CR to DR Optimizing Image Quality and Dose
Transcript of CR to DR Optimizing Image Quality and Dose
CR to DR – Optimizing Image Quality and Dose
Bruce Apgar B.S.
George Curley R.T.
Agfa Healthcare Inc.
Greenville, SC
What is DR? Mobile Radiography:
What is DR? Retrofit DR Panels:
Floor mounted
Ceiling mounted
R/F rooms
What is DR? In Room Systems:
Benefits:
• Workflow Improvement / Automation
• Dose Reduction
• Improved Image Quality
Radiology Today: Radiology Today July 2014
Benefits:
• Cost coming down
• “Need for speed”
• Detector sharing
Radiology Today: Radiology Today April 2012
Why the Move from CR to DR?
Workflow Improvements:
Loma Linda University Medical Center Case Study
• DR Retrofit conversion from distributed multi-plate CR
Outcomes:
• Streamlined workflow & time savings
• 8.16 seconds less time per exam (portable chest, bone
surveys)
• More patients per day / year
• 100% FTE gain
• Rapid ROI
Why the Move from CR to DR?
Best Practices, ASRT Advisory Council 2012:
• Background
• Dose / ALARA
• Social Marketing and Safety Initiatives
• ACR Guidelines before getting to the scope of the white paper
ASRT HCIAC:
ASRT White Paper
Making the Move from CR to DR…
Challenges: Minimizing Patient Dose with Pediatric DR
– Equipment variation
– Technique guides not always provided
– Large detectors (AED target area)
– Artifacts from pediatric positioning devices may be more easily seen
– Repeat/Reject Analysis implementation critical
ASRT Directed Reading: Radiation Safety
Compliance; Radiologic Technology,
volume 87 - number 5 May/June 2016:
page 522
Making the Move from CR to DR…
DR: Image Capture Technologies
Electronic control
Triggers the switching diodes
X-ray scintillator screen or charge collector
Converts x-rays to light
or to electric charges
TFT array
Collects charges from the upper
layer
Switching diodes
Connects each pixel to readout device
in
out
Analog-to digital
conversionMultiplexer
Readout the electronic signal
*Lança L., Silva A. (2013) Digital Imaging Systems for Plain Radiography, New York: Springer
DR Panel Technology*
1. Absorption of X-rays in the phosphor screen
2. X-ray is converted into visible light
3. Photo diodes read light signal and generates charges
4. Signals are amplified, digitized, processed and archived
5. Soft Copy display or hardcopy optional
Image Capture Technologies
Image Quality ImprovementRequires Improved Phosphor Technology
Calcium Tungstate for Film
Barium Fluoride Bromide (BaFlBr) for CR
Gadolinium Oxysulfide (GOS) for DR
High Efficiency Needle PhosphorsTraditional Powder Phosphors
Cesium Iodide (CsI) for DR
Cesium Bromide (CBr) for CR
The thickness of CR BaFlBr powder phosphor layer is limited to less than 300 µm, because of light
scattering. This thickness limits the X-ray absorption.
Due to the low light scattering of CsI a thicker phosphor layer can be used without jeopardizing the
sharpness of the imaging system. Higher X-ray absorption is possible with DR detectors using CsI
needle crystalline phosphors. This results in lower dose and better image quality.
Phosphor Technology
CR- BaFlBr DR- CsI
70 kVp 10 mAs70 kVp 10 mAs
CR DR
70 kVp 10 mAs
Image Quality Improvement
Image Quality Improvement and Lower Dose
70 kVp 10 mAs 70 kVp 5 mAs
CR DR
DR Cesium Iodide (CsI) Phosphors*
• Significantly reduce internal light scatter
• Improve image sharpness
• Enable thicker phosphor layers
• Improve X-ray absorption
• Reduce Dose ~ 2 X
• Reduce scatter sensitivity
* When compared to BaFlBr CR phosphor plates
Image Processing Technology
Image Processing Technology
14 µGy 55 µGy 225 µGy
Image processing can improve usable diagnostic
information at lower dose
Standard Processing 14 µGy Multi Scale
Neonatal Processing
Image Processing Can Reduce Image Noise
Standard Processing Noise Reduction Technology
Proper Image Processing Should:
• Provide consistent performance • For patient body habitus & age: Neonatal, Pediatric, Adult, Bariatric • Over a wide range of exposure factors
• Increase productivity, not create more work• Little or no post processing should be needed • Should require few window / level adjustments, electronic masking, manual ROI selection
• Should not create artifacts
• Should be dose tolerant & low dose friendly
• Be easy to configure and set up • Work well out of the box• Provide simple understandable adjustment settings• Not complex parameters requiring imaging specialists for set up
Anti-Scatter Grids for DR “Key Factors”
• Grid Ratio
• Bucky factor: amount by which the exposure has to be adapted
• Grid Positioning • Angle and Distance
• Stationary vs. Reciprocating
• Line Rate (frequency) in lines/cm or lines/inch
• Grid line direction
• Focus vs. parallel
Grid Performance
Factors
• Very important clinically
• Very important clinically
• Very important clinically
• Critical
• Critical
• Critical
• Least important to digital
How it will Impact Digital
Image Quality:
Purpose:
Identifies repeating patterns caused by grid/panel interference and removes them
Result:
Improved viewing conditions
Grid Line Suppression Algorithms
DR Grid Recommendations*
* With Agfa Grid Line Suppression Software - GLS10% variation In
grid specification
Panel Type DX-D 30 / 35 C DX- D 40 C DX-D 45 C CR
Plate
Resolution 125 micron 139 micron 124 micron 100-150 micron
32 lines / cm
80 lines / inch
36 lines / cm
90 lines/ inch
40 lines / cm
103 lines / inch
50 lines / cm
132 lines / inch
70 lines / cm
178 lines / inch
80 lines / cm
215 lines / inch
Poor Results
Not Recommended
Poor Results
Not Recommended
Poor Results
Not Recommended
Better Results
Recommended for Use
Better Results
Recommended for Use
Better Results
Recommended for Use
Better Results
Recommended for Use
Better Results
Recommended for Use
Good Results
Acceptable for Use
Good Results
Acceptable for Use
Good Results
Acceptable for Use
Poor Results
Not Recommended
Best Results
Recommended for Use
Best Results
Recommended for Use
Best Results
Recommended for Use
Best Results
Recommended for Use
Poor Results
Not Recommended
Poor Results
Not Recommended
Poor Results
Not Recommended
Poor Results
Not Recommended
Poor Results
Not Recommended
Poor Results
Not Recommended
Poor Results
Not Recommended
Good Results
Acceptable for Use
Tube / collimator angle
- 20 degrees
Grid Angle -5 degrees
Grid
Incorrect Grid Alignment = Poor Image Quality
Angle of tube collimator is not the same as the grid/panel Tube/collimator is not parallel to the grid
• Most “portable” stationary grids are the same size as the cassette being used
• Grid lines are often oriented along the long dimension
• May also be oriented along the short dimension (usually called “decubitus” grids)
• Grids in tables and upright buckys are usually 17 – 20 inch “square” grids
Typical grid line orientation “Decubitus” grid orientationSquare table or upright
bucky grid
Tube Collimator and Grid are NOT Parallel (6:1)
High Scatter Poor Lung and Spine Detail
Grid
Proper Grid Alignment = Good Image Quality
Angle of tube the same as the grid/panel
Tube collimator parallel to the grid
Tube Collimator and Grid ARE Parallel (6:1)
Low Scatter Improved Lung and Spine Detail
Tube Collimator and Grid are not parallel
High Scatter Poor Lung and Spine Detail
“Decubitus” grid orientation
Angled positioning but Good Lung and Spine Detail
Tube / collimator angle
- 20 degrees
Grid Angle -5 degrees
Grid
Incorrect Alignment = Poor Image Quality
Angle of tube collimator is not the same as the grid/panel Tube collimator is not parallel to Grid
Tube / collimator angle
- 20 degrees
Grid Angle -5 degrees
Grid
Angled Tube = Good Image Quality
Angle of tube collimator is off but it is centered A “Decubitus Grid” was used
“Decubitus” grid orientation
Grid Positioning and Orientation
• Most “portable” stationary grids are the same size as the panel being used
• Grid lines are often oriented along the long dimension
• May also be oriented along the short dimension (usually called “decubitus” grids) these good for landscape chests
• Grids in tables and upright buckys are usually 17 – 20 inch “square” grids
Typical grid line orientation“Decubitus” grid orientation Square table or upright
bucky grid
“Non-Grid”, Scatter Suppression Software
• Product Types • Agfa MUSICA Chest +
• Fuji Virtual Grid
• Philips Skyflow
• Other ?
• How does it work ? • Reduces the need for a grid using advanced image processing
• Scatter radiation is low frequency in an image
• Scatter Suppression Software reduces the visualization of low frequencies while enhancing medium to high frequencies
Low Frequency Scatter Contribution
Scatter Suppression Software extracts low frequency scatter
information while enhancing relevant clinic information
Insert Non Grid Beside Chest Image
GenRad Soft
Non Grid Chest with Standard Processing
Low Frequency Scatter Contribution
Scatter Suppression Software extracts low frequency scatter
information while enhancing relevant clinic information
Insert Non Grid Beside Chest Image
GenRad Soft
Non Grid Chest with Scatter Suppression
Software
Scatter Suppression Software can improve lung field detail without
the use of an anti-scatter grid
Chest Phantom with 6:1 Grid
MUSICA 2 GenRad Soft Chest Processing
100 kVp 3.2 mAs
Chest Phantom no Grid
MUSICA 3 Chest + Processing
100 kVp 1.6 mAs
flipped for comparison
* When compared to non grid exposures , grid exposures may require up to 50% higher exposure depending on conditions
Scatter Suppression Software may also be used with a Grid to give the
best overall results but requires proper positioning at a higher dose*
Chest Phantom with 6:1 Grid
MUSICA 2 GenRad Soft Chest Processing
100 kVp 3.2 mAs
Chest Phantom with 6:1 Grid
MUSICA 3 Chest Processing
100 kVp 3.2 mAsflipped for comparison
MUSICA 2 GenRad Soft Processing
Non Grid
Clinical Examples
MUSICA 3 + Chest Processing
Non Grid
Can Image Processing Software Eliminate the Need for a Grid with DR?
• It Depends• On the application
• Mobile or In room
• On the Patient Size• Pediatric, Normal, Obese
• On the workflow requirements• Grid and Panel Weight
• On the examination criteria • Image Quality
• Dose
• “Non-grid” image processing is another option in the toolbox that should be considered and used when appropriate to improve image quality and reduce dose.
Managing Exposure and Dose to Optimize Image Quality
5 mAs 10 mAs 20 mAs
1/2 ref dose ref dose 2x ref dose
Dynamic Range: Film
Dynamic Range: CR
2.5 mAs 10 mAs 160 mAs
Dynamic Range: DR
1/4 ref dose ref dose 4x ref dose
2.5 mAs 10 mAs 40 mAs
Managing Exposure and Dose to Optimize Image Quality
70 kVp 160 mAs 70 kVp 160 mAs
CR DR
• Better image quality at lower dose • DR images can be higher in contrast and sharpness
• DR panels (CsI) require less exposure to achieve equal to or better image quality
• Significantly less Exposure Latitude with DR• CR Exposure Range -4x to +16 x mAs
• DR Exposure Range ± 4x mAs
• Image Saturation can occur with DR !!!! Data is not recoverable !!!
• With DR Accurate exposure is key (similar to film)
CR vs DR Image Quality and Dose
To maintain proper exposure
One Uniform Standard required for measuring exposure
!!!!Exposure in
µGyFuji Canon Agfa Carestream
IEC
Exposure
Index *
2.5 710 30 1.96 1451 250
5 355 60 2.26 1751 500
10 177 120 2.56 2051 1000
20 89 240 2.86 2351 2000
*Flatfield RQA5 beam quality
Managing Exposure and Dose to Optimize Image Quality
• A standard way to measure the exposure to a digital detector • Developed by IEC - International Electrotechnical Commission
• IEC 62494-1 “Exposure index of digital X-ray imaging systems”
• Standard Calibration Condition – RQA-5
• Designed to monitor exposure consistency within an exam type
• Consists of three values • Exposure Index – EI
• Target Exposure Index - TEI
• Deviation Index - DI
• In a perfect situation the DI would be zero (0)
A Better Approach; Exposure Index (IEC 62494-1)
DI =10 * Log TEI[ EI_ ] 500500_[DI =10 * Log ] = 0
Actual
Exposure
Index
Target
Exposure
Index
Deviation
Index
Exposure
Factor
%
Change
1300 500 4 2.6 160%
1000 500 3 2 100%
800 500 2 1.6 60%
630 500 1 1.26 26%
500 500 0 1 0%
400 500 -1 0.8 -20%
300 500 -2 0.6 -40%
250 500 -3 0.5 -50%
200 500 -4 0.4 -60%
Target Exposure Index and Deviation Index
“On Target” Exposure
EI = TEI
DI = 0
60 kVp 1 mAs
. Exposure Monitoring Software
1. Color Coded Exposure Bar
2. Exposure Index Performance Table
60 kVp 1.75 mAs
Green
within range
Overexposed within range Less
than + or - 2X exposure
Less than + or – 3.0 DI
The dose bar is an indication to see how far the applied Exposure Index
(Detector exposure) is away from the reference exposure (Target Exposure
Index) identified for this exposure.
Overexposed outside of range
Greater than 2X exposure
Greater than 3.0 DI
The bar gives a relative indication of the exposure to the plate and it is a good
measure of the variation of exposure to the plate within a given exam type, but it
is not an absolute dose measurement value.
60 kVp 2.88 mAs
Yellow Slightly
Overexposed
60 kVp 4.75 mAs
Overexposed far
outside of range.
Greater than 4X exposure
Greater than 6.0 DI
Red Significantly
Overexposed
Underexposed far
outside of range.
Less than 1/4X exposure
Less than - 6.0 DI
60 kVp 0.23 mAs
Red Significantly
Underexposed
Automatic ROI Selection
Deviation Index in range
EI 264
DI -0.6
TEI 300
Region of Interest ROI Selection Affects Exposure Index
Manual ROI Selection of appropriate
region for examination
Deviation Index in range
EI 375
DI 1.0
TEI 300
Region of Interest ROI Selection Affects Exposure Index
Manual ROI Selection of incorrect
region for examination
Deviation Index out of range
EI 64DI -6.7
TEI 300
Region of Interest ROI Selection Affects Exposure Index
Region of Interest ROI Selection Affects Exposure Index
Manual ROI Selection of
incorrect region for examination
ROI outside of clinical region
Deviation Index out of range
EI 1924
DI 8.1
TEI 300
Ongoing Exposure - Trend Analysis
Dose Monitoring
DICOM Mapping of EI, TEI, DI
The DICOM committee defined tags:
• EI : (0018,1411)
• TEI : (0018,1412)
• DI : (0018,1413)
• “DAP” Dose Area Product: (0018,115E)
Proper X-Ray Collimation for
region of interest
If the X-ray collimation is done correctly the area of
interest should be detected automatically and minimal
manual cropping should be required
Proper X-Ray Collimation is Key to Dose and Image Quality with DR
Proper Automatic image Collimation
(black borders) for region of interest
No Manual Collimation Required
Collimation Issues
Original Plate size 24 x 30 cm
X-Ray Collimated area size 20 x 28 cm (77%)
Region of interest size only 16 x 15 cm (33%)
Automatic collimation with black borders applied based on X-Ray collimation
The technologist does additional manual
collimation (cropping) at the workstation
The Radiologist may be unaware of the actual
patient exposure – quantity and anatomically
If the X-ray collimation is correct, minimal
manual cropping should need to be done
The actual region of interest size is 16 x 15 cm (33%)
Repeat Rates are Changing with DR
1. Panel Technology and Phosphor Type - Can reduce dose 50% or more
2. Proper Image Processing - Can significantly reduce dose and improve image quality
3. Proper Grid selection and position – Will greatly influence image quality
a. Grid lines per inch
b. Grid position and distance
c. Non Grid Scatter Suppression software - Is an option to improve workflow and dose
4. Repeat Rates with DR are increasing
a. Repeats for positioning are increasing because it is so easy to repeat
CR to DR – Optimizing Image Quality and Dose
5. Proper Technique Selection is more important than ever
a. DR Dynamic Range is much less than CR
b. DR images can be saturated (image recovery is not possible)
c. The IEC Exposure Index can be used to monitor and control exposure
6. X-ray Collimation
a. Influences scatter, image processing and overall image quality
b. Improper Electronic collimation or masking reduces image quality and increasesdose
CR to DR – Optimizing Image Quality and Dose