cursus LC-MS 2001

154
1 AO-MO-CHROM - 14 februari 2001 cursus LC-MS 2001 - WJAN

Transcript of cursus LC-MS 2001

Page 1: cursus LC-MS 2001

1AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Page 2: cursus LC-MS 2001

2AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Cursus LC-MS-2001

1. Introductie massaspectrometrie2. Problematiek LC-MS3. Historisch overzicht LC-MS4. Electrospray5. APcI6. Bespreking MS toepassing AO-MO-CHROM7. Aanhalen praktische problemen

Page 3: cursus LC-MS 2001

3AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Sir Joseph John Thomson (1856 - 1940)

‘I feel sure that there are many problems in chemistry which could be solved with far more greater ease by this than by any other method. The method is surprisingly sensitive, requires an infinitesimal amount of material and does not require this to be specially purified…’

Statement about the technique of mass spectrometry, made by J.J.Thomson in 1913

Page 4: cursus LC-MS 2001

4AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Deel 1

1.1 Wat is MS?1.2 Instrumentatie1.3 Ionisatie1.4 Quadrupoolfilter1.5 Tandem massaspectrometrie

Page 5: cursus LC-MS 2001

5AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.1 Wat is MS?

MS is een analysetechniek die ionen in de gasfase bestudeert.Toepassingen:

Kwalitatief en kwantitatiefOrganisch en anorganischStructuuropheldering van onbekende verbindingen

Eerste massaspectrometersir Francis William Aston (1877-1945)Cambridge, 1919

Page 6: cursus LC-MS 2001

6AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Eerste MS in 1919

Page 7: cursus LC-MS 2001

7AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Eerste MS spectrum

Page 8: cursus LC-MS 2001

8AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.2 Opbouw MS

InlaatsysteemBronMassa-analysator in vacuümsysteemDetector

Page 9: cursus LC-MS 2001

9AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Inlaatsystemen

Doel: moleculen introduceren in de bronHoe:

Direct Insertion ProbeGCCE, CECLC

Page 10: cursus LC-MS 2001

10AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Bron

Doel: moleculen ioniserenHoe:

Electron Impact (EI)Chemische Ionisatie (CI)Desorptie Chemische Ionisatie (DCI)Field Desorption (FD)Fast Atom Bombardment (FAB)Fast Ion Bombardment (FIB)Thermospray (TSP)Thermal Desorption Laser Desorption (LD)Matrix Assisted Laser Desorption (MALDI)

Page 11: cursus LC-MS 2001

11AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Bron

Electrospray (ES)Atmospheric Pressure Chemical Ionisation (APcI)

Atmospheric Pressure Photon Ionisation (APPI)Sonic Spray Ionisation (SSI)

Page 12: cursus LC-MS 2001

12AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Analysator

Doel: ionen scheiden op basis van m/zm/z = massa over lading = Thomson (Th)

Hoe:Quadrupoolfilter (Q of q)Iontrap (IT)Magneet Sector (M)Time of Flight (TOF)Fourrier Transform Ion Cyclotron (FTMS)

Page 13: cursus LC-MS 2001

13AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Parameters van de analysator

Resolutie (R) R= m/∆m10% valley ofwel 50% (FWHM)

Massa-accuraatheidMassa-bereikGevoeligheidScan snelheidPrijs

Page 14: cursus LC-MS 2001

14AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Vacuümsysteem

Waarom vacuümsysteem?Ionen in de gasfase botsen met luchtmoleculenfragmentatie, verlies van gevoeligheid, verlies van resolutie

Hoe?Vacuümpompen

Voorvacuümpompen (tot 10-3 mb)– rotatievoorvacuümpompen

Hoogvacuümpompen (tot 10-9 mb)– Turbomoleculaire pompen

– Oliediffusiepompen

– Cryogene pompen

Page 15: cursus LC-MS 2001

15AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Detectoren

Doel: omzetten van ionen in een meetbare stroomHoe?

Electron multipliers (EM)Korte levensduur (enkele jaren)

Photonmultipliers (PM)Lange levensduur (10 jaar)Vnl Q

Page 16: cursus LC-MS 2001

16AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Werking Detector

Page 17: cursus LC-MS 2001

17AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Detectoren

Array detectoren (AD)Uiterst gevoeligKlein lineair bereikKwetsbaar en duurMagneet SectorTOFVb: MCP (Q-TOF)

Page 18: cursus LC-MS 2001

18AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Detectoren: MCP

Page 19: cursus LC-MS 2001

19AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Detectoren: MCP

Page 20: cursus LC-MS 2001

20AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Schema electrospray bron

Page 21: cursus LC-MS 2001

21AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Schema ES bron

Page 22: cursus LC-MS 2001

22AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Schema APcI bron

Page 23: cursus LC-MS 2001

23AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.3 Ionisatie

Gemiddeld Moleculair Gewicht (Average MW)MW berekend met behulp van de gewogen gemiddelden van de natuurlijke isotopen van elk atoom (bvb zoals vermeld in MI, stoechiometrische berekeningen, ed)

Mono-isotopische Moleculair Gewicht (Nominal MW)

MW berekend met behulp van de massa van het meestvoorkomend isotoop van elk atoom (zoals gemeten met een Q)

Page 24: cursus LC-MS 2001

24AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Moleculair gewicht (MW)

Exacte Moleculair Gewicht (Exact MW)Is het mono-isotopisch MW, rekening gehouden met de massa-defecten (wordt gemeten via een hoge resolutie MS)

Voorbeeld voor 3 moleculen:Azijnzuur (C2H4O2; 60.0211)Ureum (CH4N2O; 60.0324)2-propanol (C3H8O; 60.0575)

Page 25: cursus LC-MS 2001

25AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Moleculair gewicht (MW)

Isobaren zijn moleculen met hetzelfde mono-isotopisch MW

Massa-eenheid: u of Da=1/12 van de massa van het 12C atoom=1/NA (NA=6.022 1023 mol-1)=1.66043 10-27 kg

12C = 12.000u

Page 26: cursus LC-MS 2001

26AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Moleculair gewicht (MW)

N

O

N

N

N

NOO

O

N

Cl

N N

Cl R051211 C-06.79.19.00 MW 705.64 Factor 1C35H38Cl2N8O4(+-)-(CIS)CDU : 3906 mg, 3 unitsITRACONAZOLE

Page 27: cursus LC-MS 2001

27AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Moleculair gewicht (MW)

Voorbeeld R051211Average MW: 705.6436Mono-isotopisch MW: 704Exact MW: 704.2393

Page 28: cursus LC-MS 2001

28AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Stikstofregel

Even MW: geen of een even aantal N-atomenOneven MW: oneven aantal N-atomenVoorbeeld R051211:mono-isotopisch MW 704, dus even aantal N-atomen (8)

Page 29: cursus LC-MS 2001

29AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Ionisatie (beperkt tot ES)

Positief of negatief:M → (M+H)+

M → (M-H)-

Nevenreacties (clustervorming)[M+NH4]+, [M+Na]+, [M+K]+

Page 30: cursus LC-MS 2001

30AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Fragmentatie (zeer beperkt)

Ion → ion + neutraal door toevoegen van extra inwendige energie

Extra spanningsveldBotsingsenergie (CID, CAD, …)

Page 31: cursus LC-MS 2001

31AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Massa spectrum

Curve van relatieve intensiteit (RI) in fuctie van m/z

Voorbeeld: R051211MW=704

[M+H]+=705m/z=705

Page 32: cursus LC-MS 2001

32AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Massa spectrum R051211

Page 33: cursus LC-MS 2001

33AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Isotopen

Hoofdpiek → mono-isotopisch→ alleen meest voorkomende isotopen

Koolstof: 12C (98.9%) en 13C (1.1%)Voorbeeld: R051211 (C35H38Cl2N8O4)

1x 13C in het skelet →signaal bij 704+1+1=706intensiteit: 1.1% x #C-atomen

2x 13C in het skelet →signaal bij 704+1+1+1=707intensiteit: 1.1% x 1.1% x 35 →verwaarloosbaar

Page 34: cursus LC-MS 2001

34AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Isotopen

Wat met Cl?35Cl (75.77%) en 37Cl (24.23%)algemeen 3:1voor R051211: piek bij 707Intensiteit: 3:1 x 3:1 = 9:6:1

Page 35: cursus LC-MS 2001

35AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Isotopen R051211

Page 36: cursus LC-MS 2001

36AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Isotopen

Wat met Br?79Br (50.69%) en 81Br (49.31%)algemeen 1:1twee Br-atomen in de moleculeIntensiteit: 1:1 x 1:1 = 1:2:1

Page 37: cursus LC-MS 2001

37AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Isotopen

Page 38: cursus LC-MS 2001

38AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Isotopen

Page 39: cursus LC-MS 2001

39AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Clustervorming

Naast [M+H]+

soms ook[M+NH4]+, [M+Na]+, [M+K]+

Voor- en nadeelvoordeel: duidt gemakkelijk MW aannadeel: verlies intensiteit [M+H]+

Page 40: cursus LC-MS 2001

40AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Massa spectrum R051211

Page 41: cursus LC-MS 2001

41AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Quizvraag

Page 42: cursus LC-MS 2001

42AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Meervoudig geladen ionen

OorsprongVoorbeeld R051211: [M+2H]2+

m/z bij (704+1+1)/2 = 353Scheiding 12C en 13C 0.5 uScheiding 35Cl en 37Cl 1 u

Page 43: cursus LC-MS 2001

43AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Meervoudig geladen ionen

15:19:4502-Oct-1998

AdministratorOpl1

150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z0

100

%

FKC-222-001-002 621 (32.117) Cm (619:625-(610:611+632:633)) 1: Scan ES+ 9.15e7705.5

91495072

353.418687148

256.44453118

325.53254019

354.513249136

671.55920441

707.570918080

708.527327742

727.418685352

729.513366653

730.55518147

Page 44: cursus LC-MS 2001

44AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Interpretatie MS

API gekendMeten onbekende

verschil → info over onbekendevb: verschil=+16hydroxy of N-oxF → Cl[M+Na]+ → [M+K]+

vb: verschil=+14methyleringCH2 →C=O

Verschil discrete waardenvb: verschil 11?

Page 45: cursus LC-MS 2001

45AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.4 Quadrupoolfilter

º1953, Paul ( Nobelprijs 1989) en SteinwedelRelatief goedkoopEenvoudig en gebruiksvriendelijkResolutie = eenheidScan, gebaseerd op electrostatische velden

Page 46: cursus LC-MS 2001

46AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Quadrupoolfilter

Page 47: cursus LC-MS 2001

47AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Quadrupoolfilter

Page 48: cursus LC-MS 2001

48AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Quadrupoolfilter

DC en RF spanningEnkel DC: blokkageEnkel RF: geen filtering, wel focusScannen?

Page 49: cursus LC-MS 2001

49AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Scan technieken

Twee verschillende: full-scan en SIR(M)Full-scaninstelling scan van laag naar hoogideaal voor identificatie werkSIRslechts 1 ion, voor- en nadeelvoordeel: enorm gevoelig (100x)nadeel: geen structuur informatie

Te vergelijken met Single Wavelength en PDA

Page 50: cursus LC-MS 2001

50AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.5 Tandem massaspectrometrie

Combinatie van ‘2 MS’Techniek waarbij fragmentatie spectrum bekomen wordt van een ‘precursor’

Precursor ion (parent ion)Product ion (daughter ion)

Zowel in space als in timeVoorstelling:

MS1 selecteert precursorBotsingscel (fragmenten opwekken)MS2 bepaald fragmentatie spectrum

Voorbeelden zijnQqQ, QqTOF, BqQ, BqB, BqTOF, BqIT, IT, FTMS, QqQqQ, …

Page 51: cursus LC-MS 2001

51AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.5 Tandem massaspectrometrie

Gebruik:Structuur ophelderingQuantificatie!!!!

Voordeel:Uitzonderlijk selectiefZeer gevoelig

Nadeel:Bijzonder duur (uitgenomen IT)Ingewikkeld?

Page 52: cursus LC-MS 2001

52AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.5 Tandem massaspectrometrie

Enkele voorbeelden:Zonder noemenswaardige voorscheiding:

Geef het zuivere massa spectrum van verbinding ZGeef alle verbindingen die fragment Y bevatten (Precursor Scan)Geef alle verbindingen die X verliezen (Constant Neutral Loss scan)vb: alle Cl-verbindingen opsporen, verlies van water, verlies van MeOH, …

Page 53: cursus LC-MS 2001

53AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.5 Tandem massaspectrometrie

Page 54: cursus LC-MS 2001

54AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

1.5 Tandem massaspectrometrie

Kwantificatie:Kiezen twee ionen (overgang)

Page 55: cursus LC-MS 2001

55AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Deel 2: Problematiek LC-MS

2.1 Waarom koppeling HPLC met MS?

2.2 Vloeistof → gas2.3 Polaire moleculen2.4 Niet-vluchtige buffers

Page 56: cursus LC-MS 2001

56AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

2.1 Waarom koppeling met MS?

Page 57: cursus LC-MS 2001

57AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

2.2 Vloeistof →gas

Algemene gaswet leert:0.1 mL/min →135 cm3 gas bij 1 atmMS onder hoog vacuumvacuumpompen maar 5 cm3 gas bij 1 atmOplossingionisatie bij atmosferische drukdifferentieel pompen!Vloeistofstroom splitten!

Page 58: cursus LC-MS 2001

58AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

2.3 Polaire moleculen

HPLC = ideale scheidingstechniek voor polaire moleculenPolaire moleculen zijn meestal thermolabiel Zachte ionisatie techniek gebruikenES, APcI, (APPI, SSI), …

Page 59: cursus LC-MS 2001

59AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

2.4 Niet-vluchtige buffers

Worden gebruikt om HPLC methode te optimaliserenTe vermijden bij LC-MS

contaminatie ionenbron en MSsignaal onderdrukking (ion supression) MS spectrum

Beter met moderne ionenbronnen (Z-spray)Vluchtige buffers te prefereren

Page 60: cursus LC-MS 2001

60AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Niet-vluchtige buffersystemen

Vluchtige buffers:Ammonium acetaat, ammonium formiaat, ammonium carbonaatMierenzuur, azijnzuur, TFA

Te vermijden bij LC-MSAlle fosfaten, boraten en sulfatenZwavelzuur, fosforzuur

Opletten met modifiers, zoalsHPβCD, DEA, TEA, TRIS, …

Page 61: cursus LC-MS 2001

61AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Deel 3

Historisch overzicht LC-MS

Page 62: cursus LC-MS 2001

62AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Historisch overzicht LC-MS

Koppeling met gaschromatografie en massaspectrometrie reeds in 1958(gas, apolaire moleculen, laag moleculair gewicht, geen buffers, …)Vanaf ’70 koppelingspogingen HPLC en MSMeest succesvol en meest verbreid: ES (’90)

Page 63: cursus LC-MS 2001

63AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Historisch overzicht LC-MS

Moving belt: EI en CI (MacFadden, 1976)(Janssen 1983 -1986)Direct Liquid Introduction (1974)Thermospray (Marvin Vestal, 1983)(Janssen 1986 -1994)Particle Beam (Willoughby, 1984, EI en CI)Dynamic FAB (Caprioli, 1986)(Janssen 1994 -1996)

Page 64: cursus LC-MS 2001

64AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Historisch overzicht LC-MS

Electrospray (1984, Doyle & Fenn)(Janssen 1994 - )APcI (1974, Horning, opvolger thermospray, doorbraak ’90)(Janssen 1997-)APPI en SSI (eind ’90)nog niet commercieel

Page 65: cursus LC-MS 2001

65AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Deel 4: Electrospray

4.1 Algemene kenmerken4.2 Reden van succes4.3 Principe4.4 Experimentele gegevens

Page 66: cursus LC-MS 2001

66AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.1 Algemene kenmerken

Electrospray is een fysisch verschijnsel beschreven in 1917 door Zeleny:door het aanleggen van een hoog-electrisch veld op een vloeistof in een capillair wordt een spray van geladen druppels gevormdWordt dagdagelijks toegepast in de verfindustrie (autolak)Dole, 1968, ionisatie-techniekFenn, 1984, introductie electrospray

Page 67: cursus LC-MS 2001

67AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.1 Algemene kenmerken

Ionisatie gebeurt bij atmosferische druk en in de vloeistof faseZachte ionisatie techniek

Weinig inwendige energieLage kinetische energieDus weinig kans op fragmentatie

Ionisatie gebeurt bij lage temperatuur (<100°)Ideaal voor polaire, thermolabiele producten

Page 68: cursus LC-MS 2001

68AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.1 Algemene kenmerken

Debiet vanaf 30 nL/min (nano-ES) tot 1 mL/minES is een ‘concentratiegevoelige’ detector, zoals UV, dus splitten mag!Bepalen van kleine moleculen (MW<100) tot macromoleculenrecord is het tetrameer van een hemoglobineachtige verbindingMW~4 000 000

Page 69: cursus LC-MS 2001

69AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.1 Algemene kenmerken

Bij grotere moleculen meervoudige ladingen mogelijkKan gecombineerd worden met alle MS analysatorenZowel + als –Vb: HHM

Page 70: cursus LC-MS 2001

70AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.1 vb: HHM

Page 71: cursus LC-MS 2001

71AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.2 Reden van succes

Analyse van proteïnenZeer gevoeligGebruiksvriendelijkRobuustRelatief goedkoop

Page 72: cursus LC-MS 2001

72AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.3 Principe

Aanleggen van een hoog electrisch veld (106 V/m, 3kV op een metalen tip van 100 µm)Spray van geladen druppels wordt gevormd Analyt-ion wordt gevormd in de vloeistoffase (klassieke zuur-base chemie)Geladen druppels verkleinen door solvent evaporatie

Page 73: cursus LC-MS 2001

73AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.3 Principe

Page 74: cursus LC-MS 2001

74AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

4.3 Theorie

Ion-evaporation model (Iribarne and Thomson)Druppel verkleint, Coulomb afstoting vergrootRaleigh limiet overschreven, druppel explodeert: ion in de gasfase (alleen kleine moleculen)

Mechanisme van Doyledoor solvent-evaporatie blijft uiteindelijk alleen ion over (alleen grote moleculen)

Page 75: cursus LC-MS 2001

75AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

5.0 APcI

5.1 Algemene kenmerken5.2 Principe

Page 76: cursus LC-MS 2001

76AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

5.1 Algemene kenmerken

Toepasbaar voor moleculen die apolairder zijn en minder ioniseerbaar dan deze voor ESMW<2000Vloeistof debiet 1 tot 2 mL/minMassa-gevoelig (kolom splitting reduceert gevoeligheid)In combinatie met alle massa-analysatorenIonisatie bij hoge temeratuur (~400°C)dus geen zachte ionisatie techniek

Page 77: cursus LC-MS 2001

77AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

5.1 Algemene kenmerken

Ionisatie gebeurt in de gasfase, echte CIIonisatie gebeurt bij atmosferische druk+ en - ionenToepasbaar voor pesticiden, hormonen, …

Page 78: cursus LC-MS 2001

78AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

5.2 Principe

Er wordt geen electrisch veld aangelegd aan de tipHoog vloeistofdebiet (1 mL/min – 2 mL/min)Een spray wordt gevormd door het eluens sterk te verhittenIonisatie van de solvent moleculen gebeurt dmv een onladingselectrodeDe geïoniseerde solvent moleculen ondergaan zuur – base reacties in de gasfase

Page 79: cursus LC-MS 2001

79AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

5.2 Principe

Page 80: cursus LC-MS 2001

80AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

5.2 Principe

Page 81: cursus LC-MS 2001

81AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6 Algemene toepassingen

6.1 Methode Ontwikkeling: SCM6.2 Methode Validatie: specificiteit6.3 Forced Degradation Studies6.4 Troubleshooting: identificatie onzuiverheden/afbraakproducten6.5 Gebruik van MS als supportieve techniek

Page 82: cursus LC-MS 2001

82AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.1 Overzicht HPLC MO

Pro-Active Method

Development - Method Validation

Phase 1- Phase 2 methods

"Locked method"

Phase 3 Method

Development and validation

Supportive methods

Final Method

Development and

Validation

Secondarymethods

Page 83: cursus LC-MS 2001

83AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Gedetailleerd overzicht

(4)InitialBatchStress

samples (17)Placebo Stability

Excipients (9) ReportStress

(1) (5) solutionNotification of Method screening Optimization Optimization Method descriptionDevelopment Experiments Phase 1 - 2 Method Method is "Locked"

(6) Stress MV level 1 - 2 (16)dry (12)

Peak purity (10) Selectivity SamplesLC-MS/PDA

(7) (15)Retain

Plan PMD the samples Identification/Peak purity ID Sheetstudy (11) LC-MS/PDA/NMR/Others

(3) (13) (14)

(8) Supportive method(s) HPLC CE TLC LC-MS CE-MS etc…

(2)

"Teamleader" co-ordinates the Method Development Process

Page 84: cursus LC-MS 2001

84AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Screeningsmodule deel 1

Page 85: cursus LC-MS 2001

85AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Purity bepalingen

Op basis van PDAOp basis van MS

Page 86: cursus LC-MS 2001

86AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Purity met PDA

Via MassLynx routineEerst integrerenPeak Purity berekenen

0<P<100

Interpretatie P !!!P > 95 OK?

Page 87: cursus LC-MS 2001

87AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Purity met PDA

Visueel (via overlay)Op basis van 3 PDA spectra

Up-slopeApexDown-slope

Page 88: cursus LC-MS 2001

88AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Purity met MS

Visueel (via overlay)Op basis van 3 MS spectra

Up-slopeApexDown-slope

Page 89: cursus LC-MS 2001

89AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Purity met MS

Via MassLynxBepalen van RIC/TIC

Page 90: cursus LC-MS 2001

90AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

RIC/TIC

RIC = som van alle ionen die product gerelateerdzijnproduct + fragmenten + clustersTIC = som van alle ionenook achtergrond ionen + onbekende ionen + ...

RIC/TIC ≅ 1 (of 100%)

Page 91: cursus LC-MS 2001

91AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

BPI

BPI = curve van alle basis ionen in de spectraIs soms beter dan TICVb: R149524

Page 92: cursus LC-MS 2001

92AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

BPI

Page 93: cursus LC-MS 2001

93AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Identificatie

Beperkte mogelijkhedenInstrumentatieAfspraak spectro

Page 94: cursus LC-MS 2001

94AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Toepassing SCM deel 1

PurityVoorbeeld R184798

Page 95: cursus LC-MS 2001

95AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Chemische structuur R184798

HO

O

NO O

O

Cl

HN

OH

Page 96: cursus LC-MS 2001

96AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld van een sample list in SCM deel I

Page 97: cursus LC-MS 2001

97AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 chromatogram

AO-053017:54:37

05-Oct-2000pH=10; 70 gradenSample stress WJAN

0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00Time-1

100

%

MO-90-R184798-1-1-4-16 2: Diode Array 230 1.00Da

1.06e612.35

13.27

Page 98: cursus LC-MS 2001

98AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 purity

AO-053017:54:37

05-Oct-2000pH=10; 70 gradenSample stress WJAN

12.00 12.10 12.20 12.30 12.40 12.50 12.60 12.70 12.80 12.90Time0

100

%

MO-90-R184798-1-1-4-16 Sm (Mn, 2x3) 2: Diode Array 230 1.00Da

1.01e6Area

Area%100.00

Area220117.77

Height1005566

Time12.36

12.36;99%;220118

Page 99: cursus LC-MS 2001

99AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 PDA-spectra

AO-053017:54:37

05-Oct-2000WJANpH=10; 70 gradenSample stress

200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 360 370 380 390 400nm0

100

%

0

100

%

0

100

%

MO-90-R184798-1-1-4-16 1490 (12.568) Cm (1490-1441) 2: Diode Array 6.26e5210.0

MO-90-R184798-1-1-4-16 1465 (12.360) Cm (1465-1441) 2: Diode Array 2.11e6214.0

296.0

MO-90-R184798-1-1-4-16 1454 (12.268) Cm (1454-1441) 2: Diode Array 5.60e5210.0

Page 100: cursus LC-MS 2001

100AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 PDA-spectra overlay

AO-053017:54:37

05-Oct-2000WJANpH=10; 70 gradenSample stress

200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 360 370 380 390 400nm0

100

%

MO-90-R184798-1-1-4-16 1490 (12.568) Cm (1490-1441) 2: Diode Array 6.26e5210.0

Page 101: cursus LC-MS 2001

101AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Screeningsmodule deel 2

Page 102: cursus LC-MS 2001

102AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Toepassing SCM deel II

Purity en IdentificatieVoorbeeld R184798

Page 103: cursus LC-MS 2001

103AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld van een sample list in SCM deel II

Page 104: cursus LC-MS 2001

104AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 chromatogram peak purity

AO-053019:36:29

14-Nov-2000stress pH 2, 70C, 28 dagensample LVOS

12.00 12.10 12.20 12.30 12.40 12.50 12.60 12.70 12.80 12.90 13.00Time0

100

%

MO-90-009-008 Sm (Mn, 2x3) 3: Diode Array 230 1.00Da

8.47e5Area

Area%100.00

Area181795.14

Height845203

Time12.44

12.44;100%;181795

Page 105: cursus LC-MS 2001

105AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 PDA overzicht

AO-0530stress pH 2, 70C, 28 dagensample LVOS 15-Nov-2000

200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 360 370 380 390 400nm0

100

%

0

100

%

0

100

%

MO-90-009-030 1497 (12.582) Cm (1497-1451) 3: Diode Array 7.25e5210

MO-90-009-030 1474 (12.390) Cm (1474-1451) 3: Diode Array 2.00e6212

295

MO-90-009-030 1462 (12.290) Cm (1462-1451) 3: Diode Array 8.08e5210

Page 106: cursus LC-MS 2001

106AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R184798 PDA overlay

AO-0530stress pH 2, 70C, 28 dagensample LVOS 15-Nov-2000

200 210 220 230 240 250 260 270 280 290 300 310 320 330 340 350 360 370 380 390 400nm0

100

%

MO-90-009-030 1497 (12.582) Cm (1497-1451) 3: Diode Array 7.25e5210

Page 107: cursus LC-MS 2001

107AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Overzicht massa spectra

AO-0530stress pH 2, 70C, 28 dagensample LVOS 15-Nov-2000

200 300 400 500 600 700 800 900 1000m/z0

100

%

0

100

%

0

100

%

MO-90-009-030 340 (12.739) Cm (340-326x2.000) 2: Scan ES+ 1.78e6427

411409130

171 213229

294 341

429

430 465 875853

569500585

669663693

837775728

877 998995921

MO-90-009-030 332 (12.434) Cm (332-326x2.000) 2: Scan ES+ 3.63e6427

409130

294182 186213393

296

429

430

853449 487 569528

747693571674

758 855 891 997975

998

MO-90-009-030 329 (12.320) Cm (329-326x2.000) 2: Scan ES+ 2.01e6427

409130

154 258193 296 341391

429

430 853465 569500 571 689679704

757772

819875

899 995939 997

Page 108: cursus LC-MS 2001

108AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Overlay massaspectra

AO-0530stress pH 2, 70C, 28 dagensample LVOS 15-Nov-2000

200 300 400 500 600 700 800 900 1000m/z0

100

%

MO-90-009-030 340 (12.739) Cm (340-326x2.000) 2: Scan ES+ 1.78e6427

411409130171 213

229294 341

429

430449

465 875853

467 569500585

669663693

837775728

877 998995921

Page 109: cursus LC-MS 2001

109AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

RIC/TIC

Via MassLynx(vrij ingewikkelde) procedure

Page 110: cursus LC-MS 2001

110AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

RIC/TIC

AO-053019:44:24

15-Nov-2000LVOSstress pH 2, 70C, 28 dagensample

150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z0

100

%

MO-90-009-030 331 (12.396) Cm (327:331-317:322x2.000) 2: Scan ES+ 1.97e5427.2

409.2

294.2186.2 217.2267.8

393.3378.9

302.4

428.2

449.1

853.3

451.1

465.1

467.1

487.1 500.1 569.0 689.3586.6

822.6770.0

855.3

875.3

877.3

878.4

995.1900.1

Page 111: cursus LC-MS 2001

111AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

RIC/TIC

Sample Name: MO-90-009-030 Sample ID: sample

# Name RT Area1 RIC4 12.396 540242 RIC3 12.473 1787453 RIC2 12.396 1722274 RIC1 12.473 35829655 TIC 12.473 3955590

Page 112: cursus LC-MS 2001

112AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Opstellen van ID-sheets

Doel: bundelen van gegevensOnzuiverheden APIAfbraakproducten

Start bij AO-MO-CHROMAO server-Applications-ID-sheetsVroeger: overdracht SpectroNu: MO-CHROM blijft verantwoordelijk

Voorbeeld: R184798

Page 113: cursus LC-MS 2001

113AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld ID-sheet R184798

Page 114: cursus LC-MS 2001

114AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld ID-sheet R184798

Page 115: cursus LC-MS 2001

115AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld ID-sheet R184798

Page 116: cursus LC-MS 2001

116AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Eindresultaat R184798

Page 117: cursus LC-MS 2001

117AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.2 Methode validatie: Specificiteit

SpecificiteitVoorbeeld Galantamine

Page 118: cursus LC-MS 2001

118AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Specificiteit

Specificity is the ability to assess unequivocally the analyte in the presence of components which may be expected to be present. Typically, these might include impurities, degradation products, matrices, etc…

Page 119: cursus LC-MS 2001

119AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Acceptatie criteria

All placebo and blank peaks should be separated from degradants, impurities and the main compoundAll specified and selectivity related substances should be separated from the main compoundAll specified degradants should be separated from each otherAll selectivity degradants and impurities should be separated from the specified degradants and impurities

Page 120: cursus LC-MS 2001

120AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Acceptatie criteria

All specified impurities should be separated from the degradants and preferably be separated from each otherThe PDA spectra and the ES-MS peak purity analysis should confirm that the API peak is indeed pureThe PDA spectra are comparableThe ES-MS spectra are comparable

Page 121: cursus LC-MS 2001

121AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Sample list van de specificiteitsstudie van

Galantamine

Page 122: cursus LC-MS 2001

122AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study(FDS)

Momenteel nieuwe richtlijnen op komst

Page 123: cursus LC-MS 2001

123AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study (FDS)

FDS is an experiment in which the stability indicating power of the assay/purity method for DP has to be demonstrated

Page 124: cursus LC-MS 2001

124AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study (FDS)

Wordt meestal uitsluitend toegepast voor HPLC Is geen stabiliteitstudie!Wordt uitgevoerd door de ontwikkelaar zelf

Page 125: cursus LC-MS 2001

125AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study(FDS)

Zowel placebo als DP in oplossingTijdsduur experiment: 7 dagen4 meetpunten:

0 en 7 dagenTwee intermediairen

Maximale afbraak 15% na 7 dagenWelke condities?

Page 126: cursus LC-MS 2001

126AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study (FDS)

0.01N HCl 70°C 0.01N NaOH 70°C0.03% H2O2 25°CLicht Xe-lamp (cfr. ICH)

Page 127: cursus LC-MS 2001

127AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study (FDS)

Peak purity APIPeak purity gespecifieerdeonzuiverheden/afbraakproductenIdentificatie ‘belangrijkste’ onbekenden

Page 128: cursus LC-MS 2001

128AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.3 Forced Degradation Study(FDS)

LC-MS apparatuurPDA 200 - 400 nmMS:

ES ⇔ APcI+ ⇔ -Enkel Q

Page 129: cursus LC-MS 2001

129AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.4 Troubleshooting

Identificatie van onzuiverheden, afbraakproducten, ….

Relatief beperkt (Q) ⇒ hulp SpectroLC-MS niet afhankelijk van UV-chromofoor!

Page 130: cursus LC-MS 2001

130AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

6.5 Supportieve methode LC-MS

Bepaling van palmitinezuur in project R092670 d.m.v. ES negatief

Page 131: cursus LC-MS 2001

131AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

R092670

ProbleemPalmitinezuur geen UV-absorptiePalmitinezuur ES neg respons

Aantonen mogelijk?Semi-kwantitatief?

Afbraakproduct na bestraling?Afbraak in stabiliteit DP?

Page 132: cursus LC-MS 2001

132AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

UV + MS chromatogram palmitinezuur

AO-039517:17:57

22-Jun-20000.5% palmitinezuur lot 00310275

WJAN

5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00Time1

100

%

-9

100

%

palm_40 2: Diode Array 275 1.00Da

9.59e31.13

palm_40 1: Scan ES- TIC

4.86e651.86

Page 133: cursus LC-MS 2001

133AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

UV + m/z 255 van MR092670PUA021

AO-039502:06:36

23-Jun-2000MR092670PUA021

WJAN

5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00Time-5

100

%

0

100

%

R092670B 1: Scan ES- 255 1.00Da

1.26e351.98

19.5314.01

11.319.335.691.51

7.71

18.24

34.8623.28

26.13

30.83

29.01

44.8537.11

39.50

46.4350.07

58.3754.0355.38

60.72

R092670B 2: Diode Array 275 1.00Da

6.47e3

11.85

8.732.40

24.88

17.8441.31

37.53

55.95

48.16

Page 134: cursus LC-MS 2001

134AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

UV + m/z 255 van ZR092670EXA002

AO-039503:22:30

23-Jun-2000ZR092670EXA002

WJAN

5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00Time-3

100

%

0

100

%

R092670C 1: Scan ES- 255 1.00Da

4.48e351.98

16.7812.93

7.144.732.21 9.63

34.3319.4932.0828.7623.11

44.8337.20 40.65 50.5648.25 54.18 56.90

R092670C 2: Diode Array 275 1.00Da

8.04e3

24.8413.38

2.38 8.5317.77

19.52

37.53

34.78

29.35

41.30

39.42

46.55

48.15

55.91

54.31

52.49

Page 135: cursus LC-MS 2001

135AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

UV + m/z 255 van 00J16/F10

AO-053019:17:43

24-Nov-200000J16/F10 start

LVOS

5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00Time-1

100

%

1

100

%

MO-8-003-004 1: Scan ES- 255 1.00Da

6.61e651.90

4.191.03 9.036.69 14.5611.76 21.7918.16 43.7025.66 33.1329.16 37.56 40.13 47.56 49.9060.9054.67 57.63 64.03

MO-8-003-004 2: Diode Array 275 1.00Da

1.09e4

13.29

11.618.37

40.9634.60

23.6419.8237.26

46.24

54.4647.67

Page 136: cursus LC-MS 2001

136AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

SIR m/z 255

Page 137: cursus LC-MS 2001

137AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

UV van ZR92670EXA005 + SIR m/z 255

Page 138: cursus LC-MS 2001

138AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

7. Enkele praktische problemen

Page 139: cursus LC-MS 2001

139AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 1

Opletten met het gebruik van cyclodextrinen in stabiliteitsstalen

Page 140: cursus LC-MS 2001

140AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

HPβCD

Page 141: cursus LC-MS 2001

141AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 1:UV-chromatogramR160371

AO-053014:48:38

12-Oct-2000pH=6; 70 graden

WJAN

2.50 5.00 7.50 10.00 12.50 15.00 17.50 20.00 22.50 25.00 27.50 30.00 32.50 35.00Time-20

100

%

MO-R160371-1-1-005-3 2: Diode Array 230 1.00Da

5.00e424.04

1.14

Page 142: cursus LC-MS 2001

142AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 1:UV+MS chromatogram R160371

Stress sample

AO-053014:48:38

12-Oct-2000pH=6; 70 graden

WJAN

2.50 5.00 7.50 10.00 12.50 15.00 17.50 20.00 22.50 25.00 27.50 30.00 32.50 35.00Time8

100

%

-81

100

%

MO-R160371-1-1-005-3 2: Diode Array 230 1.00Da

1.25e424.041.14

MO-R160371-1-1-005-3 1: Scan ES+ TIC

5.09e97.03

6.13

4.781.28

3.43

2.83

8.088.48

24.08

23.7317.98

34.08

Page 143: cursus LC-MS 2001

143AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 1: MS spectrum

AO-053014:48:38

12-Oct-2000WJANpH=6; 70 gradenStress sample

150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900 950 1000m/z0

100

%

MO-R160371-1-1-005-3 140 (7.026) 1: Scan ES+ 6.18e7279.1

221.0

203.0

185.0

155.2

222.0

441.1

383.1

280.1337.0

325.0

384.0

385.1

603.0

545.0

499.0

442.1

487.2

500.1546.1

557.0

823.2661.2

604.1

605.0

765.1

662.2

719.3

663.4

766.2

767.0

985.3

881.2824.2

927.4

882.2928.2

986.2

Page 144: cursus LC-MS 2001

144AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 2: fosfaat buffer

Opletten met het gebruik van fosfaatbuffer in stabiliteitsstalen

Page 145: cursus LC-MS 2001

145AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 2: fosfaat buffer massaspectrum

Page 146: cursus LC-MS 2001

146AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 3: polymeren

Gebruik van polymeren in formuleringen, zoals OpaDry (PEG)

Page 147: cursus LC-MS 2001

147AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 3: polymeren

Page 148: cursus LC-MS 2001

148AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 4: modifiers en zouten

Modifiers beperken in concentratieS/N daalt (ion supression)Organische zouten kunnen bepaald worden

Page 149: cursus LC-MS 2001

149AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 4: modifiers en zouten

Voorbeeld: T002187DEA gebruikt in de HPLC methode

Page 150: cursus LC-MS 2001

150AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 4: modifiers en zouten

Page 151: cursus LC-MS 2001

151AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 4: modifiers en zouten

Page 152: cursus LC-MS 2001

152AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 4: modifiers en zouten

Page 153: cursus LC-MS 2001

153AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Voorbeeld 4: Na-acetaat

Page 154: cursus LC-MS 2001

154AO-MO-CHROM - 14 februari 2001 – cursus LC-MS 2001 - WJAN

Besluit

MS toegevoegde waarde MO-CHROMOnafhankelijk van UV-chromofoorZeer gevoeligSemi-kwantitatief (Q)Koppeling met GC,LC,CE,CEC,…MS/MS

identificatiekwantificatieverhoogde selectiviteit (HR)