CHAPTER-4 A VALIDATED STABILITY-INDICATING...
Transcript of CHAPTER-4 A VALIDATED STABILITY-INDICATING...
-
158
CHAPTER-4
A VALIDATED STABILITY-INDICATING
ANALYTICAL METHOD FOR THE
DETERMINATION OF IMPURITIES IN
FLORFENICOL
-
159
4.1 Introduction of Florfenicol and survey of analytical Methods
Florfenicol is a synthetic derivative of thiamphenicol. It has a
wide spectrum of antibacterial properities, including efficacy against
Gram-positive and Gram-negative bacteria [1]. The mechanism of
action of Florfenicol similarly to chloramphenicol and thiamphenicol,
relies on slowing down the biosynthesis of bacterial proteins through
reversible binding with an active center of peptydylotransferase on a
subunit 50S of bacterial ribosomes. It is chemically designated as 2,2-
dichloro-N-[(1S,2R)-1-(fluoromethyl)-2-hydroxy-2-[4-(methylsulfonyl)
phenyl]ethyl] acetamide. Florfenicol is a white to almost white,
crystalline powder. It is very soluble in N,N-Dimethylformamide and
soluble in methanol. The molecular formula is C12H14Cl2FNO4S. The
molecular weight of Florfenicol is 358.22.
Fig: 4.1 Chemical structure of Florfenicol
H
Cl
S FO
N
OH Cl
O
H3C
O
2,2-dichloro-N-[(1S,2R)-1-(fluoromethyl)-2-hydroxy-2-[4-
(methylsulfonyl)phenyl]ethyl]acetamide
Molecular formula C12H14Cl2FNO4S
Molecular weight 358.22
-
160
The different analytical techniques reported so far for the
determination of this drug and its metabolites in biological samples
include capillary electrophoresis [2] and spectrometry [3]. The
determination of Florfenicol in plasma by RP-LC [4], stability of
Florfenicol [5-6], pharmacokinetics studies of Florfenicol in plasma [7-
8] was also reported.
Organic impurities can arise during the manufacturing process
and storage of the drug substances and the criteria for their
acceptance up to certain limits are based on pharmaceutical studies
or known safety data [9]. As per regulatory guidelines, the
pharmaceutical studies using a sample of the isolated impurities can
be considered for safety assessment. It is, therefore, essential to
isolate and characterize unidentified impurities present in the drug
sample. Recently we have developed a process for the synthesis of
Florfenicol in our laboratory. During the development of an analytical
procedure, the LC method was developed for the determination of in-
house synthesized Florfenicol and the impurities arising during its
manufacturing. In the present study, we describe a reverse phase
column liquid chromatography method for the separation and
quantification of process related and degradation impurities of
Florfenicol. The accuracy, precision, limit of detection (LOD), limit of
quantification (LOQ) and robustness of the method were determined in
accordance with ICH guidelines [10]. The target is to develop a suitable
stability-indicating HPLC related substances method for Florfenicol in
-
161
this chapter we describe a stability-indicating LC method for the
determination of Florfenicol and its potential and degradation
impurities and also the method validation.
4.2 Development of a stability-indicating analytical method for
Florfenicol
4.2.1 Materials
Reference standard of Florfenicol and seven impurities namely,
Imp-A, Imp-B, Imp-C, Imp-D, Imp-E , Imp-F and Imp-G (Fig: 4.1)
were synthesized and characterized by use of LC-MS, NMR and IR in
Aurobindo Pharma Ltd., Hyderabad, India. The commercial samples of
Florfenicol are also manufactured by Aurobindo Pharma Ltd. All
reagents used were of analytical reagent grade unless stated
otherwise. Milli Q water, HPLC-grade acetonitrile, HPLC-grade
orthophosphoric acid (OPA) were purchased from Merck (Darmstadt,
Germany).
4.2.2 Equipment
The LC system was equipped with quaternary gradient pumps
with autosampler and auto injector (Alliance, Waters 2695, Milliford,
MA, USA) controlled with Empower software (Waters).
-
162
Fig: 4.2 Chemical structures of impurities of Florfenicol
Me O2S F
NH2
OH
(1R,2S)-1-[4-(Methylsulfonyl) phenyl]-2-amino-3-fluoro-1-propanol
(Imp-A)
Fig 4.2 (a)
FMe O2S
N
OH
CH3
O
H
(1R,2S)-2-Acetamido-3-fluoro-1-[4-(methyl-sulfonyl)phenyl]-1-
propanol (Imp-B)
Fig: 4.2 (b)
Cl
Me O2S OHO
N
OH ClH
(1R,2R)-2-Dichloroacetamido-1-[4-(methylsulfonyl)phenyl]-1,3-
propanediol (Imp-C)
Fig: 4.2 (c)
Cl
Me O2S FO
N
OHH
(1R,2S)-2-Chloroacetamido-3-fluoro-1-[4-(methylsulfonyl)phenyl]-1-
propanol (Imp-D)
Fig: 4.2 (d)
-
163
Cl
Me O2S ClO
N
OH ClH
(1R,2S)-2-Dichloroacetamido-3-fluoro-1-[4-(methylsulfonyl)phenyl]-1-
propanol (Imp-E)
Fig: 4.2 (e)
Cl
MeO2S FO
N
OH ClCH3
(1R,2S)-2-(N-Methyl)dichloroacetamido-3-fluoro-1-[(4-methylsulfonyl)-
phenyl]-1-propanol (Imp-F)
Fig: 4.2 (f)
Cl
Me O2S FO
N
OH ClClH
(1R,2S)-2-Trichloroacetamido-3-fluoro-1-[4-(methylsulfonyl)-phenyl]-1-
propanol
(Imp-G)
Fig: 4.2 (g)
4.2.3 Preparation of sample and stock solutions
The stock solutions of Florfenicol (0.5 mg/ml) and spiked with
0.3% of Imp-A of Imp-B, Imp-D, Imp-E, Imp-F and Imp-G with respect
to the Florfenicol analyte concentration. The stock solutions was
further diluted with diluent to obtain a standard solution of 0.0005
-
164
mg/ml (0.5 µg/ml) for related substances determination and 0.5
mg/ml ( 500 µg/ml) for assay determination.
4.2.4 Generation of stress samples
One lot of Florfenicol drug substance selected for stress
testing. From the ICH stability guideline: “stress testing likely to be
carried out on a single batch of material. Different kinds of stress
conditions (i.e., acid hydrolysis, base hydrolysis, oxidative stress, heat
and humidity) were employed on one lot of Florfenicol drug substance
based on the guidance available from ICH stability guideline (Q1AR2).
The details of the stress conditions are as follows:
a) Acid Degradation: drug in 5.0 M HCl solution was kept at 85°C for
60 mins.
b) Base Degradation: drug in 5.0 M NaOH solution was kept at room
temperature.
c) Oxidative stress: drug in 30% H2O2 solution was kept at 80°C for
120 mins.
d) Thermal Degradation: drug was subjected to dry heat at 105°C for
188 hrs.
e) Phtolytic degradation: drug was subjected to UV at 254 nm (10 K
Lux ) for 188 hrs.
4.2.5 Optimization of chromatographic conditions
The main objective of the chromatographic method was to
seperate Florfenicol from Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F
and Imp-G impurities were coeluted using different stationary phases
-
165
such as C8, phenyl and cyano as well as different mobile phases
containing buffers like phosphate, sulfate and acetate with different
pH and using organic modifiers like acetonitrile and methanol in the
mobile phase. Apart from the co-elution of impurities, we have also
observed poor peak shapes for Florfenicol, some impurities and
degradants. The chromatographic separation was achieved on a
Hypersil BDS, C18, (250 mm x 4.6 mm), 5 particle size. The gradient
LC method employs solution A and B as mobile phase. The solution A
contains phosphate buffer pH 7.0 and acetonitrile as solution B. The
flow rate of the mobile phase was 1.0 ml/min. The HPLC gradient
program was set as: time% solution B: 0.01/10, 15/20, 30/30,
40/40, 50/60, 60/60, 61/10, 70/10 with a post run time of 10 min.
The detection was monitored at a wavelength of 225 nm. The injection
volume was 20 µl. Standard and test solutions were prepared in pH
3.0 buffer was used as diluent. In the optimized chromatographic
conditions of Florfenicol, Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F
and Imp-G were separated with a resolution greater than 3, typical
relative retention times were approximately 0.29, 0.35, 0.59, 0.61,
1.35, 1.36, 1.72 with respect to Florfenicol eluted at 22.45 min.
-
166
Optimized liquid chromatographic conditions:
Column : Hypersil BDS, C18, (250 mm x
4.6 mm), 5
Mobile phase : The solution A contains
phosphate buffer pH 7.0 and
solution B contains acetonitrile
Pump mode : Gradient
Flow rate : 1.0 ml/min
UV of detection : 225 nm
Injection volume : 20 µl
Run time : 60 min
Retention time : 22.45min
Relative Retention Time (RRT) : Imp-A about 0.29
Imp-B about 0.35
Imp-C about 0.59
Imp-D about 0.61
Imp-E about 1.35
Imp-F about 1.36
Imp-G about 1.72
Diluent : Dissolve 1.36 g of potassium dihydrogen orthophosphate in
1000 ml of water. Adjust pH to 3.0±0.05 with orthophosphoric acid.
No considerable degradation was observed in Florfenicol bulk
samples under stress conditions such as acid hydrolysis, photolytic
and thermal. The degradation of drug substance was observed during
base hydrolysis and oxidative stress condition. Florfenicol was
degraded to Imp-A (7.85%) under acidic conditions (5M HCl/180°C/60
-
167
min). Florfenicol was degraded to Imp-A (7.80%), Imp-C (1.14%) under
base conditions (5M NaOH/Initial) and it was confirmed by co-
injection with a qualified Imp-A and Imp-C standards. Mild
degradation was observed under oxidative environment (treated with
30% H2O2/85°C/240 min) leads to the formation of some unknown
degradation peaks (3.5%).
Peak purity test results obtained by using a PDA detector
confirmed that the Florfenicol peak is homogenous and pure in all the
analyzed stress samples. The mass balance of Florfenicol in all stress
samples was close to 99.6% (%Assay + %Degradation). This clearly
demonstrates that the developed HPLC method was found to be
specific for Florfenicol in presence of its impurities (Imp-A, Imp-B,
Imp-C, Imp-D, Imp-E, Imp-F and Imp-G) and degradation products.
Figures:
Fig: 4.3 to Fig: 4.7 is the typical HPLC chromatograms showing
the degradation of Florfenicol in various stress conditions and also the
corresponding peak purityplots.
-
168
Fig: 4.3 Typical HPLC chromatograms of Acid hydrolysis
Fig: 4.3 (a)
Fig: 4.3 (b)
Blank Chromatogram of Acid hydrolysis (5N HCl)
Florfenicol stressed with 5N HCl at 85°C for 60 mins
-
169
Fig: 4.3 (c) Peak purity plot of Acid hydrolysis
Purity Angle Purity Threshold Purity Flag Peak Purity
0.048
0.249 No Pass
Fig: 4.3 (c)
-
170
Fig: 4.4 Typical HPLC chromatograms of Base hydrolysis
Fig: 4.4 (a)
Fig: 4.4 (b)
Blank Chromatogram of Base hydrolysis ( 5N NaOH )
Florfenicol stressed with 5N NaOH at room temperature
-
171
Fig: 4.4 (c) Peak purity plot of Base hydrolysis
Purity Angle Purity Threshold Purity Flag Peak Purity
0.047
0.247 No Pass
Fig: 4.4 (c)
Fig: 4.5 Typical HPLC chromatograms of Peroxide Degradation
Fig: 4.5 (a)
Fig: 4.5 (b)
Blank Chromatogram of Peroxide Degradation ( 30% H2O2 )
Florfenicol stressed with 30%H2O2 at 80°C for 120 mins
-
172
Fig: 4.5 (c) Peak purity plot of Peroxide Degradation
Purity Angle Purity Threshold Purity Flag Peak Purity
0.067
0.272 No Pass
Fig: 4.5 (c)
Fig: 4.6 Typical HPLC chromatograms of Thermal Degradation
Fig: 4.6 (a)
Blank
-
173
Fig: 4.6 (b)
Fig: 4.6 (c) Peak purity plot of Thermal Degradation
Purity Angle Purity Threshold Purity Flag Peak Purity
0.063
0.261 No Pass
Fig: 4.6 (c)
Florfenicol stressed at 105°C for 188 hours
-
174
Fig: 4.7 Typical HPLC chromatograms of Photolytic Degradation
Fig: 4.7 (a)
Fig: 4.7 (b)
Blank
Florfenicol stressed with 12,000 Lux for 188 hours
-
175
Fig: 4.7(c): Peak purity of Photolytic Degradation
Purity Angle Purity Threshold Purity Flag Peak Purity
0.061
0.259 No Pass
Fig: 4.7 (c)
Peak purity test performed for the Florfenicol peak using photo
diode array (PDA) detector data confirmed the purity of the peak for
all the stressed samples. This clearly demonstrates that the developed
HPLC method was found to be specific for Florfenicol in presence of
its impurities (Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F, Imp-G
and degradant products.
No degradants were observed after 15 min in the extended run
time of 60 min for all the Florfenicol stressed samples (acid
hydrolysis, base hydrolysis, oxidation stress, heat 120°C, photolysis)
with 90% can in mobile phase.
-
176
4.2.6 Validation of analytical method and its results:
The developed and optimized HPLC method was taken up to
validation. The analytical method validation was carried out in
accordance with ICH guideline [11].
4.2.6.1 System suitability: A mixture of Florfenicol standard
injections were injected into HPLC system and good resolution was
obtained between the impurities and Florfenicol [Fig: 4.8). The system
suitability results are tabulated (Table: 4.1).
-
177
Fig: 4.8 Typical Blank, Florfenicol sample and SST
chromatograms
Fig: 4.8 (a)
Fig: 4.8 (b)
Fig: 4.8 (c)
Blank
Florfenicol sample spiked with impurities
Florfenicol Sample
-
178
Table: 4.1 System suitability results
S.NO Impurity
name USP Plate
count USP
Tailing USP
Resolution
1 Imp-A 4179 1.23 -
2 Imp-B 10231 1.10 3.93
3 Imp-C 29350 1.12 16.68
4 Imp-D 32575 1.13 1.92
5 Florfenicol 56327 1.18 24.53
6 Imp-E 80882 1.10 18.89
7 Imp-F 88309 1.15 0.85
8 Imp-G 173154 1.11 22.69
4.2.6.2 Precision:
The precision of an analytical process experiment the closeness
of agreement between a series of measurements obtained from
multiple sampling of the some homogeneous same under prescribed
conditions.
Precision may be considered at three levels: System precision,
Method precision and Intermediate Precision. Assay method precision
study was evaluated by carrying out six independent assays of
Florfenicol test sample against qualified reference standard and RSD
of six consecutive assays was 0.6% (Table: 4.2 to Table: 4.4).
The results showed insignificant variation in measured
response. Which demonstrated that the assay method was repeatable
with RSD’s below 0.4%.
-
179
Table: 4.2 System Precision results of the Assay method
Injection ID Florfenicol
1 2194243
2 2174916
3 2189161
4 2184508
5 2192197
6 2191167
Mean 2187699
SD 7082
% RSD
95% Confidence
Interval
0.3
± 7433
Table: 4.3 Method Precision results of the Assay method
Sample ID Assay (% w/w)
1 98.8
2 98.8
3 99.9
4 98.8
5 100.2
6 99.9
Mean 99.4
SD 0.67
% RSD 0.7
95% Confidence Interval
± 0.7
-
180
Table: 4.4 Intermediate Precision results of the Assay method
Sample ID Assay (% w/w)
1 99.1
2 100.3
3 99.4
4 100.7
5 100.5
6 100.6
Mean 100.1
SD 0.68
% RSD 0.7
95% Confidence
Interval ± 0.7
The precision of the related substance method was checked by
injecting six individual preparations of Florfenicol (0.5 mg/ml) spiked
with 0.3% of Imp-A, Imp-B, Imp-C, Imp-D, Imp-E, Imp-F and Imp-G
with respect to the Florfenicol analyte concentration. The % RSD of
the area percentage of each impurity (impurities-A, -B, -C, -D, -E, -F
and –G) for six consecutive determinations was respectively as below
(Table: 4.5 to Table: 4.7).
The results showed insignificant variation in measured
response. Which indicated that the related substance method was
repeatable with RSD’s below 1.8%.
-
181
Table: 4.5 System Precision results of the Related Substance
method
Injection ID Florfenicol
1 64644
2 64842
3 64434
4 64178
5 64439
6 64029
Mean 44290
SD 44423
% RSD 130
95% Confidence
Interval 0.3
Table: 4.6 Method Precision results of the Related Substance
method
Preparation
Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F Imp-G
1 0.217 0.334 0.461 0.311 0.394 0.341 0.471
2 0.218 0.332 0.462 0.309 0.386 0.344 0.463
3 0.225 0.333 0.464 0.309 0.391 0.347 0.465
4 0.226 0.334 0.469 0.311 0.397 0.343 0.474
5 0.222 0.329 0.463 0.309 0.391 0.346 0.464
6 0.224 0.330 0.466 0.312 0.392 0.346 0.467
Mean 0.222 0.332 0.464 0.310 0.392 0.345 0.467
SD 0.004 0.002 0.003 0.001 0.004 0.002 0.004
%RSD 1.8 0.6 0.6 0.3 1.0 0.6 0.9
95%Confidence
interval ± 0.004 ±0.002 ±0.003 ±0.001 ±0.004 ±0.002 ±0.004
-
182
In intermediate precision method six preparations individually
using single batch of Florfenicol drug substance spiked with related
substances at specification level (0.3%) as per test method and
injected each solution as per methodology using different column,
system and by another analyst. Results showed insignificant variation
in measured response. Which demonstrated that the related
substance method was repeatable with RSD’s below 1.9%.
Table: 4.7 Intermediate Precision results of the Related substance
method
Preparation Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F Imp-G
1 0.217 0.341 0.469 0.309 0.373 0.347 0.466
2 0.211 0.341 0.466 0.305 0.378 0.349 0.456
3 0.215 0.346 0.468 0.310 0.377 0.354 0.470
4 0.217 0.344 0.468 0.309 0.389 0.356 0.472
5 0.209 0.337 0.460 0.303 0.388 0.350 0.458
6 0.215 0.339 0.466 0.297 0.372 0.351 0.466
Mean 0.215 0.341 0.466 0.306 0.380 0.351 0.465
SD 0.004 0.003 0.003 0.005. 0.007 0.003 0.006
%RSD 1.9 0.9 0.6 1.6 1.8 0.9 1.3
95%Confidence interval ± 0.004 ±0.003 ±0.003 ±0.005 ±0.007 ±0.003 ±0.006
4.2.6.3 Limit of Detection (LOD) and Limit of Quantification
(LOQ)
LOD and LOQ values of each related substance were predicted
from a separate linearity data at lower concentrations. Each predicted
concentration was verified by preparing the solutions at about
predicted concentration and injecting each solution six times into the
HPLC.
-
183
4.2.6.4 Limit of Detection (LOD):
The detection limit of an individual analytical procedure is the
lowest amount of analyte is a sample, which can be detected but not
necessarily quantitated as an exact value (Table: 4.8).
Table: 4.8 LOD values of the Florfenicol and its impurities
Injection
ID
Area
Imp-A Imp-B Imp-C Imp-D Florfenicol Imp-E Imp-F Imp-G
1 2820 4918 3541 4858 3144 2615 3920 3307
2 1786 4611 3098 4579 2306 2608 3426 3170
3 2544 4219 3404 4957 2659 2048 2889 2973
4 3132 5132 2663 4304 2905 2376 3041 3656
5 2091 4370 2614 3798 3474 1627 2785 3093
6 2395 3388 2664 3624 3524 2858 2818 2485
Mean 2461 4440 2997 4353 3002 2355 3147 3114
SD 486 616 410 550 475 449 446 387
% RSD 19.7 13.9 13.7 12.6 15.8 19.1 14.2 12.4
Conc.
(µg/mL) 0.113 0.090 0.086 0.084 0.072 0.082 0.077 0.075
Conc. (%
w/w) 0.011 0.018 0.014 0.019 0.013 0.011 0.013 0.018
4.2.6.5 Limit of Quantification (LOQ):
The quantitation limit of (LOQ) of an analytical procedure is the
lowest amount of analyte in a sample, which can be quantitatively
determined with suitable precision and accuracy. The quantitative
limit is a parameter of quantitative assays for low levels of compounds
-
184
in sample matrices, and is used particularly for the determination of
impurities and/ or degradation products (Table: 4.9).
Table: 4.9 LOQ values of the Florfenicol and its impurities
Injection ID
Area
Imp-A Imp-B Imp-C Imp-D Florfenicol Imp-E Imp-F Imp-G
1 4619 8359 6337 8966 5818 6471 6691 6310
2 4489 8470 5986 9014 6006 6230 6093 6536
3 4166 8110 6212 9270 5961 5987 6560 6386
4 4727 8450 6135 8775 6264 5786 6387 6472
5 4175 8646 6031 9487 5948 6288 6558 6106
6 4774 8691 6325 9254 6269 6133 6155 6861
Mean 4492 8454 6171 9128 6044 6149 6407 6445
SD 267 210 147 256 183 240 241 252
% RSD 5.9 2.5 2.4 2.8 3.0 3.9 3.8 3.9
Conc. (µg/mL)
0.227 0.179 0.172 0.169 0.144 0.165 0.154 0.151
Conc. (% w/w)
0.044 0.034 0.028 0.039 0.027 0.030 0.027 0.037
-
185
4.2.6.6 Linearity
The linearity of an analytical procedure is its ability to obtain
test results, which are directly proportional to the concentration of
analyte in the test sample. The linearity of the assay method was
developed by injecting test sample at 80%, 90%, 100%, 110% and
120% of Florfenicol assay concentration (i.e.100 µg/ml). Each solution
injected twice (n=2) into HPLC and the average area at each
concentration calculated (Table: 4.10).
Calibration curve drawn by plotting average area on the Y-axis and
concentration on the X-axis (Fig: 4.9).
-
186
Table: 4.10 Linearity results of the Assay method
% Concentration Average area
80 1706050
90 1939894
100 2166393
110 2361512
120 2602089
Slope 21874
Intercept -58290
Residual Sum of Squares
12208
Correlation Coefficient 0.9995
Linearity Plot (Concentration Vs Response
Fig: 4.9 Linearity Plot for Assay method
1706050
1916050
2126050
2336050
2546050
80.96 90.86 100.76 110.66 120.56
Are
a
Conc.(µg/mL)
-
187
Linearity of the related substance method were carried out by
preparing the Florfenicol sample solutions containing Imp- A, B, C, D,
E , F and G from LOQ to 250% (i.e. LOQ 25%, 50%,150%) with respect
to their specifications limit (0.3%). Calibration curve was drawn by
ploting average value of the impurities. (Imp- A, B, C, D, E , F and G)
on the y-axis and concentrations on the X-axis (Fig: 4.10 to 4.17).
-
188
Linearity results of the Related Substance method
Table: 4.11 Linearity results of the Imp-A
Imp-A
Concentration
(µg/mL) Area Statistical Analysis
0.227 4478 Slope 20120
0.550 9575
0.824 16077 Intercept -669
1.374 27528
1.649 33017 Residual sum of
squares 898
1.924 37772
2.473 47188 Correlation
Coefficient 0.9994
2.748 55545
3.298 65435 Response Factor 2.21
Linearity Plot (Concentration Vs Area)
Fig: 4.10 Linearity Plot for Imp-A
4478
19478
34478
49478
64478
79478
0.227 1.127 2.027 2.927 3.827
Are
a
Conc.(µg/mL)
-
189
Table: 4.12 Linearity results of the Imp-B
Imp-B
Concentration
(µg/mL) Area Statistical Analysis
0.180 8975 Slope 49241
0.504 24310
0.756 37725 Intercept -99
1.260 62100
1.512 73753 Residual Sum of
Squares 415
1.764 86477
2.268 111431 Correlation
Coefficient 0.9999
2.520 124521
3.024 148654 Response 0.90
Linearity Plot (Concentration Vs Area)
Fig: 4.11 Linearity Plot for Imp-B
8975
43975
78975
113975
148975
183975
0.180 1.080 1.980 2.880 3.780
Are
a
Conc.(µg/mL)
-
190
Table: 4.13 Linearity results of the Imp-C
Imp-C
Concentration
(µg/mL) Area Statistical Analysis
0.172 6691 Slope 43285
0.500 21425
0.749 33058 Intercept -88
1.249 54391
1.499 64653 Residual Sum of
Squares 577
1.749 75611
2.248 97452 Correlation
Coefficient 0.9999
2.498 108358
2.998 128583 Response 1.03
Linearity Plot (Concentration Vs Area)
Fig: 4.12 Linearity Plot for Imp-C
6691
36691
66691
96691
126691
156691
0.172 1.022 1.872 2.722 3.572
Are
a
Conc.(µg/mL)
-
191
Table: 4.14 Linearity results of the Imp-D
Imp-D
Concentration (µg/mL)
Area Statistical Analysis
0.166 9162 Slope 45771
0.507 23428
0.760 35326 Intercept 866
1.266 59613
1.520 70227 Residual Sum of Squares
558 1.773 81942
2.279 104834 Correlation Coefficient
0.9999 2.533 116927
3.039 139386 Response 0.97
Linearity Plot (Concentration Vs Area)
Fig: 4.13 Linearity Plot for Imp-D
9162
39162
69162
99162
129162
159162
0.166 1.066 1.966 2.866 3.766
Are
a
Conc.(µg/mL)
-
192
Table: 4.15 Linearity results of the Florfenicol
Florfenicol
Concentration
(µg/mL) Area Statistical Analysis
0.144 6253 Slope 44463
0.528 23091
0.792 36323 Intercept 81
1.321 59019
1.585 70100
Residual sum of
squares 713 1.849 82184
2.377 105209
2.641 118687
Correlation
coefficient 0.9999 3.170 140187
3.962 176500
Linearity Plot (Concentration Vs Area)
Fig: 4.14 Linearity Plot for Florfenicol
6253
39253
72253
105253
138253
171253
0.144 1.094 2.044 2.994 3.944
Are
a
Conc.(µg/mL)
-
193
Table: 4.16 Linearity results of the Imp-E
Imp-E
Concentration (µg/mL)
Area Statistical Analysis
0.165 8391 Slope 40988
0.512 20235
0.768 31118 Intercept 889
1.280 53878
1.536 64690 Residual sum of
squares 1194
1.792 75261
2.304 95501 Correlation
Coefficient 0.9997
2.560 107453
3.072 125550 Response Factor 1.08
3.840 157581
Linearity Plot (Concentration Vs Response)
Fig: 4.15 Linearity Plot of for Imp-E
8391
37391
66391
95391
124391
153391
0.165 1.065 1.965 2.865 3.765
Are
a
Conc.(µg/mL)
-
194
Table: 4.17 Linearity results of the Imp-F
Imp-F
Concentration (µg/ml)
Area Statistical Analysis
0.154 8835 Slope 47172
0.493 23432
0.739 34887 Intercept 411
1.232 59155
1.478 70031 Residual sum of squares
1043 1.725 81219
2.218 105102 Correlation Coefficient
0.9998 2.464 115059
2.957 139092 Response Factor 0.94
3.696 176568
Linearity Plot (Concentration Vs Response)
Fig: 4.16 Linearity Plot of for Imp-F
8835
40835
72835
104835
136835
168835
0.154 1.004 1.854 2.704 3.554
Are
a
Conc.(µg/mL)
-
195
Table: 4.18 Linearity results of the Imp-G
Imp-G
Concentration
(µg/mL) Area Statistical Analysis
0.151 8863 Slope 34332
0.504 16368
0.756 25112 Intercept 1302
1.261 44784
1.513 54162 Residual Sum of
Squares 1828
1.765 62349
2.269 79927 Correlation
Coefficient 0.9990
2.521 89814
3.025 102709 Response Factor 1.30
3.782 131349
Linearity Plot (Concentration Vs Area)
Fig: 4.17 Linearity Plot of for Imp-G
5230
14230
23230
32230
41230
50230
59230
0.200 0.550 0.900 1.250 1.600 1.950
Are
a
Con. (µg/mL)
-
196
4.2.6.7 Accuracy/Recovery
The accuracy of an analytical procedure expresses the closeness
of agreement between the value, which is accepted either as a
conventional true value or an accepted reference value and the value
found.
Accuracy of the assay method
Accuracy of the assay method was established by injecting three
preparations of test sample at 80%, 100% and 120% of analyte
concentration (i.e.100 µg/ml). Each solution was injected twice (n=2)
into HPLC and the mean peak area of Florfenicol peak was calculated.
Assay (%w/w) of test solution was determined against three
injections (n=3) of qualified Florfenicol reference standard (Table:
4.19).
The method showed consistent and high absolute recoveries at
all three concentration (80%, 100% and 120%] levels with mean
absolute recovery ranging from 99.3 % to 100.5%. The obtained
absolute recoveries were normally distributed around the mean with
uniform RSD values. The method was found to be accurate with low %
bias (< 1.0).
-
197
Table: 4.19 Accuracy results of the Assay method
S.NO %
Concentration
Mean recovery
(%) (n=3) %RSD
1 80 100.7 0.6
2 100 100.2 0.3
3 120 100.5 0.3
4.2.6.8 Acuracy/Recovery of the Related substance method
Acuracy of the related substance method established 50%,
100%, 150% the impurities specification limit (0.30%).
Accuracy at 50% impurity specification limit:
Test solution prepared in triplicate (n=3) with impurities (Imp-A,
B, C, D, E, F and G ) at 0.15% level w.r.s analyte concentration (i.e 0.5
mg/m l). Each solution was injected thrice into HPLC. Mean %
recovery of impurities calculated in the sample solution using the area
of impurities standard at 0.3% level with respect to analyte (Table:
4.20).
-
198
Table: 4.20 Accuracy at 50% level
S.NO Impurity
name Mean
recovery(%) SD %RSD
1 Imp-A 98.2 0.60 0.6
2 Imp-B 97.8 0.40 0.4
3 Imp-C 99.5 0.81 0.8
4 Imp-D 104.6 1.18 1.1
5 Imp-E 101.6 0.75 0.7
6 Imp-F 97.3 1.18 1.2
7 Imp-G 99.2 0.35 0.4
Accuracy at 100% impurity specification limit:
Test solution prepared in triplicate (n=3) with impurities (Imp-A,
B, C, D, E, F and G) at 0.3 % level w.r.s analyte concentration (i.e 0.5
mg/m l). Each solution was injected thrice into HPLC (Table: 4.21).
Table: 4.21 Accuracy at 100% level
S.NO Impurity
name
Mean
recovery(%) SD %RSD
1 Imp-A 106.3 0.70 0.7
2 Imp-B 95.0 0.20 0.2
3 Imp-C 96.1 0.42 0.4
4 Imp-D 99.3 0.58 0.6
5 Imp-E 96.8 0.36 0.4
6 Imp-F 93.2 0.50 0.5
7 Imp-G 103.3 1.65 1.6
Accuracy at 150% impurity specification limit:
Test solution in triplicate (n=3) with impurities (Imp-A, B, C, D,
E, F and G) at 0.9 % level w.r.s analyte concentration (i.e 0.5 mg/m l).
Each solution was injected thrice into HPLC (Table: 4.22).
-
199
Table: 4.22 Accuracy at 150 % level
S.NO Impurity
name
Mean
recovery(%) SD %RSD
1 Imp-A 99.9 0.42 0.4
2 Imp-B 95.3 0.21 0.2
3 Imp-C 95.7 0.10 0.1
4 Imp-D 98.0 0.06 0.1
5 Imp-E 98.2 0.89 0.9
6 Imp-F 94.4 1.03 1.1
7 Imp-G 96.3 1.45 1.5
The related substance method showed consistent and high obsolute
recoveries of all six impurities at all three different concentrations (50,
100, 150%) levels in drug substance.
4.2.6.9 Solution state stability
The solution state stability of Florfenicol in diluent in the assay
method was carried out by leaving both the test solutions of sample
and reference standard in tightly capped volumetric flasks kept at
room temperature for two days. The same sample solutions were
assayed for every one hour interval up to the study period. The % RSD
of assay of Florfenicol during solution stability experiments was with
in 1.0%.
The solution state stability of Florfenicol related substance
method was carried out by leaving sample solution in tightly capped
volumetric flask at room temperature for two days. Content of Imp A,
B, C, D, E, F and G were checked for every six hours internal up to
the study period. No significant change was observed in the content
-
200
of all six impurities in drug solution stability experiments up to the
study period. Hence Florfenicol sample solutions are stable for atleast
48 hours in the developed method. In assay method standard and test
solutions injected at each 0h, 1h, 2h, 3h, 4h, 5h, 6h, 7h, 8h, 9h, 10h,
11h, 12h. (Table: 4.23).
Table: 4.23 Solution stability results of the Assay method
S.No Time in Hours Assay (% w/w)
1 initial 99.9
2 1 99.8
3 2 99.2
4 3 99.6
5 4 99.5
6 5 99.6
7 6 99.7
8 7 99.2
9 8 99.3
10 9 99.2
11 10 99.4
12 11 99.6
13 12 99.5
% RSD 0.24
In related substances method solution stability studies of
Florfenicol in diluent was done for 15 hrs by injecting sample solution
for every one hour interval up to the study period. The impurity
profiles obtained at different interval were very consistent and
matched with initial value.
-
201
4.2.7.0 Robustness
The robustness of an analytical procedure is a measure of its
capacity to remain unaffected by small, but deliberate variations in
method parameters and provides an indication of its reliability during
normal usage. To determine the robustness of the developed method
experimental conditions were purposely altered and the resolution
between Imp-C and Imp-D was evaluated. In each of the deliberately
altered chromatographic condition (flow rate 1.3 ml/min and
1.7ml/min, acetonitrile 23% and 27% in the mobile phase, column
temperature 25 C and 35C) the resolution between Imp-B, Imp-C
and Imp-D, Imp-E and Imp-F was greater than 2.0,, illustrating the
robustness of the method.
4.3 Mass balance
The mass balance it is a process of adding together the assay
value and the levels of degradation products to see how closely these
add up to 100% of the initial value, with due consideration of the
margin of analytical error [10]. Its establishment hence is a regulatory
requirement. The mass balance is very closely linked to the
development of stability-indicating assay method as it acts as an
approach to establish its validity. The stressed samples of Florfenicol
bulk drug were assayed against the qualified reference standard and
the results of mass balance obtained in each condition is presented
below (Table: 4.24).
-
202
Table: 4.24 Mass balance of the assay method
Degradation Mechanism
Degradation Condition
% Assay of active
substance)
Mass balance (% Assay+ % impurities+ % degradants)
Remarks
Acid 5M HCl/80°C
/60 min 91.5 99.5
Degraded to Imp-A and some unknown
degradants observed
Base 5M NaOH/
Initial 90.6 99.6
Degraded to Imp-A, Imp-C and some
unknown degradants observed
Peroxide 30% H2O2/
80°C/120 min 96.1 99.5
Some unknown degradants observed
Thermal 105°C/188
Hours 99.5 99.6 No degradation observed
Photolytic 12,000 Lux/ 188 Hours
99.5 99.6 No degradation observed
4.4 Analysis of Florfenicol drug substance stability samples
One manufacturing lot of Florfenicol drug substance was placed
on stability study in chambers maintained at ICH set conditions [11].
The analysis of stability samples were carried up to 12 months period
using the above optimized method. The stability data results obtained
are presented in Table: 4.25 to 4.26. The developed HPLC method
performed satisfactorily for the quantitative evaluation of stability
samples.
-
203
Table: 4.25 Accelerated stability data (Storage conditions:
40°C/75%RH)
Batch No: KMR(510)128 Packing & storage conditions: Each sample
packed in a polyethylene bag in a triple laminated bag and kept in a HDPE drum.
Stability study duration: 6 months Temperature
%Relative humidity
40°C/75%RH
Description
Loss on
drying (%w/w)
Identification
Assay (By HPLC,
%w/w, on
dried basis,)
Specifications
A white to almost
white
crystalline powder
NMT 0.5
IR spectrum should
concordant
with that of standard
NLT 98.0 and NMT 102.0
Initial
A white
crystalline
powder
0.15 Complies 99.5
1M A white
crystalline
powder
0.14 Complies 99.5
2M
A white
crystalline powder
0.13 Complies 99.2
3M
A white
crystalline
powder
0.15 Complies 99.4
6M A white
crystalline
powder
0.14 Complies 99.3
Related substances details on next page.
-
204
Related
Substances
LOQ
(%w/w)
LOD
(%w/w)
Related Substances (By HPLC, %w/w)
INITIAL 1M 2M 3M 6M
Imp-A 0.044 0.011 ND ND ND ND ND
Imp-B 0.034 0.018 ND ND Below
LOQ
Below
LOQ
Below
LOQ
Imp-C 0.028 0.014 0.08 0.09 0.09 0.09 0.08
Imp-D 0.030 0.011 ND ND ND ND ND
Imp-E 0.027 0.013 Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Below
LOQ
Imp-F 0.032 0.014 ND ND ND ND ND
Imp-G 0.037 0.018 ND ND ND ND ND
Highest
unknown - - ND ND ND ND ND
Total
unknown - - NA NA NA NA NA
Total RS - - 0.08 0.09 0.09 0.09 0.08
ND: Not detected
NA: Not applicable
-
205
Table: 4.26 Long-term stability data (Storage conditions:
25°C/60%RH)
Batch No: KMR(510)128 Packing & storage conditions: Each sample
packed in a polyethylene bag in a triple laminated bag and kept in a HDPE drum
Stability study duration: 12 months Temperature
%Relative humidity
25°C/60%RH
Tests Description
Loss on
drying (%w/w)
Identification
Assay (By HPLC,
%w/w, on
dried basis,)
Specifications
A white to almost
white
crystalline powder
NMT 0.5
IR spectrum should
concordant
with that of standard
NLT 98.0 and NMT 102.0
Initial
A white
crystalline
powder
0.15 Complies 99.5
1M A white
crystalline
powder
0.14 Complies 99.6
2M
A white
crystalline powder
0.13 Complies 99.4
3M
A white
crystalline
powder
0.14 Complies 99.5
6M A white
crystalline
powder
0.13 Complies 99.2
9M
A white
crystalline powder
0.13 Complies 99.3
12M
A white
crystalline
powder
0.12 Complies 99.2
Related substances details on next page.
-
206
Related Substances
LOQ (%w/w)
LOD
(%w/w)
Related Substances (By HPLC, %w/w)
Initial 1M 2M 3M 6M 9M 12M
Imp-A 0.044 0.011 ND ND ND ND ND ND ND
Imp-B 0.034 0.018 ND Below LOQ
ND Below LOQ
Below LOQ
ND Below LOQ
Imp-C 0.028 0.014 0.07 0.08 0.07 0.06 0.06 0.07 0.08
Imp-D 0.030 0.011 ND ND ND ND ND ND ND
Imp-E 0.027 0.013 Below LOQ
Below LOQ
Below LOQ
Below LOQ
Below LOQ
Below LOQ
Below LOQ
Imp-F 0.032 0.014 ND ND ND ND ND ND ND
Imp-G - - ND ND ND ND ND ND ND
Highest unknown
- - ND ND ND ND ND ND ND
Total unknown
- - NA NA NA NA NA NA NA
Total RS - - 0.08 0.09 0.08 0.08 0.08 0.08 0.09
ND: Not detected NA: Not applicable
-
207
4.5 Summary and conclusions
Validated stability-indicating HPLC method was developed for
Florfenicol after subjecting the samples to stress testing under ICH
recommendes conditions. The RPLC method was developed for
quantitative and related substance determination of Florfenicol is
rapid precise, accurate and selective. The method was completely
validated showing satisfactory data for all the method validation
parameters tested. The developed method was found to be ‘specific’ to
the drug, as the peaks of the degradation products did not interfere
with the degradation peak. Thus the proposed method can be
employed for assessing the stability of Florfenicol bulk drug samples.
-
208
Table: 4.27 Summary of Analytical method validation data
Test Parameter
Related Substances method Assay
method
Imp-A Imp-B Imp-C Imp-D Imp-E Imp-F Imp-G
Precision (RSD) 1.8 0.6 0.6 0.3 1.0 0.6 0.9 0.7
LOD (µg/ml) 0.113 0.090 0.086 0.084 0.082 0.077 0.075 N/A
LOQ (µg/ml) 0.044 0.034 0.028 0.039 0.030 0.027 0.037
N/A
Linearity (corre coefficient
0.9994 0.9999 0.9999 0.9999 0.9997 0.9998 0.9990 0.9995
Accuracy (%) 98.2-106.3 95.0-97.8 95.7-99.5 98.0-104.6 98.2-101.6 93.2-97.3 96.3-103.3 100.2-100.7
Robustness
Resolution b/w Imp-E& Imp-
F>2
Resolution b/w
Imp-E& Imp-F>2
Resolution b/w
Imp-E& Imp-F>2
Resolution b/w Imp-E& Imp-
F>2
Resolution b/w Imp-E& Imp-
F>2
Resolution b/w
Imp-E& Imp-F>2
Resolution b/w Imp-E&
Imp-F>2
Resolution b/w Imp-E&
Imp-F>2
Solution stability Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Mobile phase stability Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
Stable up to 15hr
-
209
4.6 References:
1. Sams, R. A.; in proceedings of the XVIII World Buiatrics
Congress, 1994, Bolonga, 13.
2. Piotr, K.; Lucyna, K.; Aleksandra, C.; Iiona, O.; Alina, P.; Michal
B,; Henry, L.; J. Pharm. Biomed. Anal., 2005, 39, 983.
3. Suxia, Z.; Zhongwei, L.; Xia. G.; Linli. C.; Zhanhui. W.;
Jianzhong. S.; J.Chromatogr. B, 2008, 875, 399.
4. Chue. V.; Larry, S. J.; Guy. S. R.; William, G. H.; J.Chromatogr.
B, 2002, 780, 111.
5. Hayes, J. M.; Eichman, J.; Katz, T.; Gilewicz, R.; J. AOAC,
2003, 86, 22.
6. Marciniec, B.; Stawny, M.; Kachlicki, P.; Jaroszkiewicz, E.;
Michael, N.; Analytical Sciences. 2009, 25, 1255.
7. De Craene, B. A.; Deprez, P.; Haese, E. D.; Nelis, H. J.;
Bossche, W.; De Leenheer, A. P.; Antimicrobial Agents and
Chemotherapy., 1997, 41, 1991.
8. Roy, Y. P. E.; Eric, C. W.; J. Aquatic Animal Health. 2005, 17,
129.
9. ICH, Validation of analytical Procedures: Text and methodology
Q2 (R1), International Conferences on Hormonization, IFPMA,
Geneva, 2005.
10. Steven, B. W.; Pharmaceutical Stress Testing Predicting
Drug Degradation.
11. Validation of Analytical Procedures: Methodology Q2B – ICH
Guidelines.