High Quality Groundwater and Surface Water Monitoring Instrumentation
Solinst is a registered trademark of Solinst Canada Ltd.
3001 Levelogger Series: Hardware CompatibilityRecommended Software & Firmware Congurations
PC Software FirmwareLeveloader Gold
FirmwareSDI-12
Firmware
3001 Levelogger Edge
4.1
3.003
2.001
Note: Use a fully charged 9V
battery in your Leveloader
during upgrade. A depleted
battery may cause a
communication interruption.
2.000
3001 Barologger Edge
3001 Junior Edge*
3001 LTC Junior2.005
3002 Rainlogger Edge3.000
Levelogger PC Software Versions 4.1 and up arecompatible with Windows XP and Windows 7.
(Page 1 of 2)
Solinst always recommends using the latest software and firmware release.Current releases can be downloaded at any time, free of charge, fromthe Solinst web site at: http://www.solinst.com/Downloads/.
Levelogger Software & Firmware
NOTE:
Operating System Requirements
Direct Read Cables, Optical Readers, Well Caps and other accessories are compatible with all versions of the Levelogger and Rainlogger Models.
For information on the complete list of accessories available, please visit www.solinst.com
Optical Reader
(USB)
PC Interface Cable
(USB)
PC Interface Cable
(RS232)
Direct Read Cable Assembly
with Levelogger Edge
Levelogger Accessories
Solinst always recommends that our customers downloadthe data from their Leveloggers, using the same LeveloggerSoftware version they were originally programmed with.
Example: If a Levelogger Gold unit was programmed using Levelogger Software 3.4.1,Solinst recommends first downloading the data from your Levelogger Gold usingLevelogger Software Version 3.4.1 and barometrically compensating the data in thatversion first, before upgrading your Levelogger Gold and PC Software to thelatest version. After successfully retrieving the Levelogger data andbarometrically compensating the data (if desired), you can then proceed toupgrade your Levelogger firmware and PC Software to the latest version.
Note:Always use the latest Levelogger Software version to install the latest Firmware versions, unless otherwise stated.
-
7/25/2019 Conductivity sensor 3001 Compatibility
2/2
Printed in CanadaNovember 12, 2013
3001 Levelogger Series: Hardware Compatibility
For further information contact: Solinst Canada Ltd.Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8Web Site: www.solinst.com E-mail: [email protected]
Recommended Software & Firmware Congurations
PC Software Firmware Leveloader Gold Firmware
3001 Levelogger Gold
4.1
2.007
2.001
Note: Use a fully charged 9V battery in your
Leveloader during upgrade. A depleted battery may
cause a communication interruption.
3001 Barologger Gold
3001 Junior*
3002 Rainlogger2.000
(Page 2 of 2)
3001 Mini LT Levelogger Note:The previous version of the Levelogger LT (Mini-Silver) and the LTC (Black Zirconium Oxide) are stillcompatible with all of the Solinst communication cables, except the SDI-12 Cable. These Leveloggers arecompatible only with Levelogger PC Software 3.1.1 with Windows 98, 2000 or XP,and Leveloader Firmware 1.000.
Previous Levelogger Versions
To upgrade Levelogger Gold or Barologger Gold Firmware from 1.017 or higher, use the latest Firmware Upgrade Utility, which
comes with the latest Levelogger Software version.
To upgrade Levelogger Gold or Barologger Gold Firmware from 1.016 or lower, please contact Solinst.
If you are programming Leveloggers using Levelogger Series Software 3.4.1, you must use Leveloader Firmware 1.200. Levelogger PC Software Version 3.4.1 is compatible with Windows 98, 2000, XP and Windows 7.
SDI-12 Cables sold after November 1, 2013 will be compatible with all Leveloggers and Rainlogger Models with latest firmware.
Software/Firmware Upgrade Notes:
*Note:The Levelogger Junior, looks very similar to the new Levelogger Junior Edge. To determine the difference, look at the serial number on the bodyof the Levelogger. If the fourth number in the serial number is 1, it is a Levelogger Junior. If the fourth number is a 2, it is a Levelogger Junior Edge.