Crystal16: Best practices to perform the tune step
Click here to load reader
-
Upload
technobis-crystallization-systems -
Category
Spiritual
-
view
689 -
download
0
Transcript of Crystal16: Best practices to perform the tune step
Workflow White Paper
Avantium 2009 & Confidential
C r y s t a l l i z a t i o n S y s t e m s
T u n i n g t u r b i d i t y
Tuning turbidity
Avantium Crystallization Systems 2
Content
EXPLANATION OF THE TUNE STEP .................................................3
1. EXAMPLE CRYSTAL16® ...............................................................4
2. EXAMPLE CRYSTALLINE® ...........................................................5
Tuning turbidity
Avantium Crystallization Systems 3
Explanation of the tune step
The turbidity sensor of the Crystal16® and the Crystalline® can be tuned
to make the detection of the phase transition more sensitive and accurate.
The tune step can be programmed by the user and automatically adjusts
the light source intensity of each turbidity sensor to match the light range.
Two examples are shown in Figure 1.
Each new experiment in the Crystal16® and Crystalline® can best be tuned
at the start of the experiment with a clear “blank” and/or at the point
where all solids are in solution. At the beginning of the experiment, the
turbidity sensor can best be tuned with a reactor containing only the
solvent of the solutions.
Note: In case the compound colors the solution dark, the maximum light
transmission at a clear solution will be below 100%.
Figure 1: Explanation of tuning step with
a) Clear solution, b) Colored solution.
Tuning turbidity
Avantium Crystallization Systems 4
1. Example Crystal16®
An example program for the Crystal16® is given in Figure 2. Take care to
remove the blank and place the sample inside the reactor holder after the
stirrer is switched on. If you are not interested in the dissolution point in
the first heating cycle, remove the first tuning step from the program and
tune after complete dissolution.
Figure 2: Example program Crystal16®.
Tuning turbidity
Avantium Crystallization Systems 5
2. Example Crystalline®
An example program for the Crystalline® is given in Figure 3. Take care to
remove the blank and place the sample inside the reactor holder after the
second prompt in the program. If you are not interested in the dissolution
point in the first heating cycle, remove the first tuning step from the
program and tune after complete dissolution.
Figure 3: Example program Crystalline
Tuning turbidity
Avantium Crystallization Systems 6