Transcript of Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of...
![Page 1: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/1.jpg)
Deep Temperature Deep Temperature MeasurementMeasurement
![Page 2: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/2.jpg)
RTM – 01 – RESRTM – 01 – RESImaging SystemImaging System
Items Specifications
Accuracy of measuring the averaged internal temperature, when a temperature is 32 ‑ 38 C, C
0,2
Time required for measuring internal temperature at a point, seconds
6
Antenna diameter, mm 39
Accuracy of measuring the skin temperature, C 0,2
Time required for measuring skin temperature at a point, when the temperature is 32 ‑ 38 C, seconds
1
Device mass, kg 4
Power consumption, Watt 5
![Page 3: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/3.jpg)
Temperature FieldTemperature Field
Results of the measurements can be displayed in different modesResults of the measurements can be displayed in different modes
![Page 4: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/4.jpg)
Three Types of AntennasThree Types of Antennas
![Page 5: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/5.jpg)
Main PrinciplesMain Principles
dVrPrTТ rad )(*)(
T(r)-Thermodynamic Temperature,r-Current CoordinateP(r)- The Radiometric Weighting Function
-The Vector of the Electric Field Intensity Induced in Tissue by the Antenna
-The Conductivity of the Tissueσ
dVrE
rErP
2
2
)(
)()(
(r)E
fkТP rad
KJ
k 23-10*1.38 BandFrequency_Δf
![Page 6: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/6.jpg)
Contradictions in the LiteratureContradictions in the Literature
Published Our ExperimentsCalculated penetration depth =6.6mm
For rectangular waveguide
at F=1.15 GHz a=44mm
More then 40mm
Penetration depth do not differ much with the kind of probe
Penetration depth of microstrip probe differs from cylinder waveguide probe substantially
The radiometric temperature increase mainly due skin temperature increase.
Experiments show that sometimes skin temperature is increase and
radiometric is decrease.
05.0_5.5
![Page 7: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/7.jpg)
Penetration Depth Penetration Depth for Plane Wavefor Plane Wave
![Page 8: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/8.jpg)
Contour Plot of the Penetration Contour Plot of the Penetration Depth[1]Depth[1]
a
mmDepth
mmDepth
mmaF
mmaF
4.3_
2.6
21_8.3
44_15.1_
akaK 2
1
00 )(
a
![Page 9: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/9.jpg)
RadiometricRadiometric Temperature for Layer Structure Temperature for Layer Structure
i
N
iirad CTТ *
1
U
dVrE
C iV
i
i
2)(
N
i V
i
i
dVrEU1
2)(
z
z
N
ii
N
ii
hz
CzD
1
1
)(
Ci-Weighting Coefficients for Layer Structure
D(z)- Integration Weighting Function for Layer Structure
![Page 10: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/10.jpg)
Dielectric Properties of the Dielectric Properties of the TissuesTissues
![Page 11: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/11.jpg)
Volume Under Investigation for Different Volume Under Investigation for Different Antennas (Level 85%)Antennas (Level 85%)
![Page 12: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/12.jpg)
Volume Under Investigation for Different Volume Under Investigation for Different Antennas (Level 85%)Antennas (Level 85%)
![Page 13: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/13.jpg)
Weighting Coefficients for Weighting Coefficients for Antenna F=1.15 GHz D=44mm H=67mmAntenna F=1.15 GHz D=44mm H=67mmCi
i
N
iirad CTТ *
1
![Page 14: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/14.jpg)
Weighting Coefficients for Weighting Coefficients for Antenna F=1.15 GHz D=38mm H=7mmAntenna F=1.15 GHz D=38mm H=7mm
Ci
![Page 15: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/15.jpg)
Weighting Coefficients for Weighting Coefficients for Antenna F=3.8 GHz D=38mm H=45mmAntenna F=3.8 GHz D=38mm H=45mm
Ci
![Page 16: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/16.jpg)
Integrated Weighting Functions Integrated Weighting Functions F=1.15 GHz D=44mm H=67mm Antenna F=1.15 GHz D=44mm H=67mm Antenna
0.0
0.2
0.4
0.6
0.8
1.0
0 10 20 30 40 50 60 70 80Depth, mm
Inte
gra
ted
Wei
gh
tin
g
Fu
nct
ion
h fat =8mm
h fat=28 mm
h fat=58mm
Level =0.85
Measuring DepthMeasuring Depth (D(Deffeff L=0.85L=0.85))
zN
iiCzD
1
)(
![Page 17: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/17.jpg)
Integrated Weighting Functions Integrated Weighting Functions F=1.15 GHz D=38mm H=7mm Antenna F=1.15 GHz D=38mm H=7mm Antenna
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0 10 20 30 40 50 60 70 80Depth, mm
Inte
gra
ted
Wei
gh
tin
g F
un
ctio
n D
(z)
h fat =8mm
h fat=28 mm
h fat=58mm
Measuring DepthMeasuring Depth (D(Deff L=0.85eff L=0.85))
Level=0.85
![Page 18: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/18.jpg)
Integrated Weighting Functions Integrated Weighting Functions F=3.8 GHz D=38mm H=45mm Antenna F=3.8 GHz D=38mm H=45mm Antenna
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0 10 20 30 40 50 60 70 80Depth, mm
Inte
gra
ted
We
igh
tin
g
Fu
ncti
on
h fat =8mm
h fat=28 mm
h fat=58mm
Level=0.85
Measuring Depth (DDepth (Deff L=0.85eff L=0.85))
![Page 19: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/19.jpg)
Measuring Depth (DMeasuring Depth (Deffeff L=0.85L=0.85) for Different Fat ) for Different Fat ThicknessThickness
0
10
20
30
40
50
60
70
80
90
0 10 20 30 40 50 60 70 80 90Fat Thickness, mm
Me
as
uri
ng
De
pth
(D eff
L=
0.8
5),m
m
F=1.15 GHz D=44mm H=67mmF=1.15 GHz D=38mm H=7 mmF=3.8 GHz D=38mm H=45mm
h skin=2 mm
![Page 20: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/20.jpg)
Measuring Depth (DMeasuring Depth (Deffeff L=0.85L=0.85))
0
10
20
30
40
50
60
70
F=1.15GHz D=38mmH=7mm
F=1.15 GHz D=44mmH=67mm
F=3.8 GHz D=38mmH=45mm
Mea
suri
ng
Dep
th (
mm
)
h fat=8mm
h fat=28mm
h fat=58 mm
![Page 21: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/21.jpg)
Effective Width for Different Fat ThicknessEffective Width for Different Fat Thickness
0
20
40
60
80
100
120
140
0 10 20 30 40 50 60 70
Fat Thickness, mm
Eff
ec
tiv
e W
idth
(W
eff),m
m
F=3.8 GHz D=38mm H=45mmF=1.15 GHz D=44mm H=67mmF=1.15 GHz D=38mm H=7 mm
h skin=2mm
)85.0()85.0(
2
4
)( LLeff
eff VDW
![Page 22: Deep Temperature Measurement. RTM – 01 – RES Imaging System ItemsSpecificatio ns Accuracy of measuring the averaged internal temperature, when a temperature.](https://reader033.fdocuments.us/reader033/viewer/2022061612/56649ed95503460f94be731b/html5/thumbnails/22.jpg)
Temperature Changes During the Cooling of DorsumTemperature Changes During the Cooling of Dorsum
32
32.5
33
33.5
34
34.5
35
35.5
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18Time, min
Tem
per
atu
re,
C
Palm F=1.15 GHz D=44mm H=67mm
Palm IK
Palm F=3.8 GHz,D=38mm,H=45mm