CERTIFICATE OF ACCREDITATION...ISO/IEC 17025:2005 and national standard ANSI/NCSL Z540-1-1994 while...

13
CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Continental Testing 104 S. Main Street Union, OH 45322 has been assessed by ANAB and meets the requirements of international standard ISO/IEC 17025:2005 and national standard ANSI/NCSL Z540-1-1994 while demonstrating technical competence in the field(s) of CALIBRATION Refer to the accompanying Scope(s) of Accreditation for information regarding the types of calibrations and/or tests to which this accreditation applies. AC-1647 Certificate Number ANAB Approval Certificate Valid: 03/28/2016-03/29/2018 Version No. 002 Issued: 03/28/2016 This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated January 2009).

Transcript of CERTIFICATE OF ACCREDITATION...ISO/IEC 17025:2005 and national standard ANSI/NCSL Z540-1-1994 while...

  • CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board

    500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that

    Continental Testing 104 S. Main Street Union, OH 45322

    has been assessed by ANAB and meets the requirements of international standard

    ISO/IEC 17025:2005 and national standard

    ANSI/NCSL Z540-1-1994 while demonstrating technical competence in the field(s) of

    CALIBRATION Refer to the accompanying Scope(s) of Accreditation for information regarding the types

    of calibrations and/or tests to which this accreditation applies.

    AC-1647 Certificate Number

    ANAB Approval Certificate Valid: 03/28/2016-03/29/2018 Version No. 002 Issued: 03/28/2016

    This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality

    management system (refer to joint ISO-ILAC-IAF Communiqué dated January 2009).

  • Version 002 Issued: 03/28/2016 Page 1 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994

    Continental Testing 104 S. Main Street, Union, OH 45322-3358

    DuWain Ake Phone: 800-648-5091 [email protected] www.continentaltesting.com

    CALIBRATION

    Valid to: March 29, 2018 Certificate Number: AC-1647 I. Electromagnetic - DC/Low Frequency

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    DC Current – Source

    (3.2 to 32) mA (32 to 320) mA

    320 mA to 3.2 A (3.2 to 10.5) A

    (10.5 A to 20) A

    140 µA/A + 19 nA 140 µA/A + 180 nA 140 µA/A + 1.8 A 160 µA/A + 18 A

    600 µA/A + 200 A 550 µA/A + 1.8 mA 550 µA/A + 4.8 mA

    Fluke / Wavetek 9100 (w/ Option -200)

    DC Voltage - Source

    Up to 320 mV 320 mV to 3.2 V

    (3.2 to 30) V (32 to 320) V

    (320 to 1.05) kV

    60 µV/V + 16 µV 60 µV/V + 160 µV 65 µV/V + 1.6 mV 65 µV/V + 16 mV 60 µV/V + 160 mV

    Fluke / Wavetek 9100

    DC Resistance - Source

    Up to 40 (40 to 400) 400 to 4 k (4 to 40) k

    (40 to 400) k 400 k to 4 M

    (4 to 40) M (40 to 400) M 400M to 2 G

    250 / + 11 m 150 / + 26 m

    150 / + 180 m 150 / + 1.8 180 / + 18

    200 / + 190 500 / + 2.6 k 600 / + 43 k 1.0 m / + 160 k

    Fluke / Wavetek 9100 (w/ Option -135)

    Conductance - Source

    (2.5 to 25) nS (25 to 250) nS

    250 nS to 2.5 S (2.5 to 25) S (25 to 250) S

    250 nS to 2.5 mS

    4.0 mS/S + 1.6 pS 2.0 mS/S + 16 pS

    1.2 mS/S + 160 pS 500 S/S + 1.6 nS 500 S/S + 16 nS

    400 S/S + 160 nS

    Fluke / Wavetek 9100

  • Version 002 Issued: 03/28/2016 Page 2 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    AC Current – Source

    Up to 32 A

    (32 to 320) A

    320 A to 3.2 mA

    (3.2 to 32) mA

    (32 to 320) mA

    320 mA to 3.2 A

    (3.2 to 10.5) A

    (10.5 to 20) A

    (20 to 32) A

    (32 to 200) A

    (200 to 1000) A

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 100) Hz

    (100 to 440) Hz

    (10 to 100) Hz (100 to 440) kHz

    (10 to 100) Hz

    700 A/A + 900 nA 1.0 mA/A + 1.8 A 2.0 mA/A + 6.0 A 2.5 mA/A + 9.0 A

    700 A/A + 300 nA 1.0 mA/A + 600 nA 2.0 mA/A + 2.0 A 2.5 mA/A + 3.0 A

    700 A/A + 340 nA 1.0 mA/A + 620 nA 2.0 mA/A + 2.0 A 2.5 mA/A + 3.0 A

    700 A/A + 3.6 A 1.0 mA/A + 6.6 A 2.0 mA/A + 13 A 2.5 mA/A + 23 A

    800 A/A + 36 A 1.0 mA/A + 51 A 2.0 mA/A + 66 A 2.5 mA/A + 98 A

    1.0 mA/A + 510 A 2.5 mA/A + 2.6 mA

    2.0 mA/A + 3.4 mA 5.0 mA/A + 11 mA

    2.0 mA/A + 7.1 mA 5.0 mA/A + 23 mA

    2.0 mA/A + 5.8 mA 7.8 mA/A + 27 mA

    2.1 mA/A + 92 mA

    6.7 mA/A + 250 mA

    2.1 mA/A + 480 mA

    Fluke / Wavetek 9100 (w/ Option -200)

  • Version 002 Issued: 03/28/2016 Page 3 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    AC Voltage - Source

    Up to 10 mV

    (10 to 32) mV

    (32 to 320) mV

    320 mV to 3.2 V

    (3.2 to 32) V

    (32 to 105) V

    (105 to 320) V

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 30) kHz (30 to 50) kHz

    (50 to 100) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 30) kHz (30 to 50) kHz

    (50 to 100) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 30) kHz (30 to 50) kHz

    (50 to 100) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 30) kHz (30 to 50) kHz

    (50 to 100) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 30) kHz (30 to 50) kHz

    (50 to 100) kHz

    10 Hz to 3 kHz (3 to 10) kHz

    (10 to 30) kHz (30 to 50) kHz

    (50 to 100) kHz

    (40 to 100) Hz 100 Hz to 1 kHz

    (1 to 3) kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    90 20

    0.90 mV/V + 2.0 mV 2.0 mV/V + 5.2 mV

    8 30

    2.0 mV/V + 1.3 mV

    25 30 51 100

    2.0 mV/V + 260 mV

    250 300 510 1.0 mV

    2.0 mV/V + 2.6 mV

    2.5 mV 3.0 mV 5.1 mV

    1.5 mV/V + 9.8 mV 3.5 mV/V + 32 mV

    17 mV 18 mV 22 mV

    1.5 mV/V + 35 mV 3.5 mV/V + 110 mV

    25 mV 25 mV 25 mV

    6 mV 1.2 mV/V + 51 mV 1.5 mV/V + 66 mV

    Fluke / Wavetek 9100

  • Version 002 Issued: 03/28/2016 Page 4 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    AC Voltage - Source (Cont.)

    (320 to 800) V

    (800 to 1 050) V

    (40 to 100) Hz 100 Hz to 1 kHz

    (1 to 3) kHz (3 to 10) kHz

    (10 to 20) kHz (20 to 30) kHz

    (40 to 100) Hz (100 to 1) kHz

    (1 to 3) kHz (3 to 10) kHz

    (10 to 20) kHz

    170 mV 170 mV 170 mV

    90 mV 1.2 mV/V + 220 mV 1.5 mV/V + 260 mV

    200 mV 200 mV 200 mV

    60 mV 1.2 mV/V + 350 mV

    Fluke / Wavetek 9100

    Capacitance - Simulation

    (0.5 to 4.0) nF 4.0 to 40) nF

    (40 to 400) nF 400 nF to 4 F

    (4 to 40) F (40 to 400) F

    (400 to 4 000) F (4 000 to 40 000) F

    3.0 mF/F + 15.0 pF 3.0 mF/F + 34 pF

    3.0 mF/F + 230 pF 4.0 mF/F + 2.3 nF 5.0 mF/F + 23 nF 5.0mF/F + 230 nF 5.0 mF/F + 2.3 F 10.0 mF/F + 62 F

    Fluke / Wavetek 9100

    Logic Pulses

    Pulse Width

    Period

    (0.3 to 999.9) S 0.3 S to 99.999 mS 1 S to 999.99 mS

    10 S to 1999.9 mS

    (0.6 to 999.9) S 6.0 S to 99.999 mS 1 S to 999.99 mS

    10 S to 2.0 S

    250 nS/S + 1.6 S 250 nS/S + 1.6 S 250 nS/S + 16 S

    250 nS/S + 160 S

    250 nS/S + 1.6 S 250 nS/S + 1.6 S 250 nS/S + 16 S

    250 nS/S + 160 S

    Fluke / Wavetek 9100 (w/ Option -100)

    Frequency Accuracy 10 Hz to 600 MHz 250 nHz/Hz Fluke / Wavetek 9100 (w/ Option -100)

  • Version 002 Issued: 03/28/2016 Page 5 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    Oscilloscopes

    AC Voltage

    Leveled Sine Wave

    10 Hz to 50 kHz

    50 kHz to 100 MHz 100 MHz to 250 MHz 250 MHz to 600 MHz

    10 Hz to 50 kHz

    Frequency Accuracy

    4.44 mV to 3.336 V 4.44 mV to 133.4 V

    4.44 mV to 2.78 V 4.44 mV to 133 V

    4.44 mV to 5.56 V 4.44 mV to 5.56 V 4.44 mV to 5.56 V

    10.66 mV to 3.34 V

    4.44 mV to 133 V

    10 Hz to 600 MHz

    4.4 mV/V 4.4 mV/V

    4.1 mV/V + 40 µV 4.1 mV/V + 40 µV

    4.4 mV/V 15.6 mV/V 30.3 mV/V 50.2 mV/V

    4.4 mV/V

    250 nHz/Hz

    Fluke / Wavetek 9100 (w/ Options -100 & -600)

    DC Current - Measure

    Up to 100 nA 100 nA to 1 µA

    (1 to 10) µA (10 to 100) µA

    100 µA to 1 mA (1 to 10) mA

    (10 to 100) mA 100 mA to 1 A

    30 µA/A + 41 pA 20 µA/A + 54 pA

    20 µA/A + 240 pA 20 µA/A + 2.0 nA 20 µA/A + 17 nA

    20 µA/A + 170 nA 35 µA/A + 2.4 µA 110 µA/A + 50 µA

    Agilent / HP 3458A

    AC Current - Measure

    Up to 100 µA

    100 µA to 1 mA

    (10 to 20) Hz (20 to 45) Hz

    (45 to 100) Hz 100 Hz to 5 kHz

    (10 to 20) Hz (20 to 45) Hz

    (45 to 100) Hz 100 Hz to 5 kHz

    (5 to 20) kHz (20 to 50) kHz (50 to 100) kHz

    4.0 mA/A + 44 nA 1.5 mA/A + 240 nA 600 µA/A + 40 nA 600 µA/A + 40 nA

    4.0 mA/A + 240 nA 1.5 mA/A + 240 nA 600 µA/A + 240 nA 300 µA/A + 240 nA 400 µA/A + 240 nA 2.7 mA/A + 420 nA 3.7 mA/A + 1.5 µA

    Agilent / HP 3458A

  • Version 002 Issued: 03/28/2016 Page 6 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    AC Current – Measure (Cont.)

    (1 to 10) mA

    (10 to 100) mA

    100 mA to 1 A

    (10 to 20) Hz (20 to 45) Hz

    (45 to 100) Hz 100 Hz to 5 kHz

    (5 to 20) kHz (20 to 50) kHz

    (50 to 100) kHz

    (10 to 20) Hz (20 to 45) Hz

    (45 to 100) Hz 100 Hz to 5 kHz

    (5 to 20) kHz (20 to 50) kHz

    (50 to 100) kHz

    (10 to 20) Hz (20 to 45) Hz

    (45 to 100) Hz 100 Hz to 5 kHz

    (5 to 20) kHz (20 to 50) kHz

    4.0 mA/A + 2.7 µA 1.5 mA/A + 2.7 µA 600 µA/A + 2.7 µA 300 µA/A + 2.7 µA 600 µA/A + 2.7 µA 4.0 mA/A + 4.5 µA 5.5 mA/A + 16 µA

    4.0 mA/A + 28 µA 1.5 mA/A + 28 µA 600 µA/A + 28 µA 300 µA/A + 28 µA 600 µA/A + 28 µA 4.0 mA/A + 45 µA

    5.5 mA/A + 160 µA

    4.0 mA/A + 290 µA 1.6 mA/A + 290 µA 800 µA/A + 290 µA 1.0 mA/A + 290 µA 3.0 mA/A + 290 µA 10 mA/A + 460 µA

    Agilent / HP 3458A

    DC Voltage - Measure

    Up to 100 mV 100 mV to 1 V

    (1 to 10) V (10 to 100) V

    (100 to 1 000) V

    9.0 µV/V + 510 nV 8.0 µV/V + 3.0 µV 8.0 µV/V + 28 µV 10 µV/V + 510 µV 10 µV/V + 7.3 mV

    Agilent / HP 3458A

  • Version 002 Issued: 03/28/2016 Page 7 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    AC Voltage - Measure

    Up to 10 mV

    (10 to 100) mV

    100 mV to 1 V

    (1 to 10) V

    (10 to 100) V

    (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz

    (20 to 50) kHz (50 to 100) kHz

    (100 to 300) kHz

    (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz

    (20 to 50) kHz (50 to 100) kHz

    (100 to 300) kHz 300 kHz to 1 MHz

    (1 to 2) MHz

    (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz

    (20 to 50) kHz (50 to 100) kHz

    (100 to 300) kHz 300 kHz to 1 MHz

    (1 to 2) MHz

    (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz

    (20 to 50) kHz (50 to 100) kHz

    (100 to 300) kHz 300 kHz to 1 MHz

    (1 to 2) MHz

    (1 to 40) Hz 40 Hz to 1 kHz

    (1 to 20) kHz (20 to 50) kHz

    (50 to 100) kHz (100 to 300) kHz

    300 kHz to 1 MHz

    300 µV/V + 4.1 µV 200 µV/V + 3.0 µV 300 µV/V + 3.1 µV 1.0 mV/V + 9.1 µV 5.0 mV/V + 66 µV 40 mV/V + 66 µV

    70 µV/V + 7.0 µV 70 µV/V + 6.1 µV 140 µV/V + 8.1 µV 300 µV/V + 38 µV 800 µV/V + 38 µV 300 µV/V + 70 µV 10 mV/V + 70 µV 15 mV/V + 70 µV

    70 µV/V + 64 µV 70 µV/V + 54 µV

    140 µV/V + 72 µV 300 µV/V + 140 µV 800 µV/V + 220 µV 300 µV/V + 610 µV 10 mV/V + 1.7 mV 15 mV/V + 1.7 mV

    70 µV/V + 700 µV 70 µV/V + 530 µV

    140 µV/V + 710 µV 300 µV/V + 1.4 mV 800 µV/V + 1.7 mV 300 µV/V + 5.2 mV 10 mV/V + 19 mV 15 mV/V + 19 mV

    200 µV/V + 9.3 mV 200 µV/V + 8.7 mV 200 µV/V + 13 mV 350 µV/V + 15 mV 1.2 mV/V + 35 mV 4.0 mV/V + 36 mV 15 mV/V + 36 mV

    Agilent / HP 3458A

  • Version 002 Issued: 03/28/2016 Page 8 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    AC Voltage – Measure (Cont.)

    750 V

    (1 to 40) Hz 40 Hz to 1 kHz (1 to 20) kHz

    (20 to 50) kHz (50 to 100) kHz

    400 µV/V + 83 mV 400 µV/V + 79 mV 600 µV/V + 79 mV 1.2 mV/V + 79 mV 3.0 mV/V + 79 mV

    Agilent / HP 3458A

    Resistance Measure

    (10 to 100)

    to 1 G

    15 110 12 1.0 m

    10 7.6 m 10 76

    10 760 m 15 8.2 50 180

    500 2.0 k 5.0 25

    Agilent / HP 3458A

    Electrical Simulation of RTD’s

    R0 = (10 to 60)

    R0 = 60 to1 k

    R0 = (1 to 2) k

    (-200 to -100) C (-100 to 100) C (100 to 630) C (630 to 850) C

    (-200 to -100) C (-100 to 100) C (100 to 630) C (630 to 850) C

    (-200 to -100) C (100 to 100) C (100 to 630) C (630 to 850) C

    0.23 C 0.15 C 0.30 C 0.45 C

    0.15 C 0.10 C 0.20 C 0.30 C

    0.12 C 0.082 C 0.16 C 0.24 C

    Fluke / Wavetek 9100

    Electrical Simulation of Thermocouples

    Type B

    Type C

    (500 to 800) C (800 to 1 000) C

    (1 000 to 1 400) C (1 400 to 1 820) C

    (0 to 600) C

    (600 to 1 000) C (1 000 to 1 800) C (1 800 to 2 320) C

    0.57 C 0.44 C 0.38 C 0.40 C

    0.33 C 0.31 C 0.43 C 0.44 C

    Fluke / Wavetek 9100

  • Version 002 Issued: 03/28/2016 Page 9 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    Electrical Simulation of Thermocouples (Cont.)

    Type E

    Type J

    Type K

    Type L

    Type N

    Type R

    Type S

    Type T

    (-250 to -200) C (-200 to -100) C (-100 to 100) C (100 to 1 000) C

    (-210 to -100) C (-100 to 800) C (800 to 1000) C

    (1 000 to 1 200) C

    (-250 to -200) C (-200 to -100) C (-100 to 100) C (100 to 600) C

    (600 to 1 372) C

    (-200 to -50) C (-50 to 200) C (200 to 700) C (700 to 900) C

    (-200 to -100) C (-100 to 900) C (900 to 1 100) C

    (1 100 to 1 300) C

    (0 to 100) C (100 to 200) C

    (200 to 1 600) C (1 600 to 1 767) C

    (0 to 200) C

    (200 to 1 000) C (1 000 to 1 400) C (1 400 to 1 767) C

    (-250 to -200) C (-200 to -100) C

    (-100 to 0) C (0 to 400) C

    0.48 C 0.27 C 0.23 C 0.26 C

    0.30 C 0.25 C 0.26 C 0.28 C

    0.59 C 0.31 C 0.25 C 0.28 C 0.31 C

    0.31 C 0.24 C 0.26 C 0.28 C

    0.37 C 0.28 C 0.27 C 0.29 C

    0.55 C 0.43 C 0.39 C 0.32 C

    0.52 C 0.40 C 0.39 C 0.40 C

    0.61 C 0.31 C 0.27 C 0.23 C

    Fluke / Wavetek 9100

  • Version 002 Issued: 03/28/2016 Page 10 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    II. Dimensional Calibration

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND MEASUREMENT

    CAPABILITY [EXPRESSED AS

    UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    Length Standards 1-10 in (15+6.0 L) µin Gauge Blocks, Supermicrometer,

    Calipers Up to 24 in (560 + 8.4 L) µin Gauge Blocks

    Micrometers Up to 24 in (49 + 5.3 L) µin Gauge Blocks

    Dial Indicators Up to 0.025 in (0.025 to 1) in

    (1 to 4) in

    85 µin 89 µin

    140 µin

    Gauge Blocks, Micrometer Head

    Test Indicators Up to 0.008 in (0.008 to 0.060) in 85 µin 45 µin Gauge Blocks

    Micrometer Heads Up to 1 in 59 µin Gauge Blocks, Electronic Indicator

    Height Gages Digital or Dial

    Up to 6 in (6 to 12) in

    71 µin 97 µin

    Gauge Blocks, Surface Plate

    Pin Gauges Up to 1 in 20 µin Supermicrometer

  • Version 002 Issued: 03/28/2016 Page 11 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    III. Mechanical Calibration

    PARAMETER / EQUIPMENT RANGE

    CALIBRATION AND

    MEASUREMENT CAPABILITY

    [EXPRESSED AS UNCERTAINTY(±)]

    REFERENCE STANDARD OR

    EQUIPMENT

    Torque Wrenches

    Up to 10 lbf·in (10 to 100) lbf· in

    (100 to 200) lbf· in

    Up to 100 lbf·ft (100 to 2 000) lbf·ft

    (2 000 to 2 500) lbf·ft

    0.06 lbf·in 0.23 lbf·in 0.30 lbf·in

    0.16 lbf·ft 2.4 lbf·ft 2.9 lbf·ft

    Torque Sensor

    Tensiometers Up to 200 lbf (200 to 2 000) lbf 0.74 lbf 2.8 lbf

    Dead weights Torque Sensor

    Force Gages Load Cells

    Tension

    Compression

    Up to 200 lbf (200 lbf to 1 200) lbf

    (200 to 1 000) lbf (1 000 to 10 000) lbf

    (10 000 to 30 000) lbf

    Up to 200 lbf (200 lbf to 1 200) lbf

    (200 to 1 000) lbf (1 000 to 10 000) lbf

    (10 000 to 30 000) lbf (30 000 to 100 000) lbf *

    0.036 lbf 0.18 kbf 0.47 lbf 5.5 lbf 15 lbf

    0.038 lbf 0.2 lbf 0.47 lbf 5.3 lbf 19 lbf 40 lbf

    Dead Weights Dead Weights

    Load Cell Load Cell Load Cell

    Dead Weights Dead Weights

    Load Cell Load Cell Load Cell Load Cell

    Scales

    Up to 50 lbf (50 to 200) lbf

    (200 to 1 200) lbf (200 to 1 000) lbf

    (1 000 to 10 000) lbf (10 000 to 30 000) lbf

    (30 000 to 100 000) lbf *

    0.032 lbf 0.17 lbf 0.17 lbf 0.50 lbf 6.8 lbf 19 lbf 20 lbf

    Class S-1 weights Class F weights Class F weights

    Load Cell Load Cell Load Cell Load Cell

    Pressure Gauges

    Up to 100 psi (35 to 500) psi

    (500 to 5 000) psi (5 000 to 10 000) psi

    (10 000 to 15 000) psi

    0.077 psi 0.74 psi 11 psi 7 psi 9 psi

    Setra 370 Pressure Indicator Pressure Indicator Pressure Indicator Pressure Indicator

    Vacuum Gauges Up to 19 psia (19 to 50) psia 0.0047 psi 0.032 psi

    Druck DPI 145 Setra 370

    Tachometers 200 000 rpm 0.01% of reading Frequency Counter, Signal Generator

  • Version 002 Issued: 03/28/2016 Page 12 of 12

    500 Montgomery St. Suite 625 Alexandria, VA 22314 703-836-0025 www.anab.org

    Notes: 1. Calibration and Measurement Capabilities (Expanded Uncertainty) are based on approximately a 95% confidence interval, using a

    coverage of k=2. 2. This laboratory provides in-laboratory and on-site calibration service. Since field (on-site) conditions are typically more variable

    than those in the laboratory, larger measurement uncertainties are expected in the field (on-site) than what is reported on the accredited scope.

    3. On-site calibration and testing services are not available for the (*) asterisk items. 4. The use of (L) signifies an expression of Length in inches. 5. This scope is formatted as part of a single document including the Certificate of Accreditation No. AC-1647.

    _________________________