/46.jpg)
46
Treatment on HCC (percutaneous)
Dimension of the tumor 5 X 3 cm Used 2 x 17G
![Page 47: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/47.jpg)
Padova – Open surgery
47
9 tumors in total RESECTION AND ABLATION Data of the first 4 ablations: 18 mm, 35W, 10min 19 mm, 35W, 2.3min and 40W, 1.30min 20 mm, 40W, 6min 18 mm, 40W, 6.2min
![Page 48: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/48.jpg)
Osteoid osteoma
48
Osteoid osteoma in vertebra 18G, 10W, 6min
![Page 49: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/49.jpg)
Aarau - Patient with kidney tumor
Multi-probe case Tumor size: 2.4 x 2.2 cm (+2 cm margin) Number of antennas: 4 x 14G Power: 100W total Time: 15min
49
![Page 50: Diapositiva 1 - amt-medi.com · Title: Diapositiva 1 Author: certificazioni Created Date: 5/21/2018 12:45:40 PM](https://reader035.fdocuments.us/reader035/viewer/2022071213/602dc8c0442eec52810f094b/html5/thumbnails/50.jpg)
50