Body Composition Publication

6
Innovation Single prediction equation for bioelectrical impedance analysis in adults aged 22–59 years B. R. PATIL*{, D. P. PATKAR{, S. A. MANDLIKx, M. M. KUSWARKARx and G. D. JINDALx {Department of Instrumentation Engineering, Bharati Vidyapeeth College of Engineering, Navi Mumbai, 400 614, India {MRI Centre, Dr Balabhai Nanavati Hospital, Vile Parle (West), Mumbai, 400 056, India xBio-Medical Section, Electronics Division, Modular Laboratories, B.A.R.C., Mumbai, 400 085, India (Received 17 October 2010; revised 14 December 2010; accepted 24 November 2010) The purpose of this study was to validate a single bioelectrical impedance analysis (BIA) equation in healthy Indian subjects aged 22–59 years with a body mass index (BMI) between 16.8 and 47.3 kg m 72 . Healthy subjects (34 men and 30 women) were measured by two methods: bone mineral content (BMC) was measured by a commercial body composition analyser and bioelectrical impedance at various frequencies was measured by a newly developed bioelectrical impedance measurement system. As correlations were high and prediction error was low, a single equation was developed using all subjects as follows: BMC ¼ 73.5268 þ (0.0279 6 h) þ (0.0145 6 w) þ (184 6 (h 2 /Z body50 )) (1.08 6 (w 6 h 2 /Z body6.25 )) – (0.0032 6 (age)) – (0.103 6 (sex); men ¼ 1, women ¼ 0). BMC measured from commercial instrument InBody720 was 2.552 + 0.457 kg. BMC predicted by equation was 2.554 + 0.447 kg (R ¼ 0.976, adjusted R 2 ¼ 0.948, standard error of estimate ¼ 0.104 kg, total error ¼ 0.09987 kg). The results of this study show that the newly developed multi-frequency bioelectrical impedance measurement system with the single prediction BIA equation can be used in screening the subjects suspected with osteoporosis and for follow-up study of the patient under the therapy for osteoporosis. For validation of commercial instrument InBody720, BMC of 22 healthy subjects was measured by InBody720 and dual-energy X-ray absorptiometry. High correlation (R ¼ 0.9531) and low error (total error ¼ 0.0913 kg) was found between these two methods. Keywords: Body parameters; Bioelectrical impedance analysis; Bone mineral content; Osteoporosis 1. Introduction Osteoporosis is a disease in which the bones deteriorate and become weak and fragile. Around 500 million populations worldwide are estimated to have osteoporosis. Osteoporo- sis in women is increasingly being recognized as an important health issue. In the early stage of this disease there may be no symptoms. It is frequently painless until a bone breaks. In fact, a fractured bone may be the first symptom of osteoporosis. Osteoporosis is also character- ized by an abnormal loss of bone mineral content, which leads to a tendency to non-traumatic bone fractures or to *Corresponding author. Email: [email protected] Journal of Medical Engineering & Technology, Vol. 35, No. 2, February 2011, 109–114 Journal of Medical Engineering & Technology ISSN 0309-1902 print/ISSN 1464-522X online ª 2011 Informa UK, Ltd. http://www.informahealthcare.com/journals DOI: 10.3109/03091902.2010.543751 J Med Eng Technol Downloaded from informahealthcare.com by Bharati Vidyapeeth Deemed University on 01/14/11 For personal use only.

description

bcp

Transcript of Body Composition Publication

  • Innovation

    Single prediction equation for bioelectrical impedance analysisin adults aged 2259 years

    B. R. PATIL*{, D. P. PATKAR{, S. A. MANDLIKx, M. M. KUSWARKARx and G. D. JINDALx

    {Department of Instrumentation Engineering, Bharati Vidyapeeth College of Engineering,Navi Mumbai, 400 614, India

    {MRI Centre, Dr Balabhai Nanavati Hospital, Vile Parle (West), Mumbai, 400 056, IndiaxBio-Medical Section, Electronics Division, Modular Laboratories, B.A.R.C., Mumbai, 400 085, India

    (Received 17 October 2010; revised 14 December 2010; accepted 24 November 2010)

    The purpose of this study was to validate a single bioelectrical impedance analysis (BIA)

    equation in healthy Indian subjects aged 2259 years with a body mass index (BMI)

    between 16.8 and 47.3 kg m72. Healthy subjects (34 men and 30 women) were measured

    by two methods: bone mineral content (BMC) was measured by a commercial body

    composition analyser and bioelectrical impedance at various frequencies was measured

    by a newly developed bioelectrical impedance measurement system. As correlations were

    high and prediction error was low, a single equation was developed using all subjects

    as follows: BMC73.5268 (0.02796 h) (0.01456w) (1846 (h2/Zbody50)) (1.086 (w6 h2/Zbody6.25)) (0.00326 (age)) (0.1036 (sex); men 1, women 0).BMC measured from commercial instrument InBody720 was 2.552+ 0.457 kg. BMCpredicted by equation was 2.554+ 0.447 kg (R 0.976, adjusted R2 0.948, standarderror of estimate 0.104 kg, total error 0.09987 kg). The results of this study show thatthe newly developed multi-frequency bioelectrical impedance measurement system with

    the single prediction BIA equation can be used in screening the subjects suspected with

    osteoporosis and for follow-up study of the patient under the therapy for osteoporosis.

    For validation of commercial instrument InBody720, BMC of 22 healthy subjects was

    measured by InBody720 and dual-energy X-ray absorptiometry. High correlation

    (R 0.9531) and low error (total error 0.0913 kg) was found between these twomethods.

    Keywords: Body parameters; Bioelectrical impedance analysis; Bone mineral content;

    Osteoporosis

    1. Introduction

    Osteoporosis is a disease in which the bones deteriorate and

    become weak and fragile. Around 500 million populations

    worldwide are estimated to have osteoporosis. Osteoporo-

    sis in women is increasingly being recognized as an

    important health issue. In the early stage of this disease

    there may be no symptoms. It is frequently painless until a

    bone breaks. In fact, a fractured bone may be the rst

    symptom of osteoporosis. Osteoporosis is also character-

    ized by an abnormal loss of bone mineral content, which

    leads to a tendency to non-traumatic bone fractures or to

    *Corresponding author. Email: [email protected]

    Journal of Medical Engineering & Technology, Vol. 35, No. 2, February 2011, 109114

    Journal of Medical Engineering & TechnologyISSN 0309-1902 print/ISSN 1464-522X online 2011 Informa UK, Ltd.

    http://www.informahealthcare.com/journalsDOI: 10.3109/03091902.2010.543751

    J Med

    Eng

    Tec

    hnol

    Dow

    nloa

    ded

    from

    info

    rmah

    ealth

    care

    .com

    by

    Bhar

    ati V

    idya

    peet

    h D

    eem

    ed U

    nive

    rsity

    on

    01/1

    4/11

    For p

    erso

    nal u

    se o

    nly.

  • structural deformations of bone [1]. Osteoporosis is a

    widespread medical condition aecting middle-aged and

    older people [2]. The accurate estimation of the BMC has

    been an important diagnostic indicator for determining the

    status of osteoporosis and for the follow-up study of

    patients under therapy for osteoporosis. Various BMC

    measurement techniques/methods are available, such as

    dual-energy X-ray absorptiometry (DEXA), ultrasound

    densitometry and quantitative computed tomography

    (QCT). DEXA is the most well-known method. However,

    the use of DEXA is limited because of its practicability

    and cost.

    Bioelectrical impedance analysis (BIA) is a user-

    friendly, safe, simple, inexpensive and non-invasive

    technology for clinical and non-clinical purposes. Mea-

    surements can be made quickly and can be repeated at

    short time intervals. BIA has the potential to provide

    more information regarding the functioning of the human

    body. To date the full potential of this technology has

    not been realized. Commercial instruments available for

    BIA measure electrical parameters in dierent body

    segments and yield clinical information regarding total

    body water, fat free mass, and body fat, etc. Many

    investigators have developed empiric BIA equations for

    prediction of fat-free mass, total body water and body

    fat [314]. Most of these equations have been validated in

    relatively young healthy adults against several body-

    composition techniques [10]. Studies have shown that

    BIA formulae developed for healthy, normal-weight sub-

    jects are not suitable for obese subjects [14, 15]. In

    longitudinal studies, the use of dierent BIA formulae in

    the same subject who becomes overweight introduces a bias

    into body-composition studies, making one question

    whether the dierences in body composition are due to

    changes in the BIA formula or due to changes in body

    composition. Thus, it would be advantageous to use a

    single formula that is applicable in both normal and

    overweight subjects and permits estimation of body

    composition. Roubeno et al. [14] and others concluded

    that BIA equations are subject to errors that cannot be

    determined a priori unless they are validated in the specic

    population in which they are to be applied [1618]. Thus,

    BIA equations must be validated in a representative

    population sample against a reference method before they

    can be accepted as accurate. The purpose of the present

    investigation was to validate a single BIA equation for

    the assessment of BMC, against the commercial device

    InBody720 (Biospace, Seoul, Korea) which was validated

    with DEXA [19]. DEXA is a reference method for BMC

    that has been validated against many independent methods.

    A single BIA equation that is valid in subjects with dierent

    BMIs for use in longitudinal studies would be a signicant

    advantage in screening subjects suspected with osteoporosis

    and for follow-up study of patients under therapy for

    osteoporosis.

    2. Subjects and methods

    2.1. Subjects

    The study group consisted of 64 individuals (34 men and 30

    women) in the age group of 22 to 59 years. All subjects were

    born in India and resided in Mumbai. Although subjects

    were randomly selected, statistical analysis showed not

    much dierence in (h), (w) and BMI between subjects and a

    control group of healthy age-matched 264 men and 279

    women. Body height was measured to the nearest 0.5 cm.

    The purpose of the study was explained to all subjects and

    their oral consent was taken. Each subject was measured by

    two methods: bone mineral content (BMC) was measured

    by a commercial instrument (InBody720, Biospace) and

    bioelectrical impedance at various frequencies was mea-

    sured by a newly developed bioelectrical impedance

    measurement system.

    2.2. Segmental multi-frequency bioelectrical impedance

    analysis (MF-BIA) instrument (InBody720)

    The BMC of each subject was obtained from the segmental

    multi-frequency bioelectrical impedance analysis instru-

    ment (InBody720). From segmental impedance values,

    using proprietary equations, the instrument gives an

    estimated value of BMC. The correctness of this estimated

    value is validated against DEXA measurement. This

    commercial product is approved by FDA (Food and Drug

    Administration, United States, May 2003). The instrument

    was operated at frequencies of 1, 5, 50, 250, 500 and 1000

    kHz which were pre-set by the manufacturer and intro-

    duced into the body in the ascending order of frequency.

    This device uses contact electrodes, located in the handgrips

    and the footpads. Subjects were asked to stand with the ball

    and heel of each foot on two metallic electrodes on the oor

    scale and hold handrails with metallic grip electrodes in

    contact with the palm and thumb [20, 21]; they were

    instructed to keep their arms fully extended and abducted

    approximately 208 laterally. For 1 kHz programmedfrequency, an alternating current of 100 mA and for otherprogrammed frequencies, an alternating current of 500 mAwas applied between a pair of carrier electrodes, and

    voltage was measured across a pair of sensing electrodes.

    For cross validation of InBody720, the BMC of healthy 22

    subjects was measured by InBody720 and dual-energy

    X-ray absorptiometry (Lunar Prodigy, DPX-IQ, General

    Electric Healthcare, Belgium, Europe). High correlation

    (R 0.9531) and low error (total error 0.0913 kg) werefound between these two methods (gure 1).

    2.3. Developed multi-frequency bioelectrical impedance

    measurement system

    The body impedance of each subject at various frequencies

    was measured and recorded by the new system, using the

    110 B. R. Patil et al.

    J Med

    Eng

    Tec

    hnol

    Dow

    nloa

    ded

    from

    info

    rmah

    ealth

    care

    .com

    by

    Bhar

    ati V

    idya

    peet

    h D

    eem

    ed U

    nive

    rsity

    on

    01/1

    4/11

    For p

    erso

    nal u

    se o

    nly.

  • whole body impedance measurement approach. The system

    operated at frequencies of 6.25, 12.5, 25, 50 and 100 kHz

    which were pre-set and introduced into the body in the

    ascending order of frequency. Before testing, standardiza-

    tion of a subjects posture is not required. Subjects were

    asked to sit on the chair with connected electrodes. For

    each of the programmed frequencies, an alternating current

    of 500 mA was applied between silver band currentelectrodes, which are located in the palm and feet on same

    side. The two sensing electrodes are located in the wrist and

    ankle on the same side. The amplitude of the sensed voltage

    drop is directly proportional to the impedance of the body

    (Zbody) between the sensing electrodes. Figure 2 shows a

    schematic block diagram of the multi-frequency bioelec-

    trical impedance measurement system. The Biomedical

    Instrumentation Processor Board (BMIPB) is an interface

    board used to interface the system with compatible PC. The

    100 kHz square wave signal is derived from BMIPB, which

    is further divided by the synchronous up/down counter and

    produces the four square wave singles of 50, 25, 12.5 and

    6.25 kHz frequencies. For smoothening, each square wave

    signal is ltered with the help of second order low pass lter

    followed by a band pass lter. This produces pure

    sinusoidal alternating signals. These signals are then

    multiplexed with the help of an analogue multiplexer.

    Any one of the signals can be selected at the output of the

    multiplexer by the operator through the user interface

    panel. The selected signal is applied to the VI (voltage to

    current) converter. The sinusoidal alternating current of

    constant magnitude produced by VI converter is passed

    through the body with the help of isolation transformer.

    The voltage signal developed along the current path is

    sensed with the help of sensing electrodes and amplied

    using a precision instrumentation amplier. The wide band

    pass lter removes the superimposed noise and produces a

    pure sinusoidal alternating voltage, which is proportional

    to the impedance of the body under investigation. Further

    the output of the band pass lter is rectied, ltered and

    buered to obtain a pure DC voltage which is also

    proportional to the body impedance. The injected sinusoi-

    dal alternating current and the voltage signal developed

    along the current path were multiplied by analogue

    multiplier. The two componentsdouble frequency sinu-

    soidal signal and DC signal (mean value)were separated

    by the second order low pass lter.

    2.4. Statistics

    Descriptive statistics were calculated for (h), (w), (age),

    (sex), BMI and BIA parameters, including body impedance

    (Zbody), (Zbody/h), (Zbody/w), (h2/Zbody) and (w6 h

    2/Zbody)

    at 6.25, 12.5, 25, 50 and 100 kHz frequencies. The values

    were expressed as mean+ standard deviation (SD). Simpleregressions were calculated to test correlations between

    BMC obtained from InBody720 and BIA parameters

    measured from developed bioelectrical impedance measure-

    ment system. Student t-test was used to test dierences

    between methods. BMC measured by commercial instru-

    ment InBody720 was used as the criterion measurement.

    Stepwise multiple regression analysis was used to derive a

    prediction equation by BIA. Predictor variables, entered

    into the BIA model in the order of highest correlation

    coecient and smallest standard error of estimation (SEE)

    were (h2/Zbody50), h, w, (w6 h2/ ZBody6.25), (sex) and (age)

    and a single equation was developed using the entire

    sample. In addition to correlation and regression techni-

    ques, error analysis was performed. SEEs were calculated

    and used as errors of prediction for InBody720 derived

    BMC and predicted BMC by BIA equation. The total error

    (TE) was calculated as:

    TE P

    ni BMC1i BMC2i 2

    n 2

    s; 1

    where (BMC1) is the observed value of BMC by In-

    Body720 and (BMC2) is the predicted value of BMC by

    equation. The total error of measurement estimates the

    magnitude of the error for a given measurement and is

    dened as the dierence between measurements for the

    individual (i), i 1, . . . , n, where n is the number ofindividuals [22]. The total error was compared with the

    SEE. A small dierence between total error and SEE

    indicates high accuracy of the prediction. To assess the

    agreement between the two clinical measurements the

    dierence between the values was plotted against their

    means because the mean was the best available estimate of

    the true value. This analysis allows for the calculation of

    bias (estimated by the mean dierences), the 95%

    condence interval for the bias, and the limits of agreement

    (two SDs of the dierence) [23]. Statistical signicance was

    set at p 5 0.05 for all tests.

    Figure 1. Relationship between bone mineral content

    (BMC) measured with DEXA and the commercial instru-

    ment InBody 720.

    Bioelectrical impedance analysis in adults 111

    J Med

    Eng

    Tec

    hnol

    Dow

    nloa

    ded

    from

    info

    rmah

    ealth

    care

    .com

    by

    Bhar

    ati V

    idya

    peet

    h D

    eem

    ed U

    nive

    rsity

    on

    01/1

    4/11

    For p

    erso

    nal u

    se o

    nly.

  • 3. Results

    A total of 64 healthy adults aged 2259 years were recruited

    as subjects. Table 1 shows the anthropometric character-

    istics and table 2 shows the BIA characteristics of subjects

    grouped by dierent frequencies. The prediction equation

    developed from all subjects is shown in table 3. The order

    of entry of predictor variables was (h2/Zbody50), h, w, (w6h2/Zbody6.25), (sex) and (age) for the BIA model (table 4).

    (h2/Zbody50) accounted for 84.12% of the variability

    (SEE 0.182 kg) of the equation whereas (h) aloneaccounted for only 74.51% of the variability (SEE0.231kg) and (w) alone accounted for 53.74% of the

    variability (SEE 0.311 kg). Inclusion of height (h), weight(w), (age) and (sex) without BIA parameters accounted for

    92.62% of the variability with a SEE of 0.124 kg. Thus

    inclusion of BIA parameters clearly improved the prediction

    power and decreased the SEE compared with anthropo-

    metric parameters only. Figure 3 shows the correlation and

    mean dierence, according to Bland and Altman [23], using

    the prediction equation in all subjects [23]. Subjects (data

    not shown) with BMIs above 27 kg m72 were analysed

    separately to determine whether greater error occurred with

    the BIA equation in larger subjects. BMC measured for 22

    subjects with BMIs above 27 kg m72 from commercial

    instrument InBody720 was 2.701+ 0.5473 kg. BMCpredicted by equation was 2.706+ 0.5173 kg (R 0.978,total error 0.114 kg). Thus, it is possible to estimate BMCwith the same equation for overweight and obese subjects.

    Figure 2. Schematic block diagram of the multi-frequency bioelectrical impedance measurement system.

    Table 1. Anthropometric Characteristics of healthy subjects.

    Men Women Combined

    n 34 30 64

    (age) (years) 36.15+ 8.72 43.9+ 7.83 39.78+ 9.12h (cm) 167.1+ 6.47 155.2+ 5.08 161.50+ 8.33w (kg) 73.23+ 14.44 62.08+ 8.79 68.01+ 13.26BMI (kg m72) 26.2+ 4.88 25.9+ 4.18 26.1+ 4.53

    nnumber of subjects; hheight; wweight; BMI body mass index

    112 B. R. Patil et al.

    J Med

    Eng

    Tec

    hnol

    Dow

    nloa

    ded

    from

    info

    rmah

    ealth

    care

    .com

    by

    Bhar

    ati V

    idya

    peet

    h D

    eem

    ed U

    nive

    rsity

    on

    01/1

    4/11

    For p

    erso

    nal u

    se o

    nly.

  • Table 2. Bioelectrical impedance analysis characteristics of healthy subjects (n 64).Frequency (kHz)

    6.25 12.5 25 50 100

    (Zbody) (O) 658.31+ 101.10 563.98+ 91.41 499.44+ 78.74 460.19+ 73.31 425.22+ 62.87(Zbody /h) (O cm

    71) 4.09+ 0.73 3.51+ 0.66 3.11+ 0.57 2.87+ 0.53 2.65+ 0.46(Zbody/w) (O kg

    71) 10.17+ 3.06 8.72+ 2.66 7.72+ 2.33 7.11+ 2.15 6.56 + 1.89(h2/Zbody) (m

    2 O71) 0.0041+ 0.0009 0.0048+ 0.001 0.0054+ 0.001 0.0058+ 0.001 0.0063+ 0.001(w6 h2/Zbody)(kg m2 O71)

    0.289 + .137 0.338+ 0.162 0.382 + 0.179 0.415+ 0.196 0.447+ 0.211

    nnumber of subjects; Zbody body impedance; hheight; wweight.

    Table 3. Prediction equation for BMC using all subjects(n 64).

    BMC73.5268 (0.02796 h) (0.01456w) (1846 (h2/Zbody50) (1.086 (w6 h2/Zbody6.25)) (0.00326 (age)) (0.1036 (sex);men 1, women 0)

    Observed BMC by commercial instrument (InBody720) 2.552+0.457 kg

    Predicted BMC 2.554+ 0.447 kg (R 0.976, adjusted R2 0.948,SEE 0.104 kg, TE 0.09987 kg)

    nnumber of subjects; BMC bone mineral content; Zbody50bodyimpedance at 50 kHz; h height; wweight; Zbody6.25body im-pedance at 6.25 kHz; R validity coecient; SEE standard error ofthe estimate; TE total error.

    Table 4. Contribution and order of entry of variables to thebioelectrical impedance analysis model and anthropometric

    model for bone mineral content (n 64 subjects).

    Model and variables

    Cumulative variables

    used in model Variables

    Ad.R2 SEE P Ad.R2 SEE P

    BIA

    (h2/Zbody50) 84.12 0.182 0.0001 84.12 0.182 0.0001

    h 91.87 0.130 0.0001 74.51 0.231 0.0001w 93.70 0.115 0.0001 53.74 0.311 0.0001(w6 h2/Zbody6.25) 94.39 0.108 0.0001 62.89 0.278 0.0001(sex) 94.57 0.106 0.0001 49.49 0.325 0.0001(age) 94.80 0.104 0.0001 6.24 0.442 0.0001

    BMI

    h 74.51 0.231 0.0001

    hw 91.82 0.131 0.0001hw (age)

    92.73 0.123 0.0001

    hw (age) (sex)

    92.62 0.124 0.0001

    n number of subjects; BIA bioelectrical impedance analysis;BMIbody mass index; hheight; wweight; BMCbone mineralcontent; Zbody50body impedance at 50 kHz; Zbody6.25 bodyimpedance at 6.25 kHz; SEE standard error of the estimate;P signicance of contribution of each additional individual para-meter to the stepwise multiple regression model; Ad.R2% adjustedsquared value of the validity coecient.

    Figure 3. Correlations (a) and dierences (b) of bone

    mineral content (BMC) in all subjects estimated by

    commercial instrument InBody720 and developed BIA

    equation. The dierence (BMC obtained by InBody720

    BMC calculated by BIA equation per BlandAltman) is

    plotted against the mean of the measurements of BMC by

    commercial instrument InBody720 and developed BIA

    equation. SEE, standard error of the estimate; R, validity

    coecient; TE, total error. Circles indicate men; triangles

    indicate women.

    Bioelectrical impedance analysis in adults 113

    J Med

    Eng

    Tec

    hnol

    Dow

    nloa

    ded

    from

    info

    rmah

    ealth

    care

    .com

    by

    Bhar

    ati V

    idya

    peet

    h D

    eem

    ed U

    nive

    rsity

    on

    01/1

    4/11

    For p

    erso

    nal u

    se o

    nly.

  • 4. Discussion

    BIA has been developed for eld use and has shown great

    potential for use in estimating body composition and it is

    easy, non-invasive, and inexpensive. In osteoporosis, bones

    become porous and bone mineral content reduces as

    minerals such as calcium leach out. Bones aected by

    osteoporosis are less dense than the normal bones, resulting

    in reduction in their density and an increase in their

    resistivity. Due to the change in resistivity of bones, the

    bioelectrical parameters of the body changes and provide

    quantitative information about the BMC.

    All subjects in this study were reported to be healthy. The

    BIA equation developed in this study is valid for healthy

    adults 2259 years old. Slightly lower accuracy should be

    expected in subjects older than 60 years. Validity of the BIA

    equation in subjects older than 60 years is unknown and

    requires further validation. It is also necessary in subjects

    with BMIs below 17 kg m72.

    5. Conclusion

    The results of this study show that the newly developed

    system with the single prediction BIA equation validated

    against commercial instrument (InBody720, Biospace, Seoul,

    Korea) can be used to predict BMC in subjects aged 2259

    years and with BMIs ranging from 16.8 to 47.3 kg m72. This

    low cost, portable and simple system will help in screening

    subjects with suspected osteoporosis and in follow-up studies

    of patients under therapy for osteoporosis.

    Acknowledgements

    This studywas supportedby theElectronicsDivision,Bhabha

    Atomic Research Centre (BARC),Mumbai and Dr Balabhai

    Nanavati Hospital, Mumbai. We are thankful to Aarti

    Karkera for helping in data collection and S. K. Athawale

    for proof-reading. We are also grateful to G. P. Srivastava,

    Director, E & I Group, R. K. Patil, Associate Director, E & I

    Group (C) and C. K. Pithawa, Head, Electronics Division of

    BARC for their support and encouragement.

    Declaration of interest: The authors declare that they have

    no conict of interest.

    References

    [1] Wasserman, S.H. and Barzel, U.S., 1987, Osteoporosis: The state of

    the art in 1987 a review. Seminars in Nuclear Medicine, 17, 283292.

    [2] National Institute of Health, 1984, Consensus conference: osteoporo-

    sis. Journal of the American Medical Association, 252, 799802.

    [3] Segal, K.R., Van Loan, M., Fitzgerald, P.I., Hodgdon, J.A. and Van

    Itallie, T.B., 1988, Lean body mass estimation by bioelectrical

    impedance analysis: a four-site cross over validation. American Journal

    of Clinical Nutrition, 47, 7.

    [4] Kushner, R.F. and Schoeller, D.A., 1986, Estimation of total body

    water by bioelectrical impedance analysis. American Journal of Clinical

    Nutrition, 44, 417.

    [5] Lukaski, H.C., 1985, Assessment of fat-free mass using bioelectrical

    impedance measurements of the human body. American Journal of

    Clinical Nutrition, 41, 810.

    [6] Lukaski, H.C., 1986, Validation of tetrapolar bioelectrical impedance

    measurements to assess human body composition. Journal of Applied

    Physiology, 60, 1327.

    [7] Van Loan, M. and Mayclin, P., 1987, Bioelectrical impedance

    analysis. Is it a reliable estimator of lean body mass and total body

    water? Human Biology, 2, 299.

    [8] Jackson, A.S., Pollock, M.L., Graves, J.E. and Mahar, M.T., 1988,

    Reliability and validity of bioelectrical impedance in determining body

    composition. Journal of Applied Physiology, 64, 529.

    [9] Lukaski, H.C. and Bolonchuk, W.W., 1988, Estimation of body uid

    volumes using tetrapolar bioelectrical measurements. Aviation, Space,

    and Environmental Medicine, 59, 1163.

    [10] Graves, J.E., Pollack, M.L., Calvin, A.B., Van Loan, M. and Lohman,

    T.G., 1989, Comparison of dierent bioelectrical impedance analyzers

    in the prediction of body composition. American Journal of Human

    Biology, 1, 603.

    [11] Heitmann, B.L., 1990, Prediction of body water and fat in adult danes

    from measurement of electrical impedance. A validation. International

    Journal of Obesity, 14, 417.

    [12] Deurenberg, P., van der Kooy, K., Evers, P. and Hulshof, T., 1990,

    Assessment of body composition by bioelectrical impedance

    in a population aged 60 y. American Journal of Clinical Nutrition,

    51, 3.

    [13] Deurenberg, P., van der Kooy, K., Leenen, R., Westrate, J.A. and

    Seidell, J.C., 1991, Sex and age specic prediction formulas for

    estimating body composition from bioelectrical impedance: a cross-

    validation study. International Journal of Obesity, 15, 17.

    [14] Roubeno, R., Baumgartner, R.N., Harris, T.B., et al., 1997,

    Application of bioelectrical impedance analysis to elderly population.

    Journal of Gerontology, 52A, M129.

    [15] Stolarczyk, L.M., Heyward, V.H., Van Loan, M.D., Hicks, V.L.,

    Wilson, W.L. and Reano, L.M., 1997, The fatness-specic bio-

    electrical impedance analysis equations of Segal et al: are they

    generalizable and practical? American Journal of Clinical Nutrition, 66,

    817.

    [16] Pichard, C., Kyle, U.G., Janssens, J.P., Burdet, L., Rochat, T.,

    Slosman, D.O., Fitting, J-W., Thiebaud, D., Roulet, M., Tschopp,

    J.M., Landry, M., Schutz, Y., 1997, Body composition by x-ray

    absorptiometry and bioelectrical impedance in chronic respiratory

    insuciency patients. Nutrition, 13, 9528.

    [17] Kyle, U.G., Pichard, C., Rochat, T., Slosman, D.O., Fitting, J-W.,

    Thiebaud, D., 1998, New bioelectrical impedance formula for patients

    with respiratory insuciency: comparison to dual-energy X-ray

    absorptiometry. European Respiratory Journal, 12, 960.

    [18] Pichard, C., Kyle, U., Gremion, G., Gerbase, M. and Slosman, D.,

    1997, Body composition by X-ray absorptiometry and bioelectrical

    impedance in elite female runners. Medicine & Science in Sports &

    Exercise, 29, 1527.

    [19] http://www.inbody./resources/userles/File/720manual.pdf; chapter

    no.2;page no.2-16; Last accessed date: 30 Nov.2010.

    [20] Deurenberg, P., 1994, International consensus conference on impe-

    dance in body composition. Age & Nutrition, 5, 142145.

    [21] Bedogni, G., Malavolti, M., Severi, S., Poli, M., Mussi, C., Fantuzzi,

    A.L., Battistini, N., 2002, Accuracy of an eight-point tactile- electrode

    impedance method in the assessment of total body water. European

    Journal of Clinical Nutrition, 56, 11431148.

    [22] Wellens, R., Chumlea, W.C., Guo, S., Roche, A. F., Reo, N.V.,

    Siervgel, R.M., 1994, Body composition in white adults by dual

    energy X-ray absorptiometry, densitometry, and total body water.

    American Journal of Clinical Nutrition, vol.59, no.3, 547555.

    [23] Bland, J.M. and Altman, D.G., 1986, Statistical methods for assessing

    agreement between two methods of clinical measurement. Lancet, i,

    307.

    114 B. R. Patil et al.

    J Med

    Eng

    Tec

    hnol

    Dow

    nloa

    ded

    from

    info

    rmah

    ealth

    care

    .com

    by

    Bhar

    ati V

    idya

    peet

    h D

    eem

    ed U

    nive

    rsity

    on

    01/1

    4/11

    For p

    erso

    nal u

    se o

    nly.