Biomicroscopy Tech

download Biomicroscopy Tech

of 66

Transcript of Biomicroscopy Tech

  • 7/27/2019 Biomicroscopy Tech

    1/66

    SLIT-LAMP BIOMICROSCOPY

    The slit-lamp biomicroscope is an

    important diagnostic instrument for

    the examination of the anterior eye

    structures as well as the retina.

    Being able to proficiently use the

    instrument is important. It use

    critical in the fitting of contact

    lenses.

  • 7/27/2019 Biomicroscopy Tech

    2/66

    THE SLIT-LAMP BIOMICROSCOPE

    The instrument is composed of two main components. The slit-lamp is the

    light source and the controls to modify the size, shape, direction, color and

    intensity of the light.

    The biomicroscope is the microscope used to view the eye. The

    magnification can be varied.

    The accurate term for the instrument is the slit-lamp biomicroscope,however, it is usually called the slit-lamp or the biomicroscope.

    Light source

    microscope

  • 7/27/2019 Biomicroscopy Tech

    3/66

    THE SLIT-LAMP BIOMICROSCOPE

    Each brand of biomicroscope is slightly different. They have a switch to

    turn the light source on. The intensity of the light source can be varied.

    The control is often the same dial that turns the light source on or near itdepending on the make of the instrument.

  • 7/27/2019 Biomicroscopy Tech

    4/66

    THE SLIT-LAMP BIOMICROSCOPE

    The slit-lamp biomicroscope can be moved in relation to the

    patient using the joy stick for focusing and bringingstructures into view. Other parts of the instrument are the

    head and chin rest to hold the patient still and the adjustable

    table to hold and adjust the height the instrument for doctor

    and patient comfort.

    Joy stick to

    position instrument

    To adjust the

    Vertical height

    of the instrument

    To adjust the heightof the chinrest

  • 7/27/2019 Biomicroscopy Tech

    5/66

    THE SLIT-LAMP BIOMICROSCOPE

    The instrument is designed such that when you use the joystick to move

    the instrument towards and away from the patient to focus structureswheel moves on a track. When the instrument moves side to side it

    moves along a rod. It can be locked so that one can not move it with the

    joystick by tightening down a knob on the rod. It should be locked when

    not in use or when moving the instrument.

  • 7/27/2019 Biomicroscopy Tech

    6/66

    THE SLIT-LAMP BIOMICROSCOPE

    The angle between the light source and microscope can be changed by just

    moving the arms that hold the light and microscope. They can be locked in

    place if desired, however, during the examination the angle between the two

    are being changed to obtain the proper views.

    Light source and microscope

    arms that can be rotated

    independently.

    Knobs to lock arms in place,

    usually used when

    biomicroscope in not in use

    or being moved.

  • 7/27/2019 Biomicroscopy Tech

    7/66

    THE SLIT-LAMP BIOMICROSCOPE

    Before examining a patient become familiar with the controls.

    The width of the light beam can be adjusted by a control on the slit-lamp.

    The light beam can also be moved back and forth horizontally. In addition the

    light beam can be rotated away from vertical.

    Red arrows-slit width control

    Green arrows-slit height and filter control

  • 7/27/2019 Biomicroscopy Tech

    8/66

    THE SLIT-LAMP BIOMICROSCOPE

    Different instruments have different controls to

    change the magnification. In some cases they alsohave a second set of eyepieces that can be used to

    change the magnifications.

  • 7/27/2019 Biomicroscopy Tech

    9/66

    THE SLIT-LAMP BIOMICROSCOPE

    The light beam and the microscope both have to be focused at

    the same point. A focusing rod is put in place. The light source,

    a narrow beam, is positioned on the rod. This should be at the

    click position of the light source. When looking through the

    microscope the light source should be in the center of your view

    and in focus.

  • 7/27/2019 Biomicroscopy Tech

    10/66

    THE SLIT-LAMP BIOMICROSCOPE

    You must focus the eyepieces for your eyes. With a narrow

    beam of light on the focusing rod, look through thebiomicroscope. Looking through only one eyepiece at time,

    rotate the eyepiece counter-clockwise until the light beam

    appears blurred, then rotate it clockwise until the light beam

    just comes into focus.

  • 7/27/2019 Biomicroscopy Tech

    11/66

    PATIENT POSITIONING

    Clean the headrest and chin rest with an alcohol swab

    before putting the patient behind the instrument.

    Some units have paper sheets on the chinrest where a

    sheet can be removed between patients.

    Before starting the exam the patient must be

    comfortably seated behind the instrument with their

    chin on the chin rest and head tight against the head

    rest. The height of the instrument and chin rest

    should be adjusted so the patients eyes are aligned

    with the alignment marks on the head rest.

    Alignment

    mark

    Incorrect, the

    forehead is not

    against theheadrest.

  • 7/27/2019 Biomicroscopy Tech

    12/66

    PATIENT FIXATION

    You need to control the patients fixation. The instruments have a

    fixation light that can be placed in front of the patient andpositioned so the patient is looking in the desired position. This

    may cause the patient to accommodate and converge on the

    fixation light causing some discomfort.

    Often an easier and quicker way is to have the patient just look at

    your ear or a point on the biomicroscope. Then you can just ask

    the patient to look down, up, left or right as needed.

  • 7/27/2019 Biomicroscopy Tech

    13/66

    ROUNTINE FOR EXAMINING THE ANTERIOR SEGMENT

    In order to be sure you do not miss any anomaly you must getinto the habit of using a set routine. A common order of

    examination is:

    Lids and lashesConjunctiva and cornea

    Instillation of fluorescein and BUT measurement

    Eversion of the lids

    Anterior chamber and angle measurement

    Iris

    Crystalline lens

    Anterior vitreous

  • 7/27/2019 Biomicroscopy Tech

    14/66

    DIRECT AND DIFFUSE ILLUMINATION

    The lids and anterior ocular structures are examined

    with the slit lamp beam wide open. This called directillumination. To obtain an even wider view a diffuser

    can be placed in front of the light source to give diffuse

    illumination.

  • 7/27/2019 Biomicroscopy Tech

    15/66

    LID EXAMINATION

    The skin of the lids should be examined for any lumps or bumps

    or skin lesions. It is best to start by having the patient close theireyes and scan the lids starting with the upper lid from temporal to

    nasal followed by the lower lid. It is important to get in a routine

    sequence of examination so you do not miss anything.

    Low (7X to 10X) to moderate (10X to 16X) magnification is used

    when examining the lids with higher magnifications only used to

    study an abnormality. It too high magnification is used you will

    miss some abnormalities.

  • 7/27/2019 Biomicroscopy Tech

    16/66

    EXAMINATION OF THE LASHES

    The lashes should be examinedfor abnormalities such blepharitis.

    With blepharitis there will be

    collarettes found at the base of

    the lashes.

    This photo shows nits (egg

    sacs) from lice infestation of

    the lashes.

  • 7/27/2019 Biomicroscopy Tech

    17/66

    EXAMINATION OF THE LASHES

    The lashes can also be missing. This is called

    madaroisis.

  • 7/27/2019 Biomicroscopy Tech

    18/66

    LID MARGINS

    After examining the closed lids have the patient open their eyes and

    examine the lid margins. You should see the normal structures suchas the meibomian gland openings on both the upper and lower lids.

    The puncta (opening for the tears to exit the eye) can be seen at the

    nasal side of the lid margin on the upper and lower lids.

    Meibomian gland openings puncta

  • 7/27/2019 Biomicroscopy Tech

    19/66

    LID MARGINS

    The lid margins should be examined for injection, swelling,

    lesions, plugged meibomian glands, chalazion or otheranomalies.

    Plugged gland

  • 7/27/2019 Biomicroscopy Tech

    20/66

    EYE MAKE-UP

    It is common to see make-up along the lid margins. It can also

    be seen in the tear film and may become embedded in the

    palpebral conjunctiva. Make-up can create problems with

    contact lens wear.

  • 7/27/2019 Biomicroscopy Tech

    21/66

    EXAMINATION OF THE BULBAR CONJUNCTIVA

    The conjunctiva visible within the palpebral opening without

    pulling the lids back can next be inspected with diffuse

    illumination. One should look for injection, dry spots, cysts,pigmentation, pingueculum, pterygium or other anomalies.

    Any type of irritation, trauma or infection can cause injection of the

    conjunctiva.

    Injected conjunctivaNormal conjunctiva

  • 7/27/2019 Biomicroscopy Tech

    22/66

    EXAMINATION OF THE BULBAR CONJUNCTIVA

    This photo shows a nevus onthe conjunctiva.

    This photo shows a pinquecula,

    a raised area on the

    conjunctiva that is fairly

    common.

    EXAMINATION OF THE SUPERIOR AND INFERIOR

  • 7/27/2019 Biomicroscopy Tech

    23/66

    EXAMINATION OF THE SUPERIOR AND INFERIOR

    BULBAR CONJUNCTIVA

    The superior bulbar conjunctiva isexamined by having the patient

    look down and gently pulling the

    upper lid up.

    The inferior bulbar conjunctiva

    can likewise be examined by

    having the patient look up and

    pulling the lower lid down.

  • 7/27/2019 Biomicroscopy Tech

    24/66

    EXAMINATION OF THE LIMBUS

    It is important to examine the limbus, the junction between the

    cornea and conjunctiva. A number of normal structures may bevisible including blood vessel loops, nerves, palisades of Vogt and

    pigmentation.

    A green filter can be used to

    better visualize the blood vessels

  • 7/27/2019 Biomicroscopy Tech

    25/66

    EXAMINATION OF THE LIMBUS

    This photo shows the Palisades of Vogt which are often visible

    with darkly pigmented patients. The palisades are radial

    extensions of the sclera towards the cornea at the limbus with

    epithelial cells 10 to 12 layers thick between them.

  • 7/27/2019 Biomicroscopy Tech

    26/66

    EXAMINATION OF THE LIMBUS

    This photo is of posterior

    embryotoxon, an developmental

    anomaly where Schwalbes ring

    (collagenous tissue between the

    end of Descemets membrane

    and the trabecular meshwork)

    extends further into the cornea

    than normal

    EXAMINATION OF THE CORNEA WITH

  • 7/27/2019 Biomicroscopy Tech

    27/66

    EXAMINATION OF THE CORNEA WITH

    DIRECT OR DIFFUSE ILLUMINATION

    The cornea can be scanned with a wide beam direct or diffuse illumination to

    detect anomalies.

    Corneal scars, neovascularization, infiltrates, abrasions, foreign bodies and

    other conditions can be seen.

    Anomalies should not be confused with the reflection of the light source from

    the surface of the cornea or from the surface of the crystalline lens.

    reflection fromsurface of the cornea

    reflection from surface of

    the crystalline lens

    Corneal scar

    EXAMINATION OF THE CORNEA WITH

  • 7/27/2019 Biomicroscopy Tech

    28/66

    EXAMINATION OF THE CORNEA WITH

    DIRECT OR DIFFUSE ILLUMINATION

    Arcus senilis is a build-up of cholesterol and other lipids in the

    peripheral cornea. It appears as a whitish or yellowish ring just

    inside the limbus. It is found in 60% of patients between 40 and 60

    years old and almost all patients over 80.

  • 7/27/2019 Biomicroscopy Tech

    29/66

    EXAMINATION OF THE TEAR FILM

    Diffuse or a wide beam direct illumination can also be used to

    inspect the tear film. Debris can be seen floating in the tear filmand the debris can increase in dry eye conditions, irritation and

    infection.

  • 7/27/2019 Biomicroscopy Tech

    30/66

    EXAMINATION OF THE TEAR FILM

    Bubbles, also called frothing, can sometimes be seen in the tear

    film, especially along the lid margins. This can indicatemeibomianitis.

  • 7/27/2019 Biomicroscopy Tech

    31/66

    EXAMINATION OF THE TEAR FILM

    Colored interference fringes may also beseen on the tear film over the cornea.

    These are visible when the oily layer of the

    tear film is thicker than normal.

    These are most often seen just above thelower lid margin. By pressing on the

    lower lid and forcing the oily material from

    the meibomian glands, this will enhance

    the fringes.

    INSTILLATION OF FLUORESCEIN

  • 7/27/2019 Biomicroscopy Tech

    32/66

    INSTILLATION OF FLUORESCEIN

    Following a general overview of the visible portion of the anterior

    segment and prior to significant manipulation of the lids, fluorescein

    should be instilled to measure the break-up time of the tear film(B.U.T.), look for corneal staining, conjunctival staining and papillae.

    The method of instilling fluorescein is important. The fluorescein

    strip should be moistened with saline solution. Do not touch the tip

    of the bottle to the strip. After putting a drop on the strip shake the

    excess off prior to instilling it on the eye. You do not want to add

    excessive solution t the eye, just the stain.

    Correct

    Incorrect-do

    not touch strip to

    bottle

  • 7/27/2019 Biomicroscopy Tech

    33/66

    INSTILLATION OF FLUORESCEIN

    To instill fluorescein, have the patient look down, rest your wrist on

    the patients forehead and grasp the upper lid with the index finger

    over the lashes. The patient should not be able to blink. With thepatient still looking down, pull the lid up and gently lay the flat side

    of the strip against the conjunctiva. Do not rub the fluorescein strip

    across the eye. Gently lift it off and release the lid.

    Do not use a dry fluorescein strip or dab the fluorescein strip

    against the lower lid as this unnecessarily irritates the patient and

    causes reflex tearing.

  • 7/27/2019 Biomicroscopy Tech

    34/66

    INSTILLATION OF FLUORESCEIN

    Another method of instilling fluorescein is to have the patient look

    up and while holding the lids instill the fluorescein on the inferior

    conjunctiva.

    The fluorescein strip will cause some staining of the conjunctival

    which is normal. You should not touch the strip to the cornea as

    this will cause the patient discomfort and staining of the cornea.

  • 7/27/2019 Biomicroscopy Tech

    35/66

    FLUORESCEIN STAINING

    Fluorescein is a water soluble stain that will stain the tear film and

    any disruption of the ocular surface. It will show any breaks in theepithelium and be absorbed by cells which have damaged

    membranes.

    FLUORESCEIN STAINING

  • 7/27/2019 Biomicroscopy Tech

    36/66

    FLUORESCEIN STAINING

    Fluorescein staining can be

    highlighted by the use of filters.

    The biomicroscope has a cobalt(blue) filter that can be placed in

    the light beam. The blue filter

    enhances the visibility of the

    stain.

    The visibility of the staining can

    be further enhanced by placing a

    yellow filter (ie. a Wratten #12 or

    Tiffin #2) over the objective of the

    biomicroscope.

    The top photo shows a filter being

    placed over the objective. The

    bottom photo shows some

    different filters that can be used.

    FLUORESCEIN STAINING

  • 7/27/2019 Biomicroscopy Tech

    37/66

    FLUORESCEIN STAINING

    The yellow filter filters out the

    blue light to give better

    visualization of the greenfluorescence of the fluorescein.

    The top photo is with the yellow

    filter in place.

  • 7/27/2019 Biomicroscopy Tech

    38/66

    MEASURING THE TEAR FILM BREAK-UP TIME

    The time it takes for the tear film to break-up following a blink is an

    indication of the quality of the tear film. Immediately following theinstillation of fluorescein the B.U.T. should be measured. The room

    lights should be out and filters in place. The patient is asked to blink

    3 or 4 times and then to look straight ahead and not blink. The time

    in seconds from the last blink until a dark spot or streak develops in

    the tear film is the B.U.T. Do not hold the lids open.

    TEAR FILM

  • 7/27/2019 Biomicroscopy Tech

    39/66

    TEAR FILM

    In addition to the light green staining of the tear film seen over the

    corneal surface, the tears in the tear meniscus (also termed tearprism, lacrimal lake, or tear strip) along the lid margins is easily

    seen. With dry eyes this tear meniscus is decreased in size and

    may be essentially absent in places.

    CORNEAL STAINING

  • 7/27/2019 Biomicroscopy Tech

    40/66

    CORNEAL STAINING

    If only the superficial epithelial cells are damaged, the

    fluorescein staining will be limited to the area of damage. If the

    full thickness of the epithelium is damaged, fluorescein will seepinto the stroma giving a larger area of staining. The borders will

    not be as distinct since the fluorescein is slowly diffusing

    through the stroma.

    CORNEAL STAINING

  • 7/27/2019 Biomicroscopy Tech

    41/66

    CORNEAL STAINING

    Rubbing the cornea through the eyelid can cause an epithelial

    mosaic staining pattern (Schweitzers Lines). This lasts only afew minutes.

    CONJUNCTIVAL STAINING

  • 7/27/2019 Biomicroscopy Tech

    42/66

    CONJUNCTIVAL STAINING

    Using the wide beam or diffuse illumination and the filters, the

    conjunctiva should be examined for staining. It is normal to haveareas of pooling of fluorescein, especially in older patients where

    the conjunctiva is looser and more wrinkled.

    Fluorescein pooling in

    folds of the conjunctiva Spots of staining from drying

    ROSE BENGAL STAINING

  • 7/27/2019 Biomicroscopy Tech

    43/66

    ROSE BENGAL STAINING

    Rose bengal stain is used to stain dead or degenerating cells as

    well as mucous. It is used to show staining with dry eye patients.

    It is supplied on strips like fluorescein or as a solution. Only asmall amount of stain is required. If too much is used it will be

    uncomfortable. One must be careful as it will stain clothing and is

    not as easily washed out as fluorescein. Another stain which

    stains similar cells as rose bengal but is more comfortable is

    lassime green.

    PALPEBRAL CONJUNCTIVA EXAMINATION

  • 7/27/2019 Biomicroscopy Tech

    44/66

    PALPEBRAL CONJUNCTIVA EXAMINATION

    It is important to evert the upper lid and examine the palpebral

    conjunctiva for papillae, cysts, concretions, foreign bodies,irritation and other anomalies. With the patient in the head rest,

    have the patient look down and grasp the upper lid margin over

    the lashes. With a cotton-tipped applicator or other stylus placed

    at the upper lid sulcus (the fold in the upper lid) and lift the lid up

    and evert it over the stylus.

    HOLDING THE EVERTED LID FOR EXAMINATION

  • 7/27/2019 Biomicroscopy Tech

    45/66

    HOLDING THE EVERTED LID FOR EXAMINATION

    Hold the lid in the everted

    position with your thumb and

    examine the palpebral

    conjunctiva with the

    biomicroscope.

    PALPEBRAL CONJUNCTIVA EXAMINATION

  • 7/27/2019 Biomicroscopy Tech

    46/66

    PALPEBRAL CONJUNCTIVA EXAMINATION

    The upper conjunctiva should be

    smooth and not injected. A common

    problem with contact lens wear is thedevelopment of papillae (raised areas

    of inflammatory tissue). Each

    papillae has a central blood vessel

    stalk. With fluorescein it is easier to

    detect the papillae.

    PALPEBRAL CONJUNCTIVA EXAMINATION

  • 7/27/2019 Biomicroscopy Tech

    47/66

    PALPEBRAL CONJUNCTIVA EXAMINATION

    The lower conjunctiva should also be examined. Just pull down

    the lid. Look for cysts, follicles, embedded make-up, concretions

    or other anomalies.

    concretionEmbedded make-up

    CORNEAL EXAMINATION-SCLEROTIC SCATTER

  • 7/27/2019 Biomicroscopy Tech

    48/66

    CORNEAL EXAMINATION SCLEROTIC SCATTER

    An illumination called sclerotic

    scatter is useful in detectingcentral corneal clouding (corneal

    edema) created with rigid contact

    lenses and other corneal opacities

    such as scars, foreign bodies and

    abrasions.

    A light beam about 2 to 4 mm wide

    is focused on the temporal limbus.

    The light source should be rotated

    such that it is off to the side of the

    viewing area (the microscopefocused on the center of the

    cornea and the light source on the

    limbus). When properly set up the

    limbus 180 degrees from the light

    beam will glow.

    PARALLELEPIPED

  • 7/27/2019 Biomicroscopy Tech

    49/66

    PARALLELEPIPED

    One of the most useful illuminations is the parallelepiped. This is a

    relatively wide beam of light that illuminates a block of the cornea.

    The microscope is directly in front of the patients cornea and the

    light source is approximately 45 degrees from the straight ahead

    position. Low to moderate magnification can be used to scan the

    cornea with high magnification used to examine structures in

    detail.

    Width of beam

    (front surface of cornea)

    Back surface of the

    cornea

    Light reflected

    off of the iris

    CORNEAL EXAMINATION WITH THE PARALLELEPIPED

  • 7/27/2019 Biomicroscopy Tech

    50/66

    CORNEAL EXAMINATION WITH THE PARALLELEPIPED

    The corneal should be scanned with a parallelepiped as it is the

    illumination that is used to detect and examine corneal structures

    and defects.

    Corneal nerve seen with

    a wide parallepiped.

    Pigment on the back of

    the cornea seen with a

    more narrow beam.

    A corneal scar seen with

    a relatively wide beam

    CORNEAL EXAMINATION WITH THE PARALLELEPIPED

  • 7/27/2019 Biomicroscopy Tech

    51/66

    CORNEAL EXAMINATION WITH THE PARALLELEPIPED

    The parallelepiped can be used to detect corneal striae that

    develop when corneal edema occurs with hydrogel lens wear, in

    keratoconus and normally in some older patients.

    They are seen as small, usually vertical, white lines deep in the

    corneal stroma. They are best seen with high magnification. With

    greater amounts of edema folds develop in the same area which

    appear as dark lines.

    striae

    fold

    CORNEAL EXAMINATION WITH THE OPTIC SECTION

  • 7/27/2019 Biomicroscopy Tech

    52/66

    CORNEAL EXAMINATION WITH THE OPTIC SECTION

    When scars, abrasions or other

    anomalies are detected one often

    wants to know how deep they go intothe cornea. This is when an optic

    section is used. The optic section is

    a very thin parallelepiped and

    optically cuts a very thin slice of the

    cornea. The distance from the front

    and back of the section gives you the

    location within the cornea.

    The epithelium (front surface of the

    cornea) gives a bright reflection. With

    high magnification a speckle patter isseen within the normal stroma. The

    back of the optic section is the

    endothelium.

    Front of cornea Back of cornea

    CORNEAL EXAMINATION WITH THE OPTIC SECTION

  • 7/27/2019 Biomicroscopy Tech

    53/66

    CORNEAL EXAMINATION WITH THE OPTIC SECTION

    The top photo shows a corneal scar

    with a wide beam of illumination.

    The bottom photo shows an opticalsection through the scar indicating

    the depth within the cornea-in this

    case very superficial.

    MEASURING THE SIZE OF DETAILS

  • 7/27/2019 Biomicroscopy Tech

    54/66

    MEASURING THE SIZE OF DETAILS

    It is important in following and

    documenting conditions to accurately

    measure them. The width or height of a

    structure can be measured by using a spotof known diameter or a parallelepiped of

    known width. Biomicroscope have a scale

    on the light source that indicates the width

    of the beam. They also have apertures of

    different diameters that can be used.

    It is also possible to have a reticule in the

    eyepiece to make the measurement,

    however, most biomicroscopes do not have

    such a reticule.

    MEASURING THE MOVMENT OF A CONTACT LENS

  • 7/27/2019 Biomicroscopy Tech

    55/66

    MEASURING THE MOVMENT OF A CONTACT LENS

    The amount of movement of a lens can be estimated by using a

    spot of light of known size. For example with a soft lens, a spotsize of 0.2 mm to 1 mm can be used. The amount of movement of

    the lens edge with a blink in relation to the spot size can be

    determined.

    RETRO AND INDIRECT ILLUMINATION

  • 7/27/2019 Biomicroscopy Tech

    56/66

    RETRO AND INDIRECT ILLUMINATION

    Many corneal defects are not easily seen

    with direct illumination. Such things as

    microcystic edema, microcysts andneovascularization are examples.

    Retro-illumination is where a relatively

    narrow beam (1-3 mm) is reflected off

    the iris and the object of interest isviewed against the bright iris. If the

    object is opaque it will appear dark.

    Indirect is similar except you view the

    area of the cornea next to the light beamhits the cornea and not against the

    reflection from the iris. In practice both

    are used more or less similtaneously.

    RETRO AND INDIRECT ILLUMINATION

  • 7/27/2019 Biomicroscopy Tech

    57/66

    These photos show an foreign

    body in direct illumination (top)

    and indirect illumination (bottom)

    RETRO AND INDIRECT ILLUMINATION

  • 7/27/2019 Biomicroscopy Tech

    58/66

    This photo shows microcystic

    edema and vacuoles from

    wearing a thick hydrogel contact

    lens in retro illumination.

    RETRO-ILLUMINATION

  • 7/27/2019 Biomicroscopy Tech

    59/66

    Neovascularization (new vessel growth) near the limbus is

    often difficult to visualize. Retro-illumination makes thevessels easier to see.

    RETRO ILLUMINATION

    RETRO-ILLUMINATION

  • 7/27/2019 Biomicroscopy Tech

    60/66

    This photo shows epithelialmicrocysts (small dots) in the

    corneal epithelium from the

    wearing of an extended wear soft

    contact lens. It is often easiest to

    detect such small defects alongthe edge of the retro-illumination,

    this is sometimes termed

    marginal retro-illumination.

    RETRO ILLUMINATION

    RETRO-ILLUMINATION

  • 7/27/2019 Biomicroscopy Tech

    61/66

    When the pupil is dilated, retro-illumination, can give a red reflex

    against which corneal defects can be seen. The beam is passed

    through the periphery of the pupil to obtain this view. This photoshows an area of localized edema.

    RETRO ILLUMINATION

    SPECULAR REFLECTION

  • 7/27/2019 Biomicroscopy Tech

    62/66

    SPECULAR REFLECTION

    Specular reflection is when the angle

    between the light source and microscope

    is such that a specular reflection isobtained from the cornea.

    A bright reflection is obtained from the

    surface of the cornea and a less bright

    reflection form the posterior surface ofthe cornea. This illumination is obtained

    by using a narrow parallelepiped and

    placing the microscope directly in front of

    the patients eye with the light source

    approximately 45 degrees to the side.

    The angle of the light should be moved

    until a very bright reflection is obtained

    from the corneal surface.

    SPECULAR REFLECTION

  • 7/27/2019 Biomicroscopy Tech

    63/66

    SPECULAR REFLECTION

    Specular reflection is mainly used to

    view the endothelium, the single layer

    of cells on the back surface of thecornea. The highest magnification

    available should be used. To view this

    layer look just to the side of the bright

    reflection from the font surface of the

    cornea. If the parllelepiped is too wide

    you will not be able to see the

    endothelium. To view different areas

    of the cornea have the patient change

    their gaze as well as you changing the

    angle of the microscope and light

    source.

    The endothelium will appear as a fine,

    rough surface due to the individual

    endothelial cells.

    endothelium

    reflection from the

    front surface

    SPECULAR REFLECTION

  • 7/27/2019 Biomicroscopy Tech

    64/66

    SPECULAR REFLECTION

    Endothelium guttata, polymegethism (variation in size of the

    cells) and dystrophies may be detected and studied with

    specular reflection.

    guttata

    USING THE BIOMICROSCOPE TO MEASURE ANGLES

  • 7/27/2019 Biomicroscopy Tech

    65/66

    The biomicroscope can be used to measure rotation, for example

    the rotation of soft toric contact lenses.

    To measure the rotation of a lens, locate the mark indicating thebase of the prism or horizontal marks on the lens. Using a narrow

    parallelepiped, adjust the height of the instrument such that the

    center of the beam is approximately in the center of the pupil.

    Rotate the beam until it goes through the mark on the lens and the

    center of the pupil. The amount of rotation can be read from thescale on the slit-lamp.

    lens with mark at base

    rotating nasallylens with markings in

    horizontal plane rotating

    scale on the slit-lamp

    showing degrees of

    rotation

    BIOMCROSCOPY TECHNIQUE

  • 7/27/2019 Biomicroscopy Tech

    66/66

    BIOMCROSCOPY TECHNIQUE

    You should practice using the biomicroscope until you are

    proficient at using the controls and can do an examination of theanterior surface of the eye in an efficient and smooth manner.