-DR.MAJELLA
Cardiac catheters
![Page 2: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/2.jpg)
2
In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital Vein and An angiographic film could be obtained using radiographic contrast.
![Page 3: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/3.jpg)
3
“Was the key in the lock” – Andre Cournand & Dickinson Richards said in their Nobel lecture in 1956.
Certainly this key has unlocked the door to expanded diagnostic capabilities & therapeutic interventions.
CARDIAC CATHETERS
![Page 4: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/4.jpg)
4
Ideal characteristics of catheters
Better Torque Control
Strength
Radio-opacity
Flexible
Atraumatic Tip
Low Surface frictional resistance for good trackability over guide wire.
![Page 5: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/5.jpg)
5
PARTS OF A CATHETER
HUB
BODY
TIPHUB
BODY
TIP
![Page 6: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/6.jpg)
6
FRENCH CATHETER SCALE:
The French catheter scale is commonly used to measure the outer diameter of cylindrical medical instruments including catheters, needles etc.
D(mm) = Fr/3 or Fr = D(mm)*3
MEASUREMENT:
Most commonly in adult Diagnostic Catheters of 5 – 7 Fr is used.
![Page 7: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/7.jpg)
7
![Page 8: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/8.jpg)
8
Thick walled- Accentuates pressure waveform-systolic overshoot & diastolic dips.
Thin walled _ Improves monitoring, blood sampling & flushing abilities, decrease thrombogenicity.
Disadvantage – less torque control, not suitable high pressure injection.
SIZE – wall thickness
![Page 9: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/9.jpg)
9
What is a Catheter made up of ?Materials:
A range of polymers are used for the construction of catheters, including silicone rubber latex and thermoplastic elastomers. Silicone is one of the most common choices because it is inert and unreactive to body fluids and a range of medical fluids with which it might come into contact.Materials:CATHETER:Polyvinylchloride (PVC)Polyethylene (PE)Fluoropolymers (PTFE) (TEFLON)Polyurethane (PUR)Silicone (SI)
![Page 10: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/10.jpg)
10
Very maneuverable & flexible. Covered by polyurethane coating –
reduce vascular trauma. Nylon core-increase bursting
pressure Nylon – great mechanical & physical
strength, reduced friction coefficient. Eg- USCI Cournand, birds eye,
eppendorf & sones.
DACRON
![Page 11: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/11.jpg)
11
Stiffness inbetween polyurethane & Teflon.
Selective injection. More thrombogenic than PVC,
polyurethane, silicone catheters. Stainless steel mesh braid improves
rotational control & increase bursting pressure.
Eg – pigtail angiographic catheters (cook), judkins catheters, NIH & cournand.
POLYETHYLENE
![Page 12: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/12.jpg)
12
Excellent memory – Superselective injection.
Softer than polyethylene or Teflon. Increased thrombogenicity Reshaped if immersed in boling
water. Eg – pigtail angiographic (cordis)
catheters & original judkins catheters.
POLYURETHANE
![Page 13: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/13.jpg)
13
Stiffest. Poor memory. Low friction coefficient. Eg – Brockenbrough catheters,
transducer-tip catheters & introducer sheaths.
TEFLON
![Page 14: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/14.jpg)
14
Softest & flexible. High friction coefficient- venous spasm. Increased thrombogenicity. Very poor tensile strength ( memory) Cant be reformed. Most hydrophilic. Drugs absorbed- NTG, insulin,
diazepam,thiopentone. Eg- Balloon-tip flow directed catheters.
POLYVINYL CHLORIDE
![Page 15: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/15.jpg)
15
CHARACTERISTIC
TEFLON POLYETHYLENE
POLYURETHANE PVC
FRICTION COEFFICIENT
0.04 0.21 1.35 2.0
STIFFNESS +4 +3 +2 +1
MEMORY Good Excellent Excellent Fair
MOISTURE ABSORPTION (% 24 HRS)
0 0.015 0.9 0.75
![Page 16: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/16.jpg)
16
TIP: Neither blunt nor too sharp, soft & flexible.
Bullet nose tip- least trauma, Though too taper increased tip penetration.
HUB: Metal or plastic, larger than catheter, tapered hubs – easier insertion of guidewire.
TIP & HUB
![Page 17: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/17.jpg)
17
Rt heart catheters- 100 to 125cm. Lt heart catheters- 100 to 110cm. 125cm for very tall person.
LENGTH
![Page 18: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/18.jpg)
18
Cournand Lehman Goodale–lubin Balloon floatation catheters
General purpose – RIGHT heart catheters
![Page 19: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/19.jpg)
19
Gensini NIH Eppendorf Lehman ventriculography Pigtail angiographic catheter Flow directed angiographic catheter
ANGIOGRAPHIC CATHETERS
![Page 20: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/20.jpg)
20
Judkins Amplatz Schoonmaker multipurpose catheter Coronary bypass catheter Sones Castillo
PREFORMED CORONARY CATHETERS
![Page 21: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/21.jpg)
21
Shirey transvascular catheter Brockenbrough transeptal catheter Double lumen catheter Multilumen catheter Fogarty catheter Transducer-tip catheter Angioscopic catheter Pacing catheter.
SPECIALISED CATHETERS
![Page 22: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/22.jpg)
22
COURNAND catheter
Designer: Andre Cournand,1939.End hole radio-opaque woven Dacron catheter with an outer coating of polyurethane.
Construction: very gradual distal curve Tapered tip.
Use : All purpose right heart catheter. Size : 5 to 8Fr.length – 100 & 125cm.
![Page 23: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/23.jpg)
23
SCHOOMAKER MULTIPURPOSE CATHETER Designer : Dr.Fred W Schoomaker.
Construction: polyurethane with an inner wire braid
.A-1 MP:A bend –hockey stick with straight tip 1-one end hole only A-2 MP:2side holes ,1end hole
B-1 MP :B bend gradual 90 degree curve,1 end hole only
B -2 MP: 2 sideholes and an end hole Use- CAG & LV,for crossing different lesions
PDA and MAPCAcoiling.
Size : 7 & 8 Fr. length –100cm.
![Page 24: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/24.jpg)
24
NIH catheter
Construction: USCI version- woven dacron with a nylon core.Injection at high flow rates 6 sideholesCOOK – polyethylene with a stainless steel braid, 4 to 6 sideholes.
Hole : No end hole , only sidehole catheter with a gentle curve.[excellent mixing of contrast ]
Use : visualizing RV.LV,arterial,pulmonary vasculature & great veins.
Size :USCI 5 to 8Fr.length –50, 80,100 cms 125cm.
COOK:6.5,7.3 & 8.2Fr, all 100cm. Disadvantage : perforation
![Page 25: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/25.jpg)
25
PIGTAIL CATHETER Designed : Judkins. Construction: woven dacron coated
with polyurethane or polyethylene. Hole : 4-12 non-laterally opposed
sidehole in the terminal 5cm. Terminal 5cm coiled back.
Use : most commonly used LV, aortography & pulmonary angiography.
Size : 6.5,7,3 & 8.2Fr.length – 65,80,100 & 110cm.side holes-4,6,8 or 12.
![Page 26: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/26.jpg)
26
Advantage : least traumatic, less incidence of arrythmia,catheter recoil, intramyocardial injection & cardiac perforation.
Disadvantage : Thrombogenicity, & not for prolonged haemodynamic monitoring.
Pigtail catheter
![Page 27: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/27.jpg)
27
Construction: polyurethane with stainless steel braid.
Hole : 8 nonlaterally opposed sidehole near the endhole.
Use. most commonly used LV, aortography & pulmonary angiography
Size : 7 & 8Fr. length – 110cm.
Positrol II pigtail catheter
![Page 28: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/28.jpg)
28
Construction: polyurethane over a thin nylon core.
Hole : 8 nonlaterally opposed sidehole near the endhole.
Use: LV, aortography. Size : 7 & 8Fr. length – 65,80 &
110cm. Advantage : flow rate equal to one
Fr > than designated.
Nycore high-flow pigtail catheter
![Page 29: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/29.jpg)
29
Construction: polyurethane with a stainless steel braid except in tip.
Hole : 12 nonlaterally opposed sidehole near the endhole.
Use: LV, aortography. Size : 5,7 & 8Fr. length – 65,90 &
110cm. Advantage : can withstand upto
1000psi.
Ducor high-flow femoral-ventricular Pigtail catheter
![Page 30: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/30.jpg)
30
PIG TAIL ANGIOGRAPHIC
12 Side holes evenly disperses contrast in LV
QUANTICOR [Cardiomarker pig tail] Radiopaque markers set 2 cm apartUsed for quantitative angiography
![Page 31: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/31.jpg)
31
![Page 32: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/32.jpg)
32
Grollman pigtail catheter Construction: polyethylene with for
spiraled side ports near tip. Hole : 12 nonlaterally opposed
sidehole near the endhole. 60 degree bend.
Use: RV & selective PA angiography. Size : 5,7 & 8Fr. length – 65,90 &
110cm.
![Page 33: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/33.jpg)
33
VAN TASSEL ANGLED PIGTAIL Construction: polyurethane over a thin nylon core.
Hole : 8 nonlaterally opposed sidehole near the endhole 145˚or 155˚ angle 7cm from the tip.Use: LV, aortography.
Size : 7 & 8Fr. length – 110cm. Advantage : can cross stenotic
aortic valve
![Page 34: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/34.jpg)
34
GOODALE-LUBIN CATHETER Birdseye catheter. Construction: woven dacron coated with
polyurethane. Hole : Two laterally opposed sidehole
near the endhole. Use : right heart pressure, including
wedge & blood sampling. Size : 4 to 8Fr.length – 80,100 & 125cm. Variation :Standard wall –Cournand Thin wall- Lehman
![Page 35: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/35.jpg)
35
LEHMAN CATHETER; Thin wall variation of cournand. Shorter distal curve, increased
inner diameter & decreased stiffness.
size-4 to 9Fr.length-50,80,100 &125cm.
![Page 36: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/36.jpg)
36
GENSINI catheter
Construction: woven dacron coated with polyurethane.
Hole : Three laterally opposed oval sidehole within 1.5cm of its open tip.
Use : right or left heart, pulmonary & vena cava angiographic studies.
Size : 5 to 8Fr.length – 80,100 & 125cm.
Disadvantage: More arrythmogenic
![Page 37: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/37.jpg)
37
EPPENDORF catheter
Construction: woven dacron coated with polyurethane area 20cm proximal to hub is reinforced with nylon
Hole : closed-end, six laterally opposed sidehole catheter with a gentle curve.
Use : visualizing RV.LV,arterial,pulmonary vasculature & great veins.
Size : 7 to 8Fr.length –100 & 125cm. Feature: less stiff, & more torque control.
![Page 38: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/38.jpg)
38
GUIDEWIRES
Three components Central core that tapers distally.
Flexible tip Lubricious coating.
![Page 39: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/39.jpg)
39
Fixed & movable core GW 1.spring coils 2.inner safety
wire 3.mandrel
core 4.flexible tip 5.proximal
end
![Page 40: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/40.jpg)
40
Guidewire
First available standard guidewire -0.018’ contains 2 to 3cm,safety wire in the tip
Safetywire is replaced by a ribbon steerability,trackability,torquability,kink
resistence, frictional resistance. Standard length- 175 to 190cm- usually
20cm longer than the catheter. For exchange wire-300cm. Thickness – 0.035 inch (0.9mm).
![Page 41: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/41.jpg)
41
CORE
single or multiple segment. Provides tensile strength, torque strength,
torque transmission & blood compatibility. Commonly composed of stainless steel. Nitinol – increased tractability. Disadvantage- tends to store
rather than transmit torque- wire WINDING UP.
Commonly used “workhorse wire” have moderate flexibility & support.
![Page 42: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/42.jpg)
42
Distal tip
Platinum or tungsten alloy. Radiopacity ,flexibilty & blood
compatibility. Radiopacity -2 to 3cm. Rarely 11 to
40cm. High radiopacity is a feature of more
aggressive wire, Tip load- Amount of force required
to deflect the tip into a predetermined configuration. Exp-gms of force.
![Page 43: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/43.jpg)
43
Coating
- silicone, teflon, polytetrafluoroethylene, hydrophilic polymer.
Hydrophilic wire- crosses severe stenosis & total occlusion,
![Page 44: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/44.jpg)
44
J-curve guidewire
![Page 45: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/45.jpg)
46
![Page 46: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/46.jpg)
47
Spring tip wire & Plastic wire Spring wire:A) stainless steel tip B) Nitinol tip jointed to stainless steel
shaft Steerability helps Plastic wire: Little resistence,
torquability lost , useful for severe stenosis with heavy calcification.
![Page 47: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/47.jpg)
48
![Page 48: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/48.jpg)
49
![Page 49: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/49.jpg)
50
![Page 50: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/50.jpg)
51
![Page 51: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/51.jpg)
52
![Page 52: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/52.jpg)
53
![Page 53: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/53.jpg)
54
![Page 54: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/54.jpg)
55
![Page 55: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/55.jpg)
56
![Page 56: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/56.jpg)
57
![Page 57: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/57.jpg)
58
![Page 58: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/58.jpg)
59
![Page 59: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/59.jpg)
60
![Page 60: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/60.jpg)
61
![Page 61: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/61.jpg)
62
![Page 62: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/62.jpg)
63
![Page 63: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/63.jpg)
64
![Page 64: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/64.jpg)
65
![Page 65: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/65.jpg)
66
![Page 66: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/66.jpg)
67
![Page 67: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/67.jpg)
68
![Page 68: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/68.jpg)
69
![Page 69: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/69.jpg)
70
![Page 70: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/70.jpg)
71
![Page 71: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/71.jpg)
72
![Page 72: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/72.jpg)
73
![Page 73: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/73.jpg)
74
![Page 74: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/74.jpg)
75
![Page 75: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/75.jpg)
76
![Page 76: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/76.jpg)
77
![Page 77: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/77.jpg)
78
![Page 78: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/78.jpg)
79
![Page 79: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/79.jpg)
80
![Page 80: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/80.jpg)
81
![Page 81: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/81.jpg)
82
![Page 82: Cardiac catheters. 2 In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the antecubital.](https://reader035.fdocuments.us/reader035/viewer/2022062421/56649cb75503460f9497c3cb/html5/thumbnails/82.jpg)
83
In 1929 Werner Forssmann demonstrated that a simple Rubber catheter could be passed to the pulmonary artery through the Anti- Cubital Vein and An angiographic film could be obtained using radiographic contrast.