Pharmacology & Toxicology Core Facilities SOP #: 18654.B446B.001 Version #: Author: Brian Jespersen Page of 10 Approval Date: 1 Core Equipment ID: 18654 Description: Applied BioSystems, 7500 Real Time PCR System Room: B446B (Molecular & Biochemical Core) Champion: 1.0 Purpose Standardize the process of control, maintenance and ownership, of the Applied BioSystems, 7500 Real‐Time PCR System located in Room B‐446B Life Sciences. 1.1 Applied BioSystems, 7500 Real-Time PCR System Capabilities 1 Is a powerful platform for labs requiring superior performance and maximum dye versatility. The 7500 Real‐Time PCR System is a sophisticated platform for users who require extended capabilities and maximum versatility. The 3rd generation platform features an innovative optical system that enhances sensitivity and lets you access a broader range of fluorophores. The variable excitation capability allows greater sensitivity for longer wavelength (red) dyes. a. Powerful, five‐color platform is calibrated for the broadest range of dyes available: FAM™/SYBR® Green I, VIC®/JOE™, NED™/ TAMRA™/ Cy3™, ROX™/Texas Red®, and Cy5™ dyes. b. Specialized optical system enables easy and accurate calibration to new dyes without requiring the addition of new filter sets. c. Advanced multi‐componenting algorithm minimizes spectral crosstalk – superior for multiplexing. d. User‐friendly software includes plate set‐up wizards, multi‐ plate data viewing capabilities, and advanced analysis tools to make data processing simple and straightforward. 1.2 7500 Software v2.0, Software Capabilities 2 1 7500 Real‐Time PCR System. Applied Biosystems. 25 August, 2008. https://products.appliedbiosystems.com:443/ab/en/US/adirect/ab?cmd=catNavigate2&catID=601251&tab=DetailInfo . 2 7500 Real‐Time PCR System. Applied Biosystems. 25 August, 2008. https://products.appliedbiosystems.com:443/ab/en/US/adirect/ab?cmd=catNavigate2&catID=601251&tab=DetailInfo .

Transcript of Core Equipment IDcore.phmtox.msu.edu/Scheduling/ItemDocs/13/SOP_ABS_Real_Time_P… · Core...

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

Core Equipment ID:  18654 

Description: Applied BioSystems, 7500 Real Time PCR System 

Room: B446B (Molecular & Biochemical Core) 

Champion:  

 

1.0 Purpose Standardize  the  process  of  control,  maintenance  and  ownership,  of  the Applied  BioSystems,  7500  Real‐Time  PCR  System  located  in  Room B‐446B Life Sciences.   

1.1 Applied  BioSystems,  7500  Real­Time  PCR  System Capabilities1  Is  a  powerful  platform  for  labs  requiring  superior  performance  and maximum  dye  versatility.  The  7500  Real‐Time  PCR  System  is  a sophisticated  platform  for  users  who  require  extended  capabilities and  maximum  versatility.  The  3rd  generation  platform  features  an innovative  optical  system  that  enhances  sensitivity  and  lets  you access  a  broader  range  of  fluorophores.  The  variable  excitation capability allows greater sensitivity for longer wavelength (red) dyes. 

 a. Powerful,  five‐color  platform  is  calibrated  for  the  broadest 

range  of  dyes  available:  FAM™/SYBR®  Green  I,  VIC®/JOE™, NED™/ TAMRA™/ Cy3™, ROX™/Texas Red®, and Cy5™ dyes.  

b. Specialized  optical  system  enables  easy  and  accurate calibration to new dyes without requiring the addition of new filter sets. 

 c. Advanced  multi‐componenting  algorithm  minimizes  spectral 

crosstalk – superior for multiplexing.  

d. User‐friendly  software  includes  plate  set‐up  wizards,  multi‐plate data viewing capabilities, and advanced analysis tools to make data processing simple and straightforward.  

1.2  7500 Software v2.0, Software Capabilities2 

                                                         17500 Real‐Time PCR System. Applied Biosystems. 25 August, 2008. https://products.appliedbiosystems.com:443/ab/en/US/adirect/ab?cmd=catNavigate2&catID=601251&tab=DetailInfo. 2 7500 Real‐Time PCR System. Applied Biosystems. 25 August, 2008. https://products.appliedbiosystems.com:443/ab/en/US/adirect/ab?cmd=catNavigate2&catID=601251&tab=DetailInfo. 

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

Now  the easy‐to‐use StepOne  software  is  available  for  the 7500 and with  the  7500  Software  v2.0  upgrade.  The  7500  Software  v2.0 incorporates  your  favorite  StepOne  Software  features,  such  as  a variety of plate setup wizards, standard curve dilution and master mix recipe calculators, QC flags, data filters, and email notification when a run  is  finished.  The  7500  Software  v2.0  also  includes  an  enhanced Gene  Expression  Study  package  and  has  a  variety  of  new  melting curve  protocol  options,  including  multiple  peak  detection,  step  and hold  temperature  control,  and  customizable  ramp  rates.  The  NEW Gene  Expression  Study  package  accommodates  large  studies  better than any other instrument software package: 

a. Import  an  unlimited  number  of  Comparative  CT  (relative quantitation) files to one study.  

b. Group samples and view data both by technical replicate group and biological replicate group. 

 c. Use any gene(s) as an endogenous control, including averaging 

multiple controls together.  

d. Enter  known  efficiency  values  to  be  factored  into  the  RQ results. 

 2.0 Reason for Issue 

Maintain a document that describes the Standard Operating Procedures that allows for the standard safe and maximal use of the Applied BioSystems, Real Time PCR System within the Pharmacology and Toxicology Core Facilities.   

3.0 Process Description Allow  Core  Facility  Users  within  the  Pharmacology  and  Toxicology Department  to  properly  and  effectively  use  the  Applied  BioSystems,  Real‐Time  PCR  System.  The  process  description  details  the  standard  use  of  the Applied  BioSystems,  7500  Real‐Time  PCR  System.  The  controlled  standard must maintain and adhere to proper and approved research and regulatory qualitative conditions.   3.1 SOP: 18654.B4446B.001 for Applied BioSystems, 7500 Real Time PCR 

System,  authored  by  Brian  Jespersen,  created  on  8/20/2008,  issued on 8/25/2008.   

3.2 SOP:  18654.B446B.001  applies  to  any  User  and  /  or  Trainer  of  the Applied BioSystems, 7500 Real Time PCR System. 

                                                                                                                                                                      

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

 3.3 Responsibilities 

All  Users  are  responsible  for  obtaining  the  proper  approval  and training  before  the  use  of  the  Applied  BioSystems,  7500  Real  Time PCR System. All Users are responsible for the proper use, according to defined protocol, when using the Applied BioSystems, 7500 Real Time   a. New Users need a new Applied BioSystems, Real 7500 Time PCR 

System  account  created  for  equipment  access,  before  initial  use. New  accounts  are  authorized  and  created  by  the  Equipment Champion  and  /  or  the  Core  Facility  Manager.  A  new  account maybe  created  after  training  and  equipment  approval  has occurred. New accounts must  turn off  the default  computer sleep mode.  (Under:  Start/Control  Panel/Display  Settings/Screen Saver/Monitor Power (Power)/Power Schemes)  Select “Always On.”  Click  “Ok”  and  “Ok”  again.  This  process  disables  the  default system “sleep” mode.  If  the computer enters sleep mode, Applied BioSystems, 7500 Real Time PCR System data may be lost.  

b. All  Users  must  schedule  equipment  use  in  the  Equipment Schedule Logbook prior to use. The Equipment Schedule Logbook is  located on the desk next  to  the Applied BioSystems, 7500 Real Time PCR System.  

c. All Users must record all equipment use in the Equipment Usage Logbook  post‐use  on  the  same  day  as  the  recorded  use.  The Logbook  is  located  on  the  desk  next  to  the  Applied  BioSystems, 7500 Real Time PCR System. Within  the Logbook on  the  current log  sheet,  Users  must  record  the  following:  Date,  PI,  Name,  Lab Location,  Phone  Number,  Program,  Samples,  Cycles,  Error Messages.  

 d. Only covered samples may enter Room B‐446B. 

 e. Immediately  after  use,  the  Applied  BioSystems,  7500  Real  Time 

PCR System must be turned off to preserve lamp life.  

f. Do not close the door to room B‐446B.  

g. When not occupying room B‐446B room light off should be off.   

3.4 Equipment Safety Issues  

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

a. Contamination  of  the  thermocycler  block  can  result  in  high background.  To  avoid  contamination  follow  the  below precautions:  

i. Do not open tubes in room. ii. Do not write on tubes – ink will transfer to block. iii. Do not setup reactions with tubes in ice. iv. Do not handle tubes or caps with bare hands. v. Tubes  should  only  have  contact  with  powder‐free 

gloves and clean tube rack  

b. Instrument Issues: Plates and Tubes: Only ABI optical plates, or tubes  and  caps maybe  used.  Caps must  be  the  flat  style.  Domed caps will  cause damage  to  the machine. Do not  close  the door  to room B‐446B.  

3.5 Quality Measures   a. Weekly:  Once  a  week  the  Applied  BioSystems,  7500  Real  Time 

PCR System computer must be shut down and rebooted.    

b. Monthly:  Once  a  month  a  Background  Calibration  and  Optical Calibration must  be preformed on  the Applied BioSystems,  7500 Real Time PCR System. Use the Applied Biosystems Instillation and Maintenance Guide (location of guide see section 3.10) to perform these  monthly  quality  assurance  measures.  Once  these calibrations have been completed, the date, time and person who performed  the  calibrations,  must  be  recorded  in  the  Quality Measures Logbook.   

c. Every  Six  Months:  Once  every  six  months  a  Region  of  Interest calibration and a Pure Dye Calibration must be preformed on the Applied BioSystems, 7500 Real Time PCR System. Use the Applied Biosystems  Instillation and Maintenance Guide  (location of guide see  section  3.10)  to  perform  these  monthly  quality  assurance measures. Once these calibrations have been completed, the date, time  and  person  who  preformed  the  calibrations,  must  be recorded in the Quality Measures Logbook.  

 3.6 Procedure; Applied BioSystems, 7500 Real­Time PCR System Use 

 Setup 

 a. Log  into  computer.  Sign  in  under  User  Account  Name  and 

password. User accounts are sectioned by Laboratory Names.  

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

b. Click  on  ABI  7500  Software  desktop  icon  to  start  use.  This software  runs  the  Applied  BioSystems,  7500  Real  Time  PCR System. 

 c. Within the ABI 7500 Software, click on the “New Document” icon 

in the left hand corner of the screen.  

d. Define  the  document  as  either  Absolute  or  Standard  /  Relative Quantization. 

 e. Choose  Plate  –  96  clear well  plate  and  enter  a  plate  name.  Click 

“Next” button.  

f. Choose  detectors,  the  gene  of  interest  being  analyzed.  Select  the detector and add it to the document. This must be done for every well on the plate. Click “Next” button. 

 g. The plate setup screen will appear. Assign each well a label. 

 h. Highlight  the  box  next  to  the  well  display  on  the  screen.  Assign 

individual well task and enter the quantity of DNA in the quantity sample box for all wells. Make sure to properly label each well. ie; Unknown, Control, Standard, etc. 

 i. Turn  the  Instrument ON. This must be done before pressing 

the  “Finish”  button  or  the  computer  cannot  detect  the instrument. The Applied BioSystems, 7500 Real Time PCR System ON button is located on the front of the system. 

j. When all well parameters are defined, click “Finish.” The computer will  detect  the Applied BioSystems, 7500 Real Time PCR System.  Noise  from  the Applied BioSystems, 7500 Real Time PCR System should be heard at this point.  

k. In  the ABI 7500  software  the 96 Well Template  Screen will  now open. Under the “Setup” tab, individual wells / samples can now be labeled.  Multiple  wells  /  samples  can  be  labeled  by  highlighting multiple boxes at once. 

 l. The default analysis settings can be found under the “Instrument” 

tab.  These  settings  may  be  modified  prior  to  use.  The  default settings are: Sample Volume 50 µL, Stage 3, Step 2, 60 °C, 1 minute. An  important default  deviation  is  the  addition of  the dissociation stage.  This  stage  is  necessary  when  testing  primers  for  the  first time;  assay validation  is  achieved. The data  collection  time point can also be changed. 

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

 m. Save the document as a *.spd file in the proper corresponding User 

Laboratory folder.  

Sample Preparation n. Samples  that  require  centrifuging  must  follow  the  steps  under 

Sample  Preparation.  If  centrifuging  is  not  required  proceed  to the Sample Analysis section. 

 o. To centrifuge samples, use the Eppendorf Centrifuge 8504, which 

is  located  in  room  B‐446B  on  the  desk  adjacent  the  Applied BioSystems,  7500  Real  Time  PCR  System.  96  well  plates  and individual  sample  tubes may  be  centrifuged  with  the  Eppendorf Centrifuge 8504. 

 p. To  turn  on  the  Eppendorf  Centrifuge  8504  by  pressing  the  “On” 

button.  

q. Press “Open” to open the Eppendorf Centrifuge 8504 access door. The  Eppendorf  Centrifuge  8504  will  unlock  and  the  top  access door may be lifted open. 

 r. Securely place samples in the proper tray. Two trays are available, 

a 96 well plate tray and an individual sample tube tray. To switch between the  two trays, use  the Allen Wrench,  located next  to  the Eppendorf  Centrifuge  8504,  and  remove  the  center  bolt.  Lift  the tray  out  and  replace with  required  tray.  Retighten  the  bolt  with the Allen Wrench. 

 s. Sample  must  have  a  counterbalance.  The  96  well  plate 

counterbalance  is  located  with  the  96  well  plate  Eppendorf Centrifuge 8504 tray. Individual sample tube counterbalances are the responsibility of  the User. Both samples and counterbalances must be secured and covered.  

 t. Close  the  lid  of  the  Eppendorf  Centrifuge  8504.  Set  preferred 

centrifuge speed and time.   

u. Press  the  “Start”  button  to  activate  the  centrifuge. Wait  until  the Eppendorf  Centrifuge  8504  has  stopped  before  opening  sample access door. Do not  force  the Eppendorf Centrifuge 8504  sample access door open.  

 v. All  Eppendorf  Centrifuge  8504  use  must  be  recorded  in  the 

Eppendorf Centrifuge 8504 Logbook, which  is  located next to the 

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

Eppendorf  Centrifuge  8504.  Users  must  record  the  following  on the  current  Logbook  form:  Date,  PI,  Name,  Lab  Location,  Phone, Rotor, Time, Number, and Problems / Comments.  

 w. Eppendorf Centrifuge 8504 manuals and user guides are located in 

the second, left desk door.  

      Sample Analysis 

x. Open  the  door  to  the  Applied  BioSystems,  7500  Real  Time  PCR System by pressing the circular indent on the front of the machine. 

 y. Insert the well plate in the aligned position. Make sure to use the 

proper  plate;  either  96  well  plate  or  the  8  strip‐tube  plate  for individual  tubes.  Make  sure  the  plate  is  properly  inserted  and secured. 

 z. Close  the Applied BioSystems,  7500 Real  Time PCR  System door 

by pushing on the circular indent located on the front of the open door until the door is securely closed. 

 aa. When the 96 well plate settings are properly defined, click “Start” 

in the ABI 7500 software. Once started, the ABI 7500 Software will estimate  the  total  read  time.  Once  the  proper  temperature  is reached the Applied BioSystems, 7500 Real Time PCR System will start. 

 bb. The ABI 7500 Software, under the “Methods” Section, will in real‐

time  monitor  the  block,  heated  cover,  sample  temperature, repetition  and  display  the  current  cycle  location.  At  this  point previous data may be manipulated while the system performs the selected task. 

 cc. When finished with the Applied BioSystems, 7500 Real Time PCR 

System, close all programs and completely log off the computer.  

dd. Make sure Applied BioSystems, 7500 Real Time PCR System is off. 

 3.7 Core Materials  

Currently,  there  are  no  Core  Materials  provided  for  the  Applied BioSystems, 7500 Real Time PCR System.  

3.8 User Materials Applied  BioSystems,  7500  Real  Time  PCR  System  User  Materials maybe  located  and  purchased  from  the  below  web  addresses.  The 

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

available  materials  include,  but  are  not  limited  to:  Assays,  Buffers, Enzymes,  Master  Mixes,  Primers,  PCR  Controls,  Reagents,  Services, Thermal Cyclers, Trays, Tubes, Etc. These materials will allow for the proper  and  optimal  use  of  the  Applied  BioSystems,  7500  Real  Time PCR System. Users are  individually responsible  for  the compatibility, maintenance and purchase of these materials.   Related Products https://products.appliedbiosystems.com:443/ab/en/US/adirect/ab?cmd=catNavigate2&catID=601251&tab=RelatedProd.  PCR / RT­PCR Products https://products.appliedbiosystems.com/ab/en/US/adirect/ab;jsessionid=GXtCLySTsPdBh1mTX8cCprJybWVbZvv92GHFcGMp0LQ4zPz0f5fG!1185967524?cmd=catNavigate2&catID=600961.  

3.9 User Use Records  a. Records  of  Use  ­  All  Applied  BioSystems,  7500  Real  Time  PCR 

System use must be recorded. Refer to 3.3b, 3.3c.   

b. Error Messages / System Issues ­ All error messages and system issues  must  be  relayed  to  the  Equipment  Champion  and  the Pharmacology  &  Toxicology  Core  Facility  Manager.  This information must  be  relayed on  the  same day  as  equipment use. Error messages / system issues must be recorded, refer to 3.3c. 

 3.10 Resource Index 

Applied  BioSystems,  7500  Real  Time  PCR  System  literature  and resources for the follow items can be found at the link below. Printed versions of these resources can also be found in the top left desk door adjacent the Applied BioSystems, 7500 Real Time PCR System in room B‐446B. 

 a. Brochure. b. Installation and Maintenance Guide c. Starting Guide: Absolute Quantitation Using Standard Curve. d. Starting Guide: Allelic Discrimination Experiments e. Starting Guide: Plus / Minus Assay Experiments f. Starting Guide: Relative Quantitation Using Comparative CT g. Computer Setup h. Specification Sheet i. User Bulletins j. Data Comparison Guide k. Etc.  

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

 https://products.appliedbiosystems.com:443/ab/en/US/adirect/ab?cmd=catNavigate2&catID=601251&tab=Literature  Applied Biosystems Company Representatives  Field Service Engineer Devin Millions  [email protected] (800) 831‐6844  Customer Support / Sr. Field Application Specialist Chris Miller [email protected] (800) 248‐0281  x7611  

4.0 Competences, Authorization and Training New  Users  must  receive  proper  authorization  from  either  the  Equipment Champion and / or Pharmacology & Toxicology Core Facility Manager before equipment  use.    A  new  User  may  contact  the  Equipment  Champion  or Pharmacology  &  Toxicology  Core  Facility  Manager  to  schedule  training. Training  includes  SOP  and  instrument  familiarization  and  any  additional required or specialized training. Once training is complete authorization may be  issued  and  a  system account  and password may be  setup. All Users  are individually responsible for current SOP familiarization. All New Users must refer  to  3.3a  during  new Applied BioSystems,  7500 Real‐Time  PCR  System account creation.  

 5.0 SOP Performance and Equipment Review 

The  effectiveness  of  the  SOP:  18654.B4446B.001 will  be monitored  by  the Pharmacology & Toxicology Core Facility Manager, Equipment Champion and All Users. Any procedural or qualitative deviations will be reflected within an updated  SOP.  Any  Approved  User  should  aptly  report  any  procedural  or qualitative  issues  and  /  or  errors  to  the  Pharmacology  &  Toxicology  Core Facility Manager or Equipment Champion. The Equipment Champion’s name and  contact  information  can  be  found  on  the  Core  Facility  Identification Sticker.  The  Core  Facility  Identification  Sticker  is  located  on  the  Applied BioSystems,  7500  Real‐Time  PCR  System.  The  Core  Facility  Manager’s contact information maybe found on the Core Facility Contact Posting within room  B‐446.  Updated  SOPs  will  be  published  and  Approved  Users  will  be notified. SOP: 18654.B4446B.001 review will occur every two years.   5.1 Update Date:     

Reason for SOP Change:     New Version #: 

Pharmacology & Toxicology Core Facilities    SOP #: 18654.B446B.001     Version #:  

Author: Brian Jespersen  Page      of  10  Approval Date:      

10 

 5.2 Update Date 

Reason for SOP Change New Version #: 

 6.0 Definitions 

 a.  SOP  Standard  Operating  Procedure,  which  is  a  standard  guide  that officially standardize  the process of control, maintenance and ownership of the Applied BioSystems, 7500 Real‐Time PCR System. The SOP number stand for (xxx . xxx . xxx) equipment number . room number . SOP version number.  b. Originator / Author  The  individual  representing  the Pharmacology and Toxicology Core Facility that created SOP: 18654.B446B.001.  c. Stakeholder Any individual that uses the Applied BioSystems, 7500 Real‐Time PCR System,  including the Pharmacology and Toxicology Core Facility Department.   d.  New  User  An  individual  who  has  not  completed  the  requirements  of section 4.0.  e.  Approved  User  An  individual  who  uses  the  Applied  BioSystems,  7500 Real‐Time  PCR  System  and  has  fulfilled  section  4.0.  This  title may  only  be given  by  the  Equipment  Champion  and  /  or  the  Pharmacology  and Toxicology Core Facility Manager.   f.  Champion  An  individual  who’s  direct  expertise  with  the  Applied BioSystems,  7500  Real‐Time  PCR  System  has  been  recognized  by  the Pharmacology and Toxicology Core Facility Committee. This title may only be awarded by the Pharmacology and Toxicology Core Facility Committee.   

7.0 Appendix  This SOP was written by:       Brian Jespersen ___________________________________ 

 This SOP was reviewed by:      Dr. Stephanie Watts ______________________________  This SOP was authorized by:    Dr. J.R. Haywood __________________________________    Issue Date: _____________________________________