Brachytherapy
Dosimetry in the
ClinicMargaret Bidmead
Royal Marsden NHS Trust
![Page 2: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/2.jpg)
Two Curies
Bequerel
![Page 3: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/3.jpg)
Brachytherapy
What is it?
“Sealed radioactive source(s) placed directly
into, or immediately adjacent to the volume
to be treated.”
Treatment of melanoma, 1910, Sydney, Australia
![Page 4: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/4.jpg)
Uses for Radium
1904
“Guaranteed to
Contain Real
Refined Radium
And to be
Perfectly
Harmless”
Self-
luminous
paint-
stopped
in 1950’s
Radium Girls-
Occupational disease
Labour law
![Page 5: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/5.jpg)
Brachytherapy Statistics:
4 % of all new cancer cases receive brachytherapy (12% in USA)
Gynae - 68%
Head & Neck - 5%
Bronchus - 7%
Rectum - 3%
prostate-25%
![Page 6: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/6.jpg)
Types of Brachytherapy
IntracavitaryCervix
vagina
oesophagus/bronchus
![Page 7: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/7.jpg)
Why do we use it?
localised treatment, i.e. high dose to the target
volume, whilst sparing adjacent normal tissue, and
minimal whole body dose
![Page 8: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/8.jpg)
Types of BrachytherapyInterstitial Prostate
Head and neck
Breast
Rectum
![Page 9: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/9.jpg)
Types of Brachytherapy
Surface Moulds
Angiosarcoma
Melanoma
Basal Cell carcinoma
Recurrent disease
![Page 10: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/10.jpg)
Implantation Rules
System of work for source implantation
How best to cover defined volume
How best to minimize dose to organs at
risk
What is practically achievable
What AKR of source(s) available
![Page 11: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/11.jpg)
Clinical Dosimetry Systems
Intracavitary ……Historically Modified
Manchester System, Point A: now volumetric
Interstitial with Ir….Paris System
Line sources…..prescription at distance from a
plane
Permanent seeds……volumetric isodose
coverage (D90)
Moulds…….different Manchester System,
treatment at a specified distance
![Page 12: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/12.jpg)
Dosimetry Overview
Purpose of source calibration and dosimetry?
To verify source strength independently and improve precision.
To achieve national and international agreement
(DOH requirement: check measured source strength against manufacturer’s certificate)
Source specification:
AKR in air @ 1m - RAKR
![Page 13: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/13.jpg)
Brachytherapy Source Specification
Reference Quantity Definition Units
Reference
Air Kerma
Rate
(RAKR)
Air kerma rate due to
photons of energy
greater than , at a
point located at a
distance of 1m from
the source centre, in
vacuo.
Gyhr-1
at 1m
AAPM TG 43
Update:
Med. Phys. 31
(3), March
2004
Air-kerma
strength
Sk
The air kerma rate,
in vacuo, and due to
photons of energy
greater than , at
distance d,
multiplied by the
square of the
distance.
Gym2hr-1
(U)
![Page 14: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/14.jpg)
Source Strength
Past:
Activity
Effective
Activity
mg Ra
equivalent
Present:
Air Kerma Rate
Under standard
conditions
NB: Link to algorithm
in TPS or MC model
of each source
Uncertainties in conversion factors
![Page 15: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/15.jpg)
Reference Standards & Traceability
Direct calibration method:
source or calibrator has been calibrated at a national standards laboratory (NPL in the UK, NIST in the USA).
Secondary method:
source is calibrated in comparison with a source of the same design and comparable strength, which has direct traceability.
![Page 16: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/16.jpg)
Reference Standards & Traceability
Remote traceability : if the user relies on the manufacturer’s calibration as the only standard. (not recommended)
![Page 17: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/17.jpg)
Calibration Methods
Ion chamber in air: AKR at known
distance from the source
Calibrated re-entrant well chamber
Ion chamber and source inside a
solid phantom. (2nd check?)
![Page 18: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/18.jpg)
Linear response with source strength.
Accurately measured energy response.
Response dependent on geometric
configuration (size, shape, orientation
and position), filtration
and encapsulation of
source.
Properties of Re-entrant Well
Chambers
![Page 19: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/19.jpg)
Properties of Re-entrant Well
Chambers
Reproducibility better than 2%.
Proven long-term stability (+-1%)
Signal-to-noise ratio > 100:1.
Collection efficiency better than
99%.
Easy to set up and use
![Page 20: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/20.jpg)
Issues With Well Chambers
Effect of source type used for standard calibration
Sweet spot of chamber
Saturation effects of well chambers at high currents
Temperature stabilisation
Method of constancy check
Second, independent check of source calibration
![Page 21: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/21.jpg)
HDR/PDR Afterloading Systems
Nucletron HDR
Micro Selectron
Ir192 (370Gbq)
Ir192 (370GBq)
Co60 (74GBq)
source availableFlexitron
Dual source
Ir192
(Co60/Yb169)
Varian / GammaMed
Single Ir192 source
![Page 22: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/22.jpg)
Currents from HDR sources
The BS EN 60731:1997 standard sets a "maximum input current" for a secondary standard dosemeter at 5 nanoamp
currents of 10-30nA per Citherefore possibly 500nA from a very hot HDR source
![Page 23: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/23.jpg)
RAKR Stability Analysis
Data collected by Chris Lee
0.980
0.990
1.000
1.010
1.020
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Source number
Ratio of well chamber to Farmer measurement
Reference air kerma rate (RAKR) comparison between
a Standard Imaging 1000 Plus well chamber calibrated using the new COP,
and a Farmer chamber, calibrated at 280 kVp according to the old 1992 BIR/IPSM
recommendations.
![Page 24: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/24.jpg)
RAKR PTW 22004 Well chamber, calibrated according to the new COP,
and an NE 2571 Farmer Chamber , calibrated according to the IPEM 1992
Recommendations,
manufacturer’s source certificate for a Nucletron microSelectron-v1 (classic) HDR
192Ir source
0.98
0.99
1.00
1.01
1.02
Aug 04 Nov 04 Feb 05 May 05 Nov 05 Feb 06 May 06 Aug 06 Nov 06
Ratio of Well Chamber to Farmer
MeasurementsRatio of Well Chamber Measurement to
Certificate Value
![Page 25: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/25.jpg)
Constancy checks of well-
chambers (long t1/2 source)
Cs 137 source, available but custom made jig for well chamber
required.
Activity and associated current to be selected
Cs137HTSL Cs
0.980
0.985
0.990
0.995
1.000
1.005
1.010
1.015
1.020
01-S
ep-0
7
01-D
ec-0
7
01-M
ar-
08
01-J
un-0
8
31-A
ug-0
8
01-D
ec-0
8
02-M
ar-
09
01-J
un-0
9
01-S
ep-0
9
Date
Measu
rem
en
t vari
ati
on
no
rmali
sed
to
th
e
refe
ren
ce
. Error bars give estimates of the 95% confidence intervals
based on the uncertainties of the temperature, pressure
and fluctuations of the measured charge.
![Page 26: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/26.jpg)
Constancy checks of well-
chambers (2)
Calculate ratio for the two measured currents and
compare with the original reference value. Ratio
should agree with reference to within 2%.
If difference greater than 2% should be investigated
Problem may lie with either chamber
![Page 27: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/27.jpg)
Ir-192 Clinical Source
Calibration
Objective:
Determine reference AKR of an HDR
source to ensure that the agreement
with the RAKR stated on the source
certificate agrees within +/- 5%.
+/- 3% should be possible
![Page 28: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/28.jpg)
Ir-192 Source Calibration
When to Calibrate?
Field chamber should be cross calibrated with
secondary standard chamber immediately after
return from NPL
At machine Commissioning.
Following major repair or modification.
Following Source Replacement.
![Page 29: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/29.jpg)
Associated QC
Timer checks, source position and
room safety checks should also be
done following service, source
replacement.
Possibly some applicator
checks(especially rings)
![Page 30: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/30.jpg)
Quality AssuranceRemote Afterloading Systems
1. Source IntegritySources inaccessibleHigh activityWipe source path or transfer cable at machine serviceCheck catheters for activity
2. Source Position CheckFollowing position calibration or after source change. AutoradiographyUse reference jig viewed by in room camera
![Page 31: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/31.jpg)
Quality AssuranceRemote Afterloading Systems
HDR source stepping checks during a single transfer: 10mm steps 1500mm-1470mm
1500mm
1490mm
1480mm
1470mm
![Page 32: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/32.jpg)
Required Measurements for
Calibration
2 independent measurements:A and B
Measure AKR to derive activity.
Uses measured AKR to deliver a set dose.
2 physicists.
2 dosimeter / chamber combinations in ideal situation.
![Page 33: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/33.jpg)
Ir-192 Source Calibration
How?
–Using a re-entrant chamber traceable to NPL
–Prior to measurement leave well chamber connected to
electrometer in treatment room for at least half an hour.
![Page 34: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/34.jpg)
Published 2010
Recommends use of a well chamber
Defines the factor ksg to account for different source types.
No consensus so set to 1.
Comparison with source cert may be used as the second
check if cert comes from a approved and registered
calibration centre.
Well chamber stability check – ratio or check source
Electrometer calibration in terms of current.
Ir192 HDR Source Calibration in ClinicThe IPEM Code of Practice based on the NPL Air Kerma Standard
![Page 35: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/35.jpg)
Set up well chamber 1m from walls and floor on low scatter
surface
Thermal equilibrium
Connect to calibrated electrometer
300V (collecting electrode is +ve wrt outer electrode)
Find position of max response
Measure stable current from at least 3 source transfers
Measure ion recombination using half voltage technique,
typically around 1.001 for 370GBq source
Correct for temp/pressure, ion recombination, source decay
Ir192 HDR Source Calibration in ClinicThe IPEM Code of Practice based on the NPL Air Kerma Standard
METHOD
![Page 36: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/36.jpg)
Ir-192 Source Calibration
Part A: Measure Reference Air Kerma Rate (RAKR) of new source using the NPL calibrated secondary standard well chamber and electrometer, using chamber calibration coefficient
Connect to the HDR unit using the same transfer tube for sake of consistency
Minimize scatter conditions
RKN
RKN
RKN
![Page 37: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/37.jpg)
If the water equivalent thickness of the
tabletop (at least 1 m above floor level) is
between 2.5 mm and 15 mm, the scatter
contribution to the measured ionisation
current is within ±0.1% of the scatter
contribution during calibration of the well
chamber at the NPL.
![Page 38: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/38.jpg)
Position of maximum chamber response
(Sweet Spot) of chamber
Each centre must find its own sweet spot for a particular transfer tube
NPL will also test for sweet spot location
“Peak” is broad (15-20mm) so any positional error (<+-1mm) in source is not significant
Min and Max measurement positions should be at least +-10mm from expected SS position
Range in position should be within +- 2.5mm of previously determined value
![Page 39: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/39.jpg)
Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2
Source
Source Number D36B 1458 7/2/2007
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
1.05
128
0
128
5
129
0
129
5
130
0
130
5
131
0
131
5
132
0
132
5
133
0
133
5
134
0
134
5
135
0
135
5
136
0
136
5
137
0
137
5
138
0
MicroSelectron Source Position in Channel 3
Cu
rren
t (n
A)
PT
W U
nid
os 2
0525
Position
Postion + 1mm
Postion -1mm
![Page 40: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/40.jpg)
Part A
• Send the source to the position of
maximum chamber response where
the NPL calibration is valid.
• Take at least 3 current readings ,
each from separate source transfers
and correct for background. There
should be no significant trend in the
readings.
• Calculate mean corrected current and
its standard deviation (SD< 0.1%)
![Page 41: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/41.jpg)
Ion Recombination
Recommend Attix two voltage method for determination of recombination factor within the COP for users to determine their own factor. Therefore Part A readings repeated with chamber voltage at 150V
Factor ~ 1.001for 300GBq source but does depend on AKR
F H Attix, AAPM radiotherapy dosimetry protocol, Med. Phys. 11, 714 (1984)).
![Page 42: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/42.jpg)
![Page 43: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/43.jpg)
Part B
• Use RAKR measured in Part A to
calculate the time required to deliver a
specified air kerma at 1m. Set time on
HDR unit and measure the delivered air
kerma with an independent chamber
and electrometer
• The time set must be corrected for the
additional air kerma recorded during
source transit
![Page 44: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/44.jpg)
Second “independent” check
Is the AKR from the source certificate a
second check?
Yes in Austria, the Netherlands, Belgium and
Norway! (unless the measurement is >5%
different from the certificate)
AAPM TG-56 (Nath et al 1997) for a
confirmatory check of source strength
![Page 45: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/45.jpg)
Phantom end-to-end check
Small phantom drilled to take a tertiary
instrument (e.g. Farmer chamber) at a
fixed distance of 1cm-5cm from a drilled
channel which carries the source
catheter would be suitable for a “second
check”
(3-5cm probably best for HDR)
This would allow “end-to-end” check
![Page 46: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/46.jpg)
Calibration Outcome
Source accepted if:the activity calculated by AKR measurement is
within +/-5% of that stated on the source certificate.
dose delivery measurement calculated from the RAKR measurement is also within +/-3%.
If source is from a nationally accredited lab then we can use the source certificate supplied by the manufacturer as a “second check”.
![Page 47: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/47.jpg)
However it is suggested that the AKR
measured at the Hospital is used in the
treatment planning system, rather than
the AKR from the source certificate.
Generally do not know the measurement
conditions/parameters of source
certifying institute
![Page 48: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/48.jpg)
Error Frequency
![Page 49: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/49.jpg)
When calculating dosimetry - it is
important to consider:
Method of dose calculation used and standardisation of input data
![Page 50: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/50.jpg)
If activity is used in the TPS or for “convenience” then dose rate is usually calculated via:
Where k is a specific gamma ray constant or an air kerma rate constant
If A is determined from the RAKR, then k is a dummy variable which must be identical to that in the TPS.
How does a TPS calculate dose?
2r
AkD
![Page 51: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/51.jpg)
Update of AAPM Task Group No.43 Report: A
revised AAPM protocol for brachytherapy
dose calculations
Rivard M J, Coursey B M, DeWerd L A,
Hanson W F, Huq M S, Ibbott G S,
Mitch M G, Nath R and Williamson J F
2004 Med. Phys. 31 633-74
![Page 52: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/52.jpg)
the dose rate constant (λ)
radial dose function g(r,) (absorption and
scattering)
geometry factors or source data (spatial
resolution of radioactivity within source)
G (r,Ө)
anisotropy function & factors F (r, Ө)
,],/,, 00 rFrgrGrGSrD k
![Page 53: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/53.jpg)
Schematic diagram illustrating the geometry considered by the Task Group 43
dose calculation formalism for a linear radioactive source (adapted from AAPM
1995).
TG43: Medical Physics 22 (2) 209-235 Feb 1995
Med Phys. 2007 Jun;34(6):2187-205
![Page 54: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/54.jpg)
Cylindrically symmetric source
,],/,, 00 rFrgrGrGSrD k
Air Kerma Strength Dose rate constant: dose rate to water at 1cm
![Page 55: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/55.jpg)
Website for validated MC source data (TG-
43)
Currently accessible through Valencia
University
New data continuously validated by a
BRAPHYQS/AAPM subgroup
Easy access through ESTRO website:
www.estro.org
![Page 56: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/56.jpg)
“Insurance”
Good QA procedures
Second checks
Work instructions
Thorough familiarity with TPS
Communication between multi-
disciplinary staff groups
![Page 57: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/57.jpg)
MRI at time of
Brachytherapy
AFTER:
45Gy EXTB and
200mg/m2 cis-platin
GTV measured
2cm w, 3cmt, 3cmh
![Page 58: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/58.jpg)
6 cm tube
+ medium ovoid
2cm 15
10
10
20 20
AR AL
1.5
1
12 2
HDR
But R < 65Gy
or < 60 % dose to A
![Page 59: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/59.jpg)
Standard loading pattern
Point A
Point A
45 Gy EBRT + 4 x BT:
85 Gyαβ10
60 Gyαβ10
15 mg10 mg10 mg
20
mg
20
mg
Courtesy : Kirisits, Vienna
![Page 60: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/60.jpg)
Standard loading pattern
Courtesy : Kirisits, Vienna
![Page 61: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/61.jpg)
E-lfh website
![Page 62: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/62.jpg)
Patient Results
Before Treatment After Treatment
?
2mths after treatmentPrior to any treatment
40cc HR-CTV
Courtesy S Aldridge and A Winship
![Page 63: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/63.jpg)
Clinical dosimetry issues
Problem; organ motion, brachy applicator motion Multiple imaging sequences and parameters (fusion)
Varying dose rates (complex biologically equivalent doses)
Application of more complex algorithms (TG 186) Acuros and ACE are two commercial planning systems that implement this now
![Page 64: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/64.jpg)
Prostate Brachytherapy
![Page 65: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/65.jpg)
![Page 66: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/66.jpg)
HDR needles (Ir 192)I 125 seeds
HDR with u/s guided needles and remote afterloading or
LDR Iodine seeds
![Page 67: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/67.jpg)
Test the whole process
Plan analysis
(DVH)
Isodoses
Dose prescription
Applicator and sources
identified
Phantom imaged
Digitised
![Page 68: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/68.jpg)
Aspects that need checking
include:
Input of patient and applicator data
source and applicator localization
dose calculation around sources
use of multi-modality imaging,
evaluation and visualization of 3D dose
distributions
optimization routines
output of patient plan and associated
information
![Page 69: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/69.jpg)
Useful references
http://www-naweb.iaea.org/nahu/dmrp/pdf_files/TECDOC1274.pdf
![Page 70: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/70.jpg)
Use of phantoms and external audit
ESTRO ESQUIRE PROJECTEducation, Science and QUality Assurance
for RadiotherapyBrachytherapy Physics Quality assurance
System
5 cm
Source TLD
Reconstruction Dosimetry
UK: Alanine audit and Gaf chromic
Applicators and line sources
![Page 71: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/71.jpg)
![Page 72: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/72.jpg)
![Page 73: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/73.jpg)
- radiation absorption- signal emission
- signal transmission
- signal collection- dose rate
measurement
Important requirementson in-vivo dosimeter:• small size• real-time signal• dose linearity• large dynamic range• bendable
Real time luminescence
dosimetry
Courtesy Gustavo Kertzscher, DTU, Denmark
![Page 74: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/74.jpg)
References
Wilkinson JM (2001) Source specification and dosimetry. In: Principles and practice of Brachytherapy using afterloading systems. Chapter 2 (Joslin CAF, Flynn A, Hall EJ eds.) Arnold, London
The IPEM code of practice for determination of the reference air
kerma rate for HDR (192)Ir brachytherapy sources based on the
NPL air kerma standard.(2010)A M Bidmead, T Sander, S M Locks,
C D Lee, E G A Aird, R F Nutbrown, A Flynn Physics in Medicine
and Biology 06/2010; 55(11):3145-59.
www.e-lfh.org.uk
![Page 75: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/75.jpg)
TG43:
Medical Physics 22 (2) 209-235 Feb 1995
Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.
Rivard MJ, Coursey BM, DeWerd LA, Hanson WF, Huq MS, Ibbott GS, Mitch MG, Nath R, Williamson JF.
Med Phys. 2004 Mar;31(3):633-74. Erratum in: Med Phys. 2004 Dec;31(12):3532-3.
Supplement to the 2004 update of the AAPM Task Group No. 43 Report.
Rivard MJ, Butler WM, DeWerd LA, Huq MS, Ibbott GS, Meigooni AS, Melhus CS, Mitch MG, Nath R, Williamson JF; American Association of Physicists in Medicine.
Med Phys. 2007 Jun;34(6):2187-205
![Page 76: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source](https://reader031.fdocuments.us/reader031/viewer/2022030421/5aa908d57f8b9a6c188c4d58/html5/thumbnails/76.jpg)
Report of the Task Group 186 on model-
based dose calculation methods
in brachytherapy beyond the TG-43
formalism: Current status and
recommendations for clinical
implementation: Med. Phys. 39 (10),
October 2012
Luc Beaulieua,Åsa Carlsson Tedgren,
Jean-François Carrier, Stephen D. Davis,
Firas Mourtada, Mark J. Rivard, Rowan M.
Thomson, Frank Verhaegen, Todd A.
Wareing, Jeffrey F. Williamson