bitirmeee
-
Upload
ersin-tukenmez -
Category
Documents
-
view
222 -
download
0
Transcript of bitirmeee
-
8/6/2019 bitirmeee
1/37
I
BLINK SIGNAL ANALYSIS
by
Aylin Nuholu
Engineering Project Report
Department of Biomedical Engineering
Faculty of Engineering and Architecture
Yeditepe University
June 2011, Istanbul, Turkey
-
8/6/2019 bitirmeee
2/37
II
BLINK SIGNAL ANALYSIS
APPROVED BY:
Assoc.Prof.Dr. Ali mit KESKN (Supervisor)
Prof. Dr. Fuat BAYRAKEKEN .
Assist.Prof. Yusuf YUSUFOLU
DATE OF APPROVAL: 02.06.2011
-
8/6/2019 bitirmeee
3/37
III
ACKNOWLEDGEMENTS
I want to thank to my supervisor Prof. Dr. Ali mit Keskin at Yeditepe University of
Biomedical Engineering Department for his help, sharing his experience and superior
support during this project. I also thank Res. Assistant Mr. Hakan BOZKURT from
the Department of Biomedical Engineering for his generous help.
-
8/6/2019 bitirmeee
4/37
IV
ABSTRACT
Electrodes placed on the skin adjacent to the eyes measure changes in standing potential
between the front and back of the eyeball as the eyes move; a sensitive electrical test for
detection of retinal pigment epithelium dysfunction called oculography.
Eye movement measurements: Usually, pairs of electrodes are placed either above and
below the eye or to the left and right of the eye. If the eye is moved from the center position
towards one electrode, this electrode "sees" the positive side of the retina and the opposite
electrode "sees" the negative side of the retina. Consequently, a potential difference occurs
between the electrodes. Assuming that the resting potential is constant, the recorded potential
is a measure for the eye position.
The eye acts as a dipole in which the anterior pole is positive and the posterior pole is
negative.1.Left gaze; the cornea approaches the electrode near the outer canthus resulting in a
positive-going change in the potential difference recorded from it. 2.Right gaze; the cornea
approaches the electrode near the inner canthus resulting in a positive-going change in the
potential difference recorded from it (A, an AC/DC amplifier).
This Project is focused on design of an EOG device and experimental results. Proteus are
used for simulation and the signals received from the scalp are transferred into digital form.
-
8/6/2019 bitirmeee
5/37
V
ZET
Gz kaslarnn hareketine dayanarak elde edilen deney sonularndan yaplan karmlardorultusunda uyuukluk halinin etkisi llmtr. stemsiz gz krpma hareketinin hangiartlar altnda hangi iddette gerekletii dikkate alnarak deneklerde yaplan anketlerinsonular deerlendirilmitir.
Din ve uyuukluk (yorgun) hallerde alnan datalar kyaslanarak 3 farklhaldeki iaretlergz nnde bulundurularak sonuca ulalmtr.
Denek zerinde deney yaplmadan nce alnan anket sonular dorultusunda ortam
artlar da gz nnde bulundurularak alnan sonularda karlatrmaya gidilmitir.
Bu sayede gz kaslma hareketinin (istemsiz gz krpma) winkten farkn ortaya koyarak blink hareketininn kendi iinde farkl durumlarda deiik artlar altnda ayrldna dikkatekilmitir.
-
8/6/2019 bitirmeee
6/37
VI
TABLE OF CONTENTS
ACKNOWLEDGEMENTS .................................................................................................. III
ABSTRACT ............................................................................................................................ IV
ZET........................................................................................................................................ V
TABLE OF CONTENTS ....................................................................................................... VI
LIST OF FIGURES ............................................................................................................ VIII
LIST OF ABBREVIATIONS ................................................................................................ IX
1. INTRODUCTION ............................................................................................................. 1
1.1 WHAT IS EOG ? .............................................................................................................. 1
1.1.2 PUPPILARY DILATATION .................................................................................... 1
1.1.3 SACCADES .............................................................................................................. 1
1.1.5 LIGHT RESPONSE ................................................................................................... 2
1.1.6 MEASUREMENT OF AN EOG ............................................................................... 3
1.1.7 NORMAL VALUES ................................................................................................. 6
1.1.8 REPORTING THE EOG ........................................................................................... 6
1.2 ELECTRODE LOCATION ............................................................................................. 6
1.2.1 SKIN ELECTRODES ................................................................................................ 7
1.2.2 LIGHT SOURCES ..................................................................................................... 7
1.2.3. ELECTRONIC RECORDING EQUIPMENT ......................................................... 8
1.3 WHAT IS EMG? .............................................................................................................. 9
1.4 ELECTROMYOGRAM (EMG) AND NERVE CONDUCTION STUDIES ............... 10
2. CIRCUIT DESIGN OF AN EOG ..................................................................................... 11
2.1 INSTRUMENTATION AMPLIFIER ............................................................................ 11
2.2 NON-INVERTING AMPLIFIER ............................................................................... 12
2.2.1 NON-INVERTING AMPLIFIER CONFIGURATION .......................................... 13
2.2.3 EQUIVALENT POTENTIAL DIVIDER NETWORK ........................................... 14
2.2.4 NON-INVERTING AMPLIFIER CIRCUIT ........................................................... 15
3. ANATOMY OF AN EYE .................................................................................................. 16
3.1 EYE STRUCTURE ........................................................................................................ 16
-
8/6/2019 bitirmeee
7/37
VII
3.2 EYE SIGNALS............................................................................................................... 16
3.3 OPTIC AND CILIARY MUSCELS .............................................................................. 16
3.4 WHAT IS BLINK? ......................................................................................................... 23
4. EXPERIMENTAL RESULTS .......................................................................................... 24
5.CONCLUSION .................................................................................................................... 26
6.REFERENCES .................................................................................................................... 28
-
8/6/2019 bitirmeee
8/37
VIII
LIST OF FIGURES
FIGURE 1 Appearance of EOG saccadic recordings with DC (top) or AC (bottom)
amplification. .............................................................................................................................. 3
FIGURE 2. Examples of an idealized (top) and a practical (bottom) EOG record .................... 5
FIGURE 3 International 10-20 electrode placement .................................................................. 6
FIGURE 4. Non-inverting Amplifier ....................................................................................... 13
FIGURE 5.Potential Divider Network ..................................................................................... 14
FIGURE 6 . Connection of non-inverting amplifier ................................................................ 15
FIGURE 7. Non-inverting Amplifier circuit ............................................................................ 15
FIGURE 8 . Ekstraocular muscles ........................................................................................... 16FIGURE 9 Motor Nerve ........................................................................................................... 17
FIGURE 10 Oblique Muscles .................................................................................................. 17
FIGURE 11 Oblique Muscles .................................................................................................. 18
FIGURE 12 Orbita Fasya System ............................................................................................ 18
FIGURE 13 Eye Cover ............................................................................................................ 19
FIGURE 14 Orbita Fasya System ............................................................................................ 19
FIGURE 15 Eye Cover ............................................................................................................ 20
FIGURE 16 Orbitalis Oculi...................................................................................................... 20
FIGURE 17 Orbital Septum ..................................................................................................... 21
FIGURE 18 Lid Retractors ...................................................................................................... 21
FIGURE 19 Lid Retractors ...................................................................................................... 22
FIGURE 20 Tarsal Disc and Tarsal Sempatic Muscles ........................................................... 22
FIGURE 21 Example 1 ............................................................................................................ 24
FIGURE 22 Example 2 ............................................................................................................ 24
FIGURE 23 Example 3 ............................................................................................................ 24
FIGURE 24 data 1 .................................................................................................................... 25
FIGURE 25 data 2 .................................................................................................................... 25
FIGURE 26 data 3 .................................................................................................................... 25
FIGURE 27 Simulation with Proteus ....................................................................................... 26
FIGURE 28 Example of Proteus .............................................................................................. 26
-
8/6/2019 bitirmeee
9/37
IX
LIST OF ABBREVIATIONS
EOG: Electrooculogram
EMG:Electromyogram
ISCEV:International Society for Clinical Electrophysiology of Vision
-
8/6/2019 bitirmeee
10/37
1
1. INTRODUCTION
Biomedical signals used in the present study are EOG: horizontal and vertical eye
movements and voluntary eye blinks generate electrical activity. Therefore, positive potential
is charged at cornea side, whereas negative potential is charged at the retina side, constant
potential difference (corneal-retinal potential) is charged between the cornea and the retinas.
Body surface electrodes situated around the eyeball socket can detect potential changes
according to eye movements. (1)
On the other hand, eye blinks are classified into two categories: one is involuntary eye
blink, which occurs frequently or is evoked spontaneously by an external stimulus such as a
flash light; the other is a voluntary wink that is caused by intentional eye closing. These
phenomena can also be detected with the same electrodes as a potential change with opening
and closing of eyelids. (1)
1.1 WHAT IS EOG ?
1.1.2 PUPPILARY DILATATION
EOGs may be performed with pupils either dilated or undilated. Pupillary dilatation
provides better control of illumination levels, but adds time to the test and may make the test
somewhat more uncomfortable for certain patients. The critical parameter in producing a light
response of the standing potential is the level of retinal illumination. Thus, the light
levels which are used to perform the test will be different according to the state of the pupil.
(2)
1.1.3 SACCADES
Saccades are typically induced by illuminating the fixation lights alternately, but
patients could be instructed by other means to look back and forth at a steady rate between
fixation targets. ISCEV suggests that the eyes alternate direction every 1 to 2.5 seconds
(equivalent to a complete back and forth cycle every 2-5 seconds). Faster alterations become
uncomfortable and unstable, whereas alternations at the slow end make it more difficult for
subjects to maintain a steady alternating rhythm. Since continuous saccadic movement
becomes tiresome, it is recommended that a set number of saccades (minimum of 10) be
recorded once per minute throughout the test. Testing at least once per minute is necessary to
recognize the relevant peaks and troughs.(2)
-
8/6/2019 bitirmeee
11/37
2
1.1.4 PRE-ADAPTATION
Patients should be in ordinary room light, or be pre-adapted to room light levels, for at
least 15 minutes prior to the dark phase of testing. This pre-adaptation phase light should
measure between 35 and 70 lux looking ahead (not at the ceiling or close to a white wall).
Dimmer pre-adaptation light levels may fail to suppress rod function and will diminish the
size of the dark trough (although they would be acceptable for the baseline method of
recording described below). Stronger light levels or sudden changes in illumination may
excessively stimulate the dark trough and slow oscillations, and they will make it more
difficult to reach a steady baseline. Unusual bright light exposure such as sunlight,
ophthalmoscopy or fluorescein angiography should be avoided within 60 minutes of EOG
testing.(2)
1.1.5 LIGHT RESPONSEThe light stimulus should be turned on, and the EOG recorded, until the light peak has
occurred and the signal amplitudes have clearly begun to descend (at which point the
recording can be stopped). If there is no clear light peak, then recording should continue for at
least 20 minutes to insure that a delayed light peak is not missed. The choice of luminance
levels for the stimulus will depend on whether the pupil is dilated.
a. Dilated pupils: The stimulus intensity should be a fixed value (for each laboratory) within
the range of 50 and 100 cd/m2.
b. Undilated pupils: The stimulus intensity should be a fixed value (for each laboratory)
within the range of 400 and 600 cd/m2.
c. Technical note: These ranges of illumination are recommended to accommodate the typical
range of dilated or undilated pupil sizes. Technically speaking, the EOG would be better
standardized by the use of trolands (cd/m2
x mm2
of pupillary area) which would insure
similar levels of retinal illumination regardless of pupil size. ISCEV considers the range of
1000 to 3000 trolands (3 to 3.5 log trolands) to be optimal. For practical purposes, however, it
is difficult to measure pupillary diameter during the test, while the patient is looking into the
diffusing sphere, and most commercial stimulus units do not have the capability of finely
adjusting the luminance of the sphere. Thus, the Standard recommends separate fixed light
-
8/6/2019 bitirmeee
12/37
3
intensities for dilated and undilated pupils, as the best compromise between an adjustable
light level and the reality of clinical testing.
FIGURE 1 Appearance of EOG saccadic recordings with DC (top) or AC (bottom)
amplification. Measurement of saccade amplitude (brackets) should avoid the artifact of
overshoot.
1.1.6 MEASUREMENT OF AN EOG
a. Saccadic amplitudes: The measurement of EOG oscillations must take into account
the potential artifacts of overshoot (positive or negative) and falloff (Figure 1). Overshoot
occurs when subjects look beyond the fixation target and then return to a stable position.
Falloff occurs during AC recording, as the amplitude fades from its maximal value. Sharp
overshoots can be ignored and the stable amplitude level used for measurement. If there is
significant falloff from AC filtering of the signal, we recommend measurement of the leading
edge, or at least the initial stable value.(3)
-
8/6/2019 bitirmeee
13/37
4
b. The ratio of light peak to dark trough (Arden ratio): Measurement of the lowest dark-
adapted point (dark trough) and highest light point (light peak) should be made. However,
examiners should be aware that there is often some random variation in these values, and the
curves should be visually or otherwise "smoothed" to identify the true trough and peak points.
c. The ratio of light peak to dark baseline: The average stable baseline value in the dark is
determined. The light peak is determined as in the preceding section. The value of the light
peak to dark baseline ratio will typically be lower than the Arden ratio.
d. Latency (implicit time) of the light peak: The latency is the time between the onset of the
light phase and the peak of the light response. It can be of clinical relevance in addition to the
Arden- or baseline-ratios.
e. Amplitude of dark trough or dark baseline: It is important to measure the standing potential
amplitude in microvolts at either the bottom of the dark trough (if the Arden ratio method is
used) or at the dark-adapted baseline, since low values may have clinical significance and
may lead to the calculation of ratios of uncertain physiologic meaning. (These baseline
amplitudes should be normalized to microvolts per degree if a visual angle other than the
standard 30 is used). (3)
-
8/6/2019 bitirmeee
14/37
5
FIGURE 2. Examples of an idealized (top) and a practical (bottom) EOG record of saccadic
amplitude versus time. The points often must be "smoothed" visually or mathematically to
accurately estimate the dark trough and light peak.
-
8/6/2019 bitirmeee
15/37
6
1.1.7 NORMAL VALUES
ISCEV recommends that each laboratory establish or confirm normal values for its
own equipment and patient population, and that EOG reporting (whether for local records or
publication) include normal values. Some manufacturers may choose to distribute norms for
their standard protocols. An effort will be underway to establish world-wide norms.
Although normal values for the Arden ratio may be larger than those for the light peak vs.
dark baseline method, the two methods are roughly comparable in terms of variability of the
response (in the range of 10% variance about the mean for repeated testing on an experienced
subject). Norms from one method cannot and must not be used for the other.(4)
1.1.8 REPORTING THE EOG
ISCEV recommends that reports or communications of EOG data state clearly whether
the ratio of light peak to dark trough (Arden ratio), or light peak to dark baseline, method has
been used. EOG reports should include the latency of the light peak and the amplitude of the
dark trough or dark baseline in microvolts per degree of visual cycle. Published clinical
reports should indicate whether the recording technique meets the International Standard.
Research reports should indicate pupil size, spatial arrangement and measured intensity of the
light stimulus, conditions of pre-and dark-adaptation, time intervals of stimulation and filtercharacteristics of the recording equipment. (4)
1.2 ELECTRODE LOCATION
FIGURE 3 International 10-20 electrode placement
-
8/6/2019 bitirmeee
16/37
7
We determine the optimum electrode position with a minimum number of electrodes for the
EOG communication system. Two necessary conditions exist to decide electrode disposition:
i) Potential changes for eye movements of the four orientation axesup, down, right, and left
must be detected separately to generate four directional inputs. Thereby, two directional
channels are required to clearly distinguish vertical and horizontal eye movements, including
their directions.
ii) A potential change of voluntary eye blink is used for selection in the present system. The
blink simulates a mouse button click. Voluntary and involuntary eye blinks should be
distinguishable for this reason. Moreover, the detected signal from a voluntary eye blink
should be repeatable whether both eyes blink or one eye blinks; also, they should not be
affected by an electromyogram.
Two skin electrodes should be used for each eye, placed as close to each canthus as possible.
Avoid large size electrodes that fit poorly and increase the distance of separation. A ground
electrode should be attached to the middle of the forehead, or some other neutral site.
1.2.1 SKIN ELECTRODES
A. Construction: Electrodes should be made of relatively non-polarizable material such as
silver-silver chloride or gold.
B. Electrode application: The skin should be cleansed of oils with alcohol or a commercial
skin-preparing material. The electrodes should be applied with a conductive paste.
C. Cleaning: If non-disposable electrodes are used, they should be suitably cleaned after each
use to prevent transmission of infectious agents. The cleaning protocol should follow current
standards for devices that contact the skin. (4)
1.2.2 LIGHT SOURCES
a. Luminance: Illumination may be provided by one or several lamps, but they should produce
visibly white light, and be situated so that the full-field diffusing sphere is evenly illuminated
from the vantage point of the patient. Areas of focal intensity ("hot spots") or shadows shouldbe avoided.
-
8/6/2019 bitirmeee
17/37
8
b. Adjustment: The light source should be adjustable by the use of filters or other means, to
allow for calibration of the unit. Large variation in stimulus intensity will need to be available
if patients with both dilated and undilated pupils are to be studied.
c. Calibration: The luminance produced by the full-field stimulator should be measured in
cd/m2
with a photometer that meets international standards for photometric measurements
based on the photopic luminosity curve. ISCEV recommends that in the future, manufacturers
of stimulators provide a suitable photometer as a part of the equipment. Since light output
may vary with time, from changes in light bulbs or filters, it is important that the luminance
be periodically recalibrated. Self-calibrating units are to be encouraged. (4)
1.2.3. ELECTRONIC RECORDING EQUIPMENT
a. Basis of measurement of the standing potential: Because of the dipole effect of the eye,
with the cornea positive to the back of the globe, saccadic eye movements result in current
flows around the orbit that are proportional to the magnitude of the standing potential of each
eye. These voltage changes can be measured from skin electrodes placed at the nasal and
temporal canthal regions of the eye.
b. Amplification systems: Direct current (DC) amplification most faithfully reproduces the
square wave voltage changes that occur when a subject looks back and forth. However, for
practical purposes the use of alternating current (AC) recording systems is easier since the
problems of drift and stability are minimized. In general, it is recommended that an AC
system with a low frequency cutoff at 0.1 Hz or lower, and a high frequency cutoff no lower
than 20 Hz (but preferably below 50 or 60 Hz to minimize interference). DC recording may
be utilized by experienced laboratories, but will usually require some type of electronic
baseline compensation to avoid drift
.c. Display system: It is very important that the original waveforms be displayed during the
recording of the EOG. This allows the individual performing the test to judge whether the
signals are stable, and to observe artifacts, jerky saccades, unacceptable overshoot, etc., which
might necessitate re-application of electrodes, re-instruction of the patient, or an altered
interpretation of the results. In systems which automatically measure the amplitude of
collections (epoches) of waveforms, and plot the values, it is important that the raw data be
displayed transiently as it is gathered, to allow for these judgments.
-
8/6/2019 bitirmeee
18/37
9
d. Patient isolation: It is recommended that the amplifiers be electrically isolated from the
patient, according to current standards for safety of biological recording systems used
clinically.
1.3 WHAT IS EMG?
Electromyography is a term meaning "electrical muscle recording." Because muscle is an
electrical tissue, the needle electrode in the muscle, when connected to an amplifier and an
oscilloscope screen, can show the muscle's electrical activity as a series of voltage-
fluctuations. Moreover, the signal is also fed through a speaker, so you and the
electromyographer can listen to what the muscle has to say. Each muscle is analyzed while
you tense it, and again while you relax it. (5)
Normal muscles display typical patterns to the eye and ear through the oscilloscope and
speaker. Abnormal muscles show altered patterns. Sometimes the abnormality is more
evident on the oscilloscope, and sometimes more evident on the speaker. Muscles that are
themselves sick (myopathy) show one pattern, while muscles that are connected to sick
nerves (neuropathy) or spinal roots (radiculopathy) can show yet another pattern. (5)
Electromyography is often paired with nerve conduction studies performed at the same
testing session by the same doctor and with the same equipment. Each testEMG and nerve
conduction studieshas its own story to tell, as well as its own strengths and weaknesses in
its ability to show signs of disease. The results of the electromyography and nerve conduction
studies are considered together to come up with a more complete, combined test-outcome and
report. (5)
The time required for testing can depend on the nature of the problem. Also, the
electromyographer might add or subtract additional testing depending on how the initialcomponents turn out. Overall, a typical session might last between 45 and 90 minutes. That
doesn't mean that the patient is subjected to unpleasantries during that whole period of time.
Actually, much of the time is devoted to getting all the little pieces and parts of equipment
organized and in place for each "mini-test" comprising the overall testing session. (5)
-
8/6/2019 bitirmeee
19/37
10
1.4ELECTROMYOGRAM (EMG) AND NERVE CONDUCTION STUDIES
An electromyogram (EMG) measures the electrical activity of muscles at rest and during
contraction. Nerve conduction studies measure how well and how fast the nerves can send
electrical signals. Nerves control the muscles in the body by electrical signals (impulses), and
these impulses make the muscles react in specific ways. Nerve and muscle disorders cause the
muscles to react in abnormal ways.
Measuring the electrical activity in muscles and nerves can help find diseases that damage
muscle tissue (such as muscular dystrophy) or nerves (such as amyotrophic lateral sclerosis or
peripheral neuropathies). EMG and nerve conduction studies are often done together to give
more complete information. (6)
Why It Is Done
An electromyogram (EMG) is done to:
Find diseases that damage muscle tissue, nerves, or the junctions between nerve andmuscle (neuromuscular junctions). These disorders may include a herniated disc,
amyotrophic lateral sclerosis (ALS), or myasthenia gravis (MG).
Find the cause of weakness, paralysis, or muscle twitching. Problems in a muscle, thenerves supplying a muscle, the spinal cord, or the area of the brain that controls amuscle can cause these symptoms. The EMG does not show brain or spinal cord
diseases.
Nerve conduction studies are done to:
Find damage to the peripheral nervous system, which includes all the nerves that leadaway from the brain and spinal cord and the smaller nerves that branch out from those
nerves. Nerve conduction studies are often used to help find nerve disorders, such as
carpal tunnel syndrome or Guillain-Barr syndrome.
Both EMG and nerve conduction studies can help diagnose a condition called post-polio
syndrome that may develop months to years after a person has had polio. (6)
http://www.webmd.com/hw-popup/duchenne-muscular-dystrophyhttp://www.webmd.com/hw-popup/amyotrophic-lateral-sclerosis-alshttp://www.webmd.com/hw-popup/peripheral-neuropathyhttp://www.webmd.com/hw-popup/electromyogramhttp://www.webmd.com/hw-popup/herniated-disc-7991http://www.webmd.com/hw-popup/myasthenia-gravishttp://www.webmd.com/hw-popup/nerve-conduction-studieshttp://www.webmd.com/hw-popup/carpal-tunnel-syndrome-8276http://www.webmd.com/hw-popup/guillain-barre-syndromehttp://www.webmd.com/hw-popup/guillain-barre-syndromehttp://www.webmd.com/hw-popup/guillain-barre-syndromehttp://www.webmd.com/hw-popup/post-polio-syndromehttp://www.webmd.com/hw-popup/post-polio-syndromehttp://www.webmd.com/hw-popup/poliohttp://www.webmd.com/hw-popup/poliohttp://www.webmd.com/hw-popup/post-polio-syndromehttp://www.webmd.com/hw-popup/post-polio-syndromehttp://www.webmd.com/hw-popup/guillain-barre-syndromehttp://www.webmd.com/hw-popup/carpal-tunnel-syndrome-8276http://www.webmd.com/hw-popup/nerve-conduction-studieshttp://www.webmd.com/hw-popup/myasthenia-gravishttp://www.webmd.com/hw-popup/herniated-disc-7991http://www.webmd.com/hw-popup/electromyogramhttp://www.webmd.com/hw-popup/peripheral-neuropathyhttp://www.webmd.com/hw-popup/amyotrophic-lateral-sclerosis-alshttp://www.webmd.com/hw-popup/duchenne-muscular-dystrophy -
8/6/2019 bitirmeee
20/37
11
2. CIRCUIT DESIGN OF AN EOG
2.1 INSTRUMENTATION AMPLIFIERAn instrumentation (or instrumentational) amplifier is a type of differential amplifier
that has been outfitted with input buffers, which eliminate the need for input impedance
matching and thus make the amplifier particularly suitable for use in measurement and test
equipment. Additional characteristics include very low DC offset, low drift, low noise, very
high open-loop gain, very high common-mode rejection ratio, and very high input
impedances. Instrumentation amplifiers are used where great accuracy and stability of the
circuit both short- and long-term are required.(7)
Although the instrumentation amplifier is usually shown schematically identical to a
standard op-amp, the electronic instrumentation amp is almost always internally composed of
3 op-amps. These are arranged so that there is one op-amp to buffer each input (+,), and one
to produce the desired output with adequate impedance matching for the function.[1][2]
The most commonly used instrumentation amplifier circuit is shown in the figure. The gain of
the circuit is
(Equ.1)
The rightmost amplifier, along with the resistors labelled R2 and R3 is just the standard
differential amplifier circuit, with gain = R3 / R2and differential input resistance = 2R2. The
two amplifiers on the left are the buffers. With Rgain removed (open circuited), they are simple
unity gain buffers; the circuit will work in that state, with gain simply equal to R 3 / R2 and
high input impedance because of the buffers. The buffer gain could be increased by putting
resistors between the buffer inverting inputs and ground to shunt away some of the negative
feedback; however, the single resistor Rgain between the two inverting inputs is a much more
elegant method: it increases the differential-mode gain of the buffer pair while leaving the
common-mode gain equal to 1. This increases the common-mode rejection ratio (CMRR) of
the circuit and also enables the buffers to handle much larger common-mode signals without
clipping than would be the case if they were separate and had the same gain. Another benefitof the method is that it boosts the gain using a single resistor rather than a pair, thus avoiding
http://www.answers.com/topic/differential-amplifierhttp://www.answers.com/topic/electronic-test-equipmenthttp://www.answers.com/topic/electronic-test-equipmenthttp://www.answers.com/topic/direct-currenthttp://www.answers.com/topic/drift-telecommunicationhttp://www.answers.com/topic/noise-level-1http://www.answers.com/topic/open-loop-gainhttp://www.answers.com/topic/common-mode-rejection-ratiohttp://www.answers.com/topic/input-impedancehttp://www.answers.com/topic/input-impedancehttp://www.answers.com/topic/accuracyhttp://www.answers.com/topic/bibo-stabilityhttp://www.answers.com/topic/electrical-networkhttp://www.answers.com/topic/instrumentation-amplifier#cite_note-0http://www.answers.com/topic/instrumentation-amplifier#cite_note-0http://www.answers.com/topic/instrumentation-amplifier#cite_note-0http://www.answers.com/topic/instrumentation-amplifier#cite_note-0http://www.answers.com/topic/instrumentation-amplifier#cite_note-0http://www.answers.com/topic/electrical-networkhttp://www.answers.com/topic/bibo-stabilityhttp://www.answers.com/topic/accuracyhttp://www.answers.com/topic/input-impedancehttp://www.answers.com/topic/input-impedancehttp://www.answers.com/topic/common-mode-rejection-ratiohttp://www.answers.com/topic/open-loop-gainhttp://www.answers.com/topic/noise-level-1http://www.answers.com/topic/drift-telecommunicationhttp://www.answers.com/topic/direct-currenthttp://www.answers.com/topic/electronic-test-equipmenthttp://www.answers.com/topic/electronic-test-equipmenthttp://www.answers.com/topic/differential-amplifier -
8/6/2019 bitirmeee
21/37
12
a resistor-matching problem (although the two R1s need to be matched), and very
conveniently allowing the gain of the circuit to be changed by changing the value of a single
resistor. A set of switch-selectable resistors or even a potentiometer can be used for R gain,
providing easy changes to the gain of the circuit, without the complexity of having to switch
matched pairs of resistors.
The ideal common-mode gain of an instrumentation amplifier is zero. In the circuit
shown, common-mode gain is caused by mismatches in the values of the equally-numbered
resistors and by the mis-match in common mode gains of the two input op-amps. Obtaining
very closely matched resistors is a significant difficulty in fabricating these circuits, as is
optimizing the common mode performance of the input op-amps.
An instrumentation amp can also be built with 2 op-amps to save on cost and increase
CMRR, but the gain must be higher than 2 (+6 dB).
Instrumentation amplifiers can be built with individual op-amps and precision resistors,
but are also available in integrated circuit form from several manufacturers (including Texas
Instruments, National Semiconductor, Analog Devices, Linear Technology and Maxim
Integrated Products). An IC instrumentation amplifier typically contains closely matched
laser-trimmed resistors, and therefore offers excellent common-mode rejection. Examples
include AD620, MAX4194, LT1167 and INA128.
Instrumentation Amplifiers can also be designed using "Indirect Current-feedback
Architecture", which extend the operating range of these amplifiers to the negative power
supply rail, and in some cases the positive power supply rail. This can be particularly useful in
single-supply systems, where the negative power rail is simply the circuit ground (GND).
Examples of parts utilizing this architecture are MAX4208/MAX4209 and AD8129/AD8130.
2.2 NON-INVERTING AMPLIFIER
The second basic configuration of an operational amplifier circuit is that of a Non-
inverting Amplifier. In this configuration, the input voltage signal, (Vin) is applied directly to
the non-inverting (+) input terminal which means that the output gain of the amplifier
becomes "Positive" in value in contrast to the "Inverting Amplifier" circuit we saw in the last
tutorial whose output gain is negative in value. The result of this is that the output signal is
"in-phase" with the input signal.(8)
Feedback control of the non-inverting amplifier is achieved by applying a small part of
the output voltage signal back to the inverting (-) input terminal via a Rf - R2 voltage divider
network, again producing negative feedback. This closed-loop configuration produces a non-
http://www.answers.com/topic/resistor-1http://www.answers.com/topic/integrated-circuithttp://www.answers.com/topic/texas-instrumentshttp://www.answers.com/topic/texas-instrumentshttp://www.answers.com/topic/national-semiconductorhttp://www.answers.com/topic/analog-devices-inchttp://www.answers.com/topic/linear-technology-corphttp://www.answers.com/topic/maxim-integrated-products-1http://www.answers.com/topic/maxim-integrated-products-1http://www.answers.com/topic/laser-trimminghttp://www.analog.com/en/amplifiers-and-comparators/instrumentation-amplifiers/ad620/products/product.htmlhttp://www.maxim-ic.com/quick_view2.cfm/qv_pk/2006http://www.linear.com/1167http://focus.ti.com/docs/prod/folders/print/ina128.htmlhttp://www.maxim-ic.com/quick_view2.cfm/qv_pk/4925http://www.analog.com/en/audiovideo-products/video-ampsbuffersfilters/ad8129/products/product.htmlhttp://www.analog.com/en/audiovideo-products/video-ampsbuffersfilters/ad8129/products/product.htmlhttp://www.maxim-ic.com/quick_view2.cfm/qv_pk/4925http://focus.ti.com/docs/prod/folders/print/ina128.htmlhttp://www.linear.com/1167http://www.maxim-ic.com/quick_view2.cfm/qv_pk/2006http://www.analog.com/en/amplifiers-and-comparators/instrumentation-amplifiers/ad620/products/product.htmlhttp://www.answers.com/topic/laser-trimminghttp://www.answers.com/topic/maxim-integrated-products-1http://www.answers.com/topic/maxim-integrated-products-1http://www.answers.com/topic/linear-technology-corphttp://www.answers.com/topic/analog-devices-inchttp://www.answers.com/topic/national-semiconductorhttp://www.answers.com/topic/texas-instrumentshttp://www.answers.com/topic/texas-instrumentshttp://www.answers.com/topic/integrated-circuithttp://www.answers.com/topic/resistor-1 -
8/6/2019 bitirmeee
22/37
13
inverting amplifier circuit with very good stability, a very high input impedance, Rin
approaching infinity, as no current flows into the positive input terminal, (ideal conditions)
and a low output impedance, Rout as shown below.
2.2.1 NON-INVERTING AMPLIFIER CONFIGURATION
FIGURE 4. Non-inverting Amplifier
In the previous Inverting Amplifier tutorial, we said that "no current flows into the input"
of the amplifier and that "V1 equals V2". This was because the junction of the input and
feedback signal (V1) are at the same potential in other words the junction is a "virtual earth"
summing point. Because of this virtual earth node the resistors, Rf and R2 form a simple
potential divider network across the non-inverting amplifier with the voltage gain of the
circuit being determined by the ratios of R2 and Rf as shown below. (8)
http://www.electronics-tutorials.ws/opamp/opamp_2.htmlhttp://www.electronics-tutorials.ws/opamp/opamp_2.html -
8/6/2019 bitirmeee
23/37
14
2.2.3 EQUIVALENT POTENTIAL DIVIDER NETWORK
FIGURE 5.Potential Divider Network
Then using the formula to calculate the output voltage of a potential divider network, we
can calculate the closed-loop voltage gain (Av) of the Non-inverting Amplifier as follows:
that the overall closed-loop gain of a non-inverting amplifier will always be greater but
never less than one (unity), it is positive in nature and is determined by the ratio of the values
of Rf and R2. If the value of the feedback resistor Rf is zero, the gain of the amplifier will be
exactly equal to one (unity). If resistor R2 is zero the gain will approach infinity, but in
practice it will be limited to the operational amplifiers open-loop differential gain, (Ao).
We can easily convert an inverting operational amplifier configuration into a non-inverting
amplifier configuration by simply changing the input connections as shown.
-
8/6/2019 bitirmeee
24/37
15
FIGURE 6 . Connection of non-inverting amplifier
2.2.4 NON-INVERTING AMPLIFIER CIRCUIT
The basic circuit for the non-inverting operational amplifier is relatively straightforward. In
this circuit the signal is applied to the non-inverting input of the op-amp. However the
feedback is taken from the output of the op-amp via a resistor to the inverting input of the
operational amplifier where another resistor is taken to ground. It is the value of these two
resistors that govern the gain of the operational amplifier circuit. (8)
FIGURE 7. Non-inverting Amplifier circuit
-
8/6/2019 bitirmeee
25/37
16
3. ANATOMY OF AN EYE
3.1 EYE STRUCTUREVision is one of our most valued senses and during the course of each day our eyes are
constantly moving. Attached to the globe of the eye, there are three antagonistic muscle pairs,
which relax and contract in order to induce eye movement. These pairs of muscles are
responsible for horizontal, vertical and torsional (clockwise and counter clockwise)
movement. (9)
3.2 EYE SIGNALSTwo specific categories exist which can be used to classify the four different types of
conjugate eye movements: 1. Reflex eye movements - These provide stabilization of eye
position in space during head
movement. 2. Voluntary eye movements - These are conscious eye movements involved in
the redirection of
the line of sight in order to pursue a moving target (pursuit movement) or to focus on a new
target of interest. (9)
3.3 OPTIC AND CILIARY MUSCELS
FIGURE 8 . Ekstraocular muscles
-
8/6/2019 bitirmeee
26/37
17
FIGURE 9 Motor Nerve
FIGURE 10 ObliqueMuscels
-
8/6/2019 bitirmeee
27/37
18
FIGURE 11 Oblique Muscles
FIGURE 12 Orbita Fasya System
-
8/6/2019 bitirmeee
28/37
19
FIGURE 13 Orbita Fasya System
FIGURE 14 Orbita Fasya System
-
8/6/2019 bitirmeee
29/37
20
FIGURE 15 . Eye Cover
FIGURE 16 Orbitalis Oculi
-
8/6/2019 bitirmeee
30/37
21
FIGURE 17 Orbital septum
FIGURE 18 Lid Retractors (10)
-
8/6/2019 bitirmeee
31/37
22
FIGURE 19 Lid Retractor
FIGURE 20 Tarsal Disc and Tarsal Sempatic Muscels
-
8/6/2019 bitirmeee
32/37
-
8/6/2019 bitirmeee
33/37
24
4. EXPERIMENTAL RESULTS
FIGURE 21. Example 1
FIGURE 22. Example 2
FIGURE 23. Example 3
-
8/6/2019 bitirmeee
34/37
25
FIGURE 24. data 1
FIGURE 25. data 2
FIGURE 26. data 3
V
ol
t
Sampling Period : 10ms
V
o
l
t
Sampling Period : 10ms
A
a
-
8/6/2019 bitirmeee
35/37
26
FIGURE 27. simulation with proteus
FIGURE 28. Example of proteus
-
8/6/2019 bitirmeee
36/37
27
5.CONCLUSION
In this Project a system is designed to detect blink signal analysis. Main purposes of this
project is identify the signal from various people with using EOG. Many factors have been
determined to affect blink rate. Dry eye patients have been reported to have in- creased blink
rate. Contact lenses disrupts the tear film, resulting in an increased blink rate. Fatigue,
anxiety, and mental activities are known correlate to blink rate. Blink rate has even been
reported as an index for visual efficiency. Research findings indicate there are physiologic
and psychological/perceptual factors as- sociated with blinking. The fact that blindness and/or
binocular enucleation did not reduce the blink rate to zero suggests there may be
undetermined factors that cause endogenous blinking. It has been suggested that a blink
clock is located in the brainstem. This may be the reason for the effect of mental activities
on blink rates. The mechanism involved in blinking is, however, not fully understood.
-
8/6/2019 bitirmeee
37/37