Concerns in pH # 4: Reference Contamination
-
Upload
orlando-finley -
Category
Documents
-
view
14 -
download
0
description
Transcript of Concerns in pH # 4: Reference Contamination
![Page 1: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/1.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
Concerns in pH # 4:
Reference Contamination
CONTROLS
![Page 2: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/2.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
pH electrodes have porous reference junctions whichallows KCl to migrate through the junction to complete the electrochemical connection and cause a mV signal to be produced.
Since the junction is porous the process can also migrate into the reference causing referencecontamination and premature failure of the pH electrode
PROBLEM
![Page 3: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/3.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
CAUSEAs the process migrates into the reference it attacks theSilver/Silver Chloride wire changing the chemistry which results in the electrode drifting. This is seen as anoffset from zero mV during calibration.
![Page 4: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/4.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
Silver/Silver ChlorideWire
Porous Frit
Solid Spool Piece
Porous ReferenceJunction
Potassium Chloride (KCl)Internal Fill Solution
Typical pH Electrode Internal
![Page 5: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/5.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
Process
With no protection in the reference the KCl and Silver/Silver Chloride wire becomes contaminated very quickly.
Contamination with Standard Internal
![Page 6: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/6.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
Process
By adding the Double Junction, the path length is increased and the rate of contamination is decreased.
Contamination with Double Junction Internal
![Page 7: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/7.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
Process
By adding the Plasticized Reference, which is solid up to 150oC, the migration is slower because it will take longer to migrate through a solid in comparison to a liquid.
Contamination with PlasticizedInternal
![Page 8: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/8.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
Process
By combining the Double Junction and the plasticized Reference, the path length is increased and the rate of contamination is decreased which will give maximum life in applications where reference contamination is a problem. Life expectancy can be months longer when this combination is used.
Contamination with Double Junctionand Plasticized Internal
![Page 9: Concerns in pH # 4: Reference Contamination](https://reader036.fdocuments.us/reader036/viewer/2022072016/56813258550346895d98e353/html5/thumbnails/9.jpg)
IC Controls Quality Water Solutions for pHIC Controls Quality Water Solutions for pHwww.iccontrols.com
pH / ORPConductivityDissolved OxygenChlorineStandards
R1.0 © 2004 IC CONTROLS
IC Controls Plasticized, DoubleJunction Reference
Longer Path Length
Slower Migration
Solid up to 150oC
Longer Sensor LifeProven Technology
Superior DesignBetter Results