A New Rapid Acquisition Solid State Imaging System for Cardiac
SPECT Imaging
S.S. Gambhir M.D., Ph.D.Nuclear Medicine Grand Rounds
September 2007
![Page 2: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/2.jpg)
SPECT vs. PET
• SPECT has potential for multiplexing• SPECT can achieve very high spatial
resolution• SPECT isotopes have some advantages
but also key disadvantages• SPECT is limited by sensitivity and
inability to do dynamic rapid acquisition
![Page 3: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/3.jpg)
Gamma Camera/SPECT
GAMMA CAMERA (GC)
GC GC
COLLIMATOR
NaI CRYSTALS
PHOTOMULTIPLIER TUBES
TO COMPUTER
GAMMA RAYS
SUBJECT INJECTED WITH GAMMA-EMITTING PROBE
SCHEMATIC OF A GAMMA CAMERA
![Page 4: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/4.jpg)
Nuclear Cardiac Imaging
Stress
Rest
![Page 5: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/5.jpg)
Scintillation Cameras for Cardiac Imaging
A. Dual Head, Variable Angle
GE Infinia
B. Dual Head, Fixed 90o
GE Ventri
C. Compact Dual Head, Fixed 90o
Siemens C-Cam
![Page 6: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/6.jpg)
Current SPECT Technology
• Detectors– Dual-Head– NaI(Tl) Scintillation
Cameras– Fixed 90o geometry
• Data Acquisition– Rotation – 90o
– No. of frames – 60 – Acquisition – 180o
![Page 7: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/7.jpg)
Objectives of New SPECT System• Designed specifically for cardiac imaging• Improved energy resolution• Improved spatial resolution• Significant reductions in imaging times (x10)• Improved patient comfort
– Imaging in an upright or reclining position– Reduction of claustrophobia– Elimination of need for arms over head
• Small footprint to fit in ED, offices, exam rooms, etc.
![Page 8: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/8.jpg)
D-SPECT™ Cardiac ScannerSpectrum-DynamicsHaifa, Israel
Dynamic SPECT(D-SPECT)
![Page 9: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/9.jpg)
Detector Configuration
10 rotatingcolumns
![Page 10: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/10.jpg)
Detector Configuration
D-SPECT™ detector head
![Page 11: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/11.jpg)
D-SPECT Detector Configuration
QuickTime™ and aYUV420 codec decompressor
are needed to see this picture.
![Page 12: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/12.jpg)
64 elements~160 mm
16 elements~40 mm
2.46 mm
2.46 mm
Detector Column Detail
5 mm thick CZTDetector Element
1024 elementsper column
DetectorColumn
Column Collimator
Detector Array Direct conversion of photonenergy to voltage pulse
![Page 13: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/13.jpg)
Detector Characteristics
Detector ThicknessEffective
Atomic No.Density
(gm/cm3)Intrinsic Efficiency
(140 keV)
5 mm 88%
90%9.5 mm(3/8 in)
Energy Resolution(140 keV)
CZT 49 5.7 6%
NaI(Tl) 50 3.7 10%
![Page 14: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/14.jpg)
Rel
ativ
e co
unts
Tc-99mHALF LIFE = 6.02 Hours
140.5 KeV18.5 KeV
100
7
CZT Energy Resolution6%
NaI(Tl) Energy Resolution10%
![Page 15: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/15.jpg)
Sources of Improvements in Sensitivity
• Sensitivity– Use of wide angle collimators with high geometric efficiency– Unique ability to concentration acquisition on region-of-
interest (ROI-centric scanning)
![Page 16: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/16.jpg)
Sources of Improvementsin Spatial Resolution
• Spatial Resolution– Data collection from many angles– Decreased distance between detectors and ROI– Improved energy resolution– Use of proprietary BroadviewTM reconstruction
algorithms based on the ML-EM method
![Page 17: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/17.jpg)
D-SPECT Collimator
GE LEHR Collimator
Collimator HoleShape
Material Septum(mm)
Length(mm)
Hole Dia.(mm)
GeometricEfficiency
D-SPECT Square Tungsten 0.2 21.7 2.26 0.000748
GE LEHR Hex Lead 0.2 35 1.5 0.0000985
Ratio 7.6
Collimator Comparison
No lead x-rays to interfere with 201Tl imaging
![Page 18: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/18.jpg)
15 cm
7.5 cm
Spatial Resolution (mm)
Point # X Y
1-outer 4.7 5.2
2-outer 5.5 4.7
3-central 7.0 6.5
Average 5.7 5.5
13.612.63-central
12.813.4Average
12.412.82-outer
12.414.81-outer
YXPoint #
Spatial Resolution (mm)
2.11.83-central
2.32.3Average
2.62.32-outer
2.43.11-outer
YXPoint #
RatioA-SPECT/D-SPECT
15 cm
Spatial Resolution NEMA NU1 2001
A-SPECTD-SPECTAnger Camera SPECT
(A-SPECT)Dual Head 90O
60 frames, 180O
Circular orbitLEHR Collimators
![Page 19: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/19.jpg)
Resolution and SensitivityCo-57 line source 2.5 million counts
A-SPECT
1 mm
D-SPECT
Measured at 100 mm
System SpatialResolution
(mm)
Time(sec)
A-SPECT 10.4 600
D-SPECT 5.5 49
Ratio 1.9 12.3
![Page 20: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/20.jpg)
Gated Acquisition
• Acquired data are encoded during acquisition with EKG signal so that gated images may be reconstructed after data acquisition is complete
![Page 21: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/21.jpg)
Torso Phantom Studies
![Page 22: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/22.jpg)
D-SPECTA-SPECT
Anthropomorphic torso phantomActivity in liver, background, myocardium
No defect
D-SPECTA-SPECT
![Page 23: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/23.jpg)
D-SPECTA-SPECT
Anthropomorphic torso phantomActivity in liver, background, myocardium
20 mm; 50% activity defect
D-SPECTA-SPECT
![Page 24: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/24.jpg)
D-SPECTA-SPECT
Anthropomorphic torso phantomActivity in liver, background, myocardium
10 mm defect
D-SPECTA-SPECT
![Page 25: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/25.jpg)
Dual Isotope Phantom Studies
D-SPECT
SPECT
Virgin Thallium
Dual Thallium
Virgin Thallium
Dual Thallium
![Page 26: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/26.jpg)
10.58.47.0SensitivityGain (SD/GE)
*990,00095,000*920,000110,000*950,000136,000Counts/minfrom Myocardium
73.1%8.3%71.1%8.9%71%9.5%% MyocardiumCounts
1.981.18*1.831.43*1.901.70MyocardiumCounts (M)
*2.7114.26*2.6116.15*2.6817.08Total Counts (M)
212.5212.5212.5Duration (min)
10:339:2514:2511:5519:2117:08Start Acq
26.5 mCi at 08:4023.4 mCi at 11:1028.3 mCi at 16:13Administration
Raw Data
SDGESDGESDGE
Patient #130Patient #126Patient #125
*After decay correction
![Page 27: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/27.jpg)
47 y/o male, overweight, hypertensive,chest pain - Stress/Rest
AXIS, Philips Medical Systems
D-SPECT, Spectrum Dynamics
Rest2 min
Stress4 min
Rest11 min
Stress19 min
![Page 28: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/28.jpg)
44 y/o male, hypertensive, elevated cholesterol levels, asymptomatic - Stress/Rest
AXIS, Philips Medical Systems
D-SPECT, Spectrum Dynamics
Rest2 min
Stress4 min
Rest11 min
Stress19 min
![Page 29: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/29.jpg)
Future ProductsD-SPECT/CT
![Page 30: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/30.jpg)
Patient personalized RFID-tagged Syringe
Information transmitted to
D-SPECT Scanner; settings
adjusted automatically
Patient demographics acquired and transmitted to
®-Scan Wristband
Verification & Authentication
®-Scan Wristband:Patient demographics integrated
with dose and injection time
D-SPECT Cardiac Scanner System features the ®-SCAN data transfer system based on RFID technology.
![Page 31: D-SPECT: Design and Performance](https://reader038.fdocuments.us/reader038/viewer/2022102722/55616c2bd8b42a35458b56d9/html5/thumbnails/31.jpg)
Conclusions• D-SPECT is a new cardiac SPECT system that has
been developed using a radically new data acquisition technology. This system exhibits:– Factor of ~2 improvement in spatial resolution– Factor of ~10 improvement in sensitivity– Factor of ~1.7 improvement in energy resolution
• Systems will be in nuclear medicine laboratories in late 2007
• Clinical trials are just starting