Calibration software for the HADES electromagnetic calorimeter (EMC)
description
Transcript of Calibration software for the HADES electromagnetic calorimeter (EMC)
![Page 1: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/1.jpg)
Calibration software for the HADES electromagnetic calorimeter (EMC)
Dimitar MihaylovExcellence Cluster ‘Universe’, TU Munich
HADES collaboration meeting XXVGSI, November 2012
![Page 2: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/2.jpg)
2
Overview
The proposed electromagnetic calorimeter (EMC)
Data analysis
Motivation for a calibration procedure
Calibration procedure – Mathematical model
Calibration procedure – Realization
Summary
Ongoing improvements and outlook
![Page 3: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/3.jpg)
3
Overview of EMC
Parameters and benefits:• Identification of light neutral pseudoscalar mesons (π0 and η).
• Good e/π-separation at high momenta.
• Six sectors, each containing 163 lead-glass blocks.
• High refractive index of the lead-glass blocks (n = 1.708).
• Radiation length (X0) of 2.51 cm.
• Almost full azimuthal coverage.
• The polar angle is covered between 12˚ and 45˚.
• The energy resolution is 𝜎𝐸 𝐸Τ = 6% ඥ𝐸 𝐺𝑒𝑉ΤΤ . Czyzycki et al., arXiv:1109.5550v2
![Page 4: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/4.jpg)
4
Data analysis
Simulation
Reconstruction Output data(reconstr. γ)
Invariant massspectrum (IMS)
Input data(initial γ)
Experiment
• Unaccounted energy losses.
• Undetected photons.
• Reconstruction of a non-existing photon (false signals).
• Large combinatorial background in the IMS (polynomial part of the fit).
![Page 5: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/5.jpg)
5
Overview of the calibration procedure
Fitting the IMS
Calibration of single γ
Final IMS𝑴𝝅 = 𝒎𝝅 ? ?
𝑝Ԧ𝐶 = 𝑀𝑝Ԧ, The general form of the calibration procedure is:
where M is a calibration matrix.
A calibration procedure capable of compensating for those inaccuracies is needed.The calibration procedure will be performed on individual photons.
![Page 6: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/6.jpg)
6
Is only energy calibration enough?
• The error in the energy is significantly larger compared to the error in the polar and azimuthal angles.
Distribution of ES/E Distribution of θS/θ Distribution of ϕS/ϕ
• Only spherical coordinates will be used.
• Comparison between the initially simulated and reconstructed values of the components of the photon momentum.
![Page 7: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/7.jpg)
7
Choice of calibration function
The general form of the calibration function is:
The exact form of the expressions Fi can be determined by reconstructing simulated data and analyzing the resulting deviations.
𝑓ሺ𝐸,𝜃,𝜑ሻ= 𝑒𝑥𝑝൝ 𝐴𝑖𝐹𝑖(𝐸,𝜃,𝜑)𝑖 ൡ
D.J. Tanner, MSc thesis, University of Manchester, Oct 1998
The general form of the calibration procedure is:
𝐸𝐶 = 𝐸𝑓ሺ𝐸,𝜃,𝜑ሻ
![Page 8: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/8.jpg)
8
Mathematical model
𝝀𝒊 ≔ 𝟏𝟐ሺ𝑭𝟏𝒊 + 𝑭𝟐𝒊ሻ • Definition of some variables for photon pairs
𝝆≔ 𝒍𝒏𝑴𝜸𝜸 − 𝒍𝒏𝒎𝝅 = 𝒍𝒏𝑴𝜸𝜸𝒎𝝅
• The calibration is performed by minimizing the function
𝓛≔ ൣ�𝒍𝒏𝑴𝜸𝜸𝒋− 𝒍𝒏𝒎𝝅൧𝟐𝑵
𝒋=𝟏 = 𝝆𝒋𝟐𝑵
𝒋=𝟏
𝑹𝒊 ≔ 𝟏𝟐 𝝏𝓛𝝏𝑨𝒊 = 𝝆𝒋𝝀𝒊𝒋𝑵
𝒋=𝟏 = 𝟎
𝑀𝛾𝛾 = ඥ2𝐸𝐶1𝐸𝐶2(1− 𝑐𝑜𝑠𝜂)
![Page 9: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/9.jpg)
9
Energy dependence
𝐸𝑓𝑖𝑡1 = 𝑒𝑥𝑝ሺ𝑃0 + 𝑃1𝑙𝑛𝐸+ 𝑃2𝑙𝑛2𝐸ሻ with 𝜒2 𝑁𝐷𝐹Τ = 5.34
𝐸𝑓𝑖𝑡2 = 𝑒𝑥𝑝ሺ𝑃0 + 𝑃1𝑙𝑛𝐸+ 𝑃2𝑙𝑛2𝐸+ 𝑃3𝑙𝑛3𝐸ሻ with 𝜒2 𝑁𝐷𝐹Τ = 2.96
The energy dependence of the calibration function can be well described with function of the type:
𝑓𝐸ሺ𝐸ሻ= 𝑒𝑥𝑝൝ 𝑃𝑖𝑙𝑛𝑖𝐸)𝑖 ൡ
𝑓ሺ𝐸,𝜃,𝜑ሻ= 𝑒𝑥𝑝൝ 𝐴𝑖𝐹𝑖(𝐸,𝜃,𝜑)𝑖 ൡ
![Page 10: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/10.jpg)
10
Result of the calibration procedure
The results below are obtained with the following calibration functions: 𝑓1 = 𝑒𝑥𝑝ሺ𝐴0 + 𝐴1𝑙𝑛𝐸+ 𝐴2𝑙𝑛2𝐸ሻ, 𝑓2 = 𝑒𝑥𝑝൛𝐴0 + 𝐴1𝑙𝑛𝐸+ 𝐴2𝑙𝑛2𝐸+ 𝐴3𝑙𝑛3𝐸+ 𝐴4𝑐𝑜𝑠൫𝜃෨𝜔𝜃𝑐𝑜𝑠𝜑 − 𝜃𝑠ℎ𝑖𝑓𝑡൯+ 𝐴5𝑐𝑜𝑠൫𝜑𝜔𝜑𝑠𝑖𝑛𝜃− 𝜑𝑠ℎ𝑖𝑓𝑡൯ൟ, 𝑓3 = 𝑒𝑥𝑝ሺ𝐴0 + 𝐴1𝑙𝑛𝐸+ 𝐴2𝑙𝑛2𝐸+ 𝐴3𝑙𝑛3𝐸ሻ.
![Page 11: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/11.jpg)
11
Accuracy
Example of the calibration functions f1, f2 and f3 as a function of the energy.
Error in the energy of single photons before and after calibration.
![Page 12: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/12.jpg)
12
Accuracy
Distribution of the error in the energy before and after calibration.
Position of the η peak after calibration.
![Page 13: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/13.jpg)
13
Implementation in C++
• The described calibration procedure has been implemented in a standalone C++ program called IMS-expert.
• Only standard C++ and ROOT libraries have been used.
• IMS-expert allows the user to perform the calibration with different types of functions.
• After a calibration has been performed, the calibration parameters are saved and can be used for further calibrations of different data.
![Page 14: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/14.jpg)
14
Summary
• The presented calibration procedure is able to accurately correct the IMS for invariant masses in the range of the mass of π0 (135 MeV/c2) and of η (548 MeV/c2).
• The systematic shift of the IMS is compensated by the calibration procedure but the statistical error, which is mostly related to the energy resolution, remains approximately the same.
• The error in the energy of individual photons decreases, but it still remains significant.
• The energy dependence of the calibration function has the biggest influence on the final result.
• The calibration procedure is implemented in a standalone program called IMS-expert and can be used for any EMC.
![Page 15: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/15.jpg)
15
Ongoing improvements
𝐸𝐶 = 𝐸𝑓𝐸ሺ𝐸,𝜃,𝜑ሻ 𝜃𝐶 = 𝜃+𝑓𝜃ሺ𝐸,𝜃,𝜑ሻ 𝜑𝐶 = 𝜑+𝑓𝜑ሺ𝐸,𝜃,𝜑ሻ
Look-up table approach
Analytical angle calibration
![Page 16: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/16.jpg)
Thank You For Your Attention
![Page 17: Calibration software for the HADES electromagnetic calorimeter (EMC)](https://reader036.fdocuments.us/reader036/viewer/2022062811/568161b8550346895dd1877f/html5/thumbnails/17.jpg)
17
Backup slide
If not all Ri=0, then:𝑹𝒊 + 𝜹𝑹𝒊 = 𝟎
For small δRi𝜹𝑹𝒊𝜹𝑨𝒍 ≈ 𝝏𝑹𝒊𝝏𝑨𝒍 = 𝑮𝒊𝒍
𝑮𝒊𝒍 ≔ 𝝏𝑹𝒊𝝏𝑨𝒍 = 𝟏𝟐 𝝏𝟐𝓛𝝏𝑨𝒊𝝏𝑨𝒍 = 𝝀𝒍𝒋𝝀𝒊𝒋𝑵
𝒋=𝟏 𝑹𝒊 ≔ 𝟏𝟐 𝝏𝓛𝝏𝑨𝒊 = 𝝆𝒋𝝀𝒊𝒋𝑵
𝒋=𝟏 = 𝟎
𝑮𝒊𝒍𝜹𝑨𝒍 = 𝜹𝑹𝒊 −𝑮𝒊𝒍𝜹𝑨𝒍 = 𝑹𝒊 𝜹𝑨𝒍 = −𝑮𝒊𝒍−𝟏𝑹𝒊
𝑓ሺ𝐸,𝜃,𝜑ሻ= 𝑒𝑥𝑝൝ 𝐴𝑖𝐹𝑖(𝐸,𝜃,𝜑)𝑖 ൡ