Structure RayCAm C.A 1884Structure RayCAm C.A 1884

Power supply interface

USB interface.

Standard USB interface

connect to PC

Video output interface.

Standard RCA interface

Main LCD

Key panel

Definable trigger

Manual focus lens

Cover of battery compartment

Page 3: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Structure RayCAm C.A 1884Structure RayCAm C.A 1884

Instructions on Key OperationsInstructions on Key Operations

INSTRUCTION OF IMAGE DISPLAY STATUS BARINSTRUCTION OF IMAGE DISPLAY STATUS BAR

Power switch

Power LED

Control panel  “A”: Auto Adjust key “C“: Cancel key / correction    “S”: Freeze/live key, saving key      4 direction keys Menu - Enter

«  »: user is in dynamic mode«  »: user is in freeze mode

Operation condition : NULL, MENU, SP1, ISO or CAPT

Emissivity of the target point

Power status

Sound alarm on or off

The color bar in the middle represents pseudocolors

Page 4: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the Menu Items and Functions of the Menu

Sub menu « Analysis »Sub menu « Analysis »

Analysis►FileAuto Adj.Manual Adj.Obj. Par.Image OnlySetup

Spot 1Spot 2Spot 3Max Sp.IsothermRemove all

The camera will come into the temperature measurement condition. Move the camera or move the direction key to point the cursor to the target, the target’s temperature value will be displaying on the right-up of the screen

Maximal or minimal temperature: one spot is tracking the highest or lowest temperature in the area.. The user can change the size of the area.

The camera will do the isotherm analysis of the infrared image. If the temperature is in one setting area, it will be displayed in one color.

Cancel all the analysis

Page 5: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the MenuItems and Functions of the Menu

Sub menu « File »Sub menu « File »

AnalysisFile ►Auto Adj.Manual Adj.Obj. Par.Image OnlySetting

OpenSaveDelete

Open the images of the Flash Disk

Save the image to the Flash Disk

Delete the images of the Flash Disk

Page 6: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the MenuItems and Functions of the Menu

Sub menu «Manual Adj. »Sub menu «Manual Adj. »

Level30°C

Span 10°C

Temp. Range-20 +1500 °C

Filter Off

it represents the median value of the temperature scale

it represents the amplitude of the scale, i.e. the difference between the max. temperature and the min. temperature

it represents the max. temperature range of the RayCAm

Page 7: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the MenuItems and Functions of the Menu

Sub menu « Obj. Par. »Sub menu « Obj. Par. » (Object Parameters) (Object Parameters)

Emiss: 0.95

Distance: 5m

Temp Amb: 25.0°C

Humidity: 50%

Material property. Setting between 0 to 1

Distance between the thermography user and the target *

It seems the temperature around the target

Parameter of the ambiant air *

* You have to set these parameters only if you are > 10m of the target.

Page 8: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the MenuItems and Functions of the Menu

Sub menu « Auto. Sub menu « Auto. Adjustment »Adjustment » the camera will do the auto-adjustment. The key “A” is

the same function as this.

Sub menu « Image only »Sub menu « Image only » the information will be concealed. You need to push on

any key to make reappear all the information.

Page 9: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the MenuItems and Functions of the Menu

Sub menu « Setup »Sub menu « Setup » Analysis Analysis

Capture spot : Maximum

Alert: Activé

Temp alert: 100°C

Correct temp: 0° C

Isotherm width : 0.7° C

Isotherm color : Vert

Isotherm alert : 50° C

Laser Adjust : Activé

Set the option “Max Sp” : auto-track the highest or the lowest T°C

Turn on or turn off the temperature alarm function Preset the alarm temperature value

Correct the value of the measured temperature

Set the width of the isotherm range

Set the color of the isotherm

Sound alarm of the isotherm

Define the position of image of the laser point and measure spot SP1

Page 10: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Items and Functions of the MenuItems and Functions of the Menu

Sub menu « Setup »Sub menu « Setup » Local Local

Sub menu « Setup »Sub menu « Setup » Date and hour Date and hourLanguage : FrenchVideo output : PALTemp unit .: °CDist unit : MeterDate format : DD/MM/YY

Year: 2009Month: 3Day: 5Hour: 7Minute: 15

Page 11: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Show the current file / directory name

Set the control switch of the trigger button

Items and Functions of the MenuItems and Functions of the Menu

Sub menu « Setup »Sub menu « Setup » Other Other

Palette : Métal

Auto Adjust : Lumin. et con.

Continus Adj : Lumin. et con.

Shut down : Jamais

LCD display : Jamais

Shutter period : Courtes

File name :

Directory name :

Trigger Button: Laser on

Palette : It provides a number of artificial color options

Define the adjustment way of the key “A”.

Set whether the camera needs to auto-adjust continuously according to the change of the image

Set the camera that whether it needs to auto-shut down

Set the screen that whether it needs to auto-shut down

Set the period of auto-adjustment

Page 12: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Characteristics of the RayCAm C.A 1884Characteristics of the RayCAm C.A 1884

Detector characteristicsDetector characteristics Size of the matrix: 120 x 160 Detection frequency: 50 Hz Thermic sensibility: 0,1°C Temperature range: -20°C to 250 / 600 / 1000 / 1500°C Modelization thanks to the settings of all the parameters

which affect the measurement Emissivity, ambiant temperature, relative humidity,

distance between the user and the target

Page 13: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

RayCAm C.A 1884 characteristicsRayCAm C.A 1884 characteristics

Optic characteristicsOptic characteristics Field of view: 20° x 15° IFOV: 2,2 mrad(A detector measure an object of 2,2 mm at a distance of 1m)

HFOV VFOV PPOM 3*IFOV

 IFOV

Distance minimale

focalisation

Imaging factor

0.1m

0.3m

0.5m

1.0m

2.0m 6.0m 10m 30m 100m

20°×15°

 2.2 mrad 10 cm

HFOV (m) 0.03 0.10 0.2 0.35 0.70 2.11 3.52 10.57 35.26

VFOV (m) 0.02 0.07 0.13

0.26 0.52 1.57 2.63 7.89 26.33

IFOV (mm) 0.22 0.66 1.1 2.20 4.40 13.2 22.04 66.12 220.40

PPOM 0.66 1.98 3.3 6.60 13.2 39.7 66.12 198.4 661.20

Page 14: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

RayCAm C.A 1884 presentationRayCAm C.A 1884 presentation

Analysis toolsAnalysis tools

Automatic recognition T°C max … in a adjustable area

3 manual cursors: creation of an isotherm width

And a sound alert to indicate a value have been overrun

P01:46,1°C

P02:41,5 °C

P03:37,5 °C

Page 15: Confidentiel –Usage interne Presentation RayCAm CA 1884.

Confidentiel –Usage interne

Other specifications of the C.A 1884Other specifications of the C.A 1884

Picture capacityPicture capacity 1000 radiometric pictures could be saved Creation of 250 folders

Physical caracteristics :Physical caracteristics : Weight: < 700 g Chocs resistivity: 25 G Vibration resistivity: 2 G IP 54

Environmentals requirementsEnvironmentals requirements Operating temperature : -20° to 50°C Relative Humidity: 10 to 95 %HR Storage temperature : -40°C to 70°C