Detecting Electrons: CCD vs Film
Practical CryoEM Course
July 26, 2005
Christopher Booth
![Page 2: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/2.jpg)
Overview
• Basic Concepts
• Detector Quality Concepts
• How Do Detectors Work?
• Practical Evaluation Of Data Quality
• Final Practical Things To Remember
![Page 3: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/3.jpg)
Basic Concepts
• Fourier Transform and Fourier Space
• Convolution
• Transfer Functions– Point Spread Function– Modulation Transfer Function
• Low Pass Filter
![Page 4: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/4.jpg)
Fourier Transform
The co-ordinate (ω) in Fourier space is often referred to as spatial frequency or just frequency
![Page 5: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/5.jpg)
Graphical Representation Of The Fourier Transform
![Page 6: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/6.jpg)
Convolution
![Page 7: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/7.jpg)
Convolution In Fourier Space
• Convolution in Real Space is Multiplication in Fourier Space
• It is a big advantage to think in Fourier Space
![Page 8: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/8.jpg)
Low Pass Filter
• Reducing or removing the high frequency components
• Only the low frequency components are able to “pass” the filter
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
x =
![Page 9: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/9.jpg)
Transfer Functions
• A transfer function is a representation of the relation between the input and output of a linear time-invariant system
• Represented as a convolution between an input and a transfer function
dyyxtyfxf
xtxfxf
inputoutput
inputoutput
)()()(
)()()(
![Page 10: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/10.jpg)
Transfer Functions
• In Fourier Space this representation is simplified
)()()( sTsXsX inputoutput
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
x =
0
0.5
1
1.5
2
2.5
3
0 5 10 15 20 25
0
0.5
1
1.5
2
2.5
3
0 5 10 15 20 25
![Page 11: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/11.jpg)
Point Spread Function (PSF)
• The blurring of an imaginary point as it passes through an optical system
• Convolution of the input function with a
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25 0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
![Page 12: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/12.jpg)
Modulation Transfer Function (MTF)
• A representation of the point spread function in Fourier space
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
x =
0
0.5
1
1.5
2
2.5
3
0 5 10 15 20 25
0
0.5
1
1.5
2
2.5
3
0 5 10 15 20 25
![Page 13: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/13.jpg)
Summarize Basic Concepts
• Fourier Transform and Fourier Space• Convolution describes many real processes • Convolution is intuitive in Fourier Space• Transfer Functions are multiplication in Fourier
Space• MTF is the Fourier Transform Of the PSF• MTF is a Transfer Function• Some Filters are easiest to think about in Fourier
Space
![Page 14: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/14.jpg)
Detector Specific Concepts
• Nyquist Frequency
• Dynamic Range
• Linearity
• Dark Noise
![Page 15: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/15.jpg)
Nyquist Frequency
• Nyquist-Shannon Sampling Theorem
• You must sample at a minimum of 2 times the highest frequency of the image
• This is very important when digitizing continuous functions such as images
![Page 16: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/16.jpg)
Example Of Sampling Below Nyquist Frequency
![Page 17: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/17.jpg)
Quantum Efficiency
• The Quantum Efficiency of a detector is the ratio of the number of photons detected to the number of photons incident
![Page 18: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/18.jpg)
Dynamic Range
• The ratio between the smallest and largest possible detectable values.
• Very important for imaging diffraction patterns to detect weak spots and very intense spots in the same image
![Page 19: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/19.jpg)
Linearity
• Linearity is a measure of how consistently the CCD responds to light over its well depth.
• For example, if a 1-second exposure to a stable light source produces 1000 electrons of charge, 10 seconds should produce 10,000 electrons of charge
![Page 20: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/20.jpg)
Summarize CCD Specific Terms
• Nyquist Frequency, must sample image at 2x the highest frequency you want to recover
Quantum Efficiency (%)
Dynamic Range
Linearity
CCD 50 – 90 10,000 Very linear
Film 5 – 20 100 Limited linearity
![Page 21: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/21.jpg)
So Why Does Anyone Use Film?
• For High Voltage Electron Microscopes, the MTF of Film is in general better than that of CCD at high spatial frequencies.
• If you have an MTF that acts like a low pass filter, you may not be able to recover the high resolution information
![Page 22: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/22.jpg)
How a CCD Detects electrons
![Page 23: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/23.jpg)
Electron Path After Striking The Scintillator
100 kV 200 kV 300 kV 400 kV
![Page 24: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/24.jpg)
How Readout Of the CCD Occurs
![Page 25: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/25.jpg)
How Film Detects Electrons
Silver Emulsion
Film
Incident electrons
![Page 26: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/26.jpg)
Silver Grain Emulsion At Various Magnification
![Page 27: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/27.jpg)
How Film Is Scanned
Developed Silver Emulsion
Film
Incident Light
Scanner CCD Array
![Page 28: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/28.jpg)
Options For Digitizing Film
![Page 29: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/29.jpg)
Summary Of Detection Methods
• Scintillator and fiber optics introduce some degredation in high resolution signal in CCD cameras
• Film + scanner optics introduce a negligible amount of degredation of high resolution signal
![Page 30: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/30.jpg)
Practical Evaluation Of The CCD Camera
![Page 31: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/31.jpg)
Decomposing Graphite Signal
0
0.5
1
1.5
2
2.5
3
0 5 10 15 20 25
0
0.2
0.4
0.6
0.8
1
1.2
0 5 10 15 20 25
0
0.5
1
1.5
2
2.5
3
0 5 10 15 20 25-1.5
-1
-0.5
0
0.5
1
1.5
0 5 10 15 20 25
x x
![Page 32: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/32.jpg)
Calculating Spectral Signal To Noise Ratio
• Signal To Noise Ratio is more meaningful if we think in Fourier Space
)(
)()()(
sNoise
sNoisesrumPowerSpectsSNR
![Page 33: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/33.jpg)
Calculating The Fourier Transform Of an Image
Also called the power spectrum of the image
Image Of Carbon Film• amorphous (non crystalline) specimen• not beam sensitive• common
![Page 34: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/34.jpg)
Power Spectrum Of Amorphous Carbon On Film and CCD
![Page 35: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/35.jpg)
Comparing The Signal To Noise Ratio From Film and CCD
![Page 36: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/36.jpg)
Film Vs CCD Head-To-Head
CCD Film
Linearity
Quantum Efficiency
Dynamic Range
MTF
![Page 37: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/37.jpg)
Calculating SNR for Ice Embedded Cytoplasmic Polyhedrosis Virus
![Page 38: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/38.jpg)
Reconstruction To 9 Å Resolution
![Page 39: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/39.jpg)
Confirming A 9 Å Structure
![Page 40: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/40.jpg)
Relating SNR(s) To Resolution
2/5 Nyquist Frequency
![Page 41: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/41.jpg)
Further Experimental Confirmation Of 2/5 Nyquist
Table 2: Comparison of Reconstruction Statistics between Several Different Ice Embedded Single Particles Collected On the Gatan 4kx4k CCD at 200 kV at the Indicated Nominal Magnification
ComplexNumber Of Particles
Nominal Microscope Magnification
Expected Resolution (Å) at
2/5 Nyquist
Final Resolution (0.5 FSC cutoff, Å)
Software Package For Reconstructi
on
CPV 5,000 60,000 9 9 SAVR
GroEL 8,000 80,000 6.8 7-8 EMAN
Ryr1 29,000 60,000 9 9.5 EMAN
Epsilon Phage 15,000 40,000 13.6 13 EMAN/SAVR
![Page 42: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/42.jpg)
Evaluate Your Data To Estimate The Quality Of Your Imaging
• You can use ctfit from EMAN to calculate a spectral signal to noise ratio– Built In Method– Alternate Method Presented Here
)(
)()()(
sNoise
sNoisesrumPowerSpectsSNR
![Page 43: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/43.jpg)
Final Practical Things to Remember…
• Good Normalization Means Good Data– Dark Reference– Gain Normalization– Quadrant Normalization
• Magnification Of CCD relative to Film
• Angstroms/Pixel
![Page 44: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/44.jpg)
Normalization
• Standard Normalization
• Quadrant Normalization
)(
)()()(
_
_
xI
xIxIxI
referencegain
referencedarkacquiredfinal
![Page 45: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/45.jpg)
Quadrant Normalization
![Page 46: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/46.jpg)
Dark Reference
![Page 47: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/47.jpg)
Gain Normalization
![Page 48: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/48.jpg)
How Do I Tell If Something Is Wrong?
![Page 49: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/49.jpg)
Magnification Of CCD relative to Film
• 2010F Mag x 1.38 = 2010F CCD Mag• 3000SFF Mag x 1.41 = 3000SFF CCD Mag
• This has to be calibrated for each microscope detector.
![Page 50: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/50.jpg)
How Do I Calculate Angstroms/Pixel?
• Å/pixel = Detector Step-Size/Magnification
• For a microscope magnification of 60,000 on the 3000SFF:
• Å /pixel = 150,000 Å / (microscope magnification x 1.41)• Å /pixel = 150,000 Å / (60,000 x 1.41)
Å /pixel = 1.77
![Page 51: Detecting Electrons: CCD vs Film Practical CryoEM Course July 26, 2005 Christopher Booth.](https://reader036.fdocuments.us/reader036/viewer/2022062314/56649de45503460f94adabc7/html5/thumbnails/51.jpg)
Conclusion
• Understand what you are trying to achieve and use the detector that will make your job the easiest
• Check Your Own Data!