Deviating Samples UKAS TPS 63 Tim Platt Technical Manager.

14
Deviating Samples UKAS TPS 63 Tim Platt Technical Manager

Transcript of Deviating Samples UKAS TPS 63 Tim Platt Technical Manager.

Deviating SamplesUKAS TPS 63

Tim PlattTechnical Manager

Introduction

• EA LC identified ISO/IEC 17025 REQUIREMENTS not being met Clause 5.8 and 5.10

• Integrity of test samples maintained• Results reported – Accurately– Clearly– Unambiguously– Objectively– Include comments on sample quality or adequacy

that may have compromised the result

Terminology• Deviating – an abnormality or departure from what is

expected• Deviating samples– Those not correctly cared for or sampled

inappropriately.

Deviating Samples

• Exceeded their holding time• Lack pertinent information– Date– Time– Location– Identity of sampler/contact

• Not retained at appropriate temperature

Deviating Samples

• Presented in inappropriate containers /packaging

• Inappropriate headspace• Insufficient quantity• Denatured through heat/ light/ humidity• Suffered microbiologically

UKAS TPS 63

• 4.1 Policies and procedures in place – Sample integrity maintained– Ensure results reported accurately– Include comments on quality /adequacy

where this may have impacted on results

UKAS TPS 63

• 4.2 Laboratory to define “deviating sample”– Holding time (stability)– Storage /transit conditions– Pertinent sample information (date/time/identity)– Use of appropriate containers– Use of preservatives– Sample type as defined in accreditation schedule

Holding timesParameter Sample Size

(ml)Holding time (days)BS/ISO 5667-3

Holding time (days) USEPA

Metals 100 28 180

BOD 250 1 2

Total and free Cyanide

100 7 14

Sulphide 100 7 7

PAHs 500 7 7 pre 40 post extraction

VOCs 2x60 1 14

pH 50 6Hours Immediate

Sample Containers

• Waters – Volatiles ,BTX, GRO – 60ml amber glass

containers*– Other organics 1x 1 litre amber glass container*– Inorganics 1x1 litre plastic container– Sulphide, Phenols, Cyanide, Metals , individual

plastic containers with preservative

– * no headspace

Preservatives

• Analyte - Preservative• Cyanide - Sodium hydroxide• Phenol - Sodium hydroxide• Metals - Nitric acid• Mercury - Potassium dichromate• BOD - Unpreserved, no headspace• Sulphide - Zinc acetate

Temperature control• Volatile and Organic samples or those affected

by biological activity should be stored at 5°C• Samples should be stored in the dark• MCERTS for waters “the sample storage

environment shall maintain a temperature of 5 ± 3°C”

• Cool transportation – cool boxes, ice packs, fill box, pre-chill, min/max thermometer.

UKAS TPS 63

• 4.3 On identifying a deviating sample.–Consistent application/actions taken–Consult with customer and record– Initiate investigation into non- conforming

work procedures if allowed to deviate in laboratory–Report results in appropriate manner.

UKAS TPS 63

• 4.4 Laboratory required to demonstrate– Sample assessed upon receipt– If found to be deviating – contact customer

for instructions– If customer requests to proceed, report will

include a disclaimer indicating deviating sample and results may be invalid– If deviation occurred after receipt –

investigation required

Deviating SamplesUKAS TPS 63

Tim PlattTechnical Manager