CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1 INSTRUCTION MANUAL CoolSafe 110 Freeze Dryer

Transcript of CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe...

Page 1: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

INSTRUCTION MANUAL

CoolSafe 110 Freeze Dryer

SV
Image importée
Page 2: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

Important user information Please read this entire manual to fully understand the safe and effective use of this product. In case you have any comments about this manual we will appreciate receiving them at the address below. Warranty and Liability ScanVac A/S guaranties that the product delivered has been thoroughly tested to ensure that it meets its published specifications. The warranty included in the conditions of delivery is valid only if the product has been installed and used in accordance with the instructions supplied by ScanVac A/S. ScanLaf A/S shall in no event be liable for incidental or consequential damages, including without limitation, lost profits, loss of income, loss of business opportunities, loss of use, and other related exposures, caused by e.g. incorrect use of the product. Symbols used in this manual

WARNING Used in case of danger of a serious accident or when documentation needs to be consulted.

NOTE Used to direct attention to a special item.

Page 3: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 3/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

Table of contents 1. Introduction --------------------------------------------------------------------- 4 2. Safety precautions--------------------------------------------------------------- 4 2.1. General information------------------------------------------------------------ 4 2.2. Safety----------------------------------------------------------------------------- 4 3. Installation ----------------------------------------------------------------------- 4 3.1. Installation of CoolSafe 110 -------------------------------------------------- 4 3.2. Installation of vacuum pump-------------------------------------------------- 5 3.3. Installation of accessories ----------------------------------------------------- 5 4. Functional description --------------------------------------------------------- 5 4.1. Mains----------------------------------------------------------------------------- 5 4.2. Control panel -------------------------------------------------------------------- 5 4.3. Valves and connections-------------------------------------------------------- 6 5. Operation ------------------------------------------------------------------------- 6 5.1. Overview ------------------------------------------------------------------------ 7 5.2. Preparations --------------------------------------------------------------------- 7 5.3. Freeze-drying using manifolds ----------------------------------------------- 8 5.4. Freeze-drying using chambers ------------------------------------------------ 9 5.5. De-ice ---------------------------------------------------------------------------- 10 5.6. Stopping the vacuum pump --------------------------------------------------- 10 6. Alarms and errors--------------------------------------------------------------- 10 7. Maintenance---------------------------------------------------------------------- 11 7.1. CoolSafe 110-------------------------------------------------------------------- 11 7.2. Vacuum pump------------------------------------------------------------------- 11 7.3. Cleaning ------------------------------------------------------------------------- 11 8. The Freeze-drying process----------------------------------------------------- 11 9. Technical Specifications-------------------------------------------------------- 13 9.1. General--------------------------------------------------------------------------- 13 10. Accessories ---------------------------------------------------------------------- 13

Enclosure: Declaration of conformity

Page 4: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 4/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

1. Introduction The CoolSafe 110 is a small, compact and easy to use Freeze Dryer. Using the wide range of accessories including a number of chambers and manifolds, the CoolSafe 110 covers a large number of different applications. Freeze-drying is often used for sample preparations and for preservation and storage of biological and pharmaceuticals material. Freeze-drying is suited for substances that are heat sensitive. If you are not familiar with the freeze-drying process, please read Section 7, which introduces the freeze-drying process. 2. Safety precautions 2.1. General information Please read this entire manual to fully understand the safe and effective use of this product. Should you have any comments about this manual we will be pleased to receive them.

2.2. Safety In case of overload, an automatic safety device stops the freeze dryer. The CoolSafe 110 is electrically protected with an over current circuit beaker. The safety system is designed according to Safety requirements for electric equipment, control and laboratory use (EN 61010-1), which assumes that two separate faults do not occur simultaneously.

NOTE ScanVac A/S recommends using safety glasses for personal protection when connecting flasks or other equipment to the freeze-dryer – due to the risk of implosion.

3. Installation 3.1. Installation of CoolSafe 110

• Make sure that the mains supply is according to the specifications on the type label at the rear panel. • It is important that free air circulation is insured at the air in- and outlets. If the CoolSafe 110 is

positioned against a wall, there must be at least 100 mm free space at the rear of the unit. • The unit is intended to be located at a tabletop or a trolley.

Page 5: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 5/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

3.2. Installation of vacuum pump A vacuum pump (two-stages with 2-5 m3/hour capacity) must be connected to the CoolSafe 110 in order to allow cryo pumping. ScanVac recommends Ilm Vac2 rotary vane pump for which a variety of protection equipment against aggressive materials is available. If the pump outlet is not connected to e.g. a fume hood, an oil mist filter should always be connected. See accessories section for recommended rotary vane pump and oil mist filter.

• Check that the mains supply for the pump is according to the specifications on the type label. • Connect the vacuum hose to the suction side of the pump. • Connect the pump to the vacuum connection pipe on the acrylic plate of the CoolSafe. • Check the oil level on the pump and fill it if necessary. The oil level sight glass should be 1/2 - 3/4

full. • ScanVac A/S recommends that the pump is running with the gas ballast valve open at all times.

3.3. Installation of accessories The CoolSafe 110 is a basic unit and must be equipped with accessories in order to run a specific freeze-drying process. Select from the range of accessories listed in the accessories section. Make sure to mount the proper manifold, chamber and shelf arrangements and always check that all connections are tight. 4. Functional description For reference, see the overview in section 6. 4.1. Mains The mains switch and mains inlet is located at the rear of the unit. 4.2. Control panel CoolSafe 110 has the following indicators:

Display A four-digit display shows condenser temperature in the range from ambient temperature down below 110 ºC.

AWO-indicator Display

Page 6: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 6/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

Alarm-Wait-Ok (AWO) indicator A multicolor LED shows one of the following states State LED Color Description Alarm* Red

Flashing Condenser temperature rising above –20 °C or sensor malfunction. See Alarms and Error section.

Wait Yellow Flashing Condenser temperature between –20 °C and –70 °C.

Ok Green Condenser temperature below –70 °C / -90 °C initially. *Alarm state is both visual and acoustic. 4.3. Valves and connections

• CoolSafe must be used with a top lid of transparent acrylics. • The vacuum connection pipe for the vacuum pump connection is located on the acrylic lid of the

CoolSafe. • Choose to use manifolds or chambers from accessory section 9. • The drain valve is located at the front, and is used for draining the ice condenser after de-icing.

5. Operation The freeze drying process is a balance between the heat absorbed by the sample to vaporize the ice, and the heat removed by the condenser to convert the water vapor into ice. It is important to remember following factors when trying to obtain efficient utilization of the freeze dryer system. The efficiency of the freeze drying process depends upon:

• The surface area and the thickness of the sample. • The temperature of the condenser and the vacuum obtained. • The eutectic point (melting temperature) and the concentration of the sample solution.

Page 7: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 7/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

5.1. Overview

Rear view:

5.2. Preparations Check that the CoolSafe ice condenser is free from ice and water, and that all valves i.e. the drain valve is closed.. Check that their is no water in the drain valve.

• Check that the gas ballast valve on the vacuum pump is open at all times. Check the oil level in the pump and the oil mist filter. The filter must be operated with maximum 10-mm oil level (preferably empty).

• Switch on the CoolSafe and the vacuum pump. The CoolSafe will enter the WAIT-STATE indicated

Control panel

Drain valve

Air inlet

Page 8: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 8/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

by a flashing yellow LED, and start the cooling process. Just after switch on, the display shows the actual software version and then all display segments are switched on for test of display. After a few seconds, the temperature measurement is stabilized and the actual condenser temperature will be displayed.

• The vacuum pump should now be allowed to run for at least 30 minutes before opening the (external) vacuum valve. This period of time also allows the condenser to reach operational temperatures (below –90 ºC initially) indicated by a green LED, ready for freeze-drying.

• The product has to be in frozen state using auxiliary equipment, freezer, liquid nitrogen etc or use the condenser as a freezer. To avoid product melting within the time it takes to move the product to the freeze dryer and establish a pressure below the triple point, the temperature of the product should be brought down to as low a temperature as possible.

5.3. Freeze-drying using manifolds

1. Check that all rubber valves are closed (A) or in vacuum-release position (C), and that the flange and lid are tight.

2. Start the creation of vacuum by opening the vacuum valve and check that the AWO-Indicator stays

green indicating OK for freeze-drying. 3. Connect a flask to an appropriate cone on the rubber valve. Turn the rubber valve to position (B).

The AWO-Indicator should always be green prior to connecting the next flask. The freeze-drying process is now in progress.

Page 9: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 9/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

4. To stop the freeze-drying process, turn the rubber valve handle 180º to position (C) to equalize the

pressure in the flask, and remove the flask from the cone.

5. Close the vacuum valve after removal of the last flask.

6. To allow the system to equalize the vacuum in condenser, use the rubber valve.

5.4. Freeze-drying using chambers

1. Make sure that the (external) vacuum valve is closed. 2. Mount the chamber and shelf arrangement on the CoolSafe. Check that all connections are tight and

the chamber is closed. If heated shelves are to be used, refer to the controller manual. 3. Open the chamber; place the frozen material on the shelves in the chamber. Close the chamber and

the drain valve.

Page 10: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 10/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

4. Open the (external) vacuum valve. The freeze-drying process is now in progress. 5. To stop the freeze-drying process, close the (external) vacuum valve and equalize the pressure by

opening the release valve at the lid. 6. Remove the material from the shelves.

5.5. De-ice The freeze dryer may be de-iced by:

• Leaving the ice condenser open at room temperature. • Filling the ice condenser with very hot water leaving the drain valve closed until the ice has melted.

NOTE Do not fill the ice condenser with too much water

5.6. Stopping the vacuum pump

Close the (external) vacuum valve and let the pump run for a minimum of 45 minutes after the freeze-drying has ended. This is to avoid any possible condenses remaining in the pump oil. If the freeze dryer is used regularly during the week, it is recommended that the vacuum pump run constantly with open gas ballast valve and closed vacuum valve. 6. Alarms and errors In case of alarms or errors, an alarm/error code will be displayed. E_1: Error indicating that the temperature sensor is short circuit. E_2: Alarm indicating that the temperature during freeze-drying is above 20 C. E_3: Error indicating that the temperature sensor is disconnected.

Page 11: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 11/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

7. Maintenance 7.1. CoolSafe 110 The air-cooled condenser must be cleaned minimum twice a year. Use a vacuum cleaner and remove any dust in the cooling ribs on the front. Rubber valves, black O-rings and valves on the manifold and black O-rings on chambers should be greased with a ultra thin layer of vacuum grease and wiped off with a soft cloth every 3-4 months. The rubber must appear shiny, smooth and without cracks. All silicone gaskets needs only dust removal (do not use any vacuum grease). 7.2. Vacuum pump See the vacuum pump instruction manual for maintenance according to supplier’s recommendation. Maintenance for a vacuum pump in general will be to:

• Check the oil volume and color conditions regularly. The oil should be changed every 400-500 working hours, more often if the pump has been exposed to aggressive vapors. Keep a Log book

• Drain the oil mist filter if the oil level is more than 10 mm. • Drain, clean and eventually change the filter insert if smoke comes from the oil mist filter.

7.3. Cleaning

If a sample is spilled in the concentrator chamber, it must be cleaned up immediately. 8. The Freeze-drying process To understand the freeze-drying process, this section describes the theory for the process in short. For more detailed information about freeze-drying please consult available literature e.g. “ An introduction to freeze drying by James M. Flink & Henning Knudsen, ISBN 87-872-0084-8. All material in free form will exist in one of three states: Vapor, liquid or solid. As may be seen in the figure where water is given as an example, the actual material state is dependent on the surrounding pressure and the temperature. The point where all three Liquid phases co-exist is called the triple point. Freeze-drying is basically a change in state space properties of the material that is to be removed. In principle the process is very simple, but in reality it is very complex as the product handled consists of many components, which will interact.

Page 12: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 12/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

Moving around in the state space of a certain material requires supply or removal of energy to or from the product. This energy flux will have an impact on the structure of the material. Freeze-drying is achieved using equipment that controls temperature and pressure in the surroundings of the product and is a 3-step process.

• In the first step, the product is frozen solid so that the water present in the material is converted to ice (pre-freeze).

• In the second step, the ice formed in pre-freezing is removed from the product by direct conversion from solid to vapor. This process is called sublimation. In order to start removing moisture in the sublimation process, the pressure surrounding the material has to be brought down to a value below the triple point value while still keeping the temperature of the product below the freezing point. The condenser surface temperature must also be brought down to a temperature inferior to the temperature of the material handled. When the product is heated it will place itself somewhere on the transition line between the solid and vapor states. The heating will initiate sublimation of moisture from the product to the surrounding environment. The vapor will reach the condenser surface and condense because of the lower vapor pressure here.

• In the third step, water that was strongly bound to the solids of the sample (called sorbed water) is converted to vapor and removed from the product, this process being referred to as desorption. Removing moisture in desorption phase or secondary drying phase is done at highest product temperatures and lowest possible pressure.

Page 13: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

INSTRUCTION MANUAL CoolSafe 110 Page 13/ 13

CoolSafe 110-manual-UK.doc – 17. June 2006 Revision 1

9. Technical Specifications 9.1. General

Specifications CoolSafe 110 Type CoolSafe 110

Ultimate temperature: -110 °C at 20 °C room temperature Condenser dimension, mm: 162 × 180 Total volume, L: 4 Cabinet dimensions H×W×D, mm: 530 × 400 × 500 Insulation, cm: 9 Condenser capacity per 24 hours, kg: 2 Condenser capacity/total, kg: 3 Cooling media: R507/R1150 Power, V/Hz: 230/50, 115/60 Power consumption, W: 600 Materials: Cabinet polyester coated steel,

Condenser stainless steel AISI 316 Weight, kg: 55 Drain tap: Yes Digital temperature read-out: Yes Microprocessor control with alarms: Yes Start delay for the compressors: Yes Ambient temperature 5 – 32 ºC

10. Accessories See attached catalogue

Page 14: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

SCANLAF A/S

Declaration of conformity

DS/EN 45014 Annex A (recommended form for declaration of conformity

We: ScanLaf A/S

Nøglegårdsvej 20, Vassingerød, DK- 3540 Lynge, Denmark

declare under our sole responsibility that the product

Model: ScanLaf CoolSafe110

to which this declaration relates is in conformity with the following standard(s) or other normative document(s):

EN 292-1:1993 Safety of machinery (Basic concepts – General principles)

EN 292-2:1993 Safety of machinery (Technical principles and specifications)

EN 60204-1:1999 Safety of machinery (Electrical equipment of machines – General requirements)

EN 60000-6-3:2001 and EN 60000-6-1:2001 Electromagnetic compatibility. (Generic emission / immunity standard – Residential, commercial and light industry)

EN 60000-6-4:2001 and EN 60000-6-2:2001 Electromagnetic compatibility. (Generic standards - Emission / immunity for industrial environment)

EN 1050:1996 – Safety of machinery. (Principles for risk assessments)

(if applicable) following provisions of:

Directive 89/392/EEC Machinery

Directive 91/368/ Machinery (1. amendment)

Directive 93/44/EEC Machinery (2. amendment)

Directive 73/23/EEC Low voltage

Vassingerød, 10. May 2006 Finn Kokholm, President

Page 15: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

CoolSafe™ facilitates versatile large, fast and easy freeze drying of even large Samples.

Chamber CCS300

Large strong acrylic freeze drying chamber for up to 10 trays with individuel height adjustment possibility. Allows individual prefreeze in the trays.With 10 trays Ø 250 mm, product area will be 0,50 m². With maximum height of 350 mm with one tray or 35 mm with 10 trays.Each tray has a dimension of Ø 250 × 20 mm and is of high grade 316 L stainless steel.Chamber dimensions H×Ø: 350 ×300 mm.Materials: Stainless steel AISI 316 rack, shelf and tray. Chamber in Acrylic, gasket for acrylic plate.

Chamber CCS300/4 Manifold

Same as CCS300 but with extra 4 ¾ inches rubber valves on top. For attachments of other freeze drying accessories.Chamber dimensions H×Ø: 350×300 mm.Materials: Stainless steel AISI 316 rack, shelf and tray. Chamber in Acrylic, gasket for acry-lic plate. 4 rubber valves ¾ inches.

Cat. No. Code Description

7.001.000.085 CCS300 AC Chamber with rack, 3 stainless steel shelves 3 trays and gasket

7.001.000.185 Shelf300 Extra shelf for fi xation of extra product tray

7.001.000.285 Tray300 Extra product tray Ø 250 × 20 mm

Cat. No. Code Description

7.001.000.085 CCS300 4V AC Chamber with rack, 3 stainless steel shelves, gasket and 4 rubber valves

7.001.000.185 Shelf300 Extra shelf for fi xation of extra product tray

7.001.000.285 Tray300 Extra product tray Ø 250 x 20 mm

Page 16: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

Chamber CCS200

Compact strong acrylic chamber system takes up to 6 trays with individual height adjustment possiblity. Allows individual prefreeze of the trays or the whole rack in a freezer.With 6 trays Ø 180 mm product area will be 0,16 m² .With maximum height 210 mm with one tray or 30 mm with 6 trays.Each tray has a dimension of Ø 180 × 10 mm and is of high grade 316 L Stainless steel.Chamber dimensions H×Ø: 210 × 200 mm.Materials: Stainless steel AISI 316 rack, shelf and tray. Chamber in acrylic. Rubber gasket for acrylic plate.

4Tree Manifold

Strong easy accessible manifold tree with ¾ inches rubber valves. The system can be extended with extra 4Tree manifold to the height and number of valves wanted. The acrylic lid acts as a drip trip tray. The 4Tree Manifold can also be used for connection of up to 4 selfstanding cham-bers (ACex) or vacuum concentrators among others.

Cat. No. Code Description

7.001.000.064 CCS200 AC Chamber with rack, 2 shelves and 2 trays and gasket

7.001.000.164 Shelf200 Extra shelf for fi xation of extra product tray

7.001.000.264 Tray200 Extra product tray Ø 180 x 10 mm

Cat. No. Code Description

7.001.000.062 M4Basic 4 × ¾ inches Manifold, 40 cm heigh

7.001.000.063 M4Ext 4× ¾ inches Manifold, 20 cm heigh for extension of M4Basic

Page 17: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

Cones for Manifold drying

Cones in aluminium for connection of freezdrying fl ask to manifold tree with ¾ inch rubber valves. Includes 2 silicone tightening rings and one silicone rubber for safe handling of the fl asks.

Acrylic plate for all chambers and manifolds

When choosing a chamber also choose the correct adaptation plate to ensure correct fi tting of the chamber, vacuum centrifuge and manifold to the CoolSafe. For freeze drying : The strong acrylic plate includes connector to the vacuum pump and vacuum release. Takes all chambers and mani-folds. (The gasket for tight vacuum fi tting follows the chamber).For use as a cooling trap: The strong acrylic lid includes two separate pipes for the vacuum pump and e.g. the vacuum centrifuge.

Support stand

Standard. Provided with rubber feet for easy and stable accesment on standard laboratory tables.Option: Castors for easy operation and / or electrical elevation stand for the highest operator comfort.

Cat. No. Code Description

7.001.000.065 El. Stand

Compact electrical elevation stand with room for CoolSafe, vacuum pump and VacControl. Easy change of working height from 70 cm to 110 cm for best operator comfort.

7.001.000.066 4 Castors4 castors of which 2 are lockable. For direct attach-ment to CoolSafe or to electrical elevation stand

Cat. No. Code Description

7.001.000.072 C29/32Aluminium cones for adapting glass fl asks of 29/32 mm

7.001.000.073 C32/35Aluminium cones for adapting glass fl asks of 32/35 mm

7.001.000.074 C26/29Aluminium cones for adapting glass fl asks of 24/29 mm

Cat. No. Code Description

7.001.000.061 CTpl Cooling trap plate with 2 pipes

7.001.100.061 ACplAcrylic plate. Basic for use on top on CoolSafe. Includes connection for vacuum pump and vacuum release. Takes both M4 Basic, AC 200 and 300 chambers

7.001.200.061 ACexAcrylic plate basic for use beside on CoolSafe includes connection for vacuum pump and vacuum release. Takes both AC 200 and 300 chambers

ACexCTpl ACpl

Page 18: CoolSafe 110 Freeze Dryer - LabMakelaar · CoolSafe 110 Freeze Dryer . INSTRUCTION MANUAL CoolSafe 110 Page 2/ 13 CoolSafe 110-manual-UK.doc – 17. ... full. • ScanVac A/S recommends

DuoCoolSafe

Double CoolSafe with double ice condenser volume of 8 litres allows almost unlimited total ice capacity with max. 5 kg ice per 24 hours, and pre-freezing possibility:◆ One condenser can prefreeze while the other

begin freeze drying. ◆ One condenser can be deiced while the other

condenser still freeze dry.◆ Both condensers freeze dry at the same time.

Other Accessories

Cat. No. Code Description

7.001.000.077 T-kitT-kit including 2 ¾ inch rubber valves and vacuum hose connectors. For an extra CoolSafe connected to the vacuum pump

7.001.000.076 AmpMan 16Ampoule manifold fi tting ¾ inches rubber valves. Takes 16 ampoules. Uses 6 mm silicone tubes.

7.001.000.067 VC Vacuum hose including two clamps

7.001.000.075 PC Vacuum hose connector for connection of ¾ inches rubber valves and vacuum hose

7.001.001.069 VP2 30 m²/min, 1,8 m³/hours, 0,001 mBar

7.001.201.069 VP5 77 m²/min, 4,6 m³/hours, 0,002 mBar

7.001.001.070 Oil mist Oilmist fi lter for VP2 / VP5

7.001.001.071 HN Hose nozzle for VP2 / VP5

7.001.000.168 VacCon For checking leakages and end drying. 1 atm to 1 x 10 m³/mBar

7.001.000.078 VacCon Trans Transportable vacControl for checking leaking

Cat. No. Code Description

2 x CS 55 or CS 110 CoolSafe 55 or CoolSafe 110

2 x Acpl Acrylic plate

2 x CSS300 or/and Man4 Chamber or manifold as preferred

3 x VH Vacuum hose with clamps

1 x T-kit T-Kit connection

1 x VacCon Trans Vacuum controller

1 x VP 5 Vacuum pump

1 x Oilmist Hose nozzle

1 x HN Connection kit

PC

VC

Beside of choosing chamber or manifold, recommended confi guration is

VacControl

AmpMan 16

VacControl

Nøglegårdsvej 20, Vassingerød, DK-3540 LyngePhone +45 3940 2566, Mobile +45 3016 0990, Telefax +45 4498 [email protected], www.scanlaf.dk

T-kitVacCon Trans

Scan

Laf®

, Sca

nVac

™ a

nd V

acSa

fe™ a

re T

rade

mar

ks o

f Sca

nLaf

A/S

, Den

mar

k. S

canL

af® is

reg

iste

red

in E

urop

e.Pr

inte

d in

Den

mar

k by

Ath

ene

Gra

fi sk