CHAPTER-4 A VALIDATED STABILITY-INDICATING...

52
158 CHAPTER-4 A VALIDATED STABILITY-INDICATING ANALYTICAL METHOD FOR THE DETERMINATION OF IMPURITIES IN FLORFENICOL

Transcript of CHAPTER-4 A VALIDATED STABILITY-INDICATING...

  • 158

    CHAPTER-4

    A VALIDATED STABILITY-INDICATING

    ANALYTICAL METHOD FOR THE

    DETERMINATION OF IMPURITIES IN

    FLORFENICOL

  • 159

    4.1 Introduction of Florfenicol and survey of analytical Methods

    Florfenicol is a synthetic derivative of thiamphenicol. It has a

    wide spectrum of antibacterial properities, including efficacy against

    Gram-positive and Gram-negative bacteria [1]. The mechanism of

    action of Florfenicol similarly to chloramphenicol and thiamphenicol,

    relies on slowing down the biosynthesis of bacterial proteins through

    reversible binding with an active center of peptydylotransferase on a

    subunit 50S of bacterial ribosomes. It is chemically designated as 2,2-

    dichloro-N-[(1S,2R)-1-(fluoromethyl)-2-hydroxy-2-[4-(methylsulfonyl)

    phenyl]ethyl] acetamide. Florfenicol is a white to almost white,

    crystalline powder. It is very soluble in N,N-Dimethylformamide and

    soluble in methanol. The molecular formula is C12H14Cl2FNO4S. The

    molecular weight of Florfenicol is 358.22.

    Fig: 4.1 Chemical structure of Florfenicol

    H

    Cl

    S FO

    N

    OH Cl

    O

    H3C

    O

    2,2-dichloro-N-[(1S,2R)-1-(fluoromethyl)-2-hydroxy-2-[4-

    (methylsulfonyl)phenyl]ethyl]acetamide

    Molecular formula C12H14Cl2FNO4S

    Molecular weight 358.22

  • 160

    The different analytical techniques reported so far for the

    determination of this drug and its metabolites in biological samples

    include capillary electrophoresis [2] and spectrometry [3]. The

    determination of Florfenicol in plasma by RP-LC [4], stability of

    Florfenicol [5-6], pharmacokinetics studies of Florfenicol in plasma [7-

    8] was also reported.

    Organic impurities can arise during the manufacturing process

    and storage of the drug substances and the criteria for their

    acceptance up to certain limits are based on pharmaceutical studies

    or known safety data [9]. As per regulatory guidelines, the

    pharmaceutical studies using a sample of the isolated impurities can

    be considered for safety assessment. It is, therefore, essential to

    isolate and characterize unidentified impurities present in the drug

    sample. Recently we have developed a process for the synthesis of

    Florfenicol in our laboratory. During the development of an analytical

    procedure, the LC method was developed for the determination of in-

    house synthesized Florfenicol and the impurities arising during its

    manufacturing. In the present study, we describe a reverse phase

    column liquid chromatography method for the separation and

    quantification of process related and degradation impurities of

    Florfenicol. The accuracy, precision, limit of detection (LOD), limit of

    quantification (LOQ) and robustness of the method were determined in

    accordance with ICH guidelines [10]. The target is to develop a suitable

    stability-indicating HPLC related substances method for Florfenicol in

  • 161

    this chapter we describe a stability-indicating LC method for the

    determination of Florfenicol and its potential and degradation

    impurities and also the method validation.

    4.2 Development of a stability-indicating analytical method for

    Florfenicol

    4.2.1 Materials

    Reference standard of Florfenicol and seven impurities namely,

    Imp-A, Imp-B, Imp-C, Imp-D, Imp-E , Imp-F and Imp-G (Fig: 4.1)

    were synthesized and characterized by use of LC-MS, NMR and IR in

    Aurobindo Pharma Ltd., Hyderabad, India. The commercial samples of

    Florfenicol are also manufactured by Aurobindo Pharma Ltd. All

    reagents used were of analytical reagent grade unless stated

    otherwise. Milli Q water, HPLC-grade acetonitrile, HPLC-grade

    orthophosphoric acid (OPA) were purchased from Merck (Darmstadt,

    Germany).

    4.2.2 Equipment

    The LC system was equipped with quaternary gradient pumps

    with autosampler and auto injector (Alliance, Waters 2695, Milliford,

    MA, USA) controlled with Empower software (Waters).

  • 162

    Fig: 4.2 Chemical structures of impurities of Florfenicol

    Me O2S F

    NH2

    OH

    (1R,2S)-1-[4-(Methylsulfonyl) phenyl]-2-amino-3-fluoro-1-propanol

    (Imp-A)

    Fig 4.2 (a)

    FMe O2S

    N

    OH

    CH3

    O

    H

    (1R,2S)-2-Acetamido-3-fluoro-1-[4-(methyl-sulfonyl)phenyl]-1-

    propanol (Imp-B)

    Fig: 4.2 (b)

    Cl

    Me O2S OHO

    N

    OH ClH

    (1R,2R)-2-Dichloroacetamido-1-[4-(methylsulfonyl)phenyl]-1,3-

    propanediol (Imp-C)

    Fig: 4.2 (c)

    Cl

    Me O2S FO

    N

    OHH

    (1R,2S)-2-Chloroacetamido-3-fluoro-1-[4-(methylsulfonyl)phenyl]-1-

    propanol (Imp-D)

    Fig: 4.2 (d)

  • 163

    Cl

    Me O2S ClO

    N

    OH ClH

    (1R,2S)-2-Dichloroacetamido-3-fluoro-1-[4-(methylsulfonyl)phenyl]-1-

    propanol (Imp-E)

    Fig: 4.2 (e)

    Cl

    MeO2S FO

    N

    OH ClCH3

    (1R,2S)-2-(N-Methyl)dichloroacetamido-3-fluoro-1-[(4-methylsulfonyl)-

    phenyl]-1-propanol (Imp-F)

    Fig: 4.2 (f)

    Cl

    Me O2S FO

    N

    OH ClClH

    (1R,2S)-2-Trichloroacetamido-3-fluoro-1-[4-(methylsulfonyl)-phenyl]-1-

    propanol

    (Imp-G)

    Fig: 4.2 (g)

    4.2.3 Preparation of sample and stock solutions

    The stock solutions of Florfenicol (0.5 mg/ml) and spiked with

    0.3% of Imp-A of Imp-B, Imp-D, Imp-E, Imp-F and Imp-G with respect

    to the Florfenicol analyte concentration. The stock solutions was

    further diluted with diluent to obtain a standard solution of 0.0005

  • 164

    mg/ml (0.5 µg/ml) for related substances determination and 0.5

    mg/ml ( 500 µg/ml) for assay determination.

    4.2.4 Generation of stress samples

    One lot of Florfenicol drug substance selected for stress

    testing. From the ICH stability guideline: “stress testing likely to be

    carried out on a single batch of material. Different kinds of stress

    conditions (i.e., acid hydrolysis, base hydrolysis, oxidative stress, heat

    and humidity) were employed on one lot of Florfenicol drug substance

    based on the guidance available from ICH stability guideline (Q1AR2).

    The details of the stress conditions are as follows:

    a) Acid Degradation: drug in 5.0 M HCl solution was kept at 85°C for

    60 mins.

    b) Base Degradation: drug in 5.0 M NaOH solution was kept at room

    temperature.

    c) Oxidative stress: drug in 30% H2O2 solution was kept at 80°C for

    120 mins.

    d) Thermal Degradation: drug was subjected to dry heat at 105°C for

    188 hrs.

    e) Phtolytic degradation: drug was subjected to UV at 254 nm (10 K

    Lux ) for 188 hrs.

    4.2.5 Optimization of chromatographic conditions

    The main objective of the chromatographic method was to

    seperate Florfenicol from Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F

    and Imp-G impurities were coeluted using different stationary phases

  • 165

    such as C8, phenyl and cyano as well as different mobile phases

    containing buffers like phosphate, sulfate and acetate with different

    pH and using organic modifiers like acetonitrile and methanol in the

    mobile phase. Apart from the co-elution of impurities, we have also

    observed poor peak shapes for Florfenicol, some impurities and

    degradants. The chromatographic separation was achieved on a

    Hypersil BDS, C18, (250 mm x 4.6 mm), 5 particle size. The gradient

    LC method employs solution A and B as mobile phase. The solution A

    contains phosphate buffer pH 7.0 and acetonitrile as solution B. The

    flow rate of the mobile phase was 1.0 ml/min. The HPLC gradient

    program was set as: time% solution B: 0.01/10, 15/20, 30/30,

    40/40, 50/60, 60/60, 61/10, 70/10 with a post run time of 10 min.

    The detection was monitored at a wavelength of 225 nm. The injection

    volume was 20 µl. Standard and test solutions were prepared in pH

    3.0 buffer was used as diluent. In the optimized chromatographic

    conditions of Florfenicol, Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F

    and Imp-G were separated with a resolution greater than 3, typical

    relative retention times were approximately 0.29, 0.35, 0.59, 0.61,

    1.35, 1.36, 1.72 with respect to Florfenicol eluted at 22.45 min.

  • 166

    Optimized liquid chromatographic conditions:

    Column : Hypersil BDS, C18, (250 mm x

    4.6 mm), 5

    Mobile phase : The solution A contains

    phosphate buffer pH 7.0 and

    solution B contains acetonitrile

    Pump mode : Gradient

    Flow rate : 1.0 ml/min

    UV of detection : 225 nm

    Injection volume : 20 µl

    Run time : 60 min

    Retention time : 22.45min

    Relative Retention Time (RRT) : Imp-A about 0.29

    Imp-B about 0.35

    Imp-C about 0.59

    Imp-D about 0.61

    Imp-E about 1.35

    Imp-F about 1.36

    Imp-G about 1.72

    Diluent : Dissolve 1.36 g of potassium dihydrogen orthophosphate in

    1000 ml of water. Adjust pH to 3.0±0.05 with orthophosphoric acid.

    No considerable degradation was observed in Florfenicol bulk

    samples under stress conditions such as acid hydrolysis, photolytic

    and thermal. The degradation of drug substance was observed during

    base hydrolysis and oxidative stress condition. Florfenicol was

    degraded to Imp-A (7.85%) under acidic conditions (5M HCl/180°C/60

  • 167

    min). Florfenicol was degraded to Imp-A (7.80%), Imp-C (1.14%) under

    base conditions (5M NaOH/Initial) and it was confirmed by co-

    injection with a qualified Imp-A and Imp-C standards. Mild

    degradation was observed under oxidative environment (treated with

    30% H2O2/85°C/240 min) leads to the formation of some unknown

    degradation peaks (3.5%).

    Peak purity test results obtained by using a PDA detector

    confirmed that the Florfenicol peak is homogenous and pure in all the

    analyzed stress samples. The mass balance of Florfenicol in all stress

    samples was close to 99.6% (%Assay + %Degradation). This clearly

    demonstrates that the developed HPLC method was found to be

    specific for Florfenicol in presence of its impurities (Imp-A, Imp-B,

    Imp-C, Imp-D, Imp-E, Imp-F and Imp-G) and degradation products.

    Figures:

    Fig: 4.3 to Fig: 4.7 is the typical HPLC chromatograms showing

    the degradation of Florfenicol in various stress conditions and also the

    corresponding peak purityplots.

  • 168

    Fig: 4.3 Typical HPLC chromatograms of Acid hydrolysis

    Fig: 4.3 (a)

    Fig: 4.3 (b)

    Blank Chromatogram of Acid hydrolysis (5N HCl)

    Florfenicol stressed with 5N HCl at 85°C for 60 mins

  • 169

    Fig: 4.3 (c) Peak purity plot of Acid hydrolysis

    Purity Angle Purity Threshold Purity Flag Peak Purity

    0.048

    0.249 No Pass

    Fig: 4.3 (c)

  • 170

    Fig: 4.4 Typical HPLC chromatograms of Base hydrolysis

    Fig: 4.4 (a)

    Fig: 4.4 (b)

    Blank Chromatogram of Base hydrolysis ( 5N NaOH )

    Florfenicol stressed with 5N NaOH at room temperature

  • 171

    Fig: 4.4 (c) Peak purity plot of Base hydrolysis

    Purity Angle Purity Threshold Purity Flag Peak Purity

    0.047

    0.247 No Pass

    Fig: 4.4 (c)

    Fig: 4.5 Typical HPLC chromatograms of Peroxide Degradation

    Fig: 4.5 (a)

    Fig: 4.5 (b)

    Blank Chromatogram of Peroxide Degradation ( 30% H2O2 )

    Florfenicol stressed with 30%H2O2 at 80°C for 120 mins

  • 172

    Fig: 4.5 (c) Peak purity plot of Peroxide Degradation

    Purity Angle Purity Threshold Purity Flag Peak Purity

    0.067

    0.272 No Pass

    Fig: 4.5 (c)

    Fig: 4.6 Typical HPLC chromatograms of Thermal Degradation

    Fig: 4.6 (a)

    Blank

  • 173

    Fig: 4.6 (b)

    Fig: 4.6 (c) Peak purity plot of Thermal Degradation

    Purity Angle Purity Threshold Purity Flag Peak Purity

    0.063

    0.261 No Pass

    Fig: 4.6 (c)

    Florfenicol stressed at 105°C for 188 hours

  • 174

    Fig: 4.7 Typical HPLC chromatograms of Photolytic Degradation

    Fig: 4.7 (a)

    Fig: 4.7 (b)

    Blank

    Florfenicol stressed with 12,000 Lux for 188 hours

  • 175

    Fig: 4.7(c): Peak purity of Photolytic Degradation

    Purity Angle Purity Threshold Purity Flag Peak Purity

    0.061

    0.259 No Pass

    Fig: 4.7 (c)

    Peak purity test performed for the Florfenicol peak using photo

    diode array (PDA) detector data confirmed the purity of the peak for

    all the stressed samples. This clearly demonstrates that the developed

    HPLC method was found to be specific for Florfenicol in presence of

    its impurities (Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F, Imp-G

    and degradant products.

    No degradants were observed after 15 min in the extended run

    time of 60 min for all the Florfenicol stressed samples (acid

    hydrolysis, base hydrolysis, oxidation stress, heat 120°C, photolysis)

    with 90% can in mobile phase.

  • 176

    4.2.6 Validation of analytical method and its results:

    The developed and optimized HPLC method was taken up to

    validation. The analytical method validation was carried out in

    accordance with ICH guideline [11].

    4.2.6.1 System suitability: A mixture of Florfenicol standard

    injections were injected into HPLC system and good resolution was

    obtained between the impurities and Florfenicol [Fig: 4.8). The system

    suitability results are tabulated (Table: 4.1).

  • 177

    Fig: 4.8 Typical Blank, Florfenicol sample and SST

    chromatograms

    Fig: 4.8 (a)

    Fig: 4.8 (b)

    Fig: 4.8 (c)

    Blank

    Florfenicol sample spiked with impurities

    Florfenicol Sample

  • 178

    Table: 4.1 System suitability results

    S.NO Impurity

    name USP Plate

    count USP

    Tailing USP

    Resolution

    1 Imp-A 4179 1.23 -

    2 Imp-B 10231 1.10 3.93

    3 Imp-C 29350 1.12 16.68

    4 Imp-D 32575 1.13 1.92

    5 Florfenicol 56327 1.18 24.53

    6 Imp-E 80882 1.10 18.89

    7 Imp-F 88309 1.15 0.85

    8 Imp-G 173154 1.11 22.69

    4.2.6.2 Precision:

    The precision of an analytical process experiment the closeness

    of agreement between a series of measurements obtained from

    multiple sampling of the some homogeneous same under prescribed

    conditions.

    Precision may be considered at three levels: System precision,

    Method precision and Intermediate Precision. Assay method precision

    study was evaluated by carrying out six independent assays of

    Florfenicol test sample against qualified reference standard and RSD

    of six consecutive assays was 0.6% (Table: 4.2 to Table: 4.4).

    The results showed insignificant variation in measured

    response. Which demonstrated that the assay method was repeatable

    with RSD’s below 0.4%.

  • 179

    Table: 4.2 System Precision results of the Assay method

    Injection ID Florfenicol

    1 2194243

    2 2174916

    3 2189161

    4 2184508

    5 2192197

    6 2191167

    Mean 2187699

    SD 7082

    % RSD

    95% Confidence

    Interval

    0.3

    ± 7433

    Table: 4.3 Method Precision results of the Assay method

    Sample ID Assay (% w/w)

    1 98.8

    2 98.8

    3 99.9

    4 98.8

    5 100.2

    6 99.9

    Mean 99.4

    SD 0.67

    % RSD 0.7

    95% Confidence Interval

    ± 0.7

  • 180

    Table: 4.4 Intermediate Precision results of the Assay method

    Sample ID Assay (% w/w)

    1 99.1

    2 100.3

    3 99.4

    4 100.7

    5 100.5

    6 100.6

    Mean 100.1

    SD 0.68

    % RSD 0.7

    95% Confidence

    Interval ± 0.7

    The precision of the related substance method was checked by

    injecting six individual preparations of Florfenicol (0.5 mg/ml) spiked

    with 0.3% of Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F and Imp-G

    with respect to the Florfenicol analyte concentration. The % RSD of

    the area percentage of each impurity (impurities-A, -B, -C, -D, -E, -F

    and –G) for six consecutive determinations was respectively as below

    (Table: 4.5 to Table: 4.7).

    The results showed insignificant variation in measured

    response. Which indicated that the related substance method was

    repeatable with RSD’s below 1.8%.

  • 181

    Table: 4.5 System Precision results of the Related Substance

    method

    Injection ID Florfenicol

    1 64644

    2 64842

    3 64434

    4 64178

    5 64439

    6 64029

    Mean 44290

    SD 44423

    % RSD 130

    95% Confidence

    Interval 0.3

    Table: 4.6 Method Precision results of the Related Substance

    method

    Preparation

    Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F Imp-G

    1 0.217 0.334 0.461 0.311 0.394 0.341 0.471

    2 0.218 0.332 0.462 0.309 0.386 0.344 0.463

    3 0.225 0.333 0.464 0.309 0.391 0.347 0.465

    4 0.226 0.334 0.469 0.311 0.397 0.343 0.474

    5 0.222 0.329 0.463 0.309 0.391 0.346 0.464

    6 0.224 0.330 0.466 0.312 0.392 0.346 0.467

    Mean 0.222 0.332 0.464 0.310 0.392 0.345 0.467

    SD 0.004 0.002 0.003 0.001 0.004 0.002 0.004

    %RSD 1.8 0.6 0.6 0.3 1.0 0.6 0.9

    95%Confidence

    interval ± 0.004 ±0.002 ±0.003 ±0.001 ±0.004 ±0.002 ±0.004

  • 182

    In intermediate precision method six preparations individually

    using single batch of Florfenicol drug substance spiked with related

    substances at specification level (0.3%) as per test method and

    injected each solution as per methodology using different column,

    system and by another analyst. Results showed insignificant variation

    in measured response. Which demonstrated that the related

    substance method was repeatable with RSD’s below 1.9%.

    Table: 4.7 Intermediate Precision results of the Related substance

    method

    Preparation Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F Imp-G

    1 0.217 0.341 0.469 0.309 0.373 0.347 0.466

    2 0.211 0.341 0.466 0.305 0.378 0.349 0.456

    3 0.215 0.346 0.468 0.310 0.377 0.354 0.470

    4 0.217 0.344 0.468 0.309 0.389 0.356 0.472

    5 0.209 0.337 0.460 0.303 0.388 0.350 0.458

    6 0.215 0.339 0.466 0.297 0.372 0.351 0.466

    Mean 0.215 0.341 0.466 0.306 0.380 0.351 0.465

    SD 0.004 0.003 0.003 0.005. 0.007 0.003 0.006

    %RSD 1.9 0.9 0.6 1.6 1.8 0.9 1.3

    95%Confidence interval ± 0.004 ±0.003 ±0.003 ±0.005 ±0.007 ±0.003 ±0.006

    4.2.6.3 Limit of Detection (LOD) and Limit of Quantification

    (LOQ)

    LOD and LOQ values of each related substance were predicted

    from a separate linearity data at lower concentrations. Each predicted

    concentration was verified by preparing the solutions at about

    predicted concentration and injecting each solution six times into the

    HPLC.

  • 183

    4.2.6.4 Limit of Detection (LOD):

    The detection limit of an individual analytical procedure is the

    lowest amount of analyte is a sample, which can be detected but not

    necessarily quantitated as an exact value (Table: 4.8).

    Table: 4.8 LOD values of the Florfenicol and its impurities

    Injection

    ID

    Area

    Imp-A Imp-B Imp-C Imp-D Florfenicol Imp-E Imp-F Imp-G

    1 2820 4918 3541 4858 3144 2615 3920 3307

    2 1786 4611 3098 4579 2306 2608 3426 3170

    3 2544 4219 3404 4957 2659 2048 2889 2973

    4 3132 5132 2663 4304 2905 2376 3041 3656

    5 2091 4370 2614 3798 3474 1627 2785 3093

    6 2395 3388 2664 3624 3524 2858 2818 2485

    Mean 2461 4440 2997 4353 3002 2355 3147 3114

    SD 486 616 410 550 475 449 446 387

    % RSD 19.7 13.9 13.7 12.6 15.8 19.1 14.2 12.4

    Conc.

    (µg/mL) 0.113 0.090 0.086 0.084 0.072 0.082 0.077 0.075

    Conc. (%

    w/w) 0.011 0.018 0.014 0.019 0.013 0.011 0.013 0.018

    4.2.6.5 Limit of Quantification (LOQ):

    The quantitation limit of (LOQ) of an analytical procedure is the

    lowest amount of analyte in a sample, which can be quantitatively

    determined with suitable precision and accuracy. The quantitative

    limit is a parameter of quantitative assays for low levels of compounds

  • 184

    in sample matrices, and is used particularly for the determination of

    impurities and/ or degradation products (Table: 4.9).

    Table: 4.9 LOQ values of the Florfenicol and its impurities

    Injection ID

    Area

    Imp-A Imp-B Imp-C Imp-D Florfenicol Imp-E Imp-F Imp-G

    1 4619 8359 6337 8966 5818 6471 6691 6310

    2 4489 8470 5986 9014 6006 6230 6093 6536

    3 4166 8110 6212 9270 5961 5987 6560 6386

    4 4727 8450 6135 8775 6264 5786 6387 6472

    5 4175 8646 6031 9487 5948 6288 6558 6106

    6 4774 8691 6325 9254 6269 6133 6155 6861

    Mean 4492 8454 6171 9128 6044 6149 6407 6445

    SD 267 210 147 256 183 240 241 252

    % RSD 5.9 2.5 2.4 2.8 3.0 3.9 3.8 3.9

    Conc. (µg/mL)

    0.227 0.179 0.172 0.169 0.144 0.165 0.154 0.151

    Conc. (% w/w)

    0.044 0.034 0.028 0.039 0.027 0.030 0.027 0.037

  • 185

    4.2.6.6 Linearity

    The linearity of an analytical procedure is its ability to obtain

    test results, which are directly proportional to the concentration of

    analyte in the test sample. The linearity of the assay method was

    developed by injecting test sample at 80%, 90%, 100%, 110% and

    120% of Florfenicol assay concentration (i.e.100 µg/ml). Each solution

    injected twice (n=2) into HPLC and the average area at each

    concentration calculated (Table: 4.10).

    Calibration curve drawn by plotting average area on the Y-axis and

    concentration on the X-axis (Fig: 4.9).

  • 186

    Table: 4.10 Linearity results of the Assay method

    % Concentration Average area

    80 1706050

    90 1939894

    100 2166393

    110 2361512

    120 2602089

    Slope 21874

    Intercept -58290

    Residual Sum of Squares

    12208

    Correlation Coefficient 0.9995

    Linearity Plot (Concentration Vs Response

    Fig: 4.9 Linearity Plot for Assay method

    1706050

    1916050

    2126050

    2336050

    2546050

    80.96 90.86 100.76 110.66 120.56

    Are

    a

    Conc.(µg/mL)

  • 187

    Linearity of the related substance method were carried out by

    preparing the Florfenicol sample solutions containing Imp- A, B, C, D,

    E , F and G from LOQ to 250% (i.e. LOQ 25%, 50%,150%) with respect

    to their specifications limit (0.3%). Calibration curve was drawn by

    ploting average value of the impurities. (Imp- A, B, C, D, E , F and G)

    on the y-axis and concentrations on the X-axis (Fig: 4.10 to 4.17).

  • 188

    Linearity results of the Related Substance method

    Table: 4.11 Linearity results of the Imp-A

    Imp-A

    Concentration

    (µg/mL) Area Statistical Analysis

    0.227 4478 Slope 20120

    0.550 9575

    0.824 16077 Intercept -669

    1.374 27528

    1.649 33017 Residual sum of

    squares 898

    1.924 37772

    2.473 47188 Correlation

    Coefficient 0.9994

    2.748 55545

    3.298 65435 Response Factor 2.21

    Linearity Plot (Concentration Vs Area)

    Fig: 4.10 Linearity Plot for Imp-A

    4478

    19478

    34478

    49478

    64478

    79478

    0.227 1.127 2.027 2.927 3.827

    Are

    a

    Conc.(µg/mL)

  • 189

    Table: 4.12 Linearity results of the Imp-B

    Imp-B

    Concentration

    (µg/mL) Area Statistical Analysis

    0.180 8975 Slope 49241

    0.504 24310

    0.756 37725 Intercept -99

    1.260 62100

    1.512 73753 Residual Sum of

    Squares 415

    1.764 86477

    2.268 111431 Correlation

    Coefficient 0.9999

    2.520 124521

    3.024 148654 Response 0.90

    Linearity Plot (Concentration Vs Area)

    Fig: 4.11 Linearity Plot for Imp-B

    8975

    43975

    78975

    113975

    148975

    183975

    0.180 1.080 1.980 2.880 3.780

    Are

    a

    Conc.(µg/mL)

  • 190

    Table: 4.13 Linearity results of the Imp-C

    Imp-C

    Concentration

    (µg/mL) Area Statistical Analysis

    0.172 6691 Slope 43285

    0.500 21425

    0.749 33058 Intercept -88

    1.249 54391

    1.499 64653 Residual Sum of

    Squares 577

    1.749 75611

    2.248 97452 Correlation

    Coefficient 0.9999

    2.498 108358

    2.998 128583 Response 1.03

    Linearity Plot (Concentration Vs Area)

    Fig: 4.12 Linearity Plot for Imp-C

    6691

    36691

    66691

    96691

    126691

    156691

    0.172 1.022 1.872 2.722 3.572

    Are

    a

    Conc.(µg/mL)

  • 191

    Table: 4.14 Linearity results of the Imp-D

    Imp-D

    Concentration (µg/mL)

    Area Statistical Analysis

    0.166 9162 Slope 45771

    0.507 23428

    0.760 35326 Intercept 866

    1.266 59613

    1.520 70227 Residual Sum of Squares

    558 1.773 81942

    2.279 104834 Correlation Coefficient

    0.9999 2.533 116927

    3.039 139386 Response 0.97

    Linearity Plot (Concentration Vs Area)

    Fig: 4.13 Linearity Plot for Imp-D

    9162

    39162

    69162

    99162

    129162

    159162

    0.166 1.066 1.966 2.866 3.766

    Are

    a

    Conc.(µg/mL)

  • 192

    Table: 4.15 Linearity results of the Florfenicol

    Florfenicol

    Concentration

    (µg/mL) Area Statistical Analysis

    0.144 6253 Slope 44463

    0.528 23091

    0.792 36323 Intercept 81

    1.321 59019

    1.585 70100

    Residual sum of

    squares 713 1.849 82184

    2.377 105209

    2.641 118687

    Correlation

    coefficient 0.9999 3.170 140187

    3.962 176500

    Linearity Plot (Concentration Vs Area)

    Fig: 4.14 Linearity Plot for Florfenicol

    6253

    39253

    72253

    105253

    138253

    171253

    0.144 1.094 2.044 2.994 3.944

    Are

    a

    Conc.(µg/mL)

  • 193

    Table: 4.16 Linearity results of the Imp-E

    Imp-E

    Concentration (µg/mL)

    Area Statistical Analysis

    0.165 8391 Slope 40988

    0.512 20235

    0.768 31118 Intercept 889

    1.280 53878

    1.536 64690 Residual sum of

    squares 1194

    1.792 75261

    2.304 95501 Correlation

    Coefficient 0.9997

    2.560 107453

    3.072 125550 Response Factor 1.08

    3.840 157581

    Linearity Plot (Concentration Vs Response)

    Fig: 4.15 Linearity Plot of for Imp-E

    8391

    37391

    66391

    95391

    124391

    153391

    0.165 1.065 1.965 2.865 3.765

    Are

    a

    Conc.(µg/mL)

  • 194

    Table: 4.17 Linearity results of the Imp-F

    Imp-F

    Concentration (µg/ml)

    Area Statistical Analysis

    0.154 8835 Slope 47172

    0.493 23432

    0.739 34887 Intercept 411

    1.232 59155

    1.478 70031 Residual sum of squares

    1043 1.725 81219

    2.218 105102 Correlation Coefficient

    0.9998 2.464 115059

    2.957 139092 Response Factor 0.94

    3.696 176568

    Linearity Plot (Concentration Vs Response)

    Fig: 4.16 Linearity Plot of for Imp-F

    8835

    40835

    72835

    104835

    136835

    168835

    0.154 1.004 1.854 2.704 3.554

    Are

    a

    Conc.(µg/mL)

  • 195

    Table: 4.18 Linearity results of the Imp-G

    Imp-G

    Concentration

    (µg/mL) Area Statistical Analysis

    0.151 8863 Slope 34332

    0.504 16368

    0.756 25112 Intercept 1302

    1.261 44784

    1.513 54162 Residual Sum of

    Squares 1828

    1.765 62349

    2.269 79927 Correlation

    Coefficient 0.9990

    2.521 89814

    3.025 102709 Response Factor 1.30

    3.782 131349

    Linearity Plot (Concentration Vs Area)

    Fig: 4.17 Linearity Plot of for Imp-G

    5230

    14230

    23230

    32230

    41230

    50230

    59230

    0.200 0.550 0.900 1.250 1.600 1.950

    Are

    a

    Con. (µg/mL)

  • 196

    4.2.6.7 Accuracy/Recovery

    The accuracy of an analytical procedure expresses the closeness

    of agreement between the value, which is accepted either as a

    conventional true value or an accepted reference value and the value

    found.

    Accuracy of the assay method

    Accuracy of the assay method was established by injecting three

    preparations of test sample at 80%, 100% and 120% of analyte

    concentration (i.e.100 µg/ml). Each solution was injected twice (n=2)

    into HPLC and the mean peak area of Florfenicol peak was calculated.

    Assay (%w/w) of test solution was determined against three

    injections (n=3) of qualified Florfenicol reference standard (Table:

    4.19).

    The method showed consistent and high absolute recoveries at

    all three concentration (80%, 100% and 120%] levels with mean

    absolute recovery ranging from 99.3 % to 100.5%. The obtained

    absolute recoveries were normally distributed around the mean with

    uniform RSD values. The method was found to be accurate with low %

    bias (< 1.0).

  • 197

    Table: 4.19 Accuracy results of the Assay method

    S.NO %

    Concentration

    Mean recovery

    (%) (n=3) %RSD

    1 80 100.7 0.6

    2 100 100.2 0.3

    3 120 100.5 0.3

    4.2.6.8 Acuracy/Recovery of the Related substance method

    Acuracy of the related substance method established 50%,

    100%, 150% the impurities specification limit (0.30%).

    Accuracy at 50% impurity specification limit:

    Test solution prepared in triplicate (n=3) with impurities (Imp-A,

    B, C, D, E, F and G ) at 0.15% level w.r.s analyte concentration (i.e 0.5

    mg/m l). Each solution was injected thrice into HPLC. Mean %

    recovery of impurities calculated in the sample solution using the area

    of impurities standard at 0.3% level with respect to analyte (Table:

    4.20).

  • 198

    Table: 4.20 Accuracy at 50% level

    S.NO Impurity

    name Mean

    recovery(%) SD %RSD

    1 Imp-A 98.2 0.60 0.6

    2 Imp-B 97.8 0.40 0.4

    3 Imp-C 99.5 0.81 0.8

    4 Imp-D 104.6 1.18 1.1

    5 Imp-E 101.6 0.75 0.7

    6 Imp-F 97.3 1.18 1.2

    7 Imp-G 99.2 0.35 0.4

    Accuracy at 100% impurity specification limit:

    Test solution prepared in triplicate (n=3) with impurities (Imp-A,

    B, C, D, E, F and G) at 0.3 % level w.r.s analyte concentration (i.e 0.5

    mg/m l). Each solution was injected thrice into HPLC (Table: 4.21).

    Table: 4.21 Accuracy at 100% level

    S.NO Impurity

    name

    Mean

    recovery(%) SD %RSD

    1 Imp-A 106.3 0.70 0.7

    2 Imp-B 95.0 0.20 0.2

    3 Imp-C 96.1 0.42 0.4

    4 Imp-D 99.3 0.58 0.6

    5 Imp-E 96.8 0.36 0.4

    6 Imp-F 93.2 0.50 0.5

    7 Imp-G 103.3 1.65 1.6

    Accuracy at 150% impurity specification limit:

    Test solution in triplicate (n=3) with impurities (Imp-A, B, C, D,

    E, F and G) at 0.9 % level w.r.s analyte concentration (i.e 0.5 mg/m l).

    Each solution was injected thrice into HPLC (Table: 4.22).

  • 199

    Table: 4.22 Accuracy at 150 % level

    S.NO Impurity

    name

    Mean

    recovery(%) SD %RSD

    1 Imp-A 99.9 0.42 0.4

    2 Imp-B 95.3 0.21 0.2

    3 Imp-C 95.7 0.10 0.1

    4 Imp-D 98.0 0.06 0.1

    5 Imp-E 98.2 0.89 0.9

    6 Imp-F 94.4 1.03 1.1

    7 Imp-G 96.3 1.45 1.5

    The related substance method showed consistent and high obsolute

    recoveries of all six impurities at all three different concentrations (50,

    100, 150%) levels in drug substance.

    4.2.6.9 Solution state stability

    The solution state stability of Florfenicol in diluent in the assay

    method was carried out by leaving both the test solutions of sample

    and reference standard in tightly capped volumetric flasks kept at

    room temperature for two days. The same sample solutions were

    assayed for every one hour interval up to the study period. The % RSD

    of assay of Florfenicol during solution stability experiments was with

    in 1.0%.

    The solution state stability of Florfenicol related substance

    method was carried out by leaving sample solution in tightly capped

    volumetric flask at room temperature for two days. Content of Imp A,

    B, C, D, E, F and G were checked for every six hours internal up to

    the study period. No significant change was observed in the content

  • 200

    of all six impurities in drug solution stability experiments up to the

    study period. Hence Florfenicol sample solutions are stable for atleast

    48 hours in the developed method. In assay method standard and test

    solutions injected at each 0h, 1h, 2h, 3h, 4h, 5h, 6h, 7h, 8h, 9h, 10h,

    11h, 12h. (Table: 4.23).

    Table: 4.23 Solution stability results of the Assay method

    S.No Time in Hours Assay (% w/w)

    1 initial 99.9

    2 1 99.8

    3 2 99.2

    4 3 99.6

    5 4 99.5

    6 5 99.6

    7 6 99.7

    8 7 99.2

    9 8 99.3

    10 9 99.2

    11 10 99.4

    12 11 99.6

    13 12 99.5

    % RSD 0.24

    In related substances method solution stability studies of

    Florfenicol in diluent was done for 15 hrs by injecting sample solution

    for every one hour interval up to the study period. The impurity

    profiles obtained at different interval were very consistent and

    matched with initial value.

  • 201

    4.2.7.0 Robustness

    The robustness of an analytical procedure is a measure of its

    capacity to remain unaffected by small, but deliberate variations in

    method parameters and provides an indication of its reliability during

    normal usage. To determine the robustness of the developed method

    experimental conditions were purposely altered and the resolution

    between Imp-C and Imp-D was evaluated. In each of the deliberately

    altered chromatographic condition (flow rate 1.3 ml/min and

    1.7ml/min, acetonitrile 23% and 27% in the mobile phase, column

    temperature 25 C and 35C) the resolution between Imp-B, Imp-C

    and Imp-D, Imp-E and Imp-F was greater than 2.0,, illustrating the

    robustness of the method.

    4.3 Mass balance

    The mass balance it is a process of adding together the assay

    value and the levels of degradation products to see how closely these

    add up to 100% of the initial value, with due consideration of the

    margin of analytical error [10]. Its establishment hence is a regulatory

    requirement. The mass balance is very closely linked to the

    development of stability-indicating assay method as it acts as an

    approach to establish its validity. The stressed samples of Florfenicol

    bulk drug were assayed against the qualified reference standard and

    the results of mass balance obtained in each condition is presented

    below (Table: 4.24).

  • 202

    Table: 4.24 Mass balance of the assay method

    Degradation Mechanism

    Degradation Condition

    % Assay of active

    substance)

    Mass balance (% Assay+ % impurities+ % degradants)

    Remarks

    Acid 5M HCl/80°C

    /60 min 91.5 99.5

    Degraded to Imp-A and some unknown

    degradants observed

    Base 5M NaOH/

    Initial 90.6 99.6

    Degraded to Imp-A, Imp-C and some

    unknown degradants observed

    Peroxide 30% H2O2/

    80°C/120 min 96.1 99.5

    Some unknown degradants observed

    Thermal 105°C/188

    Hours 99.5 99.6 No degradation observed

    Photolytic 12,000 Lux/ 188 Hours

    99.5 99.6 No degradation observed

    4.4 Analysis of Florfenicol drug substance stability samples

    One manufacturing lot of Florfenicol drug substance was placed

    on stability study in chambers maintained at ICH set conditions [11].

    The analysis of stability samples were carried up to 12 months period

    using the above optimized method. The stability data results obtained

    are presented in Table: 4.25 to 4.26. The developed HPLC method

    performed satisfactorily for the quantitative evaluation of stability

    samples.

  • 203

    Table: 4.25 Accelerated stability data (Storage conditions:

    40°C/75%RH)

    Batch No: KMR(510)128 Packing & storage conditions: Each sample

    packed in a polyethylene bag in a triple laminated bag and kept in a HDPE drum.

    Stability study duration: 6 months Temperature

    %Relative humidity

    40°C/75%RH

    Description

    Loss on

    drying (%w/w)

    Identification

    Assay (By HPLC,

    %w/w, on

    dried basis,)

    Specifications

    A white to almost

    white

    crystalline powder

    NMT 0.5

    IR spectrum should

    concordant

    with that of standard

    NLT 98.0 and NMT 102.0

    Initial

    A white

    crystalline

    powder

    0.15 Complies 99.5

    1M A white

    crystalline

    powder

    0.14 Complies 99.5

    2M

    A white

    crystalline powder

    0.13 Complies 99.2

    3M

    A white

    crystalline

    powder

    0.15 Complies 99.4

    6M A white

    crystalline

    powder

    0.14 Complies 99.3

    Related substances details on next page.

  • 204

    Related

    Substances

    LOQ

    (%w/w)

    LOD

    (%w/w)

    Related Substances (By HPLC, %w/w)

    INITIAL 1M 2M 3M 6M

    Imp-A 0.044 0.011 ND ND ND ND ND

    Imp-B 0.034 0.018 ND ND Below

    LOQ

    Below

    LOQ

    Below

    LOQ

    Imp-C 0.028 0.014 0.08 0.09 0.09 0.09 0.08

    Imp-D 0.030 0.011 ND ND ND ND ND

    Imp-E 0.027 0.013 Below

    LOQ

    Below

    LOQ

    Below

    LOQ

    Below

    LOQ

    Below

    LOQ

    Imp-F 0.032 0.014 ND ND ND ND ND

    Imp-G 0.037 0.018 ND ND ND ND ND

    Highest

    unknown - - ND ND ND ND ND

    Total

    unknown - - NA NA NA NA NA

    Total RS - - 0.08 0.09 0.09 0.09 0.08

    ND: Not detected

    NA: Not applicable

  • 205

    Table: 4.26 Long-term stability data (Storage conditions:

    25°C/60%RH)

    Batch No: KMR(510)128 Packing & storage conditions: Each sample

    packed in a polyethylene bag in a triple laminated bag and kept in a HDPE drum

    Stability study duration: 12 months Temperature

    %Relative humidity

    25°C/60%RH

    Tests Description

    Loss on

    drying (%w/w)

    Identification

    Assay (By HPLC,

    %w/w, on

    dried basis,)

    Specifications

    A white to almost

    white

    crystalline powder

    NMT 0.5

    IR spectrum should

    concordant

    with that of standard

    NLT 98.0 and NMT 102.0

    Initial

    A white

    crystalline

    powder

    0.15 Complies 99.5

    1M A white

    crystalline

    powder

    0.14 Complies 99.6

    2M

    A white

    crystalline powder

    0.13 Complies 99.4

    3M

    A white

    crystalline

    powder

    0.14 Complies 99.5

    6M A white

    crystalline

    powder

    0.13 Complies 99.2

    9M

    A white

    crystalline powder

    0.13 Complies 99.3

    12M

    A white

    crystalline

    powder

    0.12 Complies 99.2

    Related substances details on next page.

  • 206

    Related Substances

    LOQ (%w/w)

    LOD

    (%w/w)

    Related Substances (By HPLC, %w/w)

    Initial 1M 2M 3M 6M 9M 12M

    Imp-A 0.044 0.011 ND ND ND ND ND ND ND

    Imp-B 0.034 0.018 ND Below LOQ

    ND Below LOQ

    Below LOQ

    ND Below LOQ

    Imp-C 0.028 0.014 0.07 0.08 0.07 0.06 0.06 0.07 0.08

    Imp-D 0.030 0.011 ND ND ND ND ND ND ND

    Imp-E 0.027 0.013 Below LOQ

    Below LOQ

    Below LOQ

    Below LOQ

    Below LOQ

    Below LOQ

    Below LOQ

    Imp-F 0.032 0.014 ND ND ND ND ND ND ND

    Imp-G - - ND ND ND ND ND ND ND

    Highest unknown

    - - ND ND ND ND ND ND ND

    Total unknown

    - - NA NA NA NA NA NA NA

    Total RS - - 0.08 0.09 0.08 0.08 0.08 0.08 0.09

    ND: Not detected NA: Not applicable

  • 207

    4.5 Summary and conclusions

    Validated stability-indicating HPLC method was developed for

    Florfenicol after subjecting the samples to stress testing under ICH

    recommendes conditions. The RPLC method was developed for

    quantitative and related substance determination of Florfenicol is

    rapid precise, accurate and selective. The method was completely

    validated showing satisfactory data for all the method validation

    parameters tested. The developed method was found to be ‘specific’ to

    the drug, as the peaks of the degradation products did not interfere

    with the degradation peak. Thus the proposed method can be

    employed for assessing the stability of Florfenicol bulk drug samples.

  • 208

    Table: 4.27 Summary of Analytical method validation data

    Test Parameter

    Related Substances method Assay

    method

    Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F Imp-G

    Precision (RSD) 1.8 0.6 0.6 0.3 1.0 0.6 0.9 0.7

    LOD (µg/ml) 0.113 0.090 0.086 0.084 0.082 0.077 0.075 N/A

    LOQ (µg/ml) 0.044 0.034 0.028 0.039 0.030 0.027 0.037

    N/A

    Linearity (corre coefficient

    0.9994 0.9999 0.9999 0.9999 0.9997 0.9998 0.9990 0.9995

    Accuracy (%) 98.2-106.3 95.0-97.8 95.7-99.5 98.0-104.6 98.2-101.6 93.2-97.3 96.3-103.3 100.2-100.7

    Robustness

    Resolution b/w Imp-E& Imp-

    F>2

    Resolution b/w

    Imp-E& Imp-F>2

    Resolution b/w

    Imp-E& Imp-F>2

    Resolution b/w Imp-E& Imp-

    F>2

    Resolution b/w Imp-E& Imp-

    F>2

    Resolution b/w

    Imp-E& Imp-F>2

    Resolution b/w Imp-E&

    Imp-F>2

    Resolution b/w Imp-E&

    Imp-F>2

    Solution stability Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Mobile phase stability Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

    Stable up to 15hr

  • 209

    4.6 References:

    1. Sams, R. A.; in proceedings of the XVIII World Buiatrics

    Congress, 1994, Bolonga, 13.

    2. Piotr, K.; Lucyna, K.; Aleksandra, C.; Iiona, O.; Alina, P.; Michal

    B,; Henry, L.; J. Pharm. Biomed. Anal., 2005, 39, 983.

    3. Suxia, Z.; Zhongwei, L.; Xia. G.; Linli. C.; Zhanhui. W.;

    Jianzhong. S.; J.Chromatogr. B, 2008, 875, 399.

    4. Chue. V.; Larry, S. J.; Guy. S. R.; William, G. H.; J.Chromatogr.

    B, 2002, 780, 111.

    5. Hayes, J. M.; Eichman, J.; Katz, T.; Gilewicz, R.; J. AOAC,

    2003, 86, 22.

    6. Marciniec, B.; Stawny, M.; Kachlicki, P.; Jaroszkiewicz, E.;

    Michael, N.; Analytical Sciences. 2009, 25, 1255.

    7. De Craene, B. A.; Deprez, P.; Haese, E. D.; Nelis, H. J.;

    Bossche, W.; De Leenheer, A. P.; Antimicrobial Agents and

    Chemotherapy., 1997, 41, 1991.

    8. Roy, Y. P. E.; Eric, C. W.; J. Aquatic Animal Health. 2005, 17,

    129.

    9. ICH, Validation of analytical Procedures: Text and methodology

    Q2 (R1), International Conferences on Hormonization, IFPMA,

    Geneva, 2005.

    10. Steven, B. W.; Pharmaceutical Stress Testing Predicting

    Drug Degradation.

    11. Validation of Analytical Procedures: Methodology Q2B – ICH

    Guidelines.