A Drug Multiplexing infusion system
Biomedical Sensor- and Devices Laboratory Lübeck, Germany
Saif Abdul-Karim, M.Sc Prof. Dr. Bodo Nestler
msgt.fh-luebeck.de
![Page 2: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/2.jpg)
Introduction Aim Flow Setup Optical Setup Model design Conclusions
Medical errors are a leading
cause of death in hospitals.
About 210.000 are killed by
preventable hospital errors
each year in the U.S [1].
Between years 2005 and 2009,
more than 56.000 infusion
pump incidents were reported,
including 710 deaths [3]
![Page 3: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/3.jpg)
Infusion problems
• Hygiene
• Drugs incompatibility
• Patient transportation
• Chaos and Complexity
• Delivery errors
• Human errors
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 4: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/4.jpg)
Delivery errors*
* A. Timmermann, University Medical Centre, Utrecht, Netherlands.
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 5: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/5.jpg)
Examination of a multi infusion system
Introduction Aim Flow Setup Optical Setup Model design Conclusions
7mL/h; C = 0.13 g/L
5mL/h; C = 0.06 g/L
![Page 6: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/6.jpg)
Better delivery
Multiplex system
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 7: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/7.jpg)
Questions / tasks
Separation by means of gas bubbles (CO2 or air)
Cleaning the inner surface
size of the bubbles as a function of the time t or the position x along the tube
Accuracy of the volume of the separated units
Reproducibility of a liquid volume
Optimized size of the separated units
Compatibility of different drugs
Contamination of the following volumes behind a separation bubble
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 8: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/8.jpg)
Drug multiplexing test bench - flow
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 9: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/9.jpg)
boundary conditions 1
Volume of a 1mm gas bubble V = 800 nL
Radius of the tube: R = 0.5 mm
Volume of the tube: V = 1.2 mL
Fill time by flow min: t = 144 s
Fill time by flow max: t = 1.2 s
Flow range: 0.5 ml/min - 60 mL/min (= 3.6 L/h = 86 L/d)
Max pressure: 1000 hPa
Length of the tube: 1.5 m
Max. Gas CO2 dosage: 300 ml *
* Department of Radiology, University Hospitals Gasthuisberg, Leuven, Belgium.
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 10: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/10.jpg)
Result: The length of the bubble will be reduced to 30% along the tube
Dissolving CO2 in water
And Boyle Mariotte: pV = const. (T=const.)
during the flow
Introduction Aim Flow Setup Optical Setup Model design Conclusions
Bubble size as a function of the position
![Page 11: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/11.jpg)
boundary conditions 2
Flow range: 0.05 ml/min - 6 mL/min ( = 8.6 L/d)
Max pressure: 1000 hPa
Length of the tube: 1.5 m
Max. Gas CO2 dosage: 300 ml *
Volume of a 1mm long gas bubble VB = 200 nL
Radius of the tube: R = 0.25 mm
Volume of the tube: V = 0.3 mL
Fill time by flow min: t = 360 s
Fill time by flow max: t = 3 s
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 12: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/12.jpg)
Quantitative and qualitative measuring equipments
UV-VIS Spectroscopy
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 13: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/13.jpg)
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 14: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/14.jpg)
14
Model of Drug Multiplexing
• Compact module
• Patient’s side
• Mobile
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 15: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/15.jpg)
Drug Multiplexing
Hygiene
Drugs incompatibility reactions
Patient transportation
Chaos and Complexity
Delivery errors
Human errors
Better flow and concentration control
Pulsated dosing
(Restricted Performance with filter)
Separation by gas bubbles
Introduction Aim Flow Setup Optical Setup Model design Conclusions
![Page 16: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/16.jpg)
Thank you for your attention
![Page 17: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/17.jpg)
![Page 18: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/18.jpg)
18
Typical flow rates at clinics?
Mimics clinic conditions
Reliable
A mobile delivery station helps to reduce
preventable infusion errors.
![Page 19: Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif … · 2016-09-12 · Biomedical Sensor- and Devices Laboratory Lübeck, Germany Saif Abdul-Karim, M.Sc Prof. Dr. Bodo](https://reader030.fdocuments.us/reader030/viewer/2022041023/5ed4223ef6cab513ec15a57c/html5/thumbnails/19.jpg)
Flow
PVC Tube
Length: 1.5 m
Inner Ø: 1 mm
Volume: 11.786 mL (for 0.6 mm inner Ø: 4.243 mL)