Cameron George Technical Support Engineer 26 April 2001 An Optimized Dual Column System for the...
-
Upload
jerry-gloster -
Category
Documents
-
view
213 -
download
1
Transcript of Cameron George Technical Support Engineer 26 April 2001 An Optimized Dual Column System for the...
Cameron GeorgeTechnical Support Engineer26 April 2001
An Optimized Dual Column System for the Analysis of Chlorinated Pesticides, Herbicides and PCBs by GC-ECD
Cameron GeorgeApplications ChemistApril 26, 2001
An Optimized Dual Column System for the Analysis of Chlorinated Pesticides, Herbicides and PCBs by GC-ECD
11:00 a.m. ESTTelephone Number: 904-779-4715Chair Person: Lisa Lloyd
Cameron GeorgeApplications ChemistApril 26, 2001
An Optimized Dual Column System for the Analysis of Chlorinated Pesticides, Herbicides and PCBs by GC-ECD
11:00 a.m. ESTTelephone Number: 904-779-4715Chair Person: Lisa Lloyd
Starts in Five Minutes
Cameron GeorgeApplications ChemistApril 26, 2001
An Optimized Dual Column System for the Analysis of Chlorinated Pesticides, Herbicides and PCBs by GC-ECD
11:00 a.m. ESTTelephone Number: 904-779-4715Chair Person: Lisa Lloyd
Starts in One Minute
CLP Pesticides B-0447March 8, 2001
AREAS OF FOCUS
• Injector
• Detector
• Column (including guard column)
CLP Pesticides B-0447March 8, 2001
TRACE ANALYSIS INJECTION TECHNIQUES5-2000 pg On-Column
• Megabore direct
• Splitless
• PTV
• On-Column (cold & hot)
• Large volume
CLP Pesticides B-0447March 8, 2001
SAMPLE INJECTIONGoals
• Introduce sample into the column
• Reproducible
• No efficiency losses
• Representative of sample
CLP Pesticides B-0447March 8, 2001
ShortConcentrated
LongDiffuse
Solute Bands
Same column, same chromatographic conditions
INFLUENCE OF INJECTION EFFICIENCY
CLP Pesticides B-0447March 8, 2001
FIRST, SOME BASIC DEFINITIONS:
Regarding Inlets
• Backflash
• Discrimination
CLP Pesticides B-0447March 8, 2001
A FEW WORDS ABOUT BACKFLASH
Definition:
In a vaporization injection, a phenomenon wherein a portion of the sample expands beyond the boundary of the injection port liner towards the septum face and incoming gas line
CLP Pesticides B-0447March 8, 2001
BACKFLASHEffects
• Ghost peaks
• Erratic quantitation
• Poor accuracy
CLP Pesticides B-0447March 8, 2001
BACKFLASHCauses
• Highly volatile solvent
• Excessive inlet temperature
• Excessive injection volume
• Small liner volume
CLP Pesticides B-0447March 8, 2001
BACKFLASHPreventative Measures
• Lower inlet temperature
• Less volatile solvent
• 1-2 µL injection volume
• Chambered liner
• Using pulsed split or pulsed splitless
CLP Pesticides B-0447March 8, 2001
Yes NO!!!
INJECTION PORT LINERSSplitless
CLP Pesticides B-0447March 8, 2001
INLET DISCRIMINATION
• Injected sample Sample into the column
• Due to compound volatility differences
• Higher volatility = More into the column
CLP Pesticides B-0447March 8, 2001
INLET DISCRIMINATIONIn Pesticide Analyses
Generally inlet discrimination is not a concern for
pesticide analyses when utilizing appropriate GC
inlets such as Splitless, Pulsed Splitless and PTV
CLP Pesticides B-0447March 8, 2001
OTHER INLET CONSIDERATIONS
• Silylation
• Using glass wool
• Cleaning and reusing liners
CLP Pesticides B-0447March 8, 2001
A FEW WORDS ABOUT RESIDUES
• Semivolatile vs nonvolatile
• Worst offenders
• Sample prep
CLP Pesticides B-0447March 8, 2001
GUARD COLUMNSUse 0.53mm Or 0.32mm ID Deactivated Tubing
• Fit all inlets
• Enhanced deposition of residues
• Easy to work with
• Minimum 1 meter
CLP Pesticides B-0447March 8, 2001
Usually 1-5 meters long and same diameter as the column
Detector
DeactivatedFused SilicaTubing
UnionColumn
Injector
GUARD COLUMNTraps non-volatile sample residues
CLP Pesticides B-0447March 8, 2001
COLUMN CONNECTORSThermal Mass, Inertness, Seal Integrity
• Stainless steel
• VU-Tight
• Press-fits
• Integral guard columns (Duraguard)
CLP Pesticides B-0447March 8, 2001
Let’s Be SensitiveANALYTE DETECTION
CLP Pesticides B-0447March 8, 2001
The minimum amount of an analyte that can still be confidently identified as a peak (S/N > 4)
SENSITIVITYAnalytical Definition
CLP Pesticides B-0447March 8, 2001
SENSITIVITYAnalytical Definition
• Detector Sensitivity: No sample influence (standard)
• Method Sensitivity: Matrix influence (sample)
CLP Pesticides B-0447March 8, 2001
N
S
N
S x
"Normal" response
1. High noise, detector, background
2. Absorption, breakdown of analyte
3. Sample prep losses
"Improved" response
1. High quality circuit components and reagents
2. Inertness
3. Method optimization
RESPONSE TO A CONSTANT ANALYTE AMOUNT2 Cases
HP 6890 Series Micro-ECD Design
Comparison of HP 6890 ECDs
Standard
< 0.040
> 10^4
no spec.
5 - 80
5
Micro-ECD
< 0.008
> 5 x 10^5
> 5 x 10^4
1 - 500
50
pg/sec lindane
lindane
lindane
pg (ppb)
Hz
MDL
Dynamic range
Linear range
Linear Range, CLP pesticides
Maximum data rate
CLP Pesticides B-0447March 8, 2001
Break Number 1
For Questions and Answers
Press *1 on Your Phone to
Ask a Question
CLP Pesticides B-0447March 8, 2001
LETS GET TO THE CHROMATOGRAPHY
CLP Pesticides B-0447March 8, 2001
WHAT EXACTLY ARE WE TRYING TO ACHIEVE
• The best resolution possible
• Minimize analysis time
• Get MDLs as low as possible
CLP Pesticides B-0447March 8, 2001
RESOLUTION VS SEPARATION
• Separation: Time between the 2 peaks
• Resolution: Describes how well 2 peaks are separated with regard to their widths
CLP Pesticides B-0447March 8, 2001
10 11 12 13 14 15
10 11 12 13 14 15
11.24 12.72
(tm = 95.5)
(tm = 95.5)
11.6111.14
WHICH PAIR OF SOLUTES HAVE BETTER SEPARATION?
CLP Pesticides B-0447March 8, 2001
10 11 12 13 14 15
10 11 12 13 14 15
11.24 12.72
11.6111.14
Better Separation
= 1.17 Rs = 0.6
Better Resolution
= 1.05 Rs = 2.7
K = 7.00K = 6.07
K = 6.00 K = 6.30
RESOLUTION VS SEPARATION
CLP Pesticides B-0447March 8, 2001
RESOLUTION AND ANALYSIS TIME
• Improving resolution often results in the opportunity to shorten analysis times
• Many variables can affect resolution
CLP Pesticides B-0447March 8, 2001
N = (L, rc)
k = (T, df, rc)
= (T, phase)
RN kks =
4 11
RESOLUTION
CLP Pesticides B-0447March 8, 2001
The column must provide sufficient retention of the early eluting compounds without excessive retention of the late eluting compounds
IMPROVING RESOLUTION
Retention
CLP Pesticides B-0447March 8, 2001
IMPROVING RESOLUTIONFilm Thickness
Decreasing Film Thickness Results In:
• Increased efficiency
• Elution of analytes at lower temperatures
• Decreased analysis time
• Decreased bleed interference
• Increased column activity
• Decreased capacity
CLP Pesticides B-0447March 8, 2001
IMPROVING RESOLUTIONEfficiency
• High column efficiency is necessary to resolve large numbers of compounds
• Improperly operated injectors and/or improperly optimized carrier gas can result in efficiency losses
CLP Pesticides B-0447March 8, 2001
IMPROVING RESOLUTIONColumn Length
Increasing Column Length Results In:
• Increased efficiency
• Increased analysis time
• Increased bleed
• Big increase in cost
CLP Pesticides B-0447March 8, 2001
IMPROVING RESOLUTION Column Inner Diameter
Decreasing Column Inner Diameter Results In:
• Increased efficiency
• Increased head pressure
• Decreased capacity
• Decreased carrier gas flow rates
CLP Pesticides B-0447March 8, 2001
IMPROVING RESOLUTION Stationary Phase
• Stationary phase selectivity has the largest impact on separation, thus resolution
• Optimization of stationary phase selectivity should be approached cautiously
CLP Pesticides B-0447March 8, 2001
For Years Environmental Laboratories Have Suffered With Pesticide Analyses Using Non-
Optimal Stationary Phases
CLP Pesticides B-0447March 8, 2001
TRADITIONAL STATIONARY PHASES
•5% Phenyl-methylpolysiloxane
•35-50% Phenyl-methylpolysiloxane
•Trifluoropropyl-methylpolysiloxane
•14% Cyanopropylphenyl-methylpolysiloxane
CLP Pesticides B-0447March 8, 2001
PROBLEMS WITH TRADITIONAL PHASES
•Long analysis times (Over 30 minutes!)
•High bleed resulting in decreased sensitivity
•Poor resolution and confirmation capabilities
•Poor inertness of some phases
CLP Pesticides B-0447March 8, 2001
TRADITIONAL PESTICIDE COLUMNS
High Bleed at 280°C
0 10 20 30 40Time (min.)
0 10 20 30 40Time (min.)
608 Type Phase
1701 Type Phase
Coelution of peaks 10 & 11
40 min
Poor peak shape for Endrin aldehyde
CLP Pesticides B-0447March 8, 2001
EFFORTS TO IMPROVE PESTICIDE ANALYSES
Application Specific Phases
Stationary phases designed with a primary focus placed upon maximizing separation () for a specific group of target analytes
CLP Pesticides B-0447March 8, 2001
APPLICATION SPECIFIC PHASES
C2H4CF
CH3m n
Si SiO O
3
Phase 1Trifluoropropyl-dimethylpolysiloxane
3CH
CH3
m n
Si SiO O
C2H4CF
CH
3
3
Common CLP Pesticide Phases
Phase 2Trifluoropropyl-diphenyl-dimethylpolysiloxane
Dimethyl functionality of Phase 2 not shown
CLP Pesticides B-0447March 8, 2001
EFFECT OF PHASE POLARITY
Polarity
Thermal Stability
CLP Pesticides B-0447March 8, 2001
DRAWBACKS OF APPLICATION SPECIFIC PHASES
•Limited thermal stability resulting in longer analysis times and hampered sensitivity
•Excessive column conditioning times leading to increased column activity
•Decreased column lifetimes
CLP Pesticides B-0447March 8, 2001
PHASES DESIGNED WITH OPTIMUM PERFORMANCE IN MINDArylene Phase Technology
CLP Pesticides B-0447March 8, 2001
CH
CH3
CH3
CH3
CH3
CH3
CH3
Si
Si
Si
Si
O
O
O
OBiphenyl
CH3
CH3
CH3
CH3 CH
3
CH3 CH
3
3
Si
Si
Si
SiO
O
Arylene
LOW BLEED STATIONARY PHASESArylene Structure
CLP Pesticides B-0447March 8, 2001
DB-35ms
DB-35
8.00 10.00 12.00 14.00 16.00 18.00 20.00
8.00 10.00 12.00 14.00 16.00 18.00 20.00
320°C
300°C
CLP Pesticides Analysis
LOW BLEED STATIONARY PHASESDB-35ms vs. DB-35
CLP Pesticides B-0447March 8, 2001
STATIONARY PHASE AFFECT ON SENSITIVITY
Both 30 m x 0.25 mm I.D., x 0.25 µm5 ng Decachlorobiphenyl
S/N =10 S/N =3
DB-35ms, 320°C DB-35, 300°C
CLP Pesticides B-0447March 8, 2001
LOW BLEED STATIONARY PHASESSpectral Purity
m/z-->
51
78 96 119 144
207
197
214
250
253
286
315
346
356 377
405
428
455470
498
50 100 150 200 250 300 350 400 450 500
Primary IonDB-35ms
m/z-->
52
78
96 119
135156
197
207
249
253
281
315 331
356
377
405
428
451
470
498
50 100 150 200 250 300 350 400 450 500
Primary Ion
DB-35
CLP Pesticides B-0447March 8, 2001
BENEFITS OF LOW BLEED STATIONARY PHASES• Improved sensitivity
• Improved spectral purity
• Reduced analysis times
• Potential increase in column lifetimes
• Rigorous inertness testing
• Less Detector maintenance
CLP Pesticides B-0447March 8, 2001
Break Number 1
For Questions and Answers
Press *1 on Your Phone to
Ask a Question
CLP Pesticides B-0447March 8, 2001
DEMANDING PESTICIDE APPLICATIONS BENEFIT FROM ADVANCED PHASE TECHNOLOGY
DB-35ms and DB-XLB capillary columns contain second generation arylene stationary phases giving them enhanced thermal stability
CLP Pesticides B-0447March 8, 2001
DB-35MS PRIMARY COLUMNCLP Pesticides
•Should provide adequate resolution of all 22 CLP Pesticides and surrogates
•Short analysis time ideal
•Easily realized analysis conditions preferable
•Should give minimal column bleed
CLP Pesticides B-0447March 8, 2001
1. Tetrachloro-m-xylene (SS)2. -BHC3. -BHC4. -BHC5. Heptachl
or6. -BHC7. Aldri
n8. Heptachlor
epoxide9. -
Chlordane
10. -Chlordane
11. Endosulfan I12. 4,4’-DDE13. Dieldrin14. Endrin15. 4,4’-DDD16. Endosulfan II17
. 4,4’-DDT18
. Endrin
aldehyde19. Endosulfan
sulfate20. Methoxychlor21. Endrin
ketone 22.
Decachlorobiphenyl (SS)
CLP PESTICIDESColumn: DB-35ms
30 m x 0.32 mm I.D., 0.25 µmP/N: 123-3832
Carrier: Helium at 45 cm/sec (EPC in constant flow mode)Oven: 110°C for 0.5 min
110-320°C at 15°C/min320°C for 2 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
min2 4 6 8 10 12 14 16
6 10 14 168 12
1
2
3
4
56
7
89 10
11
1213
14
15 16
17
18 19
20
21
22
Short Analysis Time
Low Bleed At 320ºC
CLP Pesticides B-0447March 8, 2001
DB-XLB CONFIRMATION COLUMNWith Regard To The Primary Column
•Should resolve any co-elutions
•Needs to provide adequate resolving power when run under the same analytical conditions
•Should possess at least equivalent thermal stability
•Should provide a different relative elution pattern
•Ideally, elution order changes should be realized
CLP Pesticides B-0447March 8, 2001
CLP PESTICIDESColumn: DB-XLB
30 m x 0.32 mm I.D., 0.50 µmP/N: 123-1236
Carrier: Helium at 45 cm/sec (EPC in constant flow mode)Oven: 110°C for 0.5 min
110-320°C at 15°C/min320°C for 2 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
min2 4 6 8 10 12 14 166 10 14 168 12
11
2
3
4
6
5
7
88910
11
12
14
13
15
16
19
20
21
22
18
17
100% Confirmation In Under 16 Minutes
1. Tetrachloro-m-xylene (SS)2. -BHC3. -BHC4. -BHC5. Heptachl
or6. -BHC7. Aldri
n8. Heptachlor
epoxide9. -
Chlordane
10. -Chlordane
11. Endosulfan I12. 4,4’-DDE13. Dieldrin14. Endrin15. 4,4’-DDD16. Endosulfan II17
. 4,4’-DDT18
. Endrin
aldehyde19. Endosulfan
sulfate20. Methoxychlor21. Endrin
ketone 22.
Decachlorobiphenyl (SS)
CLP Pesticides B-0447March 8, 2001
OPTIMIZING FOR SPEED WITHOUT SACRIFICING ANALYTE RESOLUTION
•Use Hydrogen as the carrier gas
•Set linear velocity at 65 cm/sec (3.4 mL/min)
•Increase oven ramp rate from 15°/min to 25°/min
CLP Pesticides B-0447March 8, 2001
1. Tetrachloro-m-xylene (SS)2. -BHC3. -BHC4. -BHC5. Heptachlor6. -BHC7. Aldrin8. Heptachlor epoxide9. -Chlordane
10. -Chlordane11. Endosulfan I12. 4,4’-DDE13. Dieldrin14. Endrin15. 4,4’-DDD16. Endosulfan II17. 4,4’-DDT18. Endrin aldehyde19. Endosulfan sulfate20. Methoxychlor21. Endrin ketone
22. Decachlorobiphenyl
CLP PESTICIDES USING HYDROGEN CARRIER GASColumn: DB-35ms
30 m x 0.32 mm I.D., 0.25 µmP/N: 123-3832
3 4 5 6 7 8 9 1010
1
2
3
4
5 6
7
8 910
11
12
13
14
1516
17
19
20
21
22
4 85 6 7 9
18
Baseline Resolution In Under 9 Minutes!
Carrier: Hydrogen at 65 cm/sec (EPC in constant flow mode)Oven: 110°C for 0.5 min
110-320°C at 25°C/min320°C for 2 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
CLP Pesticides B-0447March 8, 2001
1. Tetrachloro-m-xylene (SS)2. -BHC3. -BHC4. -BHC5. Heptachlor6. -BHC7. Aldrin8. Heptachlor epoxide9. -Chlordane
10. -Chlordane11. Endosulfan I12. 4,4’-DDE13. Dieldrin14. Endrin15. 4,4’-DDD16. Endosulfan II17. 4,4’-DDT18. Endrin aldehyde19. Endosulfan sulfate20. Methoxychlor21. Endrin ketone
22. Decachlorobiphenyl
Carrier: Hydrogen at 65 cm/sec (EPC in constant flow mode)Oven: 110°C for 0.5 min
110-320°C at 25°C/min320°C for 2 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
3 4 5 6 7 8 9 10
1
2
3
4
6 5
7
8
9
10
11
12
14
13
15
16
19
20
2122
17
18
105 7 96 8
CLP PESTICIDES USING HYDROGEN CARRIER GASColumn: DB-XLB
30 m x 0.32 mm I.D., 0.5 µmP/N: 123-1236
Excellent Resolution And Confirmation In Under 10 Minutes
CLP Pesticides B-0447March 8, 2001
OTHER COMMON ECD APPLICATIONS THAT THIS DUAL COLUMN SHOULD BE ABLE TO PERFORM
•EPA 8151A (Phenoxy acid herbicides)
•EPA 508.1 (Pesticides)
•EPA 552.2 (Haloacetic acids)
•EPA 8082 (PCB congeners and Aroclors)
CLP Pesticides B-0447March 8, 2001
2 4 6 8 10 12 14 16Time (min)
EPA 8151A PHENOXY ACID HERBICIDES
DB-XLB
C785
2 4 6 8 10 12 14 16Time (min)
DB-35ms
C784
2 4 6 8 10 12 14 16
12
34
5
6 7
8
9 10
11
12 131415
16 17
18
19 20
min2 4 6 8 10 12 14 16
1 23 4
5
6 7
8
910
11
121314 15
16 17
1819 20
CLP Pesticides B-0447March 8, 2001
EPA 8151A PHENOXY ACID HERBICIDESDB-35ms
30m x 0.32 mm I.D., 0.25 µmP/N: 123-3832
DB-XLB30m x 0.32 mm I.D., 0.50 µmP/N: 123-1236
Carrier: Helium at 45 cm/sec (EPC in constant flow mode)Oven: 50°C for 0.5 min
50-100°C at 25°C/min100-320°C at 12°C/min320°C for 2 min
Injector: Splitless, 250°C
30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
1. Dalapon2. 3,5-Dichlorobenzoic acid3. 4-Nitrophenol4. Methyl-2,4-dichlorophenylacetate (SS)5. Dicamba6. MCPP7. MCPA8. 4,4´, Dibromooctafluorobiphenyl (IS)9. Dichloroprop
10. 2,4-D11. Pentachlorophenol12. 2,4,5-T,P13. 2,4,5-T14. Chloramben15. Dinoseb16. 2,4-DB17. Bentazone18. DCPA19. Picloram20. Acifluorofen
CLP Pesticides B-0447March 8, 2001
EPA 508.1 PESTICIDES
DB-XLB
DB-35ms
8 10 12 14 16 18 20 22 24Time (min)
min8 10 12 14 16 18 20 22 24
min8 10 12 14 16 18 20 22 24 8 10 12 14 16 18 20 22 24
Time (min)
11
2
3
4
5
6
7
8
9,10
11
12
14
13
1516
17
18
19
20
21,22
23
2425
2627
28
29
303132
33
34
35
3738
36
11
2
3
4
5
67
8,10
9,11
12
14
13
1516
17
18
19
2022
23
2425
262728
29
3031
32
33
34
35
3738
3621
C787
C786
CLP Pesticides B-0447March 8, 2001
EPA 508.1 PESTICIDES
DB-35ms30m x 0.32 mm I.D., 0.25 µmP/N: 123-3832
DB-XLB30m x 0.32 mm I.D., 0.50 µmP/N: 123-1236
Carrier: Helium at 45 cm/sec (EPC in constant flow mode)Oven: 75°C for 0.5 min
75-300°C at 10°C/min300°C for 2 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
1. Hexachloropentadiene2. Etridiazole3. Chloroneb4. Trifluralin5. Propachlor6. Hexachlorobenzene7. a-BHC8. Atrazine9. Pentachloronitrobenzene (IS)
10. Simazine11. g-BHC12. b-BHC13. Hepatachlor14. Alachlor15. d-BHC16. Chlorothalonil17. Aldrin18. Metribuzin19. Metolachlor20. DCPA
21. 4,4´-Dibromobiphenyl (SS)22. Heptachlor epoxide23. Cyanazine24. g-Chlordane25. a-Chlordane26. Endosulfan I27. 4,4´-DDE28. Dieldrin29. Chlorobenzilate30. Endrin31. 4,4´-DDD32. Endosulfan II33. 4,4´-DDT34. Endrin Aldehyde35. Endosulfan sulfate36. Methoxychlor37. cis-Permethrin38. Trans-Permethrin
CLP Pesticides B-0447March 8, 2001
min4 5 6 7 8
EPA 552.2 HALOACETIC ACIDS
DB-XLB
DB-35ms
4 5 6 7 8Time (min)
min4 5 6 7 8
14
25
6
78,9
10
11
12
4 5 6 7 8Time (min)
1 4
2,3
5
6
7
8
9
10
11
12
3
C789
C788
CLP Pesticides B-0447March 8, 2001
EPA 552.2 HALOACETIC ACIDS
1. Chloroacetic acid2. Bromoacetic acid3. Dichloroacetic acid4. Dalapon5. Trichloroacetic acid6. 1,2,3-Trichloropropane (IS)7. Bromochloroacetic acid8. Bromodichloroacetic acid9. Dibromoacetic acid
10. 2,3-Dibromopropionic acid (SS)11. Chlorodibromoacetic acid12. Tribromoacetic acid
DB-35ms30m x 0.32 mm I.D., 0.25 µmP/N: 123-3832
DB-XLB30m x 0.32 mm I.D., 0.50 µmP/N: 123-1236
Carrier: Helium at 45 cm/sec (EPC in constant flow mode)Oven: 40°C for 0.5 min
40-200°C at 15°C/min200°C for 2 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
CLP Pesticides B-0447March 8, 2001
EPA 8082 PCB CONGENERS
DB-XLB
DB-35ms
6 8 10 12 14 16Time (min)
min6 8 10 12 14 16
min6 8 10 12 14 16
6 8 10 12 14 16Time (min)
1
2
34
56
78
9
101112
13
1415
16
1718
19
20
21
1
2
3
4
56
78
9
101112
13
1415
16
1718
19
20
21
C791
C790
CLP Pesticides B-0447March 8, 2001
EPA 8082 PCB CONGENERS1. IUPAC 12. Tetrachloro-m-xylene (IS/SS)3. IUPAC 54. IUPAC 185. IUPAC 316. IUPAC 527. IUPAC 448. IUPAC 669. IUPAC 101
10. IUPAC 8711. IUPAC 11012. IUPAC 15113. IUPAC 15314. IUPAC 14115. IUPAC 13716. IUPAC 18717. IUPAC 18318. IUPAC 18019. IUPAC 17020. IUPAC 20621. Decachlorobiphenyl (IS/SS)
DB-35ms30m x 0.32 mm I.D., 0.25 µmP/N: 123-3832
DB-XLB30m x 0.32 mm I.D., 0.50 µmP/N: 123-1236
Carrier: Helium at 45 cm/sec (EPC in constant flow mode)Oven: 110°C for 0.5 min
110-320°C at 15°C/min320°C for 5 min
Injector: Splitless, 250°C30 sec purge activation time50 pg per component
Detector: µECD, 350°CNitrogen makeup gas (column + makeup flow = 30 mL/min constant flow)
CLP Pesticides B-0447March 8, 2001
WHAT DOES THIS MEAN FOR ENVIRONMENTAL LABORATORIES?
•Improved productivity
•Increased sample throughput
•Superior results
•Total solution for ECD systems
CLP Pesticides B-0447March 8, 2001
CONCLUSION
•Excellent selectivity
•Exceptional inertness
•Superb thermal stability
DB-35ms and DB-XLB provide excellent resolution and confirmation for CLP Pesticides and other critical GC/ECD methods providing total performance:
CLP Pesticides B-0447March 8, 2001
CONCLUSION
When optimized columns are combined with effective injector setup and detector design improved testing results are realized while sample throughput and laboratory efficiency are improved dramatically
CLP Pesticides B-0447March 8, 2001
AGILENT’S J&W TECHNICAL SUPPORT
It’s Free!(800) 552-0413 (phone)(916) 985-7888 (phone)(916) 608-1964 (FAX)
web site:www.agilent.com/chem
CLP Pesticides B-0447March 8, 2001
Wrap-up E-Seminar Questions
Thank you for attending Agilent e-Seminars.
Our e-Seminar schedule is expanding, please
check our website frequently at:
www.agilent.com/chem