PRESBYOND Laser Blended VisionCustomized. All distances. Immediate.
// PRESBYOND MaDE BY CaRl ZEiSS
The moment you bridge the gapto clear vision at all distances.This is the moment we work for.
PRESBYOND Laser Blended VisionA clear choice for presbyopic patients
PRESBYOND® Laser Blended Vision from Carl Zeiss is an advanced method for treating patients with age-related loss of accommodation, also known as presbyopia. It offers the opportunity to achieve freedom from glasses by combining the simplicity and accuracy of corneal refractive surgery with the benefits of increased depth of field in retaining visual quality. As a surgical solution based on the naturally occurring spherical aberrations of the eye, PRESBYOND Laser Blended Vision extends the scope of customized ablation beyond the limits of conventional monovision laser methods in several ways.
Whether for its customized treatment profiles, its visual acuity at all distances, its indications range or its immediate impact, PRESBYOND Laser Blended Vision is a clear treatment choice for the fast growing demographic of presbyopic patients.
PRESBYOND Laser Blended VisionCustomized. All distances. Immediate.
PRESBYOND Laser Blended Vision
lets you greatly expand your MEL® 80
excimer laser treatment repertoire
and your patient base
4
Conventional monovision
With conventional monovision treatment
methods (both surgical and contact lens),
the dominant eye is corrected for distance
vision to almost plano. The non-dominant
eye, on the other hand, is corrected for
near vision, usually to be myopic up to
-3.0 D. Accommodation is thereby often
diminished. The method aims to achieve
optimal vision at distance and near range.
Following the procedure, the brain must
contend with two separate images with
different levels of correction.
According to scientific studies, only about
60 % of patients are able to successfully
merge the two images to experience
sharp binocular vision at near and distance
range.1 Patients that tolerate the method
are left with an uncorrected, compromised
gap in the intermediate range, the so-
called Blur Zone. In addition to the fuzzy
image, it can also cause other side effects
such as reduced contrast sensitivity and
stereoacuity. Also, it often takes patients a
long time to get used to merging the two
images into one picture.
Conventional monovision is a correction
method with compromises. For a number
of patients, it is not a suitable option.
PRESBYOND Laser Blended Vision
As a new, physiologically optimized laser
treatment method for presbyopic patients,
PRESBYOND Laser Blended Vision
represents the next stage in eye care
excellence. Similar to conventional
monovision, the dominant eye is corrected
for distance vision to almost plano, whereas
the non-dominant eye is corrected to be
slightly myopic for near vision to -1.5 D.
This micro-monovision strategy is further
enhanced by a decisive difference: an
increase in the depth of field of each eye
using a wavefront-optimized ablation
profile to create a continuous refractive
power gradient for the whole optical zone
of the cornea. PRESBYOND Laser Blended
Vision is an absolutely individualized
treatment plan based on preoperative
spherical aberrations and the functional
age of the eye. As a result, a customized
fusion of the two images for near and
distance vision is created for each patient –
the so-called Blend Zone.
Although similar to conventional monovision laser treatments in terms of the workflow, PRESBYOND® Laser Blended Vision takes customized vision correction a step beyond, particularly with respect to the outcomes.
Next-level vision correctionbeyond conventional monovision
The unique Blend Zone
Essentially, the Blend Zone makes it easier for the brain to merge the images of both
eyes, thereby achieving true binocular vision. This is a significant advantage over
conventional monovision. In addition to excellent near and far vision, PRESBYOND
Laser Blended Vision patients also experience very good visual acuity and contrast
sensitivity in the intermediate range. They adapt faster and most are delighted about
not having to wear glasses anymore. Not surprisingly, PRESBYOND Laser Blended
Vision is tolerated by a high number of patients – up to 97 %.1
Increase in depth
of field
Increase in depth of field
Dominant eye
±0.0 D
Non-dominant eye
-1.5 D
NEAR
DISTANCE
No increase in depth of field
No increase in depth
of field
Dominant eye
±0.0 D
Non-dominant eye
up to -3.0 D
NEAR
DISTANCE
76
Individualized ablations
PRESBYOND Laser Blended Vision is a truly customized
approach for treating presbyopic patients. It incorporates
preoperative wavefront data to fine-tune the depth of
field for each eye individually. The functional age of the
eye is also factored in. In result, a personalized ablation
profile is created for optimized target refraction. The mono-
vision component can be pre-adjusted for the patient’s
tolerance level. Also, different optical zone sizes can be
selected to account for the patient’s pupil size.
Ideal for a growing demographic
As an optimized laser method for age-related loss of
accommodation, PRESBYOND Laser Blended Vision is
ideally suited for serving the needs of patients 40–60 years
of age – a fast-growing demographic group interested in
sophisticated options. It is also one of the least invasive
methods for addressing presbyopia.
Familiar procedure
Following the same workflow as established LASIK
procedures, PRESBYOND Laser Blended Vision combines the
convenient binocular treatment planning of the CRS-Master®
with the proven comfort and workflow of the MEL® 80
excimer laser from Carl Zeiss.
Customized
Generally speaking, conventional monovision methods utilize standard ablation profiles. In the case of PRESBYOND® Laser Blended Vision, they are customized for each eye individually because of the increased depth of field.
PRESBYOND Laser Blended Vision provides the most natural form of presbyopic
correction, including 97 % of all refractive errors with preservation of contrast
sensitivity, night vision and stereoacuity; it is easily enhanced, adjusted and even
reversed, if needed. It’s the Olympic Gold medalist in presbyopic surgery.
Prof. Dan Reinstein, London Vision Clinic, United Kingdom, August 2012
»
»
98 9
All distances
Strictly in terms of the visual outcomes, PRESBYOND® Laser Blended Vision transcends conventional monovision methods by offering clear sight at all distances: near, intermediate and far. Better yet, virtually all patients are freed from the need to wear reading glasses.1-4, 6
Outstanding visual acuity
By customizing each eye individually, PRESBYOND Laser
Blended Vision provides excellent visual acuity for near and
distance vision. Unlike conventional monovision methods,
PRESBYOND Laser Blended Vision also offers good inter-
mediate vision in the Blend Zone. According to clinical
studies, there is virtually no loss of contrast sensitivity and
stereoacuity is maintained. Also, side effects such as
multiple images in one eye are almost eliminated.
An all-natural approach
PRESBYOND Laser Blended Vision is a physiological solution
and a true binocular method for treating presbyopic patients.
Wide indication range
PRESBYOND Laser Blended Vision is a proven and effective
method for treating indications ranging from -8.0 D to
+2.0 D, including emmetropic and astigmatic presbyopic
patients (up to 2.0 cyl).
I have been impressed by the very high patient tolerance rate for
PRESBYOND Laser Blended Vision which is much higher than what we see
for conventional monovision methods.
Christian de Courten, MD, MV Sante Vision, Switzerland, July 2012
»
»
10
Immediate
What also clearly sets PRESBYOND® Laser Blended Vision apart from other laser techniques is the immediate positive impact it offers patients and refractive surgeons alike – visually for the former, economically for the latter.1–4
Appropriate for most patients
A key advantage of PRESBYOND Laser Blended Vision is
that it is proven to be tolerated by more patients than
conventional monovision. It is effective for treating up to
97 %1 of all presbyopia-related forms of impairment as
compared to only 59–67 % for conventional monovision.
Even presbyopic patients with emmetropia and astigmatism
can be treated. In fact, it has the potential to achieve a far
greater success rate than any comparable treatment.1,5
A competitive edge
PRESBYOND Laser Blended Vision allows refractive practices
already using a MEL® 80 excimer laser and CRS-Master®
from Carl Zeiss to significantly expand their LASIK repertoire
and increase the patient base. For most patients, a positive
effect is that they can read without glasses the very same
day. As such, PRESBYOND Laser Blended Vision offers a
decisive competitive advantage over LASIK practices only
specializing in monovision treatment methods.
ZEISS Laser Blended Vision is the answer to the missing platform in LASIK refractive
surgery for myopic, hyperopic and even emmetropic patients.
Andrew Taylor, MD. F.R.C.S.C., LASIK Eye Surgery Niagara Falls, Canada, January 2012
»»
12
Technical data
Dimensions (W x D x H)
Max. 1060 x 420 x 1510 mm
Weight Max. 110 kg
input voltage with table
120, 230 V AC ± 10 %
without table 100, 120, 230 V AC; adjustable, single phase
input current Max. 6.3 A
Rated frequency 50/60 Hz
ambient conditions Temperature +15 . . . +30 °CRelative humidity 30 . . . 75 % Non condensing Air pressure 700 . . . 1060 hPa
Data backup CD/DVD
Data transfer USB flash memory drive (USB memory stick)
Data printout Via network connection with Ethernet cable and optional network isolator
approval CE mark as per Medical DeviceDirective 93/42/EEC
000000-1429-382-04-GB-Vs01 MWIB
VISIBLE AND INVISBLE
LASER RADIATION AVOID EYE
OR SKIN EXPOSURE TO DIRECT
OR SCATTERED RADITION
CLASS 4 LASER PRODUCT
ACCORDING
DIN EN 60825-1:2003 UND :2008
193 nm, <2 mJ, 4-7 ns
MEL 80
Laser data
Type ArF excimer laser
Wavelength 193 nm
Frequency 250 Hz
aiming beam diode 650 nm (laser class 1 according IEC 60825-1)
Device data
Weight of MEl 80 290 kg incl. gas cylinder
Weight of patient supporting system
232 kg
Dimensions (laser, W x D x H) 800 x 1550 x 1490 mm
Dimensions including patient supporting system (W x D x H)
1800 x 3140 x 1490 mm
Power supply 100 V AC; 50/60 Hz; 17.5 A 120 V AC; 50/60 Hz; 14.6 A 208, 220, 230, 240 V AC; 50/60 Hz; 7.9 A
approval CE mark as per Medical Device Directive 93/42/EEC
Gas supply Integrated ArF-Premix cylinder 10 l
Equipment
Surgical microscope OPMI® pico with integrated video camera
active eye tracker Infrared, pupil and limbus tracking, 1050 frames per second (fps)
CCa+ (plume removal system) Integrated in device
Spot scanning parameters
Beam dimensions 0.7 mm FWHM (Full width at half maximum), Gaussian beam profile
Phototherapeutic keratectomy
area ablation Programmed PTK shaping
References
Claims made in this document are supported by
information provided in the following publications:
1. Reinstein DZ, Couch DG, Archer TJ. LASIK for Hyperopic Astigmatism and
Presbyopia Using Micro-monovision With the Carl Zeiss Meditec MEL 80.
J Refract Surg. 2009;25(1):37-58.
2. Reinstein DZ, Archer TJ, Gobbe M. LASIK for Myopic Astigmatism and
Presbyopia Using Non-Linear Aspheric Micro-Monovision with the
Carl Zeiss Meditec MEL 80 Platform. J Refract Surg. 2011;27(1):23-37.
3. Reinstein DZ, Carp GI, Archer TJ, Gobbe M. LASIK for the correction of
presbyopia in emmetropic patients using aspheric ablation profiles and a
micro-monovision protocol with the Carl Zeiss Meditec MEL 80 and VisuMax.
J Refract Surg. 2012 [In Press].
4. Reinstein DZ, Archer TJ, Gobbe M. Stereoacuity after Corneal Presbyopic
LASIK in Myopic, Hyperopic and Emmetropic Patients. ESCRS Annual Meeting,
Vienna, September 2011.
5. Evans BJ. Monovision: a review. Ophthalmic Physiol Opt.
2007;27(5):417-439.
6. Reinstein DZ , Archer TJ, Gobbe M. LASER BLENDED VISION
FOR PRESBYOPIA CORRECTION.
Pallikaris I, Plainis S, Charman WN.
Presbyopia: Origins, Effects, and Treatment (pp. 191-196).
© 2012 SLACK Incorporated.
A perfect combination: CRS-Master and MEL 80
PRESBYOND® Laser Blended Vision is an optional software
upgrade for the CRS-Master® from Carl Zeiss. It forms a
perfect fit with the MEL® 80 excimer laser, expanding the
repertoire of the customized refractive laser corrections
far beyond the limits of conventional monovision methods.
CRS-Master
PRESBYOND and CRS-Master are not intended for
sale in the United States.
14 15
Publ
icat
ion
No:
000
000-
2000
-981
Pr
inte
d in
Ger
man
y CZ
-VIII
/201
2 Th
e co
nten
ts o
f the
bro
chur
e m
ay d
iffer
from
the
curre
nt s
tatu
s of
app
rova
l of t
he p
rodu
ct in
you
r cou
ntry
. Ple
ase
cont
act o
ur re
gion
al re
pres
enta
tive
for m
ore
info
rmat
ion.
Subj
ect t
o ch
ange
in d
esig
n an
d sc
ope
of d
eliv
ery
and
as a
resu
lt of
ong
oing
tech
nica
l dev
elop
men
t. Pr
inte
d on
ele
men
tal c
hlor
ine-
free
blea
ched
pap
er.
© 2
012
by C
arl Z
eiss
Med
itec
AG. A
ll co
pyrig
hts
rese
rved
.
Your local contact:
ArgentinaCarl Zeiss Argentina S.A.Calle Nahuel Huapi 4015 / 25C1430 BCO Buenos AiresArgentinaPhone: +54 11 45 45 66 [email protected]
AustraliaCarl Zeiss Pty Ltd Tenancy Office 4, Level 1 40-52 Talavera Road North Ryde NSW 2113 AustraliaPhone: +61 2 9020 [email protected]
AustriaCarl Zeiss GmbHLaxenburger Str. 21100 ViennaAustriaPhone: +43 1 79 51 [email protected]
BelgiumCarl Zeiss NV-SAIkaroslaan 491930 ZaventemBelgiumPhone: +32 2 719 39 [email protected]
BrazilCarl Zeiss do Brasil Ltda.Av. Naçoes Unidas, 21711CEP04795-100 São PauloBrazilPhone: +55 11 5693 [email protected]
CanadaCarl Zeiss Canada Ltd.45 Valleybrook DriveToronto, ON M3B 2S6CanadaPhone: +1 800 387 [email protected]
ChinaCarl Zeiss Shanghai Co. Ltd.1/f., Ke Yuan Building11 Ri Yin Nan RoadWaigaoqiao Free Trade Zone2005 Yang Gao Bei RoadShanghai 200131ChinaPhone: +86 21 5048 17 [email protected]
Czech RepublicCarl Zeiss spol. s.r.o.Radlická 14/3201150 00 Prague 5Czech RepublicPhone: +420 233 101 [email protected]
FranceCarl Zeiss Meditec France SAS60, route de Sartrouville78230 Le PecqFrancePhone: +33 1 34 80 21 [email protected]
GermanyCarl Zeiss Meditec VG mbHCarl-Zeiss-Strasse 2273447 OberkochenGermanyPhone: +49 7364 20 [email protected] Ophthalmology:Phone: +49 800 470 50 [email protected]
Hong KongCarl Zeiss Far East Co. Ltd.Units 11-12. 25/FTower 2, Ever Gain PlazaNo. 88 Container Port RoadKwai ChungHong KongPhone: +852 2332 [email protected]
IndiaCarl Zeiss India Pvt. Ltd.Plot No.3, Jigani Link RoadBommasandra Industrial Area Bangalore - 560 099IndiaPhone: +91 80 4343 [email protected]
ItalyCarl Zeiss S.p.A.Viale delle Industrie 2020020 Arese (Milan)ItalyPhone: +39 02 93773 [email protected]
JapanCarl Zeiss Meditec Japan Co. Ltd.Shinjuku KuTokyo 160-000322 Honchio-ChoJapanOphthalmic instruments:Phone: +81 3 33 55 [email protected] instruments:Phone: +81 3 33 55 [email protected]
MalaysiaCarl Zeiss Sdn Bhd.Lot2, Jalan 243/51 A46100 Petaling JayaSelangor Darul EhsanMalaysiaPhone: +60 3 7877 50 [email protected]
MexicoCarl Zeiss de México S.A. de C.V.Avenida Miguel Angel de Quevedo 49604010 Mexico CityMexicoPhone: +52 55 59 99 [email protected]
NetherlandsCarl Zeiss B.V.Trapezium 300Postbus 3103364 DL SliedrechtNetherlandsPhone: +31 184 43 34 [email protected]
New ZealandCarl Zeiss (N.Z.) Ltd.15B Paramount DriveP.O. Box 121 - 1001Henderson, Auckland 0650New ZealandPhone: +64 9 838 [email protected]
PolandCarl Zeiss sp. Z o.o.ul. Lopuszanska 3202-220 WarsawPolandPhone: +48 22 858 [email protected]
SingaporeCarl Zeiss Ptd. Ltd.50 Kaki Bukit PlaceSingapore 415926SingaporePhone: +65 6741 [email protected]
South AfricaCarl Zeiss (Pty.) Ltd.363 Oak AvenueFerndaleRandburg 2194South AfricaPhone: +27 11 886 [email protected]
South KoreaCarl Zeiss Co. Ltd.Seoul 121-828Mapo-gu141-1, Sangsu-dong2F, BR Elitel Bldg.South KoreaPhone: +82 2 3140 [email protected]
SpainCarl Zeiss Meditec Iberia S.A.U.Ronda de Poniente, 15Tres Cantos28760 MadridSpainPhone: +34 91 203 37 [email protected]
SwedenCarl Zeiss ABTegeluddsvaegen 7610254 StockholmSwedenPhone: +46 84 59 25 [email protected]
SwitzerlandCarl Zeiss AGFeldbachstrasse 818714 FeldbachSwitzerlandPhone: +41 55 254 [email protected]
ThailandCarl Zeiss Thailand90 CyberWorld Tower A, 36th Floor, Unit A 3601230 Ratchadapisek RoadHuaykhwang, Bangkok 10310ThailandPhone: +66 2 248 [email protected]
United KingdomCarl Zeiss Ltd. 509 Coldhams lane Cambridge CAMBS CB1 3JS, United KingdomPhone: +44 1707 [email protected]
United States of AmericaCarl Zeiss Meditec, Inc.5160 Hacienda DriveDublin, CA 94568USAPhone: +1 925 557 [email protected]
Carl Zeiss Meditec AG Goeschwitzer Strasse 51–5207745 JenaGermanywww.meditec.zeiss.com/PRESBYOND
0297