CRITICAL ASPECT ANALYTICAL TEST REVIEW · 2017. 12. 28. · Monitoring (EM), Microbial Limit Test...
Transcript of CRITICAL ASPECT ANALYTICAL TEST REVIEW · 2017. 12. 28. · Monitoring (EM), Microbial Limit Test...
-
CRITICAL ASPECT ANALYTICAL TEST REVIEW
Jakarta 14 December 2017
Speaker:
HERU PURNOMO, ST
-
ispe.orgConnecting Pharmaceutical Knowledge 2
Start
Review Lab Result
Submission
Lab Record to QA
Non Conformance or Out
of Specification
Investigation
Meet Specification
Stop
Sample Received
(In-Process & Finished
Good)
Testing
BATCH DISPOSITION
Yes
No
QC WORK FLOW
-
ispe.orgConnecting Pharmaceutical Knowledge 3
Analytical Test Result
Analytical Worksheet
Analytical Report
Certificate of Analysis
(CoA)
The analytical test report
is a compilation of the
results and states the
conclusions of the
examination of a sample
Internal document to be
used by the
analyst for recording
information about the
sample, the test
procedure,
calculations and the
results of testing. It is to
be complemented by
the raw data obtained in
the analysis
The list of test procedures
applied to a particular
sample with the results
obtained and the
acceptance criteria
applied. It indicates
whether or not the
sample complies with the
specification
-
ispe.orgConnecting Pharmaceutical Knowledge 4
Certificate of Analysis (CoA)
• A specification is defined as a list of tests, references to
analytical procedures, and appropriate acceptance criteria that
are numerical limits, ranges, or other criteria for the tests
described
• It establishes the set of criteria to which a drug substance or
drug product should conform to be considered acceptable for
its intended use. “Conformance to specifications” means that
the drug substance and/or drug product, when tested according
to the listed analytical procedures, will meet the listed
acceptance criteria
-
ispe.orgConnecting Pharmaceutical Knowledge 5
Review of Analytical Test Result
Written procedure should be in place when reviewing analytical result with
the following information:
• Provide general guideline for checking of analytical raw data, worksheet
and associated document in QC Lab
• Applies to GMP analytical raw data in physical – chemical, microbiology
and raw material
• Preparation steps, observation and calculation step should be
documented and show on the worksheet, these associated
documentations have to be reviewed for completeness and accuracy as
well
• The date of execution and analyst / chemist / microbiologist who perform
the testing is to be identified on worksheet
• All printout of raw data should be attached to the analytical worksheet.
Printout must signed or initialled across the edge of attachment
-
ispe.orgConnecting Pharmaceutical Knowledge 6
HPLC / GC Chromatographic Test
The following review for chromatographic data should be done for completeness
and accuracy:
1. Program setup:
Flow rate, detector wavelength, column temperature, mobile phase (e.g.
isocratic / gradient) & composition, need wash / seal wash setup
2. Sequence Setup
Injection Volume, Column ID, Sample Type, Sample Description, Sample
Weight, Reference accuracy weight
3. Integration Setup
SST Setup, Reference Standard Weight Entry, retention setup and
allowable window, proper integration peak
4. Chromatogram
Examine the blank chromatogram and note any extra peaks, particularly
those that co-eluate with any peak interest
5. Audit trail
Check data integrity of sequence by verifying the audit trail for any
unauthorized changes
-
ispe.orgConnecting Pharmaceutical Knowledge 7
HPLC / GC Chromatographic Test (cont)
6. Check the expiry of standard and test mix, use in the analysis, against the
relevant reference material certificate or preparation. Ensure they are within
the expiry date and preparation correct
7. Check that specified system suitability result are met against the relevant
testing method
8. Check the accuracy of all values that are manually calculated
-
ispe.orgConnecting Pharmaceutical Knowledge 8
Dissolution
Check should be carried out on both the dissolution procedure and method
analysis:
1. Dissolution medium used and the operating temperature
2. Type of Aparatus
3. Preparation standard and sample
4. Sampling procedure and technique (e.g. intervals, filtration, handling of
samples, dilution)
5. Calculations
6. The blank and standard solution spectra or chromatograms supporting the
dissolution test should either be available electronically or documented in lab
worksheet
-
ispe.orgConnecting Pharmaceutical Knowledge 9
Microbiological Testing for Water, Environment
Monitoring (EM), Microbial Limit Test (MLT)
1. Check that the analytical worksheet and SOP used were correct and the
sampling points were accurately recorded. Check the presence and
accuracy of the following minimum information:
• Date of sampling / performed date of testing
• Date of reading plate / result recording date
• Indicate the lot reference and expiry date of media used and verified that
the lot reference used is within expiry date at point of usage
• Date and temperature of incubation
• Equipment ID and ensure equipment are still within their qualification /
calibration status at point of usage
• Test result of negative or positive control which was carried out in
accordance to the test method
• Water sampling point (for microbiological testing for water) / legend (for
environment monitoring)
• Initial of analyst & reviewer and date
2. Check the result against the acceptance criteria in the relevant test
procedure
-
ispe.orgConnecting Pharmaceutical Knowledge 10
Other Tests
Hard copy, generated from test (for example: UV, infrared absorption)
should bear the following information:
Initial / name of analyst
• Sample Identity
• Instrument Number
• Spectrum (where applicable)
• Testing date
• Instrument Parameter (where applicable)
• Result / data obtained
-
ispe.orgConnecting Pharmaceutical Knowledge 11
Tolerance Limit & Specification Check
• If the test method requires tolerance limit check for the replicate
runs, check the replicate value against the pre-defined tolerance
limit in the test method
• Check the reported results against specification and trend limit (if
any)
• The review period of a GMP analytical result should be recorded in a
timely manner (e.g. 24 working hours from the test completion date)
-
ispe.orgConnecting Pharmaceutical Knowledge 12
Trend Rule Violation – Out of Trend
There are three (3) trend rule violations:
1. Result out of trend limits
2. Nine (9) or more consecutive data points shifted above or below the center line
3. Six (6) or more data points consecutively increasing or decreasing
-
ispe.orgConnecting Pharmaceutical Knowledge 13
OUT OF TREND (OOT) - NELSON TEST
Rule # 1
Any point outside
control limit
Rule # 2
9 consecutive point on
same side centreline
Rule # 3
6 consecutive point
decreasing or
increasing
-
ispe.orgConnecting Pharmaceutical Knowledge 14
What is an Audit Trail?
1. Systematic “story” of the data from creation, through interpretation
and final assessment and report
2. Each functional area that have GxP computer systems need to
perform audit trail
3. The information should be recorded in Audit Trail:
• Date and Time
• Name of person making change
• Original Change and value
• Reason of change made
• Invalid attempts to log on the system
If audit trail procedure is not available, the company has no
possibility to detect violations to data integrity executed by site
Operators due to errors / malfeasances, unintended or deliberate
-
ispe.orgConnecting Pharmaceutical Knowledge 15
Regulatory Guidance for Audit Trail Review
-
ispe.orgConnecting Pharmaceutical Knowledge 16
Audit Trail Review
• Evaluation of process risk associated to the system intended use
• An assessment must be performed for each system in order to
identify the most adequate method
-
ispe.orgConnecting Pharmaceutical Knowledge 17
Audit Trail Review
Risk class identification to define critical point:
• Frequency of audit trail
• Sampling Plan: The number of record to be reviewed
-
ispe.orgConnecting Pharmaceutical Knowledge 18
Audit Trail Review
Create form to define the
verification to be executed
and the relevant
acceptance criteria
-
ispe.orgConnecting Pharmaceutical Knowledge 19
Common Question During Inspection
• Is the audit trail activated? SOP?
• Record of review?
• How to prevent or detect any deletion of modification of audit trail
data? Training staff?
• Is predicate rule principle followed for changed?
Preserve of original data
Corrected data
Date of correction
Name of person who correction the data
Justification comment for correction
• What if there is no audit trail function?
• Can you prove data manipulation did not occur?
-
ispe.orgConnecting Pharmaceutical Knowledge 20
-
ispe.orgConnecting Pharmaceutical Knowledge 21
-
ispe.orgConnecting Pharmaceutical Knowledge 22
QA Responsibility
Do I need to review Analytical test result?
System & Facility
Culture
Confident
• Create robust procedure to review analytical test
result
• Qualified personnel
• Allocate budget to reduce potential data integrity
issue
• Develop management practises and supervision to
encourage reporting problem
• Identify root causes without finger pointing
Random check based on risk assessment
approach
-
ispe.orgConnecting Pharmaceutical Knowledge
http://thegledaproject.com/thank-you/http://thegledaproject.com/thank-you/
-
ispe.orgConnecting Pharmaceutical Knowledge