Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr...

31
Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova

Transcript of Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr...

Page 1: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Current UK legislation and guidelines for radiation protection

of patients and staffPrepared by: Dr D. Mladenova

Page 2: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

General principles of the International Commission on

Radiological Protection (ICRP)

1.No practice shall be adopted unless its introduction produces

positive net benefit (Justification)

2.All exposures shall be kept as low as reasonably practicable

(ALARP)

Page 3: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Principles of ICRP- cont’d

Taking economic and social factors into account( Optimization)

• The dose equivalent to individuals shall

not exceed the limits recommended by the

ICRP (Limitation)

Page 4: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Regulations for intraoral radiography

• Tube voltage should not be lower than 50

kV preferably 70-90 kV

• Beam diameter should nor exceed 60 mm

• Rectangular collimation should be used

Page 5: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Intraoral radiography-cont’d

• Total beam filtration ( inherent and added )

• -1.5 mm aluminium disc for sets operating bellow 70 kV

- 2.5 mm aluminium for sets operating

above 70 kV

Page 6: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Intraoral radiography-cont’d

• The focal spot should be marked on

the outer casting of the tubehead

• Focal spot to skin distance ( FSD) should be at least 100 mm for sets operating below 60 kV and 200 mm for sets operating above 60 kV.

Page 7: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Intraoral radiography-cont’d

• Film speed controls and finely

adjustable exposure time settings should be provided

• The fastest film available( E or F speed) that will produce satisfactory diagnostic images should be used

Page 8: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Panoramic radiography

• Equipment should have a range of tube potential settings, preferably 60-90 kV.

• The beam height at the receiving slit of cassette holder should not be greater than

Page 9: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Panoramic radiography-cont’d

• greater than the film in use (normally 125 mm or 150 mm).

• The width of the beam should not be greater than 5 mm

Page 10: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Panoramic radiography-cont’d

• Equipment should be provided with adequate patient-positioning aids incorporating light beam markers

• New equipment should provide facilities for field-limitation techniques

Page 11: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Cephalometric radiography

• Equipment must be able to ensure the precise alignment of X-ray beam, cassette and patient

• The beam should be collimated to include only the diagnostically relevant area

Page 12: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Cephalometric radiography-cont’d

• To facilitate the imaging of the soft

tissues, an aluminium wedge filter

should be provided at the X-ray tube

Page 13: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

All equipment

• Should have a light on the control panel to show that the mains supply is switched on

• Should be fitted with a light and audible warnings that gives a clear and visible indication to the operator that an exposure is taking place

Page 14: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

All equipment-cont’d

• Exposure switches (timers) should only function while continuous pressure is maintained on the switch and terminate if pressure is released

• Exposure switches should be positioned so that the operator can

Page 15: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

All equipment-cont’d

• remain outside the controlled area

and at least 2 m from the X-ray tube and patient

Page 16: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Justification

-the availability and/or findings of previous radiographs

- the specific objectives of the exposure in relation to the history and examination of the patient

Page 17: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Justification-cont’d

- the total potential diagnostic benefit to the patient

- the radiation risk associated with the radiographic examination

Page 18: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Justification-cont’d

• the efficacy, benefits and risks of

alternative techniques having the

same objective but no or less

exposure to ionizing radiation

Page 19: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Lead protection

• There is no justification for the routine use of lead aprons for the routine use of lead aprons for patients in dental radiography

• Thyroid collar to be used only in maxillary occlusal radiography

Page 20: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Lead protection-cont’d

• Lead aprons do not protect against scattered radiation for adult who support a patient during exposure

• Lead aprons should not be folded

Page 21: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Specific requirements for pregnant women

• Only radiographs that are absolutely necessary are taken

• ALARP and the patient is given the option to delay the radiography

Page 22: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Selection criteria in Dental Radiography 1998

• No radiographs should be taken without a history and clinical examination

• New patient

- Child (primary and mixed dentition)- panoramic and bitewings radiographs

Page 23: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Criteria for radiographs-cont’d

- Adult (dentate patient)-patient-specific radiographs depending on clinical examination

- Edentulous patients- panoramic and/or periapical radiographs in selected areas

Page 24: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Dose limitations and annual dose

• Patients

• Radiation workers ( classified and non- classified)

• General public

Page 25: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Radiographic investigations for patients

• Examinations directly associated with illness

• Systematic examinations

• Examinations for occupational, medico-legal or insurance purposes

• Examinations for medical research

Page 26: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Annual dose limits-cont’d

- Non-classified workers 6 mSv

- General public 1 mSv

• Dose Constraints - non-classified workers 1 mSv

Page 27: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Annual dose limits-cont’d

- For employee not directly

Involved with radiography and

General public 0.3 mSv

- Pregnant staff member 1 mSv

Page 28: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Source of radiation to dentist and their staff

• The primary beam, if they stands in its path

• Scattered radiation from the patients

• Radiation leakage form the tubehead

Page 29: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Protective measures• Personal monitoring is recommended for practice exceeding 100 intraoral or 50 panoramic film

• Staff should stand outside the controlled area and not in the line of the primary beam

Page 30: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Protective measures-cont’d

• Safe use of equipment

• Safe use of radiographic techniques

Page 31: Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.

Main methods of monitoring and measuring radiation dose• Film badges

• Thermoluminescent dosemeters

• Ionizing chambers