Cordis EMPIRA NC - Cardinal Health...Inches EMPIRA NC Hiryu DURA STAR Voyager NC RX Quantum Maverick...
Transcript of Cordis EMPIRA NC - Cardinal Health...Inches EMPIRA NC Hiryu DURA STAR Voyager NC RX Quantum Maverick...
-
Deliverability
Durability
Cordis EMPIRA™ NC
Control the Intervention with EMPIRA™ NC
RX PTCA Post-Dilatation Catheter
-
Efficient & EffectiveEMPIRA™ NC Balloon Catheter : Designed to deliver in challenging interventions & post-dilatation procedures
Treat a wide range of complex lesions & help optimize stent deployment with:
• Crossability: Proprietary folding & pleating reduce the primary profile, helping you reach even hard-to-treat lesions1
• Recrossability: Innovative hydrophilic coating improves lubricity & durability through multiple crosses
• Flexibility: Next-generation DURALYN™ Flex Balloon Material & kink-resistant shaft improve crossability & trackability
• Best-in-class controlled growth: Perform intervention with the most non-compliant post-dilatation balloon available1
Next-generation DURALYN™ Flex Balloon Material is 50% more flexible than DURALYN™1 Balloon Material
Ease of use
• Red color-coded strain relief for quick product identification
• Clear hub for easy spotting of bubbles or fluid in the lumen
Cordis EMPIRA™ NCRX PTCA Post-Dilatation Catheter
-
Inch
es
EMPIRA™ NC DURA STAR™Hiryu™ Voyager™ NC RX
Quantum™ Maverick™ Monorail™
Mill
imet
ers
0.038
0.037
0.036
0.035
0.034
0.033
0.032
0.9652
0.9398
0.9144
0.889
0.8636
0.8382
0.8128
*All balloons reach P
-
Outer Diameter (OD) growth percentage
• Best-in-class controlled growth – 3.09% between nominal and rated burst pressures (RBP)1‡
Nominal and Rated Burst Pressures2
• HIgh rated burst pressure – 20 atm for all sizes
• Mean Burst Pressure of 28.14 atm‡‡1
EMPIRA™ NC Balloon Catheter: Best-in-class controlled growthPo
urc
enta
ge
DURA STAR™ Voyager™ NC RX
9
6
3
0
bPourcentage d’augmentation = DE à la Pression d’éclatement (RBP) - DE à la pression nominale / DE à la pression nominale x 100.
aTous les ballonnets atteignent P < 0,05 vs le ballonnet EMPIRA™ NC.
Hiryu™Quantum™Maverick™Monorail
EMPIRA NC™
EMPIRA™ NC
9 10 11 1210 13 14 15 16 17 18 19 20
NC Trek
Voyager™ NC
NC Quantum Apex™
Quantum® Maverick™
atm
Hiryu™
NC Sprinter™
Co
ntr
olle
d G
row
th
‡Measured for a 3.00 x 20 EMPIRA™ NC RX PTCA Post-Dilatation Catheter‡‡Tested on 3.00 x 20 EMPIRA™ NC RX PTCA Post-Dilatation Catheter
-
EMPIRA™ NC Balloon Catheter: Best-in-class controlled growth
EMPIRA™ NC: More Non-Compliant than other PTCA Balloons2
• Limited dog boning ensures uniform pressure through entire length of the balloon for optimal stent deployment1
* 3.0 x 20 mm Data from product labels for 3.0 x 20 mm balloon sizes. RBP is Rated Burst Pressure. "-" indicates data not listed on data sources.
High pressures combined with low longitudinal growth
EMPIRA™ NC 3.00 x 20
14 atm
RBP: 20 atm
-
EMPIRA™ NC Balloon Catheter delivers crossability & best-in-class controlled growth to a wide range of lesions
EMPIRA™ NC RX PTCA Post-Dilatation Catheter - Ordering CodesBalloon
Diameter (mm)
Balloon Length (mm)
2.00
2.25
2.50
2.75
3.00
3.25
3.50
3.75
4.00
6 10 12 15 20 25 30
75R06200NP 75R10200NP 75R12200NP 75R15200NP 75R20200NP
75R10225NP 75R12225NP 75R15225NP 75R20225NP
75R06250NP 75R10250NP 75R12250NP 75R15250NP 75R20250NP 75R25250NP 75R30250NP
75R10275NP 75R12275NP 75R15275NP 75R20275NP
75R06300NP 75R10300NP 75R12300NP 75R15300NP 75R20300NP 75R25300NP 75R30300NP
75R10325NP 75R12325NP 75R15325NP 75R20325NP
75R06350NP 75R10350NP 75R12350NP 75R15350NP 75R20350NP 75R25350NP 75R30350NP
75R10375NP 75R12375NP 75R15375NP 75R20375NP
75R10400NP 75R12400NP 75R15400NP 75R20400NP
EMPIRA™ NC RX PTCA Post-Dilatation Catheter - Technical Information
Primary crossing profile* 0.0344"
Strain relief color Red
Balloon material DURALYN™ Flex Balloon Material, non-compliant
Shaft diameters Proximal 1.9F / distal 2.7F
Shaft length 139 cm
Minimum guiding catheter 5F
Maximum guidewire .014"
Nominal/Rated burst pressure (RBP) 14/20 atm
¼ size pressure 3.0 mm balloon at 26 atm pressure reaches 3.261 mm in diameter
Kissing balloon technique (KBT) compatible with Cordis 6F Guiding Catheters1
*Measured across the distal markerband of a 3.00 x 20 EMPIRA™ NC RX PTCA Post-Dilatation Catheter
PRODUCT CODE LEGEND7 5 R 0 6 2 0 0 NP
Product
Rapid Exchange
Length
DiameterNon-Compliant
Legal Manufacturer of Empira:Creganna-Tactx Medical1353 Dell AvenueCampbell, California 95008, USA
References: 1. Cordis data on file, 2010 and 2011 test report. 2. All data from manufacturer's product instructions for use for 3.0 x 20 mm balloon sizes.
Important Information:Prior to use, refer to the instructions for use supplied with this device for indications, contraindications, side effects, suggested procedure, warnings and precautions. As part of its continuous product development policy, Cordis reserves the right to change product specifications without prior notification. For healthcare professionals only. © Cardinal Health 2018. All Rights Reserved. CORDIS, Cordis LOGO and EMPIRA are trademarks of Cardinal Health and may be registered in the US and/or in other countries. All other marks are the property of their respective owners. Please contact your Cordis representative for additional product availability information. 09/18 - 100524171
Cordis EMPIRA™ NCRX PTCA Post-Dilatation Catheter