cd1800sm

download cd1800sm

of 198

Transcript of cd1800sm

  • 7/25/2019 cd1800sm

    1/198

    ABBOTT

    CELL-DYN 1800

    Automated Hematology Analyzer

    Service & Support Manual

  • 7/25/2019 cd1800sm

    2/198

    ABBOTTCD1800 SM

    Theory of Operation............................................................................................................................................................................ 4

    System Overview .................................................................................................................................................................... 4

    Major Subsystem Descriptions ........................................................................................................................................... 6

    Circuit Descriptions ............................................................................................................................................................. 15

    Signal Processor Module (SPM)....................................................................................................................................... 18

    Cell Count Module (CCM) ................................................................................................................................................... 21

    Troubleshooting.................................................................................................................................................................................31

    Troubleshooting Charts ...................................................................................................................................................... 33

    Raw Data Description .......................................................................................................................................................... 40

    CCM On-Board Diagnostic LEDs ...................................................................................................................................... 41

    CPU Hardware/Software Configuration .......................................................................................................................... 44

    Service Special Commands ............................................................................................................................................... 45

    Sample Probe Description................................................................................................................................................. 49

    CELL-DYN 1800 Error Messages ...................................................................................................................................... 60

    Software Commands and Sequence ............................................................................................................................... 72

    Engineering Drawings and Schematics............................................................................................................................................. 76

    CELL-DYN 1800 PCB Reference .................................................................................................................................................... 77

    Removal & Replacement .................................................................................................................................................................. 96

    Service Equipment Required ............................................................................................................................................. 96

    Covers (CD18-A1) .................................................................................................................................................................... 97

    Flow Panel (CD18-B1)............................................................................................................................................................ 105

    Fluid Power Supply (CD18-C1).............................................................................................................................................. 115

    Syringe Assembly (CD18-E1)................................................................................................................................................. 119

    RR-E1.04 Sample Syringe Driver Assembly .................................................................................................................... 125

    RR-E1.06 Lyse Syringe Driver Assembly ..................................................................................................................... 127

    Electronics / Card Cage (CD18-F1) ........................................................................................................................................ 129

    RR-F1.01 PAM (Pre-Amplifier Module) ............................................................................................................................. 129

    RR-F1.02 MPM (Motor Processor Module) Board .......................................................................................................... 131

    RR-F1.03 CDM (Cable Distribution Module) Board ........................................................................................................ 133

    RR-F1.04 Hard Disk Drive................................................................................................................................................... 135

    RR-F1.05 Floppy Disk Drive ............................................................................................................................................... 138

    RR-F1.06 Signal Processor Module (SPM)...................................................................................................................... 140

    RR-F1.07 Cell Count Module (CCM)................................................................................................................................. 142RR-F1.08 Data Link Adapter (DLA) ................................................................................................................................... 143

    RR-F1.09 Single Board Computer (SBC) ......................................................................................................................... 145

    RR-F1.10 Card Cage Backplane PCB .............................................................................................................................. 147

    LCD/Keyboard (CD18-G1)..................................................................................................................................................... 149

    Power Supply (CD18-H1).................................................................................................................................................... 151

    Verification Procedures ................................................................................................................................................................... 154

    VP-01 Preparation for Alignment and Verification ................................................................................................................. 154

    VP-01 Preparation for Al ignment and Verification ..................................................................................................... 154

    Test Equipment and Supplies Required ....................................................................................................................... 154

    Preparation for Alignment/Verification ......................................................................................................................... 155

  • 7/25/2019 cd1800sm

    3/198

    Order of Alignment/Verification ...................................................................................................................................... 156

    VP-02 Decontamination.................................................................................................................................................... 157

    VP-03 Vacuum and Pressure Adjustments .................................................................................................................. 158

    Regulator Alignment.......................................................................................................................................................... 159

    Pressure Adjustment (0.5 psi)......................................................................................................................................... 160

    Pressure Verification (High) ............................................................................................................................................ 161

    Vacuum Adjustment (8 inch) ........................................................................................................................................... 161

    VP-04 Metering System Timing Adjustments - RBC and WBC.............................................................................. 162

    Metering Timing Fault Report .......................................................................................................................................... 163

    RBC Metering System Timing Adjustment .................................................................................................................. 165

    WBC Metering System Timing Adjustment ................................................................................................................. 166

    VP-05 CMOS Setup Verification/Adjustment ............................................................................................................... 167

    VP-06 Card Cage Backplane Test Points ..................................................................................................................... 171

    VP-07 Cable Distribution Module Test Points ............................................................................................................. 172

    VP-08 Pre-Amplifier Module (PAM) Adjustment ......................................................................................................... 173

    VP-09 Signal Processor Module (SPM) Verification/Adjustment ........................................................................... 176

    RBC Gain.............................................................................................................................................................................. 178

    RER Adjustment................................................................................................................................................................. 179

    WBC Gain............................................................................................................................................................................. 182

    PLT Gain............................................................................................................................................................................... 184

    VP-10 Diluent and Sample Verification/Adjustment ........................................................................................................ 187

    Diluent Volume Verification ............................................................................................................................................. 188

    Sample Volume Verification............................................................................................................................................ 188

    VP-11 Stepper Motor Power Test and Verification.................................................................................................... 189

    VP-12 Sample Probe Alignment Check........................................................................................................................ 190

    VP-13 Serial Transmit to LIS Verification ..................................................................................................................... 191

    VP-14 Hard Disk Drive Setup and Verification ............................................................................................................ 194

    VP-15 Software Installation/Upgrades.......................................................................................................................... 197

  • 7/25/2019 cd1800sm

    4/198

    Theory of Operation

    System Overview

    The CELL-DYN 1800 Automated Hematology Analyzer is a complex system. Analyzer performance depends on several

    components that together make up the complete hematology system. The system is comprised of the following components

    and subsystems:

    Flow Panel System [3]

    Fluid Power Supply [2]

    Reagent Inlet Panel [6]

    Syringe Drive Assemblies [7]

    Electronics Card Cage [8]

    LCD Display System [4]

    Touchpad (Membrane) Keyboard [5]

    Power Supply Assembly [1]

    Flow Panel

  • 7/25/2019 cd1800sm

    5/198

    The Flow Panel consists of tubing, solenoid valves, and other hardware components used for sample aspiration, dilution,

    measurement and waste removal.

    Fluid Power Supply

    The Fluid Power Supply contains the vacuum and pressure pumps, accumulators, waste bottles, and associated solenoids

    and hardware.

    Reagent Inlet Panel

    The Reagent Inlet Panel provides connections for incoming reagents and outgoing waste. The Lyse solenoid is also mounted

    on this panel.

    Syringe Drive Assembies

    The Syringe Drive Assemblies include the Sample Syringe for aspirating samples, the Diluent Syringe for supplying Diluent

    throughout the Flow Panel, and the Lyse Syringe for dispensing Lyse to the WBC transducer for the HGB measurement

    process.

    Electronics Card Cage

    The electronics card cage, with associated PCBs, provides command and control signals for the various electronic

    components of the instrument. This assembly contains the Backplane PCB, Cell Count Module, Signal Processor Module,

    Data Link Adaptor and Single Board Computer.

    LCD Disp lay Screen and Keyboard

    The LCD Display Screen provides a visual data display and the keyboard provides data input by the operator.

    Power Supply Assembly

  • 7/25/2019 cd1800sm

    6/198

    The Power Supply Assembly provides an AC and DC voltage source to various components on the CELL-DYN 1800 System.

    The DC Regulator PCB provides power to the Backplane PCB for use with the digital circuitry on various PCBs. It also

    provides power to the Cable Distribution Module and fans.

    The AC Regulator PCB provides power to the Backplane PCB (for use with the analog circuitry on various PCBs). It also

    provides power to the Pump Relay Module.

    The +28VDC Switching Power Supply provides power for the stepper motors via the Motor Processor Module. It also

    provides power to initially energize solenoids.

    Major Subsystem Descriptions

    To aid in understanding the overall system, the electronic modules are divided into the following major functional

    subsystems:

    Data Interface and Control Subsystem

    Measurement Subsystem

    Solenoid Motor and Pump Subsystem

    Single Board Computer Subsystem

    AC and DC Power Distribution Subsystem

    Data Interface and Control Subsystem

    The purpose of this subsystem is to interface the user data, control data, measurement data, and system status data in the

    system. This data is connected via four independent data busses:

    DLA/CCM (Data Link Adapter/Cell Count Module)

    CCM/SPM (Cell Count Module/Sample Processor Module)

    CCM/CDM (Cell Count Module/Cable Distribution Module)

    CCM/MPM (Cell Count Module/Motor Processor Module)

    See the figure below for a diagram showing the data connections.

    Data Interface and Control Block Diagram

  • 7/25/2019 cd1800sm

    7/198

    When power to the instrument is turned ON, the system is operating software is loaded from the hard disk into RAM on the

    SBC (Single Board Computer). The SBC then uses various handshaking signals and data bytes to communicate with the

    CCM (Cell Count Module) via the DLA (Data Link Adapter).

    The CCM functions as the master controller with all system functional commands residing in firmware (PROM). The CCM

    sends control data and receives status data from the CDM (Cable Distribution Module).

    The CCM provides current control to the von Behrens RBC and WBC Transducers and the two metering PCBs and serves as

    the system's analog voltmeter for use in converting the HGB signal. Data is written and read via the CCM/CDM and

    CCM/MPM data buses.

    The CDM (Cable Distribution Module) acts as a controller for the solenoids and also interfaces data from various system

    sensors.

    The MPM (Motor Processor Module) acts as a controller for all Stepper Motor Drive PCBs.

    Measurement Subsystem

  • 7/25/2019 cd1800sm

    8/198

    The measurement subsystem provides detection, amplification, and processing of the signals from the von Behrens

    RBC/PLT Transducer, von Behrens WBC Transducer, and HGB Flow Cell. RBC/PLT and WBC metering is also included in

    this subsystem.

    See the figure below for a diagram of the measurement process.

    Measurement Block Diagram

    The PAM (Pre-Amplifier Module) supplies constant current for the von Behrens RBC/PLT and WBC Transducers and HGB

    LED voltage.

    The RBC/PLT and WBC cell pulses are input to the PAM where they are amplified and routed to the SPM (Signal Processor

    Module).

    When the SPM receives signals from the RBC/PLT and WBC, the following occurs:

    The RBC/PLT signal is amplified (gain) and split into independent RBC and PLT signals.

  • 7/25/2019 cd1800sm

    9/198

    The WBC signal is amplified (gain).

    The RBC signal is routed to the cell editing circuitry.

    Cell editing is performed on the RBC signal to eliminate invalid RBC pulses.

    The SPM discriminates cell size by converting pulse height to a proportional digital value. The amplitude of each valid pulse

    is measured by a fast A/D converter then sent across the data bus to the CCM.

    The A/D data for RBC, PLT, and WBC are individually divided by the CCM into 256 discrete size channels. The cell count in

    each channel is accumulated in discrete memory locations and is used to generate count data, percentage data, and

    histogram data for RBC, PLT, WBC, and other derived parameters.

    Signals from the upper and lower detectors on the RBC/PLT and WBC metering PCBs are converted to TTL levels by

    comparators on the CDM. The signals are then routed to the CCM where they are used to control RBC/PLT and WBC

    sample timing.

    The HGB analog signal from the HGB Flow Cell is captured by the PAM where it is amplified and routed to the CCM. The

    HGB signal is then measured and converted to a digital format by a voltmeter-A/D converter.

    Solenoid, Motor Drive, and Pump Subsystem

    Solenoid control commands reside in firmware on the CCM. These commands are sent to the CDM where they are

    multiplexed to the appropriate SDM (Solenoid Drive Module). The SDM then provides the current to open and close

    individual drive solenoids.

    Stepper Motor commands are handled in much the same manner as described above. However, the final multiplexing of theStepper Drive PCBs is controlled by the MPM.

    There are two pressure pumps and one vacuum pump in the CELL-DYN 1800 System. These pumps are described as

    follows:

    A pressure pump provides air to bubble-mix samples in the Pre-Mixing Cup and the mixing chambers of the von

    Behrens RBC/PLT and WBC Transducers. A pressure regulator regulates the 0.5 psi in the pressure accumulator for

    this process.

    An unregulated pressure pump provides air to push waste from the waste bottles inside the instrument to the waste

    container attached to the instrument and to apply back pressure to clear the apertures in the von Behrens RBC/PLT

    and WBC Transducers.

    An 8" Hg vacuum accumulator, vacuum sensor, and vacuum pump supply a constant vacuum to the entire system to

    transport Diluent, Detergent, and Lyse throughout the flow system and to maintain a constant vacuum to the

    RBC/PLT and WBC metering tubes. A vacuum regulator maintains a constant vacuum source to both metering

    tubes.

    See Solenoid, Motor Drive and Pump Block Diagramfor a diagram of the solenoid and motor drive connections.

    Solenoid, Motor Drive and Pump Block Diagram

  • 7/25/2019 cd1800sm

    10/198

    Single Board Computer Subsystem

  • 7/25/2019 cd1800sm

    11/198

    The Single Board Computer subsystem consists of the following components:

    Single Board Computer PCB

    Data Link Adapter

    Input/Output Ports (serial/parallel)

    Keyboards (PC and membrane)

    LCD Display Screen

    Disk Drives (hard and floppy)

    The figure below illustrates the major components of the User Interface Computer.

    User Interface Computer

    SBC (Single Board Computer) PCB

    The Single Board Computer (SBC) PCB is a complete Celeron 850 Megahertz PC computer system with 128 Megabytes of

    RAM that is self-contained on one board. It utilizes a PC compatible BIOS with DOS capability that directly interfaces with the

    LCD Display Screen, Hard Disk Drive, Floppy Disk Drive, PS/2 Keyboard, one (1) Parallel Port and two (2) Serial Ports. It

    connects directly into the Backplane PCB along with the Data Link Adapter (DLA) and utilizes the ISA bus and Backplane

    PCB to communicate with one another.

    DLA (Data Link Adapter)

  • 7/25/2019 cd1800sm

    12/198

    The DLA PCB connects directly into the Backplane PCB. It provides a program-controlled data channel from the SBC to the

    CCM. The DLA performs the following three functions:

    Receives measurement and analyzer status data from the CCM

    Transfers commands from the SBC to the CCM

    Sends control data from the UIC to the CCM.

    Serial I/O Ports

    The CELL-DYN 1800 contains two (2) Serial I/O Ports for transferring data to other computer systems:

    COM1RS-232 Data Output

    COM2Spare

    Parallel I/O Port

    The CELL-DYN 1800 contains one parallel I/O port for transferring data to a printer.

    PC/2 Keyboard

    The external 101 key PC keyboard is used to enter alphanumeric data, such as demographic information, into the instrument.

    Touch Pad (Membrane) Keyboard

    The touch pad (membrane) keyboard is located below the LCD Display Screen. The keyboard includes a row of eight (8)

    unmarked rectangular-shaped keys corresponding to labels displayed at the bottom of the screen. These keys activate the

    indicated function or display the indicated submenu.

    LCD Display Screen (Color)

    The LCD Display Screen has the following characteristics:

    Size: 8.5 x 6.4 inches (10.4 inches measured diagonally)

    Number of colors: 16

    Pixels: 640 width x 480 height (or 800 x 600)

    Backlight on/off control (software controlled screen saver)

    An LCD Adapter, connected directly on the SBC PCB, drives the LCD Display Screen. The adapter supports SVGA 640 x

    480 and 800 x 600 graphics modes.

    Hard Disk Drive

    The hard disk drive stores the User Interface Software program and the Patient Data Log.

  • 7/25/2019 cd1800sm

    13/198

    Floppy Disk Drive

    The 1.44MB 3.5" floppy disk drive is used for program installation and provides the QC (Quality Control) data

    upload/download capability.

    Printer

    The User Interface Software supports Epson ESC-P or PCL-3 languages. The printer has its own buffer and is capable of

    printing on 8.5" x 11" (letter size) or A4 paper size. The printer supports alphanumeric and graphics reports from stored data

    and screen displays.

    Speaker

    The PC speaker or buzzer is controlled by software and is mounted on the SBC PCB.

    AC and DC Power Distribution Subsystem

  • 7/25/2019 cd1800sm

    14/198

    The Power Supply Assembly is comprised of three components: AC Regulator PCB, DC Regulator PCB and +28VDC

    Switching Power Supply. These components are mounted together as an assembly and are located at the right/rear side of

    the instrument.

    When the system is turned on, the AC line is routed into the AC Regulator PCB and +28VDC Switching Power Supply. The

    AC Regulator PCB automatically accommodates line voltages of 90 - 130VAC and 200 - 260VAC by sensing the input

    voltage and utilizing an internal comparator bank and power transformer to produce the 120VAC necessary for the

  • 7/25/2019 cd1800sm

    15/198

    subsystem's function. The Power Supply Assembly then provides an AC and DC voltage source to various components on

    the CELL-DYN 1800 System.

    +28VDC Switching Power Supply

    The +28VDC Switching Power Supply provides the voltage source to the DC Regulator PCB, which the board then uses to

    produce other voltages. The power supply cooling fan is thermistor controlled, which means that when the internal

    temperature rises above 70C, the fan operates at full speed. The fan is then turned off at 50C and kept off until the

    temperature rises above 70C.

    AC Regulator PCB

    The AC Regulator PCB provides the 120VAC used by the Pump Relay Module (PRM) for vacuum and pressure pump

    operation.

    The 12VDC (analog) is provided to the Backplane PCB, which is used by the Cell Count Module (CCM) and Signal

    Processor Module (SPM). The 12VDC is also provided to the Cable Distribution Module (CDM), Motor Processor Module

    (MPM) and Pre-Amplifier Module (PAM) (for its circuitry).

    The +100VDC is provided to the PAM for use in its constant current circuitry. In between the AC Regulator PCB and the PAM

    is the Pre Amp Filter that is used to filter out noise.

    DC Regulator PCB

    The DC Regulator PCB provides +5VDC, 12VDC (digital), +14VDC and +28VDC. The +5VDC is provided to the BackplanePCB, which is used by the SPM, CCM, Data Link Adapter (DLA) and Single Board Computer (SBC). This voltage is also used

    by the CDM, MPM, Hard and Floppy Disk Drives.

    The 12VDC (digital) is provided to the Backplane PCB for use on the CPU fan. The Hard and Floppy Disk Drives receive

    +12VDC for their operation.

    The +14VDC is provided through the CDM to the Solenoid Driver Modules (SDM) and is used to hold the solenoids closed or

    open for normally closed solenoids.

    The +28VDC is provided to the MPM for operating stepper motors and through the CDM to the SDMs to initially closesolenoids or open for normally closed solenoids. This voltage is also used for the system's internal cooling fans. The cooling

    fans are thermistor controlled, which means that when the ambient temperature inside the instrument drops below 25C, the

    fans operate at half speed. Once the temperature rises above 25C, the speed is increased linearly until it reaches 35C,

    when the fans operate at full speed.

    The DC Regulator PCB also provides +5VDC and +14VDC to the AC Regulator PCB.

    Circuit Descriptions

    This section contains a description of the circuitry for the following printed circuit boards:

  • 7/25/2019 cd1800sm

    16/198

    Pre-Amplifier Module (PAM)

    Signal Processor Module (SPM)

    Cell Count Module (CCM)

    Cable Distribution Module (CDM)

    Solenoid Driver Module (SDM)

    Motor Processor Module (MPM)

    Stepper Drive Printed Circuit Board (SDP)

    Pressure/Vacuum Regulator Module (PVRM)

    Power Supply Assembly

    Pump Relay Module (PRM)

    Single Board Computer (SBC)

    Data Link Adapter (DLA)

    Pre-amplif ier Module (PAM)

    Note

    Refer to PAM PCB Diagram.

    PAM PCB Diagram

  • 7/25/2019 cd1800sm

    17/198

    The PAM performs the following functions:

    Provides constant current control to the von Behrens RBC/PLT and WBC Transducers. Amplifies the initial RBC/PLT, WBC and HGB signals.

    The constant current bias (+100VDC) is switched by U6, then routed to Q2 and Q3, which supplies constant current to the

    von Behrens RBC/PLT transducers. The RBC/PLT current is adjusted with R72. Once received, the RBC/PLT signals are

    initially amplified by U7, then routed to U5, where they are re-inverted and further amplified. The combined RBC/PLT signal is

    then routed to the SPM PCB.

    Transistors Q4, Q5 and associated circuitry provide constant current for the von Behrens WBC Transducer. The WBC

    current is adjusted via R66. Once received, the WBC signal is initially amplified by U12, then routed to U11 where it is

    re-inverted and further amplified. The WBC signal is then routed to the SPM PCB.

  • 7/25/2019 cd1800sm

    18/198

    The output of the HGB Flow Cell is amplified by U1 and U2 prior to being routed to the SPM PCB. The HGB self test and gain

    voltages are adjusted with R14 and R4 respectively.

    Signal Processor Module (SPM)

    Note

    Refer to SPM Architecture.

    SPM Architecture

  • 7/25/2019 cd1800sm

    19/198

    EPLD (SPM)

    The EPLD is an enhanced programmable logic device. The SPM EPLD is used to control data acquisition.

    There are three state machines that run in the EPLD. One each is used for controlling data acquisition elements such as

    Multiplexers and Peak/Hold Amplifiers. The inputs to the EPLD state machines are the outputs of the various Threshold and

    Slope Detectors as well as Bubble and Area Comparators.

  • 7/25/2019 cd1800sm

    20/198

    There is a data transfer state machine which controls data flow to the CCM.

    Signal Processing

    There are two signals from the RBC and WBC transducers that are processed through two basic data acquisition circuits.

    The Platelet signal is derived from the RBC signal. There are three gain adjustments associated with each of these signals.

    In addition to the three gain adjustments there is a integration adjustment for detecting pulses that are too long.

    WBC Signal

    The WBC signal is received through a differential amplifier to reduce noise. There is a gain adjustment after the

    differential amplifier but before the Test Signal injection point.

    The WBC signal is then "Baseline Restored" to remove DC components and Baseline fluctuations due to varying

    duty cycle of the blood cells.

    A threshold detector signals to the EPLD the presence of valid pulses.

    Simultaneously the Bubble detector signals if the pulse is determined to be a bubble. The bubbles are much larger

    than pulses from blood cells. The EPLD discards the information from the pulse if it is determined to be a bubble.

    The data from the pulse is processed on the falling edge of the Threshold Detector unless the Slope Detector senses

    another rising edge before the falling edge of the Threshold Detector. The data is processed immediately if there is

    another rising edge before the falling edge of the Threshold Detector.

    The WBC Held Peak is converted and sent to the CCM on a 15S cycle. Average pulses from the impedance

    transducers are 35S.

    RBC and PLT Signals

    There are two modes for RBC/PLT data acquisition:

    RBC MCV

    RBC Count/PLT

    RBC MCV

    RBC MCV is for RBC MCV data only. In this mode, an integrator is enabled to determine if the cell is moving through the

    middle of the aperture. If the cell is not moving through the center of the aperture, the data is falsely high and the pulse is

    abnormally long. The integrated signal is compared to a proportion of its height. If the pulse is too long for its height, it is

    discarded.

    RBC Count/PLT

    RBC Count mode is for count information. During this run, Platelets are counted simultaneous with RBCs and no integration

    is used. The Platelet signal is derived from the RBC signal by an adjustable gain.

    The RBC signal is received through a differential amplifier to reduce noise. There is a gain adjustment after the

    differential amplifier but before the Test Signal injection point.

    The RBC signal is then "Baseline Restored" to remove DC components and Baseline fluctuations due to varying duty

    cycle of the blood cells. At this point an additional gain stage is added to create the Platelet signal.

  • 7/25/2019 cd1800sm

    21/198

    A threshold detector(s) signals to the EPLD the presence of valid pulses. The Platelet Threshold Detector is used if in

    RBC Count mode.

    If in Count mode, the Platelet Lower and Upper Threshold detectors are used. If the Upper Threshold Detector is

    triggered, the pulse is a RBC and the RBC pulse is accumulated into the RBC Count histogram. If only the Lower

    Detector is triggered, the PLT pulse is accumulated into the PLT histogram.

    The data from the pulse is processed on the falling edge of the Threshold Detector unless the Slope Detector(s)

    senses another rising edge before the falling edge of the Threshold Detector. If this occurs, the data is processed

    immediately.

    The RBC Held Peak or the PLT Held Peak is converted and sent to the CCM on a 15s cycle.

    Cell Count Module (CCM)

    Note

    Refer to PAM PCB Diagram.

    CCM Architecture

  • 7/25/2019 cd1800sm

    22/198

    System Clock and Microprocessor Description

    The CCM uses a 16 MHz crystal as a time base. The fundamental cycle time for the MC68HC11KW1 processor is 4 MHz.

    The processor has many built in functions such as:

    16 bit address bus

    8 bit data bus 7 - 8 bit multi-purpose I/O ports (CCM and SBC ports)

  • 7/25/2019 cd1800sm

    23/198

    Small amounts of Internal RAM and Internal EEPROM (analyzer serial number)

    Internal Timers (system timer)

    Internal UART (debug port)

    EPLD (CCM)

    The EPLD is an enhanced programmable logic device and is used for address decode and histogram building functions. Its

    program is used in coordination with the processor software.

    LEDs

    Note

    Refer to CCM On-Board Diagnostic LEDs.

    Scratch Pad RAM

    This RAM is used for general purpose programming.

    Histogram Memory

    This RAM is used for histogram building.

    Histogram Building (cell counting)

    The following sequence is executed to build histograms:

    1. Histogram Memory is cleared.

    2. Counting is enabled.

    This disconnects Histogram Memory from the processor buss.

    3. Wait for an End of Conversion (EOC) signal from SPM.

    4. At the EOC, data is read from the SPM ADC

    5. The data then becomes the address for the histogram memory.

    6. The data from the histogram memory is read.

    This data is the count information for the respective pulse height.

    7. The count is incremented and stored back into histogram memory.

    8. Counting is disabled. And the histogram can be read by the processor.

    HGB Measurement

    A 12 bit ADC with an input multiplexer is used to measure the HGB signal from the PAM. This data is read by the processor.

    Along with the HGB measurement, various DC voltages are read by the 12 bit ADC as a diagnostic.

  • 7/25/2019 cd1800sm

    24/198

    CDM Port

    The CDM port is connected directly to microprocessor ports E, G and K.

    SBC Port

    The SBC port is connected directly to microprocessor ports J and part of port A.

    Cable Distribution Module

    Note

    Refer to Solenoid, Motor Drive and Pump Block Diagram.

    Solenoid, Motor Drive and Pump Block Diagram

  • 7/25/2019 cd1800sm

    25/198

    The CDM performs the following functions:

  • 7/25/2019 cd1800sm

    26/198

    Status Sensor Interface

    Control of Solenoid Driver Module

    Pump Relay Module interface and control

    Start Board (Touch Plate) Interface

    The CDM communicates with the CCM via the CCM/CDM data bus at J2. Analog outputs of the Metering Modules are

    converted to TTL levels by comparators (U12) and placed directly on the CCM/CDM data bus. Signals from the Pump Relay

    board, Probe Position Switches, and Start Board (Touch Plate) are interfaced by Data Drivers (U5, U10).

    Data is interfaced to the Solenoid Driver Modules via J32. This data is then multiplexed by One-of-Eight Decoders (U1, U2)

    via J3, J4, J6, J7, and J9.

    Vacuum and pressure control data is latched by U14 and routed to the Pump Relay Module via J11. Pump status signals

    (Vac On, Pres On) are converted to TTL levels by U3 and placed on the data bus by U5.

    LED drive signals are routed to the Start Board (Touch Plate) via J17. The start signal enters at J17 and is placed on the data

    bus by U5.

    Solenoid Driver Module (SDM)

    Note

    Refer to Solenoid, Motor Drive and Pump Block Diagram.

    The purpose of the SDM is to provide drive current to the solenoids. Each SDM has eight Darlington drivers (Q1-Q8) which

    are individually controlled by data bits (D0-D7) and data latch (U3).

    There are two power modes available for each solenoid - activate (+28V) and hold (+14V). This is controlled by the Hi CLK

    signal in conjunction with data bits (D0-D7) and the current control latch (U1).

    Motor Processor Modu le (MPM)

    Note

    Refer to Solenoid, Motor Drive and Pump Block Diagram.

    The MPM controls drive data to the Stepper Drive printed circuit boards and also provides self-test capability for motorwinding current. The MPM is comprised of the following major circuits:

    Microprocessor

    Program Control EPROM

    I/O Peripheral Interface Adapter (PIA)

    Direct Memory Access (DMA) Control

    Motor Phase Latches

    Motor Current Latches

    Motor Winding Self Test

    Control functions of the MPM are performed by microprocessor (U5).

  • 7/25/2019 cd1800sm

    27/198

    The operating program for the microprocessor is stored in Program Control EPROM (U3).

    Data communications between the CCM and MPM are controlled by I/O PIA (U6) and serial data is interfaced via ACIA (U2)

    and Data Bus Connector (J1).

    Phase data, motor direction, and step rate are stored in RAM (U7). This data is sent to the Motor Phase Latches under

    control of the DMA Control circuitry, which consists of U11, U12, U15, U16, U18, U21 and associated circuitry. The data is

    strobed into the appropriate Motor Phase Latch by ALG0 through ALG2.

    The Motor Phase Latches U23, U26, and U29 provide phase data to the Stepper Drive printed circuit boards. Each is an 8-Bit

    Addressable Latch which can control up to four Stepper Drive printed circuit boards and subsequently four Stepper Motors.

    Four levels of motor current for each motor is controlled by the Motor Current Latches (U22, U25, and U28). Each latch can

    control up to four stepper drive printed circuit boards. Data is strobed into the appropriate latch by WR0 through WR2.

    The Feedback- and Feedback+ inputs at J3 through J14 are connected, via resistors on the Stepper Drive printed circuit

    board, to the stepper motor windings. This allows the circuitry consisting of U30, U31, and U32 to monitor the winding current

    during an internal self-test. These values can be read by the CCM to isolate a defective Stepper Drive or Stepper Motor.

    Stepper Drive Printed Circuit Board

    Note

    Refer to MPM section, blocks 20, 23, and 24 of Solenoid, Motor Drive and Pump Block Diagram.

    The Stepper Drive printed circuit board consists of two PBL 3717 motor drive chips. Each chip drives a winding of theStepper Motor. Bits I0 and I1 are used to control four motor current levels:

    P0 - High Current

    P1 - Medium Current

    P2 - Low Current

    P3 - Current Off

    Bits PH0 and PH1 control motor phase and, therefore, direction and step-rate (velocity). Feedback+ and Feedback- are used

    to generate a motor self-test.

    Pressure/Vacuum Regulator Module

    Note

    Refer to Solenoid, Motor Drive and Pump Block Diagram.

    Pressure (or vacuum) is sensed by a transducer that is internally configured as a Wheatstone Bridge. Transistor Q1and

    resistors R4 and R5 are used to generate a stable reference voltage for the Wheatstone Bridge. The output of the

    Wheatstone Bridge is partially amplified (U1-7), stabilized against long term drift (voltage follower U1-1) and made

    offset-adjustable by R18 and associated resistors.

  • 7/25/2019 cd1800sm

    28/198

    Maximum transducer sensitivity can only be achieved when the output is zero volts at TP-1 and when there is no pressure

    differential across the transducer. To accomplish this, R18 is adjusted for zero volts when both transducer inlet ports (P1 and

    P2) are open to atmospheric pressure.

    In order to maintain the operating point of comparator U2-14 at the fixed 2-volt trip level, it is necessary to maintain the output

    of U2-8 within a relatively narrow range. This is accomplished by making the differential amplifier (whose inputs are U1-10

    and U2-10) adjustable by selecting 1 of 4 possible jumper positions. A stable reference point for the DC operating level of

    U1-8 and U2-8 is established by U1-14 in conjunction with R6 and resistor network RP1.

    Measurement of pressure in the range of approximately 0.5 lbs/sq. inch is accomplished by using transducer inlet port P1 in

    conjunction with jumper setting A/B.

    Vacuum pressure in the range of 8 inch Hg is accomplished by using inlet port P2 in conjunction with jumper setting C/D. The

    regulation point for either vacuum or pressure is established by the setting of potentiometer R16.

    When the output of comparator U2-14 goes positive, the collector of Darlington transistor Q2 is pulled to ground, thereby

    turning on either the pressure or vacuum pump. When the pumps are running, LED DS1 is lit and stays lit until either the

    pressure or vacuum increases past the hysteresis point established by R8 of U2-14.

    The output of the pressure/vacuum regulator can be inhibited by a logic low at J1-1. This completes the circuit description.

    Power Supply Assembly

    Note

    Refer to major subsystem descriptions ofAC and DC Power Distribution Subsystem.

    Pump Relay Module (PRM)

    Refer to Solenoid, Motor Drive and Pump Block Diagram.

    The PRM provides drive to the vacuum and pressure pumps, via three Solid State Relays: K1, K2, and K3.

    Single Board Computer (SBC)

    The Single Board Computer (SBC) is designed around a Celeron 850MHz microprocessor and connects directly into theBackplane PCB. The SBC receives power from the Power Supply Assembly via the Backplane PCB and receives status and

    measurement data from the CCM through the DLA. The board contains EPROM, CMOS, RAM, input/output circuitry for the

    interface ports, disk drives (hard and floppy) circuitry, LCD Display Screen Interface, and PS/2 Keyboard connectivity.

    Data Link Adapter (DLA)

    The Data Link Adapter provides interfaces from the SBC to both the CCM (Cell Count Module) and the Touch Pad

    (membrane) keyboard. These two interfaces function independently under user interface software control. However, some of

    the circuits on the DLA board are shared. The DLA uses an 82C55 PPI (Parallel Peripheral Interface) IC. Refer to the figure

    below for an illustration of the DLA board.

  • 7/25/2019 cd1800sm

    29/198

    DLA (Data Link Adapter) Block Diagram

    Interface to CCM

    A 20-pin ribbon cable connects the DLA to the CCM. This interface is a bi-directional, parallel interface that is

    software-controlled at both ends. Data is transferred in 8-bit bytes on 8 data lines (D0 - D7) in one direction at a time. TheSBC (Single Board Computer) initiates an inquiry message handshake on a periodic basis or when it has a command to send.

    The CCM responds by sending data or replying to the command sent.

    The interface is controlled cooperatively by the UIC and CCM according to the state of the handshake signals. Refer to Cell

    Count Module (CCM). The UIC always sends data first. The CCM responds by sending data back (if any is available). Thus,

    the 82C55 switches its A-port from input to output and back to input during every message transaction.

    The key handshake signals are REQ1 and REQ2. REQ2 going low initiates the communication protocol. The CCM responds

    by bringing REQ1 low. After the DLA has sent its data, it brings REQ2 high. Then the CCM sends its data.

  • 7/25/2019 cd1800sm

    30/198

    Each byte received by the DLA generates an IRQ. The application software responds to the IRQ by putting the byte into a

    buffer. When the CCM brings REQ1 high, the communication is complete; the DLA returns to an idle state, and port A is set

    to input. Both REQ1 and REQ2 remains high until the next message/data transfer.

    The two LEDs on the printed circuit board, DS1 and DS2, indicate the active state of REQ2 and REQ1, respectively. They

    should always be flickering when the application software is running because the User Interface Software program is

    constantly polling the CCM to check its state. (There are some exceptions to this during power up and certain diagnostics/test

    modes.)

    The I-O address of the DLA board is assigned by four jumpers. The default I-O address (pins 3 - 6 on S1 hard-wired) is: [off

    off on off], 340 hexadecimal. The DLA interrupt level is assigned by a jumper.

    Jumpers W1 - W6 assign the DLA interrupt to one of the following:

    W # IRQ #

    W1 IRQ 5 *

    W2 IRQ 10

    W3 IRQ 10

    W4 IRQ 11

    W5 IRQ 12

    W6 IRQ 15

    * Since IRQ 5 is used by the DLA, W1 is hard-wired (this is the default).

    Interface to Touch Pad (membrane) Keyboard

    A 14-pin ribbon cable connects the DLA to the key panel. The key panel is also polled by the User Interface Softwareprogram. This interface is not interrupt-controlled.

    The keyboard is interfaced as a parallel switch matrix circuit with 4 lines out and 8 lines in. Four scan rows are driven active

    one at a time via an LS175 latch. Then the columns of the key panel matrix are read from port B of the 82C55. The software

    interprets a low active signal as a key is pressed.

    The DLA board must be plugged into a 16-bit slot. There are no adjustments on the DLA board.

  • 7/25/2019 cd1800sm

    31/198

    Troubleshooting

    Diagnostics Menu Usage

    Utilization of the DIAGNOSTICSMenu enables the operator and/or service representative to identify and correct both

    operator-correctable and service-correctable faults. When the computer senses a fault, the message displays in the System Status Box. The following keys are available in the DIAGNOSTICSMenu.

    Level One

    INITIALIZATION: Used to perform an Initialization cycle: returns movable components to home position and

    performs internal self-tests.

    RAW DATA: Used to display raw measurement data for the last specimen.

    COUNT TEST: Used to run specimens without returning to RUN Menu and display Raw Data. MORE: Used to display additional functions (levels).

    PRINTER OUTPUT: Used to toggle printer output ON and OFF.

    HELP/ERROR: Used to display help information regarding the diagnostics menu screens. The fault log can also be

    accessed through this key function.

    MAIN: Used to return to Main Menu.

    Level Two

    WBC HISTOGRAM: Used to display WBC count and histogram data accumulated in each of 256 size channels.

    RBC HISTOGRAM: Used to display RBC histogram data accumulated in each of 256 size channels.

    PLT HISTOGRAM: Used to display PLT count and histogram data accumulated in each of 256 size channels.

    SMOOTHING ON/OFF: Used to toggle histogram display status. With Smoothing Off, only raw counts are displayed.

    With Smoothing On, channels are numbered, data is normalized and the number of the peak channel displays.

    Smoothing On/Off (example)gives an example of Smoothing On/Off.

    MORE: See Level Onedescription.

    PRINTER OUTPUT: See Level Onedescription.

    HELP/ERROR: See Level Onedescription.

    MAIN: Used to return to Main Menu.

    Level Three

    PROBE HOME: Moves Sample Probe up and above RBC Cup. Displays probe assembly sensor status.

    PROBE UP: Moves Sample Probe up. Displays probe assembly sensor status.

    MORE: See Level Onedescription.

    PRINTER OUTPUT: See Level Onedescription.

    HELP/ERROR: See Level Onedescription.

    MAIN: Used to return to Main Menu.

    Level Four

  • 7/25/2019 cd1800sm

    32/198

    SYSTEM STATUS: Used to display all pending alarms.

    FAULT REPORT: Used to display all pending faults or warnings.

    SERVICE HEX CODES: Hex system codes. Not used for operator or service troubleshooting.

    SERVICE DEC CODE: Used to initiate individual actions in the CELL-DYN 1800 hardware and software.

    MORE: See Level Onedescription.

    PRINTER OUTPUT: See Level Onedescription.

    HELP/ERROR: See Level Onedescription.

    MAIN: Used to return to Main Menu.

    Smoothing On/Off (example)

    Fault Report Description

    A detailed list of all faults generated by the CELL-DYN 1800 System software and hardware is contained in CELL-DYN 1800

    Error Messages. The fault classifications reported in the Fault Report primarily contains data pertaining to the last CCM fault.

  • 7/25/2019 cd1800sm

    33/198

    If a fault occurs, pressing the [HELP/ERROR] key immediately displays the [FAULT LOG]in the DIAGNOSTICSMenu. This

    log may contain up to 16 faults, with the most current fault at the top of the list. An alternative procedure is to go to the MAIN

    MENUand press [DIAGNOSTICS] . In this case, the [FAULT REPORT], not the Fault Log, is immediately displayed.

    The Fault Log can be viewed from any of the menus, except SETUP. To view the Fault Log, enter the desired menu, followed

    by [HELP/ERROR] and [FAULT LOG]. The system displays up to 16 past faults. From the MAIN MENU, press

    [DIAGNOSTICS] followed by [MORE] three (3) times and [FAULT REPORT]to display the FAULT REPORT screen. A

    display of indicates that all faults have been cleared.

    CELL-DYN 1800 Troubleshoot ing Guide

    A list of symptoms, probable causes, and corrective actions for the most common problems encountered on the

    CELL-DYN 1800 System is given in the Troubleshooting Chart. The probable causes and corrective actions for each

    symptom are arranged in descending order from most likely to least likely. When troubleshooting a problem, start with the

    most likely cause first.

    If possible, thoroughly verify that a component is defective before replacement. Some problems can be verified visually, but

    other problems may require a measurement tool such as a DVM (Digital Volt Meter).

    When troubleshooting DATA PROBLEMS, only the measured parameters RBC, PLT, WBC, HGB, and MCV should be used

    for reference. Using the calculated parameters can become confusing when trying to isolate a problem.

    When troubleshooting CLOG AND FLOW ERROR PROBLEMS, refer to VP-04 Metering System Timing Adjustments - RBC

    and WBCfor the MIN and MAX specifications for the RBC and WBC Upper (T1) and Lower (T2) times.

    Troubleshooting Charts

    Nonfunctional Instrument Problems

    Symptom Probable Cause Corrective Act ion

    Power Cord Check Power Cord

    No Functions. No Fans Power Source Check Power Source

    No 5VDC Check that 5VDC LED on Card Cage Backplane is On

    Note

    If off replace Power Supply Assembly.

    Defective CCM Replace CCMNo Functions. Fans Run

    Defective SBC Replace SBC

    Video Display Problems

  • 7/25/2019 cd1800sm

    34/198

    Symptom Probable Cause Corrective Act ion

    Loose Connection Check all LCD Display Screen Connections

    Flash BIOS or

    CMOS settings

    reset to Default

    Connect a computer monitor to the top of the SBC PCB. Power ON

    instrument and open CMOS setup screen. Check that Advanced

    Chipset Setup is configured correctly. Refer to VP-05 CMOS Setup

    Verification/Adjustment.

    Defective Backlight

    Interconnect PCB

    Replace Backlight Interconnect PCB

    Defective LCD

    Interconnect PCB

    Replace LCD Interconnect PCB

    LCD Display Screen

    Blank/Solenoids OK

    Note

    Do not measure voltage

    on backlight inverter

    PCB. Measuring voltage

    can damage PCB.

    Defective LCD

    Display Screen

    Replace LCD Display Screen

    Characters Out Of Focus Defective LCD

    Display Screen

    Replace LCD Display Screen

    Defective SBC

    PCB

    Replace SBC PCB

    Characters OK But GarbledDefective LCD

    Display Screen

    Replace LCD Display Screen

    Defective SBC

    PCB

    Replace SBC PCB

    Missing CharactersDefective LCD

    Display Screen

    Replace LCD Display Screen

    Displayed Error and Fault Problems

    Symptom Probable Cause Corrective Action

    Restriction Check Lines

    Check In-Line Sensor

    Defective In-Line SensorReplace Sensor PCB

    Detergent Empty

    Defective CDM Replace CDM

    Solenoid 3-1, 3-3, 3-4 Tubing Check Solenoid 3-1, 3-3, 3-4

  • 7/25/2019 cd1800sm

    35/198

    Restriction Check Lines

    Check In-Line Sensor

    Defective In-Line SensorReplace Sensor PCB

    Diluent Empty

    Defective CDM Replace CDM

    Check Solenoid 3-3

    Solenoid 3-3Replace Solenoid 3-3

    Check Solenoid 3-4

    Solenoid 3-4Replace Solenoid 3-4

    Check Solenoid 3-1

    Solenoid 3-1

    Replace Solenoid 3-1

    Check Pressure Switch

    Defective Pressure SwitchReplace Switch

    Pressure Overlimit

    Defective CDM Replace CDM

    Leak 8" Hg Check for air leaks in Fluid Power Supply and

    Flow Panel

    Defective Pump Replace PumpVacuum Low Error

    Defective Vac Regulator Replace Vacuum Regulator

    Leak 0.5 Psi Check for air leaks in Fluid Power Supply and

    Flow Panel

    Defective Pump Replace PumpPressure Low Error

    Defective Pres Reg Replace Pressure Regulator

    Check Unreg Pressure

    Replace Unreg PumpNo Air Pressure

    Replace CDM

    Check Solenoid 5-3

    Solenoid 5-3Replace Solenoid 5-3

    Check Solenoid 5-7

    Solenoid 5-7Replace Solenoid 5-7

    Check Solenoid 1-6

    Waste Overflow Into

    Accumulators

    Solenoid 1-6Replace Solenoid 1-6

  • 7/25/2019 cd1800sm

    36/198

    Check Accumulator Sensor Connections

    Flush Accumulator with DI WaterAccumulator Sensor Falsely Detecting

    LiquidReplace CDM

    Check Internal Waste Sensor Connections

    Defective Internal Waste SensorReplace Sensor

    Check Unreg Pressure

    Unreg Air Pressure LowReplace Pump

    Restriction Check tubing at Fluid Power Supply, Waste

    Bottles, A and B

    Waste Empty Timeout

    Defective CDM Replace CDM

    DLA/CCM Cable connection Check DLA/CCM Cable connection

    Defective DLA/CCM Cable Replace DLA/CCM Cable

    Defective CCM Board Replace CCM Board

    Defective DLA Board Replace DLA Board

    CCM/DLA Communication Error

    Defective SBC PCB Replace SBC PCB

    Power Source Check 5VDC and +12VDC (Digital) at Card Cage

    Backplane PCB

    Check CMOS Setup

    Defective SBC PCBReplace SBC PCB

    Disk Errors (Hard Disk or Floppy

    Drives)

    Defective Disk Drive (Hard Disk or

    Floppy Drives)

    Replace Disk Drive

    Defective Switch Replace Switch and Perform Alignment

    Defective CDM Replace CDM

    Run Motor Power testDefective Stepper Drive Printed

    Circuit BoardReplace Chopper Drive PCB

    Exercise Probe (Diagnostic Menu, Probe Home)

    Defective Sample Probe AssemblyReplace Sample Probe Assembly

    Run Motor Power Test

    Position Faults

    Defective MotorReplace Motor

  • 7/25/2019 cd1800sm

    37/198

    Data Problems

    Symptom Probable Cause Corrective Action

    Check 12VDC (Analog) LEDs on Card Cage Backplane

    PCB

    Replace Power Supply ModuleAll Results Are "0" No 12VDC (Analog)

    Replace CCM

    Check that PAM 100VDC LED is On

    No 100VDCReplace Power Supply Assembly

    Defective SPM Replace SPMHGB OK All Others "0"

    Defective PAM Replace PAM

    Reference lower than

    sample reading

    Check for bubbles in Detergent line and Solenoid 2-6

    No sample aspiration Check associated tubing, solenoids and sample syringe

    Perform VP-08 Pre-Amplifier Module (PAM) Adjustment

    Defective PAMReplace PAM

    Defective CCM Replace CCM

    HGB "0" all others OK

    Defective HGB Flow Cell Replace HGB Flow Cell

    Check 0.5 psi (Bubble Mix) pressure

    Check associated tubing and hardware0.5 psi (Bubble Mix)

    pressurePerform Pressure Adjustment (0.5 psi)

    Inadequate Probe Cleaning Check associated wash block tubing and hardware

    Perform Sample Volume VerificationImprecise Sample Aspiration

    Check associated sample aspiration tubing and hardware

    Imprecise Diluent Dispense Perform Diluent Volume Verification

    Imprecision on all Parameters

    Carryover Check for correct draining of Pre Mix Cup

    Defective SPM Replace SPM

    Imprecision on all Parameters,

    HGB OKDefective PAM Replace PAM

    Dirty RBC/PLT Transducer Clean RBC/PLT Transducer

  • 7/25/2019 cd1800sm

    38/198

    Check 0.5 psi (Bubble Mix) Pressure

    Check associated tubing and hardwareIncorrect 0.5 psi (Bubble

    Mix) PressurePerform Pressure Adjustment (0.5 psi)

    Defective SPM Replace SPM

    Imprecision on RBC and PLT,

    WBC/HGB OK

    Carryover Check for correct draining of RBC/PLT Transducer

    Dirty WBC Transducer Clean WBC Transducer

    Check 0.5 psi (Bubble Mix) Pressure

    Check associated tubing and hardwareIncorrect 0.5 psi (Bubble

    Mix) PressurePerform Pressure Adjustment (0.5 psi)

    Defective SPM

    Replace SPM

    Imprecision on WBC/HGB,

    RBC and PLT OK

    Carryover Check for correct draining of WBC Transducer

    Dirty Flow Cell Clean Flow Cell

    Check reference reading (Diagnostics Menu, Raw Data)

    Reference reading out of

    specificationPerform HGB Ref Adjustment on PAM VP-08 Pre-Amplifier

    Module (PAM) Adjustment

    Defective Flow Cell Replace Flow Cell

    Imprecision on HGB, Others OK

    Solenoid operation Check Solenoid 3-6, 2-6 and 2-7

    Dirty Aperture. Clean RBC/PLT Transducer and Aperture Plate

    RBC RER Perform VP-09 Signal Processor Module (SPM)

    Verification/AdjustmentImprecision on MCV

    Defective SPM Replace SPM

    Dirty Transducer Clean WBC Transducer and Aperture Plate

    Replace Lyse Syringe

    Incorrect Lyse Volume Replace Lyse Syringe Drive Assembly

    WBC Gain Perform VP-09 Signal Processor Module (SPM)

    Verification/Adjustment

    Replace SPM

    WBC "R" Codes, Reagents OK

    Defective PCBsReplace CCM

    Check power

    Check overhead lightingHigh Electrical Backgrounds Environmental

    Check for other items on same power line

  • 7/25/2019 cd1800sm

    39/198

    Install filter (line conditioner)

    Isolate line (dedicated line)

    Check grounding cable on Front Cover

    Replace grounding cablePoor instrument cover

    groundingCheck EMI Shielding on Card Cage Backplane PCB

    Check correct routing of WBC and RBC/PLT Transducer

    cables to PAMDefective PAM

    Replace PAM

    Check Power Supply Module

    "Noisy" Power Supply

    Module

    Replace Power Supply Module

    Check Connections on Pre-Amplifier Filter PCB

    Defective Pre-Amplifier Filter

    PCBReplace Pre-Amplifier Filter PCB

    Defective Transducer(s) Replace Transducer(s)

    Clog and Flow Error Problems

    Symptom Probable Cause Corrective Act ion

    Perform VP-03 Vacuum and Pressure Adjustments

    Incorrect vacuumCheck Solenoid 1-4 and 1-5 (closed during metering)

    Diluent and Detergent lines

    reversed

    Check Reagent lines"CLOG" both sides constant

    Metering Tube position Top of Metering Tube (tapered edge) should be visible on

    top of upper DET sensor

    Check Vent Tubing

    "CLOG" T1 = Max T2 = 0, No VentRestrictionCheck Vent Solenoid 3-6, 2-1 and 1-3

    Dirty Transducer Clean Transducer and Aperture Plate

    Check plumbing

    Check Count Solenoid 1-2 and 4-3

    "CLOG" T1 = Max T2 = 0, No

    MeniscusRestriction

    Check Vent Solenoid 1-3 and 2-1

  • 7/25/2019 cd1800sm

    40/198

    Check for bubbles on right side of Transducers

    Dirty Transducer Clean Transducer and Aperture Plate

    Check associated tubing

    Restriction Check Vent Solenoid 3-6

    Perform VP-03 Vacuum and Pressure Adjustments

    "CLOG" T1 = Max T2 = 0, Slow

    Meniscus

    Incorrect VacuumCheck Solenoid 1-4 and 1-5 (closed during metering)

    Defective Upper Detector Replace Metering Printed Circuit Board

    "CLOG" T1 = Max Meniscus

    speed OKDefective CDM Replace CDM

    Defective Lower Detector Replace Metering Printed Circuit Board

    "CLOG" T1 =OK T2 = MaxMeniscus speed OK

    Defective CDM Replace CDM

    Defective Upper Detector Replace Metering Printed Circuit Board

    "FLOW ERR" T1 = Min T2 = MaxDefective CDM Replace CDM

    Defective Lower Detector Replace Metering Printed Circuit Board

    "FLOW ERR" T1 = OK T2 = MinDefective CDM Replace CDM

    Miscellaneous Problems

    Symptom Probable Cause Corrective Act ion

    Defective SPM Replace SPM

    Garbled HistogramsDefective CCM Replace CCM

    Raw Data Descr ipt ion

    From the MAIN MENU, press [DIAGNOSTICS] followed by [RAW DATA]. The [RAW DATA]key displays raw data obtained

    from the last count cycle.

    When a single count is done, all data is contained in the first column. When a PLT recount occurs, data from the first cycle

    displays in column #2 and data from the recount displays in column #1.

    Raw Data Display Description

  • 7/25/2019 cd1800sm

    41/198

    RBC, WBC and PLT counts are RAW, uncorrected total counts.

    HGB Error is not used.

    HGB Reference is the output of the A/D Converter when reading reference (2000 = 5 volts).

    HGB Sample is the output of the A/D Converter when reading sample (2000 = 5 volts).

    WBC and RBC Up Times are the upper times in milliseconds for the last sample.

    WBC and RBC Count Times are the times in milliseconds for the last sample.

    Flow Error is coded Clog or Flow Error data.

    RBC RER is RBC Cell Editing percentage.

    WBC and RBC Upper max and Upper min are the maximum and minimum Upper Times, respectively.

    WBC and RBC Avg. Time are the averages of the previous count times. The WBC and RBC Avg. time is reset when

    the [Clear Orifice]key is pressed.

    WBC and RBC Time-Outs are the floating Upper Clog Alarm Limits calculated by the "Running Average Program".

    Note

    An example of a raw data report displays in Raw Data (example).

    Raw Data (example)

    CCM On-Board Diagnostic LEDs

    The seven LEDs on the CELL-DYN 1800 CCM can reveal much about the fundamental CCM and overall machine state. In

    general, the LEDs indicate whether the CCM is in a normal functioning mode or in a fault state, and in either case, help to

    characterize the CCM state. Also, LED2 gives some information about the state of the SBC.

    The CCM tests itself on power-up. These fundamental tests include MC68HC11KW1, RAM, and SPM interfaces. If any test

    fails, the CCM attempts to execute an endless loop routine which flashes the green LED on the board. Also, it places a 4-bit

    fault code into the adjacent yellow LEDs (see Power-on LED Patterns - Fault States on Startup).

    LED Function & Counting Control

  • 7/25/2019 cd1800sm

    42/198

    The LEDs on the CCM (labeled LED1 through LED7) are entirely under program control. Their use is as follows:

    Note

    Refer to CELL-DYN 1800 CCMfor location of LEDs on PCB.

    LED7, a green LED, should not be not flashing after the CCM has successfully completed its internal power-on

    self-check diagnostics, otherwise if it is flashing slowly (~1Hz.), there is a fundamental CCM fault.

    LED5and LED6, for CER and CEW, indicate the state of the CCM firmware generated signals CER (Count Enable

    Red) and CEW (Count Enable White). These signals enable cell counting. When the associated LED is on, the DMA

    cell counting circuitry is active.

    LED3and LED4, for WCP and RCP, are on during the metered count time intervals, LED3for the white count time

    interval and LED4for the red count time interval.

    LED2is driven by the signal NREQ1, and indicates the UIC/CCM communication activity. This LED state directly

    relates to the hi/lo state of NREQ1. When LED2is flickering, which should always be the case during normal system

    operation, it indicates that the UIC/CCM communications link is active. (There is a corresponding LED on the DLA(UIC comm. board) that will also be flickering in sync with LED2; it is controlled by the DLA output signal NREQ2.)

    LED1is used to indicate that a self-test is in progress. This indicates that the tests for the pulse processing A to D

    circuitry on the SPM with the pulse counting circuitry on the CCM are active.

    Power-on LED Patterns - Normal Setup

    Step

    LED2

    NREQ1

    LED3

    WCP

    LED4

    RCP

    LED5

    CEW

    LED6

    CER

    LED7

    READY

    State

    0 ON ON ON ON Power on

    1 RAM testing

    2 Flashing RBC/PLT testing

    3 Flashing WBC testing

    4 ON ON CCM tests done; Homing flowscript running

    5 Flashing ON ON UIC program comm. started

    6 Flashing Initialized (running/idle)

    Power-on LED Patterns - Fault States on Startup

    LED2 LED3 LED4 LED5 LED6 LED7 CCM

  • 7/25/2019 cd1800sm

    43/198

    NREQ1 WCP RCP CEW CER

    READY Fault

    ON ON FlashingMC68HC11KW1 CONFIG reg

    ON FlashingMC68HC11KW1 RAM test

    ON Flashing8K RAM bit test

    Flashing8K RAM clear test

    ON ON ON FlashingHistogram RAM test

    ON ON FlashingMC68HC11KW1 timer test

    ON FlashingCDM init. test

    CELL-DYN 1800 CCM

  • 7/25/2019 cd1800sm

    44/198

    1 Cell Count Monitor (CCM) PCB 8 J6 - NOT USED

    2 S1 RESET 9 LEDs 2-7

    3 S3 BUFFALLO 10 E2 DGND

    4 J2 MPM 11 J8 - NOT USED

    5 J1 CDM 12 E3 DGND

    6 J3 SPM 13 E1 DGND

    7 J5 DLA 14 LED 1

    CPU Hardware/Software Configuration

    RS-232 Communications Test Procedure

    Detailed information on the CELL-DYN 1800 System Interface Specifications is available on the WWCS Intranet website.

  • 7/25/2019 cd1800sm

    45/198

    CMOS Setup

    The CMOS Setup contains all the information needed by the Basic Input/Output System (BIOS) to establish proper

    communications between the single board computer (SBC) and the various computer system devices. Refer to VP-05 CMOS

    Setup Verification/Adjustment:

    Special Function

    Probe Check

    There are two probe check functions activated by softkeys in the DIAGNOSTICSmenu.

    The [PROBE UP]softkey moves the probe up and maintains position without rotational motion. (The [PROBE

    DOWN]softkey then displays to restore the probe to the down position.)

    The [PROBE HOME]softkey places the probe over the RBC/PLT transducer. (The [PROBE DOWN]softkey then

    displays to restore the probe to the down position.)

    Note

    Neither procedure puts the probe in the STANDBY position (on the left).

    Service Special Commands

    Discussion

    Several commands are available to initiate individual actions in the CELL-DYN 1800 System hardware and software. These

    commands are used for troubleshooting and/or alignment when a single action is desired or required to be repeated several

    times.

    The special command mode resides in the DIAGNOSTICSMenu. From the MAINMenu, press [DIAGNOSTICS] followed by

    [MORE] three (3) times, then [SERVICE DEC CODE]. When this softkey is pressed, the message Test Select --- FOR

    SERVICE USE ONLY --- displays.

    A command can now be entered. Pressing the Enterkey on the keyboard initiates the action. Only one command can be

    entered at a time and [SERVICE DEC CODE]must be pressed before a command is entered.

    All commands available by direct softkey can be accessed by pressing [MORE].

    Note

    Use only the commands listed in DEC Service Commandsand always verify that the correct number has been

    entered before initiating the action. Use only those numbers listed in DEC Service Commands. Other numbers may

    refer to engineering commands which are not used in the field and which may cause damage if used improperly. Be

    fully aware of the purpose of any of the DEC Service Commandsbefore using them. This is a direct-activation

    method which should be used with caution because the physical state of the CELL-DYN 1800 System may not be in

    agreement with the function to be performed. After using service commands, always re-initialize the system by

    turning the power OFF then ON again or by pressing the [INITIALIZATION]key in the DIAGNOSTICSMenu to

    ensure the instrument is in the proper configuration for normal operation.

  • 7/25/2019 cd1800sm

    46/198

    DIAGNOSTICS Menu Service Code Funct ion List

    When the [SERVICE DEC CODE]key is pressed, the (Enter number (currently, 102):____prompt displays.

    The number above corresponds to the decimal code for the last code entered.

    DEC Service Commandslists the decimal-coded (DEC) service commands that can be invoked by pressing the [SERVICE

    DEC CODE]key in the DIAGNOSTICSMenu and entering the appropriate number.

    DEC Service Commands

    UIC DEC Codes Function

    07 NOT USED

    08 NOT USED

    09 NOT USED

    11 NOT USED

    15 fill lyse into system

    16 NOT USED

    17 NOT USED18 NOT USED

    19 fill Diluent & detergent

    20 mini-wash

    22 NOT USED

    23 NOT USED

    24 NOT USED

    25 NOT USED

    26 NOT USED

    33 NOT USED

    34 NOT USED

    36 NOT USED

    37 pre-dilute sample run setup

    38 pre-dilute sample run exit

  • 7/25/2019 cd1800sm

    47/198

    39 aperture current off (uses whole blood script)

    40 open all valves

    41 NOT USED

    47 platelet recount

    48 initialization (homing)

    49 open sample run

    50 clean orifice (back-flushing)

    51 pre-dilute sample run

    52 background count run

    53

    prime system with all reagents

    54 daily shutdown

    55 empty transducers and cups

    56 gain adjust

    57 unpinching normally closed valves

    59 fill transducers and cups after empty

    60 gain adjustment setup

    61 dispense 10 ml saline

    62 open sample wash

    63 clean-for-shipping

    64 clean sample syringe setup

    65 aspirate 40 l sample for 1/250 dilution

    66 dispense 10 ml saline for 1/250 dilution

    67 aspirate 100 l sample for 1/50 dilution

    68 dispense 5 ml for 1/50 dilution

    69 NOT USED

    71 lyse syringe down

    72 NOT USED

    73 NOT USED

    74 lyse syringe up and home

    75 lyse syringe down restore

  • 7/25/2019 cd1800sm

    48/198

    76 pre-dilute sample wash

    77 NOT USED

    78 NOT USED

    81 NOT USED

    83 diluent syringe down

    84 enzyme clean setup

    85 probe up and rotate and home

    86 back to ready position from probe home

    87 probe up for probe adjustment

    88

    probe down (when finished, operator should initialize the instrument to place the probe in the homeposition)

    89 sample syringe up and restore

    90 sample syringe down and home

    91 enzyme clean the system

    92 diluent syringe up and home

    93 diluent syringe down and restore

    117 NOT USED

    118 NOT USED

    119 NOT USED

    120 NOT USED

    121 cycle solenoids on waste assy

    122 cycle solenoids on flow panel assy

    123 sample syringe aspirate

    124 sample syringe dispense

    125 vacuum test

    126 check mixing pressure

    127 check backflush pump

    128 motor power test (see Service DEC Code 128)

    129 motor power level test (see Service DEC Code 129)

    130 exercise motors (see Service DEC Code 130)

  • 7/25/2019 cd1800sm

    49/198

    999 auto-cycle (seeAuto-Cycling (Code 999))

    NoteCertain commands are not sent to the CCM when the system is in an interlock state, such as STANDBY or

    UNINITIALIZED.

    Auto-Cycling (Code 999)

    The CELL-DYN 1800 can be pre-set to do a specified number of RUN cycles without user intervention. This capability

    applies only to normal RUN Count Test, Pre-Dilute RUN, (PRE-DIL TEST), Gain Adjust (GAIN ADJ), and Electrical

    Background (ELEC BKGD). This capability helps reduce test time for the instrument. The following entry screen displaysafter entering code 999:

    -- Auto Cycle Test Set Up --

    Use Spacebar to accept current number

    Use "

  • 7/25/2019 cd1800sm

    50/198

    This code runs a computer generated test (Motor Power Test) of all stepper motors, motor driver boards, and associated

    circuitry.

    The Motor Power Test should be run whenever a problem is suspected with any assembly that is driven by a stepper motor.

    The following entry screen displays after entering code 128:

    Motor Power Test Started.

    To MPM: {I }

    To MPM: {pD32}

    To MPM: {mC1!2} AC}

    To MPM: {C1}

    inp: 0415

    A report (Motor Power Test (example)) automatically displays and can be printed. Refer to VP-11 Stepper Motor Power Test

    and Verification.

    Note

    Press the [INITIALIZE]key before leaving the DIAGNOSTICSMenu.

    Motor Power Test (example)

    Service DEC Code 129

  • 7/25/2019 cd1800sm

    51/198

    This code allows the Run and Idle power levels to be set when exercising a stepper motor. The four levels are:

    0) Full Power

    1) Medium Power

    2) Low Power

    3) Off

    This code tests mechanical assemblies at various power levels or to remove idle power so the mechanism can be more

    easily moved or checked manually. The Motor Power Level Test (example)screen displays after entering code 129(press

    the ENTERkey after each entry):

    Motor Power Level Test (example)

    Note

    After the entries are made, a message, such as Motor "A" set to running power of 1 and idle power of 3,

    displays.

    Service DEC Code 130

    This code allows the direction, speed, and number of steps to be set when exercising a stepper motor.

    The Motor Check (example)screen displays after entering code 130:

    Motor Check (example)

  • 7/25/2019 cd1800sm

    52/198

    Note

    After the entries are made, a message, such as Motor "B" : motion in direction " 0" at speed "&" for 100 steps,

    displays.

    Motor Direction Commands

    The table below contains information on the motor designation, command and direction of the motor to be tested. Motor

    Speed Commandslists the motor speed commands to determine the speed of the motor being tested. Both tables are

    needed to properly test the motor.

    Motor Direction Commands

    Motor Designations Function Command

    Direction

    0 Down/Aspirate

    A/1 Sample Syringe1 Up/Dispense

    0 Up

    B/2 Probe Up/Down1 Down

    0 CCW/To RBC cup

    C/3 Probe Rotation1 CW/To Pre-Mixing Cup

    0 Down/Aspirate

    D/4 Diluent Syringe1 Up/Dispense

    0 CCW/Dispense

  • 7/25/2019 cd1800sm

    53/198

    E/5 Directional Valve 1 CW/Aspirate

    F/6 Spare

    G/7 Spare

    0 Down/Aspirate

    H/8 Lyse Syringe1 Up/Dispense

    Motor Speed Commands

    Command Speed in Steps per Second

    1 50

    2 75

    3 283

    4 300

    5 166

    6 200

    7 250

    8 10

    9 151

    10 222

    11 25

    12 182

    13 100

    14 125

    15 91

    16 67

    17 111

    Sample Probe Normal Operation

  • 7/25/2019 cd1800sm

    54/198

    Probe Up/Down INITILIZE and RUN Modesillustrates the Sample Probe's up/down sequence during the INITIALIZE and

    RUN cycles. Probe Rotate "INITALIZE" Modeshows the probe's rotation movement during the INITIALIZE cycle.

    Probe Up/Down INITILIZE and RUN Modes

    Probe Rotate " INITALIZE" Mode

    Initialization Mode

  • 7/25/2019 cd1800sm

    55/198

    The Initialization cycle places mechanical and electrical components in the "home" position, drains any liquid in the tubing,

    Pre-Mix Cup, and the Mixing Chamber of the von Behrens RBC Transducer to the Waste System, then places the instrument

    in the INITIALIZED state.

    Stepper Motor Homing

    Homing a stepper motor is the process of setting up the initial position from which all future movement is referenced. In the

    CELL-DYN 1800 System, this is accomplished by forcing the motor to move against a physical stop (Hard Stop). When the

    mechanical assembly, driven by the motor, reaches the Hard Stop, the stepper motor electrically slips until it is forced to stop.

    This mechanical position then becomes the zero reference position for the motor.

    Operation:

    1. The Sample Probe moves up at a fast speed until the Upper Switch (#2) is activated. It is then changed to a slow

    speed, and homed against the Upper Hard Stop, which is the metal plate at the top of the Sample Probe Assembly.

    2. The probe moves down six steps and the Upper Switch (#2) is checked.

    3. The probe moves CCW at a fast speed until the Right Switch (#4) is activated. It is then changed to a slow speed,

    and homed against the Right Hard Stop, which is the mounting bracket for Right Switch (#4).

    4. The probe moves CW to the Pre-Mix Cup and Left Switch (#3) is checked. The probe then moves into the Pre-Mixing

    Cup.

    5. The probe moves up and Upper Switch (#2) is checked.

    6. The probe moves CCW to center and down positions; and the Lower Switch (#1) is checked.

    7. This completes the Initialization cycle.

    Run Mode

    The figure below illustrates the probe's movements during the RUN cycle.

    Probe Rotate "RUN" Mode

  • 7/25/2019 cd1800sm

    56/198

    Operation:

    1. When the Start Switch is pressed, 30 L of sample is aspirated and Lower Switch (#1) is checked.

    2. The Sample Probe then moves up to a position six steps from Upper Hard Stop, and Upper Switch (#2) is checked.

    3. The probe moves CW to Pre-Mix Cup and Left Switch (#3) is checked.

    4. The probe moves down eight steps and into the Pre-Mix Cup, where dispense, probe shake, and aspiration of RBC

    sample takes place.

    5. The probe then moves up to a position six steps from Upper Hard Stop, and Upper Switch (#2) is checked.

    6. The probe moves CCW to the Mixing Chamber of the von Behrens RBC/PLT Transducer, stops three steps from

    Right Hard Stop, and Right Switch (#4) is checked.

    7. The probe moves down into the RBC/PLT Mixing Chamber and RBC sample is dispensed.

    8. The probe moves up to a position six steps from Upper Hard Stop, and Upper Switch (#2) is checked.

    9. After completion of the count cycle, the probe moves CW to center position.

    10. The probe moves down and Lower Switch (#1) is checked.

    11.This completes the RUN cycle.

    Switch Failure Descript ions

    Example of fault reports are shown in the following figures:

    Lower Switch (#1) Fault Report

    Upper Switch (#2) Fault Report

    Left Switch (#3) Fault Report

    Right Switch (#4) Fault Report

  • 7/25/2019 cd1800sm

    57/198

    When a switch is checked by the computer and found to be deactivated (open) in normal operation, the message "Not Ready:

    SEE DIAGNOSTICS" displays on the RUN Menu.

    From the MAINMENU, press [DIAGNOSTICS] . The screen immediately displays one of the Fault Reports shown in Lower

    Switch (#1) Fault Report, Upper Switch (#2) Fault Report, Left Switch (#3) Fault Report, and Right Switch (#4) Fault Report.

    The message indicates that Lower Switch (#1) failed when checked. The message indicates that none of the switches were activated when the failure occurred. Refer to Lower Switch (#1) Fault

    Report.

    Lower Switch (#1) Fault Report

    The message indicates that Upper Switch (#2) failed when checked. The message indicates that none of the switches were activated when the failure occurred. Refer to Upper Switch (#2) Fault

    Report.

    Upper Switch (#2) Fault Report

    The message indicates that Left Switch (#3) failed when checked. The message indicates that Left Switch (#3) was activated when the failure occurred. Refer to Left Switch (#3) Fault Report.

  • 7/25/2019 cd1800sm

    58/198

    Left Swi tch (#3) Fault Report

    The message indicates that Right Switch (#4) failed when checked. The message indicates Right Switch (#4) was activated when the failure occurred. Refer to Right Switch (#4) Fault Report.

  • 7/25/2019 cd1800sm

    59/198

    Right Switch (#4) Fault Report

    The above conditions do not necessarily indicate that a switch has actually failed. They only indicate that the switch was not

    read as activated when checked by the computer. A failure could also be caused by improper switch alignment, an electronic

    hardware failure, or a mechanical hardware failure.

  • 7/25/2019 cd1800sm

    60/198

    CELL-DYN 1800 Error Messages

    The table below lists error messages on the CELL-DYN 1800 System.

    Error Message

    (Status box)

    Description

    Time-out at N seconds A CCM process initiated by the user took longer to complete than allowed (usually indicating

    a failure of the CCM). The process ran approximately N seconds before the time-out

    occurred.

    Process Aborted A count test was stopped either by the user or because of a fault detected by the CCM.

    Fix then press [CLEAR

    ALARM]

    A user-correctable fault condition was detected.

    Process Monitoring Aborted A process was stopped by the user (using the asterisk (*) key.

    Error Message

    (Display Area II)

    Description of the error.

    Printer Time-out The Printer Output option was ON and the printer did not print the requested report in the

  • 7/25/2019 cd1800sm

    61/198

    expected time.

    Code N is Invalid The user has entered a command for the CCM whose numeric value exceeds 127. The

    value entered was N.

    Cannot do this Function The user has attempted to issue a command to the CCM that cannot be executed becauseof a pending fault condition.

    WBC Meniscus Detection,

    RBC Meniscus Detection

    During the most recent count, a meniscus was not detected or was detected at an

    unexpected time.

    WBC Count Time-out (clog),

    RBC Count Time-out (clog)

    During the most recent count, a Clog occurred.

    CCM Pulse Height Memory

    Saturation Warning

    During the most recent count, there was an overflow in one of the pulse-height arrays

    (histograms).

    External Waste Full The external waste bottle has been filled.

    Lyse Empty, Detergent

    Empty, Diluent Empty

    The indicated reagent has run out.

    Invalid Alarm Set A bit was set in the fault message from the CCM that has no valid interpretation.

    *NOT ON ANY SWITCH* After some mechanical motion, a reading of all the position sensors indicates that none are

    activated. (This message does not necessarily mean that a mechanical fault has occurred.)

    Waste Overflow Into

    Accumulators

    A reading of the sensor in the Waste Accumulator suggests that there is liquid in the

    accumulator.

    Error Message (Status box) Description

    Vacuum There was a vacuum failure during power-up or the instrument is unable to maintain vacuum

    level while in the READY state.

    Pressure There was a pressure failure during power-up.

    Position Fault A mechanical assembly is not in the correct position for the most recent function to be

    performed, as indicated by position sensors.

    Sensor Fault - Internal Waste

    Empty

    A Time-Out fault occurred in draining one of the waste bottles. This error is also associated

    with positive pressure.

    Canceling Auto-Cycling This message displays when the user cancels Auto-Count Testing.

    Invalid UIC Command Sent to

    CCM

    The UIC sent a command to the CCM that it cannot interpret.

    Error in Flow System Timing An error in the timing of a flow script has occurred. This occurs during instrument

    initialization when the flow script takes more than three minutes to complete.

    Histogram Memory Clear The CCM was unable to clear the pulse-height memory.

    CCM Program, RAM Memory The CCM detected a failure in its RAM.

  • 7/25/2019 cd1800sm

    62/198

    CCM/MPM Message Fault

    Other error-related messages

    MPM to CCM, Message

    Transmit Error

    Command to be Sent to MPM

    is Incorrect

    Abnormal Time-out/No MPM

    Response

    Fault Response from MPM

    Incorrect Command to be

    Sent to MPM

    Attempt to Send MPM a New

    Command while Busy

    An error in CCM/MPM interprocessor communications occurred. A fault was generated in an

    attempt to send or receive motor or other command to or from MPM, or the MPM was unable

    to perform the function.

    No Such Script in ROM or

    RAM

    Error in loading a flow script.

    CCM/UIC Message Fault An error in UIC/CCM interprocessor communications occurred.

    No Response from CCM The CCM is not functioning or the signal cable connecting the CCM and UIC is faulty or

    disconnected. Turn the instrument OFF, check the CCM/UIC cable, then turn the instrument

    ON.

    CCM is Initializing The CCM is in the middle of its Initialization process.

    Undefined Event An undefined event or process occurred.

    Count Test

    The [COUNT TEST] key in the DIAGNOSTICSMenu is used to run specimens and display Count Check data without

    returning to the RUN Menu. Coded data relating to specific cycle functions, raw measurement, and flow count time are

    displayed for use in troubleshooting or service.

    Event Messages During Diagnostic Menu Count Testlists the event messages that are displayed during the Diagnostic Menu

    Count Test.

    Event Messages During Diagnostic Menu Count Test

    Event Messages Descript ion

  • 7/25/2019 cd1800sm

    63/198

    SampSw. pressed Touch Plate was pressed.

    Remove specimen Specimen should be removed.

    Count valve open The counting valve is open.

    RBC histogram availRBC histogram is available.

    WBC upper det WBC upper meniscus detection.

    WBC lower det WBC lower meniscus detection.

    RBC upper det RBC upper meniscus detection.

    RBC lower det RBC lower meniscus detection.

    Plt recount strt Platelet recount starts.

    CCM initing

    CCM initializing.

    Data invalid Data entered is invalid.

    WBC histo avail WBC histogram is available.

    Proc complete Process is completed.

    Data avail Data is available.

    CCM init done CCM initialization is completed.

    Canceled Canceled operation.

    Operator-Correctable Alarm or Fault Messages

    This table lists operator-correctable alarm or fault messages.

    Event Messages Descript ion

    External Waste FullWaste full sensor is activated.

    Detergent Low Detergent is low as dete