CPB 5000Pressure Balance
-Product overview-
2
CPB product line
CPB 5000„Standard“
CPB 3800„Compact Line“
CPB 5000„Premium“
HighEnd / Laboratory
Accuracy up to 0.008 % Accessory package included CPU 5000 included
Industrial / Standard
Accuracy up to 0.015 % High-performance Optional upgrade for up to 3 systems
Starter
Accuracy up to 0.025 % Compact instrument base Only basic functions without options
3
Applications of Pressure balances
Primary standard for defining the pressure scale in ranges of: up to 100 bar / 1,500 psi pneumatic respectively up to 400 bar / 5,000 psi pneumatic (with gas to oil separator) and up to 5,000 bar / 70,000 psi hydraulic
Reference instrument for testing, adjusting and calibrating pressure measuring instruments in factory and calibration laboratories
Self-sustaining complete system also suitable for on-site measurements / calibrations (no need of power supply and external pressure supply)
4
• Pressure P is applied on a known Area A of a piston and balanced by the Force F generated by a known Mass M
• P= F/A, which is then corrected by “environmental parameters”: Local gravity Distortion coefficient Temperature effect, etc …
Fundamental pressure standard
PistonPiston
CylinderCylinder
Working principle
5
ConTect – System
= Piston-Cylinder System (patented)
... connect and protect
Protection for the high-quality piston-
cylinder
Quick change – without any tool or
M30 x 2 male thread
Available for different pneumatic and
hydraulic ranges
6
The 3 basic components of a pressure balance
The instrument baseAdaptation of the piston cylinder systemConnection of the test deviceControl elements for pressure generation
The piston cylinder systemMeasuring system
The mass-setApplies the force to the measuring system
7
Specifications: Accuracy: up to 0.008% of reading Pressure ranges: up to 100 bar
(pneumatic)
up to 400 bar (pneumatic
with gas to oil separator)
up to 1,000 bar (hydraulic)
Pressure Balance CPB5000
Features: Very high long term stability Accuracy of reading Pressure generation with priming pump included Patented ConTect-systems for an easy
change of the measuring range
… high ease of use with best performance
Further specifications see data sheet CT 31.01
8
Assembly CPB 5000
Piston cylinder system with ConTect quick connection
Piston cylinder system with ConTect quick connection
Quick connection for test item
Quick connection for test item
Test itemTest item
Water levelWater level
Removable star handleRemovable star handle
Priming pumpPriming pump
ConTect fixtureConTect fixture
Spindle pump for pressure generation/fine adjustment
Spindle pump for pressure generation/fine adjustment
Needle valveNeedle valve
WeightsWeights
Pressure gaugePressure gauge
Rotatable feetRotatable feet
9
Accessory for Pressure Balances
CalibratorUnit CPU5000
Highest accuracy achieved through permanent
measurement
and correction of current ambient conditions
(air pressure, ambient temperature, air humidity)
Calculation of mass-load
incl. power supply and multimeter for
transmitter calibration as an option
10
CalibratorUnit - Packages
CPU5000-Basic (Computer only)
calculation of mass-load manual input of all values
Metrology-Extension
measurement of: humidity air pressure ambient temperature piston-cylinder-temp.
Transmitter-Extension
measurement of: power supply transmitter signal (V, mA) incl. conversion to pressure
Visual-Extension
touchlessmeasurement of:
piston float-position