CERTIFICATE OF ANALYSIS/media/files/coa/k00... · CERTIFICATE OF ANALYSIS COA-07417 Page 3of hu...

3
CERTIFICATE OF ANALYSIS COA-07417 Page 1 of 3 Product Description: V-PLEX® and V-PLEX Plus Vascular Injury Panel 2 (human) Kit Kit Catalog Numbers: K15198D-1, K15198D-2, K15198D-4, K15198G-1, K15198G-2, K15198G-4 Kit Lot Number: K0080578 Expiration Date: 31 Jul 2018 Acceptance Data/ Percent Recovery Specifications: Calibrator Vascular Injury Control 1 Vascular Injury Control 2 Vascular Injury Control 3 Spot Name Recovery CV Recovery CV Recovery CV Recovery CV 1 hu SAA 80% - 120% < 20% 75% - 125% < 20% 75% - 125% < 20% 75% - 125% < 20% 2 hu CRP 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 3 hu VCAM-1 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 4 hu ICAM-1 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% LLOQ ULOQ Spot Name Value Recovery CV Value Recovery CV 1 hu SAA 54.0 80% - 120% < 20% 138000 80% - 120% < 20% 2 hu CRP 27.6 80% - 120% < 20% 49600 80% - 120% < 20% 3 hu VCAM-1 37.6 80% - 120% < 20% 32000 80% - 120% < 20% 4 hu ICAM-1 6.40 80% - 120% < 20% 32700 80% - 120% < 20% Acceptance Data/ Non Specific Binding Specifications: Spot Name NSB Spec 1 hu SAA < 1.0% 2 hu CRP < 1.0% 3 hu VCAM-1 < 1.0% 4 hu ICAM-1 < 1.0%

Transcript of CERTIFICATE OF ANALYSIS/media/files/coa/k00... · CERTIFICATE OF ANALYSIS COA-07417 Page 3of hu...

Page 1: CERTIFICATE OF ANALYSIS/media/files/coa/k00... · CERTIFICATE OF ANALYSIS COA-07417 Page 3of hu ICAM-1 - Spot 4 Conc. pg/mL ECL Signal %CV 53900 576,000 5.1 10800 127,000 4.8 2160

CERTIFICATE OF ANALYSIS

COA-07417

Page 1 of 3

Product Description: V-PLEX® and V-PLEX Plus Vascular Injury Panel 2 (human) Kit

Kit Catalog Numbers: K15198D-1, K15198D-2, K15198D-4, K15198G-1, K15198G-2, K15198G-4

Kit Lot Number: K0080578

Expiration Date: 31 Jul 2018

Acceptance Data/ Percent Recovery Specifications:

Calibrator Vascular Injury Control 1 Vascular Injury Control 2 Vascular Injury Control 3

Spot Name Recovery CV Recovery CV Recovery CV Recovery CV

1 hu SAA 80% - 120% < 20% 75% - 125% < 20% 75% - 125% < 20% 75% - 125% < 20%

2 hu CRP 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20%

3 hu VCAM-1 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20%

4 hu ICAM-1 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20% 80% - 120% < 20%

LLOQ ULOQ

Spot Name Value Recovery CV Value Recovery CV

1 hu SAA 54.0 80% - 120% < 20% 138000 80% - 120% < 20%

2 hu CRP 27.6 80% - 120% < 20% 49600 80% - 120% < 20%

3 hu VCAM-1 37.6 80% - 120% < 20% 32000 80% - 120% < 20%

4 hu ICAM-1 6.40 80% - 120% < 20% 32700 80% - 120% < 20%

Acceptance Data/ Non Specific Binding Specifications:

Spot Name NSB Spec

1 hu SAA < 1.0%

2 hu CRP < 1.0%

3 hu VCAM-1 < 1.0%

4 hu ICAM-1 < 1.0%

Page 2: CERTIFICATE OF ANALYSIS/media/files/coa/k00... · CERTIFICATE OF ANALYSIS COA-07417 Page 3of hu ICAM-1 - Spot 4 Conc. pg/mL ECL Signal %CV 53900 576,000 5.1 10800 127,000 4.8 2160

CERTIFICATE OF ANALYSIS

COA-07417

Page 2 of 3

Stock Concentration of Material Tested :

Spot Name Item Conc Units

1 hu SAA

20X Calibrator 4440000 pg/mL

Vascular Injury Control 1 55800 pg/mL

Vascular Injury Control 2 4940 pg/mL

Vascular Injury Control 3 659 pg/mL

2 hu CRP

20X Calibrator 4320000 pg/mL

Vascular Injury Control 1 26300 pg/mL

Vascular Injury Control 2 5100 pg/mL

Vascular Injury Control 3 624 pg/mL

3 hu VCAM-1

20X Calibrator 1140000 pg/mL

Vascular Injury Control 1 13700 pg/mL

Vascular Injury Control 2 1280 pg/mL

Vascular Injury Control 3 154 pg/mL

4 hu ICAM-1

20X Calibrator 1080000 pg/mL

Vascular Injury Control 1 16100 pg/mL

Vascular Injury Control 2 1700 pg/mL

Vascular Injury Control 3 197 pg/mL

Controls tested may not be the same lot as shipped.

Concentrations provided in this table refer to the stock concentration of the calibrators. For top of the standard curve concentrations, please refer to the Standard Curve Data

Standard Curve Data:

hu SAA - Spot 1

Conc. pg/mL

ECL Signal

%CV

222000 767,000 2.2

44400 136,000 3.6

8880 24,800 5.7

1780 4,950 5.6

355 1,060 5.3

71.0 291 5.1

14.2 129 7.5

0.00 90 6.7

hu CRP - Spot 2

Conc. pg/mL

ECL Signal

%CV

216000 1,100,000 2.6

43200 439,000 2.5

8640 102,000 2.6

1730 21,300 3.0

346 4,350 3.2

69.1 964 3.6

13.8 268 3.2

0.00 98 12.0

hu VCAM-1 - Spot 3

Conc. pg/mL

ECL Signal

%CV

56800 474,000 2.1

11400 121,000 2.3

2270 20,200 3.1

454 3,170 2.9

90.9 651 2.5

18.2 184 4.1

3.64 92 3.7

0.00 68 6.1

Page 3: CERTIFICATE OF ANALYSIS/media/files/coa/k00... · CERTIFICATE OF ANALYSIS COA-07417 Page 3of hu ICAM-1 - Spot 4 Conc. pg/mL ECL Signal %CV 53900 576,000 5.1 10800 127,000 4.8 2160

CERTIFICATE OF ANALYSIS

COA-07417

Page 3 of 3

hu ICAM-1 - Spot 4

Conc. pg/mL

ECL Signal

%CV

53900 576,000 5.1

10800 127,000 4.8

2160 26,300 7.0

431 5,270 7.5

86.2 1,150 8.2

17.2 284 6.0

3.45 104 8.1

0.00 66 18.0

Components:

Description Lot Number Storage Temperature Expiration Date

Vascular Injury Panel 2 Plate Z0055645 2-8°C 31 Jan 2019

Anti-hu SAA Antibody D0080687 2-8°C 31 Jul 2018

Anti-hu CRP Antibody D0080731 2-8°C 30 Sep 2018

Anti-hu VCAM-1 Antibody D0080732 2-8°C 30 Sep 2018

Anti-hu ICAM-1 Antibody D0080705 2-8°C 31 Aug 2018

Vascular Panel 2 (human) Calibrator Blend A0040150 ≤ -70°C 31 Aug 2018

Diluent 101 M1010023 ≤ -10°C 30 Sep 2018

MSD Read Buffer T (4X) Not Kit Specific Not Kit Specific N/A

Additional Components Included in V-PLEX Plus

Plate Seals Not Kit Specific

Wash Buffer (20X), 100mL Not Kit Specific

Vascular Injury Control 1 Not Kit Specific

Vascular Injury Control 2 Not Kit Specific

Vascular Injury Control 3 Not Kit Specific

Statement:

The above kit was tested using MSD procedures and it conforms to MSD's specifications. The above product is intended for

Research Use Only. Not for use in Diagnostic Procedures.

Name Function Signature Date

Review/Approval Dawn Stollar Quality Assurance

14 OCT 2016