Comparison of the reliability of laser Doppler flowmetry,
description
Transcript of Comparison of the reliability of laser Doppler flowmetry,
![Page 1: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/1.jpg)
Comparison of the reliability of laser Doppler flowmetry,
pulse oximetry and electric pulp tester in assessing the
pulp vitality of human teeth
H. KARAYILMAZ & Z. KIRZIOGLU Faculty of Dentistry, Department of Pedodontics, Suleyman Demirel University,
Isparta, Turkiye
SUMMARY This study was designed to evaluate and
compare the reliability of laser Doppler flowmetry
(LDF), pulse oximetry (PO) and electric pulp tester
(EPT) for assessing the pulpal status as a vitality test
method by calculating their sensitivity, specificity
and predictive values. Data were collected from
59 pairs of maxillary anterior teeth (38 pairs of
central, 21 pairs of lateral incisors) in 51 patients
(range 12–18 years, mean age 14Æ6 � 1Æ73 years, 28
women, 23 men). The teeth with complete end-
odontic fillings constituted the study group, and the
healthy, contralateral teeth of the same patients
were constituted the control group. The calculated
sensitivity was 0Æ915 for the EPT and 0Æ813 for the
PO. And the specificity of EPT was 0Æ881 and PO was
0Æ949. The difference between the LDF values
obtained from the study, and control group was
statistically significant (P = 0Æ0001). The findings of
this study indicated that LDF can reliably discrim-
inate the vitality of the teeth with a sensitivity and
specificity of 1Æ0 for this sample. Laser Doppler
flowmetry was found to be a more reliable and
effective method than PO and EPT of assessing the
pulpal status of human teeth.
KEYWORDS: laser Doppler flowmetry, pulse oxime-
try, pulp vitality testing, pulpal blood flow
Accepted for publication 26 August 2010
Introduction
The vitality assessment of teeth is a critical diagnostic
procedure in the practice of dentistry. But it is compli-
cated by the fact that the dental pulp is enclosed within
calcified tissue. As the pulp tissue cannot be directly
inspected, the dentist has to use indirect methods (1).
However, the most widely used traditional pulp vitality
test methods, such as electric pulp testers (EPT) and
thermal stimulus, determine only the pulp sensitivity to
the stimuli used but given no direct indication of blood
flow within the pulp. Consequently, false-positive or
false-negative responses can be obtained from traumat-
ically injured teeth or from immature teeth (2, 3).
Furthermore, each of these subjective methods is
depends on the patient’s perceived response to a
stimulus, as well as the dentist’s interpretation of that
response (4). Also, these methods have a potential
to produce an unpleasant and occasionally painful
sensations. Thus, the reliability of these methods can
vary, and they are of limited use with children (5).
Therefore, new, improved diagnostic methods are
needed to assist in the diagnosis of teeth with pulpal
pathosis and consequently to aid in their correct
treatment. Pulse oximetry (PO) and laser Doppler
flowmetry (LDF) are non-invasive methods for assess-
ing blood flow in microvascular systems, which have
been recently introduced as a new method to diagnose
pulp vitality in human teeth.
The PO measurement technique
Pulse oximetry is a relatively recent advancement in
non-invasive monitoring of oxygen saturation (SaO2) of
the blood and pulse rate of the patient. It is effectively
and routinely used in medical applications through the
use of finger, toe, ear and foot probes. Its wide acceptance
in the medical field results from its ease of application
ª 2010 Blackwell Publishing Ltd doi: 10.1111/j.1365-2842.2010.02160.x
Journal of Oral Rehabilitation 2011 38; 340–347
J o u r n a l o f Oral Rehabilitation
![Page 2: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/2.jpg)
and its capability of providing vital information about the
patient’s status (6, 7). The principles of PO are based on a
modification of Beer’s law and the absorbance charac-
teristics of haemoglobin in the red and infrared range.
The PO probe consists from a photo-detector and two
light-emitting diodes (LED). One of the LED transmits
red light (640 nm), and the other transmits infrared light
(960 nm) to the vascular tissue. Oxygenated and deox-
ygenated haemoglobin absorbs different amount of red
and infrared light. The pulsate change in the blood
volume causes periodic changes in the amount of red and
infrared light absorbed by the vascular tissue before
reaching the photo-detector, and PO uses this informa-
tion to calculate the pulse rate and SaO2 (5–7).
The LDF measurement technique
Laser Doppler flowmetry is non-invasive electro-optical
technique, which has been shown to have potential as a
method of assessing the vitality of teeth by detecting the
presence or absence of a pulpal blood flow (PBF) (8, 9).
The LDF technique utilises a beam of infrared (780–
820 nm) or near infrared (632Æ8 nm) light that is
directed to the tissue by optical fibres within a special
designed probe. Monochromatic laser light is transmit-
ted through the crown of the tooth to the dental pulp
via the probe, and it is scattered by moving red blood
cells and stationary tissue cells. Photons which are
reflected from the moving red blood cells are scattered
and frequency shifted according to the Doppler princi-
ple. Photons that interact with stationary tissue cells are
scattered but not Doppler shifted. The proportion
of shifted to unshifted light within the reflected light
gives a semi-quantitative measurement of blood flow
through the tissue and recorded as a voltage output
from the LDF (8–11). The measured voltage was
linearly related to the PBF and expressed in arbitrary
perfusion units (PU) (1 PU = 10 mV) in accordance
with general consensus (European Laser Doppler Users’
Groups, 1992) (8–14).
This study was designed to evaluate and compare the
reliability of LDF, PO and EPT for assessing the pulpal
status as a vitality test method by calculating their
sensitivity, specificity, and predictive values.
Materials and methods
The study sample was constituted from the archives of
the patients who were attending to the Department
of Pedodontics, Suleyman Demirel University, Faculty
of Dentistry. The participants selected for the study had
one or two endodontically treated (at least 1 year
before) maxillary incisor teeth; the contralateral of the
same teeth was healthy. Also, the participants were
non-smokers and had no history of systemic vascular or
cardiovascular disease or any evidence of hypertension,
and none of them were taking any medication. The
research ethical committee of Suleyman Demirel Uni-
versity, Faculty of Medicine, approved the study
(09.03.2006-02 ⁄ 17). The experimental purpose and
methodology were explained to the patients or their
parents ⁄ carers, and an informed consent was obtained.
Data were collected from 59 pairs of maxillary
anterior teeth (38 pairs of central, 21 pairs of lateral
incisors) in 51 patients (range 12–18 years, mean age
14Æ6 � 1Æ73 years, 28 women, 23 men) (Fig. 1). The
teeth with complete endodontic fillings constituted
the study group, and the healthy, contralateral teeth
of the same patients were constituted the control group.
A full history was taken, and clinical and radiographical
examinations of patients were carried out by the same
examiner in accordance with normal clinical practice.
All the teeth selected for the study had clinically intact
crowns or small restorations, which were located in
areas away from cervical region, where the vitality tests
were performed. Sample criteria required teeth to be
free of caries, developmental defects, discolouration or
root resorption.
Patients were warned to abstain from hot or cold
foods and beverages for at least 2 h before attending. All
measurements were performed in a temperature-con-
trolled room (24 � 1�C) using the same unit and in
keeping the same position. The patients rested for
10 min in the unit before the measurements.
Fig. 1. The distribution of 51 participants according to age and
gender.
C O M P A R I S O N O F T H E R E L I A B I L I T Y O F L D F , P O A N D E P T 341
ª 2010 Blackwell Publishing Ltd
![Page 3: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/3.jpg)
The PO apparatus
A commercially available ‘Life Scope I, Multiparameter
Bedside Monitor’ (Model BSM-2301K*) and a modified
infant probe were used to record SaO2 levels. Special
probe holders designed taking into consideration the
morphology of the maxillary incisors to hold the
modified probe on the tooth. Stainless steel clips and
rubber dam clamps were used as the base for the
holders. The tapes enclosing the LEDs and photo-
detector of the infant probe were removed and attached
to the holder in parallel to each other (Figs 2 and 3).
The PO measurement procedures
The probe was positioned on the cervical region of the
crown of the tooth using probe holder. So that light
would travel from the facial to the palatinal side
through the middle of the crown. Pulse oximetry
values were recorded after 45 s of monitoring each of
the teeth. If there was no response at the end of the
measurement period, the PO values of the tooth were
recorded as negative (Fig. 4).
The LDF apparatus
The PBF of the teeth was measured by a commercially
available LDF device (BLF21A†; wavelength 780 nm)
and a custom made (ext. diam.; 1Æ5 mm, two fibre in
0Æ2 mm diam., centres 0Æ5 mm apart) dental probe in
this study (Figs 5 and 6).
The LDF measurement procedure
Silicon-impression-based personal splints were prepared
to ensure accurate and reproducible positioning of the
probe on the tooth for each of the participants. On the
labial side of the splints, holes 2 mm above the gingival
margins were drilled to insert and to fix the probe. The
probe was held perpendicular to the surface of the crown
2 mm from the gingival margin using these splints.
Evaluation took 45 s for each tooth, and data were
collected by a PC connected to the LDF device while
maintaining a real-time display on the monitor. The
20 s of the data, which was optimum part of the
measurement, were selected for the study by a special
software package (Windaq ver. 2.36‡), and the average
Fig. 2. The appearance of modified infant probe.
Fig. 4. The placement the modified infant probe with probe
holder on the tooth and the measurement procedure.Fig. 3. The appearance of special designed probe holder.
*Nihon Kohden Corp., Tokyo, Japan.
†Transonic Systems Inc., Ithaca, NY, USA.‡DATAQ Instruments Inc., Akron, OH, USA.
H . K A R A Y I L M A Z & Z . K I R Z I O G L U342
ª 2010 Blackwell Publishing Ltd
![Page 4: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/4.jpg)
PBF of the teeth was calculated in PU by the same
software (Figs 7 and 8).
The EPT measurement procedures
A conventional EPT (Pulptester, Model PT-20§) was
used for EPT, and the response was recorded as positive
if the teeth tested showed any response on the
aforementioned scale or negative if there was no
response.
Other measurements
The following measurements were performed to exam-
ine the effects of the values obtained from the teeth by
PO and LDF:
1 The systemic SaO2 level and pulse rate of the patients
were measured by PO device from the index finger
using a finger probe.
2 Systolic and diastolic arterial blood pressures were
also measured and recorded.
Statistical analyses (paired t test, Mann–Whitney U,
Pearson’s correlation) were performed using SPSS (Ver.
13.0¶) at P < 0Æ05 significance level. Also the sensitivity,
Fig. 5. The appearance of commer-
cially available laser Doppler flow-
metry.
Fig. 7. Silicon-impression-based personal splints were used to
ensure accurate and reproducible positioning of the probe on the
tooth.
Fig. 6. The appearance of device and a custom made dental probe
used in this study.
§Parkell Inc., Edgewood, NY, USA. ¶SPSS Inc., Chicago, IL, USA.
C O M P A R I S O N O F T H E R E L I A B I L I T Y O F L D F , P O A N D E P T 343
ª 2010 Blackwell Publishing Ltd
![Page 5: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/5.jpg)
specificity, positive predictive value (PPV) and negative
predictive value (NPV) were calculated.
Results
Electric pulp tester identified 5 of the 59 teeth with
complete endodontic fillings as vital, while 54 as non-
vital. In control group consisted from the healthy,
contralateral tooth of the same patient, 52 teeth were
identified as vital and 7 as non-vital with EPT.
The 59 tooth with complete endodontic fillings, 11
teeth gave positive response, whereas 48 teeth
responded negatively to the PO. But in the control
group, 56 gave a positive response, whereas three
responded negatively.
The average PO and LDF values obtained from the
healthy, contralateral teeth and teeth with complete
endodontic fillings have been summarised in Table 1.
In general, it was determined that approximately a
1 ⁄ 10 ratio between the PBF values measured by LDF
Fig. 8. The display of the waveform that synchronised with heart beat.
Table 1. The distribution of laser Doppler flowmetry (LDF) and pulse oximetry (PO) values obtained from healthy, contralateral teeth
and teeth with complete endodontic fillings
Central incisors Lateral incisors Total
n Min. Max. Mean � s.d. n Min. Max. Mean � s.d. n Min. Max. Mean � s.d.
Healthy, contralateral teeth
LDF (PU) 38 5Æ84 18Æ17 10Æ43 � 3Æ25 21 6Æ23 18Æ11 12Æ11 � 3Æ71 59 5Æ84 18Æ17 11Æ02 � 3Æ48
PO (%) 38 81 93 86Æ32 � 3Æ33 21 80 92 87Æ47 � 3Æ06 59 80 93 86Æ71 � 3Æ26
Teeth with complete endodontic fillings
LDF (PU) 38 0Æ13 2Æ21 0Æ99 � 0Æ54 21 0Æ8 2Æ35 1Æ29 � 0Æ66 59 0Æ13 2Æ35 1Æ09 � 0Æ60
PO (%) 38 00 79 17Æ5 � 31Æ86 21 00 76 6Æ95 � 21Æ97 59 00 79 13Æ74 � 28Æ99
PU, perfusion unit.
H . K A R A Y I L M A Z & Z . K I R Z I O G L U344
ª 2010 Blackwell Publishing Ltd
![Page 6: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/6.jpg)
from the tooth with complete endodontic fillings and
the healthy contralateral tooth of the same patient. The
difference between the study and the control group was
statistically significant (P = 0Æ0001). Thereupon, it was
deduced that LDF can successfully differentiate the
vitality of the teeth inspected in the study.
On the basis of these findings, the sensitivity, spec-
ificity and predictive values were calculated for each of
the method. The sensitivity and the specificity of EPT
were 0Æ915 and 0Æ881, respectively. But for the PO, the
sensitivity was 0Æ813 and the specificity was 0Æ949. The
calculated PPV and NPV for EPT were 0Æ885 and 0Æ912,
respectively. For the PO, the PPV was 0Æ941 and the
NPV was 0Æ835.
Under the findings from vitality assessment using
LDF in this study would give LDF a sensitivity and
specificity of 1Æ0 in discrimination between teeth with
complete endodontic fillings and healthy, contralateral
tooth of the patients.
In addition to these findings, statistically significant
positive correlations were observed between the LDF
values obtained from the study and the control group
(r = 0Æ59, P = 0Æ0001) and between PO and LDF values
obtained from healthy, contralateral teeth (r = 0Æ32
P = 0Æ016).
Although the PO values obtained from the healthy,
contralateral teeth were showed a positive correlations
only with pulse rate (r = 0Æ28 P = 0Æ034) of the patients,
the LDF values were showed statistically significant
correlations with the systemic SaO2 levels (r = 0Æ34
P = 0Æ009) and pulse rate (r = 0Æ30 P = 0Æ019) of the
patients. The gender and arterial blood pressures of the
patients were not significant.
Discussion
The perfect diagnostic test would always be positive in
the presence of disease and negative in the absence of
disease. The extent to which a test correctly classifies
patients defines its accuracy. The concepts of sensitiv-
ity, specificity, PPV and NPV have been developed to
characterise test accuracy and to compute the benefits
of test usage (15). Sensitivity denotes the ability of a
test to detect disease in patients who actually have the
disease. Conversely, specificity describes the ability of
a test to detect the absence of disease. The ideal pulp
test method would have sensitivity and specificity of
1Æ0. The PPV is the probability that a positive test
result actually represents a disease-positive person.
Negative predictive value is the probability that a
person with a negative test result is actually free of
disease (15).
This study was designed to evaluate and compare the
reliability of EPT, PO and LDF as a vitality test method
by calculating their sensitivity, specificity and predictive
values. To our knowledge, this is the first study
evaluating and comparing the reliability of EPT, PO
and LDF together in same study as a vitality test
method.
The calculated sensitivity and specificity of PO and
EPT in our study denoted that 91% of the teeth with
complete endodontic fillings were identified as non-
vital by EPT, while 81% of these teeth were identified
as non-vital by PO. Likewise, 88% of the healthy
contralateral teeth were identified as vital by EPT,
while 95% of these teeth were identified as vital by
PO. Thus, the calculated PPV and NPV denoted that
there was a probability of 88% that no sensitive
reaction represented a root-canal filled tooth, and a
there was a probability of 91% that sensitive reaction
represented a healthy, vital tooth when EPT was used.
Likewise, the probability that no sensitive reaction
represented a root-canal filled tooth was 94%, while
the probability that a sensitive reaction represented a
healthy, vital tooth was 83% with PO. These results
imply that the ability of PO in determining the vitality
of healthy teeth was found better than EPT, but PO
was found insufficient in determining the vitality of
teeth with complete endodontic fillings. The reasons
for this could be that the scattering of the light
(transmitted from PO) from the composite restorations
of the root-canal-treated teeth to the adjacent tissues
(gingiva, etc.) and the physical limitations of PO.
Therefore, the efficacy of PO in teeth with composite
restorations that located in cervical region of the tooth
caused suspicion.
Although some of the studies had shown the effec-
tiveness and accuracy of PO in determining the PBF (5,
16–20), a few studies reported that the overall accuracy
of the commercially available PO instruments was
disappointing and, in its present form, was not consid-
ered to have predictable diagnostic value in determin-
ing the PBF of the teeth (21–23).
Recently, Gopikrishna et al. (19) evaluated the effi-
cacy of PO in comparison with the conventional pulp
vitality tests. They were reported that the sensitivity
and the specificity of PO were 1Æ00 and 0Æ95, respec-
tively. Although specificity of PO reported in both of
C O M P A R I S O N O F T H E R E L I A B I L I T Y O F L D F , P O A N D E P T 345
ª 2010 Blackwell Publishing Ltd
![Page 7: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/7.jpg)
the studies was similar, the sensitivity of PO reported in
our study groups was low. It was considered that the
disagreements arise from the differences in study
groups selected and the methods preferred for the
studies. Single-rooted incisors requiring endodontic
therapy were selected for their study, but the teeth
with complete endodontic fillings constituted the study
group and the healthy, contralateral teeth of the same
patients constituted the control group in our study.
In view of these results, it was determined that LDF
can reliably discriminate between vital and non-vital
dental pulps of maxillary anterior teeth with a sensi-
tivity and specificity of 1Æ0 for this sample, and LDF was
found to be a more reliable method than PO and EPT.
There were a number of studies in the literature that
have proved these results (8, 10–14, 24–26). Of these
studies, Evans et al. (12) reported in accordance with
our results, the sensitivity and specificity of LDF were
1Æ0 in their study. Using LDF as a vitality test method, it
appeared that we could diagnose pulpal necrosis more
accurately than with conventional test methods. This
may lead to less pain and discomfort for the patient, less
cost and earlier treatment. However, despite its advan-
tages, LDF still has a number of specifically technical
limitations (27, 28). These include, motion artefact
noise, multiple Doppler shifting, variations in instru-
ment and probe specifications, lack of quantitative units
and knowledge of depth of measurements, lack of the
knowledge of normal PBF values of the healthy teeth,
environmental effects to the measurements and high
costs of the instrument.
Conclusion
Laser Doppler flowmetry was found to be a more
reliable and effective method than PO and EPT for
assessing the pulpal status of human teeth especially in
paediatric patients where patient co-operation and
incomplete pulp innervations reduce the effectiveness
and reliability of conventional test methods. However,
the improvement of the LDF method and apparatus
with further researches is indicated to become a
valuable clinical diagnostic tool in practice of dentistry.
Acknowledgment
This study was supported by the scientific research
grant given by Suleyman Demirel University (SDUBAP,
1212-D-05).
References
1. Rowe AHR, Pitt Ford TR. The assessment of pulp vitality. Int
Endod J. 1990;23:77–83.
2. Bhaskar SN, Rappaport HM. Dental vitality tests and pulp
status. J Am Dent Assoc. 1973;86:409–411.
3. Klein H. Pulp response to electrical pulp stimulator in the
developing permanent dentition. J Dent Child. 1978;45:
199–202.
4. Ehrmann EH. Pulp testers and pulp testing with particular
reference to the use of dry ice. Aust Dent J. 1977;22:
272–279.
5. Goho C. Pulse oximetry evaluation of vitality in primary
and immature permanent teeth. Pediatr Dent. 1999;21:
125–127.
6. Samraj RV, Indira R, Srinivasan MR, Kumar A. Recent
advances in pulp vitality testing. Endodontology.
2003;15:14–19.
7. Jafarzadeh H, Rosenberg PA. Pulse oximetry: review of a
potential aid in endodontic diagnosis. J Endod. 2009;35:
329–333.
8. Wilder-Smith PE. A new method for the non-invasive
measurement of pulpal blood flow. Int Endod J. 1988;
21:307–312.
9. Jafarzadeh H. Laser Doppler flowmetry in endodontics: a
review. Int Endod J. 2009;42:476–490.
10. Ramsay DS, Artun J, Martinen SS. Reliability of pulpal blood
flow measurements utilizing laser Doppler flowmetry. J Dent
Res. 1991;70:1427–1430.
11. Gazelius B, Olgart L, Edwall B. Restored vitality in luxated
teeth assessed by laser Doppler flowmeter. Endod Dent
Traumatol. 1988;4:265–268.
12. Evans D, Reid J, Strang R, Stirrups D. A comparison of laser
Doppler flowmetry with other methods of assessing the
vitality of traumatized anterior teeth. Endod Dent Traumatol.
1999;15:284–290.
13. Ingooolfsson AR, Tronstad L, Hersh EV, Riva CE. Efficacy of
laser Doppler flowmetry in determining pulp vitality. Endod
Dent Traumatol. 1994;10:83–87.
14. Ingooolfsson AR, Tronstad L, Riva CE. Reliability of laser
Doppler flowmetry in testing vitality of human teeth. Endod
Dent Traumatol. 1994;10:185–187.
15. Weinstein MC, Fineberg HV. The use of diagnostic informa-
tion to revise probabilities. In: Weinstein MC, Fineberg HV,
eds. Clinical decision analysis. Philadelphia, PA: WB Saunders,
1980:75–130.
16. Radhakrishnan S, Munshi AK, Hedge AM. Pulse oximetry: a
diagnostic instrument in pulpal vitality testing. J Clin Pediatr
Dent. 2002;26:141–145.
17. Gopikrishna V, Kandaswamy D, Gupta T. Assessment of the
efficacy of an indigenously developed pulse oximeter dental
sensor holder for pulp vitality testing. Indian J Dent Res.
2006;17:111–113.
18. Gopikrishna V, Tinagupta K, Kandaswamy D. Comparison of
electrical, thermal, and pulse oximetry methods for assessing
pulp vitality in recently traumatized teeth. J Endod.
2007;33:531–535.
H . K A R A Y I L M A Z & Z . K I R Z I O G L U346
ª 2010 Blackwell Publishing Ltd
![Page 8: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/8.jpg)
19. Gopikrishna V, Tinagupta K, Kandaswamy D. Evaluation of
efficacy of a new custom-made pulse oximeter dental probe in
comparison with the electrical, and thermal, tests for assessing
pulp vitality. J Endod. 2007;33:411–414.
20. Calil E, Caldeira CL, Gavini G, Lemos EM. Determination of
pulp vitality in vivo with pulse oximetry. Int Endod J.
2008;41:741–746.
21. Kahan RS, Gulabivala K, Snook M, Setchell DJ. Evaluation of
a pulse oximeter and customized probe for pulp vitality
testing. J Endod. 1996;22:105–109.
22. Schnettler JM, Wallace JA. Pulse oximetry as a diagnostic tool
of pulpal vitality. J Endod. 1991;17:448–490.
23. Schmitt JM, Webber RL, Walker EC. Optical determination of
dental pulp vitality. IEEE Trans Biomed Eng. 1991;38:
346–352.
24. Gazelius B, Olgart L, Edwall B, Edwall L. Non-invasive
recording of blood flow in human dental pulp. Endod Dent
Traumatol. 1986;2:219–221.
25. Fratkin DR, Kenny DJ, Johnston DH. Evaluation of a laser
Doppler flowmeter to assess blood flow in human primary
incisor teeth. Pediatr Dent. 1999;21:53–56.
26. Ingooolfsson AR, Tronstad L, Hersh EV, Riva CE. Effect of
probe design on the suitability of laser Doppler flowmetry in
vitality testing of human teeth. Endod Dent Traumatol.
1993;9:65–70.
27. Polat S, Er K, Akpınar KE, Polat NT. The sources of laser
Doppler blood-flow signals recorded from vital and root canal
treated teeth. Arch Oral Biol. 2004;49:53–57.
28. Leahy MJ, de Mul FF, Nilsson GE, Maniewski R. Principles
and practice of the laser-Doppler perfussion technique. Tech-
nol Health Care. 1999;7:143–162.
Correspondence: Dr Huseyin Karayilmaz, Suleyman Demirel Univers-
itesi, Dis Hekimligi Fakultesi, Pedodonti Anabilim Dalı, Cunur
Kampusu, Isparta, Turkiye. E-mail: [email protected]
C O M P A R I S O N O F T H E R E L I A B I L I T Y O F L D F , P O A N D E P T 347
ª 2010 Blackwell Publishing Ltd
![Page 9: Comparison of the reliability of laser Doppler flowmetry,](https://reader036.fdocuments.us/reader036/viewer/2022081908/552ceee74a7959b95b8b4593/html5/thumbnails/9.jpg)
This document is a scanned copy of a printed document. No warranty is given about the accuracy of the copy.
Users should refer to the original published version of the material.